220222 JR Confocal Laser Scanning Microscopy Confocal fluorescence image (actin, nucleus, mitochondria) of BPAE cell J. Rheinlaender Instruction Manual Institut für Angewandte Physik AG Prof. Tilman Schäffer Universität Tübingen Assistant room number: C9 A38 / C7 P31 Experiment room number: C9 P13 / H36 Table of Contents I. Preparation ........................................................................................................................ 3 II. Background ........................................................................................................................ 3 1. History of optical microscopy ......................................................................................... 3 2. Basics of wave optics ...................................................................................................... 4 2.1. Interference ............................................................................................................. 4 2.2. Refraction ................................................................................................................ 4 2.3. Diffraction ................................................................................................................ 5 2.4. The Airy disk............................................................................................................. 6 3. Experimental setup ......................................................................................................... 7 3.1. Fluorescence microscopy ........................................................................................ 7 3.2. Confocal microscopy................................................................................................ 7 3.3. Immersion oil objectives ......................................................................................... 9 4. Resolution in optical microscopy .................................................................................... 9 4.1. The point spread function (PSF) .............................................................................. 9 4.2. The Rayleigh criterion .............................................................................................. 9 4.3. Resolution in confocal microscopy ........................................................................ 10 4.4. Nyquist–Shannon sampling theorem ................................................................. 11 III. Experiment ....................................................................................................................... 12 1. Laboratory ..................................................................................................................... 12 2. Protocol ......................................................................................................................... 14 References ............................................................................................................................... 15 2 I. Preparation The following experiment is an introduction to optical microscopy with focus on confocal fluorescence microscopy and its application in biology, biophysics and medical research. The physical background and technical details is kept to minimum. However, the general principles of optical microscopy are summarized in this instruction manual, while for detailed information the reader is referred to relevant secondary literature. In order to conduct this experiment, knowledge of the following topics should be acquired by reading this instruction manual: o o o o Interference, refraction, and diffraction of light Basics of optical and fluorescence microscopy Basics of confocal microscopy Image digitalization You should be able to answer the following questions before the experiment: II. What is interference, what is refraction of light waves? What is diffraction? How does the interference pattern behind a single slit look like? What are the main components of an optical / fluorescence microscope? What determines the physical resolution of an optical microscope? What is the point spread function? How does a (laser scanning) confocal microscope work? What determines the physical resolution of a confocal microscope? Background 1. History of optical microscopy Optical or light microscopes (from Greek μικρόν or micron for "small", and σκοπεῖν or skopein for "to look at") are the oldest type of microscopes and date back into the early 17th century, when scientists like Galileo Galilei (1564-1642) invented their basic principle. Dutch scientist Antonie van Leeuwenhoek (1632-1723) introduced them into the field of biology, but the basic technology did not change until the 20th century, where revolutionary improvements where achieves such as Köhler illumination, phase contrast or differential interference contrast microscopy. Since the late 1960s fluorescence microscopy became state of the art, allowing for functional imaging of the sample. Further improvements were then the confocal microscope1 and recently super-resolution microscopy techniques2, allowing for higher resolution and contrast. 3 2. Basics of wave optics In order to understand the principles of optical microscopy, geometric optics are not sufficient and light has to be described by wave optics. Some main concepts of wave optics are briefly noted in the following. 2.1. Interference As illustrated in Figure 1, two propagating light waves can interfere due to vector properties of the underlying electromagnetic fields. If the waves have same wavelength or frequency, interference can be constructive (Figure 1a) or destructive (Figure 1b) depending on their phase difference with constructive interference for in-phase (even multiple of π or 180°) or antiphase (odd multiple of π or 180°) superposition, respectively. a) b) Wave 1 Wave 1 Wave 2 Wave 2 Result Result Figure 1. Illustration of (a) constructive and (b) destructive superposition of two propagating waves. 2.2. Refraction Refraction is the change of the propagation direction of a light wave, when passing the interface between two media (e.g. air and glass) of different refractive index 𝑛, which relates the speed of light in vacuum 𝑐 to the speed of light in the respective medium, 𝑣 = 𝑐 ⁄𝑛. According to the phenomenological Snell’s law of refraction (Figure 2a), the ratio of the sines of the two waves’ angles of incidence, 𝜗1 and 𝜗2 , is given by the opposite ratio of their refractive indices, 𝑛1 and 𝑛2 , sin 𝜗1 𝑛2 (1) = . sin 𝜗2 𝑛1 interface n1 n2 a) 1 1 normal interface v1 v2 sphere wave b) normal 2 wave front 2 λ1 λ2 Figure 2. Refraction of a light wave at an interfact between two media according to (a) Snell’s law in terms geometric optics and (b) Huygens-Fresnel principle in terms of wave optics. 4 However, the same result can be obtained using the Huygens-Fresnel principle: From the incident wave of wavelength 𝜆1 at each point of the interface of a spherical waves of wavelength 𝜆2 is emitted, forming the propagating wave by superposition (Figure 2b). Geometrically then follows sin 𝜗1 𝜆1 𝑣1 ⁄𝑓 𝑛2 (2) = = = , sin 𝜗2 𝜆2 𝑣2 ⁄𝑓 𝑛1 since the wave’s frequency 𝑓 is the same in both media. 2.3. Diffraction The Huygens-Fresnel principle also allows to intuitively describe diffraction, which takes place when a light wave interacts with structures of size similar to the wavelength. For example, if a narrow slit is illuminated with coherent light (Figure 3a), the transmitted light exhibits a typical diffraction pattern of bright and dark regions on a distant screen (Figure 3b). single slit b) screen c) intensity I / I0 a) 1.0 0.5 0.0 -4 -2 0 2 angle sin(α) b / λ 4 Figure 3. Diffraction at a single slit. (a) Schematic experiment. (b) Photography of the diffraction pattern and (c) intensity profile according to Equation (3). The interference profile of the diffraction pattern can be derived using the Fresnel-Kirchhoff diffraction formula to calculate the superposition of the individual waves on the screen. For a slit of width 𝑏 the intensity profile is given by (3) 𝐼(𝛼) = 𝐼0 sinc 2 [π sin(𝛼) 𝑏⁄𝜆] , using sinc 𝑦 = sin(𝑦)⁄𝑦 for the so-called sinc function. While the derivation of the complete profile is relatively complicated, the location of the intensity minima can be constructed easily. In a simplified geometric consideration, for the first minima the transmitted wave is divided into two parts of width 𝑏⁄2 (Figure 4), which interfere destructively, if their path difference 𝛿 is equal to 𝜆⁄2. Therefore, the first minimum can be observed under an angle of 𝛼 = arcsin(𝜆⁄𝑏 ). For the higher order minima the wave is divided into four, six, etc. parts, which then interfere destructively on their own. Therefore, the minima can be observed under the angles 𝛼𝑛 = arcsin(𝑛 𝜆⁄𝑏 ) with 𝑛 = 1, 2, 3, … (4) (see Figure 3c). 5 Figure 4. Diffraction at a single slit of width 𝒃 and geometric consideration to explain the location of the minima under angle 𝜶. The transmitted wave is divided into two parts with path difference 𝜹. Usually, the lateral location on the screen can be apprimated for small angels using 𝐭𝐚𝐧𝜶 = 𝒙⁄𝑳 ≅ 𝜶 in radians. When the single slit is replaced with a double slit, the diffraction pattern changes, because now effectively the two beams from the two slits interfere with each other. From similar geometric considerations (Figure 5) it can be deduced that the path difference between the beams is 𝛿 = 𝑔 ∙ sin 𝛼. Constructive or destructive interference then occurs if the path difference is an even or uneven multiple of 𝜆⁄2, respectively. So the angle, under which maxima and minima can be observed is 𝜆 (5) 𝛼𝑛 = arcsin (𝑛 ) 2𝑔 with 𝑛 = 0, 2, 4, … for the maxima and 𝑛 = 1, 3, 5, … for the minima. Figure 5. Diffraction at a double slit of slit distance 𝒈 and geometric consideration to explain the location of minima and maxima under angle 𝜶 with the path difference 𝜹 of the transmitted waves. a) intensity I / I0 2.4. The Airy disk If the diffraction sample is a circular aperture, the diffraction pattern changes to characteristic two-dimensional structure (Figure 6), which is commonly denoted as Airy disk or generally Airy pattern (after George Biddell Airy). b) 1.0 0.5 0.0 -2 0 2 angle sin(α ) b / λ Figure 6. Diffraction at a circular aperture of radius 𝒂. (a) Cross section and (b) two-dimensional calculation of the Airy pattern (logarithmic color scale). 6 3. Experimental setup Due to the complexity of the topic, only the basic experimental designs of widefield and confocal fluorescence microscopes are described. For further information, the reader is referred to relevant secondary literature. 3.1. Fluorescence microscopy In fluorescence microscopes, not the absorbance or reflectance of the sample is investigated, but it is stained with specific functionalized dyes to image just certain structures of interest. The sample is then illuminated not with white light, but with light of a narrow spectrum to excite the dyes, which then emit light of a certain (generally longer) wavelength. Excitation and emission light is usually separated with a dichroic mirror and an optional emission filter (see Figure 7 for details). The dichroic mirror and the filters are often combined in a “filter cube” to allow for easy change. Figure 7. Basic experimental setup of a fluorescence microscope. Light from a broad-spectrum lamp (mercury vapor or, increasingly, LED lamps) is filtered by an exitation filter (here shown for green light) and then reflected on the sample by a dichroic mirror, which is reflecting below and transmitting above a cerctain wavelength. Sample features then fluorescently emit light (here shown in red), which is displayed on the camera. Out-of-focus objects are also projected onto the camera, but appear blurred. 3.2. Confocal microscopy Conventional widefield microscopes suffer from the fact that also all light from out-of-focus regions of the sample are collected by the objective, which massively increases the background when imaging thick samples (see the two out-of-focus objects in Figure 7). This problem is solved in the confocal imaging principle, which was invented by Marvin Minsky in 1957,3 by imaging with a diffraction-limited light spot (usually by focusing a laser beam) and blocking out-of-focus emission with an aperture (usually denoted as “pin-hole”) in the detector light pass (see Figure 8 for details). This practically eliminates the image background and slightly improves the optical resolution (see below, section 0). 7 Figure 8. Principle and basic experimental setup of a confocal microscope. The sample is illuminated with a diffractionlimited spot, often using a focussed laser beam. Excitation of the sample is then limited to “focal volume”. In the detection light path a “pin-hole” aperture is placed in the focal plane. While light emitted from in-focus sample features passes the pin-hole (dashed light path), light emitted from out-of-focus sample features is than practically blocked by the pinhole and does not reach the detector (dotted light path). However, the confocal imaging principle now results in the expense that imaging now has to be performed by somehow scanning illumination and detection region through the sample. In principle, that would be possible by moving the sample relative to the stationary optics, but is often easier and especially faster to move the confocal volume with respect to the sample, which is often realized by confocal laser scanning microscopy (CLSM, often simply named ‘confocal microscopy’) (see Figure 9 for details). The image is then recorded by pixelby-pixel and line-by-line scanning the confocal volume through the sample while recording the intensity on the detector, which is usually a photomultiplier tube (PMT) or an avalanche photodiode. Owing to the efficient blockage of out-of-focus light, CLSM is often used for optical sectioning and recording “stacks” of images through thick samples. For multiple fluorescence imaging the imaging procedure is sequentially repeated for each channel. Figure 9. Experimental setup of a confocal laser scanning microscope, where the scan movement is realized by scanning the confocal volume through the sample. This is often implemented by tilting the parallel excitation and emission beams together, for example, using separate galvano mirrors for the lateral 𝒙- and 𝒚movement. The vertical movement can be realized by moving the objective with a 𝒛scanner. 8 The instrument used in the practical work is a commercial Nikon C2 confocal microscope with four separate lasers for illumination (405 nm, 488 nm, 561 nm, and 640 nm) and three photomultiplier tubes (PMT) for detection, coupled to the scan head with optical fibers. The scan head contains the 𝑥- and 𝑦-scanning mirrors and is mounted to the microscope at the camera port, while the 𝑧-scanning is realized by the motorized focus of the microscope body. 3.3. Immersion oil objectives For achieving high image intensity and resolution (see below), it is advantageous to collect the light emitted from the sample under an angle as large as possible. If sample and objective are separated by air, the angle 𝛼, under which the sample is imaged, is effectively smaller because of refraction of the light at the air/glass interfaces (Figure 10a). Furthermore, the maximum angle is limited by the critical angle for total internal reflection (typ. 40°). If the space between sample and objective is filled with a medium of refractive index similar to the glass, refraction is minimized, which is usually achieved by immersion oil (Figure 10b). The angle of observation and the refractive index of the separating medium are often combined in the “numerical aperture” (6) NA = 𝑛 sin 𝛼 , 4 which was introduced by Abbe already. Figure 10. Difference in light path for objective (a) without and (b) with immersion oil with refractive index 𝒏 > 𝟏. 4. Resolution in optical microscopy If microscopic objects are imaged with optical microscopes, the resolution is ultimately limited by diffraction, which was first derived by Ernst Abbe in 1873.4 When a sub-wavelength object is illuminated, its image is not point-like but a diffraction pattern according to the Airy pattern (see above). Why is that so? 4.1. The point spread function (PSF) The illuminated object emits light in all directions. However, the objective only collects the light, which is emitted under an angle of 𝛼 or smaller, given by the opening angle of the objective, which thereby acts as a circular aperture. The image of the object therefore corresponds to the Airy diffraction pattern. Hence, every point-like object looks like this diffraction pattern, which is therefore denoted as “point spread function” (PSF) of the microscope. 4.2. The Rayleigh criterion Based on the work of on John William Strutt, 3. Baron Rayleigh, the two objects can just be separated from each other in the image, the location of the one object is at first minimum of second particle’s Airy pattern.5 The distance of the objects is then 9 𝜆 𝜆 (7) = 0.61 , 𝑛 sin 𝛼 NA for refractive index of the medium 𝑛 and opening angle of the objective 𝛼 or numerical aperture NA [Equation (6)]. For illustration, images of two objects with distance above, equal to, and below Rayleigh criterion are shown (Figure 11). 𝑑 = 0.61 a) b) c) Figure 11. Resolution limit according to the Rayleigh critertion. Image of two closely-separated objects according with distance (a) above, (b) equal to, or (c) below the Rayleigh criterion (logarithmic color scale). Experimentally, it is complicated to realize the Rayleigh criterion, because for that the position of sub-wavelength objects would have to be controlled accordingly. For practical reasons the resolution of optical microscopes is therefore often determined from the width of the PSF by imaging individual point-like objects such as sub-µm sized fluorescent microbeads (see for example protocol in 6). The full width at half maximum (FWHM) is very similar to the Rayleigh criterion and is given in the lateral direction by7 𝜆em (8) FWHM𝑥,𝑦 = 0.51 . NA In axial direction the FWHM is given by 𝑛𝜆em (9) FWHM𝑧 ≅ 1.77 . NA² In case of widefield fluorescence microscopy λem refers to the emission wavelength. 4.3. Resolution in confocal microscopy In confocal microscopy the resolution is slightly improved, both illumination and detection patterns are now diffraction-limited and the effective PSF is then the multiplication of the illumination and the detection PSF. If the projected pinhole diameter is equal to the diameter of the first order minimum in the Airy disk pattern, referred as to 1 Airy unit (or 1 AU), then the lateral resolution is given by6: 𝜆ex (10) FWHM𝑥,𝑦 = 0.51 NA In confocal microscopy only molecules inside the diffraction limited excitation volume can emit fluorescence light. Therefore and contrary to widefield fluorescence microscopy, the excitation wavelength λex is used to calculate the FWHM. In axial direction, a pinhole diameter of 1 AU will result in a better resolution compared to the widefield setup 6: 𝜆ex FWHM𝑧 = 0.88 (11) 𝑛 − √𝑛2 − NA2 10 Theoretically, smaller pinhole diameters will further enhance the lateral and axial resolution slightly, but at the expense of signal intensity, as more light is then blocked by the pinhole. 4.4. Nyquist–Shannon sampling theorem When converting continuous signals (“analog signals”) into discrete signals (“digital signals“) the sampling rate must be at least more than two times the maximum frequency of the signal, in order to represent the signal without loss of information. This criterion is denoted as the Nyquist-Shannon sampling theorem.8 Figure 12 shows the transformation of an analogue signal into a digital signal at different sampling rates. For example, in case of an audio signal with a maximum frequency of 20 kHz the sampling frequency must be more than 40 kHz. For the compact disc digital audio (“Audio-CD”) the sampling rate is 44.1 kHz, thus satisfying the theorem. Output Samples Input The Nyquist-Shannon theorem also applies for spatial information. In optical microscopy, to resolve a structural size of e.g. 250 nm, the pixel size of the detection system has to be less than 125 nm. To be on the save side, often 1⁄3 of the object frequency is used as pixel size (e.g. 250 nm⁄3 ≈ 80 nm). A higher sampling rate (smaller pixel size) will not result in a gain of information (“empty magnification”). Figure 12. Transformation of analoge signal into digital signal at increasing sampling rate. 11 III. Experiment 1. Laboratory The confocal microscope used for the experiment is an expensive research microscope. The experimental work may therefore be started only after a comprehensive introduction and together with the instructor. If you are not sure, always ask the instructor first! Before starting the experiments, read and sign the laser safety form for each of the two setups! In the experiment the following tasks should be performed: 1. The practical work starts with the diffraction experiments at the optical bench (see Figure 13). A set of single and double slit samples are provided. The laser wavelength is 𝜆 = 633 nm. Figure 13. Schematic setup of the optical bench with laser illuminating a slit and the diffraction pattern projected on a screen. a. Adjust the collimator lens to generate a parallel laser beam. Start with the widest single slit (𝑑 = 0.8 mm). Align the laser to the slit and investigate the diffraction pattern. i. Sketch the intensity profile. ii. Measure the angle between the center maximum and the first order minimum and compare it with the expected values. b. Repeat the experiment as in a.i and a.ii with a narrower single slit (e.g. 𝑑 = 0.4 mm) and the narrowest single slit (𝑑 = 0.1 mm). How does the diffraction pattern change? c. Replace the sample with the double slit (slit distance 𝑔 = 0.3 mm, slit width 𝑑 = 0.1 mm. Investigate the diffraction pattern and compare it with the last single slit. d. Continue the experiment with all other double slits of different slit width vs. slit distances. i. How do the diffraction patterns change? ii. How do slit width and slit distance affect the diffraction pattern? e. If time allows, insert a collecting lens and the adjustable slit according to Figure 14. 12 Figure 14. Schematic setup of the optical bench modifed by inserting a lense with focal length 𝒇 for demonstrating the contribution of diffraction orders to the image. i. Move the adjustable slit in such a way that only the 0-order maximum can pass through. Discuss your observation. ii. Move the adjustable slit in such a way that only higher order maxima (1st order and above, 0-order excluded) can pass through. Discuss your observation. f. Replace the slit sample with the pinhole and remove the collecting lens and the adjustable slit. Investigate the diffraction pattern and calculate the pinhole diameter. 2. In the second part of the practical work the optical microscope in “conventional” widefield configuration is investigated. a. Mount the commercial sample with fluorescent microspheres (“beads”) into the microscope and focus on the sample position 4 (0.2 µm beads) using a lowmagnification air objective (e.g. 20x). b. Record an image and zoom in onto a single bead to inspect the point spread function (PSF): i. Use the microscope software to create a cross section through the PSF. ii. Compare the shape of the PSF qualitatively with the expected shape. At home measure the FWHM of the PSF and compare it with the expected value [see Eq. (8), note that you need the NA of the objective for that]. c. Now change to a high-magnification oil objective (e.g. 60x). Caution! The objective is very expensive and must be handled by the assistant only! Change to position 5 of the sample (0.1 µm beads). Again, focus on the beads and inspect the PSF: i. Create a cross section through the PSF. ii. At home compare the shape and width of the PSF with the result from section 2.b and with the expected value. iii. Discuss the different factors affecting the image resolution (optical resolution, pixel resolution). Discuss whether the pixel size meets the Nyquist-Shannon criterion. What is “empty magnification”? 13 3. In the third part of the practical work the optical microscope in confocal configuration is investigated. a. Use the bead sample to investigate the confocal PSF: i. Set the pinhole to approx. 1 AU and adjust the imaging parameters. ii. Zoom in onto a single bead and record the 𝑥𝑦-PSF and create a cross section. Use appropriate pixel size to meet Nyquist-Shannon criterion. iii. At home compare the shape and width of the PSF with the expected value [see Eq. (10) and (11)]. Compare the confocal PSF with the PSF in conventional mode. iv. Now record an appropriate 𝑧-stack of the bead and plot the 𝑥𝑧-PSF. v. Change the pinhole to maximum (approx. 6 AU and compare the 𝑥𝑦and 𝑥𝑧-PSF. b. Now change to the commercial cell sample. Place the sample into the microscope and focus on the cells. i. Record a widefield image of the cells with a single channel, e.g. FITC. ii. Change to confocal configuration and record a single-channel 𝑥𝑦image. Compare it with the widefield image. iii. Set up multicolor imaging, e.g. FITC and DAPI, and record an 𝑥𝑦-image. iv. Finally record an appropriate 𝑧-stack and construct 𝑥𝑧- and 𝑦𝑧-slices, a maximum-intensity projection and a 3D view of the data. v. Use the laser power meter to measure the laser power of the 488 nm laser in the location of the sample with your last used settings. At home compare irradiance of your 488 nm laser power setting with the irradiance of the sun in central Europe. Irradiance is the radiant flux (power) received by a surface per unit area (unit W/m2 ). The illuminated area can be considered circular with the diameter estimated by the FWHM of the PSF. In central Europe, the irradiance of the sun peaks at approx. 1000 W/m2 (summer, midday, clear sky). Think about possible consequences of your results for live biological samples. At the end, clean the oil objective and sample following assistant’s instructions. (Use high-grade ethanol and Kimwipes and be very careful!) 2. Protocol The protocol should give a brief description of the basics of optical and confocal microscopy. Discuss similarities and differences of widefield vs. confocal microscopy. The major part of the protocol should be the presentation of the results obtained during the experiment and their detailed discussion and interpretation. Include error analysis for the single slit experiments and compare your results with expectations from theory. 14 Error analysis: Estimate how accurate you were able to measure the distance from slit to screen, e.g. 1 cm accuracy. This will result in a distance of e.g. 550 ± 1 cm. In a similar way, estimate your measurement error when measuring the distance between minimum and maximum on the screen. Then estimate, how these two accuracies affect your final results for the calculated angles, which should be stated in a way such as, e.g., = 0.024 ± 0.002 °. The protocol must contain a signed declaration that the elaboration of the report was performed by you and your group (see below): Name: .............................. Matriculation number: .............................. I hereby declare that my elaboration (or that of my group) was made independently and is not a copy (even partially) of an existing protocol or of the instruction manual. All external sources (also the instruction manual) have to be cited. I am aware that a violation of this rules is plagiarism and leads to the experiment being graded as failed (grade 5) and can result in an exclusion from the labwork. .............................. Date .............................. Signature References (1) (2) (3) (4) (5) (6) (7) (8) Jonkman, J.; Brown, C. M., Any Way You Slice It - A Comparison of Confocal Microscopy Techniques. J. Biomol. Tech. 2015, 26, 54-65. Hell, S. W.; Sahl, S. J.; Bates, M.; Zhuang, X.; Heintzmann, R.; Booth, M. J.; Bewersdorf, J.; Shtengel, G.; Hess, H.; Tinnefeld, P.; Honigmann, A.; Jakobs, S.; Testa, I.; Cognet, L.; Lounis, B.; Ewers, H.; Davis, S. J.; Eggeling, C.; Klenerman, D.; Willig, K. I.; Vicidomini, G.; Castello, M.; Diaspro, A.; Cordes, T., The 2015 super-resolution microscopy roadmap. J. Phys. D Appl. Phys. 2015, 48, 443001. Minsky, M., Memoir on inventing the confocal scanning microscope. Scanning 1988, 10, 128-138. Abbe, E., Beiträge zur Theorie des Mikroskops und der mikroskopischen Wahrnehmung. Archiv f. mikrosk. Anatomie 1873, 9, 413-418. Rayleigh, L., On the theory of optical images, with special reference to the microscope. Phil. Mag. 1896, 42, 167-195. Cole, R. W.; Jinadasa, T.; Brown, C. M., Measuring and interpreting point spread functions to determine confocal microscope resolution and ensure quality control. Nat. Protocols 2011, 6, 1929-1941. Amos, B.; McConnell, G.; Wilson, T., 2.2 Confocal Microscopy A2 - Egelman, Edward H. In Comprehensive Biophysics, Elsevier: Amsterdam, 2012; pp 3-23. Shannon, C. E., Communication in the Presence of Noise. Proceedings of the IRE 1949, 37, 10-21. 15