Uploaded by glyzel Agguire

AUTOMATION

advertisement
AUTOMATION
Instrumentation
4 types of analyzer
1. Continues flow - more than 1 test in one sample
2. Centrifugal analysis - multiple sample in one test
3. Discrete analyzer - random access testing
4. Dry slide - uses micro moles and reagents on slide of dry chemistry analysis (kodak electron)
BASIC APPROACHES
Continues Flow Analysis
• Reagents, diluents, and samples are pumped through a system of continuous tubing
• Air bubbles ( to minimize diffusion of reagents and mixing of samples) serve as separating
and cleaning
• All tests are performed in parallel testing. (more than one test is analyzed concurrently on
organism specimen)
Disadvantage
1. Carry over effect
2. Wasteful use of reagents
Centrifugal Analyzer
• It uses the force generated by centrifugationto transfer specimen and reagents
• Tests are performed as Batch analysis. (all samples are loaded atthe same time but one test
only)
HbA1C = whole blood (long term monitoring of DM)
Discrete Analyzer
• Most Popular & Versatile Analyzer (it only measures the test that is/are requested on a
sample)
• It employs a variety of syringe pipets to aspirate & dispense samples & reagents
• Uses random access capability / testing (any test can be performed on any sample, in any
sequence)
Allows STAT samples to be easily repeated
Sequential testing - multiple test analyse on one sample after another on a given
specimen
LIGHT ENERGY, WAVELENGTH, RADIANT ENERGY
Radiant Energy
• characterized as a spectrum from short wavelength to long wavelength: cosmic, gamma
rays, X-rays, ultraviolet, visible,infrared, microwaves, radiowaves.
Wavelength
• Distance between 2 successive peaks
• Expressed in nanometers (nm)
• Ultraviolet Light : < 400 nm
• Visible Light : 400-700 nm
• Infrared Light :>750 nm/>700 nm
X-rays = have long wavelength
Microwave = shortest wavelength
Very long wavelength = infrared
Very short = ultraviolet
Frequency
• number of vibration of wave per second
Wavelength is inversely related to Frequency & Energy
SHORTER wavelength :High Frequency, High Energy
LONGER Wavelength : Low Frequency, Low Energy
Wavelength
<400
400-430
430-480
480-560
560-590
590-630
630-750
>750
Color Absorbed
Not visible
Violet
Blue
Green
Yellow
Orange
Red
Not visible
Spectrophotometry
• Instrument that measures light energy ()
• Involves measurement of the light Transmittance by a solution to determine the concentration of
the light absorbing substances in the solution.
SPECTROPHOTOMETRY
• PRINCIPLE:Beer Tambert’s Law
• States that Concentration of Unknown Substance is directly proportional to Absorbed
Light (Absorbance) Kung gaano kconcentrated ang specimen ganun din kadami ang
maabsorb na light
• Concentration of Unknown Substance is inversely proportion to the Transmitted Light
(Transmittance) dipende ang concentration sa na-absorb na light. Kung gaano
kaconcentrated ang specimen, ganun naman ang kinakonti ng matratransmit na light
Absorbance
Light absorbed by the
solution
to the concentration
of solution
Transmittance
Amount of light that
passes through colored
solution
Indirectly proportional to
the concentration of
solution
calculate analytic concentration through
1. Molar absorptivity value
A= abc
a = molar absorptivity
b = length of light
c = concentration molecules
A= 2-log %T
A= log (100/%T)
2. Ratio of standard to the unknown
Cu= Au / As x Cs
3. Standard curve (graph paper)
A. Semilogarithmic paper
X- concentration of the standard solution
Y - transmittance or the absorbance
1. Line is straight
2. Line connects all post
3. Line goes to the origin or intersect the axes
B. Linear graph paper
SPECTROPHOTOMETRY
Parts of Spectrophotometer:
1.Light Source
• Provide Incident Light/ Radiant Energy
Types of Light Source:
• Continuum source
• Emits Radiation that changes in intensity (Widely Used:Tungsten; Deuterium,
Xenon)
• Line source
• Emits Radiation & Wavelength (Hg; Hollow-Cathode;LASER light amplification by
stimulated emission of radiation)
1. LIGHT SOURCE
Types
Incandescent lamp/ tungsten iodine lamp
Deuterium discharge lamp
Mercury lamp
Silicon carbide/ quartz halogen
Xenon lamp
Description
 Typical source in most spectrophotometers

For Visible & Near Infrared Region (320-700
nm)
 For UV spectrum

Used in UV & Fluorescent & HPLC

For Infrared Region
 Produce a continuous spectrum of fairly
uniform intensity from 300-2000 nm
 Useful for Visible, & UV applications
Parts of Spectrophotometer:
2. Entrance Slit
• Minimizes unwanted/ Stray Light and prevents entrance of scattered light into the
monochromator system.
•Stray light : Presence of any light other than the wavelength of measurement reaching the
photodetector.
• Most common cause of loss of linearity
3. Monochromator
• Isolates specific or individual wavelength of light.
Types:
• Diffraction gratings
• most commonly used because it has a better solution than prism
• Glass with grooves and slit used for UV, Visible & Infrared
• Diffraction, the separation of light into component wavelengths, is based on the
principle that wavelengths bend as they pass a sharp corner.
3. MONOCHROMATOR
Types:
• Prism glass, quartz, sodium chloride
• Produces the purest monochromatic light in UV & Separates white (Visible) Light
into a continuous spectrum
• Interference filter semi transparent silver films
• Produce monochromatic light based on the principle of constructive interference
of waves
Parts of Spectrophotometer:
4. Exit Slit
• Controls the width of light beam (Bandpass)
• Bandpass : Total Range of wavelengths transmitted
• Exit slit selects the bandpass of the monochromator that allows light of the selected
wavelength to pass through the cuvet onto the detector.
Parts of Spectrophotometer:
5.Cuvet/ Cuvette/ sample cell
• Analytical cell / absorption cell / photometer tubes
• It holds the solution whose concentration is to be measured
TYPES
DESCRIPTION
Alumina silica glass
Quartz/plastic/fused silica
Borosilicate glass
Soft glass
Most Commonly Used in Visible Region
Measurements below 340 nm (UV & Visible)
For Alkaline solution
For Acidic solution
Parts of Spectrophotometer:
6. Photodetector
• Detects and converts transmitted light into photoelectric energy
• The more light transmitted, the more energy, and the greater the electrical signal that is
measured
TYPES
Photocell
Phototube
Photomultiplier tube
Photodiode
DESCRIPTION
Aka “Barrier Layer Cell/ Photovoltaic cell”
Simplest Detector; least expensive; TemperatureSensitive
Contains cathode & anode enclosed in glass case
Most Commonly Used (Visible & UV)Detects Very
Low Levels of Light (↑ Sensitivity; Rapid
response)
Excellent linearity
Parts of Spectrophotometer:
7. Meter/Read-out Device
• Displays the output of the detection system
TYPES OF SPECTOPHOTOMETERS:
• Single Beam
• Simplest Type of Absorption Spectrometer and designed to make one measurement at a
time at one specifies wavelength.
•Double Beam
• An Instrument that splits the monochromatic light into 2 components- one beam passes
through the sample and through a reference solution or blank. Reagent blank (compensate
reading), water blank (set spectro reading to 0), sample blank(compensate for unsafe
absorbance (hemolysis, steric sample,lypinic sample))
• Double-Beam in Space : 2 photodetector. 1 for sample, 1 for reference beam
• Double-Beam in Time : 1 photodetector
Wavelength accuracy
ISSUES IN QUALITY CONTROL OF SPECTROPHOTOMETER
Means the wavelength indicated on the control
dial is actual wavelength of light passed by the
monochromator
Stray light (entrance slit)
Determined by use of Holmium Oxide &
Didynium Filter
Presence of any light other than the wavelength
of measurement reaching the photodetector
Most Common Cause: Reflections of light from
scratches on optical surfaces
Linearity
Remedy:use a sharp cut of filter
Change in the concentration results to a straight
line
Atomic Absorption Spectrophotometer
Principle: Ground state atoms absorb light at defined wavelength
• Measures the light absorbed by Ground State Atoms (Unexcited atoms) hollow cathode
lamp
• Measures concentration by detecting the absorption of electromagnetic radiation by atoms
rather than by molecules
•More sensitive that FES
• Reference Method for Calcium, magnesium
• Lanthanum/ Strontium Chloride: Form stable complexes with PO4 (PO4 causes interference
to Ca)
• Components
• hollow cathode lamp - chopper - flame- monochromator Flame Emission Spectrophotometry
• Measures the light emitted by single atom burned in a flame
• Measures the light absorbed by Excited State Atoms
• Lithium / cesium - internal standard correct variation in flame and atomizer
characteristics
• Measurement of Na; K; Li (Excited atoms)
Turbidimetry
• Measures the amount of Light Blocked by suspension of particles at 180 Angle
• Measures abundant large particles (proteins) & bacterial suspension.
• Ex: Protein Measurement; Clot formation Detection; Bacterial growth
Nephelometry
• Measures the amount of Light Scattered by small particles at 90 degree angle to the
incident light
• The amount of scatter is Dependent to the number and size of particles
present in
the solution
• Light Source : Merciry arc, tungsten, light emitting diode
• Measurement of CRP; Antigen-Antibody complexes, proteins.
Rayleigh’s law
__________________
• Measurement of light emitted by a substance that has been previously excited by a source
of UV light
• Measures the amount of light present over a zero background
• Emitting Light energy of a ______________________
• Light Source : _____________________
• ______________________
• Primary Filters : _________________________
• Secondary Filters : _______________________
__________________
• Advantages: • Fluorometry is about 1000 times more sensitive than
absorption techniques and has increased specificity
because optimal wavelengths are chosen both for
absorption (excitation) and for monitoring emitted
fluorescence. • Limitations: • Changes from the established protocol that affect pH,
temperature, and solvent quality; self-absorption; quenching
• _______________ : Process by which fluorescence of an
analytes is reduced due to excited molecule losing some of
its energy by interacting with other substances in a solution
FLUOROMETRY
• The emission of light produced by certain
substances after they absorb energy. • It is similar to fluorescence except that the time
delay is ______________________ between
absorption of radiant energy and release of
energy as photons of light.
__________________
• Diff. from Fluorometry since it does not require
Excitation Energy (Light source) &
Monochromators
• The process where the chemical energy of a
reaction produces excited atoms, and upon
electron return to ground state, photons of light
are emitted
• Reactions are oxidation reactions of
_________________
__________________
• A technique where solutes in a sample are
separated for identification based on physical
differences that allow their differential distribution
between a mobile phase and a stationary phase. • Components: • __________________ : carries the
sample
• __________________ : Through w/c the mobile phase
flows. • __________________ : Holding the stationary phase
• __________________ : Separated Components
__________________
CHROMATOGRAPHY
TYPES OF CHROMATOGRAPHY
PLANAR CHROMATOGRAPHY
1. ____________________
• Used for the fractionation of sugar and Amino
Acid
2. ____________________
• ____________________________________
TYPES OF CHROMATOGRAPHY
COLUMN CHROMATOGRAPHY
1. ____________________
• Separate mixtures of compounds that are ____________ or
can be converted to volatile form
• It is used for separation of steroids, Barbiturates, Blood,
Alcohol, Lipids
• An inert carrier gas (mobile phase) carries the vaporized
sample into the column. Carrier gases commonly used
include hydrogen, helium, nitrogen, and argon
CHROMATOGRAPHY
TYPES OF CHROMATOGRAPHY
COLUMN CHROMATOGRAPHY
1. Gas Chromatography
• ____________________ : separation occurs in absorption at the
solid phase
• ____________________ : Occurs by diff. in solute between the
gas mobile & Liquid stationary
CHROMATOGRAPHY
TYPES OF CHROMATOGRAPHY
COLUMN CHROMATOGRAPHY
2. _______________________
• ________________________________
• ___________________ : Isolate drug
• ___________________ : Based on the fragmentation &
ionization of molecules using energy
CHROMATOGRAPHY
TYPES OF CHROMATOGRAPHY
COLUMN CHROMATOGRAPHY
2. _______________________
• ________________________________
• ___________________ : Isolate drug
• ___________________ : Based on the fragmentation &
ionization of molecules using energy
CHROMATOGRAPHY
TYPES OF CHROMATOGRAPHY
COLUMN CHROMATOGRAPHY
3. ______________________
• Based on the distribution of solutes between a mobile Liquid phase &
Liquid Stationary Phase
• _____________________________
• ______________________________________
• Uses pressure for fast separations, controlled temp, in-line
detectors
• Used for Fractionation of Drugs, Hormones, Lipids, CHO, CHON
• Separation of Hemoglobins (Thalassemia) & for HbA1C Test
CHROMATOGRAPHY
TYPES OF CHROMATOGRAPHY
COLUMN CHROMATOGRAPHY
3. LIQUID CHROMATOGRAPHY
• _____________________________
• Used for detecting Non-Volatile Subs
• Confirm positive drug screening tests;
• Complementary to GC-MS
CHROMATOGRAPHY
CHROMATOGRAPHY
Separation Mechanisms Used In Liquid Chromatography
• Exchange of sample ions and mobile-phase ions w/ the
charged group of stationary phase
• Separation of Amino Acids, proteins & Nucleic Acids
• Liquid-Liquid Chromatography
• Based on solubility (Polar/Non-Polar)
• Separation of lipoproteins, CHO, HBA1C, ab
• Liquid-Solid Chromatography (Separation is based on the
differences bet. absorption and desorption of solutes at
the surface of a solid particle
The measurement of current or voltage generated by the activity
of specific ion.
___________________
• Determine the concentration of a substance in solution
employing an electrochemical cell that consists of two halfcells, where the potential difference
between an indicator
electrode and a reference electrode is measured.
• Measurement of electrical potential (Voltage) between 2
electrodes
• Use : ________________________
ELECTROCHEMISTRY
________________________
• Electrochemical Transducer capable of
responding to one given ion
• Direct ISE (Without Dilution):
• _______________________________
• Indirect ISE (With Dillution):
• _______________________________
POTENTIOMETRY
ISE MEMBRANES:
• Sodium : ______________________
• Potassium : ______________________
• Calcium & Lithium : ______________________
• Hydrogen : ______________________
POTENTIOMETRY
________________________
• Glass Electrode
• Silver-Silver Chloride Electrode
• Calomel Electrode (Hg2Cl2)
________________________
• Severinghaus Electrode
POTENTIOMETRY
• Amount of increase in current (i.e., the wave height)
proportional to the concentration of analyte. • Current measured; voltage change versus current
plotted to produce a polarogram.
• Voltage at which sharp rise in current occurs
characteristic of the electrochemical reaction
involved
___________________
• Measurement of Current Flow produced by
oxidation-reduction reaction
• Use : _______________________
• Also used in Glucose (Polarography); Chloride;
Peroxidase
___________________
• Measures the quantity of Electricity (Coulombs)
needed to convert an analyte to a different
oxidation state
• Use : _____________________
• Interferences : ____________________
___________________
• Method in which a potential is applied to an
electrochemical cell and the resulting current is
measured
• Use : ____________________________
___________________
• The Process of separating the charged constituents
of a sample by means of electrical current
• Electrophoresis is defined as the movement of
charged molecules in a liquid medium when an
electric field is applied
___________________
• Components:
• _________________________
• _________________________
• _________________________
• _________________________
• _________________________
ELECTROPHORESIS
ELECTROPHORESIS
SUPPORTING MEDIA
molecular size
charge
Molecular Size
surface charge
& Molecular
Size
mon
(Routine)
(Routine)
isozenzymes
used)
ELECTROPHORESIS
• Detecting System : ___________________
• Colorimetric Stains Or Fluorescent Chemicals.
• Serum proteins : _____________________________
• CSF proteins : _____________________________
• Lipoproteins : _____________________________
• Lactate dehydrogenase isoenzymes : __________
ELECTROPHORESIS
Commonly Encountered Problems In
Electrophoresis
• Holes in staining pattern
• _______________________________
• Very slow migration
• _______________________________
• Sample precipitates in support
• _______________________________
ELECTROPHORESIS
ELECTROPHORESIS
TYPES OF ELECTROPHORESIS
• Migration of Small Ions
• Migration of Charged Macromolecule
• The movement of buffer ions and solvent based fixed
support
• Migration through pH gradient (Isoenzymes, CSF
oligoclonal) • It is ideal for separating proteins of identical sizes with
different net charges
• Molecules are separated by electro-osmotic flow (EOF)
• Most Important Application: _______________________
• Hemoglobin Electrophoresis
• _______________ : Tetramer composed of four globin
chains, four heme groups, and four iron atoms
a. Hemoglobin A1 : __________________________________
b. Hemoglobin A2 : __________________________________
c. Hemoglobin F : __________________________________
ELECTROPHORESIS
• __________________________
A number of exist where a substitution of one amino
acid on either the alpha chain or the beta chain
causes the formation of an abnormal hemoglobin
molecule
ELECTROPHORESIS
Hemoglobinopathies Substitution
for glutamic acid in position 6 of the beta chain
chain
chain
THANK YOU!!!
JBCRMT
DO NOT PHOTOCOPY / REPRODUCE
WITHOUT PERMISSION!
WASH YOUR HANDS AND WEAR YOUR
Download