use a petri dish and put soap and water to make a bubbly solution Calibrate to: 1L/min 1.5L/min 2L/min blank is your cassette and pre-calibrate before and after flow airflow Inventory used: - Bubble tube - pump - 3 clear tubes - 1 petri dish - soap - water - cassette - 1 small screwdriver Set up for Bubble Flow Calibration: Cassette has an inlet connected to the bubble tube and outlet side to the pump where it reads inlet. Put Plug pump to the wall. Put petri dish underneath the bubble tube once you have the soap and water on them (only touch the surface of the water). Calibrate by using a screwdriver to move the flow to 1L/ minute. Use a cap on the top of the bubble tube to cover the airflow from the top of the bubble tube. 1L/minute: First Run 1000mL to 0mL: 75.00 seconds Second Run 1000mL to 0mL: 74.80 seconds Third Run 1000mL to 0mL: 73.14 seconds Calculate Flow Rate First Run (1L/min): Volume/Time: 1L / 1.25 (75/60seconds) = 0.80 Second Run (1L/min): Volume/Time: 1L/ 1.25 (74.80/60 seconds) = 0.80 Third Run (1L/min): Volume/Time: 1L/ 1.22 (73.14/ 60 seconds) = 0.82 Average Time: 0.81L/minute 1.5L/minute: First Run 1500mL to 0mL: 74.42 seconds Second Run 1500mL to 0mL: 74.11 seconds Third Run 1500mL to 0mL: 74.48 seconds Calculate Flow Rate: First Run (1.5L/min): Volume/Time: 1.5L/ 1.24 (74.42/60 seconds) = 1.21 Second Run (1.5L/min): Volume/Time: 1.5L/1.24 (74.11/60 seconds) = 1.22 Third Run (1.5L/min): Volume/Time: 1.5L/1.24 (74.48/60 seconds) = 1.22 Average Time: 1.22 L/ minute 2L/minute: First Run: 2000mL to 0mL line time: 63 seconds Second Run: 2000mL to 0mL: 74.96 seconds Third Run: 2000mL to 0mL: 73.56 seconds Calculate Flow Rate: First Run (2L/min): Volume/Time: 2L/ 1.05 (63 seconds/60 seconds) = 1.90 Second Run (2L/min): Volume/Time: 2L/ 1.25 (74.96/60 seconds) = 1.60 Third Run (2L/min): Volume/Time: 2L/ 1.23 (73.56/60 seconds) = 1.63 Average: (1.9 + 1.6 + 1.63/ 3) = 1.71 L/ minute SOP #2: Defender 510 Pump Put the tube to the suction section of the defender pump and the other end to the cassette (inlet); put another tube on the outlet end of the sample pump and the other end to the cassette end of the sampler pump. Place the pump setting to continuous 1L/minute Flow: 0.996 Average: 1.0019L/minutes 1.5L/minute Flow: 1.5866 Average: 1.5506 2L/minute Flow: 2.0372 Average: 2.0355L/minutes