Uploaded by jkazuya33

ANACHEM-LEC-REVIEWER

advertisement
INTRO TO SPECTROSCOPY
Spectroscopy – study of interactions between
the matter and electromagnetic radiations.
Types:
Atomic Spectroscopy – concerned with the
interactions of electromagnetic radiations with
atoms.
Molecular Spectroscopy - concerned with the
interactions of electromagnetic radiations with
molecules.
Methods:
-
These bands are arranged in the
increasing order of wavelength.
Classification:
Line Spectra – spectrum obtained by interaction
of electromagnetic radiations with atoms.
-
Consist of sharp well-defined lines that
correspond to a definite frequency.
Band Spectra – spectrum obtained by
interaction of electromagnetic radiations with
molecules.
-
Non – Destructive
Consist of different bands just as visible
spectrum
Destructive
Electromagnetic Radiation – is an oscillating
electric and magnetic disturbance that spreads
as a wave through empty space, the vacuum.
Characteristics:
Wavelengths – distance between two crest or
troughs.
Frequency - number of waves which can pass
through a point in one second.
Wave Number – reciprocal of wavelength
-
Total number of waves which can pass
through a space of 1 cm
Continuous Spectrum – if one color merges into
another without a gap
Discontinuous Spectrum – if one color will not
merge into another
Absorption Spectrum - if
electromagnetic
radiations are passed through a substance, the
dark pattern of lines that are obtained
corresponding to wavelengths absorbed.
Emission Spectrum - if
electromagnetic
radiations are passed through a substance, the
pattern of lines recorded after the emission of
the absorbed wavelengths.
Amplitude – measure of electric or magnetic
field strength at a maximum point in the wave
Electromagnetic Spectrum – arrangement of
electromagnetic waves in the order of the
increasing
wavelengths
or
decreasing
frequencies.
Period – time in seconds it takes for successive
maxima or minima to pass a point in space
Absorption – light is absorbed by an atom, ion or
molecule taking it to a higher energy state
Dispersion of Light – when white light passes
through a prism, it is split up into seven colors
which correspond to definite wavelengths.
Absorption Process – light is absorbed by a
substance only when the energy corresponds to
some energy need or transition of a substance
Spectrum – series of color bands obtained by
dispersion of light.
Modes:
Velocity – distance travelled by a wave in 1 sec.
Rotational – occurs in the far IR and microwave
region
-
Changes in the energy of a molecule as it
rotates about a center of gravity
Vibrational - changes in the average separation
of nuclei
-
Occurs in the mid and far IR regions
-log (T) = abc
A = abc
-
Triplet state – spins are parallel
Emission – release of a photon by an atom, ion
or molecule, taking it to a lower energy level
Methods in which the stimulus is usually
heat or electrical energy
Chemiluminescence – excitation of analyte by a
chemical reaction
Photoluminescence – emission of photon is
measured after absorption
Could
be
fluorescence
phosphorescence
-log (P/P0) = abc
Relationship between A and T
Singlet State – state in which spin QNs are
opposite
-
log (P/P0) = abc = A
A= log T
Electrical
-
Formula Relationships:
or
-
-
Measuring absorbance
-
Fluorescence – initial absorption of a photon
followed by the emission of a second photon
-
Takes place rapidly and complete in
about 10 – 5 secs from the time of
excitation
-
Must be done at the wavelength if
maximum absorbance
This is because the point of maximum
response results to better sensitivity and
lower detection limits and reduces
possibility of committing errors during
measurement
Most absorbing species will only give a
linear response over a certain range
Errors that could occur:
Phosphorescence - like fluorescence but there is
a delay before emission. Delay can take a few
more seconds or hours
-
Transmittance – is the measured light that
passes through the solution (T = P/P0)
-
Absorbance – governed by Beer Lambert’s law
which states that the amount of light absorbed
by a solution is an exponential function of
concentration and path length of the solution.
Absorbance has direct relationship with
concentration
Transmittance is the one directly
measured
Some
instruments
give
both
transmittance
and
absorbance
information, but transmittance is the
preferred parameter
As the absorbance of the solution
increases, the transmittance decreases
At low concentration, a small change in
concentration can result in a large
change in %T
At high concentration changes %T are
very small
Solution: stay in the range of 80 -20 %T
to minimize measurement errors
Limitations of Beer Lambert’s Law
-
-
-
-
Chemical: occurs when the absorbing
species is involved in an equilibrium
equation
Instrumental: Strictly valid for purely
monochromatic radiation that is for
radiation consisting of only one
wavelength. Presence of stray radiation
that arises from imperfections within the
wavelength selector that allows
extraneous light to “leak” into the
instrument. Stray radiation adds an
additional contribution, P stray to the
radiant power reaching the detector.
For smaller concentrations of analyte, P
stray is significantly smaller that Po and P,
and the absorbance is unaffected. At
higher concentrations however, P stray is
no longer significantly smaller than P and
the absorbance is smaller than expected.
The result is a negative deviation of
Beer’s law
A = log (Po + P stray / P + P stray)
Emission – light produced is directly proportional
to the concentration of species being measured
-
C = kI, k (proportionality constant), I
(intensity of light), C (concentration)
Quantitative analysis of single analyte –
concentration of analyte is determined by
measuring the absorbance of the sample and
applying Beer’s law using any of the
standardization methods such as normal
calibration curves and method of standard
additions.
-
The absorbance of a solution at any
wavelength is the sum of absorbances of
all species in the solution.
reactants by preparing solutions containing
different mole fractions of one reactant.
-
Y = nL/nM = XL/XM = XL/1-XL
Job’s method
Mole ratio method – procedure of determining
the stoichiometry between two reactants by
preparing solutions containing different mole
ratios of two reactants
-
-
Absorbance is monitored at a
wavelength at which metal ligand
complex absorbs
A plot of absorbance as a function of
metal ligand to metal mole ratio has two
linear branches that intersect at a mole
ratio corresponding to the formula of
the complex
Slope Ratio method – procedure of determining
the stoichiometry between two reactants by
measuring the relative change in absorbance
under conditions when each reactant is a limiting
reagent
-
For weak complexes
Prepare two sets of solutions: series of
M solutions (varied amount of M with
excess of L) and series of L solutions
(varied amount of L with excess of M)
UV VIS Spectroscopy
The wavelength and amount of light that a
compound absorbs depends on its molecular
structure and the concentration of the substance
used. Concentration dependence follows the
Beer’s law:
A=eBc
A (absorbance) =log10 (P0/P)
Stoichiometry of Metal-Ligand Complex
Methods of determination:
Method of continuous variations – procedure of
determining the stoichiometry between two
e (molar absorbtivity with units of L mol-1 cm-1)
B (path length of the sample, cm)
c (concentration of compound in solution, mol/L)
Spectrophotometer - is a device used to
measure the transmittance or absorbance of a
sample as a function of the wavelength of the
electromagnetic
spectrum.
The
key
components
of
UV-visible
spectrophotometers are:
-
-
-
A source which generates a broad band
of electromagnetic radiation (UV and
visible)
A dispersion device (monochromator)
which selects a particular wavelength or
wavelengths
A sample cell
One or more detectors to measure the
intensity of radiation
Atomic Spectroscopy
Optical Spectrometry – elements present in a
sample are converted to gaseous atoms or
elementary ions by atomization process and the
UV-vis absorption, emission or fluorescence of
atomic species in the vapor is then measured
Mass Spectrometry – samples are also atomized
but the gaseous atoms are converted to positive
ions and separated according to their mass-tocharge ratios and quantitative data are obtained
by counting the separated ions
X-ray Spectrometry – does not require
atomization because most elements are largely
independent of their chemical composition in a
sample thus quantitative results are based on
direct measurement of the fluorescence,
absorption or emission spectrum of a sample.
Principle: Samples vaporized at 2000 – 8000 K
decompose into atoms and concentration of
these atoms in the vapor are measured by
emission or absorption of characteristic
wavelengths of radiation
-
In atomic absorption, atoms absorb part of the
light from the source and the remainder of the
light reaches the detector
Atomic emission comes from atoms that are in
an excited state because of the high thermal
energy of the flame
To observe atomic fluorescence, atoms are
excited by an external lamp or laser. An excited
atom can fall to a lower state and emit radiation.
Characteristics of Atomic absorption:
-
Absorption of sharp lines from hollow
cathode lamp
High sensitivity
Ability to distinguish one element from
the other in a complex sample
Ability to perform simultaneous multielement analysis
Easiness in the analysis of samples
Distinguishing Features:
-
-
-
Types:
-
Emission of thermally populated excited
state
Fluorescence following absorption of
laser radiation
-
Analysis is limited to the elements
(metals)
Sample separation schemes for atomic
spectroscopy usually place the metals in
water solution. It must have a means of
converting metal ions into free gas phase
ground state atoms (atomization) in
order to measure them
The sample container used is the source
of the thermal energy needed for the
conversion of ions in solution to atoms in
the gas phase (atomizer)
Spectral line sources are used as light
sources in atomic absorption instrument
and several atomic emission techniques
require no light source at all apart from
the thermal energy source
The wavelengths used for quantitation
are well known and do not require the
analyst to ever first measure the
absorption or emission spectra
Optical Spectra – consist of narrow spectral lines
since only electronic transition occurs
-
-
Types:
Emission Spectra – analyte atoms are excited by
external energy in the form of heat or electrical
energy
-
Transition to or from ground state is
called resonance transition and the
resulting spectral line is called resonance
line
Absorption Spectra – external source of
radiation impinges on the analyte vapor
-
-
Doppler Broadening – during atomization,
species may move towards or away from the
detector
The radiation is absorbed by the analyte
atoms and promote them to excited
states
-
Fluorescence Spectra – same external source as
absorption
-
-
Radiant power of fluorescence is
measured at right angles to the source
Resonance fluorescence happens when
atomic fluorescence measured at the
same wavelength as the source of
radiation
-
-
-
Atomic spectra have narrow lines and
anything that causes broadening of the
lines can cause problems
Narrow lines are highly desirable for
both atomic and emission spectra
because they reduce the possibility of
interference due to overlapping lines
Four Sources of Broadening:
Natural broadening due to uncertainty effect
Doppler shift in the resulting line occurs
Spectral lines that are 1 – 5 nm wide
result to 100 times magnification
Nothing can be done except to recognize
that it occurs
Causes:
Atomic Line width – the width of atomic lines is
quite important in atomic spectroscopy
-
The line width of the source must be
narrower than the line width of the
atomic vapor for Beer’s law to be
obeyed.
Line width is governed by quantum
mechanical effect called Heisenberg
uncertainty principle stating that the
shorter the lifetime of the excited state,
the more uncertain is its energy relative
to the ground state
Typical radiative lifetimes of the atoms
are on the order 10-8 s which lead to the
natural line widths on the order of 10-5
nm
When atom moves toward a photon
detector it emits radiation, the detector
sees wave crest more often and defects
radiation of higher frequency
When atom moves away from photon
detector and emits radiation, the
detector sees wave crest less frequent
and detects radiation of lower frequency
Pressure broadening - arises from the collision of
the sample atoms with other species causing
some energy to be exchanged
-
-
Amount of broadening increases with
the concentration of the collision
partners
Has greater effect a temperature
increase
Electric and magnetic effect
Types of Atomizers used for Absorption
Spectroscopy:
created a plume of a particulate and vaporized
sample is transported by an inert gas
Continuous atomizer – samples are introduced
in a continuous manner
Flame Atomization – utilize large flames as the
atomizer in which the sample is drawn into the
flame via vacuum mechanism. The light beam for
the absorption measurement is directed through
the width of the flame right to left or left to right
depending on the instrument orientation and
measured
-
-
Samples are frequently introduced into
plasmas or flames via nebulizer
producing a mist of spray
Samples maybe introduced via flow
injection or HPLC or separate converted
to vapor
Discrete Atomizer – samples are introduced
using a device such as syringe or autosampler
-
-
Methods for introduction of sample:
Pneumatic Nebulizer – the liquid sample is
drawn by a capillary tube by a high-pressure
stream of gas flowing around the tip of the tube
and the process is called aspiration
Ultrasonic Nebulizer – liquid sample is drawn by
a capillary tube by a high-pressure stream of gas
flowing around the tip of the tube and the
process is called aspiration the sample is pumped
onto the surface of a piezoelectric crystal. It
produces denser and more homogenous
aerosols compared to the pneumatic nebulizer
Electrothermal vaporizers – an evaporator in
which argon flows to carry the vaporized sample
into the atomizer
Hydride generators – use to carry highly toxic
species in low concentration levels into an
atomizer as a gas.
Direct sample insertion – sample is physically
placed in atomizer
Electrothermal vaporizers (flameless) – sample
is heated conductively on or in a graphite or
tantalum rod or boat and sample is carried by an
inert gas
Arc, spark or laser ablation – sample discharge
interacts with the surface of a solid sample and
-
Flame atomizer will usually have a long
narrow burner (premix burner) that
serves as sample path
Nebulizer controls sample flow,
producing a mist
Mixing chamber assures the sample
mixes with the oxidant and fuel prior to
entry to the flame
Process:
Process occurring in the flame:
-
combination works best with a given
element
Air-acetylene flames are the most
commonly used
Radiation Source:
Hollow Cathode Lamp – source produces
emissions specific for the element used to
construct the cathode
-
Cathode must be capable of conducting
a current for it to work
Will only produce emission lines for the
cathode element
Limitation: metal is too volatile, good
cathode can’t be produced, metal may
not be a good conductor
Mechanism:
Premix burner – components of the flame (fuel,
oxidant and sample solution) are premixed as
they take a common path to the flame
-
Fuel and oxidant originate from
pressurized sources and their flow the
burner is controlled at an optimum rate
by flow control mechanisms that are
part of the overall instrument
Laminar flow burner – provides a relatively quiet
flame and long path length
Fuel and Oxidant – all flames require these
-
-
-
To achieve desirable sensitivity, hotter
flame is desirable for quantitative
analysis to sufficiently atomize sample
Aside from flame temperature, burning
velocity should also be considered
because high burning velocity decreases
the completeness of the atomization
and therefore actually lowering the
sensitivity
The choice is made based on which
flame temperature-burning velocity
Electrodeless Discharge Lamp – has no anode or
cathode
-
Small, sealed quartz tube containing the
metal or metal salt and some argon at
low pressure wrapped with a coil for the
purpose of creating a radio frequency
field
Radiation Source Modulation – used to
eliminate interferences caused by the emission
of radiation by the flame
Chopper – used to provide signal modulation - in
conjunction with a lock – in amplifier
-
-
-
It is not practical to have separate cells,
so the light is simply split, with half being
sent around the atomization source
This reduces some noise from the
atomization source and accounts for
instrumental variations
Types: rotating vane (b), rotating disk
(a), oscillating tuning fork (c) where they
cause periodic interruption of light beam
Optical Path – light source then flames then
monochromator then detector. Monochromator
is placed after flame because flame is necessarily
in an open area and is surrounded with room
light. This ensures that light from the room will
not leak to the detector and the flame emissions
will be eliminated
-
Maybe single or double beam
Single Beam
Double Beam – primary function is to eliminate
problems due to source drift and fluctuation
-
Side advantage is that source warm up
time is eliminated since changes in
intensity during warm up are
immediately compensated and thus very
rapid changeover of lamps in automated
instruments is possible
Monochromator and Detector – high resolution,
holographic grating is used to resolve the lines. It
is not designed to be used in scan mode
-
Typical detector is photomultiplier tube
Method of background correction since
signal modulation simply accounts for
flame flicker and instrument variations
but not accounts for background
absorption of emission
Interferences in AAS
Spectral Interferences – due to substances in the
flame that absorb the same wavelength as the
analyte, causing the absorbance measured to be
high
-
Absorption due to the presence of light
absorbing molecules in the flame and
light dimming due to the presence of
small particles in the flame are much
more common spectral interferences
these are called background absorption
Methods of Correction:
Zeeman Effect – magnetic field is applied to the
sample. When an atomic vapor is exposed to a
strong magnetic field, there is a splitting of the
atom’s electronic energy levels. This essentially
moves the absorption away from emission lines.
Using this method, we can measure at a fixed
and narrow wavelength. At regular intervals, we
simply move our sample component out of the
way. This allows us to directly measure the
background
Deuterium correction – continuous source
correction method wherein light from both AA
source lamp and D2 lamp alternately pass
through the sample. Because spectral width is
significantly larger than AA source line, the D2
lamp send a much broader band of light to the
sample
-
HC (narrow band width) D2 (larger band
width)
Limitations: either under correction or
over correction can occur based on the
sample, background may vary around
line, composition of background can
differ based on position in flame
(requires good alignment of HC and D2
lamp), D2 output is not very good at
greater than 350 nm
Chemical Interferences – results of problems
with the sample matrix
-
Correction: matrix matching (matching
the matrix of all of standards as these
solutions are prepared with that of the
sample so that the same matrix effect is
seen in all solutions measured and
therefore becomes inconsequential,
standard addition method
Practical Matters and Applications
Slits and Spectral lines
Linear and non-linear standard curves
Hollow cathode lamp current
Lamp alignment
Aspiration rate
Burner head position
Fuel and oxidant source and flow rates
Download