Original Research n Technical Developments Note: This copy is for your personal non-commercial use only. To order presentation-ready copies for distribution to your colleagues or clients, contact us at www.rsna.org/rsnarights. Arterial, Venous, and Cerebrospinal Fluid Flow: Simultaneous Assessment with Bayesian Multipoint Velocity-encoded MR Imaging1 Verena Knobloch, MSc Christian Binter, MSc Vartan Kurtcuoglu, PhD Sebastian Kozerke, PhD Purpose: To measure arterial, venous, and cerebrospinal fluid (CSF) velocities simultaneously by using Bayesian multipoint velocity-encoded magnetic resonance (MR) imaging and to compare interacquisition reproducibility relative to that of standard phase-contrast MR imaging for sequential measurements of arterial, venous, and CSF velocities. Materials and Methods: This study was approved by the local ethics committee, and informed consent was obtained from all subjects. Simultaneous measurement of blood and CSF flow was performed at the C1-C2 level in 10 healthy subjects (mean age, 24.4 years 6 2.7; five men, five women) by using accelerated Bayesian multipoint velocity-encoded MR imaging. Data were compared with those obtained from two separate conventional phase-contrast MR imaging acquisitions, one optimized for arterial and venous blood flow (velocity encoding range, 650 cm/sec) and the other optimized for CSF flow (velocity encoding range, 610 cm/ sec), with an imaging time of approximately 2 minutes each. Data acquisition was repeated six times. Intraclass correlation coefficient (ICC) and linear regression were used to quantify interacquisition reproducibility. Results: There was no significant difference in arterial blood flow measured with Bayesian multipoint velocity-encoded MR imaging and that measured with phase-contrast MR imaging (mean ICC, 0.96 6 0.03 vs 0.97 6 0.02, respectively). Likewise, there was no significant difference between CSF flow measured with Bayesian multipoint velocity-encoded MR imaging and that measured with phase-contrast MR imaging (mean ICC, 0.97 6 0.02 vs 0.96 6 0.05, respectively). For venous blood flow, the ICC with Bayesian multipoint MR imaging was significantly larger than that with conventional phase-contrast MR imaging (mean, 0.75 6 0.23 vs 0.65 6 0.26, respectively; P = .016). Conclusion: Bayesian multipoint velocity-encoded MR imaging allows for simultaneous assessment of fast and slow flows in arterial, venous, and CSF lumina in a single acquisition. It eliminates the need for vessel-dependent adjustment of the velocity-encoding range, as required for conventional sequential phase-contrast MR imaging measurements. 1 From the Institute for Biomedical Engineering, University and ETH Zurich, Gloriastrasse 35, 8092 Zurich, Switzerland (V. Knobloch, C.B., S.K.); the Interface Group, Institute of Physiology, University of Zurich, Zurich, Switzerland (V. Kurtcuoglu); and Division of Imaging Sciences and Biomedical Engineering, King’s College London, London, England. Received April 9, 2013; revision requested May 28; revision received July 16; accepted July 31; final version accepted August 12. Supported by the Swiss National Science Foundation, SmartShunt, The Hydrocephalus Project (K-32K1_120531); and Swiss National Science Foundation (project CR23I3_144218). Address correspondence to S.K. (e-mail: kozerke@biomed.ee.ethz.ch). RSNA, 2013 q RSNA, 2013 q 566 radiology.rsna.org n Radiology: Volume 270: Number 2—February 2014 TECHNICAL DEVELOPMENTS: Simultaneous Assessment of Arterial, Venous, and Cerebrospinal Fluid Flow Q uantification of blood supply and drainage of the cranium and measurement of cerebrospinal fluid (CSF) flow in the spinal canal are the basis for the noninvasive assessment of intracranial compliance, pressure (1,2), and cranial transfer functions (3–5). These parameters are of fundamental value in advancing the understanding and improving treatment of cerebrospinal dynamic disorders such as normal-pressure hydrocephalus (3) or Chiari malformations (6). Phase-contrast magnetic resonance (MR) imaging allows for quantification of blood and CSF velocities. The velocity-to-noise ratio of the measurement is, however, dependent on the maximum velocity-encoding value of the bipolar velocity-encoding gradients used in phase-contrast MR imaging. In practice, velocity encoding is defined on the basis of estimation of the highest velocities to be expected in the volume of interest. Velocities with an absolute value higher than the velocity-encoding value will result in aliasing and require dedicated postprocessing to unwrap the velocity data. In contrast, velocity-encoding values that are too high relative to actual velocities reduce the velocityto-noise ratio proportionally and, hence, lead to noisy measurements. Advances in Knowledge nn Bayesian multipoint velocityencoded MR imaging allows for simultaneous assessment of fast and slow flows in arterial, venous, and cerebrospinal fluid (CSF) lumina by using a single protocol and provides equal or improved interacquisition reproducibility relative to that with conventional phase-contrast MR imaging for arterial, venous, and CSF flow. nn Bayesian multipoint velocityencoded MR imaging eliminates the need for vessel-dependent adjustment of the velocity sensitivity, as required in conventional phase-contrast MR imaging. Accordingly, the issue of measurement accuracy arises if both fast- and slow-flowing fluids are present in the volume of interest. Although phasecontrast MR imaging has been extensively validated by using experiments in phantoms (7–9) and in terms of reproducibility for arterial measurements (10,11), few data are available for domains in which large differences in flow velocity are present. To address the limited velocity-tonoise ratio during the slow-flow phase of the cardiac cycle, three-point, dual velocity-encoding, and variable velocity-encoding approaches have been proposed (12–14). Despite these advances, assessment of intracranial blood and CSF flow still requires separate measurements with individual velocity-encoding values to match the large differences in fluid velocity. Initial studies have demonstrated simultaneous assessment by using a dual velocity-encoding approach (15). However, a general approach is still lacking. In particular, venous drainage pathways show strong intersubject variability, with a resulting large range of velocities (16,17). Accordingly, setting optimal velocity-encoding values remains difficult. Even though the assessment of venous dynamics has been based on jugular pulsatility only (2,9), the influence of different venous drainage pathways with different pulsatility has been found to have a substantial effect (18,19). Therefore, vessels with lower velocities cannot be neglected in venous blood flow measurements. To address the issue of simultaneously encoding fluid flows with large differences in velocities, a multipoint velocity-encoding scheme used in combination with Bayesian processing has recently been described (20). It has been demonstrated that such a scheme allows for simultaneous encoding of a wide range of velocities with an optimal velocity-to-noise ratio in a single imaging session. However, given the additional encoding steps involved in Bayesian multipoint measurements, the method must be combined with data undersampling Radiology: Volume 270: Number 2—February 2014 n radiology.rsna.org Knobloch et al techniques to achieve clinically feasible imaging times (21). The objective of our study was to measure arterial, venous, and CSF velocities simultaneously by using Bayesian multipoint velocity-encoded MR imaging and to compare interacquisition reproducibility relative to that of standard phase-contrast MR imaging for sequential measurements of arterial, venous, and CSF velocities. Materials and Methods This prospective study was approved by institutional and local ethics committees (Cantonal Ethics Commission Zurich) and performed during November 2012. Written informed consent was obtained from all study participants. Ten healthy volunteers (mean age 6 standard deviation, 24.4 years 6 2.7; five men, five women) were enrolled in the study. Any subjects with neurologic disorders, including previous traumatic brain injury, were excluded by using a questionnaire. Data Acquisition Measurements were performed by using a 3.0-T system (Ingenia; Philips Healthcare, Best, the Netherlands) with a standard 12-element head coil array. Data were acquired by following the protocol shown in Figure 1. CSF flow acquisitions in the aqueduct were part of the protocol, but they were not further considered in this study. Published online before print 10.1148/radiol.13130840 Content codes: Radiology 2014; 270:566–573 Abbreviations: CSF = cerebrospinal fluid ICC = intraclass correlation coefficient Author contributions: Guarantors of integrity of entire study, V. Knobloch, S.K.; study concepts/study design or data acquisition or data analysis/interpretation, all authors; manuscript drafting or manuscript revision for important intellectual content, all authors; manuscript final version approval, all authors; literature research, V. Knobloch, S.K.; clinical studies, V. Knobloch; statistical analysis, V. Knobloch, S.K.; and manuscript editing, all authors Conflicts of interest are listed at the end of this article. 567 TECHNICAL DEVELOPMENTS: Simultaneous Assessment of Arterial, Venous, and Cerebrospinal Fluid Flow The preparation for section planning included survey imaging and time-offlight angiography. A transverse imaging section was prescribed at the C1-C2 level for the quantification of blood and CSF flow. Bayesian multipoint velocity-encoded MR imaging data with six velocity-encoding points and one reference point were acquired with fivefold undersampling in k-space and time with principal component analysis reconstruction. Taking into account the acquisition of 15 training profiles for k-t principal component analysis reconstruction, a net acceleration factor of 3.5 was achieved. Accordingly, the total imaging time for a Bayesian multipoint encoding measurement was identical to that for single conventional two-point phase-contrast MR imaging. As reference, two conventional phase-contrast MR imaging data sets were collected in separate measurements before and after Bayesian multipoint acquisition with velocity encodings of 50 and 10 cm/sec, respectively. To test measurement reproducibility, subjects were asked to stand up for a few minutes after the first set of three measurements. The protocol was repeated once after repositioning. Heart rates before and after repositioning did not change substantially. By using prospective electrocardiographic triggering, measurement times were dependent on heart rate; ranges for imaging time are given in Figure 1. Imaging parameters were as follows: gradient-echo imaging; field of view, 180 3 145 mm; matrix size, 224 3 180; section thickness, 5 mm; repetition time, 11 msec; and echo time, 7.2 msec. Velocity encodings were interleaved over cardiac cycles, and three k-space profiles were acquired consecutively, resulting in a temporal resolution of three times the repetition time (ie, 33 msec). Depending on heart rate, 20–29 cardiac phases were recorded by using prospective electrocardiographic triggering. Data Processing Data processing was implemented by using software (Matlab; Mathworks, 568 Knobloch et al Figure 1 Figure 1: Measurement protocol. After three repetitions of arterial (A), venous (V), and CSF velocity acquisition interleaved with two repetitions of aqueductal CSF flow, volunteers were asked to stand up and walk around. Protocol was restarted after repositioning. Imaging times for phasecontrast MR imaging (PC-MRI) and multipoint acquisitions were dependent on heart rate. RARE = rapid acquisition with relaxation enhancement, TOF = time of flight, venc = velocity encoding. Figure 2 Figure 2: Data processing of data acquired at C1-C2 level. Left: Phase-contrast MR imaging (PC-MRI) and multipoint data were reconstructed individually and segmented in the same way. k-t = k-t space and time, kv = velocity k-space, PCA = principal component analysis. Right: Segmentation template threshold was set during first imaging, and vessels were numbered according to vessel scheme shown. Arteries are in red and veins are in blue. Threshold level was adapted at subsequent imaging to ensure identical vessel sizes. Natick, Mass). Data were analyzed by a reader (V. Knobloch) with 3 years of experience in MR flow imaging. An overview of processing steps is given in Figure 2. The phase-contrast MR imaging data were reconstructed by using standard radiology.rsna.org n Radiology: Volume 270: Number 2—February 2014 TECHNICAL DEVELOPMENTS: Simultaneous Assessment of Arterial, Venous, and Cerebrospinal Fluid Flow Knobloch et al Figure 3 Figure 3: MR images allow visual comparison of velocities (v) at different velocity scaling. Top: Images are scaled to 670 cm/sec to show arterial and jugular vein velocities. Insets are enlargements of internal carotid artery and jugular vein. Bottom: Images are scaled to 610 cm/sec to highlight CSF velocities in spinal canal. Insets are enlargements of spinal canal. PC-MRI = phasecontrast MR imaging, venc = velocity encoding. Fourier transformation and phase difference processing. Phase aliasing was corrected by using the Matlab unwrap algorithm applied along the temporal dimension. Multipoint data were reconstructed by using k-t principal component analysis (21), and velocities were reconstructed by using the Bayesian reconstruction algorithm (20). Each data set was segmented individually to account for volunteer motion between the two acquisitions. Segmentation of multipoint and phase-contrast MR imaging data was based on the modified angiographic image, A, as follows: A = σ [ I (t )] ⋅ ∑ v ⋅ ∑ mag, T T where s[I(t)] is the standard deviation over time of the complex image I, v the velocity, mag the image magnitude, and T the number of cardiac phases per cycle. In the first multipoint acquisition of each set (Fig 1), arteries and veins were segmented by manually setting the threshold, and the number scheme shown in Figure 2 was manually assigned as the original vessel mask. The segmentation was automatically adapted to all multipoint and phase-contrast MR images (velocity encoding = 50 cm/ sec) for each vessel, as follows: The angiographic image was masked with a dilated vessel mask of the original image. A threshold was computed to result in the same vessel size as that on the original mask. The spinal canal was manually segmented for each acquisition. Velocity profiles and stroke volumes were computed for each vessel and each acquisition. Stroke volume of the spinal canal was computed as the sum of forward and backward stroke volumes. As a measure of reproducibility, intraclass correlation coefficient (ICC) 3,1 (22) was computed for each vessel. The three repeat acquisitions with Bayesian multipoint MR imaging and Radiology: Volume 270: Number 2—February 2014 n radiology.rsna.org with standard phase-contrast MR imaging were correlated by using multiple linear regression. Velocities for each cardiac phase and in each vessel were correlated among and across the two sets of three measurements each. Deviation of the linear regression line from the identity line between measurements is given in degrees and by the intercept (ie, the distance of the fitted line from zero). The standard error of the estimate was computed as the mean distance of the data relative to the line obtained by using linear regression. Normalization relative to the mean vessel velocity was performed. Results Velocity images from phase-contrast MR imaging and multipoint data at the time point of maximal CSF flow in the spinal canal acquired at 188 msec after the R wave are shown in Figure 3. The 569 TECHNICAL DEVELOPMENTS: Simultaneous Assessment of Arterial, Venous, and Cerebrospinal Fluid Flow Figure 4 Figure 4: Graphs of ICC as a function of peak velocity (vpeak) for arteries, veins, and CSF (ie, spinal canal). Means 6 standard deviations for all values are shown in bar graphs on right. PC-MRI = phase-contrast MR imaging, venc = velocity encoding. top row is scaled to 70 cm/sec to visualize arterial velocities, and the lower row is scaled to 10 cm/sec to visualize CSF velocities. Velocities obtained with the multipoint sequence showed good velocity-to-noise ratios for both arterial and CSF flow. In contrast, CSF velocities acquired with phase-contrast MR imaging exhibited amplified noise when velocity encoding was 50 cm/ sec. Arterial and venous velocities were wrapped multiple times at phasecontrast MR imaging with velocity encoding of 10 cm/sec, as expected, and required additional unwrapping in postprocessing. Figure 4 summarizes reproducibility given as the ICCs for all vessels of all subjects as a function of peak velocity. 570 In arteries, no significant difference was found between Bayesian multipoint MR imaging and phase-contrast MR imaging with a velocity encoding of 50 cm/ sec (mean ICC, 0.96 6 0.03 vs 0.97 6 0.02, respectively), whereas significantly higher ICCs were seen for multipoint imaging in veins (mean ICC, 0.75 6 0.23 vs 0.65 6 0.26; P = .016). Flow of CSF in the spinal canal was detected equally well with multipoint MR imaging and phase-contrast MR imaging with a velocity encoding of 10 cm/sec (mean ICC, 0.97 6 0.02 vs 0.96 6 0.05, respectively). In general, it was noted that multipoint acquisitions resulted in improved performance, particularly for low peak velocities as found in venous and CSF compartments. Knobloch et al Results of linear regression analysis of velocities determined in the three repeat acquisitions are reported as the difference in slope, intercept, and standard error of the estimate. Arterial velocities were found to be close to identity for both multipoint and phasecontrast MR imaging with a velocity encoding of 50 cm/sec (mean difference in slope: 3.2° 6 2.3 vs 2.7° 6 2.0, respectively; mean intercept: 1.8 cm/sec 6 0.9 vs 1.7 cm/sec 6 1.4, respectively). The mean standard error of the estimate was 5% 6 2 for both multipoint and phase-contrast MR imaging. For venous velocities, the mean difference in slope was 13.2° 6 13.1 for multipoint MR imaging and 12.9° 6 13.1 for phase-contrast MR imaging with a velocity encoding of 50 cm/sec; the mean intercept was 4.2 cm/sec 6 5.9 and 3.0 cm/sec 6 3.6, respectively. There was a significant reduction in the standard error of the estimate for multipoint MR imaging compared with phase-contrast MR imaging (mean, 8.1% 6 4.5 vs 20% 6 15.2, respectively; P , .001). Linear regression of CSF velocities resulted in a mean differnce in slope of 2.6° 6 1.7 for multipoint MR imaging and 7.3° 6 5.8 and 3.7° 6 4.6 for phase-contrast MR imaging with velocity encodings of 50 and 10 cm/sec, respectively. The mean intercept of CSF was 0.1 cm/ sec 6 0.1 with multipoint MR imaging and 0.3 cm/sec 6 0.2 and 0.1 cm/ sec 6 0.1 for phase-contrast MR imaging with a velocity encodings of 50 and 10 cm/sec, respectively. Significant differences were found with regard to the standard error of the estimate between multipoint and phase-contrast MR imaging with a velocity encoding of 10 cm/sec (mean, 17.0% 6 6.0 vs 11.8% 6 4.2, respectively; P , .001) and between multipoint and phase-contrast MR imaging with a velocity encoding of 50 cm/sec (mean, 17.0% 6 6.0 vs 37.5% 6 14.7, respectively; P , .001). Stroke volumes for all arteries and veins are shown in Figure 5. There were no significant differences between phase-contrast MR imaging and multipoint acquisition. The intrasubject standard deviation was considerably radiology.rsna.org n Radiology: Volume 270: Number 2—February 2014 TECHNICAL DEVELOPMENTS: Simultaneous Assessment of Arterial, Venous, and Cerebrospinal Fluid Flow smaller than the intersubject standard deviation in all vessels (P , .001). Both inter- and intrasubject standard deviations were higher for veins than for arteries (P , .001). The correlation matrix between all vessels and CSF in the spinal canal is shown in Figure 6. Correlation between arteries was stronger than intervenous correlation (P , .001). CSF velocities correlated mostly positively with those of the venous system. Knobloch et al Figure 5 Discussion In this study, we demonstrated the simultaneous assessment of arterial, venous, and CSF velocities with use of Bayesian multipoint velocity encoding and showed it could provide values similar to those obtained with conventional phase-contrast MR imaging. Overall, Bayesian multipoint velocity encoding was found to provide similar or improved interacquisition reproducibility relative to that of sequential phase-contrast MR imaging, with the added advantage of requiring only half the total imaging time. The technique facilitates flow measurements in clinically feasible times. Although conventional phase-contrast MR imaging requires phase unwrapping if the magnitude of actual velocities exceeds velocity encoding, data from multipoint encoding is inherently free of aliasing as long as actual velocities are lower than or equal to the maximum of the entire encoding range, which was 104 cm/sec in our study. Accordingly, a single protocol is sufficient to map a large dynamic range, simplifying imaging setup and minimizing planning overhead. The ICC as a measure of reproducibility was found to be high for arterial velocities, with a slight reduction at lower peak velocities. For venous velocities, single values of ICC were low at higher peak velocities. This is related to reduced pulsatility of venous flow and can also be associated with physiologic factors. It has been shown that venous flow is dependent on posture (23) and respiration (24). Because posture was supine during the entire examination, there was no direct influence expected Figure 5: Bar charts show stroke volumes (SV) for all arteries, veins, and CSF (ie, spinal canal). Numbers in parentheses are numbers of subjects in whom the respective vessel could be identified. PC-MRI = phasecontrast MR imaging, s = standard deviation. Figure 6 Figure 6: Correlation matrix of velocities in all arteries, veins, and CSF. from this end. However, changes in respiratory patterns may have occurred, possibly influencing venous flow. Respiratory monitoring was not performed, and, thus, data on respiratory Radiology: Volume 270: Number 2—February 2014 n radiology.rsna.org variations during imaging were not recorded in our study. Stroke volumes were similar for vessels on the left versus right side of the neck. Stroke volumes in the left 571 TECHNICAL DEVELOPMENTS: Simultaneous Assessment of Arterial, Venous, and Cerebrospinal Fluid Flow and right internal carotid arteries (1,2) were higher than those in vertebral arteries (3,4). Venous drainage was mainly through the left and right jugular veins, but other pathways were also identified. Correlation matrix analysis revealed differences in pulsatility. The highest correlation was found between intracranial arteries. In the venous system, the following regions of higher correlation were identified: Jugular veins exhibited high correlation between the left and right sides, whereas vertebral veins and spinal column venous drainage correlated positively with that of arteries. Another cluster of strong correlation was identified for veins positioned posteriorly to the spinal canal. CSF flow was found to correlate positively with that of the jugular veins while showing negative correlation with arterial velocities. This is in line with previous findings (18). However, to confirm the venous correlation dependencies, studies in larger cohorts are warranted. Interindividual variations in venous pulsatility patterns are associated with certain disorders (25). Differences in anatomy and pulsation can be linked to increased outflow resistances, as found in chronic cerebrospinal venous insufficiency (26). The simultaneous assessment of all venous outflow pathways, together with arterial and CSF flow, is also considered important for the noninvasive assessment of intracranial pressure (2). Because pulsatility can vary in amplitude and phase, the acquisition of flow waveforms in all veins is important for a complete description of intracranial pulsation. This work was a proof-of-concept and feasibility study of measuring blood and CSF velocities simultaneously by using Bayesian multipoint velocity-encoded MR imaging. This initial study was limited in that only healthy young adults were investigated. Accordingly, further studies are warranted to translate the approach to relevant patient populations. In summary, Bayesian multipoint velocity-encoded MR imaging allows 572 for simultaneous assessment of arterial, venous, and CSF velocities with equal or improved interacquisition reproducibility relative to that of sequential phase-contrast MR imaging acquisitions and holds potential for studies in patients. Disclosures of Conflicts of Interest: V. Knobloch No relevant conflicts of interest to disclose. C.B. No relevant conflicts of interest to disclose. V. Kurtcuoglu No relevant conflicts of interest to disclose. S.K. No relevant conflicts of interest to disclose. References 1.Alperin N, Mazda M, Lichtor T, Lee SH. From cerebrospinal fluid pulsation to noninvasive intracranial compliance and pressure measured by MRI flow studies. Curr Med Imaging Rev 2006;2(1):117–129. 2. Alperin NJ, Lee SH, Loth F, Raksin PB, Lichtor T. MR-intracranial pressure (ICP): a method to measure intracranial elastance and pressure noninvasively by means of MR imaging—baboon and human study. Radiology 2000;217(3):877–885. 3.Balédent O, Gondry-Jouet C, Meyer ME, et al. Relationship between cerebrospinal fluid and blood dynamics in healthy volunteers and patients with communicating hydrocephalus. Invest Radiol 2004;39(1):45– 55. 4.Schmid Daners M, Knobloch V, Soellinger M, et al. Age-specific characteristics and coupling of cerebral arterial inflow and cerebrospinal fluid dynamics. PLoS ONE 2012;7(5): e37502. 5.Tain RW, Alperin N. Noninvasive intracranial compliance from MRI-based measurements of transcranial blood and CSF flows: indirect versus direct approach. IEEE Trans Biomed Eng 2009;56(3):544– 551. 6. Alperin N, Sivaramakrishnan A, Lichtor T. Magnetic resonance imaging–based measurements of cerebrospinal fluid and blood flow as indicators of intracranial compliance in patients with Chiari malformation. J Neurosurg 2005;103(1):46–52. 7. Greil G, Geva T, Maier SE, Powell AJ. Effect of acquisition parameters on the accuracy of velocity encoded cine magnetic resonance imaging blood flow measurements. J Magn Reson Imaging 2002;15(1):47–54. 8. Lotz J, Meier C, Leppert A, Galanski M. Cardiovascular flow measurement with phasecontrast MR imaging: basic facts and im- Knobloch et al plementation. RadioGraphics 2002;22(3): 651–671. 9. Wåhlin A, Ambarki K, Hauksson J, Birgander R, Malm J, Eklund A. Phase contrast MRI quantification of pulsatile volumes of brain arteries, veins, and cerebrospinal fluids compartments: repeatability and physiological interactions. J Magn Reson Imaging 2012;35(5):1055–1062. 10.Spilt A, Van den Boom R, Kamper AM, Blauw GJ, Bollen EL, van Buchem MA. MR assessment of cerebral vascular response: a comparison of two methods. J Magn Reson Imaging 2002;16(5):610–616. 11. Koerte I, Haberl C, Schmidt M, et al. Inter- and intra-rater reliability of blood and cerebrospinal fluid flow quantification by phase-contrast MRI. J Magn Reson Imaging 2013;38(3):655–662. 12.Ringgaard S, Oyre SA, Pedersen EM. Arterial MR imaging phase-contrast flow measurement: improvements with varying velocity sensitivity during cardiac cycle. Radiology 2004;232(1):289–294. 13.Buonocore MH. Blood flow measurement using variable velocity encoding in the RR interval. Magn Reson Med 1993;29(6): 790–795. 14.Lee AT, Pike GB, Pelc NJ. Three-point phase-contrast velocity measurements with increased velocity-to-noise ratio. Magn Reson Med 1995;33(1):122–126. 15.Dhoondia HA, Alperin N. Improved MRintracranial pressure (MR-ICP) measurement using a new data acquisition technique [abstr]. In: Proceedings of the eleventh meeting of the International Society for Magnetic Resonance in Medicine. Berkeley, Calif: International Society for Magnetic Resonance in Medicine, 2003; 793. 16.Doepp F, Schreiber SJ, von Münster T, Rademacher J, Klingebiel R, Valdueza JM. How does the blood leave the brain? A systematic ultrasound analysis of cerebral venous drainage patterns. Neuroradiology 2004; 46(7):565–570. 17.Stoquart-Elsankari S, Lehmann P, Villette A, et al. A phase-contrast MRI study of physiologic cerebral venous flow. J Cereb Blood Flow Metab 2009;29(6):1208–1215. 18. Marshall I, MacCormick I, Sellar R, Whittle I. Assessment of factors affecting MRI measurement of intracranial volume changes and elastance index. Br J Neurosurg 2008;22(3):389–397. 19.Pomschar A, Koerte I, Lee S, et al. MRI evidence for altered venous drainage and radiology.rsna.org n Radiology: Volume 270: Number 2—February 2014 TECHNICAL DEVELOPMENTS: Simultaneous Assessment of Arterial, Venous, and Cerebrospinal Fluid Flow intracranial compliance in mild traumatic brain injury. PLoS ONE 2013;8(2):e55447. 20. Binter C, Knobloch V, Manka R, Sigfridsson A, Kozerke S. Bayesian multipoint velocity encoding for concurrent flow and turbulence mapping. Magn Reson Med 2013;69(5): 1337–1345. 21. Pedersen H, Kozerke S, Ringgaard S, Nehrke K, Kim WY. k-t PCA: temporally constrained k-t BLAST reconstruction using principal component analysis. Magn Reson Med 2009;62(3):706–716. 22.Shrout PE, Fleiss JL. Intraclass correla tions: uses in assessing rater reliability. Psychol Bull 1979;86(2):420–428. 23.Alperin N, Lee SH, Sivaramakrishnan A, Hushek SG. Quantifying the effect of posture on intracranial physiology in humans by MRI flow studies. J Magn Reson Imaging 2005;22(5):591–596. 24.Zamboni P, Menegatti E, Pomidori L, et al. Does thoracic pump influence the cerebral venous return? J Appl Physiol 2012;112(5):904–910. Radiology: Volume 270: Number 2—February 2014 n radiology.rsna.org Knobloch et al 25.Beggs CB. Venous hemodynamics in neurological disorders: an analytical review with hydrodynamic analysis. BMC Med 2013;11:142. 26. Zamboni P, Galeotti R, Menegatti E, et al. Chronic cerebrospinal venous insufficiency in patients with multiple sclerosis. J Neurol Neurosurg Psychiatry 2009;80(4):392– 399. 573