Agilent G3388B Leak Detector Operation Manual Agilent Technologies Notices © Agilent Technologies, Inc. 2012 No part of this manual may be reproduced in any form or by any means (including electronic storage and retrieval or translation into a foreign language) without prior agreement and written consent from Agilent Technologies, Inc. as governed by United States and international copyright laws. Manual Part Number G3388-90005 Edition First edition, June 2012 Printed in Japan Agilent Technologies, Inc. 2850 Centerville Road Wilmington, DE 19808-1610 USA Warranty The material contained in this document is provided “as is,” and is subject to being changed, without notice, in future editions. Further, to the maximum extent permitted by applicable law, Agilent disclaims all warranties, either express or implied, with regard to this manual and any information contained herein, including but not limited to the implied warranties of merchantability and fitness for a particular purpose. Agilent shall not be liable for errors or for incidental or consequential damages in connection with the furnishing, use, or performance of this document or of any information contained herein. Should Agilent and the user have a separate written agreement with warranty terms covering the material in this document that conflict with these terms, the warranty terms in the separate agreement shall control. Safety notices CAUTION A CAUTION notice denotes a hazard. It calls attention to an operating procedure, practice, or the like that, if not correctly performed or adhered to, could result in damage to the product or loss of important data. Do not proceed beyond a CAUTION notice until the indicated conditions are fully understood and met. WA R N I N G A WARNING notice denotes a hazard. It calls attention to an operating procedure, practice, or the like that, if not correctly performed or adhered to, could result in personal injury or death. Do not proceed beyond a WARNING notice until the indicated conditions are fully understood and met. Agilent G3388B Leak Detector Operation Manual The G3388B Leak Detector Introduction 4 Specifications 7 Operation 9 General information 9 Displays 10 Detecting leaks 12 Setting options 13 Maintenance 16 Agilent Technologies 3 Introduction The Agilent Technologies G3388B Leak Detector is a portable, highly sensitive instrument designed to detect helium and hydrogen gas leaks in gas chromatographs (GCs) and their accessories, such as purge and trap systems. The detector finds leaks by comparing thermal conductivity between the ambient atmosphere and a target gas. The detector indicates leaks through a dual system that consists of an indicating LCD and an audible alarm. The detector is more reliable and convenient to use than soap solutions and is capable of detecting even the smallest, hard-to-reach gas leaks. Figure 1 and Figure 2 show the controls and other parts of the meter. Reference port LED indicator LCD display Charging terminal FUNC/POWER key ZERO/ENTER key Sampled gas inlet Sample probe Figure 1 Front view of the meter 4 G3388B Leak Detector Reference port Strap eyelet (Strap not included.) USB cable (for charging battery) Figure 2 Back view of the meter, and USB cable Use the Agilent Leak Detector only for detecting gases. Never allow liquids or particles to enter the sample probe or reference port. The G3388B is not a quantitative device. It is designed to detect leaks in gas line connections commonly associated with laboratory equipment. G3388B Leak Detector 5 CAUTION WA R N I N G 6 • Do not expose the detector to shock, wet environments, or solvents. • Do not use the detector at temperatures outside its specified operating range. • Do not open the detector’s body. • Do not use the detector to detect corrosive gases. • Do not block the reference port (fresh air inlet) at the top of the detector. • The detector’s sensitivity will be adversely affected if it is used in an area containing the leaking gas. • Detector sensitivity can be affected by wind. • Do not use the detector’s micro USB cable for other devices. • During use, keep the detector temperature stable. Do not use the detector while charging. Allow the detector temperature to stabilize (cool) for 30 minutes after charging. • Do not use the detector in areas at risk for fire, explosion, or large leaks of flammable gas. • When detecting small leaks of flammable gases, make sure sufficient ventilation exists to prevent a fire hazard. • Do not use the detector in environments that may be potentially contaminated with dust or combustible fumes. • Do not use the detector as safety equipment. • This leak detector must not be used in Classified (Hazardous) locations. Use of this leak detector in a hazardous location could lead to personal injury or death. • The customer may not replace the battery. The customer must return the leak detector to an Authorized Agilent Repair Center for repair. G3388B Leak Detector Specifications Detector: Thermistor-actuated thermal conductivity cell Thermal conductivity of He, CO2, Ar, Ne, and so forth (at 0 °C, gases: nonflammable, noncorrosive atmosphere) Sensitivity: Selectable in two ranges: Std range: Helium 0.005 mL/min in atmosphere Hi range: Helium 0.0005 mL/min in atmosphere Battery: Rechargeable lithium-ion battery Charging time < 2.5 h Continuous operation: 5 hours (with a new, fully charged battery, standard range, LED off, Buzzer off, Backlight off) Display: LCD display with 8 bar graphs 8 LED lamps Recharger: USB cable, part number G3388-80000 Operating temperature 10 to 40 °C (noncondensing) range: Overall size: 50 mm (W) × 19.5 mm (D) × 111 mm (H) Net weight: 95 grams (approximate) G3388B Leak Detector 7 The sensitivity depends on the relative thermal conductivities of the target gas and the ambient atmosphere (typically air). The larger the difference, the greater the sensitivity. See Table 1. Table 1 Thermal conductivities of common gases at 0 °C, 1 atm Gas Thermal conductivity (mW/m•k) Hydrogen 168.2 Helium 142.2 Neon 46.5 Methane 30.2 Air* 24.1 Nitrogen* 24.0 Ethane 18.0 Ethylene 16.4 Argon 16.3 Carbon dioxide 14.5 Krypton 8.7 Xenon 5.2 * Not detectable 8 G3388B Leak Detector Operation General information NOTE Before first use and after a long period of disuse, fully charge the battery. See “Charging the battery” on page 16. To determine whether a GC or a GC accessory is leaking gas, place the Agilent Leak Detector’s sample probe near the suspected leak. Make sure that the sample probe remains open during leak detection. Be careful not to place the probe’s tip against the connection or fitting that you are checking. Protect the detector from vibration while it is operating. NOTE Be aware that the detector will give a positive response to water vapor, for example to skin vapor and to fittings that were recently leak checked with a soap solution. Periodically replace the filter (part number G3388-80001) that is attached at the inlet of the sample probe. See “Replacing the filters” on page 17. G3388B Leak Detector 9 Displays Warmup display: At startup, the detector warms up for 25 s when set for standard range, or 90 s when set for high range sensitivity. Press [ZERO/ENTER] to skip warmup. After warmup, the detector is ready for use. Standard and high range displays: Range Battery indicator R_Std Leak indicator (LED) Leak indicator (LED) Leak indicator (LCD) Range R_Std Standard sensitivity range R_Hi High sensitivity range (10 times more sensitive than standard range) Battery indicator The Battery indicator shows the battery charge level in 3 levels. 10 3 bars Ample charge. Greater than 50% charge remaining 2 bars Low charge. Battery charge about 20–30%. 1 bar Very low charge. Battery charge required. About 10 minutes of use remaining. G3388B Leak Detector Leak indicator (LCD display) When the leak gas is detected, displays a bar graph. The number of bars indicates the relative size of the gas leak (maximum 8 bars). Zero point display When measuring clean ambient air, the first bar alternates between ■ and □ when the detector becomes stable. Leak display ■: Indicates a gas leak for a gas with a higher thermal conductivity than air. □: Indicates a gas leak for a gas with a lower thermal conductivity than air. Leak indicator (LED display) When the leak gas is detected, the LEDs light from the bottom left to the top, then from the bottom right to the top (maximum 8 bars). The LED indicators can be turned off if desired. Zero point display When measuring clean ambient air, the green ■ or the red ■ LED blinks when the detector becomes stable. Leak display ■: Red indicates a gas leak for a gas with a higher thermal conductivity than air. ■: Green indicates a gas leak for a gas with a lower thermal conductivity than air. G3388B Leak Detector 11 Detecting leaks To use the detector: 1 Press and hold [FUNC/POWER] to turn on the detector. You will hear the detector begin to sample air. Wait until the warmup completes. 2 If desired, set the sensitivity range or other settings. See “Setting options” on page 13. 3 Check the stability at the zero point (while sampling clean air). When stable, the leak indicator in the LCD display will alternate between ■ and □. • If the detector indicates a leak (more than one bar), press [ZERO/ENTER] to reset the zero balance. • Temperature or humidity fluctuations can impact the zero point. Out of zero point ZERO adjust R_Std R_Hi or R_Std ZERO ENTER 4 Place the sample probe near the gas connection to check for a leak. Do not press the probe tip against the fittings, tubing, and so forth. • One bar (or LED) may not indicate a leak, especially in high sensitivity mode. • Two or more bars indicates a leak. • An audible alarm sounds when 3 or more bars display (with Buzzer set to On). Turn off the detector when not in use. Press FUNC/POWER and hold for 3 seconds. The detector can be set to turn off automatically after 10 minutes. To enable Auto Power Off, see “Setting options” on page 13. 12 G3388B Leak Detector Setting options Press [FUNC/POWER] to cycle through the setting options or to toggle a selected function on or off. Press [ZERO/ENTER] to select a function for editing. See Figure 3 and Figure 4. Table 2 Item Function and options Range Select standard (Std) or high (Hi) range sensitivity. Standard is the factory default. Buzzer Enable/disable audible alarm for leaks that show 3 or more bars/LEDs. LED Turn LED indicators off/on. Turn off to conserve battery charge. Backlight Turn LCD backlight on or off. Turn off to conserve battery charge. A.P.Off When on, the detector turns off after 10 min of inactivity. Turn on to conserve battery charge. G3388B Leak Detector 13 Standard Display (Std range) R_Std Standard Display (Hi range) R_Hi or FUNC POWER a) Setting the Range Param.Set Range Std Buzzer ON ON LED ZERO ENTER Param.Set Range Std Buzzer ON LED ON Param.Set Range Hi Buzzer ON ON LED ZERO ENTER Param.Set Range Std Buzzer ON ON LED FUNC POWER To “A” on Figure 4 Returns to the Standard Display ZERO ENTER Returns to the Standard Display Param.Set Range Std Buzzer ON ON LED ZERO ENTER Returns to the Standard Display ZERO ENTER Returns to the Standard Display ZERO ENTER Returns to the Standard Display ZERO ENTER Returns to the Standard Display FUNC POWER FUNC POWER c) Setting the LED lamp ZERO ENTER FUNC POWER FUNC POWER b) Setting the Buzzer Param.Set Range Std Buzzer ON ON LED Param.Set Range Std Buzzer OFF ON LED ZERO ENTER Param.Set Range Std Buzzer ON ON LED FUNC POWER Param.Set Range Std Buzzer ON OFF LED From “B” on Figure 4 Figure 3 Setting options 14 G3388B Leak Detector A d) Setting the Backlight B Param.Set Backlight ON A.P.Off OFF ZERO ENTER Param.Set Backlight ON A.P.Off OFF ZERO ENTER Returns to the Standard Display ZERO ENTER Returns to the Standard Display ZERO ENTER Returns to the Standard Display ZERO ENTER Returns to the Standard Display FUNC POWER FUNC POWER e) Setting the Automatic Power Off Param.Set Backlight ON A.P.Off OFF Param.Set Backlight OFF A.P.Off ON ZERO ENTER Param.Set Backlight ON A.P.Off OFF FUNC POWER FUNC POWER Param.Set Backlight ON A.P.Off ON Figure 4 Setting options, continued G3388B Leak Detector 15 Maintenance Charging the battery Charge the battery when the LCD display is blank or when the battery indicator shows less than 3 bars. Low battery (20-30% charge) R_Std Very low battery (≤10 min.) To charge the battery: 1 Turn off the detector. 2 Connect the detector to a PC USB port using the USB to micro USB cable (part number G3388-80000). 3 Allow to charge for approximately 2.5 hours. Do not use the detector with the cable attached. The detector and battery may become warm while charging. After charging the battery, allow the detector to stabilize (cool) for 30 minutes before use. CAUTION 16 Because the battery gets warm while charging, allow the detector to stabilize (cool) for 30 minutes before use. The detector should be thermally stable for best performance. G3388B Leak Detector Replacing the filters The detector includes filters on the sample probe and reference port to protect against particulate contamination. A decrease in sensitivity can indicate a clogged filter. Replace the filters as described below. • Replacement sample probe filter: G3388-80001 • Replacement reference port filter: G3388-80002 To replace the sample probe filter 1 Turn off the detector. 2 Unscrew the probe halves and disassemble the probe. 3 Use long-nosed pliers to remove the old filter. 4 Install the filter into the probe base as shown below. Sampling probe Probe tip Filter Probe base G3388B Leak Detector 17 5 Install the new filter over the PTFE tubing as shown below. Filter PTFE tubing 6 Reassemble the probe halves. To replace the reference port filter To replace the reference port filter, remove the old filter from the detector body, then insert a new filter. Reference port filter 18 G3388B Leak Detector Agilent Technologies G3388-90005