Design and Simulation of a Pneumatic, Stored-energy, Hybrid Orthosis for Gait Restoration William K. Durfee Adam Rivard Department of Mechanical Engineering, University of Minnesota, Minneapolis, MN 55439 The reciprocating gait orthosis 共RGO兲 links opposite joints so that extension of the hip on one side leads to flexion on the contralateral side 关30兴. Gharooni et al. proposed that stored spring energy and limb-segment potential energy could be used to replace stimulation of the hip flexors or withdrawal reflex 关31兴. In their spring brake orthosis 共SBO兲, excess quadriceps energy is stored in a mechanical spring that resists knee extension. Spring release causes knee flexion, which due to inertial properties of the leg forces the hip to flex. Our system goes further to provide decoupled hip extension and flexion that increases gait-assist performance. Energy Storing Orthosis Loss of mobility due to lower limb paralysis is a common result of thoracic level spinal cord injury. Functional electrical stimulation (FES) can restore primitive gait in the vicinity of a wheelchair by using electrical stimulation to generate muscle contractions. A new concept for FES-assisted gait is presented that combines electrical stimulation with an orthosis that contains a fluid power system to store and transfer energy during the gait cycle. The energy storage orthosis (ESO) can be driven through a complete gait cycle using only stimulation of the quadriceps muscles. The conceptual design of the ESO was completed and implemented in a dynamic simulation model and in a benchtop prototype for engineering measurements. No studies were conducted with human subjects. The results demonstrate the potential of the ESO concept for a feasible gait-assist system and the validity of the simulation model as a means for designing the system. 关DOI: 10.1115/1.2050652兴 Introduction The inability to walk because of lower limb paralysis is a common result of a thoracic-level spinal cord injury 共SCI兲. Functional electrical stimulation 共FES兲, which uses electrical stimulation of motor nerves to trigger muscle contractions, is one means of restoring rudimentary standing and gait for limited mobility in the vicinity of a wheelchair to some individuals with SCI 关1–7兴. The user must have good trunk control and a strong upper body because considerable effort is required from the arms engaging parallel bars, a walker, or crutches for support. Despite these restrictions, successful FES users are able to ambulate for hundreds of meters with many years of use from their system 关8,9兴. Two limitations of FES-aided gait systems are rapid muscle fatigue and the inability to precisely control joint torques, which leads to erratic stepping trajectories 关10兴. Hybrid systems that combine electrical stimulation with a lower limb orthotic brace have been developed to address these problems 关11–26兴. Our lab developed the controlled brake orthosis 共CBO兲, a hybrid FESaided standing and gait system that contains computer-regulated friction brakes at the knee and hip to lock the joints during stance phase and control motion during swing phase 关27,28兴. The goal of our present research is to develop an FES-aided gait system that requires only one channel of surface muscle stimulation. This can be achieved by combining FES, a mechanical orthosis, and energy storage. Excess energy generated by electrical stimulation of one muscle is harvested, stored, and transferred to other joints that cannot be conveniently driven by direct muscle stimulation. The concept of using orthotics to store energy is not new. For example, Van den Bogert used elastic exotendons for gait assist 关29兴. Greene and Granat utilized a cam-slider mechanism to transfer energy from the knee to the ankle to assist dorsiflexion 关24兴. Contributed by the Bioengineering Division for publication in the JOURNAL OF BIOMECHANICAL ENGINEERING. Manuscript received by the Bioengineering Division April 7, 2005; revision received July 14, 2005. Associate Editor: Mary Frecker. 1014 / Vol. 127, NOVEMBER 2005 The energy storing orthosis 共ESO兲 hybrid FES gait system uses stimulated muscle power to not only move the limb but also to push on the orthosis, storing energy in the process. The stored energy is piped to another joint and released to drive joint motion without having to stimulate additional muscles. The following sections of the paper describe the engineering design for an ESO concept that utilizes a fluid power system to capture excess energy from the quadriceps that is then released to drive hip and knee joints. A mathematical model was developed, and a bench top prototype was fabricated and mounted onto a simulated leg for conducting baseline engineering tests and for validating the simulation model. No experiments or device evaluations were done on humans. Design Requirements Joint Range of Motion. The system must allow flexion and extension at the hip and knee. Only sagittal plane motions are considered, and the ankle is assumed to be locked with an ankle foot orthosis. Prior work with the CBO 关27兴 showed that the orthosis must accommodate −10– 25 deg of hip flexion and 0 – 60 deg of knee flexion. Muscle Stimulation. The system must stimulate only the quadriceps muscle. Quadriceps stimulation is attractive for several reasons. First, the quadriceps 共technically four muscles兲 are large and relatively large amounts of power can be generated through stimulation. Second, the quadriceps drive knee extension, which undergoes a large excursion during normal gait, simplifying the task of designing an energy capture system. Third, the quadriceps are simple to activate with surface electrodes; users can apply and remove quadriceps electrodes relatively easily. Although completely implanted stimulation systems are the ultimate goal, easyto-use surface systems have a role as a bridge to implanted systems and for those users who do not want surgery. Available Energy. All energy to drive the leg comes from the stimulated quadriceps. Additional energy for forward progression can come from the upper body because the system will be used with a walker. The available energy from the quadriceps for one extension cycle moving was estimated from previous work by calculating the area under a typical torque-angle curve from 60 deg of flexion to full extension, resulting in 31.4 J 关10兴. To be conservative and to maximize the total number of steps before fatigue, the system was designed to extract 14 J from the quadriceps per step cycle. Efficiency. Stimulated muscles behave as nonlinear, timevarying actuators with significant power and energy limitations 关27,32兴. Thus, the mechanical system must minimize energy loss during operation. Of primary concern is the process of storing, channeling, and discharging the energy from the quadriceps. Any energy dissipation during this process will increase the amount of energy required from the quadriceps, therefore, increasing their rate of fatigue and limiting the total number of steps that can be taken. Copyright © 2005 by ASME Transactions of the ASME Fig. 1 Stored energy orthosis concept: „a… biased equilibrium position, „b… stimulation of quadriceps and storage of excess energy, and „c… discharge of stored energy Size, Weight, Don/Doff Time. Minimum size, weight, and time to don and doff the orthosis is essential for the success of a future assistive technology product. The purpose of this study was to demonstrate technical feasibility of the concept; therefore, size and weight were not considered. System Description and Physical Prototype The ESO concept is shown in Fig. 1. Elastic energy-storage elements on the orthosis hip and knee joints hold the leg in a flexed equilibrium position 共Fig. 1共a兲兲. Stimulation of the quadriceps extends the knee, placing excess energy in both the equilibrium spring and an energy transfer element 共Fig. 1共b兲兲. The stored energy is transferred to the hip where it is discharged and used to extend the hip against its equilibrium spring, and to aid in forward progression 共Fig. 1共c兲兲. A new step is initiated by releasing the hip and knee joints from the straight leg position to the flexed position shown in Fig. 1共a兲. The realization of the ESO concept is shown in Fig. 2. Gas springs crossing the hip and knee joints are the flexed equilibrium energy-storage elements. Their energy is used locally and not transferred to other joints. A pneumatic fluid power system implements the energy-storage and transfer system. An air cylinder at the knee acts as a compressor while a matching cylinder at the hip acts as a joint actuator. An accumulator stores compressed air energy, and three-way-controlled valves selectively connect the piston ports to the accumulator and to the atmosphere. A bench model physical prototype of the ESO is shown in Fig. 3. The ESO components were mounted on wooden links weighted with steel blocks to model the weight, center of mass, and inertia of a 180 cm, 76 Kg male 关33兴. The gas springs 共McMaster Carr 6465K11, 6465K12兲 had 90 lb reducible force with stroke lengths of 2.36 in. 共hip兲 and 3.94 in. 共knee兲. The springs attached to the links with miniature nylon ball sockets. The pneumatic cylinders 共Bimba D-90366-A-2.5兲 were single-acting, nonspring return, rear-pivot mount cylinders, with a 7 / 8 in. bore and 2.5 in. stroke length. Single-acting cylinders are required because they only allow air flow to and from the accumulator in a single port and minimize overall length of the unit. The pistons were pivot mounted to prevent side loads on the shaft. Two-position, threeway, 12 V dc solenoid-operated valves 共McMaster Carr 61245K2兲 with a maximum operating pressure of 120 psi were used for control. The accumulator was made from 1 / 16 in. i.d., 1 / 8 in. o.d. Teflon® tubing with a rated burst strength of 440 psi. Thickwalled Teflon tubing is stiff, and the accumulator did not visibly expand under pressure. Gas springs were chosen over mechanical springs for the bias elements because of their excellent force-to-weight and storedenergy-to-weight ratio and because they produce an almost constant force throughout their entire stroke. A pneumatic fluid power system was chosen for energy storage and transfer because large amounts of energy can be stored as highly compressed air in a Journal of Biomechanical Engineering Fig. 2 Realization of ESO. Gas springs at the hip and knee bias joints in flexion. Pneumatic cylinders at the hip and knee convert mechanical to fluid power. Compressed air stored in accumulator tubing that couples hip and knee cylinders. small volume and with almost no weight. The amount of stored potential energy in compressed air is E = PV ln共P / Po兲 where E = energy, P = pressure of the compressed air, V = volume of compressed air, and Po = atmospheric pressure. For example, pressurized air at 250 psi contains 79 J / in3, and air at 1500 psi contains 777 J / in3. Pressurized air can be stored in lightweight plastic or metal tubing, allowing for a high energy-to-weight ratio in comparison to gas springs or traditional coil springs. Energy transfer and conversion is simple because hoses can transfer compressed air to remote locations and across rotary joints, while air cylinders can convert between mechanical translation and fluid power domains. Three considerations were involved in air-cylinder size and placement. First, the cylinders must be small bore and located close to the body to avoid unnecessary weight and bulk. Smaller bore, higher pressure cylinders would be optimal, but would result in an accumulator whose volume was so small that parasitic effects would dominate. Second, the cylinders must be aligned so that they are parallel to the thigh when fully retracted to produce the highest mechanical advantage when the maximum cylinder force is exerted. This minimizes losses in the transmission during operation. Third, because the two cylinders are equal bore and have the same volume, they must have equal stroke length over the motion of the respective joint. The placement distance from the central axis of the leg can then be estimated based on a relationship between cylinder length, stroke length, and angular disNOVEMBER 2005, Vol. 127 / 1015 Fig. 4 ADAMS dynamic simulation model. Head-arm-trunk loads were simulated with a controlled, moving “floor” that applied equal and opposite loads to the bottom of the leg linkage. Fig. 3 Bench model physical prototype. The gas springs are on the back side. placement of the joint. To prevent energy loss from the system due to release of pressurized air to atmosphere, the cylinders must be at zero pressure at full extension so that all energy goes into the system and no pressurized air is exhausted into the atmosphere. The pneumatic components and operating conditions, including piston diameter, maximum pressure, and accumulator volume, were selected using an iterative design process. The resulting maximum system pressure was calculated to be 120 psi with an accumulator volume of 0.15 cu. in. Simulation Model An analytic simulation model of the ESO system was created and realized using the ADAMS 共MSC Software, Santa Ana, CA兲 dynamic systems simulation software to produce a fully coupled model of the leg operating in a gravity field 共Fig. 4兲. Although the model greatly simplified that actual system, it included sufficient detail for its primary purpose of helping select component properties and placements for the physical prototype. The human leg was modeled as a two-link, frictionless joint pendulum with geometric and inertial properties matching that of an average male. Quadriceps stimulation was approximated as a constant torque source acting on the knee joint with a magnitude set sufficiently high to move the leg and still be reasonable for the torque applied by a real muscle. For simplicity, the action of the rectus femoris about the hip was neglected. Joint locks were modeled as constraints. The gas springs and cylinders were modeled as prismatic 1016 / Vol. 127, NOVEMBER 2005 joints with controlled force profiles. Gas spring force was modeled as a constant pressure force in parallel with a damping element so that the output force changed with velocity, as it does in a real gas spring. The cylinder was modeled as a frictionless air chamber with a force that accounted for a varying accumulator pressure, the extension and retraction of the cylinder, and the on/ off timing of the three-way valves. Airflow dynamics and characteristics and energy losses of the air through the system were not incorporated into the force model. Beginning with heel strike and throughout forward progression, additional loads are placed on the foot due to the weight and accelerations of the head, arms, and trunk 共HAT兲 during gait. To avoid modeling the balance stability of upright humans, the thigh was pinned to ground at the hip in the ADAMS model; therefore, it was not possible to directly simulate the HAT loads. Instead, HAT loads were implemented through an artificial moving floor that applied loads to the bottom of the linkage equivalent to those applied by the HAT, resulting in equivalent effects on the hip and knee. The floor was modeled as a square block coupled to ground with two frictionless translation joints and a link to generate twodimensional floor movement in the sagittal plane. A positive vertical force was applied to the floor to simulate the equal and opposite force that the HAT normally applies to the hip joint 关33兴. Measurements were performed on components of the physical prototype to verify whether the physical model matched the model. The static force-displacement curve for the gas springs and for the piston compressing air into the accumulator were measured by pressing the shaft against a calibrated scale and recording force and position at several points. Friction was estimated by measuring the force of an uncharged gas spring and an unpressurized cylinder against the scale while compressing at ⬃2 in./ s. Trajectories of the prototype leg as it moved from full extension to Transactions of the ASME Fig. 5 Simulation model: hip and knee trajectories for one gait cycle the flexed equilibrium position when released from a straight configuration were measured through frame-by-frame analysis of a video recording. Results A simulation was run of the ADAMS model to show how the leg responds. Figure 5 shows the change in hip and knee joint angles throughout a simulation cycle. The simulation cycle is one step in a gait cycle, but with artificial time intervals added to the simulation to allow for adjustment of model parameters, such as air-line solenoids opening and closing, and for simpler visualization of the cycle, resulting in a 15 s simulation time. Simulation begins with the leg straight and vertical. In the first 0.9 s, the gas springs force the hip and knee to their equilibrium positions of 25 and 60 deg of flexion. The knee then fully extends over the time interval from 5.5 to 6.6 s, resisting against the pneumatic system. Over the time interval from 10 to 11.25 s, the pneumatic system discharges, causing the hip to extend to −10 deg. The simulation ends after the leg is brought to equilibrium from 11.25 to 11.8 s. Figure 6 shows the force and displacement for the hip and knee air cylinders. Not shown are the additional simulation outputs of cylinder pressure and gas spring force and displacement. The electrically stimulated quadriceps must provide enough energy to fuel the operation of the system. The simulation model was used to calculate how the 14 J extracted from the quadriceps during each step was allocated during a cycle which, in turn, was used to select ESO components. Figure 7 shows the start and finish positions of the leg during knee extension and the corre- sponding torque necessary for the movement. The required energy from the quadriceps was estimated at 8.9 J by calculating the area under the curve corresponding to the movement. Subtracting 8.9 J from the initial 14 J leaves 5.1 J that can be stored in the pneumatic accumulator for driving hip extension later in the gait cycle. Driving the hip from 25 deg of flexion to 10 deg of extension stores energy in the hip gas spring. Gravity provides an assist in the first part of the motion but resists the motion once the hip passes zero degrees. Analysis of the hip motion showed that 1.4 J of the 5.1 J available was needed to move the leg against gravity and the gas spring, leaving 3.7 J to be used for actively driving hip extension to assist forward progression and increase gait speed. Measurements of the physical gas springs and air cylinders showed that ⬃0.25 J will be lost to friction during one gait cycle, 2% of the available 14 J. In the real system, there will be additional losses from the orthosis joints and fluid drag. Figure 8 shows the measured force-displacement curves for the hip and knee air cylinders. Superimposed are two curves from the simulation model. The dashed line accounts for the extra 0.052 cu. in. of dead volume that is between the cylinder and the valves at each end of the accumulator in the prototype. The dead volume lowers the force in the system and is a loss of pressurized air and, therefore, wasted energy that is never stored in the accumulator nor transferred to the hip. Figure 9 shows the time course of the hip and knee joint angles as the limbs of the physical prototype moved into their equilibrium position. The equivalent output from the simulation model is Fig. 6 Simulation model: force „top… and displacement „bottom… for hip „left… and knee „right… air cylinders for one gait cycle. Gas springs force leg to equilibrium during t = 0 – 0.9 s; knee extends during t = 5.5– 6.6 s; pneumatic system forces hip extension during t = 10.0– 11.25 s; leg returns to equilibrium during t = 11.25– 11.8 s. Journal of Biomechanical Engineering NOVEMBER 2005, Vol. 127 / 1017 Fig. 7 Simulation model of knee extending at constant angular velocity: „a… t = 2.0 s; 60 deg knee flexion, „b… t = 4.0 s; knee fully extended, „c… knee torque versus time. Required energy is 8.9 J. superimposed. The difference between the two curves is because the model assumes the force in the gas springs is independent of stroke length, whereas the force in the real spring varies by 10% over its stroke. Discussion and Conclusions The ESO appears to be a viable concept for a FES-aided gait system. Storing energy with gas springs and storing and transferring energy with a pneumatic fluid power system was shown to work in the dynamic system model and the physical prototype. The simulation model was essential for system design and was exercised extensively to determine system component specifica- tions and mounting positions. The evaluation experiments demonstrated that the simulation model could predict the performance of the physical system with sufficient accuracy. Significant technical issues must be addressed for the next design iteration. A rigorous energy loss analysis using a refined model must be conducted that includes more accurate musculoskeletal dynamics and head losses in the fluid power system. The system must then be optimized to minimize these losses because they directly draw on the total available muscle energy. The dead volume in the fluid power system must be eliminated along with all other losses of compressed air due to valve switching. Finally, the next design iteration must address total system size and weight Fig. 8 Force-displacement properties of hip and knee air cylinders. Measured „markers… and two simulation results „lines…. The dashed line accounts for dead volume between the cylinder and the valve. Fig. 9 Prototype „markers… and simulation „line…: hip „a… and knee „b… knee angle trajectories as joints move from full extension to equilibrium 1018 / Vol. 127, NOVEMBER 2005 Transactions of the ASME as well as body attachment points for the orthosis. Once these challenges are met, a study can be conducted to evaluate the efficacy of the ESO concept using human subjects. References 关1兴 Marsolais, E. B., and Kobetic, R., 1987, “Functional Electrical Stimulation for Walking in Paraplegia,” J. Bone Jt. Surg., Am. Vol., 69A, pp. 728–733. 关2兴 Kobetic, R., Triolo, R. J., and Marsolais, E. B., 1997, “Muscle Selection and Walking Performance of Multichannel FES Systems for Ambulation in Paraplegia,” IEEE Trans. Rehabil. Eng., 5共1兲, pp. 23–29. 关3兴 Cybulski, G., Penn, R., and Jaeger, R., 1984, “Lower Extremity Functional Neuromuscular Stimulation in Cases of Spinal Cord Injury,” Neurosurgery, 15共1兲, pp. 132–146. 关4兴 Graupe, D., and Kohn, K., 1994, Functional Electrical Stimulation for Ambulation by Paraplegics, Krieger, Melbourne, FL. 关5兴 Kralj, A., and Bajd, T., 1989, Functional Electrical Stimulation: Standing and Walking After Spinal Cord Injury, CRC Press, Boca Raton. 关6兴 Marsolais, E. B., and Kobetic, R., 1983, “Functional Walking in Paralyzed Patients by Means of Electrical Stimulation,” Clin. Orthop. Relat. Res., 175, pp. 30–36. 关7兴 Creasey, G. H., Ho, C. H., Triolo, R. J., Gater, D. R., DiMarco, A. F., Bogie, K. M., and Keith, M. W., 2004, “Clinical Applications of Electrical Stimulation After Spinal Cord Injury,” J. Spinal Cord Med., 27共4兲, pp. 365–375. 关8兴 Kobetic, R., Triolo, R. J., Uhlir, J. P., Bieri, C., Wibowo, M., Polando, G., Marsolais, E. B., Davis, J. A., Jr., and Ferguson, K. A., 1999, “Implanted Functional Electrical Stimulation System for Mobility in Paraplegia: A Follow-Up Case Report,” IEEE Trans. Rehabil. Eng., 7共4兲, pp. 390–398. 关9兴 Kralj, A., Bajd, T., and Turk, R., 1988, “Enhancement of Gait Restoration in Spinal Injured Patients by Functional Electrical Stimulation,” Clin. Orthop. Relat. Res., 233, pp. 34–43. 关10兴 Hausdorff, J. M., and Durfee, W. K., 1991, “Open-Lop Position Control of the Knee Joint Using Electrical Stimulation of the Quadriceps and Hamstrings,” Med. Biol. Eng. Comput., 29, pp. 269–280. 关11兴 Hirokawa, S., Grimm, M., Le, T., Solomonow, M., Baratta, R., Shoji, H., and D’Ambrosia, R., 1990, “Energy Consumption in Paraplegic Ambulation Using the Reciprocating Gait Orthosis and Electric Stimulation of the Thigh Muscles,” Arch. Phys. Med. Rehabil., 71共9兲, pp. 687–694. 关12兴 Petrofsky, J., and Smith, J., 1991, “Physiologic Costs of Computer-Controlled Walking in Persons With Paraplegia Using a Reciprocating-Gait Orthosis,” Arch. Phys. Med. Rehabil., 72共11兲, pp. 890–896. 关13兴 Solomonow, M., Aguilar, E, Reisin, E., Baratta, R., Best, R., Coetzee, T., and D’Ambrosia, R., 1997, “Reciprocating Gait Orthosis Powered With Electrical Muscle Stimulation 共RGO II兲. Part I: Performance Evaluation of 70 Paraplegic Patients,” Orthopedics, 20共4兲, pp. 315–324. 关14兴 Solomonow, M., Baratta, R., Hirokawa, S., Rightor, N., Walker, W., Beaudelte, P., Shoji, H., and D’Ambrosia, R., 1989, “The RGO Generation II: Muscle Stimulation Powered Orthosis as a Practical Walking System for Thoracic Paraplegics,” Orthopedics, 12共10兲, pp. 1309–1315. 关15兴 Isakov, E., Douglas, R., and Berns, P., 1992, “Ambulation Using the Reciprocating Gait Orthosis and Functional Electrical Stimulation,” Paraplegia, 30, Journal of Biomechanical Engineering pp. 239–245. 关16兴 McClelland, M., Andrews, B., Patrick, J., Freeman, P., and el Masri, W., 1987, “Augmentation of the Oswestry Parawalker Orthosis by Means of Surface Electrical Stimulation: Gait Analysis of Three Patients,” Paraplegia, 25, pp. 32–38. 关17兴 Nene, A., and Jennings, S., 1989, “Hybrid Paraplegic Locomotion With the Parawalker Using Intramuscular Stimulation: A Single Subject Study,” Paraplegia, 27, pp. 125–132. 关18兴 Andrews, B., Baxendale, R., Barnett, R., Phillips, G., Yamazaki, T., Paul, J., and Freeman, P., 1988, “Hybrid FES Orthosis Incorporating Closed Loop Control and Sensory Feedback,” J. Biomed. Eng., 10, pp. 189–195. 关19兴 Popovic, D., Tomovic, R., and Schwirtlich, L., 1989, “Hybrid Assistive System: The Motor Neuroprosthesis,” IEEE Trans. Biomed. Eng., 36共7兲, pp. 729– 737. 关20兴 Popovic, D., Schwirtlich, L., and Radosavljevic, S., 1990, “Powered Hybrid Assistive System,” Advances in External Control of Human Extremities, D. Popovic, ed., Nauka, Belgrade, pp. 177–186. 关21兴 Popovic, D., Schwirtlich, L., and Radosavljevic, S., 1990, “Powered Hybrid Assistive System,” Advance External Contr. Human Extremities, D. Popovic, ed., Nauka, Belgrade, pp. 177–186. 关22兴 Marsolais, E. B., Kobetic, R., Polando, G., Ferguson, K., Tashman, S., Gaudio, R., Nandurkar, S., and Lehneis, H. R., 2000, “The Case Western Reserve University Hybrid Gait Orthosis,” J. Spinal Cord Med., 23共2兲, pp. 100–108. 关23兴 Ferguson, K. A., Polando, G., Kobetic, R., Triolo, R. J., and Marsolais, E. B., 1999, “Walking With a Hybrid Orthosis System,” Spinal Cord, 37共11兲, pp. 800–804. 关24兴 Greene, P. J., and Granat, M. H., 2003, “A Knee and Ankle Flexing Hybrid Orthosis for Paraplegic Ambulation,” Med. Eng. Phys., 25共7兲, pp. 539–545. 关25兴 Yang, L., Granat, M. H., Paul, J. P., Condie, D. N., and Rowley, D. I., 1997, “Further Development of Hybrid Functional Electrical Stimulation Orthoses,” Artif. Organs, 21共3兲, pp. 183–187. 关26兴 Kobetic, R., Marsolais, E. B., Triolo, R. J., Davy, D. T., Gaudio, R., and Tashman, S., 2003, “Development of a Hybrid Gait Orthosis: A Case Report,” J. Spinal Cord Med., 26共3兲, pp. 254–258. 关27兴 Goldfarb, M., and Durfee, W. K., 1996, “Design of a Controlled-Brake Orthosis for FES-Aided Gait,” IEEE Trans. Rehabil. Eng., 4共1兲, pp. 13–24. 关28兴 Goldfarb, M., Korkowski, K., Harrold, B., and Durfee, W., 2003, “Preliminary Evaluation of a Controlled-Brake Orthosis for FES-Aided Gait,” IEEE Trans. Neural Syst. Rehabil. Eng., 11共3兲, pp. 241–248. 关29兴 van den Bogert, A. J., 2003, “Exotendons for Assistance of Human Locomotion,” BioMedical Engineering OnLine, 2共17兲. 关30兴 Jefferson, R., and Whittle, M., 1990, “Performance of Three Walking Orthoses for the Paralysed: A Case Study Using Gait Analysis,” Prosthet. Orthot Int., 14, pp. 103–110. 关31兴 Gharooni, S., Heller, B., and Tokhi, M. O., 2001, “A New Hybrid Spring Brake Orthosis for Controlling Hip and Knee Flexion in the Swing Phase,” IEEE Trans. Neural Syst. Rehabil. Eng., 9共1兲, pp. 106–107. 关32兴 Durfee, W. K., and Palmer, K. I., 1994, “Estimation of Force-Activation, Force-Length, and Force-Velocity Properties in Isolated, Electrically Stimulated Muscle,” IEEE Trans. Biomed. Eng., 41共3兲, pp. 205–216. 关33兴 Winter, D., 1979, Biomechanics of Human Movement, Wiley, New York. NOVEMBER 2005, Vol. 127 / 1019