Lorad FFDM QC Procedures for Medical Physicist Tao Wu, Ph.D. Hologic, Inc. Lorad Selenia FFDM Tests Following Lorad QC Manual • • • • • • • Collimation Assessment Artifact Evaluation System Resolution Phantom Image Quality Evaluation Automatic Exposure Control Function Performance Automatic Exposure Control Reproducibility Signal-To-Noise and Contrast-To-Noise Measurements • Softcopy Workstation QC Acquisition Workstation Software OS Login Medical Physicist username: phys password: phys Detector Warm-Up After logging in OS, a dialog showing detector warm-up status appears If the system has been running - Dismiss If the system has been shut down - Wait for 30-40 min (or dismiss it and start with tests in which detector is not involved, e.g. HVL, tube output) Application Login Username:Andrea Password: Andrea Turn Off Auto-decompression Send Images to Output Devices Send Images to Output Devices Send Images to Output Devices Non-Imaging Mode Create a New Patient Account Select “No Output” Create a New Patient Account Create a New Patient Account Selenia QC Tests Tests Following Lorad QC Manual • • • • • • Collimation Assessment Artifact Evaluation System Resolution Phantom Image Quality Evaluation Automatic Exposure Control Function Performance Signal-To-Noise and Contrast-To-Noise Measurements • Softcopy Workstation QC Order of Testing • Artifact Evaluation • Phantom Image Quality Evaluation • SNR and CNR Measurements • AEC Function Performance • Other Tests Artifact Evaluation Evaluation of detector, x-ray tube, filter etc – 4 cm thick acrylic block – No compression paddle – Four flat field images Mo/Mo, contact mode, LFS with grid Mo/Rh, contact mode, LFS with grid Mo/Mo, mag mode, SFS without grid Mo/Rh, mag mode, SFS without grid – Review the images on acquisition workstation monitor Window width = 500, Level = 350 Artifact Evaluation Artifact Evaluation Artifact Evaluation Evaluation of Printer – Select a computer-generated flat-field image – Print the image and evaluate it – Repeat the test for each printer If a printer is connected to multiple Selenia systems, it only needs to be tested once from one Selenia. Artifact Evaluation Phantom Image Quality Evaluation • • • • ACR phantom with acrylic disc AEC position 2 Print film Measure film densities - Background density > 1.20, with control limit ± 0.20 - Density difference over the acrylic disk > 0.40, with control limit ± 0.05 • Phantom scoring (hard copy and softcopy station) • 5 fibers, 4 speck groups, 4 masses Phantom Image Quality Evaluation SNR and CNR Measurements SNR and CNR Measurements • SNR ≥ 40 • Establish the reference CNR value during the acceptance test • CNR should stay within ±15% of the reference Automatic Exposure Control (AEC) • AEC modes Auto-Filter: Auto-kV: Auto-Time: filter, kVp, mAs all automatically determined filter manually selected kVp and mAs automatically determined filter and kVp manual selected mAs automatically determined • AEC Method - Filter and kVp are determined by breast thickness (in AutoFilter and Auto-kV modes) - A pre-exposure is used to determine mAs • AEC positions Auto-AEC (The densest tissue gets enough exposure) Seven manual positions (marked on compression paddle) AEC Function Performance 1. Performance at different thickness and imaging modes 2. Performance at different compensation levels AEC Function Performance 1. Performance at different thickness and imaging modes – Acrylic or BR-12 blocks – AEC position 2 – Contact mode (LFS, grid): 2, 4, 6, 8 cm phantom Mag mode (SFS, no grid): 4 cm phantom – Variation of target pixel values ≤ 10% of the mean target pixel value AEC Function Performance AEC Function Performance 2. Performance at different compensation levels – 4 cm acrylic or BR-12 – Compensation levels: Step -5 to Step +5 – Use average target count at Step 0 as reference – Compare target count at other steps to the reference Step -5 0.23 – 0.28 Step -4 Step -3 Step -2 Step -1 Step +1 Step +2 Step +3 Step +4 Step +5 0.36 – 0.44 0.50 – 0.61 0.63 – 0.77 0.77 – 0.94 1.04 – 1.27 1.17 – 1.43 1.31 – 1.60 1.44 – 1.76 1.58 – 1.93 Collimation Assessment 1. X-Ray Field to Light Field 2. X-Ray Field to Image Receptor 3. Compression Paddle to Image Receptor Collimation Assessment • One test for 24×29 cm imaging mode • Three tests for 18×24 cm imaging modes – Left imaging position – Center imaging position – Right imaging position L M R Collimation Assessment Use “coin method” as described in ACR manual Collimation Assessment 1. X-Ray Field to Light Field – Measure deviations between the edges of the light field (exterior edge of the coin) and the xray field at al four sides – Sum of the magnitude of deviations on opposite sides ≤ 2% of the SID SID = 66 cm Æ 2% of the SID = 1.32 cm Collimation Assessment 2. X-Ray Field to Image Receptor – d1: measured on film distance from the inner edge of the coin to the edge of the x-ray field – d2: measured on digital image distance from the inner edge of the coin to the edge of the detector – (d1×1.06 ) - d2 – The x-ray field shall extend beyond the chest wall side of the image receptor – The x-ray field must NOT extend >2% of the SID at any of the four sides of the image receptor Collimation Assessment 3. Compression Paddle to Image Receptor – Measure distance from the outer edge of the coin to the edge of the image – The anterior edge of the compression paddle must be aligned just beyond the chest wall edge of the image receptor, but ≤ 1% of the SID Collimation Assessment Collimation Assessment Collimation Assessment Evaluation of System Resolution • Line pair phantom • 4 cm acrylic block • Line pair phantom placed at 45º to the detector row Evaluation of System Resolution Limiting spatial resolution ≥ 7 c/mm (lp/mm) Detector Ghosting (Optional) 1. Create ghosting effect 2. Quantify ghosting effect Detector Ghosting (Optional) 1. Create ghosting effect Ghosting area Detector Ghosting (Optional) 2. Quantify ghosting effect Ghosting area Detector Ghosting (Optional) 2. Quantify ghosting effect • | Ghosting factor | ≤ 0.3 ROI1 in background without ghosting ROI2 in Al filter without ghosting ROI3 in Al filter with ghosting Ghosting Factor = meanregion2 - meanregion3 meanregion1 - meanregion2 Thank you! Tao Wu twu@hologic.com Tel: 203-207-4574