Outline Fundamentals of Single and Multiple Row Detector Computed Tomography • • − Mahadevappa Mahesh, Ph.D. − • • The Russell H. Morgan Department of Radiology and Radiological Science The Johns Hopkins University Baltimore, MD. USA Introduction • • • Single row detector helical CT Multiple row detector helical CT Four section/rotation scanners Scanners with >4 sections/rotation X-ray tube issues Relationship between pitch, dose, noise and section thickness Conventional XX-ray Imaging A recent survey* of internists rates CT among top 5 major medical innovations over the past 30 years Non-uniform beam Nonexits opposite surface with intensity pattern due to differential attenuation of rays along different paths through patient Two major evolutionary leaps occurred during last decade, spiral or helical CT in early 90’s and multiple--row detector CT late 90s to present multiple CT has evolved considerably since its invention in 1972, the progression might be characterized as search toward the 3D radiograph X -Ray Tube Uniform xx-ray beam enters patient Image receptor captures intensity pattern *Decisions in Imaging Economics, Nov 2001 The Problem 2D Images of 3D Anatomy from Single Projection Image due to differences in x-ray attenuation along different paths through the patient Mahadevappa Mahesh, Ph.D. Johns Hopkins University, Baltimore, MD • • Resolution >5 lp lp/mm /mm Acquisition time <<1 s (stops physiologic motion) But in 2D images of 3D anatomy • Tissues are superimposed • Poor contrast resolution due to high scatter acceptance by image receptor 1 Computed Tomography Ultimate Goal: 3D Radiography • Resolution as good as conventional radiography in all planes • Method for acquiring and reconstructing an image of a thin crosscross -section of an object • • High contrast sensitivity (no scatter) • Based on measurements of x -ray attenuation through the section plane using many different projections • Can CT get us there? Fast acquisition times to stop physiologic motion Basic data acquisition in CT CT Gantry X-ray Tube X-ray Beam Gantry Opening (60--70 cm dia (60 dia.) .) Sampled region where attenuation measurements are made (50 –55 cm diameter) y z CT Table x Detectors The CT Image Pixel Voxel w Pixel value is measure of x -ray attenuation in corresponding volume element (voxel (voxel ) • Depth dimension of voxel equal to section thickness (1 -10 mm) Mahadevappa Mahesh, Ph.D. Johns Hopkins University, Baltimore, MD Field of View (FOV): user selectable location and diameter within sampled region where reconstruction matrix is located Limitations of Conventional CT • • Scan plane resolution is ~1~1-2 lp lp/mm /mm Poor zz-axis resolution − Section thickness ranges 1 to 10 mm − Volumes underunder-sampled with abutted slices • Inter-scan delay due to stopInterstop-start action necessary for table translation and cable unwinding • Section-to Sectionto--section misregistration due to variation in patient respiratory motion 2 Progress toward true 3D imaging “Possible” • • • True 3D not yet practical! “Step-like” contours “StepLarge temporal lag between sections Not useful with physiologic motion Technological Advances That Led To: Helical (Spiral) Acquisition Surface rendition of early 3D reconstruction of lower limbs using 10 mm abutted sections Slip--Ring Technology Slip • • Slip Slip--Ring gantry • High power xx-ray tubes • Interpolation algorithms Permits continuous rotation of tube and detectors while maintaining electrical contact with stationary components Projection data Power supply Slip--Ring Technology Slip Helical/Spiral CT - Principles • Segment of slip-ring Sliding Contactors Patient is transported continuously through gantry while data are acquired continuously during several 360360 -deg rotations* * Kalender WA, et.al. Radiology, 176(1):181176(1):181-3, 1990 Mahadevappa Mahesh, Ph.D. Johns Hopkins University, Baltimore, MD 3 Technology Advances Helical Path of XX-Ray Beam on Patient • Interpolation algorithms − Projection data is no longer in a crosscross -sectional plane − Interpolation of projection data into plane of interest prior to conventional filtered backback-projection* * Kalender WA, et.al. Radiology, 176(1):181176(1):181-3, 1990 Helical SingleSingle -Section Mode Helical Pitch Helical Helical Trajectory Trajectory Pitch = • • • Translation z (mm) t (s) Table increment per rotation (mm) Beam collimation (mm) Typical Pitch Ratio - 0.5, 1.0, 1.5, 2.0 Pitch <1 implies overlapping and higher patient dose Pitch >1 implies extended imaging and reduced patient dose Interpolation Interpolation using using samples samples from from single single row row detector detector ring ring Capabilities of Single Row Detector CT (SDCT) • • • • Large tissue volumes scanned in short times Inter--scan delay eliminated Inter Arbitrary section position within scanned volume permits overover-sampling without increased dose Z axis resolution improved by overover-sampling • Up to ~ 2 lp lp/cm /cm (best case), usually 0.5 to 1.0 lp lp/cm /cm Mahadevappa Mahesh, Ph.D. Johns Hopkins University, Baltimore, MD Limitations of SDCT • • • Large volume scan in short duration is limited Near isotropic resolution only over small volume Poor utilization of XX-ray tube • Multiple row detector CT (MDCT) offers substantial improvement in volume coverage, scan speed with efficient use of x -ray tube 4 SDCT versus MDCT Multiple Row Detector Helical CT (MDCT) X-ray Tube Tube Collimator • Single row of detectors replaced with multiple rows Collimated Slice Detector Collimator Single multi -element module 1-Row Detector 8-Row Detector Multiple row detector CT scanner Single row detector CT Scanner *Mahesh M, RadioGraphics RadioGraphics,, 22: 949949-962, 2002 Uniform Element Arrays Non--Uniform Element Arrays Non Possible section widths 2 4 4 4 4 2 2 16 x 1.25 mm 20 mm Z-axis x x x x x x x 0.63 mm 1.25 mm 2.5 mm 3.75 mm 5 mm 7.5 mm 10 mm Possible section widths 5 2.5 1.5 1 1 1.5 2.5 5 20 mm Z-axis x 0.5 mm x 1 mm x 2.5 mm x 5 mm x 8 mm X 10 mm Volume Zoom, Siemens Medical Systems Lightspeed , GE Medical Systems Hybrid Element Arrays 2 4 4 4 2 2 MDCT: Detector Element Arrays 16 x 1.25 mm Possible section widths GE 4 x 0.5 15 mm 15 mm 32 mm 4 4 4 4 4 4 2 x 0.5 mm x 1 mm x 2 mm x 3 mm x 5 mm x 8 mm x 10 mm Z-axis Acquilion,, Toshiba Medical Systems Acquilion Mahadevappa Mahesh, Ph.D. Johns Hopkins University, Baltimore, MD 20 mm 5 2.5 1.5 1 1 1.5 2.5 5 20 mm Siemens & Philips 4 x 0.5 Toshiba 15 mm 15 mm 32 mm Z-axis 5 How are detector elements used in MDCT? Detector Configuration: For 4 x 1.25 mm X -ray Tube X-ray Tube Focal Focal Spot Spot X -ray Beam X-ray Beam Collimator Collimator 4 x 1.25 mm Detectors Detectors 20 mm Switching Switching Array Array 4 x 1.25 mm Detector Detector 4- section scanners collect four simultaneous channels of data Switching Switching Array Array Detector Configuration: For 4 x 2.5 mm 20 mm MDCT: Detector Configurations X -ray Tube X-ray Tube Focal Focal Spot Spot X -ray Beam X-ray Beam Collimator Collimator 4 x 2.5 mm Detector Detector *GE LightSpeed , GE Medical Systems Switching Switching Array Array *Volume Zoom, Siemens Medical System 20 mm Helical Multiple Section Mode 4 helical trajectories The Detector’s Evolution… Translation z t Interpolation using samples of ALL detector rings Mahadevappa Mahesh, Ph.D. Johns Hopkins University, Baltimore, MD 6 DAS channels: Four versus Eight Detector Evolution: 4 vs 16 sections per rotation 4 x 1.25 mm 5 Detector Detector 2.5 1.5 1 1 1.5 2.5 2X0.5, 4x1, 4x2.5, 4x5 2X8, 2x10 5 20 mm 20 mm Switching Switching Array Array 16 x 0.75 8 x 1.25 mm 16x0.75, 16x1.5, 8x3, 4x6… 4 x 1.5 4 x 1.5 24 mm Z-axis GE Medical Systems Siemens & Philips Medical Systems Detector Evolution: 4 vs. 16 sections per rotation 4 x 0.5 4x0.5, 4X1, 4x2, 4x3 up to 4x8 15 mm 15 mm MDCT Episode II: Attack of the Cones 32 mm 16 x 0.5 16x0.5, 16X1, 16x2, up to 8x4 12 x 1 12 x 1 32 mm Z-axis Toshiba Medical Systems Image Reconstruction: Key Problem is Cone Angle Cone Beam Geometry Section thickness Focus cone angle fan angle Z-axis • In MDCT, widening beam aperture in zzdirection increases cone angle, that results in significant cone beam artifacts Section width ‘S’ Scan FOV Detector Array ‘’Section blurring’’ dS Scan direction 4 -section 8 -section • Up to 4 sections dS = S, cone angle may be neglected • What about more than 4 sections? Mahadevappa Mahesh, Ph.D. Johns Hopkins University, Baltimore, MD 7 Cone Beam Geometry: Alternate Reconstruction Algorithms Key Problem: Cone Angle • What happens, if the cone angle of the rays is neglected ? 4*1mm, pitch 1.5 (6) 8*1mm, pitch 1.5 (12) 12*1mm , pitch 1.5 (18) 16*1mm , pitch 1.5 (24) 12 mm/sec 24 mm/sec 36 mm/sec 48 mm/sec • Advanced Single Slice Rebinning (ASSR) • Adaptive Multiple Plane Reconstruction (AMPR) • • Pi,, Pi Pi Pi--Slant and 33-Pi methods Helical Feldkamp with weighting function (HFK) • Image results for > 4 sections are clinically unacceptable ! Kohler T, et al., Medical Physics, 29 (1): 5151-64, 2002 Courtesy Siemens Medical Systems Cone Beam Reconstruction (e.g. Adaptive Multiple Plane Reconstruction) 4*1mm, pitch 1.5 (6) 12 mm/sec Standard Reconstruction 16*1mm , pitch 1.0 (16) 32 mm/sec Standard Reconstruction 16*1mm , pitch 1.0 (16) 32 mm/sec AMPR Reconstruction Helical Pitch - Reality vs. Myth Definition, Confusion… Courtesy Siemens Medical Systems Pitch redefined for MDCT Beam Pitch vs. Detector Pitch T Beam Pitch = W T Detector Pitch = W=D T D Detector Pitch Beam Pitch = N Single Row Detector Array T - Table travel (mm)/rotation W - Beam width (mm) Table Speed (mm/rotn (mm/ rotn)) Detector combi Detector Pitch Beam Pitch 3.75 4 x 1.25 3 0.75 7.5 4 x 1.25 6 1.50 7.5 4x3 2.5 0.625 13.5 4x3 4.5 1.125 16.5 4x3 5.5 1.375 T W D Multiple Row Detector Array D - Single DAS channel width (mm) N - Number of active DAS channels Data for fourfour -section MDCT Mahadevappa Mahesh, Ph.D. Johns Hopkins University, Baltimore, MD 8 Example of Beam Pitch Options Beam Pitch • Beam Pitch >1 implies extended imaging and reduced patient dose with lower axial resolution Z • Beam Pitch <1 implies overlapping and higher patient dose with higher axial resolution dd 3d 3d Detector Detector Pitch Pitch 33 Beam Beam Pitch Pitch 0.75 0.75 dd 6d 6d Detector Detector Pitch Pitch 66 Beam Beam Pitch Pitch 1.5 1.5 Lightspeed , GE Medical Systems Beam Pitch vs. Volume Coverage • Dose in Helical CT varies as: Increase in pitch implies faster acquisition and larger volume coverage Dose • Lower pitch implies slower table speed with overlapping of tissue (for P<1) and smaller scanned volume ∝ Beam Pitch 1 (mAs/rotation) Beam Pitch vs. Dose • Varying pitch results in increase or decrease of radiation dose to patient • However in some MDCT scanner, image noise is maintained constant by varying tube current (“effective mAs mAs”), ”), resulting in patient dose independent of pitch* pitch* High Power XX-ray Tubes *Mahesh M, et.al., AJR, 177: 12731273-1275, 2001 Mahadevappa Mahesh, Ph.D. Johns Hopkins University, Baltimore, MD 9 X-ray Tubes • In helical CT, ZZ-axis resolution and scan volume place huge demands on tube • Several technical advances have been made to achieve power levels and deal with problems of heat generation, storage and dissipation X-ray tubes used for Spiral CT • • Larger anode disks allow higher tube currents Anodes of graphite based body with tungstentungstenrhenium or tungstentungsten-zircon zircon--molybdenum molybdenum** layer deposited by sintering or chemical or physical vapor process * Ammann E, et al., BJR, 70, S1 -S9, 1997 X-ray tubes used for spiral CT • X-ray tube used in Spiral CT Ceramic insulators Metal envelopes with ceramic insulators provide higher heat storage capacity • Spiral groove bearings improve heat dissipation requiring shorter cooling periods and therefore allow continuous rotation with minimal wear Direct oil cooling of spiral groove bearing Unique 200 mm anode disk Compact, all metal envelope Courtesy Philips Medical Systems Noise Modern CT XX-Ray Tubes • Heat storage capacity exceeds >3>3-8 MHU • No longer the limitations for studies demanding higher speed and larger volume coverage Noise 1 ∝ vno. of photons Tube current Scan time Section width • Double the tube current, reduces noise by √2 • Halve the section width, increases noise by √ 2 Mahadevappa Mahesh, Ph.D. Johns Hopkins University, Baltimore, MD 10 Noise vs. Pitch • For SDCT, noise is independent of pitch for constant mAs and section width • However on most MDCT scanners, system software automatically adjust scanmA scan mA per Effective Section Thickness protocol to obtain comparable image noise as user alters acquisition parameters Section and Beam Collimation • SDCT: Both are same, influences zz-axis coverage per gantry rotation • MDCT: Section thickness* thickness * is total beam collimation divided by number of active detector channels − Section Thickness • True thickness of the reconstructed image, measured as full width at half maximum (FWHM) of slice sensitivity profile • Same as beam collimation in Slice Sensitivity Profiles: conventional scanning but conventional and spiral acquisition different in spiral scanning e.g., 10 mm / 4 channels = 4 x 2.5 mm *defined at center of rotation Effective Section Thickness • • Measure of slice sensitivity profile at FWHM • In SDCT user selects section thickness, but true width of reconstructed section is influenced by pitch and interpolation algorithm (180° vs. 360°) • Affected by beam collimation, pitch and interpolation algorithm In MDCT user selects beam collimation in combination with desired section width which is affected by pitch, interpolation algorithm & ZZ-filter Mahadevappa Mahesh, Ph.D. Johns Hopkins University, Baltimore, MD Pitch vs. Effective Section Thickness • Increasing pitch broadens effective section thickness • Structures outside nominal section thickness will contribute to image 11 Evolution of Isotropic Voxel MDCT Advantages Acquisition of same region in shorter scan time or larger region in same scan time − Thinner slices yielding higher zz -axis resolution − Better tube utilization − Greater coverage per breath hold − Better use of contrast agents 1.0 mm 0.5 mm Approaching Isotropic Resolution! Conventional Helical: SDCT → MDCT Speed of Volume Acquisition Region Head Neck Chest Abdomen Pelvis 0.2 mm 0.2 mm Increased coverage per rotation 0.5 mm • 1.0 mm − 5 mm • Compared to SDCT 10 mm • Radiography (200 µ m) CT Timeline Distance (cm) Section Thickness (mm) 20 15 30 20 20 8 5 8 8 8 16.7 20.0 25.0 16.7 16.7 2.1 2.5 3.1 2.1 2.1 Total 95.1 11.9 Total scan time (sec) SDCT† MDCT‡ Slip -ring technology Slipone second scan Half second scan Sub--second scan Sub 72 . . . 85 86 87 88 89 90 91 92 93 94 95 96 97 98 99 00 01 02 Invention of CT Spiral CT Twin detector CT † 1 sec scan, pitch 1.5 sec scan, pitch ~ 1.5 for 44-section MDCT ‡ 0.5 Future Directions • • Partial rotation scan times ~150 ms possible! • Extended zz-axis coverage to cover most organs in one or two gantry rotations should be possible with large area detectors or flat panel detectors PET-CT PET16+ sections CT Conclusions • Cone beam reconstruction algorithms for 16, 40 and 64 row detectors are available Mahadevappa Mahesh, Ph.D. Johns Hopkins University, Baltimore, MD MDCT • CT technology has evolved to level where large 3D volumes can be imaged with: − isotropic resolution − acquisitions independent of most physiologic motion 3D imaging of 3D anatomy - the 3D radiograph - is becoming a reality! 12