Optimizing Image Quality and Dose in Digital Mammography: X-ray Spectrum and Exposure Parameter Selection Mark B. Williams1, Priya Raghunathan1, Mitali J. More1 Martin J. Yaffe2, Aili Bloomquist2, Gordon Mawdsley2, Joseph Lo3, Ehsan Samei3, Nicole Ranger3, J. Anthony Seibert4, Alex Kwan4, Laurie Fajardo5, Allen McGruder5, Sandra Maxwell5, Andrew D.A. Maidment6 Objective Examine the effect on image quality and radiation dose of the choice of beam quality parameters (target, filter, kVp) for current commercial FFDM systems of Virginia, 2University of Toronto University, 4University of CaliforniaDavis, 5University of Iowa, 6University of Pennsylvania 1University 3Duke What is being optimized? • Screen-film mammography maximization of contrast within the constraint of acceptable film darkening • FFDM - separation of the processes of acquisition and display permits the displayed contrast of individual structures to be adjusted when the image is viewed What is being optimized? • However, display contrast in FFDM is limited by the inherent image signal to noise ratio (SNR), because as the signal contrast is increased, so is the visibility of noise. • Thus beam optimization in digital mammography requires maximization of the image SNR. • But since the SNR can be improved almost arbitrarily by increasing the number of detected x-ray photons, exposure parameter optimization must balance increased image SNR with increased patient radiation dose 1 Previous Study FFDM systems evaluated: • Fischer • GE • Trex Williams MB, More MJ, Venkatakrishnan V, Niklason L, Yaffe MJ, Mawdsley G, Bloomquist A, Maidment ADA, Chakraborty D, Kimme-Smith C, and Fajardo LL. “Beam optimization for digital mammography”. IWDM 2000: 5th International Workshop on Digital Mammography. (Martin J. Yaffe, Editor). Medical Physics Publishing. Madison, Wisconsin, 2001; 108-119. FFDM Systems Evaluated Manufacturer Model Test Site GE Healthcare Senographe 2000D University of Virginia University of Pennsylvania Duke University Siemens Hologic Mammomat Novation DR Selenia University of Iowa Fischer Senoscan University of Toronto Fuji 5000MA University of California Davis Current Study • • • • • Fischer SenoScan GE Healthcare Senographe 2000D Hologic Selenia Siemens Mammomat Novation DR Fuji 5000 MA (Lorad M-IV) GE Senographe 2000D: amorphous silicon photodiodes with overlying x-ray converter Columnar CsI(Tl) Photodiode – TFT array CsI(Tl) p+ n+ a-Si 2 Hologic Selenia and Siemens Mammomat Novation DR: photoconducting amorphous selenium layer read out by a-Si thin film transistor array bias electrode Fischer Senoscan: Senoscan: CCDCCD-based scanned slot system Front View drain Fan Beam Collimator Swing Arm gate Compressed Breast Phosphor Fiber Optic CCD Pb Shield charge collection electrode dielectric A/D A/D Direction of Scan DMA Computer storage capacitor Dual Sided Imaging Plate Readout Display X-ray Tube amorphous selenium source Side View Focal Spot Acquisition Parameters FFDM system Fischer Target Filter kVp range W Al 28 – 37 GE Mo, Rh Mo, Rh 23 – 39 Hologic Mo Mo, Rh 23 – 39 Siemens Mo, W Mo, Rh 23 – 35 Fuji (Lorad M-IV) Mo Mo, Rh 22 – 34 Diagram courtesy of FUJIFILM 3 Methods • Common set of breast equivalent phantoms circulated among all test sites • Phantoms simulated nine breast types (3 compressed thicknesses x 3 fibroglandular/adipose compositions) • For each phantom, for each possible target/filter combination, images were obtained over a range of kVps, manual mode, average pixel value approximately constant • SNR and mean glandular dose (MGD) were calculated for each Number of Images Processed Example: Senographe 2000D at UVa • • • • 9 phantom types 3 target/filter combinations 8 kVps per target/filter combination 2 images per target/filter/kVp/phantom combination • Total of 432 images Phantom SNR Analysis Thickness 3 cm, 5 cm, 7 cm Mass stepwedge 1 cm Glandular/adipose mass fraction 30/70, 50/50, 70/30 Skins are 100% adipose Two images obtained for each phantom/target/filter/kVp combination. Raw (For Processing) images used. 1 cm 12 cm ( Signal ≡ P2 − P1 − P4 − P3 SKINS MASS STEPWEDGE Calc stepwedge 10 cm CIRS, Inc., Norfolk, VA Noise ≡ FOM ≡ σ 5, difference 5 3 4 1 2 ) 2 2 SNR MGD 4 Dose Calculation Limitations of the FOM • Boone, J. (1999). Glandular breast dose for monoenergetic and high-energy x-ray beams: Monte Carlo assessment. Radiology 213, 2337. • Sobol, WT and Wu, X. (1997) Parameterization of mammography normalized average glandular dose tables. Medical Physics 24(4), 547-555. • Stanton, L., Villafana, T., Day, J., and Lightfoot, D. (1984). Dosage evaluation in mammography. Radiology 150, 577-584. Independence of Shape of Signal vs. kVp on Step Height • SNR did not include spatial frequency dependence – Systems with higher MTFs and thus less smoothing of Poisson x-ray noise may be penalized – Inter-system comparisons can be misleading • Exposure time was not taken into account • The primary sources of uncertainty in the data were: – Noise in the measured HVL data – Fluctuating system noise; this varied greatly among manufacturers Independence of Shape of Signal vs. kVp on Step Height Normalized Signal vs kVp, Novation DR, W/Rh 7 cm 50/50 Signal v s kVp, Nov ation DR, W /Rh, 7 cm 50/50 180 180 160 160 140 signal (ADUs) signal (ADUs) 140 120 100 80 60 Step Step Step Step 40 20 0 1 2 3 120 100 80 Step 0 Step 1 Step 2 Step 3 60 40 20 0 0 20 25 30 kVp 35 40 20 22 24 26 28 30 32 34 36 38 40 kVp 5 Independence of Optimum Target/Filter on Step Height Independence of Optimum Target/Filter on Step Height Normalized Signal vs. kVp, Senographe 2000D, All Target/Filters 3 cm 30/70 Signal vs. kVp, Senographe 2000D, All Target/Filters 3 cm 30/70 1600 Normalized Signal (ADUs) Step 1 MoMo Step 2 MoMo Signal (ADUs) 1200 Step 3 MoMo 1000 Step 0 MoRh Step 1 MoRh 800 Step 2 MoRh 600 Step 3 MoRh Step 0 RhRh 400 Step 1 RhRh 200 Step 2 RhRh 0 25 30 35 Step 1 MoMo 1400 Step 2 MoMo 1200 Step 3 MoMo Step 0 MoRh 1000 Step 2 MoRh 600 Mo/Mo 400 Step 3 MoRh Mo/Rh Step 0 RhRh Step 1 RhRh 200 Step 2 RhRh 0 Step 3 RhRh 20 40 25 30 35 40 kVp kVp Contrast vs kVp, 5 cm 70/30, best performing target/filter combinations (Noise/ADU)2 vs. kVp, 5 cm 30/70, best performing target/filter combinations 0.0005 0.40 Siemens W/Rh 0.25 Hologic Mo/Rh 0.20 0.15 Fuji Mo/Mo signal avg. bkgnd . pixel value 0.0004 2 0.30 GE - Rh/Rh 0.0005 contrast = (Noise/ADU) GE Rh/Rh 0.35 Contrast Step 1 MoRh Rh/Rh 800 Step 3 RhRh 20 Step 0 MoMo 1600 Step 0 MoMo 1400 Siemens W/Rh 0.0004 0.0003 Hologic Mo/Rh 0.0003 0.0002 Fuji Mo/Mo 0.0002 0.10 Fischer 0.05 0.0001 Fischer 0.0001 0.0000 0.00 22 25 28 31 kVp 34 37 40 20 23 26 29 32 kVp 35 38 6 SNR versus kVp, 5 cm 50/50, best performing target/filter combinations 30 DgN versus kVp Example: Fuji, 50/50 composition 7 Siemens W/Rh 3 cm Mo/Mo 25 6 Hologic Mo/Rh Fuji Mo/Mo 15 10 Fischer W/Al 5 3 cm Mo/Rh DgN (mRad/R) SNR 20 5 4 5 cm Mo/Mo 3 5 cm Mo/Rh 2 7 cm Mo/Mo GE Mo/Rh 1 7 cm Mo/Rh 0 20 23 26 29 32 35 0 38 kVp 20 23 26 29 32 35 kVp Dose versus kVp Example: Fuji, 50/50 composition FOM versus kVp, fixed composition Hologic, 50/50 2.50 10 3 cm Mo/Mo 3 cm Mo/Mo 3 cm Mo/Rh 3 cm Mo/Rh 5 cm Mo/Mo 1.50 5 cm Mo/Mo 1.00 5 cm Mo/Rh 5 cm Mo/Rh FOM MGD/ADU 2.00 1 7 cm Mo/Mo 22 26 30 7 cm Mo/Mo 34 7 cm Mo/Rh AF - 3 cm 0.50 AF - 5 cm 7 cm Mo/Rh AF - 7 cm 0.00 0.1 20 25 30 kVp 35 kVp 7 FOM versus kVp, fixed composition GE, 50/50 100 FOM versus kVp, fixed composition Siemens, 50/50 3 cm Mo/Mo 10 3 cm Mo/Rh 3 cm Mo/Mo 3 cm Rh/Rh FOM 10 5 cm Mo/Mo 3 cm Mo/Rh 5 cm Mo/Rh 3 cm W/Rh 5 cm Rh/Rh 5 cm Mo/Mo 5 cm Mo/Rh 7 cm Mo/Rh 7 cm Rh/Rh FOM 7 cm Mo/Mo 22 AEC - C - 3 cm 1 22 25 28 31 34 37 40 5 cm W/Rh 1 24 26 28 30 32 34 36 7 cm W/Rh AEC - D - 3 cm AEC - 3 cm AEC - C - 5 cm AEC - 5 cm AEC - SD - 5 cm 0.1 AEC - C - 7 cm kVp AEC - 7 cm 0.1 AEC - SD - 7 cm FOM versus kVp, fixed composition Fischer, 50/50 7 cm Mo/Mo 7 cm Mo/Rh AEC - S - 3 cm kVp FOM versus kVp, fixed composition Fuji, 50/50 10 10 3 cm Mo/Mo 3 cm W/Al 3 cm Mo/Rh 5 cm Mo/Mo 5 cm W/Al FOM FOM 5 cm Mo/Rh 1 7 cm Mo/Mo 20 7 cm W/Al 22 24 26 28 30 32 34 7 cm Mo/Rh AEC - 3 cm AEC - 5 cm 1 26 29 32 35 kVp 38 41 44 AEC - 7 cm 0 kVp 8 Techniques producing maximum FOM GE 3 cm 5 cm 7 cm 30/70 Mo / Rh / 27 Mo / Rh / 27 Mo / Rh / 27 50/50 Mo / Rh / 27 Mo / Rh / 27 Rh / Rh / 31 70/30 Mo / Rh / 27 Mo / Rh / 27 Rh / Rh / 33 3 cm 5 cm 7 cm W / Rh / 27 W / Rh / 27 W / Rh / 27 W / Rh / 27 W / Rh / 27 W / Rh / 29 W / Rh / 29 W / Rh / 29 W / Rh / 29 3 cm 5 cm 7 cm Mo / Rh / 27 Mo / Rh / 27 Mo / Rh / 27 Mo / Rh / 27 Mo / Rh / 27 Mo / Rh / 28 Mo / Rh / 27 Mo / Rh / 27 Mo / Rh / 28 HVL at peak FOM, 3 cm WAl Fischer MoRh MoMo Siemens Hologic Fischer 3 cm 5 cm 7 cm W / Al / 27 W / Al / 29 W / Al / 35 W / Al / 27 W / Al / 30 W / Al / 35 Fuji RhRh GE MoRh Hologic MoMo MoRh Siemens MoMo 30/70 WRh W / Al / 27 W / Al / 30 W / Al / 35 50/50 MoRh MoMo Fuji 3 cm Mo / Mo / 24 Mo / Mo / 24 Mo / Mo / 26 5 cm 7 cm Mo / Mo / 24 Mo / Rh / 28 Mo / Mo / 24 Mo / Rh / 28 Mo / Mo / 26 Mo / Rh / 28 70/30 0.20 0.25 0.30 0.35 0.40 0.45 0.50 0.55 0.60 0.65 0.70 HVL (mm Al) HVL at peak FOM, 5 cm HVL at peak FOM, 7 cm 12 W Al WAl Fischer MoRh MoRh 10 Fischer MoMo MoMo Fuji MoRh GE Hologic MoMo MoRh RhRh 8 RhRh MoMo 6 MoMo WRh MoRh MoMo 30/70 Hologic MoRh Siemens 4 Fuji GE MoRh Siemens MoMo 30/70 WRh 50/50 50/50 MoRh 2 70/30 70/30 MoMo 0 0.20 0.25 0.30 0.35 0.40 0.45 0.50 0.55 0.60 0.65 0.70 0.20 0.25 0.30 0.35 0.40 0.45 0.50 0.55 0.60 0.65 0.70 HV L (m m Al) HVL (mm Al) 9 FOM versus kVp, 3 cm 30/70, best FOM versus kVp, 3 cm 50/50, best performing target/filter combinations performing target/filter combinations 14 Fischer W/Al 14 Fischer W/Al Fuji Mo/Mo Fuji Mo/Mo 12 12 Fuji AEC Fuji AEC GE Mo/Rh GE Mo/Rh 10 10 GE AOP - C 8 GE AOP - S GE APO - D 6 GE AOP - S FOM FOM GE AOP - C 8 4 GE AOP - D Hologic Mo/Rh Hologic Mo/Rh 6 Hologic AEC Hologic AEC 2 Siemens W/Rh 4 Siemens W/Rh Siemens AEC Siemens AEC 0 20 22 24 26 28 30 kVp 32 34 36 38 2 40 20 22 24 26 28 30 32 kVp 34 36 38 40 FOM versus kVp, 3 cm 70/30, best FOM versus kVp, 5 cm 30/70, best performing target/filter combinations performing target/filter combinations 5 12 Fuji Mo/Mo Fuji Mo/Mo 10 4 Fuji AEC Fuji AEC GE - Mo/Rh GE Mo/Rh 8 GE AOP - C GE AOP - S 6 GE AOP - D Hologic Mo/Rh 4 GE AOP - C 3 FOM FOM Fischer W/Al Fischer W/Al GE AOP - S GE AOP - D 2 Hologic Mo/Rh Hologic AEC Siemens W/Rh 2 Hologic AF 1 Siemens W/Rh Siemens AEC Siemens AEC 0 0 20 22 24 26 28 30 32 kVp 34 36 38 40 20 23 26 29 kVp 32 35 38 10 FOM versus kVp, 5 cm 50/50, best FOM versus kVp, 5 cm 70/30, best performing target/filter combinations performing target/filter combinations 4 4 Fischer W/Al Fischer W/Al Fuji Mo/Mo Fuji Mo/Mo Fuji AEC 3 Fuji AEC 3 GE - Mo/Rh GE Mo/Rh GE AOP - C GE AOP - D GE AOP - C FOM FOM GE AOP - S 2 GE AOP - S 2 GE AOP - D Hologic Mo/Rh Hologic Mo/Rh Hologic AF Hologic AF 1 Siemens W/Rh 1 Siemens W/Rh Siemens AEC Siemens AEC 0 0 20 23 26 29 32 35 38 22 25 28 31 kVp kVp 34 37 40 FOM versus kVp, 7 cm 30/70, best FOM versus kVp, 7 cm 50/50, best performing target/filter combinations performing target/filter combinations 2 1.5 Fischer W/Al Fischer W/Al Fuji Mo/Rh Fuji Mo/Rh Fuji AEC Fuji AEC 1.5 GE Mo/Rh 1 GE AOP - D Hologic Mo/Rh 0.5 GE AOP - C FOM FOM GE AOP - S GE Rh/Rh 1 GE AOP - C GE AOP - S GE AOP - D Hologic Mo/Rh 0.5 Hologic AF Hologic AF Siemens W/Rh Siemens W/Rh Siemens AEC Siemens AEC 0 0 22 26 30 34 kVp 38 42 22 25 28 31 34 kVp 37 40 43 11 FOM versus kVp, 7 cm 70/30, best performing target/filter combinations Mean Glandular Dose Using AEC Selections (max FOM settings for Fischer) 350 1.2 Fischer W/Al 300 Fuji AEC GE Rh/Rh 0.8 FOM GE AOP - C GE AOP - S GE AOP - D Hologic Mo/Rh 0.4 Hologic AF Siemens W/Rh Siemens AEC Dose Using AEC (mrads) Fuji MoRh GE Siemens Hologic Fuji Fischer 250 200 150 100 50 0 22 25 28 31 34 kVp 37 40 43 0 3 cm 30/70 3 cm 50/50 3 cm 70/30 5 cm 30/70 5 cm 50/50 5 cm 70/30 7 cm 30/70 7 cm 50/50 7 cm 70/30 Conclusions – Detector Dependence Conclusions – kVp Dependence • Optimum exposure parameters for a specific breast type are system (detector) specific – there are no universal values • The kVp dependence of the FOM and the relative performances of different target/filter combinations are independent of signal (step) size and step composition • Peaks in FOM versus kVp curves are broad, when they exist • The choice of target/filter is much more crucial than the choice of kVp 12 Conclusions – kVp Dependence • In all cases the FOM is a decreasing function of kVp at the upper end of the kVp range tested here • Thus at least for the target/filter combinations currently included in these FFDM systems, there is no advantage in expanding the available voltage range to even higher kVp. Conclusions – Noise Dependence • There is a general inverse correlation between the normalized noise and the FOM, with higher noise systems tending to have lower FOM values • This suggests that the system noise is an important determinant of FFDM system performance, irrespective of signal and dose performance. Conclusions – Target/Filter Dependence • For the Senographe 2000D, the maximum FOM was always obtained with either Mo/Rh or Rh/Rh target filter combinations, but never with Mo/Mo, even for the thinnest and most fatty breasts • The Selenia produced maximum FOM values for nearly all breast types at an exposure parameter setting of Mo/Rh, 27 kVp, with only a slightly higher kVp (Mo/Rh, 28 kVp) for the two densest 7 cm breasts • Contrary to the Senographe 2000D and the Selenia, the maximum FOM was obtained for the 5000MA with a Mo/Mo combination for most (6 of 9) breast types, with only the 7 cm phantoms benefiting more from Mo/Rh. Conclusions – HVL for Optimum Settings • The HVL of the technique producing the highest FOM tended to fall at a location within the range of available HVL values that is quite systemspecific. • For the Senoscan and 5000MA systems, the optimum techniques had HVL values that increased with increasing breast thickness, but were always in the lower half of the available HVL ranges. • For the Mammomat Novation DR, the optimum technique factors were nearly identical for all breast types, always utilized the W/Rh combination, and corresponded to a HVL that was near the top of the available range 13 Conclusions – AEC Dose Conclusions – AEC Performance • There is a substantial difference among FFDM systems in the radiation dose delivered under AEC operation. • In general, the Fuji system delivers the highest dose under its AEC settings of all systems tested, while the Novation DR delivers the least of all systems with AECs. • The Senoscan, operated with kVp settings close to those maximizing the FOM for that system, delivers the lowest radiation dose of all systems tested for 5 cm and 7 cm breasts . • The AEC systems of current commercial systems generally select technique factors close to those producing the maximum FOM value, but selection could be improved in several cases • Preset minimum kVp values may be too high in some systems Conclusions – AEC Performance Conclusions – AEC Performance • For the Senographe 2000D, the best FOM values for all 3 cm phantoms were obtained with the Mo/Rh target/filter combination. • However, the AEC selected Mo/Mo for all 3 cm phantoms in all three modes, resulting in lower FOM values • For the Selenia, the FOM for the Mo/Rh combination is superior to that for the Mo/Mo combination at virtually any voltage for all breast types. • Since the AEC (autofilter mode) selected Mo/Mo for 3 cm breasts of all compositions, dose performance could be improved without loss of image quality by programming the AEC for selection of the rhodium filter for smaller breast thickness 14 Conclusions – AEC Performance • The W/Rh target filter combination on the Mammomat Novation DR outperformed Mo/Mo and Mo/Rh for all breast types imaged. • Novation DR kVp selection is based on compressed thickness only, thus AEC selection is insensitive to changing breast composition, and the same target/filter/kVp is chosen for all breast compositions of a given thickness. • The FOM curves suggest that compared to the current AEC selections, some performance improvement may be possible through the utilization of somewhat lower kVp settings for thicker (i.e. 7 cm or thicker) breasts with only minimal increase in MGD. Acknowledgements • Patty Goodale Judy, University of Virginia • Gordon Axt, University of Iowa 15