Research Journal of Applied Sciences, Engineering and Technology 4(21): 4463-4468, 2012 ISSN: 2040-7467 © Maxwell Scientific Organization, 2012 Submitted: May 01, 2012 Accepted: June 08, 2012 Published: November 01, 2012 Testing and Analysis of Pulse Detection Circuits Based on the Concept of Chinese Medicine Xinsheng Che, Xiaoxue Gu, Dongxue Fan and Hui Xu Shenyang University of Technology, Shenyang 110870, P.R. China Abstract: In order to get real pulse information and research pulse instrument by using Chinese medicine, a method about testing pulse detection circuits is proposed by use of the definition of Chinese medicine pulse in this study and detection circuits from three different pulse instruments are analyzed using this method. The existence of distortion in circuits is proved by using electronic circuit simulation and hardware circuit experiments. At the same time, the quantified distortion errors of three pulse detection circuits are given according to describing the pulse parameters of Chinese medicine. These pulse parameters and data can be used to optimize the pulse detection circuits. Keywords: Chinese medicine pulse, electronic circuit simulation, pulse detection circuit, pulse instrument, pulse parameter Principles and compositions of pulse instrument: Pulse instrument is a kind of Chinese medicine pulse diagnosis instrument which can collect, analyze, process and obtain human pulse diagrams automatically (Zhao et al., 2006) and it consists of sensor unit, signal conditioning unit, data acquisition unit and computer unit in general (Fan, 2007). The corresponding system diagram is shown in Fig. 1. Human pulse signal is transformed to electronic signal through pulse sensor and signal conditioning unit amplify the weak signal and filter off noise. Then analog quantity is transformed to digital quantity through A/D converter. Finally, the pulse diagrams are shown by computer. INTRODUCTION Traditional Chinese Medical Pulse instrument should be objective (Lu et al., 2010). There are many methods of pulse detection and pulse instruments, but there is no pulse instrument or method of pulse detection which can be proofed to get real human pulse diagrams. Pulse instrument is a testing instrument and represents a kind of testing technology. A method of verifying pulse instrument should be put forward to build the instrument with no distortion. This is also clinical objective demand of pulse application. Many kinds of pulse instruments exist on different research directions. According to the definition of pulse, all the structures, detection circuits and processing data of pulse instruments are similar. The distortion focuses on the filter circuits of signal conditioning. According to the filter circuits which are used in pulse instruments, a method is brought up to test the pulse detection circuits by using circuit simulation and experiments. The research of this study is the basis of optimizing pulse detection circuit and evaluating the distortions by pulse parameters. Extension-rule based TP method has commended. Recovering pulse signal: According to Chinese medicine, pulse is defined as the form which doctors feel by their fingers. It is the function of changing force by time in description of curve. Pulse signal meets the Dirichlet condition in a short time and can be described by Fourier series which gives the pulse characteristic from the aspect of signal and provides basis for the design of pulse collection system (Zhu, 1995), extraction of pulse feature and access of pulse signal: DETECTION OF PULSE SIGNAL Pulse instrument is a medical instrument which can reflect pulse signal objectively. The system composition depends on the description and research direction of pulse signal (CUI INC., 2010). According to the definition of Chinese medicine, description and detection about pulse are more directive and objective based on Fourier series. f t d 0 d n sinn1 n (1) n 1 where, d0 is static stress from doctors’ fingers (stress for getting pulse) which is a direct current value in pulse Corresponding Author: Xinsheng Che, Shenyang University of Technology, Shenyang 110870, P.R. China 4463 Res. J. Appl. Sci. Eng. Technol., 4(21): 4463-4468, 2012 Fig. 1: System diagram of pulse detection instrument instrument based on pulse original definition. The second part is the doctor fingers’ feeling about the pulse (pulse waveform) which is described as harmonic waves where dn is the magnitude. In general d0 is much larger than dn. Except for the direct current, i.e., the stress for getting pulse, there are much noises mixed in pulse signal, like power frequency interference (50 Hz) and high frequency interference, which should be filtered off. To separate static stress and pulse waveform, highpass filter circuits with low cut-off frequency are used. Some pulse instrument can be seen in literatures in recent years. The first one we want to analyze is the one designed by Shenyang University of Technology in 2003. There uses a two voltage-controlled active Butterworth high-pass filter circuit to filter out the direct signal and its cut-off frequency is 0.143 Hz4. This filter circuit is named as Example 1 for short in this study. The second model is a pulse instrument designed by Tianjin University in 2007 in which high-pass filter circuit is combined with low-pass filter circuit (Che et al., 2005). The high-pass filter circuit is named as Example 2 for short. Its cut-off frequency is 0.08 Hz. The cut-off frequency of the low-pass filter circuit is 40 Hz. High frequency component of pulse can be preserved and most high frequency noise can be filtered. The pulse instrument designed by Guangdong University of Technology in 2008 also uses high-pass filter circuit combined with low-pass filter circuit (Gen, 2007). The cut-off frequency of the high-pass and the low-pass filter circuits are 0.025 and 45 Hz, respectively. The high-pass filter circuit is named as Example 3 for short. The filter circuit of the pulse instrument designed by Northeast Normal University in 2008 adopts lowpass filter and notch circuit taking into account the influence of the industrial power frequency (Li, 2008). The pulse system designed by Nanjing Information Engineering University filters out high frequency noise by using a second order low-pass filter circuit whose cut-off frequency is 40 Hz (Wang, 2008). Filter circuits which mentioned above can be divided into four categories: High-pass filter circuit High-pass filter circuit combined with low-pass filter circuit Low-pass filter circuit combined with notch circuit Low-pass filter circuit ANALYSIS ABOUT FILTERS OF PULSE INSTRUMENTS The detection precision of useful signal can be improved on the condition that the lager DC component is filtered off because of the limited digits of AD converter. The requirements of high-pass filters are the smooth amplitude in pass band, low cut-off frequency point and narrow transitional band. Analysis about high-pass filter circuit: The simulation software AD09 (Altium Designer Summer 09) can be used to calculate and test the performance of filter circuit. It can do three types of simulation directly, i.e., analog, digital and mixed-signal circuit diagrams. AC small signal analysis in AD09 can be used to analyze amplitude frequency characteristic about three high-pass filter circuits in examples mentioned in section 2. The starting frequency is set at 0.01 Hz, the stop frequency is 40 Hz and the step of frequency is 0.01 Hz. The filter circuit of Example 1 is shown in Fig. 2a4. Its amplitude and phase frequency characteristic are shown in Fig. 3a by analysis of AD9 and its cut-off frequency of this high-pass filter is 0.143 Hz. The filter circuit of Example 2 and 3 are shown in Fig. 2b7 and 2c8, respectively. Their amplitude and phase frequency characteristics are shown in Fig. 3b and 3c. For Example 2, the low end cut-off frequency of high-pass filter circuit is 0.08 Hz and the upper cutoff frequency of the low-pass filter circuit is 40 Hz. And for Example 3, the pass band of the filter circuit is from 0.025 to 45 Hz, respectively. Analysis about transmission characteristics of pulse signal: According to the above analysis, the reasons which affect pulse signal have been found that highpass filter circuit affects the transmission of pulse 4464 Res. J. Apppl. Sci. Eng. Technol., T 4(21)): 4463-4468, 2012 2 mple 1 (a) Exam (bb) Example 2 (c) Example 3 Fig. 2: Filter circuits of threee examples 75.0 50.0 25.0 -00.00 -25.00 -50.00 -75.00 -100 10.00 20.00 30.000 Frequency (Hz) F 40.00 0.900 0.800 0.700 0.600 0.500 0.400 0.300 0.200 0.100 0.00 Voltage phase 10.00 (a) ( Example 1 ( Example 2 (b) 4.500 40.00 Voltage (V) 4.000 3.500 3.000 20.00 30.000 Frequency (Hz) F Volltage phase 20.00 0.00 2.500 -20.00 2.000 -30.00 1.500 0..00 10.00 20.00 30.00 Frequency (H Hz) ( Example 3 (c) Fig. 3: Diaggrams of amplitu ude and phase freequency characteeristics of three examples 4465 4 40.00 Phase (o) Voltagee phase 100 75.0 50.0 25.0 -00.0 -25.0 -50.0 -75.0 -100 40.00 Phase (o) 1.100 1.000 Voltage (V) 0.900 0.800 0.700 0.600 0.500 0.400 0.300 0.200 0.100 0.000 0.00 Phase ( ) Voltage (V) 1.000 Res. J. Apppl. Sci. Eng. Technol., T 4(21)): 4463-4468, 2012 2 1.00 Vin Vin ,Vout (V) 0.25 122 13 14 Time (s) 0.25 Vouut Vout -0.50 Vin -0.50 15 166 13 12 14 Time (s) (a) Example 1 16 (bb) Example 2 1.00 Vin (V) 15 4..00 Vin 1.000 0.25 Vout (V) Vin ,Vout (V) 1.00 Vout -22.00 -0.50 12 13 14 Time (s) 1 15 16 (c) Example 3 Fig. 4: Input and output sign nals of filter circcuits of three exaamples (b) Example 3 (a) Exaample 1 nals of exampless Fig. 5: Input and output sign signal. Wee should have a look at the pulse p input signnal and outputt signals throug gh high-pass fillter circuits. The pulse p instrumen nt referred in Example 1 has h been used in clinical praactice for 2 yeears. During thhis 5 patients’ pulsse signals. In thhis time, it hass collected 285 study, som me of the pulse signals have been b standardizzed by Matlab to signal sourcce which can be b used in circcuit simulation software. Thee difficulty aboout signal sourrce in this studdy has been sollved. Filter circuit should be drawn a ciircuit diagram in the AD09.. Piecewise lin near source VPWL V is used to o circuit. In the Part Field TAB of VPWL V input of componnents, the File Name should be filled in a File F of import data. Discrette data of pulse p signal can c be importeed to the circuit. Using this method, m compllicated data caan be simulatedd. Human pulsse signal is irrregular signal in one cycle, so this is an effective methhod to simulatte. Traansient analysiis of AD09 cann be used to sim mulate the circcuit. Its output signal is a funnction of time.. Input and outtput signals of filter circuits can c be obtainedd. Inpput and outpuut data of thrree high-pass filters circuitss are shown in Fig. 4. 4466 Res. J. Appl. Sci. Eng. Technol., 4(21): 4463-4468, 2012 Table 1: Parameters of input and output pulse signals and relative errors h2 High-pass filter circuit h1 Parameters of input signal 1 0.4536 0.9723 0.3006 Values of output Relative error (%) 2.7700 33.730 Values of output 1.0043 0.4319 Relative error (%) 0.4300 4.7800 Values of output (h1~h4/4.157) 1.0087 0.4474 Relative error (%) 0.8700 1.3670 h3 0.2176 0.2520 15.8100 0.2041 6.2000 0.2132 2.0200 h4 0.0300 0.1234 311.33 0.0340 13.3300 0. 0291 3.0000 t1 (s) 0.4922 0.4022 18.2900 0.4945 0.4800 0.4937 0.3000 t2 (s) 0.9022 0.9198 1.9500 0.9079 0.6300 0.9051 0.3200 t3 (s) 0.0978 0.0802 18.0000 0.0921 5.8300 0.0849 13.1900 The parameters of input and output pulse signal are compared in Table 1. The input pulse signal has been normalized as t = 1s and h1 = 1. Input and output pulse data are from AD09. The filter circuit in Example 3 can amplify input signal with a ratio about 4.157, so the output signal of Example 3 should be divided by that ratio. Relative error = (Output values - Input values) /Input values100% in Table 1. Fig. 6: Pulse diagram in an ideal situation CONCLUSION Analysis about actual circuits: Actual high-pass filter circuit of Example 1 is tested and the pulse signal comes from magnetoelectric pulse generator by which many kinds of pulse signals can be reappeared (Xu, 2008; Che, 2005). The magnetoelectric pulse generator is as an actual pulse to be input into the high-pass filter circuit and then input and output signals can be observed by a dual channel oscilloscope as shown in Fig. 5a where channel 1 is input signal and channel 2 is output signal of the high-pass filter circuit. The high-pass filter circuit of Example 3 is made. The pulse condition signal is input into the circuit by reference to the above. The results are shown in Fig. 5b where channel 1 is input signal of high-pass filter circuit and channel 2 is output signal. This filter circuit can amplify input signal with a ratio about 4.157. In the process of pulse detection, the filter circuit of processing pulse signal can filter out the press for getting pulse. At the same time, it also brings the distortion of pulse signal. The filter circuits have been simulated by using VPWL building signal source in AD09. Using the pulse data obtained by the simulation, the real circuits have been tested. The results of above two methods are similar. This research has solved the problems of test and analysis of pulse instruments in method and the need of reused pulse signal source. Pulse parameters: Pulse diagram can only be understood by sense organs. The parameterization is more useful for the classification of pulses by analyzing the pulse diagrams. In an ideal situation, pulse diagram tested by pulse instrument is shown in Fig. 6. In Fig. 6, h1 is main pulse amplitude, h2 is the amplitude before dicrotic pulse, h3 is the amplitude of dicrotic notch pulse and h4 is the amplitude of dicrotic pulse according to the description of pulse characteristic parameters. Where t1 is the time period of acute ejection period, t2 is the time period of systolic period and t3 is that of diastolic period. These parameters reflect the characteristics of pulse diagram. REFERENCES ACKNOWLEDGMENT This study is supported by the Shenyang Science & Technology Planning Item No. F10-213-1-00. Che, X.S., 2005. Magnetoelectric Pulse Generator. Practical New-type Patent, Patent Number: ZL200520089617.1. Che, X.S., Z.X. Zhang and Y.J. He, 2005. The acquisition and direct description of pulse in Chinese medicine. ICEMI’ 2005 Conf. Proc., 1(6): 49-52. Fan, W.F., 2007. Research on multi-dimensional collection of Chinese medicine pulse. MA. Thesis, Shenyang University of Technology, Shenyang. 4467 Res. J. Appl. Sci. Eng. Technol., 4(21): 4463-4468, 2012 Gen, J.J., 2007. Advanced development of acquisition system on pulse tracings of traditional Chinese medicine. MA. Thesis, Tianjin University, Tianjin. Lu, X.Z., G.Q. Hu and H.M. Cao, 2010. Research progress and prospect of pulse condition instrument. The 11th National Academic Conference on TCM Diagnosis, Beijing, China, pp: 74-78. Li, Z.L., 2008. Research on pulse diagnosis instrument based on extension detecting. MA. Thesis, Guangdong University of Technology, Guangzhou. Wang, Y.B., 2008. The portable instrument of multiple pulse. MA. Thesis, Northeast Normal University, Changchun. Xu, R.Q., 2008. Design of pulse detection system and research of pulse signal processing algorithm. MA. Thesis, Nanjing Information Engineering University, Nanjing. Zhao, C.M., X.Z. Yang and Y.L. Song, 2006. The development of TCM pulse pattern diagnostic system. China J. Tradit. Chin. Med. Pharm., (supplement), pp: 260-262. Zhu, W.F., 1995. Diagnostics of Traditional Chinese Medicine. Shanghai Science and Technology Press, Shanghai. 4468