ab119631 PhosphoTracer I-κBα (pS32/36) ELISA Kit Instructions for Use

advertisement
ab119631
PhosphoTracer
I-κBα (pS32/36) ELISA Kit
Instructions for Use
For the semi-quantitative measurement of I-κBα
(pS32/36) concentrations in cell culture extracts
This product is for research use only and is not
intended for diagnostic use.
Version 2 Last Updated 19 September 2013
1
Table of Contents
1.
Introduction
3
2.
Assay Principle
7
3.
Assay Summary
8
4.
Kit Contents
9
5.
Storage and Handling
15
6.
Buffer Preparation
16
7.
Lysate Preparation
19
8.
PhosphoTracer Assay Protocol
22
9.
Data Analysis
24
10. Procedure Limitations
25
11. Technical Hints and Troubleshooting
26
2
1. Introduction
I‐κB proteins are present in the cytosol bound to NF‐κB/Rel
transcription factors, forming inactive complexes. Activation of
NF‐κB/Rel occurs via phosphorylation of I‐κB, which targets I‐kB for
proteasome‐mediated
degradation,
releasing
NF‐κB/Rel
for
phosphorylation and nuclear import.
Inflammatory mediators
Inflammatory cytokines
IKKA
IKKB
IKKE
Phosphorylation
(multiple kinases)
I-κBα
NF-κB
p65
NF-κB
p65
I-κBα
Ubiquitination
Degradation
NF-κB
p65
Nuclear import
I-κBα phosphorylation is stimulated by many extracellular signals,
such as inflammatory cytokines, growth factors and chemokines.
Phosphorylation of I-κBα at Ser32/36 is essential for the release of
active NF-κB, so phosphorylation at this site is a good marker of NFκB activation.
3
Specificity:
Abcam’s PhosphoTracer I-κBα assay kits detect endogenous levels
of I-κBα (GenBank Accession NP_065390) in cellular lysates. Total
I-κBα assay kits detect I-κBα irrespective of phosphorylation status.
Species cross-reactivity:
Tested: Human
Expected: Mouse, Rat
Other species should be tested on a case-by-case basis.
4
Figure 1. As shown, using the I-κBα assay kits, a significant
stimulation of I-κBα phosphorylation at Ser32/36 is detected in HeLa
cells treated with TNFα for 5 minutes compared with untreated cells,
while a change in total I-κBα levels is also observed over this time
period.
5
Figure 2. Day 1: HeLa cells were seeded at 25K cells/well in a
96 well tissue culture plate in medium containing 10% FBS, and
incubated overnight. Day 2: The media was removed and cells were
stimulated with various concentrations of TNFα for 10 minutes. The
media was removed, and the cells were lysed with 125 μL/well
freshly prepared Lysis Mix, with shaking for 10 min. Lysates (50 μL)
were transferred to replicate wells of a PhosphoTracer assay plate
and analyzed for either phospho or total I-κBα using the standard
PhosphoTracer protocol. Briefly, Antibody Mix specific for either
phospho-I-κBα, or Total I-κBα (50 μL/well), was added to the lysate,
and the plate was incubated for 1 hr at room temp, with shaking. The
plate was washed and Substrate Mix was added to the wells. The
plate was covered in foil and incubated for 10 min with shaking.
Signal in the wells was determined using a plate reader. After as
short a time as 10 minutes, phosphorylation-induced depletion of
I-κBα levels are observed.
6
2. Assay Principle
PhosphoTracer assays use a traditional immuno-sandwich format,
but with a major difference. For PhosphoTracer assays, both the
analyte and the assay reagents are added to the PhosphoTracer
assay microplate at the same time. After a short incubation period,
unbound assay reagents and analytes are washed away, and
immuno-complexes containing both antibodies are detected. The
process can take as little as 60 minutes to complete.
PhosphoTracer also allows a higher degree of assay flexibility. In
contrast to other ELISA formats, no antibodies are present on the
assay microplate itself, so assays for several different targets can be
performed in different wells on the same microplate. Simply mix the
lysate with your target reagents of choice, using the microplate
configuration of your choice. For example, a cellular lysate can be
analyzed on the same assay microplate in parallel for p38-MAPK
phosphorylation, ERK phosphorylation, and JNK phosphorylation,
giving fast, accurate and quantifiable information on key cell
signaling events.
A whole new way of performing cellular assays, PhosphoTracer
takes the hard work out of running a standard ELISA, while still
giving
the
high
quality
results
expected
from
a
sandwich
immunoassay. Fully self-contained kits are supplied in convenient
96-well packs. Simple to use and highly sensitive, PhosphoTracer
kits are designed to get results, fast.
7
3. Assay Summary
Add lysates to PhosphoTracer assay microplate.
Add Antibody Mix to microplate, incubate 1 hour.
Wash microplate.
Add Substrate Mix to microplate. Cover with foil, incubate
10 minutes.
Read microplate
8
4. Kit Contents
Item
Lysis Buffer (5X)
Enhancer Solution
Quantity
1 x 15 mL
1 x 1 mL
ADHP Dilution Buffer
1 x 15 mL
ADHP (100X)
1 x 120 µL
Wash Buffer (10X)
1 x 15 mL
Stop Solution
Assay Control Lysate (lyophilized)
1 x 2 mL
1 x 0.25 mL
96-well PhosphoTracer assay plate (stripwell)
1
Adherent plate seals
2
9
Target Specific Reagents
For
For
For
For
ab119674
ab119680
ab119614
ab119654 to
to
ab119673
to
ab119679
Number of
ab119653
vials per
Number of
target
vials per
target
Capture
1 x 1 mL
1 x 1.5 mL
1 x 3 mL
Antibody
3 x 0.75 mL
(phospho)
Reagent
1 x 0.75 mL
(total)
Detection
Antibody
1 x 1 mL
1 x 1.5 mL
1 x 3 mL
3 x 0.75 mL
(phospho)
Reagent
1 x 0.75 mL
(total)
10
PhosphoTracer assay microplates
Only use PhosphoTracer assay microplates for PhosphoTracer
assays. PhosphoTracer assay microplates are specifically designed
to work with PhosphoTracer assays, and cannot be substituted with
other 96-well microplates. The PhosphoTracer assay microplate is
highly optimized to form an active surface for the capture of immunocomplexes that are formed as part of the PhosphoTracer assay
procedure.
After opening, the PhosphoTracer microplate can be stored at 4°C.
The microplate should be allowed to equilibrate to room temperature
prior to opening the pouch, to minimize condensation forming in the
wells.
PhosphoTracer assay microplates are supplied in convenient 8-well
strips. Only use the amount of strips required for the assay. Remove
strips from the frame that are not needed, place back in their foil
pouch, seal with tape, and store at 4°C. Unused wells should be
used within 1 month of opening the microplate foil bag.
Antibody Mix
The Antibody Mix is comprised of a 1:1 mixture of the Capture
Antibody Reagent and the Detection Antibody Reagent, and contains
all of the antibodies necessary for the assay.
11
The Capture Antibody Reagent contains the capture antibody, and
the Detection Antibody Reagent contains the detection antibody. The
antibody reagents need to be mixed prior to use, to form the
Antibody Mix.
The Antibody Mix can be prepared by adding an equal volume of
Capture Antibody Reagent and Detection Antibody Reagent, and
mixing by inversion prior to each experiment. Alternatively, simply
combine and mix all of the contents of the Capture Antibody Reagent
and Detection Antibody Reagent before the first use of the kit, to
yield 6 mL of ready-to-use Antibody Mix.
Lysis Mix
The Lysis Mix is a combination of the Lysis Buffer and Enhancer
Solution. The 5X Lysis Buffer contains a combination of detergents,
phosphatase inhibitors, salts and buffers. 5X Lysis Buffer is
supplemented with Enhancer Solution to yield a versatile lysis
solution that can be applied to many cells and tissues.
The Lysis Mix Concentrate is used to lyse cells in the presence of
culture medium, and is typically used to lyse non-adherent cells.
The Lysis Mix is used to lyse cells after the removal of culture
medium, and is typically used to lyse adherent cells, or non-adherent
cells that have been harvested by centrifugation. Lysis Mix is
prepared by simply diluting Lysis Mix Concentrate 5-fold with water.
12
Lysis Mix should be used as the diluent for any dilution of cellular
lysates that are required.
Supplementing Lysis Mix with extra components (e.g. protease
inhibitors, chelating agents, detergents) should be tested on a caseby-case basis for compatibility with PhosphoTracer assays.
Assay Control Lysates
Assay Control Lysates are prepared from various cell types, which
have been cultured and prepared to optimize the activation of the
intracellular pathway of interest.
The Assay Control Lysates are intended for use as an assay
positive control only, and should not be used for the absolute
quantification of a particular protein or phosphorylated target. In
combination with negative control wells containing Lysis Mix only,
the Assay Control Lysates can be used to give an indication of the
expected signal range for a given assay.
The controls are supplied lyophilized, and should be reconstituted
with 250 μL of either dd H2O or high quality purified H2O. If required,
control lysates can be further diluted with Lysis Mix, and frozen in
aliquots for subsequent use.
13
Wash Buffer
The Wash Buffer, supplied as a 10X concentrate, is used for
washing the PhosphoTracer assay microplate. It is a simple mix of
buffer, salts and mild detergent. Alternatively, a PBS/0.05% (v/v)
Tween-20 solution may be substituted as a wash solution.
If washing wells with a microplate washer, use 3x washes with a 10
second mixing cycle, with a PBS/0.05% (v/v) Tween-20 wash
solution.
Substrate Mix
The
Substrate
Mix
is
a
combination
of
10-Acetyl-3,7-
dihydroxyphenoxazine (ADHP), a sensitive substrate for HRP, and
ADHP Dilution Buffer, a stabilized H2O2 solution. In the presence of
H2O2, HRP converts ADHP to the fluorogenic product, resorufin.
Resorufin is a highly fluorescent molecule, and should be used with
excitation filters in the range of 530-540nm, with bandwidths ≤ 30nm.
The emission filter should be in the range of 590-600nm, with
bandwidths ≤ 30nm.
The signal in the wells should be developed for around 10 minutes.
Best results will be obtained if the microplates are developed in the
dark, e.g. by covering the microplate with foil.
14
Stop Solution
The Stop Solution is used for stopping HRP-mediated conversion of
ADHP to resorufin. Once added to the wells, signal in the wells is
stable for 60 minutes. However, ADHP remaining in the wells is still
light-sensitive, and should be handled under low-light conditions until
the microplate is read.
5. Storage and Handling

Unopened kit: Store all kit reagents at 4°C. Do NOT freeze
the kits.

Opened/Reconstituted: Store at 4°C after opening, until
expiry date. For the PhosphoTracer assay microplate:
Return unused wells to foil pouch with desiccant pack, and
seal with tape. May be stored at 4°C for up to 1 month after
opening. For the assay control lysate: aliquot and store at
-20°C for up to 1 month.
15
6. Buffer Preparation
*Note: Bring all reagents to room temperature before use. Avoid
vortexing the Capture Antibody Reagent or Detection Antibody
Reagent, as vigorous mixing can damage some antibodies.
Item
Instructions
Enhancer Solution
Supplied as a concentrate.
Precipitation will occur during
storage at 4°C, which is normal.
To re-dissolve, warm to 37°C
and mix by inversion prior to use.
Alternatively, Enhancer Solution
can
be
stored
at
room
temperature with no loss in
activity.
Lysis Mix Concentrate
Prepare Lysis Mix Concentrate
by diluting Enhancer Solution 10fold in 5X Lysis Buffer (e.g. mix
900 µL 5X Lysis Buffer and 100
µL Enhancer Solution). Lysis Mix
Concentrate is used to lyse cells
directly in cell culture medium.
Prepare immediately prior to
use. Discard unused Lysis Mix
Concentrate.
16
Item
Instructions
Antibody Mix* (Capture Antibody Prior to each experiment,
Reagent + Detection Antibody
prepare by adding an equal
Reagent)
volume of Capture Antibody
Reagent and Detection Antibody
Reagent,
and
mixing
by
inversion. Prepare enough to
use 50 µL/well (e.g. for a single
8-well strip, prepare 400 µL
Antibody Mix).
Wash Buffer
Dilute 10X Wash Buffer 10-fold
with either dd H2O or high quality
purified H2O. (e.g. mix 1 mL of
10X Wash Buffer and 9 mL
water). Prepare enough for 3
separate exchanges with 200 µL
of 1X Wash Buffer per well.
Assay Control Lysates
The controls are supplied
lyophilized, and should be
reconstituted with 250 μL of
either dd H2O or high quality
purified H2O. If required, control
lysates can be further diluted
with Lysis Mix.
Once reconstituted, aliquot and
store at -20°C for up to 1 month.
17
Substrate Mix
Allow ADHP to thaw prior to use
– it is solid at 4°C.
Avoid preparation or use under
bright lights or direct sunlight –
ADHP is sensitive to bright light.
To prepare the Substrate Mix,
dilute ADHP 100-fold with ADHP
Dilution Buffer A (e.g. add 20 μL
ADHP to 2 mL of the ADHP
Dilution Buffer) immediately prior
to use. The Substrate Mix is then
ready to use. Prepare enough to
use 100 µL/well.
For best results we recommend
preparing the Substrate Mix
immediately prior to use. It is not
stable for long periods once
mixed.
18
7. Lysate Preparation
These lysate preparations require the preparation of cellular
lysates in a tissue culture microplate. The lysate is subsequently
transferred to a PhosphoTracer microplate for the assay. The
exception to this is the all-in-one well assays which avoid the
lysate transfer steps.
Adherent cells
1. Remove any media and cellular treatments from the
cells – optionally, wash cells with PBS.
2. For cells cultured in 96-well microplates, lyse the cells
with 100 μL* of freshly prepared Lysis Mix, with shaking
(~300 rpm) at room temp for 10 minutes.
*Lysis volume should be adjusted depending on the
desired lysate concentration. Lysates in the range of
0.1 - 0.5 mg/mL protein are usually sufficient. However,
preparing more concentrated lysates can help with the
detection of low abundance analytes.
Proceed to Section 8.
19
Non-adherent cells
1. Harvest cells by centrifugation, and resuspend at an
appropriate density in HBSS* containing 5% FBS. A cell
density
that
yields
cellular
lysate
at
a
protein
concentration of 0.1 - 0.5 mg/mL is suitable for many
cell lines.
*Avoid resuspending cells in cell culture medium
for the assay, as components of common media
can interfere with the assay. For example RPMI
contains biotin that will interfere with the assay. If
the use of media is necessary for the cells, after
treatment
the
cells
can
be
harvested
by
centrifugation, and washed with PBS or HBSS prior
to lysis.
2. Return cells to a 37°C incubator for 1-2 hours. For
certain pathways, this can allow handling-mediated
pathway activation to subside.
This step is optional,
and depends on the activation status of your cells
following re-suspension.
3. At the completion of the treatment, lyse cells with 1/5th
final volume of Lysis Mix Concentrate, with shaking
(~300 rpm) at room temp for 10 min (e.g. for 40 µL of
cells, use 10 µL of Lysis Mix Concentrate).
20
Alternatively
cells
can
be
harvested
by
centrifugation and lysed with Lysis Mix.
Proceed to Section 8.
All-in-one well assays
This method avoids lysate transfer steps
1. Harvest cells by centrifugation, and resuspend at an
appropriate density in HBSS* containing 5% FBS. A cell
density that yields 10,000-25,000 cells/well is suitable for
the analysis of many cell lines.
*Avoid resuspending cells in cell culture medium for
the assay, as components of common media can
interfere with the assay. For example RPMI contains
biotin that will interfere with the assay. If the use of
media is necessary for the cells, after treatment the
cells can be harvested by centrifugation, and
washed with PBS or HBSS prior to lysis.
2. Determine desired number of PhosphoTracer microplate
strips. Remove unused strips from frame and return to
storage pouch and seal.
3. Add 20 µL cells/well to the PhosphoTracer assay
microplate.
21
4. Return cells to a 37°C incubator for 1 - 2 hours. For
certain pathways, this can allow handling-mediated
pathway activation to subside. This step is optional, and
depends on the activation status of your cells following
re-suspension.
5. Add 20 µL treatment/well to cells (e.g. 2x agonists
and/or antagonists) for the desired time period. The final
volume in wells prior to lysis should be 40 µL.
6. At the completion of the treatment, lyse cells with 10 µL
Lysis Mix Concentrate, with shaking (~300 rpm) at room
temp for 10 minutes.
Proceed to Section 8, step 3.
8. PhosphoTracer Assay Protocol
For All-in-one well assays proceed directly to step 3.
1. Determine
desired
number
of
PhosphoTracer
microplate strips. Remove unused strips from frame
and return to storage pouch and seal.
22
2. Add 50 µL/well of lysate to the PhosphoTracer
microplate.
3. Add 50 µL/well Lysis Mix (negative control) and Control
Lysates (positive control) to separate wells for assay
controls if desired.
4. Add 50 µL/well of Antibody Mix to the wells. Cover the
microplate with adhesive seal and incubate for 1 hr at
room temp on a microplate shaker (~300 rpm).
5. Wash wells with 200 µL/well 1X Wash Buffer (repeat 3
times). After final wash, remove any remaining wash
solution from wells.
6. Immediately prior to use, prepare Substrate Mix. Add
100 µL/well of Substrate Mix. Cover microplate with foil,
and incubate for 10 minutes at room temp on a
microplate shaker (~300 rpm).
Substrate Mix should be prepared immediately prior
to use, preferably while microplate is washing, but
do not allow microplate to dry. If necessary, leave
microplate in 1X Wash Buffer for up to 30 minutes,
until Substrate Mix has been prepared.
23
7. Add 10 µL/well Stop Solution, and mix briefly
(5 - 10 sec) on a microplate shaker.
8. Read fluorescence signal with a PhosphoTracer
compatible filter set.
9. Data Analysis
To analyze the data, calculate the averaged counts for untreated and
treated cells. We recommend using at least duplicate wells (n=2) to
calculate a response.
Dose response and dose inhibition curves can be fitted to 4
parameter
non-linear
regression
equations.
These
types
of
regression analyses output key parameters such as EC50 (or IC50),
Min and Max signals, and Hillslope factors.
Ensure that samples readings are within the linear range of the
assay. This can vary based on reader performance, and analyte
concentration. If a lysate sample generates a signal outside the
linear range, the lysate samples should be diluted with Lysis Mix and
re-assayed. Absolute fluorescent units will vary from reader to
reader.
24
10.

Procedure Limitations
Variations in general operator-related procedures, such as
pipetting, washing, and incubation times, can cause variation
in the final signal.

The assay is designed to work for the detection of
endogenous cellular proteins across a wide variety of cell
lines. However, until each cell line in particular is tested, the
possibility of the presence of interfering factors cannot be
excluded.

Users should ensure that their cell line has measurable
levels of the pathway of interest. Expression levels of
signaling proteins in different cell types vary widely.
25
11.

Technical Hints and Troubleshooting
General cell handling
Cells should be harvested from flasks for seeding into
microplates when approximately 70-90% confluent. The cells
should be detached from the flasks using mild conditions
(such as trypsin-free cell dissociation solutions where
possible), accurately counted, and diluted to the appropriate
density in fresh media. If using adherent cells, allow time for
cells to regain full signaling capacity after harvesting.

Compatible cell types
The assay can be used for many adherent and non-adherent
cell types, including transfected cell lines and primary cells.
However, because kinase expression and phosphorylation
conditions can vary from one cell line to another, some cells
may be more amenable to particular assays than others.
Parameters such as stimulation time and cell number should
be optimized for each cell line used.
Cells over-expressing a receptor of interest have been
shown to elicit good phosphorylation responses. When using
over-expressed intracellular targets, ensure the expressed
target is full-length to ensure correct binding of assay
antibodies. The concentration of cell lysate should be
26
optimized to ensure the signal is within the working range of
the assay.

Lysate viscosity and handling
Lysing cells with Lysis Mix can yield lysates that are viscous
and difficult to handle, particularly when concentrated
lysates are required. Care should be taken when transferring
lysates, to minimize pipetting-related variability.
To avoid this problem when using non-adherent cells, the
cells can be transferred to the PhosphoTracer assay
microplate immediately prior to lysis, and subsequently lysed
directly in the assay microplate (e.g. transfer 40 µL cells/well
to an PhosphoTracer assay microplate, and lyse with 10 µL
5X Lysis Mix Concentrate – see Kit Contents for
preparation).
Alternatively, use the all-in-one-well assay procedure.

Assaying for multiple targets from a single lysate
One of the features of PhosphoTracer protocols is the use of
a common assay microplate for all assays. Therefore, a
cellular lysate can be transferred to several replicate wells,
and analyzed in parallel. This can enable a particular lysate
to be sampled for both total and phosphorylated target on
27
the same assay microplate, or for several assay targets on
the same microplate.

Assay scalability
The PhosphoTracer assay is optimized for equal 50 μL
volumes of both the analyte and the Antibody Mix. However,
in cases where lysate is limiting, as little as 25 μL of lysate
can be used, in combination with 50 μL of Antibody Mix.

Assay incubation times
The general assay incubation times that are recommended
are 1 hour after the addition of Antibody Mix, and 10 minutes
for
the
development
of
the
Substrate
Mix.
Longer
incubations (up to overnight) may be more convenient for
certain assays, and can enhance sensitivity in some cases.

Buffer compatibility
PhosphoTracer assays are compatible with most cell culture
media and reagents, however there are some exceptions.
Media that contain biotin (e.g. RPMI) will reduce assay
sensitivity. When it is necessary to use a media such as
RPMI for growing cells, they should be harvested and resuspended in HBSS or a similar buffer, prior to the assay.
28
Supplementing Lysis Mix with extra components (e.g.
protease inhibitors, chelating agents, detergents) is typically
okay, but should be tested on a case-by-case basis for
compatibility with PhosphoTracer assays. For example,
strong detergents such as SDS can denature proteins, and
is not recommended for use with PhosphoTracer assays.

Lysis Buffer Formulation
1X
Lysis
buffer
contains
the
following
phosphatase
inhibitors:
 Sodium pyrophosphate
 Sodium fluoride
 Sodium orthovanadate
The addition of other components to the Lysis Mix should be
tested on a case-by-case basis.

Phosphoprotein analysis
Typical analysis of phosphoproteins will involve analysis of
the up regulation of a phospho-epitope on a protein, in
conjunction with analysis of the protein itself. This ensures
that any increase in signal is due to specific phosphorylation,
and not just accumulation of the protein itself.
29
Demonstration of specific phosphorylation of a particular
protein or pathway can be typically achieved through:
 Use of specific agonists that result in activation
(phosphorylation) of the pathway/protein of interest.
 Use of specific antagonists that prevent activation or
deactivate the pathway/protein of interest.
Assay Control Lysates prepared from various cell types are
supplied with each kit. The Assay Control Lysates are
intended for use as an assay positive control only, and
should not be used for the absolute quantification of a
particular protein or phosphorylated target. In combination
with negative control wells containing Lysis Mix only, the
Assay Control Lysates can be diluted to give an indication of
the expected signal range (high/medium/low) for a given
assay.

Assay Specificity
PhosphoTracer assays are sandwich immunoassays that
use both an antibody against a specific phospho-epitope,
and a distal epitope on the same protein. Therefore, only
analyte that has both antibodies bound can contribute to
signal.
30
All antibodies are rigorously tested for specificity and
performance. Furthermore, the use of a 2-antibody format
gives extra assurance against the possibility of signal
generated from non-specific binding.

PhosphoTracer assay microplates
PhosphoTracer
assay
microplates
are
specific
to
PhosphoTracer assays, and cannot be substituted with other
96-well microplates. The PhosphoTracer assay microplate is
highly optimized to form an active surface for the capture of
immuno-complexes
that
are
formed
as
part
of
the
PhosphoTracer assay procedure. PhosphoTracer assay
microplates are supplied in 8-well strips. Remove strips from
the frame that are not needed, place back in their foil pouch,
seal with tape, and store.
After opening, the PhosphoTracer microplate can be stored
at 4°C. The microplate should be allowed to equilibrate to
room temperature prior to opening the pouch, to minimize
condensation forming in the wells. Unused wells should be
used within 1 month of opening the microplate foil bag.
31

ADHP HRP substrate
The HRP substrate used in PhosphoTracer assays is a
combination
of
10-Acetyl-3,7-dihydroxyphenoxazine
(ADHP), a sensitive substrate for HRP, and ADHP Dilution
Buffer, a stabilized H2O2 solution. In the presence of H2O2,
HRP converts ADHP to the fluorogenic product, resorufin.
Resorufin is a highly fluorescent molecule, and should be
used with excitation filters in the range of 530 - 540nm, with
bandwidths ≤ 30nm. The emission filter should be in the
range of 590 - 600nm, with bandwidths ≤ 30nm.
We recommend that the HRP-mediated generation of
resorufin be determined by fluorescence – this is the most
sensitive means of determining resorufin concentration.
However, if users do not have access to a reader that can
measure fluorescence, then resorufin can also be detected
by absorbance at 570nm, albeit with a lower sensitivity. We
estimate that there is approximately 1-1.5-log lower
sensitivity therefore use more lysate per well and develop for
longer(20 - 30 mins).
The signal in the wells should be developed for around 10
minutes. Best results will be obtained if the microplates are
developed in the dark, e.g. by covering the microplate with
foil. The Stop Solution provided can be used for stopping
HRP-mediated conversion of ADHP to resorufin. Once
32
added to the wells, signal in the wells is stable for 60
minutes. However, ADHP remaining in the wells is still lightsensitive, and should be handled under low-light conditions
until the microplate is read.

Cell Lysis
The PhosphoTracer Lysis buffer that is supplied, is a 5X
Lysis stock containing a combination of detergents,
phosphatase inhibitors, salts and buffers. 5X Lysis Buffer is
supplemented with Enhancer Solution to yield a versatile
lysis solution that can be applied to many cells and tissues.
Supplementing Lysis Mix with extra components (e.g.
protease inhibitors, chelating agents, detergents) should be
tested on a case-by-case basis for compatibility with
PhosphoTracer assays.
Lysing cells can yield lysates that are viscous and difficult to
handle, particularly when concentrated lysates are required.
Care should be taken when transferring lysates, to minimize
pipetting-related variability.
33
To avoid this problem when using non-adherent cells, the
cells can be transferred to the PhosphoTracer assay
microplate immediately prior to lysis, and subsequently lysed
directly in the assay microplate (e.g. transfer 40 µL cells/well
to an PhosphoTracer assay microplate, and lyse with 10 µL
5X Lysis Mix Concentrate - see PhosphoTracer product
manual for details).
For further technical questions please do not hesitate to
contact us by email (technical@abcam.com) or phone (select
“contact us” on www.abcam.com for the phone number for
your region).
34
UK, EU and ROW
Email: technical@abcam.com | Tel: +44-(0)1223-696000
Austria
Email: wissenschaftlicherdienst@abcam.com | Tel: 019-288-259
France
Email: supportscientifique@abcam.com | Tel: 01-46-94-62-96
Germany
Email: wissenschaftlicherdienst@abcam.com | Tel: 030-896-779-154
Spain
Email: soportecientifico@abcam.com | Tel: 911-146-554
Switzerland
Email: technical@abcam.com
Tel (Deutsch): 0435-016-424 | Tel (Français): 0615-000-530
US and Latin America
Email: us.technical@abcam.com | Tel: 888-77-ABCAM (22226)
Canada
Email: ca.technical@abcam.com | Tel: 877-749-8807
China and Asia Pacific
Email: hk.technical@abcam.com | Tel: 108008523689 (中國聯通)
Japan
Email: technical@abcam.co.jp | Tel: +81-(0)3-6231-0940
www.abcam.com | www.abcam.cn | www.abcam.co.jp
35
Copyright © 2013 Abcam, All Rights Reserved. The Abcam logo is a registered trademark.
All information / detail is correct at time of going to print.
Download