Photomultiplier Tube Calibration for the Cubic Meter Dark Matter Time Projection Chamber by Kevin Burdge Submitted to the Department of Physics in partial fulfillment of the requirements for the degree of Bachelor of Science in Physics I LCI) Coo LI < C: z .I-c at the MASSACHUSETTS INSTITUTE OF TECHNOLOGY June 2015 @ Massachusetts Institute of Technology 2015. All rights res erved. redacted e AuthorSignature A u th o r ................................................................ Department of Physics May 8, 2015 redacted Signature . . . . .. . . . . . . . . C ertified by ............................... Peter Fisher Department Head, Professor Thesis Supervisor Accepted by ................. Signature redacted ...... .. . .. ... ................... Nergis Mavalvala Associate Department Head for Education () -i 2 Photomultiplier Tube Calibration for the Cubic Meter Dark Matter Time Projection Chamber by Kevin Burdge Submitted to the Department of Physics on May 8, 2015, in partial fulfillment of the requirements for the degree of Bachelor of Science in Physics Abstract This thesis concerns measurements I performed on photomultiplier tubes (PMTs) and lenses to be used in the Cubic Meter Dark Matter Time Projection Chamber (DMTPC) experiment. DMTPC is a new generation of detector, which takes the idea of a standard time projection chamber and adds in some additional optical elements, such as CCDs and PMTs. The goal of DMTPC is the directional detection of the dark matter. During the course of my measurements, I characterized both the absolute gains of DMTPC's eight PMTs, as well as the dark currents exhibited by each of the PMTs. Seven of the eight PMTs demonstrated gains on the order of 106-107, and one PMT did not function at all. Of the seven working PMTs, six of them had dark currents under 10 kHz, and one had an excessively high dark current over 10 kHz. These gain values for the PMTs will give DMTPC the means to measure the Z dimensions of the particle tracks it intends to image, and thus when combined with the information from the CCDs will allow for full track reconstruction. DMTPC will use lenses on their CCD cameras, and I also measured the transparency of these lenses, and discovered that they are opaque below approximately 350nm. These measurements will be essential for DMTPC, because they will provide information about the relative amounts of light the PMTs and CCDs on the detector will register, and thus provide key information for track reconstruction. Thesis Supervisor: Peter Fisher Title: Department Head, Professor 3 4 Acknowledgments I would like to thank Professor Peter Fisher for his incredible mentoring, guidance, and ability to inspire. Additionally, I would like to thank Ross Corliss for helping edit this thesis, and all of his feedback and guidance on the research presented here. Also, I would like to thank Hidefumi Tomita and Michael Leyton for their mentorship in the laboratory. Finally I would like to thank Cosmin Deaconu for his explanations and assistance with my work, and Gabriela Druitt, Natalia Guerrero, and Evan Zayas for directly assisting in some of the measurements and research described in this thesis. 5 6 Contents . . . . . . . . . . . . . 13 1.2 Weakly Interacting Massive Particles . . . . . . . . . . . . . 15 1.3 Dark Matter Wind . . . . . . . . . . . . . . . . . . . . . . 16 1.4 Dark Matter Detection . . . . . . . . . . . . . . . . . . . . 16 . . . . . Time Projection Chambers . . . . . . . . . . . . . . . . . . . . . . 19 2.2 DMTPC's Chamber . . . . . . . . . . . . . . . . . . . . . . . . . . 20 2.3 PhotoMultiplier Tubes . . . . . . . . . . . . . . . . . . . . . . . . 21 . . . 2.1 . 25 Calibrating PhotoMultiplier Tubes 25 3.1.1 Light-Tightness . . . . . . . . . . . . . . . . . . . . . . . . . 26 3.1.2 PicoQuant Trigger . . . . . . . . . . . . . . . . . . . . . . . 26 3.1.3 Gain Results . . . . . . . . . . . . . . . . . . . . . . . . . . 28 Dark Current . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 30 Dark Current Measurement Results . . . . . . . . . . . . . . 31 . . . . . . . . . . . . . . . . . . . . . . . . . . . . PMT Gain Measurements 3.2.1 . 3.2 33 Optical Response to Electron Avalanches The CF 4 Secondary Scintillation Spectrum . . . . . . . . . . . . . . 33 4.2 The PMT response . . . . . . . . . . . . . . . . . . . . . . . . . . . 34 4.3 The CCD Response . . . . . . . . . . . . . . . . . . . . . . . . . . . 35 4.4 Total Responsivity . . . . . . . . . . . . . . . . . . . . . . . . . . . 35 . . . . 4.1 39 Results and Conclusions 5.1 U ncertainties . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 39 5.2 PMT Results . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 39 . 5 19 The Dark Matter Time Projection Chamber 3.1 4 . Dark Matter ................. . 3 1.1 . 2 13 Introduction . 1 7 5.3 Transparency Results . . . . . . . . . . . . . . . . . . . . . . . . . . . 41 5.4 Conclusions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 41 43 A Code 8 List of Figures 1-1 Galactic Rotation Curve of Milky Way . . . . . . . . . . . . . . . . . 14 1-2 Example of gravitational lensing in Abell 1689 . . . . . . . . . . . . . 15 2-1 Sketch of a time projection chamber . . . . . . . . . . . . . . . . . . . 20 2-2 Sketch of a PM T . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 22 3-1 Dark Current of PMT PADM with room lights off . . . . . . . . . . . 26 3-2 Dark Current of PMT PADM with room lights on . . . . . . . . . . . 27 3-3 Block diagram of gain and dark current measurement setup . . . . . . 28 3-4 PMT PGEJ responding to light from PicoQuant . . . . . . . . . . . . 29 3-5 Example of baseline electronic noise in PMT PGEJ . . . . . . . . . . 30 3-6 Histogram of Peak Amplitudes of PMT . . . . . . . . . . . . . . . . . 31 3-7 Dark current in PMT PGEJ . . . . . . . . . . . . . . . . . . . . . . . 32 4-1 Secondary Scintillation Spectrum of CF4 . . . . . . . . . . . . . . . . 34 4-2 Pressure Dependence of the Secondary Scintillation Spectrum of CF 4 35 4-3 Quantum Efficency of Hamamatsu R1408 PMTs . . . . . . . . . . . . 36 4-4 Transmission of Light through X-ray and Optical Lenses . . . . . . . 37 5-1 PMT PACI not responding to light from PicoQuant . . . . . . . . . . 40 9 10 List of Tables Table of PMT gains and dark current . . . . . . . . . . . . . . . . . . 5.1 11 40 12 Chapter 1 Introduction Of the four forces that scientists observed in the universe, gravity was the first to be described, through Newton's law of gravitation. Because of gravitational influence, astronomers understood the motion of heavenly bodies, and consequently made discoveries such as the planet Neptune. After using gravity's predictive power to document the motion of the planets, astronomers observed orbits that did not exactly agree with predictions, such as the precession of Mercury. This particular observation was ultimately accounted for in a new, revised theory of gravity-Einstein's general relativity. Then, astronomers, in their effort to understand the cosmos, encountered a new gravitational anomaly-one that is still an active topic of research. We call this anomaly dark matter. 1.1 Dark Matter Quantum mechanics allowed physicists to understand atomic spectra in great detail. In this discussion, one of the most important spectral lines is the 21cm line-produced by the hyperfine transition in hydrogen. The line is incredibly common in the galaxy because of the presence of interstellar hydrogen gas throughout it. Because the transition is forbidden by selection rules, it has a very long lifetime, and photons emitted by the transition are unlikely to interact with other matter they pass through (except for man-made antennas). This spectral line, with sources distributed throughout the 13 galaxy, allowed astronomers to observe the Milky Way in new ways. By observing 21cm photons emitted from interstellar hydrogen gas, and their Doppler shift, physicists such as Vera Rubin measured galactic velocity distribution curves [8]. Figure 1-1 on page 14 illustrates such a curve. After analyzing galactic W. W. SHANE AND o. P. SIEGER-SMITH 30' 1.0 SO- 60' 339* 320* 310- 300* km/secl 260 240 220 5 7 65 9 R, Ikpc) 9 Figure 1-1: This is an example of a rotation curve of the Milky Way. The curve is a plot of velocity in km/s vs orbital radius in kpc. Taking into account only the visible mass in the galaxy, Newtonian gravity and general relativity both predict a decreasing velocity of stars as the galactic radius increases, but as seen in the figure, observations suggest the opposite. Adapted from [9]. velocity distribution curves from numerous galaxies, physicists confirmed that both classical Newtonian gravity, and Einstein's general relativity do not account for the observed behavior if the only sources of mass are those consisting of visible matter. Moreover, several extra-galactic observations of galaxy clusters and gravitational lensing suggest sources of gravitation that cannot be accounted for by visible galaxies. One such example observed by the Hubble photographing Abell 1689 is illustrated in figure 1-2 on page 15. Because these observations both point to an invisible source of mass, physicists proposed one, and named it dark matter. Quantum mechanics has demonstrated that mass comes in a quanta-what we call particles. Thus, physicsts are searching for the quanta of dark matter: the dark matter particle. 14 LIMOUSIN ET AL Figure 1-2: In this photograph taken by Hubble of Abell 1689, a distant galaxy cluster. Notice the rings in the image around the center of the cluster. These distortions are caused by the mass of the cluster gravitationally lensing the light passing through it. GR can only explain the apparent degree of lensing if there is a large additional source of mass- dark matter. Adapted from [5]. 1.2 Weakly Interacting Massive Particles In our understanding of how particles interact , there are four forces. The Standard Mode very successfully describes the electromagnetic, weak , and strong forces. The other major force , gravity, is the subject of Einstein 's theory of general relativity. Dark matter has only been observed as a result of its gravitational effects, and our observations indicate that it does not couple to electromagnetism. Current searches for dark matter focus on finding a second interaction it exhibits other than gravitation. DMTPC seeks to detect weakly interacting massive particle, 15 or WIMPs, which are just one candidate for a type of dark matter particle. This model for dark matter proposes that it interacts via the Weak force, which is motivated by the WIMP miracle-a mathematical coincidence in which the density of dark matter in today's universe is consistent with the predicted density that a self-annihilating weakly interacting massive particle would have in the modern universe [2]. 1.3 Dark Matter Wind We believe that dark matter aggregates in a halo1 around galaxies, binding them together gravitationally and accounting for their rotation curve. In the halo model for dark matter, we expect to observe a dark matter wind. This wind simply refers to the fact that as massive bodies such as the Earth orbit the galactic center, they pass through the cloud of dark matter that permeates the galaxy, and thus dark matter particles that pass through the Earth create a wind in the direction opposite of our orbit. In our orbit, we follow the constellation Cygnus, and thus expect this wind to come from its direction. 1.4 Dark Matter Detection Many efforts to detect dark matter focus on searching direct production of the particle in accelerators, or photons emitted by the particle self-anhilating. These forms of detection are known as direction detection. The experiment in this thesis, DMTPC, seeks to detect the presence of dark matter through directional detection. Direc- tional detection simply refers to a measurement seeking to find dark matter collisions consistent with the location of Cygnus in the sky. While there are several other non-directional collision based detectors, DMTPC has an advantage because it seeks to correlate the collisions with the location of Cygnus in the sky-something which would be distinguishable from other potential sources of noise. Ultimately, DMTPC seeks to verify that collisions observed in higher density detectors in fact represent 'This is known as the Halo Model 16 dark matter collisions, and not some other source. 17 18 Chapter 2 The Dark Matter Time Projection Chamber 2.1 Time Projection Chambers Time projection chambers (TPCs) are a general class of detectors used in experimental physics; a sketch of one is illusrated in figure 2-1 on page 20. They consist of a vessel filled with a gas that releases electrons as a result of a particle passing through it. An applied electric field then causes these electrons to drift across the chamber in a region referred to as the drift region. The electrons then reach a second region of an even higher electric field at the end of the chamber, called the amplification region, where they liberate even more electrons. The electrons then produce image charges on a plate, where the charge is then read out. The experimentalist can then describe the particle tracks by studying this charge readout. Studying the temporal characteristics of the charge readout on the plate carries information about the trajectory of the particle track. If the charge readout is very narrow in time, it indicates that the track entered the detector parallel to the plate, and because all of the electrons produced by it had a very similar drift time. Alternatively, if the charge readout occurs over a larger amount of time, it indicates that the track came in with a z component, and thus the electrons were liberated varying distances from the plate, and took different amounts of time to get through 19 the drift region and reach the amplification region. The charge readout consists of an array of pads, and thus gives an XY resolution of the particle tracks. Combining the information from the charge readout and the temporal characteristics of the charge readout , TPCs obtain 3 dimensional resolution. High Polenuar Figure 2-1: This is a sketch of a time projection chamber. Note the electrons liberated from the particle track, and the arrows pointing away from the direction of higher potential, indicating the acceleration of these electrons by the chamber's applied electric field. Adapted from [6]. 2.2 DMTPC's Chamber The DMTPC's cubic meter chamber consists of a vessel filled with carbon tetrafluoride as a gaseous scintillator 1 . We use CF 4 because it scintillates in response to nuclear recoils- which is what we expect WIMPs to produce. Previous iterations of DMTPC detectors have been calibrated using neutrons and alpha particles coming from strong directional sources [l]. Using a low density gas scintillator comes with the benefit that the tracks of nuclear recoils manifest themselves in the scintillation medium because of the large mean free path of the recoiling particle. By tracking the paths of the ejected flourine in the collision (through the electrons it produces), we hope to reconstruct the incoming momentum vector of the particle that produced the event. Smaller past 1 A substance which emits light as a result of interacting with certain species of particles 20 prototype detectors have illustrated this concept by successfully reconstructing the direction of neutrons from a known source incident on the detector. An important downside of using the gas scintillation method is that while it provides directional detection because of the large mean free path of particles in the gas 2 , this also means the medium has very low density, so its effective cross section for interacting with dark matter is very small. DMTPC differs from typical time projection chambers. Instead of reading out the charge of the electrons produced in the amplification region, DMTPC looks at the scintillation that occurs as a result of the amplification process-this is known as secondary scintillation. To image the secondary scintillation, DMTPC uses charge coupled devices, or CCDs. One can imagine a CCD as essentially an array of quantum wells filled with electrons. As photons are incident on each of the wells, they scatter some of the bound electrons out of the wells. The CCDs provide DMTPC with the XY resolution, but do not give any resoultion in the Z direction because they lack the temporal resolution of charge readout. Another challenge with using CCDs lies in the fact that they continuously produce images, and have no way of easily filtering out which images contain tracks from events. Thus, we need a device that provides temporal resolution of the secondary scintillation, and also one to filter which images are useful ones containing nuclear recoils, and this involves setting up another device that can detect their occurrence-preferably an analog device that can be run continuously. This instrument is the photomultiplier tube, or PMT. 2.3 PhotoMultiplier Tubes PMTs use the photoelectric effect to detect photons. The vast majority of pho- toelectric events in PMTs involve one photon liberating a single electron from the photocathode. Exposing the photocathode to some flux of photons above its work function will produce a current proportional to the flux. The PMT allows scientists 2 The large mean free path of the recoiling flourine means that it produces drift electrons along a longer path 21 ......................... .......... to probe small amounts of flux, by amplifying the small amount of current produced by the photoelectrons produced in the photocathode. It amplifies these electrons by the application of a high voltage across dynodes. The ratio of the number of electrons produced after amplification to the number of photoelectrons before the amplification is called the gain of the PMT. A major part of my work for DMTPC consisted of measuring the gain of each of our eight PMTs. This value is significant because it will allow DMTPC to look at the characteristics of the traces from each PMT and determine the flux produced in the scintillation, and thus will help distinguish different types of scintillation events. The gain of a PMT is produced because, upon entering the externally applied electric field, the photoelectron accelerates, and scatters into a dynode, liberating more electrons. All of these electrons then also feel the applied electric field, and scatter into the next dynode, repeating the process, and multiplying the number of electrons. After a series of these dynodes, single electrons can be turned into macroscopic, measurable currents. A rough sketch of a PMT is illustrated in figure 2-2 on page 22. In our experiment, we use Hamamatsu 1408 PMTs, which Photocathode Focusing electrode Ionization track/ High energy photon Photomultiplier Tube (PMT) Low energy photons Connector Scintillator Primary electron Secondary electrons Dynode Anode pins Figure 2-2: An exmaple of a 12 dynode stage PMT responding to an external scintillator. [10] have a 9 dynode amplification with an applied potential difference of around 2000V across them. This amplification produces on the order of 10' electrons per incident photoelectron. The ratio of the number of electrons produced after amplification to the number of photoelectrons before the amplification is called the gain of the PMT. A major part of my work for DMTPC consisted of measuring the gain of each of our eight PMTs. This value is significant because it will allow us to look at the character22 ................ istics of the traces from each PMT and determine whether they represent a nuclear recoil or not. 23 afikawaitibbtdAmmiss graangra ad5$4g 24 Chapter 3 Calibrating PhotoMultiplier Tubes 3.1 PMT Gain Measurements To determine the gain of DMTPC's PMTs, I measured the current output of the device in response to a photon flux incident on the photocathode. The most challenging aspect of this measurement was accurately estimating the incident flux of photons. The challenge came from the PMTs being sensitive enough to detect extremely low levels of light-making it necessary to only expose them to very low intensity sources or risk damaging them. I thus had to ensure that my testing environment was light tight, and used a dark box to accomplish this, as described in section 3.1.2. Initially, I tried using an LED (light emitting diode) as my light source, but discovered that the intensity of light the LED fluctuated too much. I finally successfully measured the gain by switching to a device known as a PicoQuant, which is a pulsed LED picosecond laser. The PicoQuant works like a standard laser, just it uses a rapidly pulsed LED to produce the population inversion in the laser medium. The function generator used to drive the PicoQuant is finely tuned to the LED used to excite the laser medium, and the resultant laser has a very stable flux, and can be tuned to very small, even single photon levels of flux. 25 _11___-__I_ 3.1.1 - I-- - - - - I __ - . .. __ - _- - i - _- - -1 - - - q:W-:IV ._ . K: -_ X *1 . Light-Tightness I pointed the PicoQuant directly at the center of the PMTs from about a foot above them, and the entire setup took place in a dark box. To ensure there was no background light, I placed each PMT in a second dark box, covered by an additional black tarp, and turned off the lights in the room. I then recorded the rate of events1 , and compared that to the rate of events in just the outer dark box with the lights in the room turned on. Because the rates were similar in both situations, I concluded the single outer dark box was sufficiently light tight for the gain calibration measurements. I illustrate an example of dark current measurements for the PMT PADM with the lights turned off, and the lights turned on in figures 3-1 and 3-2. PADM Lights Off iO*0 . 10 a 6 U) U C 4 2 0 I -2 -4 0S 90.6 12 0 ~11 1.4 I Time (Seconds) 1.6 1.6 x 10. Figure 3-1: One thousand 10 ps sweeps of PMT PADM in the dark box with the room lights turned off. There are a total of eighteen events with an amplitude greather than l mV. 3.1.2 PicoQuant Trigger The PicoQuant had an external trigger input that was used to control when a laser pulse was fired at the PMT. This trigger was supplied by a pulser, and the signal was simultaneously fed the PicoQuant and to a Lecroy oscillioscope to trigger its sweeps as well. This setup is illustrated in figure 3-3 on page 28. 'I defined events as traces with an amplitude greater than 1 mV 26 ..... ....... ..... .. .......... ...__ ...... . ... ....... - - I - _ ' ' - -_ . . PADM Light On 12n 12 10 -- a -- V U 0 6 - (h 4 0.8 . 1 ' 1.2 ' -2 -- 14 1. 1.8 X0" Time (Seconds) Figure 3-2: One thousand 10 ps sweeps of PMT PADM in the dark box with the room lights turned on. There are a total of sixteen events with an amplitude greather than 1 mV. After one thousand sweeps, I observed the plot in figure 3-4 on page 29. The fact that the pulser triggered these sweeps as well as the PicoQuant, and that almost all of the peaks occurred at the same time, serves as strong evidence that these peaks were in fact produced by the PicoQuant firing. To further verify that these peaks were in fact produced by the PicoQuant, I blocked it from the PMT with a black tarp, and conducted the same measurement again. It yielded no peaks, reinforcing that I had been in fact observing light produced by the PicoQuant. I then adjusted the intensity of the PicoQuant such that a peak was observed only one out of every ten sweeps of the oscilloscope. This adjustment of intensity allowed me to apply the regime of low-count Poisson statistics: Equation 3.1 describes a Poisson distribution, where P(k) gives the probability of observing k photons, with A representing the rate of photons, or -. P(k) = Ak Since ninety percent of our events are null events (k (3.1) = 0), we know that A is small and can expect approximately nine to ten percent of observations to be single photon events (k = 1), and on the order of one percent double photon events (k = 2), etc. I computed the gain by integrating the traces of all non-zero events, and averaging these values, since each included trace represented on average one photon. When I 27 .. .... .. ...... ... Pulser PicoQuant Oscilloscope PMT I I HV Power Supply Figure 3-3: This schematic illustrates a block diagram of my setup for the gain and dark current measurements on the PMTs. integrated these traces, I used the known impediance of the oscilloscope (50 Ohms) to calculate the current readout from the PMT. I then performed this on all the PMTs, and recorded my gain values in table 5.1 on page 40 in the results section. 3.1.3 Gain Results In all of my measurements, I applied 2000 volts to the PMTs. For each PMT, I recorded one thousand sweeps on the oscilloscope, corresponding to one thousand trigger events from the pulser. The sweeps for one PMT, when superimposed on a plot, are illustrated in figure 3-4: As seen in figure 3-4, most of the events occurred at the same time relative to the pulser triggering, and when I covered the PMT these events disappeared; this verifies that they in fact correspond to light arriving from the PicoQuant. I computed the gain by writing a MATLAB script which integrated the peaks across all of the sweeps, and outputted the average integral of the peaks. I included my MATLAB scripts in an appendix. Because most of the sweeps contain no peak, 28 -- "... - - ...... . .... x 10·' 1000 Traces of PMT PGEJ 1 2~~--~--~-~--~--~--~-~-----~ 10 - 2 ~~ , _ 1-2--1.~1 22--~1 . 1 2-4--1.~216--~1 . 128 _ _ _1~ _ 1 3---, _ 1~ 32 --1~ . 134 ___ 1 . ~1 36-~l.138 Time (Seconds) x 10·• Figure 3-4: One thousand sweeps of the oscilloscope observing PMT PGEJ respond to one thousand triggers of the PicoQuant. The figure clearly illustrates that almost all of the events happened between 1.128 x 10- 5 and 1.132 x 10- 5 seconds, indicating that these events were produced by the PicoQuant. When I computed the gain I integrated only the sweeps that met my trigger threshold after 1.128 x 10- 5 seconds and before 1.132 x 10-5 to rule out the dark current events (for example, in this plot the first two peaks). I first told the code to isolate the sweeps with peaks exceeding a 1 mV amplitude. I chose 1 mV because the baseline noise level in all of the PMTs varied, but never exceeded 0.8 mV, so lmV allowed me to avoid triggering on it , but still capture all relevant events. Figure 3-5 gives an example of the baseline electronic noise level in PGEJ, and figure 3-6 illustrates why I chose the 1 mV cutoff for events. The code then further isolated only those sweeps with events in the region corresponding to events from the PicoQuant. Finally, the script then integrates all_ of the qualifying sweeps in the region where all of the peaks occurred. Because these peaks correspond to almost all single photons by Poisson statistics, their average represents the single photon gain , or absolute gain of the PMT. 29 Electronic Noise in PGEJ 0.8 0.6 0.4 Ci)' 0.2 .:t:::! 0 2:Q) C) 0 ns ~ -0.2 > -0.4 -0.6 -0.8 -1 1.12 1.122 1.124 1.126 1.128 1.13 Time (Seconds) 1.132 1.134 1.136 1.138 5 x 10· Figure 3-5: This is a zoomed in version of figure 3-4, and is an example of the baseline electronic noise in PMT PGEJ , which ranges from -0.7 mV to 0.8 mV on the interval in this figure. This noise was the reason I chose to use a lmV cutoff across my measurements for what constituted an event. 3.2 Dark Current In addition to determining the gain of the PMTs, I also measured their dark current. Dark current consists of events registered by the PMTs that are not produced by incident photons. Such events can be the result of thermionic emission either from the photocathode, or one of the dynodes. An idea to measure the gain was to observe dark current events, because such events represent single electron events. However , there is a flaw with this technique: thermionic emission can occur at any of the dynodes, which means that although it represents single electron events, these events do not necessarily undergo the full amplification associated with passing through all nine dynodes in the PMT. The dark current events that are not fully amplified are so small that they have little consequence on any measurements of the secondary scintillation that occurs in DMTPC 's detector. 30 Histogram of Peak Amplitudes 400 - 350 - 300 - 0 250 0 - 100 60- 0 0.6 1 1.5 2 2.5 3 3.5 Voltage (Volts) 4 6 4.6 X10 Figure 3-6: This is a histogram of the peak voltage amplitude in one thousand sweeps of the PMT PGEJ responding to the picoquant. Note that an overwhelming amount of the events occur in the bins below 1 mV. Because this occured for all of the PMTs, I decided that 1 mV serves as an appropriate cutoff to isolate events that are just electronic noise. 3.2.1 Dark Current Measurement Results I determined the dark current of the PMTs by isolating the PMT from any sources of visible light. I accomplished this by placing the PMT in a small, black, light-tight metal darkbox, covering this darkbox with a large black tarp, and enclosing all of this in an even larger dark box (as opposed to the single darkbox I used for the gain measurements). In order to ensure that everything was completely light tight, all dark current measurements involved two runs-one with the lights in the room turned on, and the other with all light sources in the room turned off, and all windows covered. The rates of dark current were similar in both cases, indicating that what was being measured was in fact dark current, and not background light. I measured the dark current by taking ten ps sweeps on the oscilloscope repeatedly one thousand times, and then counting the total number of peaks with an amplitude exceeding 1 mV on this interval. After combining the thousand traces, the interval I counted events over corresponded to ten milliseconds, so I multiplied the number of events by one hundred to obtain the dark current in units of hertz (number of events per full second). 31 . .......... ... .... ......... ..... In measuring the dark current, I used a much larger time scale than with the game measurements because at the time scale of the gain measurements dark current events are extremely rare. When I ran one thousand sweeps of the oscilloscope on one of the PMTs, I observed the plot in figure 3-7. 10 Dark Current for PGEJ x10" - 6 0 0 4 0- 0 b.6 0.7 08 09 1 1 1 lime (Seconds) X10" Figure 3-7: This is a dark current run on the PMT PGEJ. It consists of one thousand 10 ps sweeps of the PMT while it was exposed to no light sources. The dark current events are the numerous sharp peaks. 32 .. .... ...... . ... .. .............. ... .......... . .... .. .. Chapter 4 Optical Response to Electron Avalanches 4.1 The CF 4 Secondary Scintillation Spectrum The cubic meter detector relies on the light produced during secondary scintillation of CF 4 to image particle tracks in the detector. Primary scintillation refers to the light produced by the nuclear recoil in the drift region, whereas secondary scintillation refers to the light produced by the electron avalanche in the amplification region. This spectrum is significant to DMTPC because the PMTs and CCDs in the detector will respond to light from this spectrum, and have different sensitivities to various DMTPC has measured the secondary scintillation spectrum of CF4 , wavelengths. and observed the spectrum in figure 4-1. DMTPC plans to operate the detector at pressures on the order of 10-100 torr. DMTPC measured the spectrum in figure 4-1 at 180 torr. Recently, more research has focused on studying the behavior of CF 4 's secondary scintillation, and found a complicated pressure dependence, as illustrated in figure 4-2. The measurements in figure 4-2 occur at pressures between 1 and 5 bar, and thus do not extrapolate well to DMTPC's detector. However, the complicated pressure dependence could still manifest itself at low pressures, and thus merits future investigation because of its consequences on the PMTs and CCDs. Specifically, because of their different 33 sensitivities across the spectrum, the pressure dependence of the spectrum will impact the relative amount of light registered by the PMTs and CCDs. 0.016 0.014 0.012 U I C 0.01, 0.006 jb * 0.006 B - 0.004 0.002 0 200 Mo 400 110 70oo 00 M.v (rnj M Figure 4-1: This is the secondary scintillation spectrum of CF 4 , as measured by a Hamamatsu H1161 PMT at 180 torr. There are two regions of high intensity-one centered around 300nm, and another around 650nm. Adapted from [3]. 4.2 The PMT response The PMTs being used are Hamamatsu R1408s. When light enters these PMTs, it passes through the CF 4 gas in the detector, a quartz glass window, the air in the PMTs mount, and the face of the PMT. The quantum efficiency of these PMTs, as illustrated in figure 4-3, makes them responsive almost exclusively to the UV peak of CF 4 's secondary scintillation spectrum. Thus, the PMTs will be detecting only this part of the spectrum, and very little of the red line. Various manufacturers of quartz materials claim that quartz is over 80% transparent to wavelengths all the way down to below 200nm, so the transparency of the quartz windows should encompasse the entire spectrum of interest. 34 Charge gain -70 -5 bar bar -4 -3 bar 2 bar 1 bar - ( .do 0.8 E 0.6 0.40- 0.4 0.20- -0.2 0.00 200 300 600 Wavelength, nm 400 500 . 0.80- '0.0 700 800 Figure 4-2: This is a measurement of the secondary scintillation spectrum of CF 4 across a range of pressures, from 1 to 5 bars. The two major peaks in the UV and red part of the spectrum exhibit a complicated pressure dependence. Significantly, the behaviors of the two peaks are not completely the same. Notice that the intervals between the maximum height of the red peak are much more evenly spaced than those of the UV peak across the different pressures. Adapted from [7]. 4.3 The CCD Response The CCDs are senstive to the red peak, but not the UV peak. The primary reason for this lies in the lenses mounted on the CCDs. We conducted transparency measurements of these lenses across the spectrum, and observed their transparency drop off completely around 350nm, as illustrated in figure. We conducted the measurement in figure by placing our lens between a monochromator and a quartz glass mercury spectral lamp. We chose to use this light source because of its strong UV peaks, and the quartz's transparency to UV light. As highlighted in the figure 4-4, we found a drop off in transparency around 350nm. The CCDs will thus have very little response to the UV peak because of the opacity of their lenses to UV light. 4.4 Total Responsivity After considering all of the optical components involved in imaging the electron avalanches in the detector, the total response of the PMTs and CCDs are given in equations ?? and 4.3, respectively. 35 .. ........... ....... . ..... . ......... IIL I . . II II I , It,' . "11 jl ' - . - - - - , - - - 70 U Monte Carlo Absorbtivity >1 60 -~50 momooano*oooooooo n Effcincy 40 b t 00 20 0 7 o in 0 0~* 250 300 350 400 450 500 550 00 700 650 Wavelength (nrn) 600 Figure 4-3: This is a measurement of the quantum effiency of Hamamatsu R1408 PMTs. The relevant datapoints are the stars, which represent the measured quantum efficiency of the PMTs across the spectrum. Note that the spectral response drops off almost completely around 650nm, which is the center of red peak in the CF 4 secondary scintillation spectrum. The center of the UV peak occurs at 300nm, and extends up to 400nm, which is a region the PMTs are very sensitive to. Adapted from [4]. PMTSignal(A) = F(XP; A) * TQuartz(A) * TcF4 (A) * TAir(A) * PMTQE(A) * PMTGain- (4.1) Where the total signal depends on the total number of photons produced by the scintillation event that could be incident on the PMT, y(A), the transmission of the quartz window, the CF4 gas, the air in the PMT mount, the quantum efficiency of the PMT, and the PMT's gain. Note that the total photons produced, -y(A), factor into two components, as illusrated in equation 4.2. 7(yQ, P; A) 7Geometric(x) y(P; A) * YGeometric(X), (4.2) refers to the geometric cross section of light from the avalanche that the PMT sees. This factor follows a I dependence, and also depends on the angle the light strikes the window and PMT at, since it is more likely to reflect for large angles of incidence. This part of -y is independent of the other part of gamma, -Y(P; A). This term is the scintillation spectrum, where after specifying a pressure P, there is a function of lambda that gives the flux at every wavelength, as illusrated in figure 36 -Uncovered Optical Lens _X-Ray Lens 10 10 10 10 10 19'Q0 3500 3000 4000 4500 Wavelength (Angstroms) Figure 4-4: This is a measurement of the transmission of light from an elemental lamp through optical and x-ray lenses to be used on the CCDs of the detector. The figure clearly illusrates that the transmission of light through the lenses drops off at around 350nm by comparing them to the blue line, which represents the spectrum seen without any lenses in front of the lamp. 4-2. Now performing a similar analysis for the CCDs, we arrive at equation 4.3. CCDSignal(A) - y(*, P; A) *TQuartz(A) *TcF4 (A) *TAir(A) *TLensCCDQE(A). (4-3) These two equations have many terms that cancel, such as transparencies of the CF 4 , and windows. However, some terms, such as the PMT gain, quantum efficiencies, and lens transparency do not cancel. All of these non vanishing terms in the ratio of the signals represent significant information that DMTPC could use to determine the relative signals the detector expects from the PMTs and CCDs for various wavelengths and conditions. 37 . ..... .... 38 Chapter 5 Results and Conclusions 5.1 Uncertainties I computed the uncertainty in the value of the absolute gain by calculating the integral of the peaks of two different distributions for each PMT. One of these distributions was the average integral of the peaks with a 1mV cutoff defining the minimum height for the trace to qualify as an event, whereas the other was determined by setting the threshold for an event as one increment higher than the highest value the baseline electronic noise level reached on the oscilloscope (where the increment is the minimum voltage resolution of the oscilloscope). I then compared these two values: The 1 mV cutoff gave a higher value because it did not include the smallest peaks. I then took the difference between these two values to be the uncertainty in the gain. Because I computed the dark current from a simple Poisson process1 , I found that the greatest source of error was in the uncertainty of the number of events occurring in such counting processes, which goes like the square root of the number of events. 5.2 PMT Results In measuring the gain and dark current of the PMTs, I discovered that one of the PMTs, labeled PACI, produces no signal other than electronic noise. Regardless of las illusrated in equation 3.1 39 PMT PADM PAZE PEEW PGEJ PGEQ PGMQ PHDY PACI Gain (4.86 ± 0.6) x 106 (8. 75 ± 0.28) x 10 6 (6.99 ± 0.54) x 106 (3.88 ± 0.07) x 106 (5.31 ± 0.51) x 10 6 (5.61 ± 0.39) x 106 (6.04 ± 0.2) x 106 Dead Dark Current (kHz) 1.800 ± 4.24 5.800 ± 7.61 13.700 ± 1.170 1.600 ± 4.00 3.600 ± 6.00 2.100 ± 4.58 1.900 ± 4.35 Dead Table 5.1: This table documents the final measured gains and dark current for all eight PMTs, and includes the uncertainties I describe in section 4.2. what light I exposed the PMT to , it only recorded the background electronic noise voltage , and did not even produce dark current events. A plot of PACI 's response to the PicoQuant 's light is shown in figure 5-1. All the other PMTs were functional and yielded results for both measurements, as illustrated in table 5.1: PMT PACI Iii" .:I:! ~ 11) 2 Cl ra .:I:! 0 >1 ... ... ~ ., ..... . ( t~ l >• ~ "'\' ~r. ~ . ~~J ~ . .: ... . !'l: ~ "-1. ' '!l.9 . _'!· 1_; ,, ~ • ~. ~ :t• 1 ,• ~ ~ . .••~ . -1 ' - - - - - - - - - ' - - - - ' - - - - ' - - - - ' - - - - ' - - - - ' - - - - ' - - - _ _ , _ __ 1 118 1.12 1.122 1.124 1.126 1.128 1.13 1.132 1.134 __.___ 1.136 •• v •• 4\ • ~\'~ '" :t: : . ,1 Time (Seconds) ·1·· ' t • '1,, .. ~ ~ ~~ ·· ~(• ___.____, 1.1 38 x 10·• Figure 5-1: One thousand sweeps of PACI being exposed to light from the PicoQuant. The image clearly demonstrates the lack of any signal coming from the PMT other than electronic noise. PACI's dark current test also failed to yield any events. All the working PMTs exhibited gains on the order of 10 6 -10 7 , which is consistent with the expected value of 107 fromthemanuf acturer. I believe my measurements yielded results slightly lower than 10 7 because my measurement technique allowed me to recognize extremely small low gain events. I am confident in these being actual events, because of their temporal proximity to all the other events produced by the 40 PicoQuant, as illustrated in the figure of sweeps earlier. The dark noise of all the functional PMTs fell in the regime of a few kiloHertz, which is consistent with previous measurements of their dark current. However, one PMT proved an exception to this: PEEW exhibited a dark noise of 13700 Hz, which stands noticeably out among the PMTs. 5.3 Transparency Results The xray and optical lenses to be used on the DMTPC CDDs are very opaque to the UV peak of the CF 4 secondary scintillation spectrum, but transparent to the red one. Because of the low quantum efficiency profile of the PMTs at the red peak in the CF 4 spectrum, they will primarily respond to photons produced by the UV peak. In contrast, because of the lenses opacity to UV, the CCDs will respond to the red peak in the scintillation spectrum, and not the UV one. Thus, the two forms of optical detectors on the chamber will have responses to orthogonal parts of the CF 4 spectrum. 5.4 Conclusions Six of the eight PMTs intended for use on the cubic meter are in good working order. The PMT PEEW exhibited one of the highest gains of the PMTs, but with an excessive dark current of 13700 1170 Hz. Because DMTPC will not be triggering on small, single photon events, this PMT should still be a useful instrument for use on the detector. However, PACI exhibits no signal at all, but might be reparable if opened up and investigated. For the six PMTs in good working order, the gains and dark currents are all relatively similar, and thus they should exhibit relatively similar responses to secondary scintillation events. These PMTs will provide DMTPC with information about the z dimension of the nuclear recoils through the temporal characteristics of their response to avalanches. Additionally, by using the well characterized gains of these PMTs, DMTPC should be able to determine the electron avalanche's spatial 41 location in the XY plane of the amplification region because the response of each PMT can be used to calculate the relative distance of the avalanche from each PMT 2. Ultimately, with the gains in table 5.1 DMTPC will have the information necessary to identify nuclear recoils from the PMT signal. Additionally, with the information about the CF 4 spectrum, and the CCD lens transparency, DMTPC has some key information about factors that influence the relative responses of the PMTs and CCDs, since they are responding to opposite parts of the CF 4 scintillation spectrum. In light of the observations in chapter 4, future work DMTPC should consider for the cubic meter detector includes: verifying that the transparency of air in the PMT mounts does not impact the light reaching the PMTs; verifying the transparency of the CF4 gas in the chamber; testing the transparency of the quartz glass windows; and measuring the pressure dependence of the CF 4 secondary scintillation spectrum at pressures close 2 Since the flux hitting each PMT originated from the same event, and decays as 42 . to operating pressure. -,I-- ::"..::":::. ...................... - - :. I - MMIMEW . - I - - - - - -- - . - . .. . .... .. ..... . ..... Appendix A Code matfiles = dir(fullfile('D:' , 'UROP', 'PMTs', 'Tests2000V', L=[]; for i =1:length(matfiles); data = load(matfiles(i). name); if max(data(:,2))>0.001 n=integ2 (data); L=[L, n]; end end hist (L, 100) mean (L) /50 (mean(L)/50)*10^(-12)/(1.6*10^(-19)) function integ2=integ2 (A) B=A ( :, 2); C=B (98:129) 43 -31 'PGEJ', '*.dat')); W=A(2, 1) -A(1, 1); Y=O; for n = 1:length(C); Y=Y+W*C (n); end integ2=10^12*Y end end 44 ............. ...... Bibliography [1] C. Deaconu. A model of the directional sensitivity of low-pressure cf4 dark matter detectors. 2015. [2] Jonathan L Feng and Jason Kumar. Dark-matter particles without weak-scale masses or weak interactions. Physical review letters, 101(23):231301, 2008. [31 A Kaboth, J Monroe, S Ahlen, D Dujmic, S Henderson, G Kohse, R Lanza, M Lewandowska, A Roccaro, G Sciolla, et al. A measurement of photon production in electron avalanches in cf 4. Nuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment, 592(1):63-72, 2008. [4] MD Lay and MJ Lyon. An experimental and monte carlo investigation of the r1408 hamamatsu 8-inch photomultiplier tube and associated concentrator to be used in the sudbury neutrino observatory. Nuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment, 383(2):495-505, 1996. [5] Marceau Limousin, Johan Richard, Eric Jullo, Jean-Paul Kneib, Bernard Fort, Genevieve Soucail, Ardis Eliasd6ttir, Priyamvada Natarajan, Richard S Ellis, Ian Smail, et al. Combining strong and weak gravitational lensing in abell 1689. The Astrophysical Journal, 668(2):643, 2007. [61 Jeffrey T Mitchell and Upton. Time projection chamber. In ProceedingsArkansas Academy of Science, volume 49, 1995. [7] A Morozov, LMS Margato, MMFR Fraga, L Pereira, and FAF Fraga. Secondary scintillation in cf4: emission spectra and photon yields for msgc and gem. Journal of Instrumentation, 7(02):P02008, 2012. [8] Vera C Rubin and W Kent Ford Jr. Rotation of the andromeda nebula from a spectroscopic survey of emission regions. The Astrophysical Journal, 159:379, 1970. [9] WW Shane and GP Bieger-Smith. The galactic rotation curve derived from observations of neutral hydrogen. Bulletin of the Astronomical Institutes of the Netherlands, 18:263, 1966. 45 [101 Wikipedia. Photomultiplier accessed 25-April-2015j. wikipedia, the free encyclopedia, 2015. [Online; 46