Quality Control in Diagnostic Radiology Factors driving Q.C. Why do we do it? Legal Requirements Accreditation JCAHO ACR Clinical improvement equipment performance image quality Q.C. Goals Minimize dose to patients staff Optimize image quality Establish baselines More on this in a moment Why is Q.C. Important? Without a QC program the only way to identify problems is on patient images. And some problems don’t show up on images. Yeah, that’s what I always say. QC can detect Malfunctions Unpredictability may be hard to isolate clinically Inefficient use of Radiation high fluoroscopic outputs Radiation not reaching receptor inadequate filtration oversized collimation Goals of a Q.C. Program Obtain acceptable image with least possible radiation exposure to patients staff Attempt to identify problems before they appear on patient films without QC problems only detected on patient films “Acceptable” Image Image containing information required by radiologist for correct interpretation goal: minimize exposure while maintaining acceptability high exposure images often have excellent appearance Low noise Q.C. & Baselines Baselines quantitative data on equipment obtained during normal operations Baselines useful for troubleshooting isolating problem component, for example generator processor Allows efficient use of engineering / repair personnel X-Ray Quality Control Filtration Focal Spot Size Collimation Maximum Fluoroscopic Output Calibration Verification Phototimer Performance Why is Filtration Important? Tube emits spectrum of x-ray energies Filtration preferentially attenuates low energy photons low energy photons expose patients do not contribute to image low penetration Half Value Layer (HVL) We don’t measure filtration We measure HVL HVL: amount of absorber that reduces beam intensity by exactly 50% Half Value Layer Depends upon kVp waveform (single/three phase) inherent filtration Minimum HVL regulated by law Maximum HVL regulated only in mammography kVp HVL (mm Al) 30 0.3 40 0.4 49 0.5 50 1.2 60 1.3 70 1.5 71 2.1 80 2.3 90 2.5 100 2.7 110 3.0 120 3.2 130 3.5 140 3.8 150 4.1 Georgia State Rules & Regulations for X-Ray Radiographic HVL Setup Radiographic Filter R Tab leto p Checking HVL Compliance (Radiographic) Radiographic How much aluminum must be placed in beam to reduce intensity by exactly 50%? Filter R Tab leto p OK! Must add Al to reduce beam to exactly 50% 90 kVp Measurements; 2.5 mm Al minimum HVL filter mR (mm Al) ------------------0 250 2.5 133 filter mR (mm Al) ------------------0 250 2.5 125 filter mR (mm Al) ------------------0 250 2.5 111 Acceptable HVL > 2.5 mm Marginal HVL = 2.5 mm Unacceptable HVL < 2.5 mm Not OK! Must remove Al to reduce beam to exactly 50% Checking HVL Compliance (Radiographic) Radiographic Is this machine legal? 2.5 mm Al minimum filtration at 90 kVp Filter R Tab leto p 90 kVp Measurements filter mR (mm Al) ------------------0 450 2.5 205 Fluoroscopic HVL Setup Fluoroscopic Tube Filt rat ion ( Half Value Layer) Image Tube A bsorber ( t o prot ect Image Tube) R Tablet op Fi l t e r Fluoroscopic HVL Set desired kilovoltage manually measure exposure rates instead of exposure Fluoroscopic Tube Filt rat ion ( Half Value Layer) Image Tube A bsorber ( t o prot ect Image Tube) Move absorbers into beam as needed R Tablet op Fi l t e r Focal Spot Size We measure apparent focal spot Trade-off smaller spot reduces geometric unsharpness larger spot improves heat ratings Actual Focal Spot Apparent Focal Spot Focal Spot Size (cont.) Focal spot size changes with technique Standard technique required 75 kV (typical) 50% maximum mA for focal spot at kV used direct exposure (no screen) NEMA Standards defines tolerances Nominal Size Tolerance ------------------------------------>1.5 mm 30% >0.8 and <=1.5 mm 40% <0.8 mm 50% Focal Spot Measuring Tools Direct Measurement Pin Hole Camera Slit Camera Indirect Measurement of Resolving Power Star Test Pattern Bar Phantom Direct Focal Spot Measurement Measure focal spot directly in each direction Use triangulation to correct for distances formula corrects for finite tool size two exposures required for slit Slit Camera Pinhole Camera Star Test Pattern Measures resolving power infers focal spot size Dependent on focal spot energy distribution measure largest blur diameter (in each direction) magnification use equation to calculate focal spot size Bar Phantom Measures resolving power Find smallest group where you can count three bars in each direction Bar Phantom Setup Radiographic Collimation X-Ray / Light Field Alignment Beam Central Axis should be in center of x-ray beam Collimator field size indicators PBL (automatic collimation) field automatically limited to size of receptor Bucky Alignment Using longitudinal bucky light & transverse detent, x-ray field should be centered on bucky film X-Ray / Light Field Alignment Mark light field on table top with pennies X-ray / Light Field Alignment Slig ht m isal ignm ent on this edg e Lam p Lig ht Tab leto p X-Rays Tab leto p Radiographic X-Ray / Light Field Alignment Fluoroscopic Collimation image field is scale seen on monitor expose film on table above scale compare visual field (monitor) with x-ray field on film must check all magnification modes Image Tube Film Collimator Test Tool Template Ta bletop Fluoroscopic Collimation Fluoroscopic Collimation Maximum Fluoro Output put chamber in beam on tabletop block beam with lead above chamber Image Tube fools generator into providing maximum output Lead 10 R/min. limit for ABS fluoro R Tab leto p Maximum Fluoro Output Lead Calibration Performance Parameters Timer Accuracy Repeatability Linearity/Reciprocity Kilovoltage accuracy mA must be measured invasively Calibration 120 kVp mA time mAs mR mR / mAs (msec) -----------------------------------------------------.1 10 240 24 Constant 100 200 .05 10 ? ? mAs 50 .2 10 ? ? mR/mAs should stay constant for all combinations of mA & kVp at any particular kVp Calibration 120 kVp mA Double mAs Double mAs again time mAs mR mR / mAs (msec) ----------------------------------------------------100 .1 10 240 24 200 .1 20 ? ? 100 .4 40 ? ? mR/mAs should stay constant for all combinations of mA & time at any particular kVp Phototiming (check with output or film) Reproducibility Density Controls Field Placement Field Balance Phototiming Operation should be Predictable Phototimer Density Control Settings R R Tablet op Density Control -2 41 -1 49 0 62 1 76 2 96 Phototiming Density Steps should be predictable & approximately even % Step to Step Change Change 30.0 20.0 10.0 0.0 0 -2 -1 0 1 2 0 Density Control Setting 0 Phototimer Field Placement / Balance Placement Measurement of Phot ot imer Field Placement / Balance cover desired field with lead select field as indicated Lead for checking field placement Balance no fields covered select field as indicated R R Tablet op Phototimer Field Placement / Balance Field Covered Left Center Right L& R Left 355 26.9 29.9 578 Field Placement Field Selected Center Right L& 23.2 29 242 25.4 24.3 610 18.3 266 Above readings in mR R 51 25.6 56.6 354 Field Balance 15 cm Lucite, 81 kVp Field mR Left 6.6 Center 4.9 Right 6.3 L& R 6.3 "Disc" units Phototiming checked with Exposure Index kV Response phototimer pick-up attenuation may vary with kV phototimer must track kV response of rareearth film Rate Response Check with varying phantom (lucite) thickness mA Measurement of Phot ot imer kV / Rat e Response Tablet op Fi l m kV/Rate Response kV 70 81 Lucite 17.5 4.5 Depth 12.5 4.7 7.5 4.7 (cm) 4.9 90 5.2 kV Response Optical Density Optical Density Thickness Tracking 4 2 0 17.5 12.5 7.5 Lucite Thickness 4 2 0 70 81 kilovoltage 90 The End Any questions, you varmints?