ACCREDITATION COMMISSION FOR CONFORMITY ASSESSMENT BODIES Suite # 113-114, Level 1, Master Mind IV, Royal Palms, Aarey Colony, Goregaon East, Mumbai – 400 065. India. Tel/Fax: 91-22-28794410, 28794411, 28794412 E-Mail: info@accab.org Website: www.accab.org ON SITE ASSESSMENT REPORT ISO 15189:2012 - Medical Laboratories – Requirement for quality and competence SUPPLEMENTARY CHECKLIST - SPECIALIZED TESTING INSPECTION CHECKLIST Part-I: GENERAL INFORMATION ACCAB Reference No.: Assessment Type: Date(s) of Assessment: Pre – Assessment Initial Assessment Re - Assessment Extension of Scope Short Notice Visit On-site Clearance Surveillance Visit Re-Instatement Visit Assessment Team: Assessor / Technical Expert(s):: Persons Interviewed: Laboratory Details: Laboratory’s Name: Address: Country: Telephone No. Email address: Principal Contact Name: Postcode: Fax. No. Web Site: Designation: CRA-F-05-15189-S-05-STIC RD-00-01/07/2014 This report shall not be reproduced in part without the permission of ACCAB Page 1 of 15 ACCREDITATION COMMISSION FOR CONFORMITY ASSESSMENT BODIES Suite # 113-114, Level 1, Master Mind IV, Royal Palms, Aarey Colony, Goregaon East, Mumbai – 400 065. India. Tel/Fax: 91-22-28794410, 28794411, 28794412 E-Mail: info@accab.org Website: www.accab.org PART - II CLAUSE NO. DETAILED CHECKLIST: SPECIALIZED TESTING INSPECTION CHECKLIST (Doc. ASSESSMENT Ref. / COMPLIANCE REQUIREMENT Clause Y/N/NA NOTES No.) 1.0 QUALITY CONTROL 1.1 Test systems using only external controls: Are two levels of external controls run each day of use? Test systems using internal controls: Are internal controls run each day of use? Has the laboratory validated the internal quality control? Are criteria established for frequency of performing external quality control? For quantitative tests, has a statistically valid target range (mean, SD, CV) been established for each lot of control material? For qualitative tests that use a cut-off value to distinguish positive from negative; is the cut-off value established initially? Are control specimens tested in the same manner and by the same personnel as patient samples? Is quality control data reviewed at least monthly by the laboratory director or designate? Is this review documented? Are results of controls verified for acceptability before reporting results? Are criteria for what is considered to be out of control available to staff? Are criteria for referral of “out of control” results to the supervisor and/or senior staff identified? Is there documentation of corrective action taken when controls exceed defined tolerance limits? Does the documentation include: What was out of control? Why the analysis was out of control? Corrective action taken? Signature/initials of individual responsible Are control data organized and presented so 1.1.1 1.2 1.2.1 1.2.2 1.3 1.4 1.5 1.6 1.7 1.8 1.9 1.10 1.11 1.12 1.13 1.13.1 1.13.2 1.13.3 1.13.4 1.14 CRA-F-05-15189-S-05-STIC RD-00-01/07/2014 This report shall not be reproduced in part without the permission of ACCAB Page 2 of 15 ACCREDITATION COMMISSION FOR CONFORMITY ASSESSMENT BODIES Suite # 113-114, Level 1, Master Mind IV, Royal Palms, Aarey Colony, Goregaon East, Mumbai – 400 065. India. Tel/Fax: 91-22-28794410, 28794411, 28794412 E-Mail: info@accab.org Website: www.accab.org 1.15 1.16 1.17 2.0 2.1 2.2 2.3 2.3.1 2.3.2 2.3.3 2.3.4 2.3.5 2.3.6 2.3.7 2.3.8 2.3.9 2.3.10 2.3.11 2.3.12 2.3.13 2.3.14 2.3.15 2.4 2.5 2.6 2.7 they can be evaluated daily by technical staff to detect problems, trends, etc.? Is lot number change of quality control material documented on the quality control record? Are quality control records retained for at least two years? Is quality control statistics (i.e. SD and CV) calculated at specified intervals to define analytical imprecision? PROCEDURE MANUAL Is a current procedure manual available for laboratory staff? Is the procedure manual maintained as per the documented procedure ? Are the following included for each procedure: (if applicable) Purpose Specimens (type, source, amount, storage) Equipment and Materials required Preparation and storage of reagents, standards and controls Procedure – including Step by step instructions Directions for calibration Derivation of results (ie. mathematical calculations, dilutions) Quality control Linearity limits Precision Reference ranges Clinical Significance Critical values Reporting results (units, stats, critical values) Safety Is there documentation of annual review of the procedure manual by the laboratory director or designate? Is there documentation of annual review of the procedure manual by the laboratory staff? Are all changes in methodology signed and dated by whoever made the changes? Are all new procedures reviewed by the medical director or designate? CRA-F-05-15189-S-05-STIC RD-00-01/07/2014 This report shall not be reproduced in part without the permission of ACCAB Page 3 of 15 ACCREDITATION COMMISSION FOR CONFORMITY ASSESSMENT BODIES Suite # 113-114, Level 1, Master Mind IV, Royal Palms, Aarey Colony, Goregaon East, Mumbai – 400 065. India. Tel/Fax: 91-22-28794410, 28794411, 28794412 E-Mail: info@accab.org Website: www.accab.org 2.8 2.9 2.10 3.0 3.1 3.2 3.2.1 3.2.2 3.2.3 3.2.4 3.3 4.0 4.1 4.2 4.2.1 4.2.2 4.2.3 4.2.4 4.2.5 4.2.6 4.2.7 4.2.8 4.3 Are discontinued procedures retained for two years after the procedure is taken out of service? Is there a current file on manufacturer’s inserts? Are there adequate and up to date reference text books available to laboratory staff? SPECIMEN COLLECTION MANUAL Is a specimen collection manual available at all collection sites? Does the specimen collection manual include: Patient preparation Specimen type and amount (ie. midstream) Proper handling (ie. preservative/ anticoagulant) Labelling with patient first and last name and personal health number Are there written criteria for specimen rejection? REQUISITIONS Are all specimens accompanied by an adequate requisition? Does the requisition include: Adequate patient identification – name and other identification Ordering physician or authorized person ordering the test Date of birth and gender Date and time of collection Tests requested Source of specimen Clinical information, history or clinical diagnosis, when appropriate Accession number Is the date and time that the specimen was received by the laboratory recorded? 5.0 REAGENTS 5.1 5.2 Are all reagents used within their expiry date? Are all reagents labelled with: CRA-F-05-15189-S-05-STIC RD-00-01/07/2014 This report shall not be reproduced in part without the permission of ACCAB Page 4 of 15 ACCREDITATION COMMISSION FOR CONFORMITY ASSESSMENT BODIES Suite # 113-114, Level 1, Master Mind IV, Royal Palms, Aarey Colony, Goregaon East, Mumbai – 400 065. India. Tel/Fax: 91-22-28794410, 28794411, 28794412 E-Mail: info@accab.org Website: www.accab.org 5.2.1 5.2.2 5.2.3 5.2.4 5.3 5.4 5.5 5.6 6.0 6.1 6.2 6.3 6.4 6.5 6.6 6.7 6.8 6.9 6.10 6.11 6.12 Date of receipt Date prepared or opened Expiry date Content and concentration Are all reagents stored according to manufacturer’s instructions? Are common interferences evaluated for all analytes measured with each reagent system or is credible manufacturer’s information available? If there are multiple components of a reagent kit, does the laboratory use components of reagent kits only within the same kit lot, unless otherwise specified by the manufacturer? Prior to use, are new reagent lots checked against old reagent lot? INSTRUMENTS AND EQUIPMENT Is there a procedure for maintenance of all instruments and equipment? Are the maintenance records reviewed by a supervisor? Is this review documented? Are there instructions for troubleshooting? Are service records maintained for the life of the instrument, plus two years? Is there emergency power for instruments and equipment? Are instruments equipped with surge protection? (this also includes computers) If the laboratory uses more than one instrument /method to test for a given analyte, are the instruments/methods checked against each other at least twice a year for correlation of results? Are there defined tolerance limits for result agreement of inter-instrument assays? Does the laboratory have a procedure for evaluating automatic pipette systems for carryover? Are pipettors and dilutors (fixed volume or adjustable) checked at least annually for accuracy and reproducibility, and results recorded? Is the temperature of water baths and/or heat blocks checked on days of use? CRA-F-05-15189-S-05-STIC RD-00-01/07/2014 This report shall not be reproduced in part without the permission of ACCAB Page 5 of 15 ACCREDITATION COMMISSION FOR CONFORMITY ASSESSMENT BODIES Suite # 113-114, Level 1, Master Mind IV, Royal Palms, Aarey Colony, Goregaon East, Mumbai – 400 065. India. Tel/Fax: 91-22-28794410, 28794411, 28794412 E-Mail: info@accab.org Website: www.accab.org 6.13 6.13.1 6.13.2 6.13.3 6.14 6.15 6.16 6.17 6.17.1 6.18 6.18.1 6.18.2 6.18.3 6.18.4 6.19 6.19.1 6.19.2 6.20 6.20.1 6.20.2 6.20.3 6.21 6.21.1 6.22 6.22.1 6.23 6.23.1 7.0 Is the temperature of refrigerators and other temperature dependent equipment documented daily: Room temperature Freezer Refrigerator Is there evidence of active review of results of instrument maintenance and temperature? Is there a procedure available if acceptable temperature ranges are exceeded? Is this review documented? Glassware Are there appropriate documented procedures for handling and cleaning glassware, including methods for testing for detergent removal? Microscopes Are there an adequate number of microscopes for the workload? Are all microscopes clean, well maintained and suitable for their intended use? Is there an adequate selection of objective lenses for the specimens examined? Is there a procedure for Koehler illumination? Centrifuges Is there a schedule for maintenance of all centrifuges? Are operating speeds checked and documented? Balances Are balances mounted on vibration-free benches? Are balances clean and regularly serviced? Are NIST reference weights available and used for checking accuracy? Autoclaves Is the autoclave monitored weekly for proper functioning? Pipettes Does your laboratory check pipettes for calibration? Thermometers Are certified thermometers used in the laboratory? TESTING CRA-F-05-15189-S-05-STIC RD-00-01/07/2014 This report shall not be reproduced in part without the permission of ACCAB Page 6 of 15 ACCREDITATION COMMISSION FOR CONFORMITY ASSESSMENT BODIES Suite # 113-114, Level 1, Master Mind IV, Royal Palms, Aarey Colony, Goregaon East, Mumbai – 400 065. India. Tel/Fax: 91-22-28794410, 28794411, 28794412 E-Mail: info@accab.org Website: www.accab.org 7.1 7.1.1 7.1.2 7.1.3 7.1.4 7.2 7.2.1 7.2.2 7.2.3 Calibration- is the set of operations that establish, under specified conditions, the relationship between reagent system and the corresponding concentration value of an analyte. Are criteria established for frequency of calibration? Is there documentation of calibration for each analyte? Are calibrators that have method and matrix appropriate target values used? Does your laboratory verify the calibration? Linearity – is the range of analyte values that a method can directly measure on the specimen without any dilution, concentration, or other pre-treatment not part of the usual assay process. Is linearity of the instrument validated initially and thereafter every six months? Is there documentation of linearity? Is there a procedure in place to ensure correct results are reported if the result falls outside of linearity? 8.0 SPECTROPHOTOMETERS 8.1 8.2 Is linearity checked at least annually? Is this documented? Is absorbency checked periodically with filters or standard solutions, if required by the instrument manufacturer? Are filters visually inspected annually? Is wavelength calibration checked at least annually? Is stray light checked at least annually? Is this documented? Are calibration curves verified bi-annually for procedures as well as after servicing or recalibration of the instruments? Is this documented? 8.3 8.4 8.5 8.6 8.7 8.8 8.9 9.0 ATOMIC ABSORPTION SPECTROPHOTOMETERS 9.1 Are the following maintenance procedures performed daily, prior to sample analysis: CRA-F-05-15189-S-05-STIC RD-00-01/07/2014 This report shall not be reproduced in part without the permission of ACCAB Page 7 of 15 ACCREDITATION COMMISSION FOR CONFORMITY ASSESSMENT BODIES Suite # 113-114, Level 1, Master Mind IV, Royal Palms, Aarey Colony, Goregaon East, Mumbai – 400 065. India. Tel/Fax: 91-22-28794410, 28794411, 28794412 E-Mail: info@accab.org Website: www.accab.org 9.2.3 Aligning the burner head along the light path Flushing aspirator and burner with water Verify and record the lamp energy Are the following maintenance procedures performed weekly: Clean the aspirator and line Check, adjust and record the flow rate using water as a sample Clean burner, chimney and optical surface 10.0 RADIOIMMUNOASSAYS 9.1.1 9.1.2 9.1.3 9.2 9.2.1 9.2.2 10.1 10.2 10.3 10.4 11.0 11.1 11.2 11.3 11.4 12.0 12.1 12.2 12.3 Are gamma counters or scintillation counters calibrated, results recorded and compared to previous values each day of use? Is background radioactivity determined each day of use, including the background in each well of multi-well counter? Are there documented criteria for acceptable (or unacceptable) background levels? Are counting times sufficiency long for statistical accuracy and precision? THIN LAYER CHROMATOGRAPHY (TLC) Are appropriate standards, calibrators and controls included with each TLC batch plate? Is a control extracted and run through the entire procedure for each run? Are negative and positive controls extracted and run through the entire procedure at least once a week? Are solvents prepared fresh as needed or if a commercial kit is used, are the manufacturer’s instructions followed? HIGH PERFORMANCE LIQUID CHROMATOGRAPHY (HPLC) Are calibrators or standards run with each analytic batch? Are controls extracted and run through the entire procedure? If a hydrolysis step is required, does the laboratory include a control? CRA-F-05-15189-S-05-STIC RD-00-01/07/2014 This report shall not be reproduced in part without the permission of ACCAB Page 8 of 15 ACCREDITATION COMMISSION FOR CONFORMITY ASSESSMENT BODIES Suite # 113-114, Level 1, Master Mind IV, Royal Palms, Aarey Colony, Goregaon East, Mumbai – 400 065. India. Tel/Fax: 91-22-28794410, 28794411, 28794412 E-Mail: info@accab.org Website: www.accab.org 12.4 12.5 12.6 12.7 13.0 13.1 13.2 13.3 13.4 13.5 13.6 13.7 13.7.1 13.7.2 13.7.3 13.7.4 13.8 14.0 14.1 14.2 14.3 14.4 Are new columns verified for performance before use? Is there a criteria established for monitoring the performance of the column and detector on each day of use? Is there a procedure for the detection of potential carryover? Is instrument performance (ie. retention times, detector response) checked after major instrument maintenance? GAS CHROMATOGRAPHY (GC) Are calibrators or standards run with each analytic batch? Are controls extracted and run through the entire procedure? If a hydrolysis step is required, does the laboratory include a control? Is there a procedure for detection and evaluation of potential carryover? Are new columns verified before use? Is there criteria established for monitoring the performance of the column and detector on each day of use? Are the following maintenance procedures performed on each day of use: Tank pressure Flow rates Septum changes Column changes Are gas lines checked regularly for leaks? MASS SPECTROMETRY (MS) Are mass spectrometers tuned each day of patient testing or according to manufacturer’s recommendations? Are tune records maintained? Are the identification criteria for single stage mass spectrometry (ie. CG/MS, LC/MS) in compliance with recommendations? Do the identification criteria for tandem mass spectrometry (MS/MS) comply with recommendations? CRA-F-05-15189-S-05-STIC RD-00-01/07/2014 This report shall not be reproduced in part without the permission of ACCAB Page 9 of 15 ACCREDITATION COMMISSION FOR CONFORMITY ASSESSMENT BODIES Suite # 113-114, Level 1, Master Mind IV, Royal Palms, Aarey Colony, Goregaon East, Mumbai – 400 065. India. Tel/Fax: 91-22-28794410, 28794411, 28794412 E-Mail: info@accab.org Website: www.accab.org 14.5 14.6 14.6.1 14.6.2 14.7 14.8 14.9 14.10 14.11 15.0 15.1 15.1.1 15.1.2 15.1.3 15.1.4 15.1.5 15.1.6 15.1.7 15.1.8 15.2 15.3 15.4 15.5 Does the laboratory’s assay procedure for LC/MS include an evaluation for possible ionsuppression? Are mass spectrometric identification performed by: Selected Ion Monitoring (SIC) Total Ion Current (TIC) Where applicable, are TIC identification criteria based on retention time and MS Library? Are SIM spectra identification criteria based on retention time? Is there documentation indicating the precision of each assay around the cut-off? Is there evidence that the limit of detection (sensitivity) and the linearity for quantitative methods has been determined for each procedure? Are there criteria for the detection of potential carryover? MATERNAL SCREENING Do the requisitions contain the following information: Collection date Last menstrual period (LMP) or estimated gestational age by ultrasound Maternal birth date Patient race Maternal weight History of insulin-dependent diabetes mellitus Clinical evidence of multiple gestations Initial screening sample or a repeat test for this pregnancy Is there documentation that the laboratory has established its own median values or verified that the manufacturer’s package insert or other source of medians is appropriate for the population being screened? Are medians recalculated or re-verified at least annually? Are the percentages of women with screenpositive test results for both neural tube defects and Down’s Syndrome calculated and reviewed at least quarterly? If calculations are performed on a computer is CRA-F-05-15189-S-05-STIC RD-00-01/07/2014 This report shall not be reproduced in part without the permission of ACCAB Page 10 of 15 ACCREDITATION COMMISSION FOR CONFORMITY ASSESSMENT BODIES Suite # 113-114, Level 1, Master Mind IV, Royal Palms, Aarey Colony, Goregaon East, Mumbai – 400 065. India. Tel/Fax: 91-22-28794410, 28794411, 28794412 E-Mail: info@accab.org Website: www.accab.org 15.6 15.6.1 15.6.2 15.6.3 15.6.4 15.6.5 15.6.6 15.6.7 15.6.8 15.6 15.7 15.8 15.9 15.10 15.11 15.12 15.13 15.14 16.0 16.1 16.2 there documentation that these calculations were initially verified for accuracy and reverified with any software updates or change? Is the following information included in the report: Date of birth Maternal weight Maternal race First day of last menstrual period or gestational age as determined by ultrasound examination Specimen draw date Initial or repeat specimen Presence of insulin dependent diabetes Presence of multiple gestation Are test results reported as multiples of the median (MoM)? Does the report classify a pregnancy as screenpositive or screen-negative for open neural tube defects, based on the maternal screen test results? Does the report classify a pregnancy as screenpositive or screen-negative for fetal Down’s Syndrome, based on the calculated risk? If an amniotic fluid sample has an elevated AFP MoM is the fluid checked for fetal blood contamination? Is at least one amniotic fluid dilution control processed with each analytic run of amniotic fluids Is acetylcholinesterase (AChE) testing performed on all amniotic fluids having elevated AFAFP concentration? If no, explain what samples are not tested for AChE: If AChE is run in-house, are both positive and negative controls included with each analytic run? If AChE is run in-house, are positive results confirmed by addition of a specific inhibitor? NEWBORN SCREENING Is the newborn screening program monitored to ensure 100% screening? Is collection of sample required before discharge? CRA-F-05-15189-S-05-STIC RD-00-01/07/2014 This report shall not be reproduced in part without the permission of ACCAB Page 11 of 15 ACCREDITATION COMMISSION FOR CONFORMITY ASSESSMENT BODIES Suite # 113-114, Level 1, Master Mind IV, Royal Palms, Aarey Colony, Goregaon East, Mumbai – 400 065. India. Tel/Fax: 91-22-28794410, 28794411, 28794412 E-Mail: info@accab.org Website: www.accab.org 16.3 16.4 16.5 16.6 16.6.1 16.7 16.8 16.8.1 16.8.2 16.8.3 16.9 16.10 16.11 17.0 17.1 17.2 17.3 17.4 17.5 17.6 17.7 17.8 17.9 17.10 Is transit time to the laboratory monitored? Are delayed samples tested? What is the number of delayed samples each month? Are unsatisfactory samples analyzed? If yes, please explain: What is the number of NSQ/ unsatisfactory samples each month Is a repeat screen requested for: All abnormal results Early discharge from hospital Data omission What is the number of delayed re-screens each month? Is there monitoring to ensure collection of all repeat samples? Is there a registry of abnormal results? SWEAT TESTING FOR CYSTIC FIBROSIS Is the sweat test offered only to patients at an appropriate age? Does the testing protocol require a sweat stimulation and collection from the patient’s lower arm or upper leg, using a site that is free from diffuse inflammation or rash? Does the procedure specify the conditions of iontophoresis? Does the procedure specify that iontophoresis is withheld from patients receiving oxygen by an open delivery system? Does the procedure specify that the area of iontophoretic stimulation is equivalent to the area of sweat collection? Does the procedure specify the parameters of sweat collection? Does the procedure specify that multiple insufficient sweat samples are rejected and not pooled for analysis? Does the laboratory report indicate the analytes measured in the sweat analysis and the reference ranges? If the test performed is a screening test, is it indicated on the report? If the test performed is a confirmatory test, is CRA-F-05-15189-S-05-STIC RD-00-01/07/2014 This report shall not be reproduced in part without the permission of ACCAB Page 12 of 15 ACCREDITATION COMMISSION FOR CONFORMITY ASSESSMENT BODIES Suite # 113-114, Level 1, Master Mind IV, Royal Palms, Aarey Colony, Goregaon East, Mumbai – 400 065. India. Tel/Fax: 91-22-28794410, 28794411, 28794412 E-Mail: info@accab.org Website: www.accab.org 17.11 18.0 the upper limit of reportable sweat chloride results less than or equal to 160 mmol/L? Are the personnel performing sweat test collection and analysis trained to ensure proficiency with collection and analysis procedures? PROFICIENCY TESTING 18.8 Proficiency Testing (PT) – a program where multiple samples are periodically sent to a group of laboratories for analysis and/or identification, where each laboratory’s results are compared with those of other laboratories in the group and/or with an assigned value. Is the laboratory enrolled in an external PT program? For analytes where graded PT is not available, is there an alternative assessment procedure in place? Is alternative assessment performed twice a year and documented? Does the laboratory integrate all PT samples within the routine workload? Are PT samples analyzed by personnel who routinely test patient samples? Is there documentation of corrective action of unacceptable PT results? Is there documented evidence of review by the laboratory director or designate of the external PT results? Are PT survey results retained for two years? 19.0 REPORTS 18.1 18.2 18.3 18.4 18.5 18.6 18.7 19.1 19.1.1 19.1.2 19.1.3 19.1.4 19.1.5 19.1.6 19.1.7 19.1.8 Do the report forms include: patient name and other identification physician or authorized person ordering test date and time of specimen collection test result(s) and units of measurement (when applicable) reference values date and time of report release conditions of specimen that may limit adequacy of testing name of laboratory performing the test(s) CRA-F-05-15189-S-05-STIC RD-00-01/07/2014 This report shall not be reproduced in part without the permission of ACCAB Page 13 of 15 ACCREDITATION COMMISSION FOR CONFORMITY ASSESSMENT BODIES Suite # 113-114, Level 1, Master Mind IV, Royal Palms, Aarey Colony, Goregaon East, Mumbai – 400 065. India. Tel/Fax: 91-22-28794410, 28794411, 28794412 E-Mail: info@accab.org Website: www.accab.org 19.1.9 19.1.10 19.2 19.3 19.4 19.5 19.5.1 19.5.2 19.5.3 19.5.4 19.6 19.7 19.8 19.9 19.10 19.11 19.12 19.13 19.14 19.15 19.16 specimen source specimen collected by Are there established criteria for immediate notification of critical test results? Is there documentation of notification to the proper clinical individual of results of all critical values? Does the laboratory have a policy regarding “read back” of critical values that are communicated verbally or by phone? Is confidentiality ensured for laboratory results received and distributed by the laboratory via: Fax Electronic Mail Mail Delivery Is there a procedure to identify and trace late/lost reports? Is there a policy for retention of patient reports? Does the laboratory promptly notify clinical personnel and issue a corrected report when errors are detected in patient test reports? Is there a documented system to ensure that all revised reports for previously reported incorrect patient results are identified as amended on all forms of patient reports? Is there a policy regarding the communication and documentation of reportable diseases? Are all patient results reported with reference ranges? Are reference ranges verified or established by the laboratory for the population being tested? Is there a stat list in place? Are stat results available within a reasonable time? Are routine results available within a reasonable time? Is the referral laboratory identified on the final report, if specimens are referred out? CRA-F-05-15189-S-05-STIC RD-00-01/07/2014 This report shall not be reproduced in part without the permission of ACCAB Page 14 of 15 ACCREDITATION COMMISSION FOR CONFORMITY ASSESSMENT BODIES Suite # 113-114, Level 1, Master Mind IV, Royal Palms, Aarey Colony, Goregaon East, Mumbai – 400 065. India. Tel/Fax: 91-22-28794410, 28794411, 28794412 E-Mail: info@accab.org Website: www.accab.org Bibliography: FAIR USE ACT 1976 NOTICE: This document may contain copyrighted material the use of which has not always been specifically authorized by the copyright owner. such material is made available to advance understanding of political, human rights, economic, scientific, moral, ethical, and social justice issues. this constitutes a 'fair use' of any such copyrighted material as provided for in section 107 of the 1976 us fair use copyright act. in accordance with title 17 U.S.C. section 107, this material is distributed without profit, to those who have expressed a prior general interest in receiving similar information for research and educational purposes. In the event that any content of this checklist causes harm/unlawful use/loss/unhappiness to anyone, the same should be brought to the notice of Accreditation Commission For Conformity Assessment Bodies Private Limited, Mumbai India info@accab.org 1. College of Physicians & Surgeons of Saskatchewan 2. College of Physicians & Surgeons of Alberta 3. Various other internet sources. CRA-F-05-15189-S-05-STIC RD-00-01/07/2014 This report shall not be reproduced in part without the permission of ACCAB Page 15 of 15