Cardiac Implantable Devices Nursing Care: The Basics and Beyond Welcome! Terri Rhodes, RN, BSN Clinical Level III, CEP Lab Nurse Laura Hess, RN, BSN Clinical Level II, CEP Lab Nurse Please feel free to ask questions during the presentation! Objectives: Examine device terminology Examine the components, functions and indications for a pacemaker Inventory the components, indications and functions of an internal cardiac defibrillator (ICD) Compare the pacing modes using NBG pacing code system Assess patient needs preoperatively Manage patient postoperatively Analyze rhythm strips for appropriate pacemaker and ICD functioning Outline 1. Welcome and general information 2. Pacemakers 3. ICD’s 4. NBG codes 5. Biventricular Pacing 6. Nursing Considerations 7. Pacemaker Practice Strips Normal Conduction System A Brief History of Implantable Devices 1958 - First human implant Dr. Senning in Stockholm, only lasted 3 hours 1960- First clinically successful human implant Dr’s Chardack and Gage in the US William Greatbatch, engineer 1965- First VVI implanted 1972- Partially programmable 1977-Multiprogrammable 1981- Dual chamber multi-programmable Along Came ICD’s… 1980 - First human implant Thoracotomy Epicardial patch & lead Large device placed in abdomen Not programmable; i.e. only one setting Second generation ICD Transvenous electrode Bradycardia & anti-tachycardia pacing Fifth generation Dual-chamber rate responsive pacing Improved recognition of SVT The Next Generation Remote interrogation CHF Management S-ICD- subcutaneous ICD General “Device” Terms to Understand Sense Fire Capture Sense Sense: the ability of the device to recognize the presence or absence of an innate “p” wave or “qrs” complex Fire Fire: the device has sensed a missed “p” wave or “qrs” complex, and has sent energy down the pacing wire to the tissue Capture Capture: the energy has contracted the myocardial tissue, and resulted in a “p” wave or “qrs” complex on skin leads Device Terms Continued… Failure to Capture: A spike is noted on strip, but is not followed by appropriate “p” or “qrs” wave form Failure to Sense Spike (energy) is missing during absence of “p” or “qrs” Spike noted at inappropriate times R on T Occurs when device fails to sense, and delivers energy during vulnerable T wave - or – if programmed at VOO/AOO, the pacemaker delivers the energy in spite of intrinsic activity and paces on the t-wave. Failure to Fire Device does not send energy (pacer spike) when indicated ***If you notice any of these, check your patient, check pulse and notify physician*** What Do You Need To Have a Paced Beat? Atrial Paced Beat: “a” pacing spike P wave immediately following pacer spike Ventricular Paced Beat: “v” pacing spike QRS immediately follows pacing spike Examples of Paced “a”, Paced “v”, and Both Pacemakers What is a pacemaker? A internal device that regulates electrical impulses through the heart. Sense Fire Capture Single Chamber, Dual Chamber and Bi-Ventricular Pacemaker Components Pulse generator- battery which provides the energy. Controls the rate, output, and sensitivity. The “Can” Leads- carries the impulse to the heart tissue Atrial Right Ventricle Left Ventricle Coronary Sinus Indications for pacemakers Symptomatic 2nd degree, Mobitz Type II heart block Complete heart block (3rd degree) Asystole Symptomatic bradycardia Sinus node dysfunction Carotid sinus syndrome and hypersensitivity An exaggerated response to carotid sinus baroreceptor stimulation. Sometimes even mild stimulation in the neck region causes a marked decrease in heart rate, blood pressure, and causes syncope. Other Indications Hypertrophic Obstructive Cardiomyopathy (HOCM) S/P Alcohol Septal Ablation Congestive heart failure (CHF) Biventricular pacing Magnet Placement for a Pacemaker Temporarily changes the mode of pacing to asynchronous (VOO, DOO) while magnet is in place. Paces regardless of rhythm This is programmable feature of the device; NOT ONE SIZE FITS ALL Break??? Intracardiac Cardioverter Defibrillators or ICD’s What is an ICD? An internal device that can regulate electrical impulses through the heart, but its main function is to detect and terminate tachy arrhythmias. Defibrillation Override pacing Cardioversion Pacemaker Functions (Single/Dual/BiV) Components of an ICD Pulse generator- battery which provides the energy. Detects tachy arrhythmias and delivers defibrillation energy when indicated. Controls the rate, output, and sensitivity of the pacemaker function. The “Can” Leads- carries the impulse to the heart tissue Right Ventricle Endo Coil – High output lead Atrium Pacemaker lead Left Ventricle Placed via the Coronary Sinus when placed in EP lab, and epicardial when placed in OR Unipolar ICD Indications for ICDs Secondary prevention (already had event) Sudden Cardiac Death; NSVT, Sustained VT, Vfib arrest Inducible VT (EP testing) Primary prevention (trying to treat FIRST event) Cardiomyopathy (SCD-HeFT) At risk for sudden cardiac death Unknown etiology Long QT Brugada Syndrome (Na channel abnormality resulting in RBBB with J point elevation and concave ST elevation) Cardiac Sarcoid And the Latest… S-ICD The S-ICD System is intended to provide defibrillation therapy for the treatment of lifethreatening ventricular tachyarrhythmias in patients who do not have: *symptomatic bradycardia *incessant VT *spontaneous, frequently recurring VT that is reliably terminated with anti-tachycardia pacing Which one do you want? Traditional ICD *Provides effective defibrillation for ventricular arrhythmias *Provides brady pacing *Provides ATP pacing *Provides atrial diagnostics *Familiarity of implant technique S-ICD *Provides effective defib for ventricular arrhythmias *No risk of vascular injury *Low risk of systemic injury *Preserves venous access *Avoids risk of endovascular lead extraction Magnet Placement for an ICD Suspends tachycardia detection while the magnet is in place Pacing parameters remain unchanged This is a programmable feature of the ICD, and may be different Caution! Place magnet on device ONLY under guidance or supervision from a physician or Electrophysiology Department nurse. Examples of when placing magnet is appropriate: ICD “ shocking” at inappropriate times During OR procedures requiring cautery. Stat pads must be placed on patient. During a code situation when you want to take ‘control of the shocking’ Special Considerations for Pt’s with ICD’s If ICD discharges? 1. Check your pt: Think BLS/ACLS! ABC’s, is pt. responsive, what rhythm are they in? Take appropriate action if pt. is not stable 2. If pt. is stable notify EP department During a CODE? DO NOT place STAT pads directly over device UCH policy: Place external defibrillator pads 4-6 inches away from the device laterally if possible. Pt. is going for another OR procedure Notify Anesthesia that pt. has device, tell them the company and they will notify the EP department Break? NBG Codes Generic code created for NASPE and BPEG. (NASPE is the North American Society of Pacing and Electrophysiology. BPEG is the British Pacing and Electrophysiology Group.) Pacemaker programming codes that identifies how the pacemaker is programmed to function. NBG Codes: Programming the pacemaker I- What chamber do you want to pace? II- What chamber do you want to sense? III-What do you want to do with the sensed information? Inhibit pacing or trigger pacing? Tracking the Atrial activity IV-Do you want to increase the rate with the patient’s activity? NBG Code Review I Chamber Paced II Chamber Sensed III Response to Sensing IV Programmable Functions/Rate Modulation V: Ventricle V: Ventricle T: Triggered P: Simple programmable A: Atrium A: Atrium I: Inhibited M: Multiprogrammable D: Dual (A+V) D: Dual (A+V) D: Dual (T+I) C: Communicating O: None O: None R: Rate modulating S: Single S: Single (A or V) (A or V) O: None O: None The NBG pacing code Position I II III Category Chamber(s) Paced Chamber(s) Sensed Response to Sensing Letters Used Manufacturer’s Designation Only IV Programmability, rate modulation O-None O-None O-None O-None A-Atrium A-Atrium T-Triggered R-Rate modulation V-Ventricle V-Ventricle I-Inhibited D-Dual (A+V) D-Dual (A+V) D-Dual (T+I) S- Single (A or V) S- Single (A or V) The NBG pacing code Position I II III Category Chamber(s) Paced Chamber(s) Sensed Response to Sensing Letters Used Manufacturer’s Designation Only IV Programmability, rate modulation O-None O-None O-None O-None A-Atrium A-Atrium T-Triggered R-Rate modulation V-Ventricle V-Ventricle I-Inhibited D-Dual (A+V) D-Dual (A+V) D-Dual (T+I) S- Single (A or V) S- Single (A or V) The NBG pacing code Position I II III Category Chamber(s) Paced Chamber(s) Sensed Response to Sensing Letters Used Manufacturer’s Designation Only IV Programmability, rate modulation O-None O-None O-None O-None A-Atrium A-Atrium T-Triggered R-Rate modulation V-Ventricle V-Ventricle I-Inhibited D-Dual (A+V) D-Dual (A+V) D-Dual (T+I) S- Single (A or V) S- Single (A or V) The NBG pacing code Position I II III Category Chamber(s) Paced Chamber(s) Sensed Response to Sensing Letters Used Manufacturer’s Designation Only IV Programmability, rate modulation O-None O-None O-None O-None A-Atrium A-Atrium T-Triggered R-Rate modulation V-Ventricle V-Ventricle I-Inhibited D-Dual (A+V) D-Dual (A+V) D-Dual (T+I) S- Single (A or V) S- Single (A or V) Single Chamber Pacing How Do We Use The NBG Language? VOO • Ventricular pacing Ventricular lead * • No sensing • Ventricular asynchronous pacing at lower programmed pacing rate • Used for: surgical procedures with cautery VVI • Ventricular pacing • Ventricular sensing Ventricular lead I * • Sensed intrinsic QRS inhibits ventricular pacing • Used if patient is in A-fib, do not want to tract the atrial rate AOO • Atrial pacing • No sensing Atrial lead * • Atrial asynchronous pacing at lower programmed pacing rate AAI • Atrial pacing • Atrial sensing Atrial lead * • Intrinsic P wave inhibits atrial pacing Indications: Sinus Node Dysfunction Dual Chamber Pacing Tracking Mode: Both triggers and inhibits pacing Benefits of Dual Chamber Pacing Provides AV synchrony Lower incidence of atrial fibrillation Lower risk of systemic embolism and stroke Lower incidence of new congestive heart failure Lower mortality and higher survival rates DDD Atrial lead Ventricular Lead • Pacing in both the atrium and ventricle • Sensing in both the atrium and ventricle • Intrinsic P wave and intrinsic QRS can inhibit pacing * I * • Intrinsic P Wave can “trigger” a paced QRS • Maintain AV synchronization DDD pacing Dual-chamber pacing capable of pacing and sensing in both the atrial and ventricular chambers of the heart 4 distinct patterns can be observed with DDD pacing DDD pacing Sensing in both the atrium and the ventricle (inhibiting in both the atrium and the ventricle) DDD pacing Pacing in the atrium with sensing (inhibition of pacing) in the ventricle DDD pacing Sensing in the atrium (inhibition of atrial pacing) and pacing in the ventricle Also known as “P wave tracking” DDD pacing Atrial pacing and ventricular pacing (no inhibition of pacing) DDD mode May resemble other modes of pacing Will strive to maintain AV synchrony with variable atrial rates and AV conduction Dual Chamber Timing Parameters Lower rate Upper rate intervals Lower Rate The lowest rate the pacemaker will pace the atrium in the absence of intrinsic atrial events Lower Rate Interval AP DDD 60 / 120 New Slide VP AP VP Upper Tracking Rate The maximum rate the ventricle can be paced in response to sensed atrial events Lower Rate Interval { Upper Tracking Rate Limit SAV AS VA VP DDDR 60 / 100 (upper tracking rate) Sinus rate: 100 bpm New Slide SAV AS VP VA Rate responsiveness/ adaptive-rate pacing The 4th Letter in the NBG Code Rate responsiveness/adaptive-rate pacing Rate response attempts to mimic the sinus node by increasing heart rate in response to increasing metabolic demand Rate responsiveness/adaptive-rate pacing Sensor(s) detect changes in physiologic needs and increase the pacing rate accordingly Rate responsiveness/adaptive-rate pacing The sensor detects changes by: Sensing motion (crystal or accelerometer) Sensing changes in intrathoracic impedance, e.g., minute ventilation DDDR pacing Example of Dual-Chamber RateResponsive pacing Biventricular PPM or ICD A Brief Overview of What It Means To BiV Pace Biventricular pacing Three lead system: Right atrial Right ventricular Left ventricular Biventricular pacing Cardiac Resynchronization Therapy (CRT) Patient Indications Bi-Ventricular ICD Moderate to severe HF (NYHA Class III/IV) patients Symptomatic despite optimal, medical therapy QRS 130 msec LVEF 35% Biventricular pacing Also known as cardiac resynchronization therapy, keeps the right and left ventricles pumping together by sending small electrical impulses to the heart muscle coordinating their contractions. The heart is able to fill and pump blood more effectively. This along with medical therapy, helps to improve heart failure symptoms. Improves quality of life in many. Biventricular pacing Achieved by: Inhibiting intrinsic ventricular rhythm Ensure pacing in RV and LV Short A-V delays to promote pacing in the ventricle Break? When Devices Go Bad!!!! Complications of Device Implantation: Pocket hematoma Pocket infection Pneumothorax Cardiac perforation Cardiac tamponade Vascular damage Lead dislodgement Lead fracture Lead infection Inappropriate shocks Laser Lead Extraction Program Implemented at UCH in 2008 by Chancey Weaver RN and Dr. Michelle Khoo M.D. First laser lead extraction in January 2009 ~30 leads extracted/year Reasons for a lead extraction: Fractured Leads Infected Lead(s) Non-functional leads/too many leads Regaining venous access Unexplained Dents! Device Erosion Lead Fracture Intraprocedure Extracted Lead Extracted Generator and Lead Nursing Considerations Preoperative ICD Placement and Postoperative Care Preoperative Left/right arm IV Reinforce patient and family education EP department performs education prior to and after procedure, any further questions, please call the EP lab NPO Surgical site Pre-op medications Antibiotics Blood work (WBC, Platelets, INR, Basic) Anesthesia in the procedure Restrictions after procedure Postoperative Vital signs Changes may indicate pericardial effusion or pneumothorax Type of device and settings ECG interpretation and documentation, as per unit guidelines Activity HOB <30 degrees for the first 4 hours Antibiotics Incision site X-ray within 1 hour of arriving back in room and X-ray in AM as well **Pt. placed in sling for 24 hours to allow leads to adhere to tissue** Documentation According to hospital policy: University of Colorado Hospital Call report to telemetry: Include device manufacturer and model number, mode (VVI, DDD, etc.), and lower and upper programmed rates (should be given in report). Place in computerized documentation: Device manufacturer, mode, rate, rate cut off, therapies, and date of implant. If the device fires, document any therapies of the device including the precipitating dysrhythmia and outcome in your charting. Include ECG strips, if available, documenting the dysrhythmia, the delivery of the therapy via the ICD and the resultant rhythm and the patient response. Strip Documentation According to hospital policy, and individual unit guidelines. Minimal information includes “running” a strip every 12 hours or with a change in rhythm Documentation: date, time, patient's name, medical record #, heart rate, PR, QRS, and QT intervals, and rhythm analysis Patient and Family Education Wound care S/S of infection No submersion under water for 3 weeks No direct water spray (shower spray) for 1 week Coughing and deep breathing Activity All information in Post Op packet NO lifting arm above shoulder for 6 weeks Follow-up appointment Remote interrogations Electromagnetic interference Identification card Patients Admitted With a PPM or an ICD Patients admitted with a PPM/ICD Ask patient for device information, i.e. registration card EP does not need to be consulted if a patient is admitted for a non-device related problem and the device appears to be working appropriately. MRI not recommended (except new Medtronic PPM) Pre-op/Post-op patients may require device programming changes ICD- tachy therapies off, or may fire during cautery PPM- reprogram to VOO, or may fail to pace appropriately Pacemaker Practice Strips What You Need to Document Underlying rhythm? Is it “a” paced, “v” paced or both Is the device doing what it is programmed to do? Troubleshooting Failure to: Sense Fire Capture How to interpret a “paced” strip: One method of many… 1. 2. 3. 4. Is intrinsic activity present? Are pacing spikes present: “A”, “V”, or both? Is 1:1 capture present? Is intrinsic activity sensed appropriately? Over sensing- sensing of an inappropriate single Leads to underpacing Under sensing- failure to sense intrinsic cardiac signal Results in overpacing 5. What is the heart rate? 6. What is the programmed pacing rate? Compliments of Northwestern Memorial Hospital, October 2002 Is This Normal Device Operation? Is This Normal Device Operation? What Device Operation is This? Is This Normal Device Operation? Is This Normal Device Operation? What is missing? Thank You Cardiac Electrophysiology Dena Keilman, RN Kari Jackson, RN Noelle Hernandez, RN Amanda Lange, RN Heidi Huber, RN Terri Rhodes, RN Dan Sullivan, RN Claire Rutherford, RN Matt Upton, RN Laura Hess, RN Diane Ridgway, RN Ann Czyz. RN William H. Sauer, MD Duy Nguyen, MD Paul Varosy, MD Ryan Aleong, MD Joe Schulller, MD Wendy Tzou, MD Christine Tompkins, MD David Katz, MD Cathy Kenny, ANP References Burke M, et al. Safety and Efficacy of a Subcutaneous ImplantableDefibrillator (S-ICD System US IDE Study). Late-Breaking Abstract Session. HRS 2012.