Mechanized Overseer Of CVD In Integrate And Fire Pulse Train Contrivance R.Sunil Kumar PG Student, Department of Electronics & Communication, Vellammal Engineering College, Anna University, Chennai Dr.V.Latha Professor, Department of Electronics & Communication, Vellammal Engineering College, Anna University, Chennai ABSTRACT The presence of Cardiovascular Disease(CVD) has been identified by QRS frequency bands present in the Electrocardiogram(ECG) signal. Automatic diagnosing of CVD from ECG is a tedious process, therefore overseer of QRS segment is a basic to ECG mechanizing. Persistent, portable 24/7 ECG mechanizing requires wireless technology with constraints on power, bandwidth, area, and resolution. In order to provide continuous remote mechanizing of patients and fast transmission of data to medical personnel for instantaneous intervention, here methodology was proposed that converts analog inputs into pulses for ultralow power implementation. The signal encoding scheme is the time-based integrate and fire (IF) sampler from which a set of signal descriptors in the pulse domain are proposed. Furthermore, a logical decision rule for QRS segment detection based on morphological checking is derived from which the presence of CVD is detected. The algorithm was evaluated using the Massachusetts Institute of Technology Beth Israel Hospital(MIT-BIH) arrhythmia database and results show that our algorithm performance is comparable to the state-of-the art software-based detection. Key words Cardiovascular Disease(CVD, Electrocardiogram(ECG), Massachusetts Institute of Technology Beth Israel Hospital(MIT-BIH), integrate and fire (IF) I.INTRODUCTION Cardiovascular disease which has became the great threat for the people to cause death in this fast developing world. The death rate due to CVD is 1 in every 2.8 deaths[1] also called heart disease is a class of diseases that involve the heart, the blood vessels (arteries, capillaries, and veins) or both.CVD refers to any disease that affects the cardiovascular system, principally cardiac disease, vascular diseases of the brain and kidney, and peripheral arterial disease. The heart beat in the cardiac cycle of each individual can be recorded in the ECG waveform. The CVD can be identified by analysing this recorded ECG waveform[2],[3]. This analysis is a crucial task . ECG gave a clinical information regarding the heart beat rate, morphology and the proper functioning of heart at cheap rate and non-invasive test[4]. Automated ECG interpretation is the use of artificial intelligence and pattern recognition software and knowledge bases to carry out automatically the interpretation, test reporting and computer-aided diagnosis of electrocardiogram tracings obtained usually from a patient. The first automated ECG programs were developed in the 1970s, when digital ECG machines became possible by third generation digital signal processing boards. The cardiac monitoring industry has seen momentous advancements such as event monitors, AF auto trigger monitors and LOOP recorders with extended memory, and also mechanical atypical event revealing arrhythmic events [5]. The progress that promoted this power and flexibility is the use of digital signal processing chips that implement algorithms for signal detection. The hazard of these monitors is that they are still relatively bulky, inconvenient to carry around, and un-comfortable to wear on an extended basis. They also involve the recurrent stand-in of batteries as power consumption is considerable. To outwit these problems, the devices should drive with ultra-low power and be incorporated flawlessly in the day to day life of the patient, which involve advance technology innovation. For a digital signal processing elucidation, the first step is to sample the data uniformly at the Nyquist rate. This way of outmoded sampling leads to huge circuits with huge burning up of power. .Additionally, the limited use of DSP chips is inefficient because only a very small portion of its electronics is used at a given time. Although the power consumption has been progressively diminishing, one cannot look forward to foremost gains in sliding power trending because the silicon technology is in your prime. To diminish the problem of bulkier circuits, other novel sampling schemes such as compressive sensing [6] and the finite rate of originality [7] have been proposed. These methods work directly on the scrubby representations of the input and combine the density and sampling stages dropping greatly the required data rates for renovation. To gratify the ultra-low power constraints, time-based analog to pulse converters such as integrate and fire samplers (IF) have been proposed [8]. This pulse representation is as accurate as conventional A/Ds because it provides an injective mapping between analog signals and the pulses [8]–[11], i.e., it is an alternative to conventional Nyquist samplers. It also has an wellorganized hardware functioning with tiny form factor [11]. Thus, IF fulfils every bit of the constraints of wearable/portable health care arrangement such as low power, area, bandwidth, and resolution. In this paper, we propose a scheme that quantifies the time arrangement of the pulses and can be fully implemented in devoted combinatory logic, potentially diminishing the power consumption. The anticipated plot builds a set of attributes using finite-state gadget to map the pulse train into the ECG morphological elements using combinatory logic decision blocks. Thus, our suggestion is a going away from the signal processing approach based on numbers, that is enabled by the injective mapping characteristics of the IF (i.e., there is practically no loss of information) and guarantees ultralow power consumption. This provides much better trade-offs between power and accuracy. II.ANTICIPATED PROCESS The implementation of the anticipated scheme can be explained in six successive steps. The below flow chart clearly explains the processing steps. Fig 2 pre-processing of ECG signal The ECG signal is conceded through a two median filter which has two different window size one with window size 200 ms which removes the P waves. Then, an another one with a window size 600 ms removes the T waves. The filtered signal represents the baseline which is then subtracted from the novel ECG recording. Finally a notch filter cantered at 60 Hz is implemented through a 60 tap finite impulse response filter to remove power line interference [10]. Fig 1. Flow chart for the anticipated system Fig 3 The stage after the ECG signal has been passed through two levels on median filter. Which will remove the non-QRS region of the signal A.ECG SIGNAL All the ECG signal which we have used in this paper are selected from MIT-BIH Arrhythmia database[12], which as created in 1980 as a standard reference for arrhythmia detector[13]. In practice the date are collected from the 24/7 portable device which is attached with the patient's body. The database is comprised of 48 files each containing 30-minutes ECG segments selected from 24 hours recording of 47 different patients[12]. B.PRE-PROCESSING Pre-processing involves the 4 steps a)Noise removal- Electromyographic noise b)Signal Processing- Shaping the Data's Character c)Multiple Removal- Removing Unwanted Coherent Energy d)Statics Corrections- Removing Topography and Near-Surface Effects e)Removal of power line interference To spot a structure mold in the frequency domain by using time-domain data then the time-domain data must be preprocessed by estimating the frequency response function (FRF) is necessary. The pre-processing applied here can be shown below C.SAMPLING Traditional signal processors use analog to digital converters (ADC) to embody a given signal using standardized sampling, which results on a worst case condition which means Nyquist criterion can be called as a symbol of a band limited signal. This type of traditional input dependent samplers contemplate on the high-amplitude regions of interest in the signal and under signify the relatively low amplitude noise which in turn will reduce the overall bandwidth to sub-Nyquist rate.The IF model is stimulated by a simplified biological neuron operation from computational neuroscience as shown below. Fig 4 Block diagram of IF sampler Latest research has made known that the IF model can be measured a sampler [8]–[11]. The block diagram of the IF sampler is shown in Fig. 4. The uninterrupted input signal which is collected from the patient body is convolved with an averaging function u(t) with the known starting time and for a particular intervals which describes its threshold limit. The integrator is retuned as well as assumed at this situation for a explicit duration given by the refractory period to avoid two pulses from being too close to each other and then the process repeats. One of the most important merit of this sampler is the minimalism of the hardware circuitry [11] which makes it a apt device for low-power applications. assessment logic. The logical transformation is composed of three main components: 1) Morphological checking, where the morphological descriptors are transformed. A multilevel transformation plot is done to analyze morphological conditions of the ECG under different heart rhythms. In an IF sampled ECG signal, since the slope is encoded in the IPI of the pulses, evaluation of the minimum IPI of the pulse segment against a revealing threshold would be enough to determine whether it is a QRS pulse segment or not. Though it seems good but in practice due to the natural variability of heart rhythms, if assessment was done with respect to a single revealing threshold, then there is a chance of falling into the false winding up, so it is advisable to replicate the above process for two or more threshold values.. 2) Physiological blanking, where blanking descriptors are altered to point out the strong incentive. : whenever a QRS complex has been detected, there is a physiological qualified refractory period of about to 350 ms prior to the happening of next QRS complex. However, in practice some rhythms like ventricular tachycardia has an exception to this condition [17]. 3) search back, where discriminators and the search back markers are distorted. A quick and a efficient decision is made in this section which will paves the way to fetch the conclusion about the QRS region. Fig 5 The sampled output, here Integrate & Fire pulse sampling are applied instead of applying the traditional sampling method, which will convert the input signal into pulses and then the pulses are grouped together respective to their polarity. D.QRS DETECTION The preprocessed ECG signal is passed into a sequence of pulses using the IF sampler. The pulse train is aggregated online into altered pulse segments and each segment is represented by a set of features which hand out as descriptors of the pulse train. The anticipated decision logic is based on morphological checking. The features of the pulse segment are altered into a deposit of logical values. Based upon the logical values, different automata-based decision rules are executed and the QRS complexes are detected. These steps are grouped under three sections 1) descriptors, which explain the morphological setting, 2) markers, which commence a decision rule, and 3) discriminators, which differentiate between noise and isoelectric deviations. To calculate the IF pulse train of the ECG signal, a notion is defined for the pulse segment for the pulse train, in which each pulse slice is arranged by pulses respective to its polarity. The pulse segments of positive and negative polarity correspond to positive and negative waves of ECG, respectively. The pulse segments of interest in this paper is chosen as the QRS pulse segments. E.LOGICAL TRANSFORMATION AND DECISION ASSIGNMENT To enumerate the morphological structure of the ECG waves, the features of each individual pulse segment are compared against detection thresholds and converted to logical values either in the form of zeros and ones. The renovation captures information from a mixture of morphological configurations of QRS waves and the logical values serve as indicators of certain morphological conditions, which plays the major role in the anticipated Fig 6 Extracted QRS shows that R-peaks are clearly detected and its denoted by red dots, through which a further classification of CVD can be made. III.RESULTS AND DISCUSSION A.CHOICE OF DETECTION THRESHOLDS Choice of detection thresholds is of supreme importance in judgment making. Due to wide inter- and intra-patient inconsistency, habitual estimators will yield clear-cut accuracy only in certain segments of the ECG signal. Therefore, we used the Minnesota ECG coding manual [14], which provide logical criteria for happening of cardiac events, to fix the detection thresholds. B.ACCURACY The sample signal is taken from MIT-BIH Arrhythmia database and the results are compared with living methods. ACCURACY - It is the major parameter in obtaining the overall success of the proposed method. this can be defined as 𝐴𝑐𝑐𝑢𝑟𝑎𝑐𝑦(𝜌) = 𝑁𝐵 − 𝑒 𝑁𝐵 ---(1) Where NB denotes the total number of beats and 'e' denotes th total number of classification errors. compassion and specificity - These are the statistical measures of the performance of a binary classification test, also known as statistics' as classification function. 𝑐𝑜𝑚𝑝𝑎𝑠𝑠𝑖𝑜𝑛(𝛼) = 𝑆𝑝𝑒𝑐𝑖𝑓𝑖𝑐𝑖𝑡𝑦(𝛽) = 𝑛𝑢𝑚𝑏𝑒𝑟 𝑜𝑓 𝑡𝑟𝑢𝑒 𝑝𝑜𝑠𝑖𝑡𝑖𝑣𝑒 𝑛𝑢𝑚𝑏𝑒𝑟 𝑜𝑓 𝑡𝑟𝑢𝑒 𝑝𝑜𝑠𝑖𝑡𝑖𝑣𝑒 + 𝑛𝑢𝑚𝑏𝑒𝑟 𝑜𝑓 𝑓𝑎𝑙𝑠𝑒 𝑛𝑒𝑔𝑎𝑡𝑖𝑣𝑒 𝑛𝑢𝑚𝑏𝑒𝑟 𝑜𝑓 𝑡𝑟𝑢𝑒 𝑛𝑒𝑔𝑎𝑡𝑖𝑣𝑒 𝑛𝑢𝑚𝑏𝑒𝑟 𝑜𝑓 𝑡𝑟𝑢𝑒 𝑛𝑒𝑔𝑎𝑡𝑖𝑣𝑒 + 𝑛𝑢𝑚𝑏𝑒𝑟 𝑜𝑓 𝑓𝑎𝑙𝑠𝑒 𝑝𝑜𝑠𝑖𝑡𝑖𝑣𝑒 ----(02) ---(03) IV.CONCLUSION AND FUCTURE WORK The anticipated method uses a new way of processing by using the pulse train which is generated by the IF sampler. The IF representation maps the analog input signal into a pulse train which assures the trustworthiness similar to the traditional samplers and at a same time it allows for a nonnumeric signal processing line of attack which can be implemented with a few hundred combinatory logic and flip flop blocks. This paves the way for ultra small and ultralow power devices that can be encapsulated with the electrodes In future it is required to create the first archetype for clinical trials, this study shows the feasibility of this line of research and also there is a need to analyze the PQ segment and ST segment which is non QRS region because a QRS region is alone not enough to detect the presence of CVD for a particular person. In a future work, we would like to explore ECG delineation and beat classification using the proposed attribute based framework for the entire ECG signal. REFERENCES [1] American Heart Association. Heart Disease and Stroke Statistics—2009 Update. Dallas, TX: American Heart Association, 2009. [2] L. Schamroth, An Introduction to Electrocardiography, 7th ed. New York, NY, USA: Wiley, 2009. [3] Goldberg, Clinical Electrocardiography, 7th ed. Amsterdam, The Netherlands: Elsevier, 2010. [4] Dubin, Rapid Interpretation of EKG’s, 5 ed. Tampa, FL: Cover,2000. [5] The Evolution of Monitoring Slides, 2013, May. [Online]. Available: https://www.cardionet.com/medical_08.htm [6] E. Candes, J. Romberg, and T. Tao, “Robust uncertainty principles: Ex-act signal reconstruction from highly incomplete frequency information,” IEEE Trans. Inform. Theory, vol. 52, no. 2, pp. 489–509, Feb. 2006. [7] M. Vetterli, P. Marziliano, and T. Blu, “Sampling signals with finite rate of innovation,” IEEE Trans. Signal Process., vol. 50, no. 6, pp. 1417–1428, Jun. 2002. [8] H. Feichtinger, J. Principe, J. Romero, A. Alvarado, and G. Velasco, “Approximate reconstruction of bandlimited functions for the integrate and fire sampler,” Adv. Comput. Math., vol. 36, pp. 67–78, 2012. [9] A. Alvarado, J. Principe, and J. Harris, “Stimulus reconstruction from the biphasic integrate-and-fire sampler,” in Proc. 4th Int. IEEE/EMBS Conf. Neural Eng., May 2009, pp. 415–418. [10] A. Alvarado, C. Lakshminarayan, and J. Principe, “Time-based compres-sion and classification of heartbeats,” IEEE Trans. Biomed. Eng., vol. 59, no. 6, pp. 1641–1648, Jun. 2012. [11] M. Rastogi, A. S. Alvarado, J. G. Harris, and J. C. Principe, “Integrate and fire circuit as an ADC replacement,” in Proc. IEEE Int. Sym. Circuits Syst., May 2011, pp. 2421–2424. [12] G.B. Moody and R. G. Mark, “The impact of the mit/bih arrhythmia database,” IEEE Eng. Med. Biol. Mag., vol. 20, no. 3, pp. 45–50, May-Jun. 2001 [13] S. Banerjee, R. Gupta, and M. Mitra, “Delineation of ECG characteristic features using multiresolution wavelet analysis method,” Measurement, vol. 45, no. 3, pp. 474–487, Apr. 2012. [14] R. J. Prineas, R. S. Crow, and Z. Zhang, The Minnesota Code Manual of Electrocardiographic Findings, 2nd ed. London, U.K.: Springer-Verlag, 2010.