- University of Portsmouth

advertisement
1
2
Hip joint degeneration due to cam impingement: A finite element analysis
3
F.L. Hellwiga, J. Tonga*, J.G. Hussellb
4
5
6
7
8
9
10
11
12
13
14
aMechanical
Behaviour of Materials Group, School of Engineering, University of Portsmouth, Portsmouth, UK
b
Orthopaedics and Trauma Centre, Queen Alexandra Hospital, Portsmouth, UK
* Corresponded Author Tel.: 0044(0)2392842326.
Email address: Jie.Tong@port.ac.uk
15
Page 1
16
Hip joint degeneration due to cam impingement: A finite element analysis
17
The goal of this study was to investigate the impact of cam impingement, a
18
biomechanical risk factor, on hip joint degeneration and ultimately coxathrosis.
19
3D finite element solid models of a healthy and a pathologic hip were developed
20
based on clinical reports. The biphasic characteristics of cartilaginous tissues
21
were considered to identify localised solid matrix overloading during normal
22
walking and sitting down. Localised femoral intrusion at the anterior-superior
23
pelvic horn was revealed in the pathologic hip during sitting down, where the
24
radial
25
circumferential solid-stresses within the acetabular labrum increased by 3.7, 1.5
26
and 2.7 times, respectively. The increased solid-on-solid stresses, reduction in
27
fluid-load support and associated higher friction during articulation may result in
28
joint wear and other degenerative changes in the hip.
29
Keywords: Hip Joint, Cartilage Degeneration, Biphasic Model, Contact
30
Pressure, Finite Element, Impingement
and
meridional
solid-stresses
31
Page 2
in
the
acetabular
cartilage
and
32
Introduction
33
Excessive contact stresses in the hip joint due to obesity (Rečnik et al. 2009) or decreased
34
weight bearing area typical in a dysplastic hip (Mavčič et al. 2002; Russell et al. 2006) are
35
thought to result in coxathrosis. Recent clinical studies hypnotised a link between abnormal
36
anatomical conditions of femoral head-neck junction and early degenerative changes of the
37
hip joint, such as osteochondral defects and labral tear leading ultimately to idiopathic
38
osteoarthritis (Ito et al. 2001). The pathologic condition, named cam type femoroacetabular
39
impingement, is caused by jamming of a non spherical extension of the femoral head into the
40
acetabular cavity (Ganz et al. 2003, 2008). Nötzli et al. (2002) characterised this anatomic
41
abnormality by introducing an alpha angle (Figure 1). This angle is measured between the
42
femoral neck axis and a line extended from the point C, the femoral head centre, to point D,
43
which is the point of deviation from femoral head sphericity. The abnormality might emerge
44
as an anterior bump, spigot or thickened femoral neck, limiting the range of motion (ROM)
45
(Ito et al. 2001; Ganz et al. 2003, 2008).
46
Recently, a finite element study (Chegini et al. 2009) of normal and impinged
47
articulation has been presented to examine the joint mechanics during daily activities of
48
normal walking and sitting down, for which the maximum contact force was reported as
49
233% and 156% bodyweight, respectively (Bergmann et al. 2001). Chegini et al. (2009)
50
focused on the total stress responses of the cartilaginous tissues, which were represented as
51
isotropic linear elastic materials. The cartilage and labrum, however, are fully fluid saturated
52
porous mediums (Mow et al. 1980), hence the load is shared between the interstitial fluid and
53
a solid extracellular matrix (ECM):
54
πœŽπ‘‘π‘œπ‘‘ = πœŽπ‘ π‘œπ‘™π‘–π‘‘ + π›₯𝑝𝐼
55
where πœŽπ‘‘π‘œπ‘‘ is the total stress of the tissue, πœŽπ‘ π‘œπ‘™π‘–π‘‘ is the stress acting on the ECM and π›₯𝑝𝐼 is the
56
pore pressure of the interstitial fluid (Mow et al. 1980). Early macroscopic changes in
(1)
Page 3
57
articular cartilage were reported to begin with fraying at the superficial zone of the ECM
58
(Pritzker et al. 2006), a sign of joint wear accelerated with increased friction (Mow and
59
Huiskes 2005) and associated with decreased fluid-load support (Krishnan et al. 2004).
60
Ferguson et al. (2000a) utilised an isotropic poroelastic material model in ABAQUS (Wu et
61
al. 1998) to simulate the sealing function of the acetabular labrum.
62
alterations in solid-on-solid stress with the changes in fluid-load support for a static load up to
63
75% bodyweight held constant for up to 10,000s. More recently, in order to study the
64
cartilage response to elevated loading during unstable motion, Goreham-Voss et al. (2007)
65
modelled the articulating cartilage layers in knee joints, considering the directional-dependent
66
material properties of the extracellular matrix.
They correlated
67
The aim of this work is to study the stresses in the cartilage in the principal directions
68
of an orthotropic ECM and the fluid-load support during normal and 'cam'-type impinged
69
articulations under selected physiological loading conditions. To do so the cartilaginous
70
tissues were modelled as biphasic orthotropic materials and the results for normal and 'cam'-
71
type impinged articulation were compared.
72
Page 4
73
Numerical Methods
74
Geometry and Materials
75
A normal hip joint model with alpha = 40° was developed based on morphologic
76
simplifications and dimensions of Chegini et al. (2009) (Figure 2(a)), whilst a typical cam-
77
impinged hip joint with alpha = 74° was also modelled based on Nötzli et al. (2002) (Figure
78
2(b)), using ABQUS CAE. The finite element model, depicted in Figure 2(c), of the normal
79
hip joint (c) consists of acetabular labrum (1), acetabular cartilage (2), femoral cartilage (3),
80
impermeable membrane (4) and femoral head neck section (5), following (Chegini et al.
81
2009). The pathologic case was identical except the anterior hump at the head-neck junction.
82
Femoral and pelvic bones were considered rigid (Anderson et al. 2008) and the
83
cartilaginous tissues were considered as fully fluid saturated porous mediums (Mow et al.
84
1980), taking into account the orthotropic structure of the ECM following Goreham-Voss et
85
al. (2007). A spherical coordinate system was used to allocate the directional-dependent
86
material properties. The permeabilities of the femoral and acetabular cartilage and labrum are
87
9.15, 8.98 and 4.98 x 10-4 mm4/Ns, respectively (Athanasiou et al. 1994; Ferguson et al.
88
2001), and they were converted from the biphasic, k, to the poroelastic model, k', via:
89
k' = 𝛾k
90
where the volumetric weight of the interstitial water, 𝛾, was taken as 9.81 x 10-6 N mm-3 (Wu
91
et al. 1998). The solid matrix compressive radial moduli '1'of femoral and acetabular cartilage
92
and labrum were averaged to 1.23, 1.18 and 0.157 MPa (Athanasiou et al. 1994; Ferguson et
93
al. 2001), respectively. For cartilage, the tangential moduli in circumferential '2' and
94
meridional '3' directions were extracted from the toe region as 8.5 MPa (Roth and Mow,
95
1980). To permit opening between the acetabular labrum and femoral cartilage under load, a
96
tangential moduli of 26 MPa was chosen in direction '2', close to that of the toe region (20
(2)
Page 5
97
MPa) reported in Ferguson et al. (2001). According to the ratio of meridional to
98
circumferential strength of the meniscus (Tissakht and Ahmed, 1995), the tangential modulus
99
of the labrum in direction '3' was set to 3 MPa, since labrum and meniscus exhibit a similar
100
ultra structure (Ferguson et al., 2001). As shear stresses peak at the cartilage-bone interface
101
(Goreham-Voss et al. 2007), a shear modulus of 2.5 MPa (Wong et al. 2008) was chosen.
102
Shear modulus and volumetric response (ν12, ν23) (Athanasiou et al. 1994; Federico et al.
103
2005) are given for cartilage in Table 1, similar values were assigned to the labrum due to the
104
lack of data. A Summary of all material properties used in the models is given in Table 1.
105
As tissue properties were taken from different sources due to limited availability, a
106
pre-study on the sensitivity to material property variation (± 50%) was performed. The peak
107
contact force for normal walking, taken from Bergmann et al. (2001), was applied to the hip
108
joint centre during a 1 second ramp phase for the normal articulation.
109
Finite element modelling
110
The cartilaginous tissues were meshed with first order pore pressure brick elements, and the
111
femoral-head neck section and the impermeable membrane were meshed with quadrilateral
112
bilinear rigid and quadrilateral membrane elements, respectively. Reduced integration was
113
used for deformable elements to reduce the calculation time. The selected grid sizes and
114
element number are given in Table 2.
115
Since the pelvis was assumed rigid the nodes of the acetabular cartilaginous tissue on
116
the bone-cartilage interface were fully constrained. The hip joint forces for normal walking
117
and sitting down were taken from Bergman et al. (2001) and applied incrementally as a point
118
force to the centre of the femoral head. The hip joint force was synchronized with the
119
locomotion data prescribed as rotational movements of the femur. The analysis was
120
performed by first establishing contact between the femur and the acetabulum, then applying
121
an initial load of 33.31% or 55.51% of the peak load for normal walking (NW) and sitting
Page 6
122
down (SD), respectively, ramped over 1 second, and finally simulating the locomotion cycle
123
in 60 sequences. The bodyweight was assumed 86 kg.
124
Contact was implemented using the penalty method assuming frictionless finite
125
sliding, due to small coefficient of friction (Pawaskar et al. 2007). Throughout the analysis the
126
femur was the master and the acetabulum the slave surface. Further, for poroelastic mediums
127
the contact model should satisfy the biphasic jump condition, where in areas of contact fluid
128
velocity should be continuous or otherwise interstitial fluid should be free-draining (Hou et al.
129
1989). However, as contact pressure varies, local rehydration occurs (Pawaskar et al. 2007)
130
and fluid efflux and cross-flow are minimal for loadings such as NW and SD (Pawaskar et al.
131
2011), hence the biphasic jump condition was not considered.
132
Since in a biphasic medium load is shared between the fluid and the solid phase (Mow
133
et al. 1980), the total contact pressure is carried by both phases. However, in the ABAQUS
134
model for porous mediums in contact, the fluid phase contribution to the total contact pressure
135
is ignored to ensure point-wise continuity of the pore pressure at opposing sides (Hibbitt et al.
136
2007). On the other hand, contact pressure between a porous and an impermeable material
137
includes contributions from both solid and fluid phase (Hibbitt et al. 2007). In this work, the
138
total contact pressure was obtained by inserting an impermeable membrane (E = 1 MPa; ν =
139
0.49) between the cartilaginous tissues (Wilson et al. 2003; Manda et al. 2011). This
140
membrane has only a minor effect on contact pressure results (Wilson et al. 2003).
141
Page 7
142
Results
143
Normal articulation
144
For both normal walking and sitting down, the weight bearing area, peak contact and peak
145
pore pressure vary during the course of the simulated activity, and the variation is consistent
146
with that of the applied contact force. The peak contact pressures of 2.86 MPa and 3.66 MPa
147
are located in the superior region for normal walking and the posterior region for sitting
148
down, respectively. However, weight bearing area changes from superior-anterior to superior-
149
posterior for NW and from superior to posterior during SD, hence the contact and pore
150
pressure are distributed asymmetrically about the applied load. The timing of the occurrence
151
of peak contact and peak pore pressure coincided with the peak contact force. The peak
152
contact pressure (P-CPRESS), the related contact area (CAREA), the timing and the location
153
of their occurrence for NW and SD are summarised in Table 3. In addition, the peak pore
154
pressure (P-POR), which corresponded to the location of the peak contact pressure, and the
155
related compressive radial stress (S11) are also summarised in Table 3.
156
For normal walking, a plot of the results, including the stress components of the solid
157
matrix, is given in Figure 3. The locations of maximum P-CPRESS (2.87 MPa) and maximum
158
P-POR (2.85 MPa) are shown in Figure 3(a) & (b), where the distribution patterns are
159
identical. Stresses within the ECM in the three directions are highest on the articulating
160
surface. The highest stress concentrations, radial stress - 0.05 MPa and meridional stress 0.50
161
MPa, are found on the edges of the acetabular cartilage, where the pore pressure is minimal
162
(Figure 3(c) & (e)). Circumferential stress is the highest in the acetabular labral ring (0.33
163
MPa, Figure 3(e)). Shear stress, however, peaked at the cartilage-bone interface along the
164
edge of the acetabular cartilage, rising up to 1.35 MPa (Figure 3(f)). The ratio of fluid load-
165
support is 99.3% and 98.4% for NW and SD, respectively, indicating a similarity in fluid-
166
solid interaction in both cases.
Page 8
167
‘Cam’-type impinged articulation
168
The pathologic hip joint revealed localised impingement at the anterior-superior pelvic horn
169
during deep flexion. Hence, pore pressure increased significantly by 8 times from 0.42 MPa
170
to 3.76 MPa at the anterior-superior horn due to the intrusion of the femoral hump (Figure 4).
171
Compressive radial stresses, within the acetabular cartilage, increased at the impinged
172
location by 3.7 times from 0.07 MPa to 0.33 MPa. Tensile circumferential and meridional
173
stresses are at peaks within the acetabular labrum and at the cartilage-labral interface,
174
respectively (Figure 4(a) & (b)), and increased by a factor of 2.7 and 1.5, respectively. S22
175
and S33 rose from 0.84 MPa to 2.26 MPa and 0.92 MPa to 1.37 MPa (Figure 4(a) & (b)).
176
Fluid-load support in the impinged zone dropped to 91.9%, as opposed to 98.4 % in the
177
posterior region. Shear stress also doubled to 2.61 MPa on the acetabular surface and
178
surpassed the maximum at the acetabular cartilage-bone interface in the normal articulation.
179
Effect of material property variation
180
The influence of material property variation on some of the key parameters such as femoral
181
head displacement, contact pressure and pore pressure is summarized in Figure 5. The highest
182
model sensitivity was recorded for a reduction of 50% in shear modulus, with an increase of
183
46.18 % of femoral head displacement, although the contact pressure and the pore pressure
184
dropped only marginally by 2.89% and 3.68%. Reducing the tangential modulus by 50%
185
resulted in only marginal influence on femoral head displacement (-4.28%), contact (2.7%)
186
and pore pressure (1.97%), respectively. For all other variations the changes in the key
187
parameters are insignificant < 1%. This gives us confidence that the observed enhanced
188
responses in impinged case are indeed due to the abnormal articulation, and the influence of
189
model parameters are insignificant except the shear modulus.
Page 9
190
Discussion
191
The goal of this study was to explore how 'cam'-type impinged articulation can affect cartilage
192
stresses compared with those under normal articulation; and whether or not cartilage
193
structural integrity may be assessed based on a comparison of the total contact pressure, fluid-
194
load support (FLS) and generated stress components within the solid extracellular matrix
195
(ECM) in these two conditions. Special emphasis was placed on the consideration of
196
orthotropic properties of the cartilaginous tissue structures.
197
Comparisons with previous work seem to suggest that the locations identified for peak
198
contact pressure during normal walking and sitting down are consistent with those published
199
(Yoshida et al., 2006; Chegini et al., 2009). The maximum contact pressures in a normal joint
200
are found to be 2.87 MPa and 3.66 MPa during NW and SD, compared well with those (2.35
201
MPa and 3.34 MPa, respectively) reported by Chegini et al. (2009). Other subject-specific
202
models, however, predicted higher peak contact pressures, such as 6.6 MPa in Jorge et al.
203
(2014) and 10.78 MPa in Anderson et al. (2008). According to Anderson et al. (2010),
204
cartilage thickness variation accounted for more than doubled peak contact pressure compared
205
with a constant cartilage thickness. Harris et al. (2012) reported a peak contact pressure rage
206
of 7.51 ± 2.11 MPa for walking, and suggested that the peak contact pressure varies with
207
subjects. Hence, it seems reasonable that our results are within the range of reported data,
208
given the variation of the joint geometries.
209
Experimentally measured peak contact pressures seem to be higher than predicted by
210
the simplified FE models. According to Afroke et al. (1987), peak contact pressure varied
211
strongly with subjects (4.9 to 10.2 MPa). Anderson et al. (2008) reported a rather in-
212
homogeneous contact pressure distribution with a peak of 10 MPa for normal walking. In
213
contrast to that Brown and Shaw (1982) reported a peak contact pressure of 3.45 MPa, close
Page 10
214
to our findings. To the best of the authors’ knowledge, the stresses within the solid phase have
215
not been measured in a cadaveric hip joint.
216
Prior to this study, ECM of articular cartilage was assumed to deform linear elastic
217
isotropically (Ferguson et al., 2000a; 2000b; Macirowski et al. 1994) or hyper elastic
218
isotropically (Haemer et al., 2012), as the material properties of the cartilage were taken from
219
uni-axial compression tests (Mow et al., 1980; Macirowski et al., 1994; Wu and Herzog,
220
2000) where shear cannot be measured. To accommodate the multi-directional cartilage
221
deformation whilst using linear or hyper elastic isotropic properties, either lateral extension
222
was permitted (Macriowski et al., 1994) and applied load reduced (Ferguson et al., 2000a), or
223
the stress-bearing area increased, neglecting the lunate shape of the acetabular cartilage
224
(Haemer et al., 2012). None of these modelling techniques seem to be sufficient to answer the
225
current research questions.
226
Taking the different modelling techniques into account, the high FLS of 99.3% during
227
normal walking, compares reasonable well with 98.0% and 96.0% obtained by Haemer et al.
228
(2012) and Macirowski et al. (1994) for a ramp load of 1 second, but less favourably with
229
91.7% (following 1,000 static load) reported by Ferguson et al. (2000a).
230
In the pathologic case, alpha = 74°, the anterior protrusion produced an intrusion into
231
the anterior-superior acetabular cavity during sitting down, which is consistent with earlier
232
observations (Chegini et al. 2009; Jorge et al. 2014). Whilst the intrusion of the abnormal
233
protrusion did not yield significant elevated peak contact pressure, the FLS was affected and
234
the stress components of the ECM elevated. Although the FLS was still above 90%, the radial
235
compressive stress (or solid-on-solid) increased significantly (3.7 times) and the ECM was
236
not shielded from elevated solid-on-solid stress (Athesian and Mow, 2005). This may explain
237
some of the early macroscopic changes in the cartilage found in clinical observation (Ganz et
238
al. 2003, 2008; Beck et al. 2005), leading to severe osteoarthritis (Pritzker et al. 2006).
Page 11
239
Circumferential stress in the labrum was doubled, although this might not be as
240
significant due to the labral design to withstand high exposure in this direction (Ferguson et
241
al. 2001), The elevated stresses are mainly found in the cartilage-ECM, as opposed to those
242
reported by Chegini et al. (2009) in cartilage and labrum simultaneously. The increased
243
meridional stresses within the cartilage at the cartilage-labral interface seems to support the
244
hypothesis that cam impingement results in cartilage defects and subsequent separation of the
245
acetabular cartilage from the labrum (Ganz et al. 2003; 2008; Beck et al. 2005), and not
246
otherwise as discussed (McCarthy et al. 2001).
247
The emphasis of earlier numerical work (Chegini et al., 2009; Jorge et al., 2014) was
248
on severe abnormal femoral protrusions at high loads. Chegini et al. (2009) predicted,
249
dependent on the morphology of the femoral neck, peak contact pressures are in the range of
250
3.68 MPa to 12.84 MPa for 'cam'-type impingement. Jorge et al., (2014) also reported high
251
peak contact pressures up to 16.4 MPa. Due to the varied abnormal conditions presented, the
252
results cannot be readily compared with the current ones. Nonetheless, Ganz et al. (2003),
253
suggested that mild to unrecognisable, in the eye of an observer, abnormal femora are the
254
genesis of impingement. To the best of the authors knowledge, tissue stresses during 'cam'-
255
type impinged articulation have not been measured.
256
Admittedly, we only considered a typical case of impingement and limited loading
257
scenarios. The geometries considered are simplified versions based on CT images and
258
reported anatomic characteristics, although we believe they contain the essence of a typical
259
case of ‘cam’-type impingement. The finite element mesh was optimised to obtain
260
convergence during ‘cam’-type impinged articulation; the chosen grid size was the optimum
261
from a sensitivity study. Although the normal and the abnormal joints were meshed slight
262
differently, it should not affect the results. The grid-ratios on both acetabular and femoral side
263
were chosen carefully to obtain contact convergence. The slope of the contact-pressure
Page 12
264
overclosure relationship was also adjusted to minimise nodal penetration and contact pressure
265
errors. Further, loading scenarios considered here are limited to normal walking and sitting
266
down. In certain sporting activities, such as martial arts, a higher range of motions is required
267
hence higher stresses due to impingement may be expected.
268
As in all numerical work, experimental verification would be highly desirable, but it is
269
not yet possible to quantify the load sharing between the fluid and solid phase or measure the
270
strains within the ECM directly. Nonetheless, the key parameters were found to be generally
271
insensitive to the variation of material properties even by variation of ± 50% (Figure 5).
272
Because linear behaviour was considered for the fluid (permeability) and the solid phase
273
(elastic orthotropic), the error during a cam-impinged articulation should be similar to that
274
determined during static loading. The shear modulus, however, does seem to affect the
275
femoral head displacement, thus cartilage consolidation. Mansours (2009) argued that, during
276
cartilage compaction, elevated shear stresses will occur at the cartilage-bone interface, see
277
Figure 3 (f), as the cartilage layers are next to the much stiffer underlying subchondral bone
278
layer, and high lateral contraction is a result of compression of the opposing cartilage layers.
279
Hence, it is important to consider the orthotropic properties of the cartilage extracellular
280
matrix when simulating multi-directional deformation of cartilage.
281
Conclusions
282
Results from a biphasic analysis of a healthy and a cam-impinged joint show that 'cam'-type
283
impingement can elevate stresses significantly in the extracellular matrix of cartilaginous
284
tissues. During sitting down, the peak radial and the peak meridional stresses within the
285
cartilage and the circumferential stress within the acetabular labrum increased by 3.7, 2.7 and
286
1.5 times, respectively, compared with those in the normal joint. Peak pore pressure also
287
increased at the anterior-superior 'cam'-type impinged zone. The increased solid-on-solid
288
stresses and associated decrease in fluid-load support would result in higher friction during
Page 13
289
articulation and hence joint wear. In summary, 'cam'-type impingement is a risk factor that
290
could increase cartilaginous tissue strains leading to joint wear and other degenerative
291
changes in the hip, consistent with clinical findings.
292
Acknowledgement
293
The authors gratefully acknowledge access to the Sciama High Performance Computer (HPC)
294
Cluster, which was supported by the ICG, SEPNet and the University of Portsmouth and
295
support by Sciama technician Gary Burton.
Page 14
List of Figures
296
297
298
299
300
Anterior
D
Anterior Hump
C
Medial
Lateral
α
Posterior
301
302
Figure 1. Illustration of a pathologic femur (anterior hump): The deviation of femoral head
303
sphericity is measured by the α angle between the femoral neck axis, passing though the
304
femoral neck and head centre C, and a line extended from the point C to D, the point of
305
deviation from femoral head sphericity (edge of hump).
306
(1)
(1)
(2)
(3)
(4)
(5)
(a)
(b)
(c)
307
308
Figure 2. The solid models for a normal (a) and a pathologic femur with hump (b). The finite
309
element model of the normal hip joint (c) consists of acetabular labrum (1), acetabular
310
cartilage (2), femoral cartilage (3), an impermeable membrane (4) and femoral head neck
311
section (5), following (Chegini et al., 2009).
312
Page 15
contact pressure
(CPRESS)
(a)
radial stress
(S11)
(d)
A
max = 2.87 MPa
max = 2.84 MPa
pore pressure
(POR)
fluid velocity
(FLVEL)
(c)
(b)
A
A-A
circumferential stress
(S22)
A-A
meridional stress
(S11)
A-A
(e)
A-A
(f)
Tresca stress
(S, Tresca)
A-A
(g)
Figure 3. The peak contact pressure (MPa) (a) during normal walking; the related pore pressure (MPa) (b) and the stress components (MPa) in the
solid matrix: Radial (c), circumferential (d), meridional (e) and Tresca stress (f), section A-A as indicated in (a).
Page 16
(a) 'cam'-type impinged articulation
pore pressure
(POR)
radial stress
(S11)
circumferential stress
(S22)
meridional stress
(S33)
Tresca stress
(S, Tresca)
radial stress
(S11)
circumferential stress
(S22)
meridional stress
(S33)
Tresca stress
(S, Tresca)
(b) normal articulation
pore pressure
(POR)
Figure 4. Pore pressure for normal (a) and CAM-Impinged (b) acetabular labral and cartilage exposure during sitting down: Radial, circumferential,
meridional and Tresca stresses (MPa) within the ECM.
Page 17
(A)
Femoral Head Displacement
G
Ec
v12/13
v23
k
-0.11%
-0.09%
High
Low
-0.77%
-3.58%
-0.67%
-17.99%
2.5%
0.07%
0.19%
4.23%
0.86%
-2.5%
0.08%
0.0%
0%
46.18%
Deviation
25%
-25%
Et
5.0%
50%
-5.0%
-50%
Low
High
Low
High
(B)
Low
High
Low
High
Low
High
Contact Pressure
G
Ec
Et
v12/13
v23
k
0.00%
-0.02%
-0.19%
-1.79%
-0.13%
2.5%
-2.89%
High
Low
0.02%
0.14%
Low
0.11%
High
2.70%
-2.5%
0.33%
0.0%
1.95%
Deviation
5.0%
-5.0%
Low
High
Low
(C)
High
Low
High
Low
High
Pore Pressure
G
Ec
Et
v12/13
v23
k
-0.05%
-0.02%
High
Low
High
-0.01%
-1.64%
-0.75%
2.5%
-3.68%
Low
High
0.01%
0.03%
Low
0.14%
High
1.90%
-2.5%
0.97%
0.0%
2.18%
Deviation
5.0%
-5.0%
Low
High
Low
Low
High
Figure 5. The variation in the femoral head displacement, contact pressure and pore pressure
due to the variation in the material parameters (Low: - 50%; High: + 50 %).
Page 18
List of Tables
Table 1. The numbers of elements and grid sizes used in the FE models of the normal and the
abnormal hip joint.
Element Type
Healthy Joint
Abnormal Joint
Number
Grid Size
Number
Grid Size
Acetabular cartilage and labrum
C3D8RP
21,460
0.70
21,460
0.70
Femoral cartilage
C3D8RP
13,666
0.90
12,420
1.00
R3D4
9,956
0.90
8,767
1.00
M3D4R
4,004
1.15
4,166
0.95
Femoral head-neck section
Impermeable membrane
Table 2. Material properties used in this study.
Biphasic Formulation
E1
E2
E3
ν 12
ν 13
ν 23
G
k’
[MPa]
[MPa]
[MPa]
-
-
-
[MPa]
[mm/s]
Tissue
Acetabular
Labrum
Φ
-
b
0.157 a
26 a
3e
0.04
0.04
0.1
4.89
x 10-09
1.5
4
d
Acetabular
Cartilage
1.23 b
8.5 c
8.5 c
0.044 b
0.044 b
0.146 d
2.5 f
8.52
x 10-09
4
Femoral
Cartilage
1.18 b
8.5 c
8.5c
0.046 b
0.046 b
0.146 d
2.5 f
8.98 d
x 10-09
4
a
(Ferguson et al. 2001)
b
(Athanasiou et al. 1994)
f
(Tisskaht & Ahmed 1995)
g
(Wong et al. 2008)
c
d
(Roth et al. 1980)
(Frederico et al. 2003)
Table 3. Peak biphasic response for normal walking and siting down for normal articulation.
Activity
Location
Biphasic Response
-
-
P-CPRES
CAREA
P-POR
S11
Time of Cycle
-
-
[MPa]
[mm2]
[MPa]
[MPa]
[%]
Normal Walking
Superior
2.87
1798
2.84
-0.03
54
Sitting Down
Posterior - Medial
3.58
1574
3.54
-0.04
46.23
Page 19
Download