6. Beam Restricting Devices

advertisement
Beam And Scatter Radiation
Restricting Devices


DMI 50B
Kyle Thornton
Scatter Radiation

Compton interaction



Higher kVp energies in diagnostic range
As scatter increases, radiographs lose
contrast
Radiographs will look dull, blurred, and
fogged
Factors Affecting Scatter
Radiation



kVp
Field size
Patient or part thickness
kVp

Compton interaction increases with
increasing kVp



Scatter radiation increases
Scatter radiation is reduced by using
collimators and grids
Decreasing kVp is not always desirable

Increases patient dose
Beam Field Size



As field size is increased, scatter radiation
increases
Restriction of field size improves image
quality
Even more important during fluoroscopy
Patient Or Part Thickness



Scatter increases with patient or part
thickness
A chest x-ray will result in more scatter than
an arm x-ray
Compression devices reduce patient
thickness and improve quality
Compression Used to
Decrease Effects of Scatter
Radiation
Beam-Restricting Devices

Two types of devices to reduce scatter



Restricted for two reasons



Beam restricting devices
Grids
Reduce patient dose
Improve image contrast
Three types of beam-restricting devices




Aperture diaphragm
Extension cones and cylinders
Variable aperture collimator
Positive beam limitation
Different Types of Beam
Limitation Devices
Aperture Diaphragm




Simplest of all beam-restricting devices
Lead or lead-lined plate with an opening
Used with older head units
Cannot see a light
Extension Cones And
Cylinders




Modifications of aperture diaphragms
Primarily used for headwork, L5-S1
Fixed openings
Generally not light localizing
Cones and Cylinders
Comparison of Sinuses W/WOut Use of Cones
Variable-Aperture Collimator




Light localizing
Most commonly used
A series of shutters and mirrors used
Light field must coincide with x-ray beam
Variable Aperture Device
Schematic Drawing of the
Variable Aperture Device
Positive Beam Limitation
Devices


Automatically collimates to the size of film
used
Does not relieve technologist of having to
collimate to anatomy of interest
Download