O47 A Pilot Study to Investigate the effect of Routine Warming On the Level of a Glucose Degradation Product (GDP) in Conventional Peritoneal Dialysis (PD) Fluids in a Clinical Environment. Judge, C.G Kent Kidney Care Centre, East Kent Hospitals NHS University Foundation Trust INTRODUCTION: Conventional, lactate- based PD fluids are commonly used in the UK today. Heat sterilisation during manufacture causes GDPs to rise to toxic levels. A maturation period of several weeks is required for them drop to acceptable levels for clinical use. The most toxic GDP that has been detected is 3,4-di-deoxyglucosone-3ene (3,4 DGE), which is associated with delayed healing, reduced cell proliferation and cell death, even at low concentrations. While it is known that 3,4 DGE levels rise when warmed during laboratory experiments, across a range of temperatures and glucose concentrations, it is not known if this effect occurs in the routine clinical environment of the PD department. This simple pilot study compares levels of 3,4 DGE over time, in batchidentical pairs of PD bags that are either warmed to 36.5 degrees, or left at ambient (room) temperature. METHODS: Six unopened boxes of PD fluids bags from 6 different batches and sterilization dates were selected, two for each glucose concentration (conc); low (1.36%), medium (2.27%) and high (3.86%). A pair of bags were taken from each box, and labeled either ‘warmed’ or ‘room temperature’. The different batches were recorded A-F, 12 bags in total were prepared. At zero hours (before warming commenced) fluid samples were taken from each bag using aseptic technique. Then for each pair of bags, one was placed in the department bag warmer (set at 36.5 degrees Celsius) and the other on a trolley. The Ultraviolet absorbance at 228 and 284nm was recorded using the spectrophotometer at the Trust’s pathology department. The concentration of 3,4 DGE in micromols/l for each sample was calculated using an accepted formula from the literature. Sampling was repeated after 24 hours of warming and then weekly for 8 weeks. The results were plotted on a graph and analyzed using Microsoft Excel. RESULTS: At zero hours, the batch pairs showed comparable levels of 3,4 DGE, but there were differences between batches at each glucose concentration. After 8 days of warming, the levels of 3,4 DGE in all the warmed bags had increased by a minimum of 25% and maximum of 31%. The controls left at room temperature rose by between 2.86% to 5.33%. It was notable that the 3,4 DGE level of the warmed bags rose rapidly in the first 8 days, and thereafter plateaued, although warming continued at 36.5 degrees Celsius for several weeks (3 weeks shown here). Conc/Temp0C/ Batch 1.36%/ 36.5 / A 1.36%/ ambient/ A 1.36%/ 36.5 / B 1.36%/ ambient/ B 2.27%/ 36.5 / C 2.27%/ ambient/ C 2.27%/ 36.5 / D 2.27%/ ambient/ D 3.86%/ 36.5 / E 3.86%/ ambient/ E 3.86%/ 36.5 / F 3.86%/ ambient/ F 3,4 DGE Day 0 37 36 40 39 69 70 59 59 101 105 93 97 3,4 DGE day 1 39 41 42 44 74 75 62 63 105 108 97 99 3,4 DGE day 8 47 38 51 41 88 72 77 62 127 110 119 100 3,4 DGE day 15 48 38 52 42 89 73 78 62 127 108 120 102 3,4 DGE day 22 48 37 53 42 89 72 78 61 130 107 121 101 CONCLUSION: This small pilot study, part of a masters project, demonstrates that routine warming in a clinical area can cause rapid elevation of 3,4 DGE in 8 days or less. While the statistical significance of this finding cannot be demonstrated with such small numbers, it is notable the levels of 3,4 DGE found here are higher than those found by others- even before warming commenced. All 6 warmed bags showed the same response. The levels of 3,4 DGE found have been associated in the literature with toxic effects as noted above. RELEVANCE: PD fluid warming practices vary, but informal feedback suggests bags are often heated for extended periods. In an unpublished survey of 35 UK units (part of the same masters project) respondents from 12 units felt it was acceptable to warm bags for timescales ranging from 7-14 days, to > 6 weeks, and 8 units didn’t know how long their bags were warmed. PD fluid is classed as a borderline product, governed by medicines legislation. It should be stored as directed, at or below 250C. PD staff should be made aware that simple adjustment of bag warming practice could reduce patient exposure to 3,4 DGE. GDPs change in response to the temperature of their environment; this has implications for storage areas in hospitals, which may lack ventilation, and of course countries with high seasonal or year-round temperatures.