Biomedical Imaging I

advertisement

Ultrasound

Sound waves

Sounds are mechanical disturbances that propagate through the medium

Frequencies <15Hz Infrasound

15Hz<Frequencies <20KHz Audible sound

Frequencies>20Khz Ultrasound

Medical Ultrasound frequency 2 -20MHz

Some experimental devices at 50MHz

Velocity and frequency

For sound waves the relationship between frequency/velocity and wavelength is c = f x

Speed of sound depends on the material sound travels

Velocity is inversely proportional to compressibility the less compressible a material is the greater the velocity

Average velocity in tissue 1540 m/sec (air 331m/sec, fat 1450 m/sec)

The difference in speed of sound at the boundaries determines the contrast in US



Wave Speed

c air

= 331 m/s c salt water

= 1500 m/s c

B

B = Bulk Modulus

= density

Bulk modulus measures stiffness of a medium and its resistance to being compressed

Speed of sound increases with stiffness of material k = adiabatic bulk modulus

= density

Wave speed cnt

Changes in speed DO NOT affect the frequency so only the wavelength is dependent on the material.

What is the wavelength of a 2MHz beam traveling into tissue?

What is the wavelength of a 5MHz beam traveling into tissue?

Wave speed cnt

Changes in speed DO NOT affect the frequency so only the wavelength is dependent on the material.

What is the wavelength of a 2MHz beam traveling into tissue? 0.77mm

What is the wavelength of a 10MHz beam traveling into tissue? 0.15mm

The wavelength determines the image resolution

Higher frequency -> higher resolution

Penetration is higher at smaller frequencies.

Penetration and resolution

Thick body parts (abdomen)

Low frequency ultrasound (3.5 - 5 Mhz)

Small body parts (thyroid, breat)

High frequency (7.5 - 10 Mhz)

Interference

Waves can constructively and destructively interfere

Constructive interference -> Increase in amplitude (waves in phase)

Destructive interference -> Null amplitude (waves out of phase)

Acoustic Impedance

Z=

 x c [kg/m 2 /sec] SI unit ([Rayl] =1 [kg/m 2 /sec])

Independent of frequency

Air -> Low Z

Bone -> High Z

Large difference in acoustic impedence in the body generate large reflections that translate in large US signals

Example going from soft tissue to air filled lunghs ->BIG REFLECTION

Sound and pressure

Sound waves cause a change in local pressure in the media

Pressure (Pascal)=N/m 2

Atmospheric pressure 100KPa

US will deliver 1 Mpa

Intensity I (amount of energy per unit time and area) is proportional to P 2

This is the energy associated with the sound beam

Temporal and Spatial intensity when dealing with time or space

Sound and pressure

Relative sound intensity (dB) (Bels => B, 1B=10dB)

Relative intensity dB= 10 log( I/Io ) Io original intensity, and I measured intensity

Negative dB -> signal attenuation

-3dB -> signal attenuated of 50%

Attenuation

Loss by scatter or absorption

High frequency are attenuated more than low frequencies

Attenuation in homegeneous tissue is exponential

A 1Mhz attenuation in soft tissue is 1 dB/cm, 5 MHz -> 5dB/cm

Bone media attenuation increases as frequency squared.

Absorbed sound ->heat

Reflection

Echo -> reflection of the sound beam

The percentage of US reflected depends on angle of incidence and Z

Similar to light

R





Z

Z

2

2

Z

1

Z

1





2

 T

4

Z

1

Z

2

Z

1

Z

2

 2



Reflection Snell’s Law

 i

 t angle of incidence angle of transmittance

 sin sin

 

 i

 

 v

1 v

2

Transducer

Made of piezoelectric material

Crystals or ceramics

Stretching and compressing it generate V

Lead-zirconate-titanate (PZT)

A high frequency voltage applied to PZT generate high freq pressure waves

Are generators and detectors

Q factor

Q factor is the frequency response of the piezoelectric crystal

Determines purity of sound and for how long it will persist

High Q transducers generate pure frequency spectrum (1 frequency)

Q=operating frequency/BW

BW bandwidth

High Q -> narrow BW

Low Q->broad BW

Transducer backing

Backing of transducer with impedance-matched, absorbing material reduces reflections from back

 damping of resonance

Reduces efficiency

Increases Bandwidth (lowers Q)

Axial beam profile

Piston source: Oscillations of axial pressure in near-field

(e.g. z

0

= (1 mm) 2 /0.3mm = 3 mm)



Near Field

Fresnel Zone

Far Field

Fraunhofer zone

Caused by superposition of point wave sources across transducer (Huygens’ principle)

Side lobes = small beams of reduced intensity at an angle to the main beam sin(

)

1.22

/(2 r ) z

0

 r

2

4

US usually uses

Fresnel Zone

Lateral beam profile

Determined by Fraunhofer diffraction in the far field.

Given by Fourier Transform of the aperture function

Lateral resolution is defined by width of first lobe (angle of fist zero) in diffraction pattern

For slit (width a ):

I

I

0

sinc 

Minima at: sin a

 sin

  n

 a

For disc (radius r , piston source): sin

 

0.61

 r

  

 arcsin 0.61

 r

Focused transducers

Reduce beam width

Concentrate beam intensity, increasing penetration and image quality

All diagnostic transducers are focused

Focal zone – Region where beam is focused

Focal length – distance from the transducer and center focal zone

Focusing of ultrasound

Increased spatial resolution at specific depth

Self-focusing radiator or acoustic lens

Array types

a) Linear Sequential

(switched) ~1 cm

10-15 cm, up to 512 elements b) Curvilinear similar to (a), wider field of view c) Linear Phased up to 128 elements, small footprint

 cardiac imaging d) 1.5D Array

3-9 elements in elevation allow for focusing e) 2D Phased

Focusing, steering in both dimensions

Array resolution

Lateral resolution determined by width of main (w) lobe according to sin

 

 w

Larger array dimension

 increased resolution

Side lobes (“grating lobes”) reduce resolution and appear at sin

  g n

 g n

1, 2,3,...

a g w

Ultrasound Imaging

Imaging

Most ultrasound beam are brief pulses of 1 microsecond

Wait time for returning echo

Object must be large compared to wavelength

Signal is amplified when returned (echo is small signal)

A-mode (amplitude mode) I

Oldest, simplest type

Display of the envelope of pulse-echoes vs. time, depth d = ct /2

Pulse repetition rate ~ kHz

(limited by penetration depth, c

1.5 mm/

 s

20 cm

270

 s, plus additional wait time for reverberation and echoes)

A-mode (amplitude mode)

Or space! Also M mode!

depth

A-mode II

Frequencies: 2-5 MHz for abdominal, cardiac, brain; 5-15 MHz for ophthalmology, pediatrics, peripheral blood vessels

Applications: ophthalmology (eye length, tumors), localization of brain midline, liver cirrhosis, myocardium infarction

Logarithmic compression of echo amplitude

(dynamic range of 70-80 dB)

Logarithmic compression of signals

M mode or T-M mode

Time on horizontal axis and depth on vertical axis

Time dependent motion

Used to study rapid movement – cardiac valve motion

B-mode clinical example

Static image of section of tissue

Brighter means intensity of echo

Bmode (“brightness mode”)

Lateral scan across tissue surface

Grayscale representation of echo amplitude

Add sense of direction to information-> where did echo come from

Real-time B scanners

Frame rate R f

~30 Hz: t acq

2 c d

N R f

 t

1  c

2 d N d : depth

N : no. of lines

Mechanical scan: Rocking or rotating transducer

+ no side lobes

- mechanical action, motion artifacts

Linear switched array

Linear switched

CW Doppler

Doppler shift in detected frequency f shift

 c

 f v : blood flow velocity c: speed of sound

: angle between direction of blood flow and US beam

Separate transmitter and receiver

Bandpass- filtering of Doppler signal:

Clutter (Doppler signal from slow-moving tissue, mainly vessel walls) @ f<1 kHz

LF (1/f) noise

Blood flow signal @f < 15 kHz

CW Doppler bears no depth information

Frequency

Counter

Spectrum

Analyzer

CW Doppler clinical images

CW ultrasonic flowmeter measurement (radial artery) v

[10cm/s] t

[0.2 s]

Spectrasonogram:

Time-variation of Doppler Spectrum f t

CW Doppler example

Duplex Imaging

Combines real-time B-scan with US Doppler flowmetry

B-Scan: linear or sector

Doppler: C.W. or pulsed ( f c

= 2-5 MHz)

Duplex Mode:

Interlaced B-scan and color encoded Doppler images

 limits acquisition rate to 2 kHz (freezing of B-scan image possible)

Variation of depth window (delay) allows 2D mapping (4-18 pulses per volume)

Duplex imaging example (c.w.)

www.medical.philips.com

Duplex imaging (Pulsed Doppler)

US imaging example (4D)

Download