Fitting and Evaluation of FM Systems for HA Users Guidelines for Fitting and Evaluation ANSI S3.22-Specification of Hearing Aid Characteristics (2003 approved) ANSI WG 81-Electroacoustic Evaluation of Hearing Assistance Devices (Approval Pending) AAA Task Force-Developing Guidelines for Fitting and Evaluation of FM Systems Assumptions in Fitting FM Systems Properly Functioning Hearing Aid Properly Fit Hearing Aid Matched to the needs of the child or adult with hearing loss Properly Configured to Receive the FM Signal Meets Specifications-ANSI S3.22 Contacts removed on Direct Audio Input Connections Binaural Ear Level System is Optimal Arrangement Microphones at Ear Level 1. The auditory signal should be as consistent as possible This facilitates learning Ideally this is achieved with minimal changes of equipment 2. There are different signal input levels to the hearing aid mic and to the FM mic in typical use situations Typical input to HA mic - 65 dB SPL Typical input to FM mic - 80 dB SPL 3. Think in terms of output rather than gain Goal is to match SPL levels at the eardrum so that signal is consistent (exceptions will be discussed) Therefore, for same output, gains will differ because input levels are different EXAMPLE using overall signal levels Say that Mary hears best when level at the eardrum is 100 dB SPL For the HA Fitting 65 dB SPL input + 35 dB gain = 100 dB SPL For the FM Fitting 80 dB SPL input + 20 dB gain = 100 dB SPL 4. Use speech-weighted noise for frequency response and use tone sweep for SSPL90 All of the energy in complex signal sums to an overall level, there is not 70 dB at every frequency, but 70 dB overall, therefore….. The SPL at 1 kHz will be lower for 70 dB speech weighted noise than for 70 dB sweep frequency tones. But…. For a pure tone sweep, there IS 70 dB at every frequency 5. Use FM settings first, then HA, to match output for HA and HA+FM conditions Ideally want the child to put on HA and not change settings when using FM Need to check electroacoustically to determine if this is possible 6. Optimize the testing Try to do electroacoustic measures before patient arrives Perform real ear measures if possible When you have options to receive FM only & FM+ENV, start first with the signal arrangement used most often EVALUATION OPTIONS for FM SYSTEMS A. Audio-Visual Checks B. Electroacoustic Measurements Verifying Performance Matching FM’s to HA’s C. Real Ear Evaluation Matching FM’s to HA’s D. Behavioral Evaluation Speech Recognition NEW TERMINOLOGY While in a sea of strips from the test box… it occurred to me that we needed new terms Start Simple….First Letter E for Electroacoustic Measures R for Real ear Measures Then… HA for Hearing aid FM for FM system in FM mode There are four basic evaluation steps! 1) HA - 65 dB SPL complex signal 2) HA & FM - 80 dB SPL complex signal 3) HA – 90 dB pure tone signal 4) HA & FM – 90 dB pure tone signal EHA65 EFM/HA80 EHA90 EFM/HA90 SIMPLIST CASE MILD TO MODERATE HEARING LOSS 100 dB SPL Amplified Speech Gain is 40 dB for HA Unamplified HA Mic Auditory Area Gain is 25 dB for FM Speech Input to FM Mic ENV signal 10 dB below FM and still Audible 0 ENV MIC FM MIC FM+ENV MIC MORE DIFFICULT CASE SEVERE TO PROFOUND HEARING LOSS dB SPL 100 Amplified Speech Gain is 60 dB for HA due to Feedback Unamplified Speech Input to HA Mic 0 ENV MIC Auditory Area MORE DIFFICULT CASE SEVERE TO PROFOUND HEARING LOSS dB SPL 100 Amplified Speech Auditory Area Gain is 60 dB for HA due to Feedback Gain is 45 dB for FM Unamplified Speech Input to HA Mic FM Mic ENV signal 10 dB below FM and NOT Audible 0 ENV MIC FM MIC FM+ENV MIC MORE DIFFICULT CASE SEVERE TO PROFOUND HEARING LOSS Charlie So in this case, we want to amplify the FM signal to a HIGHER intensity than that achieved with the Hearing Aid Electroacoustic Measurements MORE DIFFICULT CASE SEVERE TO PROFOUND HEARING LOSS 100 Amplified Speech dB SPL Gain is 60 dB for HA due to Feedback Auditory Area Gain is 60 dB for FM Unamplified Speech Input to HA Mic FM Mic ENV signal 10 dB below FM and NOW IS Audible 0 ENV MIC FM MIC FM+ENV MIC MORE DIFFICULT CASE SEVERE TO PROFOUND HEARING LOSS 100 Amplified Speech dB SPL Gain is 60 dB for HA due to Feedback Auditory Area Gain is 60 dB for FM Unamplified Speech Input to HA Mic FM Mic ENV signal 10 dB below FM and NOW IS Audible 0 ENV MIC FM MIC FM+ENV MIC Let’s Consider …. Steps to Verify HA and FM Outputs Preliminary Setup 1. Set HA to User Settings 2. Set Test Equip for Complex Signal 3. Set Test Equip to display OUTPUT, not gain 4. Go into Multicurve Function Electroacoustic Measurements Set up for testing HA alone Steps to Verify HA and FM Outputs Billie Change setting to Curve 1-Put HA in test box Curve 1. Test HA alone with 65 dB SPL complex input Electroacoustic Measurements Set up for testing HA & FM Steps to Verify HA and FM Outputs Billie Change setting to Curve 2-Put FM Transmitter in test box, Attach FM Receiver Boot to HA Curve 2. Test HA&FM system (set to FM only) with 80 dB SPL complex input Electroacoustic Measurements Steps to Verify HA and FM Outputs Billie Compare Curves 1 and 2 : EHA65 and EFM/HA80 Adjust volume if necessary to match HA alone output, If a programmable hearing aid , make and new program and refer to that as “FM Program” Electroacoustic Measurements Steps to Verify HA and FM Outputs Checking Maximum Output- Billie Change setting to Curve 3-Put HA in Test Box Curve 3. Test HA alone with 90 dB SPL pure-tone sweep input Electroacoustic Measurements Steps to Verify HA and FM Outputs Change setting to Curve 4-Put FM Transmitter in test box, HA with FM Receiver Boot and 2cc coupler outside Curve 4. Test HA&FM system (set to FM only) with 90 dB SPL pure-tone sweep input Match Max Output- Billie Compare Curves 3 and 4, EHA90 and EFM/HA90 Adjust max output if necessary in the “FM Program”, to match HA output Electroacoustic Measurements SUMMARY These curves are closely aligned except for the high frequencies at the typical Input levels (1&2). Additional high frequency info. Is often desirable. Goals for Verfication FM Advantage of about 10 dB EFM/HA80 minus EHA65 FM Transparency of about 0 dB EFM/HA90 minus EHA90 Real Ear Measurements Same principles re: matching output apply Main difference is to place FM mic as close to reference mic as possible Curves: RHA65, RFMHA80, RHA90, RFMHA90 Real Ear Measurements Set up for Leveling Set up for testing HA alone Set up for testing HA & FM EVALUATION OPTIONS A. Audio-Visual Checks B. Electroacoustic Measurements Verifying Performance Matching FM’s to HA’s C. Real Ear Evaluation Matching FM’s to HA’s D. Behavioral Evaluation Speech Recognition Behavioral Measurements There are two main reasons why it is not advisable to do threshold testing with the FM system. The first one stems from the suggestion to place the FM mic 6 inches from the loudspeaker and place the child outside the booth….. Behavioral Measurements Behavioral Measurements 1) There is too much variability in vertical and horizontal dimensions which may result in variability in the signal levels arriving at the FM mic. Behavioral Measurements Behavioral Measurements 2) In addition, the FM systems may operate differently for low level signals than for higher level ones like speech because of compression. THEREFORE….. Threshold testing with the FM system is not recommended! Rather………. Behavioral Measurements Speech Recognition Testing Hearing Aid alone FM System FM Only FM+ENV Mode Test each arrangement in Quiet and in Background noise if time permits Behavioral Measurements Speech Recognition Testing Select appropriate test with multiple lists Set FM and HA controls to user settings Place FM mic on examiner but turned off Place HA and FM receiver on Patient Seat Patient in soundbooth at 45 degrees azimuth Behavioral Measurements INTERPRETATION Relate Laboratory findings to Clinical Applications How do you determine if a score is significantly low? Or significantly different from another score? Thornton and Raffin (1978) determined the degree to which a second speech recognition score must vary from an initial score to be significantly different Critical differences for speech recognition scores (Thornton and Raffin,1978) Initial 90 80 70 60 50 40 30 20 100 50 25 10 81-9676-9872-100 50-100 68-8964-9256-9640-100 57-8152-8648-9230-90 47-7342-7836-8420-90 37-6332-6828-7610-90 27-5322-5816-6410-80 19-4314-4812-5610-70 11-328-36 4-44 0-60 Disproportionately Low Scores? Data for these comparisons were based on the 95% confidence limits of the PB Max for NU6 word lists from a sample of 407 ears with a wide range of pure-tone averages (500, 1000 and 2000 Hz) (Dubno et al. 1995). SPeech Recognition INTerpretation SPRINT Chart May be used: 1) to compare two speech recognition scores or 2) to compare PBmax score on NU6 lists to normative data SPRINT Chart for 25-Word NU6 Lists 0 4 8 12 16 20 24 28 32 36 40 44 48 52 56 60 64 68 72 76 80 84 88 92 96 P T A d B H L 0 4 8 12 16 20 24 28 32 36 40 44 48 52 56 60 64 68 72 76 80 84 88 92 96 0 4 8 12 16 20 24 28 32 36 40 44 48 52 56 60 64 68 72 76 80 84 88 92 96 0 4 8 12 16 20 24 28 32 36 40 44 48 52 56 60 64 68 72 76 80 84 88 92 96 First Percent Correct Score S e c o n d P e r c e n t S c o r e SPRINT Chart for 50-Word NU6 Lists 0 10 20 30 40 50 60 70 80 90 0 0 10 10 20 20 30 30 e S c o n d P 40 T A 40 d B 50 H L P e r c 50 e n t 60 60 c S o r e 70 70 80 80 90 90 0 10 20 30 40 50 60 First Percent Correct Score 70 80 90 Speech recognition scores are often interpreted based on some personal internal norms because validated normative information is not conveniently available. Speech Recognition Testing ASHA Guidelines Test Conditions HA-Quiet (Speech at 55 dB HL) HA-Noise (Add Speech Noise at 50 dBHL) HA&FM-Noise (Keep levels, Turn on FM Mic) HA&FM-Quiet (Keep levels, Turn off noise) Behavioral Measurements Speech Recognition Testing Sample Scores HA-Quiet HA-Noise HA+FM-Noise If score decreases significantly, then FM gain too low HA+FM-Quiet 80% 50% 85% 80% If score increases significantly, then FM gain too low Behavioral Measurements You thought EHA65, EFM/HA90, EHAFM65, EFMHA80 were too Much! Hearing Aid Alone BHA55 BHA55/50 Hearing Aid + FM System BFM/HA55/50 BFM/HA55 Behavioral FM Verification Case Illustration Third Grader who resisted wearing FM system. Testing done to show him benefits. Hearing Aid Alone BHA55 BHA55/50 90% 60% Hearing Aid with FM BFM/HA55/50 (unlit FM) BFM/HA55/50 (bilat FM) BFM/HA55 80% 90% 100% Getting Desired Results If the desired FM Advantage and FM Transparency are not obtained, adjustments may be necessary to Hearing Aid/FM settings via Manual adjustments-screwdrivers and trimpots Programmable adjustments-software, Noah, TOASTER? WATCH WHAT YOU SAY!!! SUMMARY Goal of Fitting FM is to allow improved signalto-noise ratio so that signal from the FM transmitter is higher intensity than background noise Electroacoustic Verification is necessary to determine optimal signal to noise ratio Behavioral Verification with FM/HA combination useful for Counseling and Motivating Student