GIGO Measurement of Lead Depends on: Chemistry of Lead for Separations Chemistry of Lead for Creating a signal Chemistry of Lead for Creating Background Chemistry of Lead for the Stability of the Signal Garbage In = Garbage Out 15000 10000 Amplitude 5000 0 0 0.2 0.4 0.6 0.8 -5000 -10000 -15000 Time (s) Sample Sample Prep Instrument Instrument Out put Signal (Data) 1 1.2 Recall we mentioned that lead resides at surface of soils because of it’s insolubility Beta Recall from Quant Pb 2 OH PbOH Kf1 PbOH OH Pb OH 2 o Kf 2 Pb OH 2 OH Pb OH 3 o Kf 3 Pb OH 3 OH Pb OH 4 2 Kf 4 Mass balance Pbtotal Pb 2 PbOH Pb OH 2 Pb OH 3 Pb OH 4 o 2 PbOH K f 1 Pb 2 OH Pb OH K PbOH OH o 2 f2 Pb OH 2 K f 1 K f 2 Pb 2 OH o 2 Pb OH 3 K f 3 Pb OH 2 OH o Pb OH 3 K f 1 K f 2 K f 3 Pb 2 OH 3 Pb OH 4 K f 4 Pb OH 3 OH 2 Pb OH 4 K f 1 K f 2 K f 3 K f 4 Pb 2 OH 2 4 Mass balance Pbtotal Pb 2 PbOH Pb OH 2 Pb OH 3 Pb OH 4 o 2 PbOH K f 1 Pb 2 OH Pb OH K o 2 2 f 1 K f 2 Pb OH 2 Pb OH 3 K f 1 K f 2 K f 3 Pb 2 OH 3 Pb OH 4 K f 1 K f 2 K f 3 K f 4 Pb 2 OH 2 Pbtotal Pb 2 K f 1 Pb 2 OH K f 1 K f 2 Pb 2 OH 2 K f 1 K f 2 K f 3 Pb 2 OH K f 1 K f 2 K f 3 K f 4 Pb 2 OH 3 4 Make a definition to simplify the expression, and factor out terms 1 K f 1 2 K f 1 K f 2 3 K f 1 K f 2 K f 3 4 K f 1 K f 2 K f 3 K f 4 Pbtotal Pb 2 1 1 OH 2 OH 3 OH 4 OH 2 3 4 4 Pbtotal Pb 2 1 OH OH 1 D 1 1 OH 0 0 Pb 2 Pbtotal 2 OH 2 2 3 OH Pb 2 Pb 1 OH OH 2 3 2 1 2 3 OH 3 2 PbOH 1 Pbtotal 4 3 Pb 1 OH 1 OH OH 1 2 2 3 2 3 3 3 4 4 OH 4 2 OH 4 1 D Pb OH OH OH 1 OH 1 1 3 4 2 2 4 OH 4 OH 1 OH OH OH OH 2 3 1 1 1 3 OH 2 4 OH 4 4 OH 1 OH D 4 0 1 1 OH OH OH OH 2 1 1 2 3 4 3 1 OH OH 4 4 OH 1 OH 1 2 3 2 PbOH 2 o PbTotal PbOH 3 PbTotal PbOH 4 1 OH 1 2 PbTotal 2 4 OH OH 2 OH 2 OH 1 1 OH 2 2 OH 1 OH D 2 3 OH 2 2 1 D 3 4 OH 3 4 OH 4 3 4 OH 1 1 OH 4 3 OH 2 3 OH 3 3 4 3 4 OH 4 D 3 OH 4 3 OH 2 OH 3 D 4 OH D 4 2 1. Calculate beta values This number should be 6.4 2. Set up a column for the pH value 14 pH pOH pOH 14 pH 3. Calculate [OH-] 10 14 pH ) =10^(-(14-A3)) 4. Calculate D D 1 1 OH OH 2 2 3 OH 3 4 OH 4 =1 + $J$5*B3+$K$5*(B3^2)+$L$5*(B3^3)+$M$5*(B3^4) 5. Calculate alpha 0 1 =1/C3 0 D 6. Calculate alpha 1 7. Calculate alpha 2 1 1 OH 2 =$J$5*B3*D3 D 2 OH 2 =$K$5*B3^2*D3 D 8. Calculate alpha 3 and 4 in a similar fashion Note that at pH < 6 all of the lead is present as Pb2+ 1.2 Pb2+ Pb(OH)42- 1 Pb(OH)+ Pb(OH)2 Fraction 0.8 0.6 0.4 Pb(OH)3- 0.2 0 0 2 4 6 8 10 12 pH This graph indicates that if our instrument is measuring Pb2+ then when we Prepare the sample we need to have a pH of less than 6 14 GIGO Measurement of Lead Depends on: Chemistry of Lead for Separations Chemistry of Lead for Creating a signal Chemistry of Lead for Creating Background Chemistry of Lead for the Stability of the Signal Garbage In = Garbage Out Measurements based on PbS Measurements based on PbS 1820 Frederick Acum London “1 part of acetate of lead May be detected by means of it in 20000 Parts of water.” 1820 Sulphuretted water cupellation ppm ppb ppt 8000 B.C.E. 1000 C.E. 1900 1990 2000 More PbS measurements -log Ksp Ag2S=49 q C V q V C Internal solution fixed in Ag+ Suppose we are using a lead ion selective electrode to measure Pb2+, can We use any pH less than 6? Soln Pb -log Ksp PbS=29 Pb2+ Pb2+ S2Ag+ Pb2+ S2S2controls S2S2S2- Pb2+ Charge separation after motion of Ag+ leads to a potential across the Membrane = signal Ag+ Pb2+ S2Which controls -log Ksp Ag2S=49 Which controls Ag+ GIGO Measurement of Lead Depends on: Chemistry of Lead for Separations Chemistry of Lead for Creating a signal Chemistry of Lead for Creating Background Chemistry of Lead for the Stability of the Signal Garbage In = Garbage Out Other Alpha Plots are also useful 1.2 Lead Alpha Fractions 1 0.8 0.6 0.4 0.2 0 -5 -3 -1 1 3 5 pCl If we want to separate lead on an anion exchange column form the PbCl3- species. Which line would that be? And what conc. Cl would we want? GIGO Measurement of Lead Depends on: Chemistry of Lead for Separations Chemistry of Lead for Creating a signal Chemistry of Lead for Creating Background Chemistry of Lead for the Stability of the Signal Garbage In = Garbage Out Lead Chloride, while useful for an anion exchange separation is a problem Because of it’s low vapor pressure Water is shown for comparison. What this means is if you get about 700 oC You will have a large vapor pressure for PbCl2 which means you lose Stuff from solution GIGO Measurement of Lead Depends on: Chemistry of Lead for Separations Chemistry of preparing the sample Chemistry of Lead for Creating a signal Chemistry of Lead for Creating Background Chemistry of Lead for the Stability of the Signal Garbage In = Garbage Out only instrument you have is a….. UV-Vis Spectrophotometer UV-Vis monitors valence shell electrons Need to convert Pb to something that a. Has UV-Vis activity b. That can be selective toward Pb binding c. That can be separated from other binding metals Not water soluble Loss of a proton makes this a good Complexing agent if mixed with Aqueous Pb2+ What problems can we run into? 1. pH not high enough to remove proton 2. pH too high and results in lead hydroxide formation 1.2 Pb2+ Pb(OH)42- 1 Pb(OH)+ Pb(OH)2 Fraction 0.8 0.6 0.4 Pb(OH)3- 0.2 0 0 2 4 6 8 pH 10 12 14 Need to consider this a separation D Higher pH To get reproducible results you will need to Set a standard procedure for number of Shakes and total time. Also not water soluble Non-Water soluble Other Considerations? False Positives Any other metals (including Mg2+!!!) can cause a color change The chalk used to line the interior of your Protective gloves can cause false positives Solution? Selectively complex other metals and leave behind the lead!!!! Which complexing agent would you use? Want low value for lead High value for others CN might be good BUT!!!!! Iron ferricyanide Serves as an Oxidizing/reducting reagent Add CN to get rid Of Cd, Hg, Ni, Ag, and Zn, But Also add Citrate to pull The iron from ferricyanide To citrate form. Mild oxidation of the unreacted Dithizone results in a dimer linked by a S-S bond which absorbs at 420 (see spectra 2). More extended oxidation results in cyclization with a product that absorbs at ~610 to 620 nm. Estimated molar absorptivity of the dimer is 30000 to 49000 Some “Data Considerations” How will you choose a wavelength from which to make a calibration curve? How will you determine if you still have unreacted dithizone contributing to Your signal? How will you quantitate the absorbance at the wavelength you choose? How will you choose a wavelength from which to make a calibration curve? 1. Want a region where the signal does not change rapidly (the top of a peak) 2. Want a region where the analyte signal has the least contribution from the background (peak of Pb-complex) How will you determine if you still have unreacted dithizone contributing to Your signal? 1. Monitor wavelength of peak in the 600 region or deconvolute the data How will you quantitate the absorbance at the wavelength you choose? Ameasured ,some Aanalyte,some Abackground , some Aanalyte,some Ameasured ,some Abackground , some Method 1 Abackground , some background cmpound 1, at some bcell pathlength Cbackground cmpound 1 , Abackground , some B , 1 b CB Monitor B at wavelength where only B absorbs and at the wavelength of interest Make a calibration curve at those wavelengths with standards for The background (unreacted dithizone); determine molar absorptivities Calculate concentration of unreacted dithizone for the measurement of Pb by Use of the calibration curve for unreacted dithizone Calculate the absorbance due to unreacted dithizone A680 freedithizozne,680 b Free dithizone A555, free dithizone freedithizozne,555 b Free dithizone A680 freedithizozne ,680 b Free dithizone A555, freedithizone freedithizozne ,555 b Free dithizone A555, freedithizone freedithizozne ,555 A680 freedithizozne , 680 A555,lead dithizone Ameasured ,555 A555,unreacted , freedithizone Measurement Background A Method 2 Much easier And makes no Assumptions about What is contributing To the background Set a baseline across the bottom of the peak The difference in absorbance between the two is the background corrected signal Aanalyte,some Ameasured ,some Abackground , some Baseline estimation Use this lab to introduce another data manipulation Use this lab to introduce another data manipulation Method 3 – assess contribution by assuming Gaussians A f x e 2 1 x 2 2 Assume absorbance peak is Gaussian in the energy spread Energy of The absorption bands photons e Energy levels are randomly populated By Temperature E h Frequency = absorption E c hc Std~(first guess) width at ½ peak ht A Amax or peak s 2 2 exp 1 v vmax s 2 hc Energy of light absorbed Deconvolution 1.2 1. Get the absorption spectra 1 Absorbance 0.8 0.6 Series1 0.4 0.2 0 300 350 400 450 500 550 Wavelength (nm) 600 650 700 750 Deconvolution 1.Assume absorbance peak is Gaussian in the energy spread 2.Convert data from A vs wavelength to A vs energy 1.2 1 0.6 Series1 1.2 0.4 1 0.2 0.8 0 300 350 400 450 500 550 600 650 700 750 Wavelength (nm) A Absorbance 0.8 0.6 Series1 0.4 Notice the 2 curves look Different! 0.2 0 0 5 10 15 20 cm-1 25 30 35 Deconvolution 3.Using your data estimate: center of peak (mean) standard deviation amplitude 1.2 1 0.8 A A A f x e 2 1 x 2 2 0.6 Series1 0.4 std 0.2 0 0 5 10 15 20 cm-1 25 30 35 1.2 1 Absorbance 0.8 0.6 Series1 0.4 0.2 0 300 350 400 450 500 550 600 650 700 Wavelength, nm 750 Wavelength (nm) 1.2 1 A 0.8 0.6 Series1 E h 0.4 0.2 hc Frequency, Cm-1 0 0 5 10 15 20 cm-1 25 30 35 Sum of all the bands Conversion to energy =10000/A10 Guess four absorbance bands Value to be minimzed =(H10-B10)^2 energy wavelength Minimize Target cell Prevent solver from giving You non-plausible (negative) numbers Plot the wavelength based absorption data And superimpose the data generated by solver Now sum all the individual bands and see if you get a low sum of sq differences 1.2 1.2 sum sqrd 0.034034 11 A A 0.8 0.8 0.6 0.6 0.4 0.4 0.2 0.2 00 250 250 350 350 450 450 550 550 650 650 nm nm Some deviation here But generally pretty darn good 750 750 850 850 At the wavelength you are interested in 1.2 1 A 0.8 0.6 0.4 0.2 0 250 350 450 550 nm 650 750 850 Go to the column of data representing that Single absorbance band here I use 550 and find the max =max(data range) This will be your absorbance of the band Without the contribution from the other bands Our “signal” Go to 550 and use the A from this band Only!!! (all other absorbances represent Background contributions) Chemistry titrimetric ppm cuppellation ppb Suphuretted water dithizone ppt 8000 B.C.E. 1000 C.E. 1900 1990 2000 This method resulted in the first public health Awareness of lead as an issue for children Baltimore, Department of Public Health GIGO Measurement of Lead Depends on: Chemistry of Lead for Separations Chemistry of Lead for Creating a signal Chemistry of Lead for Creating Background Chemistry of Lead for the Stability of the Signal Garbage In = Garbage Out Convert lead to some compound which can Be measured by some instrument (what ever happens to Be available in your lab) Suppose you only have a fluorimeter! Calcein Blue H3C Chromophore – part of molecule sensitive to light HO O O + N H O O O - O - “Selectivity” arm – complexes the metal ion and turns On and off fluorescence H3C H3C O O + N H O O O - - Note role of resonance here O OH O N H O O O - - O O Emission Spectra, excitation at 320 Absorbance spectra + 1 0.4 0.9 0.35 0.8 0.3 Absorbance 0.7 0.25 0.6 0.5 480-490 nm emission 0.4 18-33 ns 0.3 duration 0.2 0.15 0.1 0.2 320 nm excitation 0 200 H3C HO O O N H O O - 250 300 350 400 wavelength, nm + O 0.05 0.1 - O pH 6-8 Carboxyl groups only deprotonated 450 500 550 0 600 Relative Fluorescence Intensity HO Excited State Proton transfer H3C O O + N H O O O - Excited State Electron transfer 0.8 0.7 O O O - 0.6 + N H - 0.5 O O O O - AU O - H3C - O 0.4 0.3 0.2 0.1 0 270 320 370 Wavelength 420 470 1 0.7 0.9 0.5 0.7 Absorbance 350-360 440-460 0.6 0.4 0.5 0.3 0.4 0.3 0.2 0.2 0.1 0.1 0 H3C H3C200 H-N HO O O + pH 8-11 O - O O + N H O - - O Ground state phenolic deprotonation O O O - O - 300 350 400 Wavelength, nm O O N H 250 450 500 550 0 600 Relative Fluorescence Intensity 0.6 0.8 Key point so far – Excitation is pH dependent Therefore the emission location and intensity is also pH dependent If the solution is fluctuating in pH will not get a linear working curve. Since you have to control pH for the chemical signal, need to also consider The role of pH in the form of the lead that is present. Why might Pb quench the emission? Lead quenches emission Structures as determined from NMR GIGO Measurement of Lead Depends on: Chemistry of Lead for Separations Chemistry of Lead for Creating a signal Chemistry of Lead for Creating Background Chemistry of Lead for the Stability of the Signal Garbage In = Garbage Out Convert lead to some compound which can Be measured by some instrument (what ever happens to Be available in your lab) Suppose you only have an IR! Need to convert Pb to some form that is amenable to IR and/or Raman spectroscopies. 1. React lead with some reagent This data can Be found in The appendix To “Sublime Lead” web page Need to convert Pb to some form that is amenable to IR and/or Raman spectroscopies. photon Change in bond length O O OH N OH HO N HO O O Pb2+ Key Data Manipulation Concepts from the Lab IR instrument allows you to set the number of waveforms that you Will average. You will need to enhance the sensitivity near the base of one peak so That you can see the background fluctuations in a single scan Repeat for 4 scans Repeat for 9 summed scans Repeat for 16 summed scans, etc. What do you think you will be asked to observe? A Case of Forensic Chemistry: Art and Forgeries Lead Tin II, Paolo Veronese, Allegory of Love Lead Antimonate Peter Rubens, The Dying Seneca Lead Tin I Forensic Art Chemistry Two Sb octahedra Linked via vertices to a) eight pointed polyhedra Of Pb & b) Hexagonal bipyramid Lead Antimonate Lead Tin I Lead Tin II Chains of Sn octahedra Joined by Pyramidally Coordinated Pb(II)