Introduction to Arson Analysis Updated 4-15-09 for an arson homicide trial (On the Web at www.denvergov.org/Crime_lab/ est. 1-?-05/ 3-10-06) The original presentation was also found as a reference link at the AAFS and at Zeno’s Forensic Science Site est. 3-30-05/3-10-06 Arson Analysis by Gas Chromatography/Mass Spectrometry Denver Police Crime Laboratory William D. McDougall II Forensic Analyst (Retired) Original Presentation September 2004 wdmll@comcast.net An abbreviated version of this PowerPoint presentation was given to a joint Denver Fire Department (Arson Bureau) and BATF Seminar for fire investigators (September 2004). This slide presentation has been expanded and will be upgraded as needed. References and Credits A number of the graphic slides and annotations were obtained from the Internet. Many of the annotations and graphic displays have been altered (and added to), to highlight the present topic. I have displayed these slides in a progressive manner of complexity and I have tried to bridge slides of similar material so that it is easy to compare the material on different slides. Most of the mass data displays were obtained from a Varian Saturn 2000 Ion Trap Mass Spectrometer and a HP 5973 Mass Spectrometer in the Denver Police Crime Laboratory. M. Jennifer Thomas, Forensic Chemist, generated most of the mass data displays obtained from the HP 5973 Mass Spectrometer. This HP GC/MSD is comprised of a HP 6890 GC interfaced to a HP 5972 Mass Selective Detector Quadrupole Mass Spectrometer. The system uses electron impact (EI) ionization and is capable of performing full mass scans or selective ion monitoring (SIM). Gas Chromatograph-Ion Trap Mass Spectrometer (MS/MS) Injector > Control Panel Column Oven Gas Chromatograph Ion Trap Mass Spectrometer Chromatography “chromato-graphy” means color writing It is a physical process of separating complex mixtures THIN LAYER CHROMATOGRAPHY The liquid (solvent or solvents) is the mobile phase NON-POLAR Silica gel SiO2.H2O POLAR STATIONARY PHASE (On plastic, glass or foil backing) MEDIUM POLAR POLAR ( NON-POLAR, MEDIUM POLAR OR SOLVENT MIXTURE ) COLUMN CHROMATOGRAPHY ADSORPTION CHROMATOGRAPHY POLAR POLAR STATIONARY PHASE MEDIUM POLAR aluminum oxide Al2O3 (Less polar than silica) Normal Phase Chromatography Interaction of the adsorptivity and solubility of the analytes relative to the two phases NONPOLAR THE MORE POLAR COMPOUNDS MIGRATE SLOWER THAN LESS POLAR COMPOUNDS. NON-POLAR LIQUID MOBILE PHASES ARE USED IN THE MIGRATION OF LESS POLAR COMPOUNDS AND MEDIUM POLAR LIQUID MOBILE PHASES ARE USED IN THE MIGRATION OF MORE POLAR COMPOUNDS. THE RIGHT MIX OF SOLVENTS ( NON-POLAR AND MEDIUM POLAR OR MODIFIER ) IS NEEDED FOR THE MIGRATION AND SPEPARATION OF ALL THE POLAR ANALYTES . NON-POLAR COMPOUNDS MIGRATE AHEAD ( ELUTE FIRST ) OF THE POLAR COMPOUNDS. MODERN HPLC SYSTEM CAN HAVE ONE OR MORE MOBILE PHASES CAN HAVE MORE THAN ONE PUMP API (ESI) M+1 Mass Spectra Controlled substances including “steroids” are excellent candidates for HPLC REVERSED PHASE HPLC THE STATIONARY PHASE IS HYDROPHOBIC (NON-POLAR). MODERATELY POLAR AND NON-POLAR COMPOUNDS START TO MIGRATE THROUGH THE COLUMN AS THE MOBILE PHASE CHANGES FROM AQUEOUS TO LESS POLAR ORGANIC SOLVENT. (Solvent Programming) High pKa (basic) compounds will have a large k’ (longer retention time) in basic buffered mobile phases. Low pKa (acidic) compounds will have a large k’ in acidic buffered mobile phases. http://www.shu.ac.uk/schools/sci/chem/tutorials/chrom/gaschrm.htm Gas Chromatography –The sample mixture is injected and vaporized. Next the mixture is transported through the heated column by the flow of an inert, gaseous mobile phase. The column contains a thin coating of a liquid stationary phase which differentially retards the migration of the sample components. After passing through the heated column, The sample components have to be in the vapor state in order to pass through the column. The separated sample components enter The detector causing an increase in the signal which is recorded as a peak for each component. The collection of peaks is called a chromatogram. < Pattern of peaks (Separated Components) (6) The outside dimension of a capillary column is about the size of a heavy fishing line. The Column The Heart Of The Gas Chromatograph Columns for arson analysis are normally 15 to 30 meters in length. A .25mm ID column is commonly use for arson analysis (Liquid stationary phase .25um thickness) HP-5 Analyte > < Air Non-retained 5% phenyl methylpolysiloxane is a relatively non-polar general purpose stationary liquid phase Retained Chromatogram DYNAMIC PARTITIONING TAKES PLACE IN GAS (LIQUID) CHROMATOGRAPHY Static Partitioning < BASIC AQUEOUS PHASE High pKa (basic) compounds such as cocaine and methamphetamine are found in higher concentrations in the lower phase after equilibration. < CHLOROFORM PHASE K= C stationary phase C mobile phase DYNAMIC PARTITIONING THE ANALYTES PARTITION BETWEEN THE STATIONARY LIQUID PHASE AND THE MOBILE GAS PHASE. N= # of theoretical plates The column is made up of a large number of partitioning (separatory) zones PACKED COLUMNS HAVE APPROX. 10,000 PLATES Each partitioning zone equals one theoretical plate CAPILLARY COLUMNS HAVE APPROX 100,000 PLATES CAPILLARY COLUMNS HAVE MORE THEORETICAL PLATES THAN PACKED COLUMNS MINIBORE COLUMNS HAVE EVEN MORE THEORETICAL PLATES PER METER (Used for packed columns) (CAPILLARY COLUMNS ARE NOT AFFECTED BY EDDY DIFFUSION) (CAPILLARY COLUMNS MINIMIZE THIS FACTOR) Carrier Gases NITROGEN, HELIUM AND HYDROGEN HAVE DIFFERENT SLOPES HYDROGEN IS USED AT HIGHER FLOW RATES THAN HELIUM. NITROGEN IS THE LEAST FAVORABLE. Resolution Resolution is proportional to the square root of (N), therefore the length of the column In order to double the resolution, the column length has to be increased by a factor of four “SELECTIVITY” IS DETERMINED BY THE STATIONARY PHASE CHEMISTRY (K’) IS OPTIMIZED BY TEMPERATURE PROGRAMMING Fire Fire is a chemical reaction. It is rapid oxidation with the release of heat and Light (plus CO2 and water). In order for the reaction (fire) to happen, there needs to be an ignition source (a spark or friction), oxygen (air) and a fuel source (such as ignitable liquids). METHANE Basic building block of hydrocarbons Straight Chain (6) Carbons (14) Hydrogens Normal Alkane HEXANE Cycloalkane Branched Alkane 2-METHYLHEXANE (Isoparaffin) METHYLCYCLOHEXANE Found is gasohol and alcoholic drinks 1-HEXENE Thermo-Decomposition ETHANOL Thermo-Decomposition Oxygenated Solvent ACETONE STYRENE < Found in Gasoline > INDANE Structures from NIST 98 version 1.6 1-METHYLINDAN BENZENE (6) Carbons (6) Hydrogens Basic building block of aromatic hydrocarbon compounds TOLUENE Benzene ring plus a methyl group C3 ALKYLBENZENE ETHYLBENZENE Hydrocarbons are fuel for arson fires and (Reference Compound) non-arson related 1,2,4-TRIMETHYLBENZENE fires NAPHTHALENE C4 ALKYLBENZENE HYDROCARBONS FOUND IN FIRE DEBRIS (Some of these compounds are found in both gasoline and thermo-decomposition). Structures from NIST 98 version 1.6 1-METHYLNAPHTHALENE < Toluene Styrene > THERMODECOMPOSITION Styrene and o-xylene have similar retention times. Each separated sample component is burnt in the flame causing an increase in the electrical signal. (PYROLYSIS PRODUCTS) Ethylbenzene > m/p -xylene Each peak represents a compound from the original sample mixture. < m/p -xylene Each time the signal is increased it is recorded as a peak in the chromatogram. Time > > > The analytes in these two chromatograms including styrene and o-xylene would superimpose Using the same time axis. < Toluene Ethylbenzene > IGNITABLE LIQUID MIXTURE Time > > > < o-xylene Toluene followed by the other analytes enter the detector one after another, after migrating through the column. (Note in both chromatograms m/pxylene co-elution) < Chromatograms ( Separated Components ) http://www.shu.ac.uk/schools/sci/chem/tutorials/chrom/gaschrm.htm The passage of three compounds through the column over time. Chromatogram ( Separated Components ) Time >> Gas Chromatography is used for separating compounds in complex mixtures (gasoline, kerosene and etc). The sample mixture is introduced into the heated injection port and an inert gas flowing through the system carries the compounds into the column. The material within the column is called the stationary phase (non-polar stationary phases are best for arson analysis) and the various sample components interact with this material to a greater or lesser degree. The greater the interaction, the slower that particular compound will move through the column. As the various components begin to migrate through the column, they undergo a series of equilibrium steps between the stationary phase and the mobile phase (the carrier gas) so that the separation becomes more pronounced as compounds progress through the column. In the ideal situation (the column length, flow rate and temperature are appropriate for the sample mixture), the differences in interaction are sufficient to allow all the components in the sample to be completely separated (I.e., resolved) from each other. However, you should note that as the compounds migrate through the column, each chromatographic peak, representing a sample component, broadens with increasing time in the column. http://www.uga.edu/srel/AACES/GCtutorial/page1.html The retention (retention time) of a compound depends not only on the column length, the type of stationary phase and flow rate, but also on the column temperature. As the temperature increases, compounds move through the column faster. Thus, one can reduce the analysis time by increasing (temperature programming) the column temperature. In this run the column temperature is increased from 40 C to 240 C at 10 C/ min during the analysis. All of the normal alkane components are completely separated into narrow symmetrical peaks. (Time of Elution) http://www.uga.edu/srel/AACES/GCtutorial/page1.html tR (Retention Time) < Peak Apex (Measured in Minutes) The retention time is the time a compound spends in the column from the time of injection to the time of elution of the compound (measured to the peak apex). < Time of Injection Time >> < n-C17 elutes before n-C18 TIC AMU AMU Additional factors determining elution order and selectivity IN ORDER TO SEPARATE M-XYLENE AND P-XYLENE, A POLAR STATIONARY PHASE WOULD BE NEEDED. In addition to molecular weight, molecular structure is a factor determining the boiling points, thus the order of elution. Time >> > Note the different profiles or patterns of mountain peaks Note the different profiles or patterns of peaks Kerosene Gasoline Pattern recognition Chromatography peaks Diesel Fuel Summery of gas chromatography (FID) used in fire debris analysis Gas chromatography – A gas carries the mixture through a column. The column is coated with a thin layer of a semi-liquid phase. The liquid phase retards the mixture based on boiling points and molecular weight (mass). The lighter components pass through the column first and the heavier components pass through the column last. As the components leave the column, they are burnt and ionized in the flame of the detector. The increased ionization produces an electrical signal that is sent to a recorder and displayed as a profile of peaks or a chromatogram. Each peak in the chromatogram is identified by its retention time. The overall pattern (chromatogram) can be compared to standard chromatograms (gasoline, kerosene, diesel and etc.). If the unknown sample displays a strong signal without background interference then the above is all that is needed. But most of the time, the signal is weak and there is background interference. Furthermore, if the unknown is weathered (ignitable liquids evaporate) then the pattern is altered and more difficult to identify. Background interference at times, can be falsely reported as ignitable liquids. A more definitive detector is needed. A mass spectrometer should be used to test fire debris cases. This type of detector produces mass spectra and extracted ion profiles in addition to chromatograms and retention times. The combination of all of these displays is a fingerprint for the identification of the ignitable liquid and the components in the ignitable liquid. The Mass Detector The mass detector used in arson analysis is typically a Quadrupole or Ion Trap Mass Spectrometer. Peak (component) Co-elution The preceding slides showed complex mixtures separated by chromatography. The separated components can be individually identified or class identified. But as shown in some of the previous slides, a number of the components were not separated. Peak coelution is a frequent occurrence. There are a number of reasons including complexity of the mixture, the chemistry (chemical makeup of the stationary and /or mobile phases) of the separation process and the chemistry of the components in the mixture. The length of the column, temperature, the analysis time and other parameters affect the separation process. New advances in chromatography are improving the separation or isolation of components in complex mixtures resulting in reduced analysis time. Mass spectrometry carries out a second separation process. Mass spectrometry is used to ionize, fragment and filter the mixture component ions (after the mixture components pass through the column), producing mass spectra. Mass spectrometry computer algorithm techniques (Quadrupole and Ion Trap) use ion extraction algorithms to produce mass chromatograms. The algorithms isolate or extract ion information about the non-resolved and resolved (separated) components. Tandem mass spectrometry (MS/MS) uses multiple quadrupole mass filtering (MS/MS) or ion trapping (MS/MS in time). In addition to ionization and fragmentation, an ion pre-isolation process and collision-induced dissociation precedes a secondary ion separation (filtering) and mass spectra formulation. New computer algorithms coupled with fast scanning detectors, such as time of flight mass spectrometry (TOF) perform peak deconvolution. Peak deconvolution, peak find algorithms plus automated library search routines, are used to extract (isolate) mass spectra, identify and confirm the presence of multiple component co-eluting compounds, in complex mixtures. In addition to a profile of the separated components (chromatogram), the mass spectrometer produces a mass fragmentation pattern (mass spectrum) for each separated component (compound). Electron impact ionization (EI) produces In order to measure the mass of a compound it has to be ionized. The mass to charge ratio is actually determined. Normally the charge is one. The molecular ion is not stable using electron impact ionization (EI) and undergoes fragmentation. The fragmentation pattern is a display of intensity versus mass of the remaining molecular ion and the newly created fragment ions. Pressurized positive ions, negative ions and neutral species. But only the positive ions are analyzed in this mode of analysis. The neutral species and molecules (not ionized), are pumped away. Vacuum Quadrupole filter Ion Trap Ionization & Quadrupole Fragmentation AC and DC Varied voltages The ions are scanned from the low masses to the high masses over time. < Mass Spectra Mixture (3) Molecular ion > Separated Components (3) (3) http://ull.chemistry.uakron.edu/gcms/MS_detector/index.html http://www.jeolusa.com/ms/docs/whatisms.html 1=H+ 16=O+ 17=[OH]+ 18=[H2O]+ Gas Chromatograph-Ion Trap Mass Spectrometer (MS/MS) Extracted ion profiles of gasoline Gasoline Toluene Toluene Aromatics Pattern of peaks Indanes Toluene Naphthalenes Chromatogram (TIC) Aliphatics Mass Spectrum TIC Sample syringe Injector > Control Panel The ions are scanned (separated) from the low masses to the high masses over time. An increasing RF (AC) voltage is used to filter the mass ions Column Column Oven Gas Chromatograph Ion Trap Mass Spectrometer Supplemental wave forms (end-cap electrodes) can be used for MS/MS enhanced CID of ions 91,105,134 & etc. found in trace gasoline containing large amounts background interference patterns. Mass Spectrometry Vacuum A high vacuum is needed to prevent unwanted collisions between the analyte ions and gas molecules during the migration of the ions from the source, through the analyzer and to the detector. The necessary “mean free path” is achieved with a vacuum of approximately 10 5 torr or less. A mechanical low vacuum pump and a high vacuum pump (diffusion or turbo) are coupled to achieve the high vacuum. The ion trap needs a partial vacuum (approximately 1 millitorr of helium) to dampen the kinetic energy of the ions in order to stabilize their trajectories. Collision-induced dissociation used in MS/MS also needs increased gas pressure (a neutral gas). The sample molecules are ionized, fragmented and trapped. The RF voltage is ramped and the ions from the lowest mass to the highest mass (with unit resolution), are ejected from the analyzer and detected by the detector. (Electron impact ionization) (End-cap) (Ring) Separated components enter here > (End-cap) http://ull.chemistry.uakron.edu/gcms/MS_detector/index.html After the positive ions pass through the mass analyzer with unit resolution, they are detected as an amplified electrical signal. http://ull.chemistry.uakron.edu/gcms/MS_detector/index.html For each positive ion that strikes the inner surface, a gain of approximately one million electrons is produced. http://ull.chemistry.uakron.edu/gcms/MS_detector/index.html Ion Trap Analyzer Showing the different electrical potentials applied to the ion trap (Electron impact ionization) With additional wave form manipulations applied to the End-cap electrodes, MS/MS can be used to further analyze fire debris containing matrices. (Figure taken from www.ivv.fhg.de/ms/ms-analyzers.html#Ion_Trap Special Note! Advanced quadrupole mass spectrometry applications use tandem mass spectrometry in space or time. Multiple quadrupole mass filters in series or a single quadrupole ion trap mass spectrometer (in time), are used to isolate one or more parent ions (MRM) which are subjected to “collision-induced dissociation”. The newly formed ions are called product ions. The product ions and remaining parent ions are then mass filtered to produce mass spectra and mass chromatograms (minus background interference). This application can be useful in arson analysis as a compliment to single quadrupole mass spectrometry. This technique can remove extra artifact peaks found in extracted ion profiles (mass chromatograms) of ignitable liquids containing complex background interference. Tandem mass spectrometry is used in a number of applications including both gas chromatography/mass spectrometry and high performance liquid chromatography/mass spectrometry. (Ion Trap and Quadrupole Mass Filter) (Ion trap) Waters Corporation These equations are used for both the Quadrupole Mass Filter and The Ion Trap (For the Ion Trap) (DC=0V for the Ion Trap) (The Quadrupole Mass Filter uses both DC and RF voltages) The AC frequency is about one megahertz and is called the fundamental RF. It is applied to the ring electrode. az DC = 0V (RF) qZ ^ || The larger the mass the smaller the q A constant RF voltage is used to trap the ions The instability along the (RF) axis is expressed as DC az qz q z z Trajectory of a single ion > . > 0 9 (unstable) q z RF Resonance allows an ion to escape Before q > 0.9 q zO RF z m c The larger the RF voltage the larger the q Z (The RF voltage is ramped (> 6000 V) to destabilize the ion trajectories) (The frequency of the AC applied to the end-caps is approximately ½ of the fundamental RF) Also called axial modulation (When the secular frequency of an ion is equal to the end-caps’ frequency it undergoes resonance) (The AC voltage applied to the end-caps is approximately 3 V) (Secular Frequency: frequency at which an ion oscillates in the trap) The secular frequency is dependent on the q z value Comparative diagram, roughly to scale, of a quadrupole mass filter and a cutaway view of a quadrupole ion trap mass analyzer. http://www.abrf.org/ABRFNews/1996/September1996/sep96iontrap.html Quadrupole mass spectrometers consist of an ion source, ion optics (lenses) to accelerate and focus the ions through an aperture into the quadrupole filter, the quadrupole filter itself with control voltage supplies, an exit aperture, an ion detector, detection electronics, and a highvacuum system. Schematic of a quadrupole filter The ions are scanned (stabilized) from the lowest mass to the highest mass (with unit resolution) through the quadrupole filter and detected by the detector. (RF voltage) http://www.chem.vt.edu/chem-ed/ms/quadrupo.html http://avogadro.chem.iastate.edu/CHEM577/CHEM577-C.pdf RLC circuits are designed as high pass filters (high frequencies) and as low pass filters (low frequencies) in many types of applications. DC and RF voltages (+) Rods RF DC (-) Rods http://www.chm.uri.edu/chm412/Chap11.ppt The Quadrupole mass analyzer consists of four cylindrical rods onto which are applied both RF and DC electrical fields. These four rods are arranged in such a manner that they form one pair in the X plane, and one in the Y plane. As ions enter the Quadrupole, they begin to oscillate in both the X and Y planes, thus causing the lower m/e ions to be destabilized in the Quadrupole whenever the alternating (RF) component of the electric field exceeds the direct (DC) component. In this condition, the lower m/e ions will be thrown out of the Quadrupole and not reach the detector, thereby creating an effective high pass filter. If the direct component exceeds the alternating component then the high m/e ions become unstable, while the lower m/e will be stabilized by the presence of the alternating component making for an effective low pass filter. In the Quadrupole system, the mass analyzer is created by connecting the two pairs of rods in such a manner that the positive pair acts as a high pass filter and the negative pair acts as a low pass filter. By carefully matching the two fields, only ions of a particular mass are able to resonate at the correct frequency allowing them to pass through the Quadrupole at any time. In this regard, the Quadrupole mass analyzer is a very fast and efficient system. http://www.elementalanalysis.com/icp/ http://www.chm.uri.edu/chm412/Chap11.ppt RF >DC + DC Potential Heavy ions pass through the rods RF Potential ( Light ions are destabilized and lost.) Positive Ions DC >RF - DC Potential Heavy ions drift in to the negative rods. http://avogadro.chem.iastate.edu/CHEM577/CHEM577-C.pdf RF Potential (Light ions pass through the rods) The proper ratio of DC and RF potentials results in unit resolution. Band Filter The AC voltage has a frequency in the radio frequency range (RF) RF>DC (+) Rods Small ion High pass filter Large ion DC>RF Low pass filter (-) Rods High and low pass filters combined Become a narrow band filter allowing ions of the same mass/charge to pass through to the detector. The combination of the high pass filtering rods and the low pass filtering rods produce a narrow band filter. With the proper DC to RF ratio, unit mass resolution is achieved. The fixed DC/RF ratio is ramped (DC and RF voltages are increased linearly), allowing all of the ions starting with low (RF) mass to high mass (with unit resolution), to pass through the rods to the detector over time. One scan in approx ½ second. A RLC circuit is used as a tuner for old style radios. When the resonance frequency of a radio tuning circuit matches the broadcast frequency of a radio station, the circuit is “tuned in” to that radio station. Variable capacitors and inductors are used to make the adjustment to the circuit to maximize the signal. http://www.kineticbooks.com/physics/17296/17329/sp.html (RF) The ratio of the DC and RF voltages is adjusted to produce unit mass resolution. The magnitudes of the two voltages (in a fixed ratio) are ramped through the entire mass range. A calibration gas is used to tune and calibrate the mass spectrometer. Masses (69,219 & 502) are tuned by choosing the proper DC to RF ratio to achieve unit resolution (scan-line). Next the unit resolved mass peaks are assigned masses from a calibration table. Unit < resolution v Tune and Calibration In addition to unit resolution and mass calibration, relative intensities are obtained. Library mass spectral searches of unknown mass spectra are now possible. From NIST 98 version 1.6 Electron Impact Ionization (70 electron volts) The relative intensity of the molecular ion found in different c lasses of hydrocarbons. The below ion fragmentation patterns are a function of molecular structure http://194.94.42.12/licensed_materials/00897/papers/0007004/7402 26ww.htm The energy needed to break the bonds in the above classes of hydrocarbons is much less than 70 electron volts. However, to maximize total ion current sensitivity and produce ion fragmentation patterns that are library searchable, 70 electron volts are the standard for electron impact ionization. (Approximately 5 eV will break a carbon, carbon single bond) ~ Next the components are ionized by electron impact (EI) and undergo fragmentation. Separated Components Filament < Toluene Mass Filter > ^ Separated by chromatography Lenses accelerate and focus the ions Toluene Mass Spectrum http://chipo.chem.uic.edu/web1/ocol/spec/MS1.htm The ions are scanned (filtered) from the lowest mass to the highest mass with unit resolution. The ions strike the detector and are recorded as a mass spectrum. The ramp display to the left is for the positive set of rods. The negative set of rods have an equal negative DC slope and a RF 180 degrees out of phase. http://ull.chemistry.uakron.edu/gcms/MS_detect or/index.html The mass spectrum of toluene (methyl benzene) is shown below. The -------------------------------spectrum displays a strong 92 molecular ion, a 91 base peak and an assortment of minor peaks m/z = 65 and below (fragmentation ions). Toluene http://chipo.chem.uic.edu/web1/ocol/spec/MS1.htm The mass spectrum is a graph of intensity versus mass/charge (m/z). The most intense ion in the mass spectrum is called the base peak (displayed as 100%) and the other fragmentation ions are displayed relative to the base peak. The highest molecular weight ion in the mass spectrum normally represents the parent molecule (with an electron removed) and is called the molecular ion (M+). Note below, as the alkyl side chain increases, the molecular ion decreases in relative intensity (see toluene in the last slide). http://chipo.chem.uic.edu/web1/ocol/spec/MS1.htm Ethylbenzene Ethylbenzene http://science.csustan.edu/tutorial/mass/ethyben5.htm Decane is a straight chain hydrocarbon with no ring structure. The molecular ion has minimal relative intensity and the base peak is of comparatively low mass. n-C10 (Decane) TIC The next set of slides show different types of non-aromatic hydrocarbons found in a Light Petroleum Distillate. Note that the Light Petroleum Distillate is compared to a gasoline reference. ( Light De-Aromatized Distillate ) TIC TIC Note the ions 43, 57 and 71 Normal Alkane Heptane n-C7 Note the ions 39, 55, 67(& 69) and 83 Methylcyclohexane Cycloalkane Note the ions 43, 57, 71 and 85 Normal Alkane Octane (n-C8) Note the ions 39, 55, 67 and 83 Cycloalkane Ethylcyclohexane The next slide is a mixture of Polystyrene and Polypropylene Decomposition The ions found in an alkene are displayed. Note that most of the ions are the same as found in a cycloalkane. The cycloalkanes are associated with ignitable liquids and the alkenes are normally associated with background interference matrices. Alkene Polystyrene and Polypropylene Decomposition TIC The next set of slides are used to identify an unknown mixture Molecular weights, mass spectra, ion averaging, library searches, retention times and references are used in the identification of the unknown. In addition, the overall pattern of the mixture is important. Are there any normal alkanes present (intense equally spaced peaks)? Is the mixture an Isoparaffin Product or a Naphthenic Paraffinic Product? Ion summed extracted ion profiling is also routinely used. But to illustrate how to differentiate between branched alkanes (Isoparaffins) and normal alkanes, the following slides are highlighted. No Apparent Equally Spaced Intense Peaks Not a normal alkane mixture? Unknown Mixture Unknown Mixture Based on the position of the unknown mixture to the reference peak (1,2,4Trimethylbenzene) found in gasoline, the mixture is a medium range product. < (n-C10 co-elutes) Ion averaging of all the ions in the TIC Peak ? > Virtually no aromatic ions (91,105,119,134) Unknown Mixture (Medium Range Product) No dominant cycloalkane ions (55,69,83,97) TIC Not a Naphthenic Paraffinic Product (Isoparaffin Product?) (Normal Alkanes??) ^|| Note Ratio || v < Ratio of ion 57 to ion 43 is found in isoparaffins The mixture is composed of compounds containing ions 43, 57, 71, 85, 99 and 113. These ions are found in both isoparaffins and normal alkanes. < Normal Alkane ? Unknown Mixture Retention Time of 7.97 min (Straight Chain Hydrocarbon) (Medium Range Product) TIC Mass Spectrum of Peak With Retention Time of 7.97 min Note Ratio Has the ion ratio of a branched alkane (isoparaffin) and < has a molecular ion of 156 Daltons The normal alkane “Undecane” (n-C11) also has a molecular ion of 156 Daltons Mass Spectrum Of Peak From the Unknown Mixture (Retention Time of 7.97 min) Appears to be an Isoparaffin ! Library Search Results Of The Above Mass Spectrum Branched Alkane (Isoparaffin) (molecular ion of 156 Daltons) Note! equally spaced peaks > Reference Mixture contains n-C11 < n-C10 < n-C11 Medium Petroleum Distillate (Normal alkane mixture plus smaller amounts of aromatics, isoparaffins and cycloalkanes) Note Ratio TIC Comparison of the unknown mixture in the previous n-C12 > slides to a mixture containing n-alkanes Retention Time 10.334 min < The normal alkane n-C11 (Undecane) has a molecular ion of 156 Daltons n-C11 (Undecane) Unknown Peak ? > Questionable peak and Undecane have different retention times. TIC Medium Isoparaffin Product? Retention Time 7.97 min Ion 57 Ion 41 Branched Alkane (Isoparaffin) 71 85 Molecular ion 99 Ion 156 Reference < n-C11 Medium Petroleum Distillate Ion 43 TIC Ion 57 71 Retention Time 10.334 min 85 Undecane (n-C11) 99 Molecular ion Ion 156 Reference Medium Petroleum Distillate Unknown In summary, the unknown is a Medium Isoparaffinic Product. The following set of slides illustrate additional ways of displaying an ignitable liquid using a mass spectrometer. Since gasoline is often encountered in arson cases it has been used for this demonstration. Ion Trap Mass Spectrometer TIC ( Total Ion Chromatogram ) The gasoline pattern in the last slide represents a very strong signal with no background interference or extreme weathering. Mass Spectrometry or a Flame Ionization Detector (FID) could have been used to analyze this particular sample. Heated head space sampling of high concentration unknowns (light range mixtures, medium range mixtures and gasoline) with minimal background interference or extreme weathering can at times be analyzed by FID. However, mixtures that cannot be easily identified by pattern recognition should be analyzed by mass spectrometry. In addition, sampling by activated charcoal strips (passive adsorption) increases the possibility of detecting trace ignitable liquids but also increases background interference patterns. Therefore, the following mass data display techniques are strongly recommended for most fire debris samples. TIC Extracted ion profiles Extracted ion profiles are electronically (by computer) simplified mass chromatograms. By using extracted ion profiles it is possible to isolate classes of organic compounds. The isolated classes are groups of isomers or chemically similar compounds found in ignitable liquids. Extracted ion profiles are paramount in isolating ignitable liquids from background interference. Ions: 91+105+119+133 Ions: 117+132 Ions: 128+142+156 Ions: 55+69+83+97+111 Ions: 57+71+85+99 TIC Library Search Results 1,2,4-TRIMETHYLBENZENE Weathered Gasoline The C3, C4 and C5 alkylbenzenes should be present in an unknown sample to be considered gasoline. Weathering can alter their relative concentrations. Additionally, the naphthalenes, indanes, aliphatics and cycloalkanes found in gasoline should also be present. Their relative concentrations are also modified by weathering. 70% Evaporated Gasoline TIC TIC 90% Evaporated Gasoline TIC TIC 99% Evaporated Gasoline TIC TIC < 1,2,4-TRIMETHYLBENZENE Weathered Patterns of Ignitable Liquids Because ignitable liquids are composed of light (flammable) and/or heavier combustible components, the composition can change when exposed to fire or air. The lighter components evaporate first, followed by the heavier components. In a chromatogram of weathered gasoline, the lighter components (peaks) in the front end (on the left side) are the first to dissipate or disappear. By the time gasoline is 99% evaporated, most of the components are missing. According to the MSDS (MATERIAL SAFETY DATA SHEET) for gasoline, washing with soap and water will remove gasoline from skin exposed to gasoline. The following slides are simplified examples of the classification of the different types of Ignitable liquids. Ignitable liquids are liquid accelerants (they fuel and enhance the flame). There are literally hundreds of commercial products containing Ignitable liquids (flammable and combustible ranging from the light range to the heavy range). Ignitable liquids can be as simple as acetone (finger nail polish remover), wood alcohol(methanol) and ethanol ((found in beer, gasohol and 85% ethanol fuel (E85)). They can be found in lighter fluids (Zippo and Ronsonol), camp fuels (Colman fuel), charcoal starters, gasoline, kerosene (used in jet airliners), diesel (used in diesel trucks and cars) and home heating oils. They range in ability to ignite from the light range (flammable) through the middle range (combustible) to the heavy range. In the following ignitable liquid classifications, the various ignitable liquids are compared to gasoline. 1,2,4-Trimethylbenzene found in gasoline is the reference peak. 1,2,4-TRIMETHYLBENZENE co-elutes or has a retention time similar to decane (nC10). A solvent or solvent mixture that is to the left of the reference peak is considered a light product or light mixture. A narrow range mixture that is centered relative to the reference peak (or is just to the right of) is a medium mixture. Any wide range mixture, whose center of peaks is to the right of the reference peak, is a heavy mixture. Light Petroleum Distillate Classification There are many commercial solvents and solvent mixtures available to the general public. Ignitable liquid classification schemes have been developed to group (classify) the various solvents and solvent mixtures. In addition, (if a mixture is light, medium or heavy) the composition will further determine the overall classification (see the classification scheme at the end of the presentation). ( Small concentration of aromatics ) ( Light De-Aromatized Distillate ) A solvent or solvent mixture that is to the left of the reference peak is considered a light product or light mixture. A narrow range mixture that is centered relative to the reference peak (or is just to the right of) is a medium mixture. Any wide range mixture, whose center of peaks is to the right of the reference peak, is a heavy mixture. The reference peak is 1,2,4-TRIMETHYLBENZENE (n-C10 has a similar retention time or may co-elute) found in the gasoline reference. Gasoline Classification THE MOST USED AND ABUSED SOLVENT MIXTURE TIC (Total Ion Chromatogram) A solvent or solvent mixture that is to the left of the reference peak is considered a light product or light mixture. A narrow range mixture that is centered relative to the reference peak (or is just to the right of) is a medium mixture. Any wide range mixture, whose center of peaks is to the right of the reference peak, is a heavy mixture. The reference peak is 1,2,4-TRIMETHYLBENZENE (n-C10 has a similar retention time or may co-elute) found in the gasoline reference. Medium Petroleum Distillate Classification TIC TIC < 1,2,4-TRIMETHYLBENZENE TIC < 1,2,4-TRIMETHYLBENZENE TIC A solvent or solvent mixture that is to the left of the reference peak is considered a light product or light mixture. A narrow range mixture that is centered relative to the reference peak (or is just to the right of) is a medium mixture. Any wide range mixture, whose center of peaks is to the right of the reference peak, is a heavy mixture. The reference peak is 1,2,4-TRIMETHYLBENZENE (n-C10 has a similar retention time or may co-elute) found in the gasoline reference. Heavy Petroleum Distillate Classification TIC TIC FUEL TIC TIC Note! Biodiesel (B20) contains petroleum diesel and 20 % fatty acid methyl esters. These esters appear primarily after nC18. (B20) may be found in some arson fires as the product is used in more trucks and cars. (see slide 112) Diesel Fuel Note the shift to the right when compared to gasoline Gasoline A solvent or solvent mixture that is to the left of the reference peak is considered a light product or light mixture. A narrow range mixture that is centered relative to the reference peak (or is just to the right of) is a medium mixture. Any wide range mixture, whose center of peaks is to the right of the reference peak, is a heavy mixture. The reference peak is 1,2,4-TRIMETHYLBENZENE (n-C10 has a similar retention time or may co-elute) found in the gasoline reference. Isoparaffinic Classification Medium Isoparaffinic Product A Medium Naphthenic Paraffinic Product is basically a Medium Petroleum Distillate with most of the aromatics and normal alkanes removed. The normal alkanes may be present in diminished amounts. The remaining components are cycloalkanes and isoparaffins. A solvent or solvent mixture that is to the left of the reference peak is considered a light product or light mixture. A narrow range mixture that is centered relative to the reference peak (or is just to the right of) is a medium mixture. Any wide range mixture, whose center of peaks is to the right of the reference peak, is a heavy mixture. The reference peak is 1,2,4-TRIMETHYLBENZENE (n-C10 has a similar retention time or may co-elute) found in the gasoline reference. Naphthenic Paraffinic Classification Kerosene has a normal alkane range from approximately C8 to C17. Diesel fuel has a normal alkane range from approximately C9 to C25. Both kerosene and Diesel Fuel are heavy petroleum products. Many lamp oils are light kerosene mixtures. A Heavy Naphthenic Paraffinic Product (lamp oil) is kerosene with most of the aromatics and normal alkanes removed. The normal alkanes may be present in diminished levels. The remaining components are cycloalkanes and isoparaffins. Kerosene std Heavy Naphthenic Paraffinic Product Normal alkanes are absent or in low concentrations Aromatics are present only in minimal amounts < Note the Cycloalkanes and Isoparaffins Weathered Gasoline std Background Interference Matrices The n-aldehydes have been found in a number of suspected arson cases. The n-aldehydes can be confused for n-alkanes under the right conditions. Toluene Styrene m/p -xylene Ethylbenzene http://people.uncw.edu/tyrellj/CHM585/c5.ppt Polystyrene thermally decomposes into the above chromatogram < Trimer Alkenes Alkenes This triplet elutes close to the triplet Of n-C13 and the Methylnaphthalenes found in gasoline. http://www.psrc.usm.edu/macrog/pp.htm Polypropylene thermally decomposes into the above chromatogram Polystyrene and Polypropylene Thermo-Decomposition V . IN ANOTHER CASE THE SAME PATTERN WITH LESS OF THE TRIMER AND STYRENE (MINUS NAPHTHALENE) WAS IDENTIFIED AS GASOLINE. THIS MIXTURE OF POLYSTYRENE AND POLYPROPYLENE DECOMPOSITION CONTAINS GASOLINE (ALSO CONTAINS TERPENES) ####### < terpene UNK1 (ION SUMMED) EXTRACTED ION PROFILES GASOLINE Ions: 91+105+119+133 Ions: 117+132 GASOLINE Ions: 128+142+156 GASOLINE POLYMER CONTAMINATION POLYMER CONTAMINATION POLYMER CONTAMINTION AND TRACE GASOLINE (HIDDEN) Ions: 55+69+83+97+111 Ions: 57+71+85+99 TIC UNK1 Gasoline std Extracted Ion 119 Terpene Extracted Ion 119 Extracted Ion 134 Extracted Ion 134 Light Medium Heavy C4-C9 C8-C13 C8-C20+ <---------------------------------Petroleum Distillates ------------------------------------> <------------------------------De-Aromatized Distillates---------------------------------> <--------------------------------Isoparaffinic Products-----------------------------------> <--------------------------Naphthenic Paraffinic Products-----------------------------> <-------------------------------- Aromatic Products------------------------------------- > <----------------------------------N-Alkanes Products----------------------------------- > <---------------------------------Oxygenated Solvents <---------------------------------Others-Miscellaneous---------------------------------- > Flammables Combustibles < 1,2,4-trimethylbenzene (& n-C10) A number of the above classes of ignitable liquids are encountered in fire debris analysis (testing). An in-house library of ignitable liquids from the the various classes should be built with chromatograms (TIC), extracted ion profiles and mass spectra. . What if an ignitable liquid is found in a Prestone container? Assuming that the laboratory tests show that the contents are an ignitable liquid, could the residual contents (ethylene glycol) have altered the ignitable liquid pattern? Is ethylene glycol an ignitable liquid? Prestone-ANTIFREEZE/COOLANT (ethylene glycol) MSDS ID: MSDSP149 Ethylene glycol is not a flammable or combustible liquid. It has a flash point above 200 F. Slight to moderate fire hazard when exposed to heat or flame. The flash point is a gauge of how easy it is to ignite a liquid. Flash point is the lowest temperature at which a liquid can form an ignitable mixture in air near the surface of the liquid. The lower the flash point, the easier it is to ignite the material. Liquids with a flash point of 100 F or less are classified as flammable liquids. Liquids with a flash point between 100 F and 200 F are classified as combustibles. For example, gasoline has a flash point of -40 degrees C (-40 F) and is more flammable than ethylene glycol (antifreeze) which has a flash point of 111 degrees C (232 F). This chromatogram is from a more polar column than used for ignitable liquids. Therefore, the retention times are greater (longer) than had these compounds been analyzed on a typical column used for ignitable liquids. CH3OH (methanol) Molecular formula of ethylene glycol C2H6O2 Above is a chromatogram showing ethylene glycol in addition to methanol and propylene glycol. All of the compounds in this chromatogram are polar compounds (soluble in water).These polar compounds are all alcohols. Knowing that methanol is a smaller polar compound than ethanol, it is apparent that ethylene glycol would come out at the beginning of a typical chromatogram (TIC) of gasoline and not compromise the gasoline pattern. Biodiesel B20 contains both Diesel fuel and Biodiesel Diesel fuel Note C17 & C18 below Biodiesel B20 Biodiesel B100