Chapter 5: Obtaining an NMR Structure

advertisement

Obtaining an NMR Spectra

Basic Requirements :

NMR sample: compound of interest dissolved in 500-600 m l of deuterated solvent.

Higher the concentration  higher the sensitivity

Magnet: differentiate spin states (aligned/unaligned).

Higher the field strength  higher the sensitivity and resolution

Requires homogeneous field over the sample

RF electronics: generate RF pulse to perturb system equilibrium and observe NMR signal.

Requires accurate control of pulse power and duration

Stability of pulse

Receiver electronics: detection of induced current from nuclear precesson

Requires high sensitivity

Conversion of analog signal to digital signal

NMR Instrumentation (block diagram)

Superconducting Magnet

• solenoid wound from superconducting niobium/tin or niobium/titanium wire

• kept at liquid helium temperature (4K), outer liquid N

2 dewar near zero resistance  minimal current lose  magnet stays at field for years without external power source

Cross-section of magnet magnet spinner sample lift

NMR Tube

RF coils cryoshims shimcoils

Probe

Superconducting solenoid

Use up to 190 miles of wire!

Liquid N

2

Liquid He

NMR Sample

Factors to Consider:

• Maximize sample concentration

Avoid precipitation or aggregation

Use a single deuterated solvent

− Reference for lock

• Avoid heterogeneous samples  distorts magnetic field homogeneity

Avoid air bubbles, suspended particles, sample separation

Avoid low quality NMR tubes

 distorts magnetic field homogeneity

− Breaks easily  damage the NMR probe

Chose appropriate temperature for the sample

Freezing or boiling the sample may break the NMR tube and damage the NMR probe.

• Properly position NMR sample in the magnet

− Position sample in homogeneous region of magnet and between detection and RF coils

Avoid positioning meniscus close to coil edge

 distorts magnetic field homogeneity

Frequency of absorption: n

= g

B o

/ 2 p

Superconducting Magnet

Problems:

− Field drifts (B

0 changes)

Field Drift over 11 Hrs (~ 0.15Hz/hr

Remember: n

= g

B o

/2 p

Lock System

• Need to constantly correct for the field drift during data collection

NMR probes contains an additional transmitter coil tuned to deuterium frequency changes in the intensity of the reference absorption signal controls a feedback circuit; a frequency generator provides a fixed reference frequency for the lock signal

Lock Feedback Circuit if the observed lock signal differs from the reference frequency, a small current change occurs in a room-temperature shim coil (Z0) to create a small magnetic field to augment the main field to place the lock-signal back into resonance

Lock Changes From

Off-resonance to

On-resonance

Lock System

Simply, the lock system can be considered as a separate NMR spectrometer that is constantly collecting a deuterium spectrum and making sure the peak doesn’t move relative to a defined chemical shift

Lock System – things to consider

Measures the resonance of the deuterated solvent

− a number of common solvents (D

2

O, methanol, chloroform) have known deuterium resonance

− Can only lock on one resonance, defined by user.

Multiple deuterium resonances may confuse lock in automated acquisition

NMR sample needs to contain at least 5-10% volume of a deuterated solvent

Consequence of locking wrong solvent

– wrong chemical shifts and missing peaks!

Lock System – things to consider

• Maximize lock signal indicates on-resonance

− Use lock signal to shim sample

• Loss of lock during experiment is problematic  data not reliable

NMR sample degraded

Instrument problem

− Started with weak lock signal

• Increase lock signal by increasing lock gain

− Amplification of the detected lock signal

Increases both signal and noise, so higher lock gain  noisier lock signal

• Increase lock signal by increasing lock power

− Strength of RF pulse to detect lock signal o Too high and lock signal is saturated  intensity of lock signal fluctuates up and down o Too low and lock signal may not be observable

Superconducting Magnet

Problems:

− Field is not constant over sample (spatial variation)

Again: n

= g

B o

/2 p

Magnetic Field Homogeneity

Frequency of absorption: n

= g

B o

/ 2 p

Poor Homogeneity  multiple peaks at different effective B o

Resonance depends on position in NMR sample

Good Homogeneity  single peak with frequency dependent on B o

Shim System

• Corrects for magnetic inhomogeneity

Spatial arrangement of 20 or more coils

Sketch of shim coils actual shim coils change current in each coil to “patch” differences in field and fix distortions in peak shape

Shim Coils

• electric currents in the shim coils create small magnetic fields which compensate for inhomogenieties in the magnet

• shim coils vary in the geometric orientation and function (linear, parabolic, etc)

Z 0 ,Z 1 ,Z 2 ,Z 3 ,Z 4 ,Z 5

X, XZ,XZ 2 ,X 2 Y 2 ,XY,Y,YZ, YZ 2 , XZ 3 ,X 2 Y 2 Z, YZ 3 ,XYZ,X 3 ,Y 3

Shim Coils

Optimize shims by i) minimizing line-width, ii) maximizing lock signal or iii) maximizing FID

Examples of poor line-shapes due to shimming errors

Shim Coils

• Examples of poor line-shapes due to shimming errors

Shim Coils

• Examples of poor FID shape due to shimming errors

Perfectly Shimmed Magnet Mis-shimmed Magnet

Spinning the Sample

Improves effective magnetic field homogeneity by averaging inhomogeneities in the magnet

− Z – shims are also known as spinning shims

• Spinning the sample causes symmetric side-bands at intervals related to spinning rate

Non-spinning shims (X,Y) problems

Samples are never spun for multi-dimensional NMR experiments

− Creates artifacts  streaks or T

1 spinning instability ridges from spinning side-bands and

Spinning side-bands symmetric about peak

Gradient Shimming

• Use pulse field gradients to automate the shimming ( TopShim )

− Gradients - spatial changes to B

0

• Gradients are used to probe (map) the Field (B 0 ) profile

• A Shim Map is unique to each probe

• Requires a Strong Signal (Solvent)

− Requires H

2

O+D

2

O, CH

3

CN+D

2

O or CH

3

OH+D

2

O solvent

Shim Map

Gradient Shimming

• Two General Approaches to Gradient Shimming

− 1D gradshim (Z-shims) seconds to minutes

− 3D gradient shimming (all shims) 5 to 30 minutes

• Shimming is accomplished by matching gradient shims for your sample to shim map

Gradient shim (red) fit to shim map

Gradient Shimming

Water resonance before and after Gradient Shimming

Gradient

Shimming

Environment Stability

Changes in the environment during data acquisition may have strong negative impacts on the quality of the NMR data

• Common causes of spectra artifacts are:

− Vibrations (building, HVAC, etc)

Temperature changes

The longer the data acquisition, the more likely these issues will cause problems

• The lower the sample concentration (lower S/N) the more apparent these artifacts will be

Noise peaks due to building vibrations

Environment Stability

Peak Chemical Shift and Shape

Change as Temperature Changes

Sample Probe

Holds the sample in a fixed position in the magnetic field

• Contains an air turbine to spin, insert and eject the sample

• Contains the coils for:

− transmitting the RF pulse

− detecting the NMR signal

− observing the lock signal

− creating magnetic field gradients

• Thermocouples and heaters to maintain a constant temperature

Sample Probe

Important to note, because of the high magnetic field, the probe has to be built with nonmagnetic material such as glass and plastics.

Thus, probes tend to be fragile and easy to break

Tuning the Probe

• Placing the sample into the probe affects the probe tuning

− Solvent, buffers, salt concentration, sample concentration and temperature all have significant impact on the probe tuning

Probe is tuned by adjusting two capacitors: match and tune

Goal is to minimize the reflected power at the desired frequency

− Tuning capacitor changes resonance frequency of probe

− Matching capacitor matches the impedance to a 50 Ohm cable

Power submitted to transmitter and receiver is maximized

Tune and Match System

• Tune- corrects the differences between observed and desired frequency

Match – correct impedance difference between resonant circuit and transmission line (should be 50

W

)

Adjust two capacitors until the tuning and desired frequency match and you obtain a null

Affects: signal-to-noise accuracy of 90 o pulse sample heating chemical shift accuracy

Tune and Match System

Tune and Match capacitors for a Bruker Probe

Tune and Match System

Changing the Distance Between the Plates or the

Amount of Plate Surface Area which overlaps in a Variable Capacitor

Physical limits to how far the capacitor can be turned in either direction.

If turned too far will easily break!!

Tuning the Probe

Side Notes: Impedance

Impedance

– any electrical entity that impedes the flow of current

 a resistance, reactance or both

Resistance – material that resists the flow of electrons

Reactance – property of resisting or impeding the flow of ac current or ac voltage in inductors and capacitors

Illustration of matching impedance

 Consider a 12V car battery attached to a car headlight

12V car battery – low impedance  high power

 Consider 8 1.5V AA batteries (12 volt total) attached to a very low wattage light bulb

8 1.5V AA batteries – high impedance  low power

 Now swap the arrangement  What happens?

Car battery can easily light the light bulb, but the headlight will quickly drain the

AA batteries  poor impedance match

Tuning the Probe

Side Notes: Quality factor (Q)

− “Q” - dimensionless and important property of capacitors and inductors

 Q - frequency of the resonant circuit divided by the half power bandwidth

All inductors exhibit some extra resistance to ac or rf

Q is the reactance of the inductor divided by this ac or rf resistance

NMR probes Q > 300

 Higher the probe Q the greater the sensitivity

− High Q for an NMR probe is required for high Signal-to-Noise

 Sample can effect the Q of the probe

The sample increases losses in the resonant circuit by inducing eddy currents in the solvent

The more conductive the sample the more the losses and the lower the probe Q.

Water, high salt lower the Q of the probe

Lower Q  longer pulse widths

Q

X

R

L

X – reactance of circuit in Ohms

R

L

– the series resistance of the circuit in Ohms

Pulse Generator & Receiver System

• Radio-frequency generators and frequency synthesizers produce a signal at essentially a single frequency.

• RF pulses are typically short-duration ( m secs)

- produces bandwidth (1/4 t

) centered around single frequency

- shorter pulse width  broader frequency bandwidth o Heisenberg Uncertainty Principal:

Du .

D t

~ 1/2p

- Shortest pulse length will depend on the probe Q and the sample property

A radiofrequency pulse is a combination of a wave (cosine) of frequency w o and a step function

* = t p

Pulse length (time, t p

)

The Fourier transform indicates the pulse covers a range of frequencies

FT

Pulse Generator & Receiver System

• RF pulse width determines band-width of excitation

- Not a flat profile

- All nuclei within ±1/4PW Hz will be equally affected

 1 H 6 m s 90 o pulse  ±41666 Hz  ±69.4 ppm at 600 MHz

 Minimizes weaker perturbations of spins a edges of spectra

- There are also null points at

±1/PW Hz where nuclei are unperturbed

 1 H 6 m s 90 o pulse first null at ±1.67e

5 Hz  ±277.8 ppm at 600 MHz

Maximum affect

Null, no affect

Invert signal, 180 o pulse

Pulse Generator & Receiver System

• RF pulse width determines band-width of excitation

- These issues become a problem at high magnetic field strengths (800 & 900

MHz) for 13 C spectra that that have a large chemical shift range (>200 ppm)

 13 C 15 m s 90 o pulse  ±16666 Hz  ±18.5 ppm at 900 MHz

 Also, complex experiments (multiple pulses) depend on the accuracy and consistency of pulse widths

- Selective pulse  long pulse width (ms)  narrow band-width.

Maximum affect

Null, no affect

Invert signal, 180 o pulse

Pulse Length Calibration

• Need to experimentally determine 90 o pulse

- Measure intenisty of major peak (solvent) in spectrum as the function of 90 o pulse length (P1)

Maximum at 90 0 and minimum at 360 o

Usually measure 90 o pulse at 360 o time point

Pulse Length Calibration

90 o pulse (12 m s)

The pulse width was arrayed from

2 m s to 60 m s in steps of 2 m s

90 o pulse is ~ 11 m s

180 o pulse (24 m s)

360 o pulse (44 m s)

270 o pulse (32 m s)

Pulse Generator & Receiver System

• A magnetic field perpendicular to a circular loop will induce a current in the loop.

90 o NMR pulses places the net magnetization perpendicular to the probe’s receiver coil resulting in an induced current in the nanovolt to microvolt range

• preamp mounted in probe amplifies the current to 0 to 10 V

• no signal is observed if net magnetization is aligned along the Z or –Z axis

Rotates at the Larmor frequency n

= g

B o

/2 p

Continuous Wave (CW) vs. Pulse/Fourier Transform

Continuous Wave –

sweep either magnetic field or frequency until resonance is observed

– absorbance observed in frequency domain

Pulse/Fourier Transform –

perturb and monitor all resonances at once

– absorbance observed in the time domain

Continuous Wave (CW) vs. Pulse/Fourier Transform

NMR Sensitivity Issue

A frequency sweep (CW) to identify resonance is very slow (1-10 min.)

Step through each individual frequency.

Pulsed/FT collect all frequencies at once in time domain, fast (N x 1-10 sec)

All modern spectrometers are FT-NMRs

Continuous Wave (CW) vs. Pulse/Fourier Transform

Fourier Transform NMR

• Observe each individual resonance as it precesses at its Larmor frequency

( w o

) in the X,Y plane.

• Monitor changes in the induced current in the receiver coil as a function of time.

FID – Free Induction Decay

Fourier Transform NMR

Signal-to-noise increases as a function of the number of scans or transients

Increases data collection time

− There are inherent limits: o Gain in S/N will eventually plateau o The initial signal has to be strong enough to signal average.

Increase signal-to-noise (S/N) by collecting multiple copies of FID and averaging signal.

/

number of scans

Fourier Transform NMR

Increase signal-to-noise (S/N) by collecting multiple copies of FID and averaging signal.

/

number of scans

But, total experiment time is proportional to the number of scans exp. time ~ ( number of scans) x (recycle delay; D1)

Fourier Transform NMR

• Recycle time (D1) – time increment between successive FID collection

− Maximum signal requires waiting for the sample to fully relax to equilibrium (5 x T

1

) o T

1

– NMR relaxation parameter that will be discussed in detail later in the course

− Most efficient recycle delay is 1.3 x T

1

Relative S/N per unit time of data collection

1.3T

1

Repetition time (t

T

/T

1

)

Optimize your repetition time …

Fourier Transform NMR

Recycle time (D1)

– time increment between successive FID collection

Typical T

1

’s for organic compounds range from 50 to 0.5 seconds o T

1 relaxation times also vary by nuclei, where 13 C > 1 H o Either estimates from related compounds or experimental measurements of T

1 is required to optimize data collection  especially for long data acquisitions.

Continuous Wave (CW) vs. Pulse/Fourier Transform

Fourier Transform NMR

• NMR signal is collected in Time Domain, but prefer Frequency Domain

• Transform from time domain to frequency domain using the Fourier function

Fourier Transform is a mathematical procedure that transforms time domain data into frequency domain

Sampling the NMR (Audio) Signal

Collect Digital data by periodically sampling signal voltage

ADC

– analog to digital converter

Continuous FID

Digitized FID

Sampling the NMR (Audio) Signal

Collect Digital data by periodically sampling signal voltage

ADC

– analog to digital converter

Sample intensity of voltage induced in coil by y-vector of net magnetization precessing in x,y-plane

Sampling the NMR (Audio) Signal

• To correctly represent Cos/Sin wave, need to collect data at least twice as fast as the signal frequency

If sampling is too slow, get folded or aliased peaks

The

Nyquist Theorem

says that we have to sample at least twice as fast as the fastest (higher frequency) signal.

Sample Rate

- Correct rate, correct frequency

-½ correct rate, ½ correct frequency

Folded peaks!

Wrong phase!

SR = 1 / (2 * SW)

SR – sampling rate

SW – sweep width

Digital Resolution – number of data points

The FID is digitized

Equal delay between points

(dwell time)

DT = 1 / (2 * SW)

Want to maximize digital resolution, more data points increases acquisition time (AQ) and experimental time (ET):

AQ = DT x NP ET = AQ x NS larger spectral width (SW) requires more data points for the same resolution

Sampling the NMR (Audio) Signal

Sweep width (Hz, ppm) needs to be set to cover the entire NMR spectra

Sweep Width

(range of radio-frequencies monitored for nuclei absorptions)

If SW is too small or sampling rate is too slow, than peaks are folded or aliased (note phase change)

Sampling the NMR (Audio) Signal

SW is decreased

The phase of folded peaks can vary:

(a) negative phase, (b) dispersive or

(c) positive phase.

Sampling the NMR (Audio) Signal

Always set SW to be slightly larger than needed to cover the entire spectrum.

Allow for blank space at both low and high chemical shifts.

Correct Spectra

Spectra with carrier offset resulting in peak folding or aliasing

Sampling the NMR (Audio) Signal

NMR data size

• Analog signal is digitized by periodically monitoring the induced current in the receiver coil

How many data points are collected?

What is the time delay between data points?

• How long do you sample for?

− Sample too long  collecting noise & wasting time

All this noise added to spectra

Higher Digital Resolution requires longer acquisition times

Sampling the NMR (Audio) Signal

NMR data size

How long do you sample for?

Sample too short

 don’t collect all the data, lose resolution & get artifacts

FID signal is truncated

Truncated

FID leads to artifacts

Sampling the NMR (Audio) Signal

NMR data size

Digital Resolution (DR)

– number of Hz per point in the FID for a given spectral width.

DR = SW / TD where:

SW – spectral width (Hz)

TD – data size (points)

TD

Dwell time DW

Sampling the NMR (Audio) Signal

NMR data size

• Dwell Time (DW) – constant time interval between data points.

SW = 1 / (2 * DW)

From Nyquist Theorem, Sampling Rate (SR)

SR = 1 / (2 * SW)

• DR, DW, SW, SR, TD are ALL Dependent Valuables

TD

Dwell time DW

Sampling the NMR (Audio) Signal

NMR data size

• Two Parameters that the spectroscopist needs to set

SW

– spectral sweep width

Should be just large enough to include the entire NMR spectra

− TD – total data points

Determines the digital resolution

Contributes to the total experiment time (acquisition time)

Should be large enough to collect entire FID

TD

Total Data Acquisition Time (AQ):

AQ = TD * DW= TD/2SWH

Should be long enough to allow complete delay of FID

Dwell time DW

Sampling the NMR (Audio) Signal

NMR data size

• Increase in the number of data points  increase in resolution

Increases acquisition time

Sampling the NMR (Audio) Signal

NMR data size

• Under sampling the data  truncated FID

Baseline distortions  sinc wiggles

FT

Sinc wiggles

Sampling the NMR (Audio) Signal

NMR Data Processing Software

• Uniform Data Sampling

Traditionally, NMR acquires EVERY data point with a uniform time-step (DW) between points voltage time

− avoids under-sampling frequencies

− FT algorithms expect uniform spacing of digital data

• Reason why nD NMR experiments take so long to collect

Why FIDS are truncated

− Why spectra have low resolution and sensitivity

• No reason why the all the points of the FID need to be collected

Sampling the NMR (Audio) Signal

NMR Data Processing Software

Non-uniform data sampling

− Significant improvement in resolution and sensitivity for nD NMR data

− Don’t need uniform sampling, just need alternative to FFT to process the data.

− The sampling non-uniform scheme is the primary decision and impact on the spectra exponential in t 1 and linear in t 2

Exponential in both t 1 and t 2 randomly sampled from an exponential distribution in t 1 and t 2

Random in t 1 and t 2 .

Graham A. Webb (ed.), Modern Magnetic Resonance, 1305 –1311 .

Sampling the NMR (Audio) Signal

NMR Data Processing Software

Non-uniform data sampling

VERY IMPORTANT POINT, t n is no longer defined by DW and number of points

− t n is now user defined since DW is no longer relevant.

Avoid FID truncation, maximize resolution voltage time

Traditional NMR

FID is truncated because number of points and DW determine how much of the FID can be collected

NUS NMR

FID is under-sampled, but the entire FID is sampled

Sampling the NMR (Audio) Signal

NMR Data Processing Software

Non-uniform data sampling

− Both noise (N) and signal to noise (SNR) are proportional to the total evolution time

− Optimal setting is 1.3T

2 of the evolving coherence

− Maximize sensitivity

Magn. Reson. Chem. 2011, 49, 483

–491

Sampling the NMR (Audio) Signal

NMR Data Processing Software

Non-uniform data sampling

What is the optimal sampling density?

Increase enhancement by increase exponential bias, eventually regenerate truncated FID

− Highly resolved spectra is pT

2

T

SMP

– time constant for the exponential weighting of the sampling.

 – enhancement lw – line width

Magn. Reson. Chem. 2011, 49, 483 –491

Sampling the NMR (Audio) Signal

NMR Data Processing Software

• Non-uniform data sampling

− A 1.5 to 2.0 bias to early data points and a 4x reduction yields a 2x enhancement

Or a 3T

2 with a 3x reduction yields a 1.7 enhancement

Truncated FID

Sampling Density/LW = T

SMP

/T

2

Magn. Reson. Chem. 2011, 49, 483–491

Sampling the NMR (Audio) Signal

NMR Data Processing Software

• Non-uniform data sampling

− Different sampling schemes have different performances at different sampling densities

− Sinusoidal Poisson Gap is currently the best – random sampling, while minimizing gap size particularly at the beginning and end of the FID

− Some drastic sampling densities at 1% or less.

Top Curr Chem. 2012 ; 316: 125

–148

Sampling the NMR (Audio) Signal

NMR Data Processing Software

Non-uniform data sampling

Dramatic gain in the quality of strychnine NMR spectrum with 25% sampling density

The spectrum was collected 4x faster (10 min. vs. 40 min.)

Uniform Sampling Non-Uniform Sampling

Nat. Prod. Rep. 2013 30: 501-524

Sampling the NMR (Audio) Signal

NMR Data Processing Software

Non-uniform data sampling

− How is the time-domain data processed?

Use the partial data to reconstruct the full Nyquist grid then process as normal

 maximum entropy reconstruction is a common approach forward maximum entropy (FM), fast maximum likelihood reconstruction (FMLR)

 multi-dimensional decomposition (MDD); and compressed sensing (CS)

MddNMR: http://www.enmr.eu/webportal/mdd.html

Newton: http://newton.nmrfam.wisc.edu/newton/static_web/index.html

− RNMRTK: http://rnmrtk.uchc.edu/rnmrtk/RNMRTK.html

− mpiPipe: Available by contacting the Wagner Group

Sampling the NMR (Audio) Signal

• Adjusting the Receiver Gain (RG) – electronic amplification of the signal

− There is an optimal setting guided by the limits of the ADC digitizer

FID intensity changes as the number of transients increase during data acquisition

RG depends on NS

Digitizer has a finite data range

Sampling the NMR (Audio) Signal

• Adjusting the Receiver Gain (RG) – electronic amplification of the signal

− If RG set too high, the digitizer is full and the FID is clipped

− Fourier transform of a clipped FID results in sinc wiggles in the spectrum baseline.

Sampling the NMR (Audio) Signal

• Adjusting the Receiver Gain (RG) – electronic amplification of the signal

If RG is set too low, the spectrum will be noisy.

RG should be set as increments of 2, where there is a maximum limit o RG may be set to higher values, but no effect on the spectra will be observed o RG may be set to non-factors of two, but adjusted to nearest factor of 2.

Sampling the NMR (Audio) Signal

• Solvent suppression

 solvent concentration is significantly larger than the sample concentration

 water is 55M compared to typical m

M – mM of compound

With Solvent Suppression

Without Solvent Suppression

Sampling the NMR (Audio) Signal

• Solvent suppression

 strong solvent signal can fill digitizer making it impossible to observe the sample signal

Dynamic range problem

16K – 32K range of intensities

 Need to suppress intense solvent signals with selective saturation pulse

 will discuss different NMR pulses in detail latter

The most intense peak is set to the largest value in the digitizer and every other peak is scaled accordingly

Sampling the NMR (Audio) Signal

• Dynamic range

 defines the range of signal amplitudes (peak intensities) observed in the spectrum

Typically 16 bit or 18 bit digitizers

16 bit digitizer – FID amplitudes range from -2 15 to 2 15

 peak smaller than 1/32768 (16 bit) or 1/131072 (18 bit) of most intense peak is lost!!

32768

Want to “see” weak peaks in the presence of intense peaks

Peak intensity has to fit between range of 1:2 15

1

Quadrature detection

• Frequency of B1 (carrier) is set to the center of the spectrum.

Small pulse length to excite the entire spectrum

Minimizes folded noise carrier

PW excites a corresponding bandwidth of frequencies same frequency relative to the carrier, but opposite sign.

carrier

Quadrature detection

• Frequency of B1 (carrier) is set to the center of the spectra.

− Rate of precession in X,Y plane is related to carrier frequency o Precession is difference from carrier frequency

− Possible to have resonances with same frequency but opposite direction same frequency relative to the carrier, but opposite sign.

carrier

Clockwise – magnetization traveling faster than rotating frame

Counter clockwise – magnetization traveling

slower than rotating frame

Quadrature detection

• How to differentiate between peaks upfield and downfield from carrier?

− observed peak frequencies are all relative to the carrier frequency

Same Frequency!

Opposite sign carrier

How to differentiate between magnetization that precesses clockwise and counter clockwise?

Quadrature detection

• If carrier at edge of spectrum, peaks are all positive or negative relative to carrier

Excite twice as much noise, decrease S/N

− Half of the digital resolution

− Half of the spectrum is irrelevant noise

PW excites a corresponding bandwidth of frequencies centered on carrier carrier

All this noise added to spectrum

Quadrature detection

Use

two

detectors

90 o out of phase.

PH = 0

B

F

B w

(B1)

F

Phase of Peaks are different.

PH = 0

F

PH = 90

F

B

B

Quadrature detection

Use two detectors 90 o out of phase.

FT is designed to handle two orthogonal input functions called the real and imaginary component

Detector along X-axis

(real component of FT)

Detector along Y-axis

(imaginary component of FT)

Phase of Peaks are different  allows differentiation of frequencies relative to carrier

Phase Correction of the NMR Spectrum

Depending on when the FID data collection begins a phase shift in the data may occur.

Phase Shift

Phase correction of the NMR spectrum compensates for this phase shift.

Phase Correction of the NMR Spectrum

Phase shift depends on the frequency of the signal

Phase Shift

Phase Correction of the NMR Spectrum

Phase Shift

Phase Correct

Manually adjust zero-order (PO) and first-order (P1) parameters to properly phase spectra.

Phase Correction of the NMR Spectrum

What is happening mathematically during manual phasing of an NMR spectrum

Fourier transformed data contains a real part that is an absorption Lorentzian and an imaginary part which is a dispersion Lorentzian

 we want to maintain the real absorption mode line-shape

 done by applying a phase factor (exp(i

Q

)) to set

F to zero

 we are effectively discarding the imaginary component of the spectrum

Phase Correction of the NMR Spectrum

If you “over-phase” the spectrum, you get baseline “roll”

Phase Correction of the NMR Spectrum

Power or Magnitude spectrum

 obtain a pure absorption NMR spectrum without manual phasing

 results in broader spectrum that can not be integrated

 not a typical or preferred approach to processing an NMR spectrum

Zero Filling of the NMR Spectrum

Improve digital resolution by adding zero data points at end of FID

 essential for n-Dimensional NMR data

 real gain in resolution is limited to zero-filling to 2AQ ( in theory) or ~ 4AQ in practice

8K data 8K zero-fill

8K FID 16K FID

No zero-filling 8K zero-filling

Zero Filling of the NMR Spectra

Better example of the resolution gain and benefits of zero-filling NMR spectra

No zero-filling 4AQ zero-filling

Applying a Window Function to NMR data

• Emphasize the signal and decrease the noise by applying a mathematical function to the FID.

• Can also increase resolution at the expense of sensitivity

Applied to the FID before FT and zero-filling

Good stuff Mostly noise

Sensitivity

Resolution

Applying a Window Function to NMR data

Simply Multiple FID with a Mathematical Function

F(t) = e - ( LB * t )

X

Applying a Window Function to NMR data

Can either increase S/N or

Resolution

Not Both!

LB = 5.0 Hz

Increase Sensitivity

LB = -1.0 Hz

Increase Resolution

FT FT

Applying a Window Function to NMR data

A Variety of Different Apodization or Window functions

Applying a Window Function to NMR data

• A main goal in applying a window function for a nD NMR spectra is to remove the truncation by forcing the FID to zero.

Truncated FID with spectra “wiggles”

Apodized FID removes truncation and wiggles

Baseline Correction of NMR Spectrum

• It is not uncommon to occasionally encounter baseline distortions in the NMR spectra

The baseline can be corrected by applying a linear fit, polynomial fit, spline fit or other function to the NMR spectrum.

Spline baseline correction

Baseline Correction of NMR Spectrum

A number of factors lead to baseline distortions:

Intense solvent or buffer peaks

Phasing problems

Errors in first data points of FID

Short recycle tines

Short acquisition times

Receiver gain polynomial baseline correction

Xi & Roche BMC Bioinformatics (2008) 9:234

NMR Peak Description

• Peak height – intensity of the peak relative to the baseline (average noise)

• Peak width – width (in hertz) at half the intensity of the peak

Line-shape

– NMR peaks generally resemble a Lorentzian function

A

– amplitude or peak height

(LW

1/2

)

– peak width at half height (Hz)

− X o

– peak position (Hz)

LW

1/2

Y

( LW

1 / 2

A (

)

2

LW

1

4 (

/ 2

X

)

2 o

X )

2

NMR Peak Integration or Peak Area

The relative peak intensity or peak area is proportional to the number of protons associated with the observed peak.

• Means to determine relative concentrations of multiple species present in an NMR sample.

Relative peak areas = Number of protons

3

Integral trace

HO-CH

2

-CH

3

1

2

NMR Peak Integration or Peak Area

Means to determine relative concentrations of multiple species present in an NMR sample.

Need to verify complete or uniform relaxation

Unknown Xylene Mixture ortho from peak heights

17.7% meta

Methyl Region of NMR Spectrum meta (21.3 ppm) para (20.9 ppm)

57.9% ortho (19.6 ppm) para impurities impurities

24.4%

NMR Peak Integration or Peak Area

NMR titration experiments are routinely used to monitor the progress of a reaction or interaction

By monitoring changes in the area or intensity of an NMR peak

Peak Picking NMR Spectra

One of the basic steps in analyzing NMR spectra is obtaining a list of observed chemical shifts

Usually refereed to as peak picking

Most programs have similar functionality, choice is based on personal preference

 display the data (zoom, traces, step through multiple spectra, etc)

Peak-picking – identify the X,Y or X,Y,Z or X,Y,Z,A chemical shift coordinate positions for each peak in the nD NMR spectra

Peak Picking List

Peak# 15 N (ppm) 1 H (ppm)

1 127.747 9.537

2 127.803 9.405

3 114.644 9.312

4 121.299 9.287

5 119.425 9.225

6 126.940 9.181

7 121.296 9.107

8 122.376 9.090

9 133.054 8.983

10 127.974 8.934

11 122.890 8.944

12 117.582 8.928

.

.

.

Peak Picking NMR Spectra

Critical for obtaining accurate NMR assignments

Especially for software for automated assignments

Only provide primary sequence and peak-pick tables

Two General Approaches to Peak Picking

Manual

– time consuming

– can evaluate crowded regions more effectively

Automated

– pick peaks above noise threshold

OR

– pick peaks above threshold with characteristic peak shape

– only about 70-80% efficient

– crowded overlap regions and noise regions (solvent, T

2 ridges) cause problems

– noise peaks and missing real peaks cause problems in automated assignment software J. OF MAG. RES. 135, 288–297 (1998)