WZZ- 3AUTOMATIC POLARIMETER OPERATING INSTRUCTION Please read through these operrating instructions before using CONTENTS I. INTRODUCTION……………………………………………… 1 II. SPECIFICATIONS………………………………………………1 III. PRINCIPLE AND STRUCTURE……………………………… 2 IV. OPERATION…………………………………………………… 2 V. MAINTENANCE……………………………………………… 4 VI. COMMON BREAKDOWN AND HANDING………………… 5 I. INTRODUCTION The polarimeter is an instrument used for determining the Optical Rotation of substances.In this way,the concentration,content ,purity,ect of certain substances can be found out.Therefore,the polarimeter has become one of the most useful instruments in laboratories of sugar industry,pharmaceutical industry,oil refining,foodstuff industry,chemical industry,universities and scientific research field. II. SPECIFCATIONS Principle: digital polarimeter based on Optical Null Principle Modulator:Faraday magneto-optical modulator Light source: monochromater sodium lamp,wavelength 589.44nm Minimum sample transmittance: 1% Range of measurement: ±45º (±120ºZ) Minimum value in reading: 0.001º(0.001ºZ) Accuracy: ±(0.01º+measuring value×0.05%) ±(0.03ºZ+measuring value×0.05%Z) Repeatability:sample transmittance>1% ≤0.002º(0.02ºZ) Available minimum sample transmittance: 10% RS232 interface Sample tube:100mm,200mm Power supply:220V±22V,50Hz±1Hz Dimensions:600mm×320mm×220mm Weight(net):30kg 1.Incandescence Lamp 5.Modulator 9.Lens 13.Auto H.V 17.Non-Linear control 21.Coder 2.Condenser 6.Collimator 3.Lens 7.Sample Tube 10.Filter 11.Aperture 14.Pre-AMP 15.FREQ-Selection 18.Speed-Feedback 19.Servo Motor 22.Display Fig.1 Block Diagram 4.Polarizer 8.Analyzer 12.Detector 16.Power-AMP 20.Drive Gear III. PRINCIPLE AND STRUCTURE Monochromatic rays(wavelength 589.44nm)from the Incandescence lamp plus interfered filter travel successively through the condenser,lens,polarizer,modulator,collimator and then change into a plane-polarized light.The plane of polarization is in swinging caused by the alternating current which flows in the Faraday coil and with the frequency F.Passing through the sample tube containing solution to be measured and the analyzer,another lens,an aperture,the light falls onto the photomultiplier.Now the optical Signal turns into an electric signal and is magnified by the Pre-Amplifier. If the analyzer deviates from the Optical Zero Position(with reference to the polarizer),an alternation optical signal with a frequency F will produce an electric signal with the same frequency F.The electric signal processed by the Selective Frequency Amplifier and the power Amplifier drives the servo-motor and by means of a mechanical gear it turns the analyzer until the Optical Zero Position is regained.At the moment,the electric signal with the frequency F disappears and the servo-motor stops running. As soon as the polarimeter is put into operation,the analyzer will run to stop at the Optical Zero Position.Now we should clear the display(Zero Set).If a sample with the Optical Rotation αis placed into the sample chamber,the analyzer will deviates from the Optical Null Position by angle α.As we stated above,the analyzer will be driven throughα angle to get a new Optical Null Position.Theα value is displayed on a digital readout and so the Optical Rotation of the sample is determined. The auto H.V system serves to change the high-voltage supplied to the photomultiplier according to the density of light fell on it and thus to meet the needs of measuring those dark-colored sample with transmittance near than 1%. IV. OPERATION i. The polarimeter should be placed in a dry place with good ventilation and kept from corrosion.It had better been used under the temperture of 20℃.The instrument should be handled with care and vibration is not allowed. ii. Insert the power plug of the intrument into the 220V power source.(It is required to use an AC electronic voltage regulation(1kVA).)And connect the grounding terminal to the earth reliably. iii. Turn on the power supply switch (in the left). Wait for 15-minute, turn on the Direct Current key iv. L C D o p e r a t e s a n d d i s p l a y . A f t e r 6 s e c o n d s , t h e L C D display ”MODE-model,C-c oncentration,L-tube length,n-time of measurement ;Default:MODE:1,L:2.0,C:0,n:6) v. Display model shift a. Display model classification MODE1-optical rotation; MODE2-specific rotation; MODE3-concentration; MODE4-sugar degree; b. If you need not shift display model,press “ ” button,display “0.000”. c. If you need shift,modify the value of “MODE,L,C,n”,After input,should press “ ” button.When input “n”,press “ ” button,display “0.000”,this means you can measure.When input “C”,if find input error,can press “←”,cursor can be moved,then modify. d. When measuring,need shift model,please press “←”. vi. Display form a. Measure optical rotation-MODE:1 Need input “n” α-value of optical rotation, αAV –average value b. Measure specific rotation-MODE:2 Need input “L(dm),C,n” [α]-value of specific rotation,[αAV]- average value c. Measure concentration-MODE:3 Need input “L, [α],n” C-value of concentration,CAV-average value d. Measure sugar degree-MODE:4 Need input “n” Z-value of sugar degree,[Z] AV- average value Sn-1-standard error of n=6,reflect discreteness of sample preparation and instrument measuring result.Low discreteness,high credence of measuring result. e. Put the test tube containing distilled water or other blank solvent into the sample chamber,and close the cover.Press the “clear” button,display “0”.(If there are air bubbles in the test tube,the first thing should be done is to make the air bubbles float on the protruded tube-neck.The atomized water drops on the both ends of test tube should be wiped dry.The screw-nut of the test tube should not be screwed down too tightly,so as to avoid stress,otherwise,the readings may be affected.When placing the test tube,care should be taken for the marked position and direction.) f. Take out the test tube,then,inject the sample to be measured into the test tube.According to the same position and direction,put the test tube into the sample chamber,then,close the cover.Now the instrument will indicate the optical rotation of the sample. g. The instrument will do repetition measurement for n times automatically,and display the value(when n=6, Sn-1 is effective).If n=1,press “repetition-measurement” button to reset.If n>1,press “repetition-measurement” button,the instrument will measure repeatedly for n times. h. If the polarized angle of the sample exceeds the measuring range,the instrument will be oscillating at ±45º,at this time,the test tube should be taken out,then the instrument will reset to zero automatically.After diluting the sample,measure again. i. Turn off the light source switch,power supply switch step by step when the measuring process is over. j. Please press “zero” button before measurement every time. k. If the zero error<0.01º,no matter how much n is,only return to zero one time. l. You can use RS232 cable to connect the instrument with computer.But first you should install the supplied software. Remark: 1. Calculate the specific rotatory power according to the formula following: [α]=100α/LC Where, α is the measured optical rotation; C is the concentration of solution (g/ml); L is the length of test tube(dm). 2. Calculate the purity of the sample according to the formula following: Purity=Measured specific rotation/Theoretical specific rotation 3. Measurement of the international saccharic concentration According to the international standard of saccharic concentration,it is specified to use 26grams of pure sugar to make up 100ml of solution,to use a 200mm test tube,and to use the sodium light to perform the measurement at 20℃.Its optical rotation is +34.626,and its saccharic concentration is 100ºZ. V. MAINTENANCE i. ii. iii. VI. Keeping The instrument should be kept away from excess moisture,corrosive gases and violent shock. Correction The instrument was calibrated before leaving our factory.If test values deviate from the correct ones,the instrument should be regulated with our standard quartz test tube(available to customer’s order)or with the sample whose Optical Rotation is accurately known.If the procedure above doesn’t succeed somehow,send the instrument back to our factory for a thorough inspection. Replacing the Incandescence lamp(12V,35W) For replacement,move off the ventilation louver on the left side of instrument loosen the screw under the lamp,then the lamp can be pulled out. COMMON BREAKDOWN AND HANDING Breakdown Appearance Analyses for Reason Handling method After turning on the power supply switch,the lamp doesn’t light up. The switch is broken. Lamp is broken. Fuse 1.6A is broken. Change the parts. Or send to our repair department. Press the “CLR” key,no display Contact not well. The display board is broken. The display board is broken. Press again. Change. Having display,not count. The digital board is broken. The photoelectric system is broken. Change. Send to our repair department. .