Introduction and Theory of Chromatography

advertisement
Introduction and Theory of
Chromatography
CU- Boulder
CHEM 5181
Mass Spectrometry & Chromatography
Prof. Jose L. Jimenez
Reading: Braithwaite & Smith Chapters 1 & 2
Last Updated: Nov. 2013
1
Business Items
• Next Tue: invited lecture from Lucas Smith, LECO
• HW due Tue (could extend some of it, but overlap with
1st LV HW)
• FCQs online + our own feedback form, response rate
counts towards class participation grade for everyone
2
Why is Chromatography So Successful?
Single-Step
Thermogram
Ziemann et al.
UC-Riverside
(now CU-Boulder)
Chromatogram
Much higher selectivity,
by doing single step
over and over
3
Schematic of Column Chromatography
Data to
Computer
Stationary Phase
Detector
Stationary Phase
Laser
Mobile
Phase
• Sequence of events
–
–
–
–
At t=0 we will open the gate and let the analyte into the column
Analyte will be carried by mobile phase
Analyte may partition to stationary phase
Analyte will be detected by its absorption of light at the detector
4
Schematic of Column Chromatography II
Stationary Phase
Stationary Phase
• For simplicity, we will assume that the mobile
phase moves in steps rather than continuously
• If analyte had no affinity to the stationary phase, it
would just follow the mobile phase
– Emerge at the detector after tM (mobile phase time)
5
Schematic of Column Chromatography III
SP
• If analyte has some affinity to the stationary
phase, it will be retarded
– Equilibrium
– Kinetics
K
Cs
CM
Distribution Ratio
• Molecular mass transfer: diffusion
– Emerge at the detector after “retention time” tR
6
Mechanisms of Partitioning to the Stationary Phase
Dissolve into the bulk
Stick to the surface
Q: (A) Left is Absorption, Right is Adsorption
(B) Left is Adsorption, Right is Absorption
(C) I don’t know
Fig. from McNair
& Miller, Basic Gas
Chromatography,
Wiley, 1998
7
Retention and Mobile Phase Time
• Simplest chromatogram w/ 2 components
– Unretained peak (tM)
– Analyte peak (tR)
– Corrected retention time (t’R)
8
Chromatography Simulator
Investigate
- Effect of K
- Effect of N
- Effect of RR
• http://www.chem.uoa.gr/Applets/AppletChrom/Appl_Chrom2.html
9
Using the Distribution Ratio: Single Step (“Plate”)
Analyte A
Mobile phase
Stationary Phase
Analyte B
• Q: If each symbol represents a
mol, given 10 cm2 of surface to
which analytes adsorb, and 100
cm3of liquid in which analytes
are dissolved, what are KA & KB?
• Large K has more affinity
for stationary phase
• Small K has more affinitiy for
the mobile phase
10
The first step in a separation
X=0,t=0
Q: does molecular-level
kinetics matter?
Pump
X=0,t=1
X=1,t=1
Key concepts:
• Enriched in
component which
prefers mobile phase
• Not very good
separation in 1-step
(like thermogram). It is
repetition that makes it
great.
11
Partitioning in Real Chromatography
12
Factors Influencing Retention…
• are those that influence distribution (equil.)
– Stationary phase: type & properties
– Mobile phase: composition & properties
– Intermolecular forces between
• Analyte & mobile phase
• Analyte & stationary phase
– Temperature
13
Intermolecular Forces I
• Based on electrostatic forces
– “Like-attracts like” or “oil and water” (similar
electrostatic properties)
• Polar/polar & non-polar/non-polar
– Molecules with dissimilar properties are not
attracted
• Polar retention forces
– Ionic interactions (IC)
– Hydrogen bonding
(permanent dipoles)
– Dipole-Induced dipole
14
Intermolecular Forces II (Dipole)
• Polar forces (cont.):
– Energy of dipole-dipole interaction
 2  A2  S2 

 D   6
 r kT 
A
 dipole moment, A: analyte,
S: stationary phase
– Factor of 10 variation on permanent dipole moment
• Factor of 104 variation on interaction energies
– As r6 => mainly at the surfaces
15
Intermolecular Forces III (London)
• London’s Dispersion Forces
– Most universal interaction between molecules
• Only one for non-polar species
– Due to induced dipoles
• Relatively weak
– Energy of interaction:
L 
3 A P I A I S
2r 6 ( I A  I S )
–  is the polarizability, I: ionization potential, A:
analyte, S: stationary phase
16
Clicker Question
A. The chemical nature of the mobile phase
affects distribution in gas-chromatography
(GC) & liquid chromatography (LC)
B. The mobile phase affects distribution in GC
but not LC
C. The mobile phase affects distribution in LC
but not GC
D. The mobile phase affects doesn’t affect
distribution in either GC or LC
E. I don’t know
17
Separating Efficiency – Peak Width
Assume Gaussian Peaks
FWHM = wh = 2.36
HWHM = 1.18
Base Width of Peak (wb)
• Width at intersection of
tangents at inflexion
points (+/- ) and the
baseline
• wb = 4 
• 95.5% of molecules are
within wb
– Assuming that the
Gaussian model holds!
When in doubt, use NORMDIST
in Excel or GAUSS in Igor
Figure 2.3, Braithwaite & Smith
18
Resolution in Chromatography
Signal Intensity (Arbitrary
Units)
t
• Objective: accurate measurement R 
s
2 * FWHM
of individual peak areas (spectra)
2.5
• What is the
resolution if this is
a mass spectrum?
• If it is a
chromatogram?
2
1.5
1
0.5
0
46
51
56
m/z OR t'R
19
Clicker Question
•
Analysis B is more desirable than A
A.
B.
C.
D.
E.
In MS & Chrom.
In MS but not Chrom.
In Chrom but not MS
In neither MS nor Chrom
I need a coffee
0.25
0.2
0.9
A
0.8
B
0.7
0.6
0.15
0.5
0.4
0.1
0.3
0.2
0.05
0.1
0
100
110
120
130
140
150
160
170
180
0
100
110
120
130
140
150
160
170
180
20
Resolution vs Peak Integration
• http://www.vias.org/simulations/simusoft_peakoverlap.html
21
Can One Have Too Much Resolution?
1.2  Rs  1.5
22
What to do if you have too much resolution?
• If you have too much resolution, you can
– Shorten the column
– Increase temperature (GC), flow rate
=> Shorten the analysis
23
Separating Efficiency – Peak Asymmetry
Q: what can cause peak
asymmetry?
24
Separating Efficiency – Peak Asymmetry
• Tailing: some part of the
stationary phase binds analyte
molecules more strongly
• Fronting: some molecules
move ahead (inject too much
sample => saturate Stat.
Phase)
• Peak Asymmetry
As 
b
at 10% h
a
• 0.9 < As < 1.2 for acceptable
chromatography
Equations for Calculation of Chromatographic Figures of Merit for Ideal and
Skewed Peaks, J.P. Foley and J.G. Dorsey, Anal. Chem., 55: 730-737, 1983
25
Separating Efficiency – Peak Width II
• Described by variance 2 (units: s2)
–  determined from Gaussian fit to peak
• Classical chromatography theory
– Separation in N discrete steps (“plates”)
2
–  
1
1


or
N
N
– Also broadening increases as tR increases:
–   tR
1 and in practice:
N
t 
N  R 
 
2
26
Height Equivalent of a “Plate”
H 
L
N
• Height Equivalent to One
Theoretical Plate (HETP)
Application:
Calculate H and N for peak 18,
assuming a 60 m column
A. H ~ 5 m
B. H ~ 50 m
C. H ~ 500 m
D. H ~ 5 mm
E. I don’t know
27
Concept of Peak Capacity ()

Time or Area Available for Separation
Time or Area of an Individual Peak
Fig. 4. Mobility–mass plot of a
complex mixture containing multiple
classes of ions. In this case peptide,
DNA, and carbon cluster ions are
observed. Lines are superimposed
onto the plot to indicate the mobility–
mass trends for each class of
molecule.
•  of IMS-MS for peptides ~ 2600
•  for LC-FTICR ~ 6 x 107
Peak capacity of ion mobility mass spectrometry: Separation of peptides in helium buffer gas. Brandon T. Ruotolo, Kent
J. Gillig, Earle G. Stone and David H. Russell. Journal of Chromatography B 782, 1-2, 25, 2002, Pages 385-392.
http://dx.doi.org/10.1016/S1570-0232(02)00566-4
28
Diffusion: Fick’s 1st Law
Concentration (y)
j A   DA
SP
Y (position in SP)
• When there is a gradient in
concentration of a species
that can diffuse in medium
dC A
dy
Concentration (y)
SP
~ 0.1-0.01 cm2 s-1 in gases
~10-5 cm2 s-1 in liquids
Y (position in SP)
• jA: molecular flux of A (moles cm-2 s-1)
• CA: concentration of A (moles cm-3)
• DAB: diffusivity of A in B (cm2 s-1)
29
Clicker Question
•
When an analyte is diffusing in the
stationary phase, equilibrium will be
reached faster
A.
B.
C.
D.
E.
When DA is small
When DA is large
When SP thickness is large
A and C
I don’t know
30
From Bird, Stewart & Lightfoot,
Transport Phenomena, 2nd Ed. 2002
Mass Transfer Kinetics: Fick’s Law
•
•
•
•
Initially no substance A
Pulse at the surface
Transient concentration
Steady state profile at long
times
WAy
 0
 DAB A0
S
Y
–
–
–
–
Stat. Phase
WAy: mass flux of A
A: mass fraction of A
DAB: diffusivity of A in B
S: surface area; : density
j Ay   DAB
d A
dy
– jAy: molecular mass flux of A
31
Situation in Column Chromatography
Stat. Phase
t=3
t=4
t=1
t=5
t=2
t=6
Mass transfer takes time => separation + limit on resolution
32
Diffusion: Fick’s 2nd Law
Concentration (y)
– Per unit area
perpendicular to
diffusion
SP
y (position in SP)
• Conservation of mass
for diffusing species in
control volume
y +y
jout
1
y1
jin
C(y1,t)
C A ( y1 , t )
 jin  jout
t
C A ( y1 , t )
C ( y , t )
C ( y  y, t )
y   D A 1  D A 1
y
t
y
33
Diffusion: Fick’s 2nd Law II
Concentration (y)
C A ( y, t )
C A ( y, t )
 DA
y 2
t
SP
y (position in SP)
• When things are
changing in time:
• Once CA is the same
everywhere, we have
reached equilibrium in
the SP
Q: can we estimate order-of-magnitude of time needed?
34
Time Scale of Mass Transfer
Concentration (y)
• For transfer across SP
thickness Y
• Start with:
SP
• “Order-of-magnitude
analysis”
C A
D
• Simplifying:
 DA
y (position in SP)
C A ( y, t )
C A ( y, t )
 DA
t
y 2
C A
Y2
Y2
D 
DA
35
Numerical Example of Mass Transfer
• Assume
– Column diameter = 100 m
– Film thickness: Y = 1 m
– Diffusivity of analyte in stationary phase
DAB= 10-6 cm2 s-1
– Solubility of A in B is 1% of volume
– MWA = 100 g mol-1
• Questions
– What is the time scale of mass transfer?
– What should be the time scale of flow along 1 mm of the
column?
– What is the max. amount of analyte that can be in the
stationary phase per 1 mm of column length?
36
Band Broadening Processes
Stationary Phase
Detector
Stationary Phase
Laser
Mobile Phase
• Non-column broadening
– Dispersion of analyte in
• Dead volume of injector
• Connection between injector & column
• Connection between column & detector
– Emphasis on minimizing dead volume
(injectors, fittings…)
• Column broadening: Van Demteer model
37
Schematic of Column Chromatography III
SP
• If analyte has some affinity to the stationary
phase, it will be retarded
– Equilibrium
– Kinetics
K
Cs
CM
• Molecular mass transfer: diffusion
– Emerge at the detector after “retention time” tR
38
GC vs. HPLC Columns
GC
•
•
•
HPLC
Most GC columns do NOT have particles
Most HPLC columns do have particles
Why?
A.
B.
C.
D.
E.
Particles are needed to prevent liquid flow for being too fast
Particles are needed in HPLC to shorten diffusion distance in MP
Particles are not needed in GC because diffusion is very fast
B&C
I don’t know
39
Effect of Mobile Phase velocity on H
• Experiment
– Repeat the same
separation, same
column and mobile
phase
– Determine H vs. flow
rate
– Observe an optimum, H
increases to both sides
– Goethe: “there is
nothing more practical
than a good theory”
Skoog & Leary, 4th Ed
40
Van Deemter Model: “B” Term
• Longitudinal Diffusion
• Basically molecular diff., as if
mobile phase was not moving
Clicker 1
A. B term  as u 
B. B term as u 
C. B term  f (u )
D. Don’t know
Clicker 2. B term is
A. more imp in GC
B. More imp in HPLC
C. Similar importance
D. I don’t know
41
Van Deemter Model:
“A Term” Broadening
• ‘Eddy diffusion’ &
unequal pathways
• Molecules may travel unequal distances
• Particles (if present) cause eddies & turbulence
• A depends on size of stationary particles (want small) and
their packing (want uniform) (or coating in TLC plate)
– GC: 150 m, HPLC: 5-10 m
A  d P Clicker
A.
B.
C.
D.
A term  as u 
A term as u 
A term  f (u )
Don’t know
42
Van Deemter Model: “C” Terms I
• Accounts for finite time
for mass transfer (equil. btw
analyte in stationary and
mobile phase not instantaneous)
• Most important effect in GC & HPLC
• CS accounts for stationary phase mass transfer
– df: stationary phase film thickness
– DS: diffusion coeff. of analyte in SP
– Thinner films reduce mass transfer time
& broadening
Cs 
d 2f
Ds
• But also reduce capacity of the column
43
Van Deemter Model: “C” Terms II
• CM accounts for mass transfer on the
mobile phase interface with the SP
• In packed columns:
d P2
– dp is particle diameter
• In open columns
– dc is column diameter
Clicker 1
A. C term  as u 
B. C term as u 
C. C term  f (u )
D. Don’t know
CM 
DM
d C2
CM 
DM
Clicker 2. CM term is
A. More important in GC
B. More important in HPLC
C. Similar importance
D. I don’t know
44
Van Deemter Model of Band Broadening
• Tries to explain previous experiment
• H: plate height
• u: average linear velocity
H  A
u
L
tM
B
 CS u  CM u
u
• H: as small as possible (calculate Hmin?)
• Some terms decrease, other increase with u
– There should be optimum u
• There are alternative models (see reading)
45
Optimum Mobile Phase Velocity
• We want N highest, H lowest
• Do we want something else?
46
Optimum Mobile Phase Velocity: GC & HPLC
Q: differences in A, B, C between Supercritical Fluid Chrom. & HPLC?
47