Product Data Sheet 00813-0100-4690, Rev GA Catalog 2005 Rosemount 2088 Rosemount 2088 Absolute and Gage Pressure Transmitter A TRADITION OF EXCELLENCE IN PERFORMANCE FROM THE INDUSTRY LEADERS • Absolute and gage pressure ranges from 0–1.5 psi to 0–4,000 psi (0–0.1 to 0–276 bar) • 0.10% reference accuracy, including linearity, hysteresis, and repeatability • 20:1 turndown • Lightweight, compact size makes for easy installation and handling • 0.075% High accuracy option Content Specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . page Pressure-169 Product Certifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . page Pressure-171 Dimensional Drawings. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . page Pressure-173 Ordering Information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . page Pressure-174 Configuration Data Sheet . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . page Pressure-176 www.rosemount.com http://www.dongatech.co.kr/board/data/cat/2088-306RT.pdf Product Data Sheet 00813-0100-4690, Rev GA Catalog 2005 Rosemount 2088 Rosemount 2088 The 2088 Smart Pressure Transmitter is an economical addition to the pressure transmitter line of instruments. The 2088 is designed with reliability, long-term performance, and maintainability in mind. The 2088 maintain a specification conformance of at least 3(1). The rugged, reliable performance for which Emerson Process Management transmitters are famous, coupled with Smart capabilities, make these transmitters exceptional values. The 2088 is available in either gauge or absolute pressure in pressure ranges to 4,000 psi (275 bar). It utilizes a solid-state, polysilicon pressure sensor with a choice of either 316L or Hastelloy isolating diaphragms. The low oil fill of this design has very little temperature effect and outstanding accuracy. (1) Features The 2088 provides accurate, stable, and reliable pressure measurement in difficult applications. Its small compact design allows it to be directly connected to a process - providing a quick, easy, and cost effective installation. The 2088 standard process connection is 1/2 inch NPT but a variety of optional connections are available. Optional connections range from multiple threaded connections to our full line of manifolds and remote diaphragm seals that provide solutions for virtually any connection. The 2088 also features a fully configurable LCD that displays pressure and diagnostic information. The information displayed is directly from the microprocessor which accounts for its accuracy and reliability. Sigma () is a statistical symbol to designate the standard deviation from the mean value of normal distribution. Rosemount Pressure Solutions Rosemount 3051S Series of Instrumentation Scalable pressure, flow and level measurement solutions improve installation and maintenance practices. Rosemount 3095MV Mass Flow Transmitter Accurately measures differential pressure, static pressure and process temperature to dynamically calculate fully compensated mass flow. Rosemount 305 and 306 Integral Manifolds Factory-assembled, calibrated and seal-tested manifolds reduce on-site installation costs. Rosemount 1199 Diaphragm Seals Provides reliable, remote measurements of process pressure and protects the transmitter from hot, corrosive, or viscous fluids. Annubar Flowmeter Series: Rosemount 3051SFA, 3095MFA, and 485 The state-of-the-art, fifth generation Rosemount 485 Annubar combined with the 3051S or 3095MV MultiVariable transmitter creates an accurate, repeatable and dependable insertion-type flowmeter. Pressure-168 Compact Orifice Flowmeter Series: Rosemount 3051SFC, 3095MFC, and 405 Compact Orifice Flowmeters can be installed between existing flanges, up to a Class 600 (PN100) rating. In tight fit applications, a conditioning orifice plate version is available, requiring only two diameters of straight run upstream. Integral Orifice Flowmeter Series: Rosemount 3051SFP, 3095MFP, and 1195 These integral orifice flowmeters eliminate the inaccuracies that become more pronounced in small orifice line installations. The completely assembled, ready to install flowmeters reduce cost and simplify installation. Orifice Plate Primary Element Systems: Rosemount 1495 and 1595 Orifice Plates, 1496 Flange Unions and 1497 Meter Sections A comprehensive offering of orifice plates, flange unions and meter sections that is easy to specify and order. The 1595 Conditioning Orifice provides superior performance in tight fit applications. Product Data Sheet 00813-0100-4690, Rev GA Catalog 2005 Rosemount 2088 Specifications Functional Specifications Zero can be suppressed between atmosphere for gage transmitters or 0 psia for absolute transmitters and upper range limit, provided the calibrated span is equal to or greater than the minimum span, and the upper range value does not exceed the upper range limit. For vacuum calibrations, the compound range option is available (Option Code CR). Service Liquid, gas, and vapor applications Ranges Range Minimum Span 1 2 3 4 1.5 psi (103 mbar) 7.5 psi (517 mbar) 40 psi (2.8 bar) 200 psi (13.8 bar) Zero Elevation and Suppression URL/Max. span/ Sensor Limit 30 psi (2.1 bar) 150 psi (10.3 bar) 800 psi (55.2 bar) 4000 psi (275.8 bar) Compound Range Capability Select the Compound Range (CR) option for use in negative gage pressure applications. An enhanced sensor is installed in transmitters with the CR option code. Time Response Output Code S: 4–20 mA dc Code N: 1-5 volt dc, low power (Outputs are directly proportional to the input pressure) Time Constant: 200 milliseconds Dead time: < 0.1 s Update rate: 20 times per second minimum Temperature Limits Rangedown Process 20 to 1 Silicone fill sensor: –40 to 250 °F (–40 to 121 °C) Inert fill sensor: –22 to 250 °F (–30 to 121 °C) Load Limitations Reverse polarity protection is standard. Maximum loop resistance is determined by the power supply voltage as described by the following equations: Max. Loop Resistance = 43.5 (Power Supply Voltage – 10.5) Process temperatures above 185 °F (85 °C) require derating the ambient limits by a 1.5:1 ratio: Ambient: –40 to 185 °F (–40 to 85 °C) –4 to 175 °F (–20 to 80 °C) with LCD display Storage: Load (ohms) –50 to 230 °F (–46 to 110 °C) –40 to 185 °F (–40 to 85 °C) with LCD display Operating Region Maximum Ambient Process Temp – 185 ) = 185 – (---------------------------------------------------------Temperature in °F 1.5 2 Power Supply Output Code S (dc Volts) Maximum Ambient ( Process Temp – 85 ) = 85 – -------------------------------------------------------Temperature in °C 1.5 Communication requires a minimum loop resistance of 250 ohms. (1) For hazardous location approvals, power supply must not exceed 36 V. Humidity Limits (2) For CENELEC Ex ia approval, the power supply must not exceed 30 V. 0–100% relative humidity Power Supply External power supply required. Transmitter operates on 10.5–36 V dc with no load (6–12V for Low Power). Reverse polarity protection is standard. Current Draw Output Code N: ≤ 3 mA without LCD display. Overpressure Limits Range 1: 120 psig max All other ranges: two times the URL Volumetric Displacement Less than 0.00042 cm3 Turn-on Time 2.0 seconds, no warm-up required Transmitter Security Activating the transmitter security function prevents changes to the transmitter configuration, including local zero and span adjustments. Security is activated by an internal switch. Pressure-169 Product Data Sheet 00813-0100-4690, Rev GA Catalog 2005 Rosemount 2088 Failure Mode Transient Protection Limits If self-diagnostics detect a sensor or microprocessor failure, the analog signal is driven either high or low to alert the user. High or low failure mode is user-selectable with a jumper on the transmitter. The values to which the transmitter drives its output in failure mode depend on whether it is factory-configured to standard or NAMUR-compliant operation. The values for each are as follows: Standard Operation Output Code Linear Output S 3.9 ≤ I ≤ 20.8 N 0.97 ≤ V ≤ 5.2 N with Code C2 0.78 ≤ V ≤ 3.44 Fail High Fail Low I ≥ 21.75 mA V ≥ 5.4 V V ≥ 4.0 V I ≤ 3.75 mA V ≤ 0.95V V ≤ 0.77 V Standard Operation NAMUR-Compliant Operation Output Code S IEEE 587 Category B 6 kV Crest (1.2 50 s) 3 kV Crest (8 20 s) 6 kV Crest (0.5 s by 100 kHz) IEEE 472 SWC 2.5 kV Crest,1 MHz waveform General Specifications Tested to IEC 801-3 Physical Specifications Electrical Connection Linear Output Fail High 3.8 ≤ I ≤ 20.5 I ≥ 22.5 mA Fail Low I ≤ 3.6 mA Performance Specifications (Zero-based spans, reference conditions, silicone oil fill, and 316L SST isolating diaphragm.) Reference Accuracy 1 /2–14 NPT, M20 1.5 (CM20), PG 13.5, or G 1/2 female (PF 1/2 female) conduit entry Process Connection 1/2–14 NPT female, DIN 16288 G 1/2 male, RC 1/2 female (PT 1/2 female), M20 1.5 (CM20) male Process Wetted Parts Isolating Diaphragm • ±0.10% of calibrated span. Includes combined effects of linearity, hysteresis, and repeatability 316L stainless steel or Hastelloy C-276 • ±0.075% of calibrated span (high accuracy option) Process Connector Ambient Temperature Effect Expressed as a total effect per 50 °F (28 °C) Total effect includes zero and span effects. ± (0.15% URL + 0.15% of span) from –40 °F to 185 °F (-40 °C to 85 °C) 316L stainless steel CF-3M (Cast version of 316L SST, material per ASTM_A743) or Hastelloy C-276 Non-wetted Parts Electronics Housing Stability Low-copper aluminum, NEMA 4X, IP65, IP67,CSA enclosure Type 4X ±0.10% of URL for 12 months Paint Vibration Effect Less than ±0.1% of URL when subjected to vibration of: peak to peak constant displacement of 4 mm (5–15 Hz) and constant acceleration of 2 g (15–150 Hz) and 1 g (150–2000 Hz). Polyurethane Cover O-rings Buna-N Fill Fluid Power Supply Effect Less than 0.01% of calibrated span per volt Mounting Position Effect Zero shift of up to 1.2 inH2O (0.30 kPa), which can be calibrated out. No span effect. RFI Effect Less than ±0.25% of upper range limit from 20–1000 MHz at 30 V/m with leads in conduit. Less than ±0.25% of upper range limit from 20-1000 MHz at 10 V/m with unshielded twisted pair (no conduit). Pressure-170 Silicone or inert fill Weight Output Code S and N: Approximately 2.44 lb (1.11 kg) Tagging The transmitter is tagged, at no charge, in accordance with customer requirements. All tags are stainless steel. The standard tag is permanently attached to the transmitter. Tag character height is 1/8 in. (0.318 cm). A wired tag is available upon request. Accessory Block and Bleed Valve (S5 Option) The Rosemount 306 Integral Manifold is pre-assembled to transmitter and leak checked. Product Data Sheet 00813-0100-4690, Rev GA Catalog 2005 Rosemount 2088 Product Certifications Approved Manufacturing Locations Canadian Standards Association (CSA) Rosemount Inc. — Chanhassen, Minnesota, USA C6 Emerson Process Management GmbH & Co. — Wessling, Germany Emerson Process Management Asia Pacific Private Limited — Singapore Intrinsically Safe for Class I, Division 1, Groups A, B, C, and D. Temp. Code T3C. Intrinsically safe when connected with approved barriers in accordance with Rosemount drawing 02088-1024. For input parameters see control drawing 02088-1024. Beijing Rosemount Far East Instrument Co., LTD — Beijing, China European Union Directive Information The EC declaration of conformity for all applicable European directives for this product can be found at www.rosemount.com. A hard copy may be obtained by contacting our local sales office. European Certifications I1 ATEX Directive (94/9/EC) Emerson Process Management complies with the ATEX Directive. European Pressure Equipment Directive (PED) (97/23/EC) Electro Magnetic Compatibility (EMC) (89/336/EEC) All 2088/2090 Smart Pressure Transmitter: EN 50081-1: 1992; EN 50082-2:1995; EN 61326-1:1997 Ordinary Location Certification for Factory Mutual II 1 G Loop/Power Input Type Ui = 30 V dc Ii = 200 mA Pi = 0.9 W Ci = 0.012 F Smart Smart Smart Smart Special Conditions for Safe Use (x): When the optional transient protection terminal block is installed, the apparatus is not capable of withstanding a 500V rms test to case. This must be taken into account on any installation in which it is used, for example by assuring that the supply to the apparatus is galvanically isolated. As standard, the transmitter has been examined and tested to determine that the design meets basic electrical, mechanical, and fire protection requirements by FM, a nationally recognized testing laboratory (NRTL) as accredited by the Federal Occupational Safety and Health Administration (OSHA). North American Certifications ATEX Intrinsically Safe Certificate No.: BAS00ATEX1166X EEx ia IIC T5 (Tamb = –55 to 40 °C) EEx ia IIC T4 (Tamb = –55 to 70°C) 1180 TABLE 1. Input Parameters 2088/2090 Pressure Transmitters — Sound Engineering Practice Hazardous Locations Certifications Explosion-Proof for Class I, Division 1, Groups B, C, and D. Dust-Ignition-Proof for Class II, Division 1, Groups E, F, G, Class III, indoor and outdoor hazardous locations. CSA enclosure Type 4X; factory sealed. Suitable for Class I, Division 2, Groups A, B, C, and D. N1 ATEX Type n Certification No.: BAS00ATEX3167X EEx nL IIC T5 (Ta = -40 °C to 70 °C) Ui = 50 V dc max II 3 G Factory Mutual (FM) Approvals E5 Explosion-Proof for Class I, Division 1, Groups B, C, and D. Dust-Ignition-Proof for Class II, Division 1, Groups E, F, G, Class III, Division 1, indoor and outdoor (NEMA 4X) hazardous locations; factory sealed. I5 Intrinsically safe for use in Class I, Division 1, Groups A, B, C, D; Class II, Division 1, Groups E, F, and G; and Class III, Division 1 when connected in accordance with Rosemount drawing 02088-1018. Non-incendive for Class I, Division 2, Groups A, B, C, and D. For input parameters see control drawing 02088-1018. Special Conditions for Safe Use (x): When the optional transient protection terminal block is installed, the apparatus is not capable of withstanding a 500 V r.m.s. test to case. This must be taken into account on any installation in which it is used, for example, by assuring that the supply to the apparatus is galvanically isolated. Pressure-171 Product Data Sheet 00813-0100-4690, Rev GA Catalog 2005 Rosemount 2088 ND ATEX Combustible Dust Certificate No.: BAS01ATEX1427X T105°C (Tamb = -20°C to 85°C) IP66 1180 Vmax = 36 V dc Max Ii = 24 mA II 1 D Japanese Certifications E4 Ex d IIC T6 (Tamb = 85 °C) Certificate Description C15879 Special Conditions for Safe Use (x): 1. The user must ensure that the maximum rated voltage and current (36 volts, 24 mA, D.C.) are not exceeded. All connections to other apparatus or associated apparatus shall have control over this voltage and current equivalent to a category “ib” circuit according to EN50020. 2. Cable entries must be used which maintain the ingress protection of the enclosure to at least IP66. C15877 C15876 C15875 C15874 3. Unused cable entries must be filled with suitable blanking plugs which maintain the ingress protection of the enclosure to at least IP66. C15873 C15872 4. Cable entries and blanking plugs must be suitable for the ambient range of the apparatus and capable of withstanding a 7J impact test. 5. The 2088/2090 sensor module must be securely screwed in place to maintain the ingress protection of the enclosure. ED ATEX Flame-Proof Certification No.: KEMA97ATEX2378 EEx d IIC T6 (Ta = -20 °C to 40°C) II 1/2 G T4 (Ta = -20 °C to 80 °C) 1180 Vmax = 36 (with Smart output option) Vmax = 14 (with low power output option) Pressure-172 JIS Flame-Proof C15871 2088 Smart with SST wetted parts (with meter) 2088 Smart with Hast wetted parts (with meter) 2088 Smart with Hast wetted parts (no meter) 2088 Smart with SST wetted parts (no meter) 2088 Smart with Hast wetted parts, CR Option (with meter) 2088 Smart with Hast wetted parts, CR Option (no meter) 2088 Smart with SST wetted parts, CR Option (with meter) 2088 Smart with SST wetted parts, CR Option (no meter) Combinations of Certifications Stainless steel certification tag is provided when optional approval is specified. Once a device labeled with multiple approval types is installed, it should not be reinstalled using any other approval types. Permanently mark the approval label to distinguish it from unused approval types. KB Combination of E5, I5, and C6 KH Combination of E5, I5, and I1 K5 Combination of E5 and I5 K6 Combination of C6, I1, and ED Product Data Sheet 00813-0100-4690, Rev GA Catalog 2005 Rosemount 2088 Dimensional Drawings Rosemount 2088 0.75 (20) Clearance for Cover Removal 5.0 (127) 4.3 (110) Max. 2 ½–14 NPT† Conduit Connection 3.9 (99) Optional Meter Cover 4.7 (120) Transmitter Circuitry Terminal Connections Certifications Tag 5.4 (140) Typical 2 ¼–20 UNC-2B Mounting Holes † M20 1.5 Female (CM20), PG 13.5, and G 1/2 Female (PF 1/2 Female) also available as options. ½–14 NPT Female Process Connection (See Ordering Table for other options.) Mounting Configurations with Optional Mounting Brackets Pipe Mounting 5.0 (127) 4.3 (110) 4.0 (102) 2.8 (71) 3.8 (97) 3.1 (78) 6.3 (159) 2-inch U-Bolt for pipe mounting (clamp shown) 4.8 (122) 3.0 (80) 7.0 (178) 2.8 (71) 1/4× 11/4 Bolts for transmitter mounting (not supplied) 1.3 (33) 6.2 (157) 2088S-2088A04A, B04A, C04A; 2088-2088A04A, A04B 3.9 (99) 5.3 (135) Panel Mounting 5/16× 1½ Bolts for panel mounting (not supplied) 1/4× 11/4 Bolts for transmitter mounting (not supplied) 1.3 (33) Dimensions are in inches (millimeters). Pressure-173 Product Data Sheet 00813-0100-4690, Rev GA Catalog 2005 Rosemount 2088 Ordering Information Model 2088 Code A G Code 1 2 3 4 Code S N Product description Pressure Transmitter Transmitter type Absolute Gage Range Minimum span URL/Max.span/Sensor limit 0–30 psi (0–2.1 bar) 0–150 psi (0–10.3 bar) 0–800 psi (0–55.2 bar) 0–4,000 psi (0–275.8 bar) 1.5 psi (103.0 mbar) 7.5 psi (517.0 mbar) 40 psi (2.8 bar) 200 psi (13.8 bar) 30 psi (2.1 bar) 150 psi (10.3 bar) 800 psi (55.2 bar) 4000 psi (275.8 bar) Output 4–20 mA dc/Digital HART® Protocol 1-5 V dc Low Power/ Digital HART protocol Materials of Construction Code Process connection Isolating diaphragm Oil fill 22(1) 33(1) 2B(1) 316L SST Hastelloy C-276 316L SST 316L SST Hastelloy C-276 316L SST Silicone Silicone Inert Code A B C(2) D(2) Code 1 2 3 4 Code Process connection ½–14 NPT Female DIN 16288 G ½ Male RC ½ Female (PT ½ Female) M20 1.5 Male (CM20 Male) Conduit thread ½–14 NPT M20 1.5 Female (CM20) PG 13.5 G ½ Female (PF ½ Female) Options Integral manifold S5 Assembly to Rosemount 306 integral manifold (Use 1/2 - 14 NPT Female Process Connection code A) Diaphragm seal assemblies S1 Attachment of one diaphragm seal Mounting brackets B4 SST mounting bracket with SST Bolts Special configuration (software) C4(4) NAMUR alarm and saturation levels, high alarm CN(4) NAMUR alarm and saturation levels, low alarm C9(3) Software configuration Special configuration (hardware) C2 0.8 - 3.2 V dc output with HART protocol, Output code N only. CR Compound range calibration capability Pressure-174 Product Data Sheet 00813-0100-4690, Rev GA Catalog 2005 Rosemount 2088 Product Certifications I1 (4) ATEX Intrinsically Safe N1(4) ATEX Type n ND(4) ATEX Combustible Dust ED ATEX Flame-Proof C6 CSA Explosion-Proof, Intrinsically Safe, and non-Incendive K6(4) CSA and ATEX Explosion-proof and Intrinsically Safe (combination of C6, I1, and ED) E4(4) JIS Flameproof (Available with Conduit Thread code 4 and with or without Digital Display codes M5 and M7) E5 FM Approvals explosion-proof I5 FM Approvals Intrinsically safe, non-incendive K5 FM Approvals Explosion-proof, Intrinsically Safe, non-incendive (combination of E5 and I5) KB FM Approvals and CSA Explosion-proof, Intrinsically Safe, non-incendive (combination of E5, I5, and C6) KH(4) FM Approvals and ATEX Explosion-Proof and Intrinsically Safe (combination of E5, I5, and I1) DW NSF drinking water approval (Requires Materials of Construction code 22 with Process Connection code A.) Digital Display M5 LCD display, scaled 0–100% M7 LCD display, special configuration Special procedures P1 Hydrostatic testing P2 Cleaning for special service P8(5) 0.075% accuracy to 10:1 turndown Special certifications Q4 Calibration certificate Q16 Surface finish certification for sanitary remote seals Terminal blocks T1 Transient protection (Available with Option codes E5, ED, I1, I5, N1, C6, and K5. (1) Materials of Construction comply with recommendations per NACE MR0175/ISO 15156 for sour oil field production environments. Environmental limits apply to certain materials. Consult latest standard for details. Selected materials also conform to NACE MR0103 for sour refining environments. (2) Not available with Hastelloy C-276, Materials of Construction code 33. (3) A Configuration Data Sheet, see page Pressure-176, must be completed. (4) Not available with low-power Output code N. (5) Available with Output code S, stainless steel isolators, silicone fill, and for spans of 10 inH2O and greater. 00806-0100-4690 English Rev. AA Pressure-175 Product Data Sheet 00813-0100-4733, Rev HA Catalog 2005 Rosemount 305 and 306 Manifolds Rosemount 305 and 306 Integral Manifolds ROSEMOUNT 305 AND 306 FEATURE... • Unique Coplanar™ design of the Rosemount 3051 and 3095 families allows “flangeless” valve integration • Coplanar, traditional, and inline styles • Compact, lightweight assembly • Factory assembled, seal-tested and calibrated • Easy in-process calibration • 50% fewer possible leak points than conventional manifold/transmitter assemblies • Direct-mount capability Content Rosemount 305R Coplanar Style Integral Manifolds. . . . . . . . . . . . . . . page Pressure-236 Specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . page Pressure-237 Dimensional Drawings. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . page Pressure-238 Ordering Information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . page Pressure-244 www.rosemount.com Product Data Sheet 00813-0100-4733, Rev HA Catalog 2005 Rosemount 305 and 306 Manifolds Rosemount 305R Coplanar Style Integral Manifolds 305R TWO-VALVE ROSEMOUNT 305R FIVE-VALVE This two-valve manifold is used with Rosemount 3051 gage and absolute pressure transmitters. The first valve provides instrument isolation. The second valve allows venting, draining, or calibration through the test port. This five-valve manifold is used with 3095 and 3051 differential pressure transmitters. It provides two blocking valves, two test/vent valves, and one equalizing valve. The two vent valves allow for 100% capture of vented or drained process, and simplified in-process calibration capability. We also offer a five-valve integral manifold with a metering pattern for Natural Gas installations. • 305RC2 • 305RT2 • 305RM2 • 305RC7 Coplanar Transmitter • 305RC5 • 305RC6 • 305RC9 • 305RM5 Coplanar Transmitter Test/Vent Test/Vent Test/Vent Isolate Equalize Process Isolate Isolate Process 305R THREE-VALVE This three-valve manifold is used with Rosemount 3095 and 3051 differential pressure transmitters. It provides two blocking valves and one equalizing valve. Two drain/vent valves are also installed at the test ports. • 305RC3 • 305RT3 • 305RM3 • 305RC8 Test/Vent Equalize Equalize Isolate Equalize Isolate Process Isolate Test/Vent Process Drain/ Vent Valve Isolate Process Pressure-236 Coplanar Transmitter Test/Vent Coplanar Transmitter Drain/ Vent Valve FIVE-VALVE NATURAL GAS NOTE Test/Vent connections receive plastic caps to protect threaded connections. Product Data Sheet 00813-0100-4733, Rev HA Catalog 2005 Rosemount 305 and 306 Manifolds 306RT TWO-VALVE MANIFOLDS 306RT BLOCK-AND-BLEED MANIFOLDS This two-valve pressure manifold is used with 3051 and 2088 gage and absolute pressure transmitters. The first valve provides instrument isolation. The second valve allows venting, draining, or calibration through the test port. Available in 1/2–14 NPT male or female process connections. This pressure manifold is used with 3051 and 2088 gage and absolute pressure transmitters. It provides a single block valve for instrument isolation. There is also a plug for drain/vent capabilities. Available in 1 /2–14 NPT male process connection. • • 306RT1 306RT2 Pressure Transmitter Pressure Transmitter Bleed Screw Test/Vent Isolate Isolate Process Process Specifications Test Connections Shipping Weights 1 /4-18 NPT TABLE 2. 305 Manifold Weights Without Options (lb/kg) Adapters DF option, CF-8M (Cast version of 316 SST, material per ASTM-A743) Bolts for Manifolds Standard material is plated carbon steel per ASTM A449, Type 1 Alternative bolt materials offered through Option Codes • L4 Austenitic 316 Stainless Steel Bolts • L5 ASTM-A-193-B7M bolts TABLE 1. Pressure Ratings(1) 305 Teflon® Graphite-based 306 Teflon Graphite-based Pressure 6092 psi (420 bar)(2) 6092 psi (420 bar)(2) Pressure Max Manifold Weight Manifold with 3051C/ 3051P Manifold with 3095 0305_C2 0305_C3 0305_C5 0305_C6 0305_C7 0305_C8 0305_C9 0305_T2 0305_T3 0305_T7 0305_T8 4.5 (2,0) 4.7 (2,1) 6.5 (3,0) 6.4 (2,9) 4.7 (2,1) 5.0 (2,3) 6.3(2,85) 6.0 (2,7) 6.0 (2,7) 6.2 (2,8) 6.2 (2,8) 9.4 (4,3) 9.6 (4,4) 11.4 (5,17) 11.3 (5,1) 9.6 (4,4) 9.9 (4,5) 11.2 (5,1) 10.9 (4,9) 10.9 (4,9) 11.1 (5,0) 11.1 (5,0) 9.9 (4,5) 10.1 (4,6) 11.9 (5,4) 11.8 (5,35) 10.1 (4,6) 10.4 (4,8) 11.7 (5,3) – – – – TABLE 3. 306 Manifold Weights Without Options (lb/kg) (3) 10000 psi (689 bar) 6000 psi (689 bar)(2) (1) Except Option HK: 2320 psig (160 bar) at 200 °F (93 °C) (2) At 200 °F (93 °C) Model Model Max Manifold Weight Manifold with 3051T Manifold with 2088 0306_T1 0306_T2 0306_T3 1.1 (0,5) 2.5 (1,1) 2.5 (1,1) 4.1 (1,9) 5.5 (2,5) 5.5 (2,5) 3.1 (1,4) 4.5 (2,0) 4.5 (2,0) (3) At 85 °F (29 °C) Pressure-237 Product Data Sheet 00813-0100-4733, Rev HA Catalog 2005 Rosemount 305 and 306 Manifolds Dimensional Drawings Rosemount 305R Two-Valve Coplanar Style Manifold 5.21 (132) 4.20 (107) 4.55 (116) 7.70 (196) 4.30 (108) Max Open 7.2 (183) Max Open 1 /2-14 NPT on Manifold for Process Connection /4-18 NPT for test/vent connection. 1 3051-3031E06Q 9.00 (229) Dimensions are in inches (millimeters) Rosemount 305R Three-Valve Coplanar Style Manifolds 5.21 (132) 4.20 (107) 4.55 (116) 7.70 (196) 9.00 (229) 5.50 (140) Max Open 10.60 (270) Max Open 1 /2-14 NPT on Manifold for Process Connection Dimensions are in inches (millimeters) Pressure-238 3051-3031E06Q Drain/Vent Valve Product Data Sheet 00813-0100-4733, Rev HA Catalog 2005 Rosemount 305 and 306 Manifolds Rosemount 305R Five-Valve Coplanar Style Manifold 5.21 (132) 4.20 (107) 4.55 (116) 7.70 (196) 5.50 (140) Max Open 3051-3031E06Q 9.00 (229) 11.50 (292) Max Open 1 /2-14 NPT on Manifold for Process Connection /4-18 NPT for test/vent connection. 1 Dimensions are in inches (millimeters) Rosemount 305RT Two-Valve Traditional Style Manifold 5.21 (132) 4.20 (107) 4.55 (116) 7.70 (196) 1/2-14 NPT on optional Process Adapter* 1.63 (41) 3.75 (95) Max Open 3.42 (87) 1 /4-18 NPT on traditional manifold for process connection without the use of a process adapter 1.10 (28) 2.125 (54) 6.40 (163) Max Open Dimensions are in inches (millimeters) Pressure-239 Product Data Sheet 00813-0100-4733, Rev HA Catalog 2005 Rosemount 305 and 306 Manifolds Rosemount 305RT Three-Valve Traditional Style Manifold 5.21 (132) 4.55 (116) 4.20 (107) 7.70 (196) 1 /2-14 NPT on optional Process Adapter* Drain/Vent Valve 1.63 (41) 3.75 (95) Max Open 1.05 (27) 3.5 (87) 1 /4-18 NPT on traditional manifold for process connection without the use of process adapters 1.10 (28) * Adapters can be rotated to give adapter connection centers of 2.0 (51), 2.125 (54), or 2.25 (57). 2.125 (54) 6.20 (158) Max Open 2.70 (69) Max 2.70 (69) Open Max 8.90 (226) Max Open Dimensions are in inches (millimeters) Rosemount 305RM Two-Valve Traditional Style Manifold 5.21 (132) 4.20 (107) 4.55 (116) 7.70 (196) 1 /2-14 NPT on optional Process Adapter* 1.63 (41) 3.75 (95) Max Open 1 /4-18 NPT on traditional manifold for process connection without the use of a process adapter 3.42 (87) 1.10 (28) 2.125 (54) 6.5 (165) Max Open Dimensions are in inches (millimeters) Pressure-240 1 /4-18 NPT Vent Connection Product Data Sheet 00813-0100-4733, Rev HA Catalog 2005 Rosemount 305 and 306 Manifolds Rosemount 305RM Three-Valve Traditional Style Manifold 5.21 (132) 4.55 (116) 4.20 (107) 0.75 (19) Clearance for Cover Removal 7.70 (196) 1 /2-14 NPT on optional Process Adapter* 1.63 (41) 3.75 (95) Max Open 1 /4-18 NPT on traditional manifold for process connection without the use of process adapters 3.5 (87) 1.05 (27) 1.10 (28) * Adapters can be rotated to give adapter connection centers of 2.0 (51), 2.125 (54), or 2.25 (57). 2.125 (54) 6.5 (165) Max Open 2.5 (63) Max Open 3051-305-3031F30B; 3031RM1 Drain/Vent Valve 9.0 (229) Max Open Dimensions are in inches (millimeters) Rosemount 305RM Five-Valve Traditional Style Manifold 5.21 (132) 4.20 (107) 4.55 (116) 7.70 (196) 1/2-14 NPT on optional Process Adapter* 1.63 (41) 1/4-18 NPT on traditional manifold for process connection without the use of process adapters 4.2 (107) Max Open 2.6 (66) Max Open 4.1 (104) 1.10 (28) * Adapters can be rotated to give adapter connection centers of 2.0 (51), 2.125 (54), or 2.25 (57). 2.125 (54) 6.5 (165) Max Open 2.5 (63) Max Open 9.0 (229) Max Open Dimensions are in inches (millimeters) Pressure-241 Product Data Sheet 00813-0100-4733, Rev HA Catalog 2005 Rosemount 305 and 306 Manifolds Rosemount 306R Two-Valve Pressure Style Manifold (3051T Shown) 4.55 (116) 4.20 (107) 8.00 (203) 1 /4-18 NPT on traditional vent connection (pipe plug shown) 4.85 (123) 4.10 (105) 1/2-14 NPT Male NPT Process Connection (code AA) 1/2-14 NPT Female NPT Process Connection (code BA) 6.25 (159) Max Open Dimensions are in inches (millimeters) Installations for Rosemount 3051T and 3051S_T Transmitters for 2-in. Pipe Mounting ROSEMOUNT 3051T 6.25 (159) ROSEMOUNT 3051S_T 3.13 (79) NOTE: Dimensions are in inches (millimeters). Dimensions are in inches (millimeters) Pressure-242 6.25 (159) 3.13 (79) Product Data Sheet 00813-0100-4733, Rev HA Catalog 2005 Rosemount 305 and 306 Manifolds Coplanar Manifold with Optional Bracket for 2-in. Pipe Mounting OPTION CODE B4 7.68 (195) 3051-305-3031E04C; F04B; 3051-3031J04A 2-inch U-Bolt for Pipe Mounting 4.9 (125) 3.54 (90) 6.25 (159) Dimensions are in inches (millimeters) Traditional Manifold with Optional Brackets for 2-in. Pipe Mounting 8.18 (208) 3.56 (90) 1.1 (28) B3/B9/BC MOUNTING BRACKET 3.42 3.42(86.9) (87) 2.62 (66) 7.7 (196) 1.94 (49.2) 13.03 (331) 0.93 (24) B1/B7/BA MOUNTING BRACKET 3.56 (90) 5.32 (135) 3.42 (87) 1.1 (28) 4.85 (123) Dimensions are in inches (millimeters) Pressure-243 Product Data Sheet Rosemount 305 and 306 Manifolds 00813-0100-4733, Rev HA Catalog 2005 Ordering Information TABLE 4. Rosemount 305R Integral Manifolds Model 0305 Code R Code C T M Code 2 3 5(1) 6(2) 7(2)(3) 8(2)(3) 9(2)(3) Product Description Integral Manifold Manufacturer Rosemount Inc. Manifold Style Coplanar Traditional Traditional (Rosemount 3095-compatible; DIN-compliant flange) Manifold Type 2-valve 3-valve 5-valve 5-valve for natural gas/metering pattern 2-valve (per ASME B31.1[ANSI] power and piping code) 3-valve (per ASME B31.1[ANSI] power and piping code) 5-valve (per ASME B31.1[ANSI] power and piping code) Materials of Construction Code Body Bonnet Stem Drain/Vent 2 316 SST 316 SST 316 SST 316 SST 3(4)(5) Hastelloy® C Hastelloy C Hastelloy C Hastelloy C (4) 4 Monel® Monel Monel Monel Code Process Connection A ¼–18 NPT (Traditional manifold styles T and M) 1 B /2–14 NPT (Coplanar manifold style only) Code Packing Material 1 Teflon 2(6) Graphite-based Code Valve Seat 1 Integral 5 Soft delrin (only available with natural gas/ metering pattern) Code 305R Options P2 Cleaning for special services (Not available with graphite-based packing) SG(5)(7) Materials of construction to meet NACE material recommendations per MR 01-75. (Not available with option code P2 L4(8) Austenitic 316 SST bolts L5 ASTM-A-193-B7M bolts Coplanar Options B4 SST bracket for 2-in. pipe mount with series 300 SST bolts Traditional Options B1 Bracket for 2-in. pipe mounting, CS bolts B3(9) Flat bracket for 2-in. pipe mounting, CS bolts B7 B1 bracket with series 300 SST bolts B9(9) B3 bracket with series 300 SST bolts BA SST B1 bracket with series 300 SST bolts BC(9) SST B3 bracket with series 300 SST bolts 1 DF /2–14 NPT flange adapters, SST (Not available with graphite-based packing or HK, HL options) HK(10) 10mm (M10) process flange bolting connection HL(10) 12mm (M12) process flange bolting connection Typical Coplanar Integral Manifold Model Number: 305RC32B11B4 Typical Transmitter Model Number: 3051CD2A02A1AS5 (1) Not available with traditional manifold style T. (2) Only available with Coplanar manifold style (3) Only available with 316SST materials of construction and graphite-based packing. (4) Not available with traditional manifold Style M. (5) Materials of Construction meet NACE material recommendation per MR 01-75. Environmental limits apply to certain materials. Consult standard for details. (6) Includes graphite tape on drains/vents and plugs. (7) Only available with Materials of Construction Code 2: 316 SST body, bonnets, and stems; Hastelloy drain/vents. (8) Not available with manifold codes 7, 8, and 9. (9) Not compatible with the Rosemount 3095 transmitter. (10) Only available with traditional manifold style M. Pressure-244 Product Data Sheet 00813-0100-4733, Rev HA Catalog 2005 Rosemount 305 and 306 Manifolds TABLE 5. Rosemount 306RT Integral Manifolds Model Product Description 0306 Code Pressure Manifold Manufacturer R Code Rosemount Inc. Manifold Style T Code Threaded Manifold Type 1 2 3(1) Block-and-bleed 2-valve 2-valve (per ASME B31.1 [ANSI] power and piping code) Materials of Construction Code Body 2 3(2)(3) Code 316 SST 316 SST Hastelloy C Hastelloy C Process Connection Bonnet AA BA(2) Code 1/2–14 1 2(4) Code Teflon Graphite-based Valve Seat 1 Code Integral 306RT Options Stem Drain/Vent Plug 316 SST Hastelloy C 316 SST Hastelloy C male NPT /2–14 female NPT Packing Material 1 P2 Cleaning for special services (Not available with graphite-based packing) SG(3) Materials of construction to meet NACE material recommendations per MR 01-75 (Not available with option P2) Typical Integral Manifold Model Number: 3 0 6 R T 2 2 B A 1 1 Typical Transmitter Model Number: 3051TG3A2B21AS5B4 (1) Only available with 316SST materials of construction and graphite-based packing. (2) Not available with block-and-bleed manifold type (3) Materials of Construction meet NACE material recommendation per MR 01-75. Environmental limits apply to certain materials. Consult standard for details. (4) Includes graphite tape on plugs. Pressure-245 Product Data Sheet 00813-0100-4733, Rev HA Catalog 2005 Rosemount 305 and 306 Manifolds OPTIONS ASME B31.1 (ANSI) Module Guard The 305 and 306 Manifolds are available in configurations that meet the requirements of ASME B31.1(ANSI) Power and Piping Code. This code specifies design criteria for most air, gas, steam, water, and oil systems used in electric generating systems, central and district heating systems, industrial power plants and geothermal plants. ASME B31.1(ANSI) includes requirements for manifolds, valves, and piping. Transmitters and other measuring devices do not fall within the scope of this code. A sensor module guard is available to protect the transmitter process isolating diaphragms. This guard should be used whenever the transmitter is removed from the integral manifold to avoid damage to the isolating diaphragms. • Part number: 00305-1000-0001 (5/pack) 3051-3031C07F Tagging P2 Cleaning for Special Services Manifolds are tagged with a part number, schematic drawing, temperature and pressure limits. Other Publications For information regarding additional vendors see www.rosemount.com. This option minimizes process contaminants by cleaning wetted surfaces with a suitable detergent. SG Sour Gas Valves can be manufactured in accordance with current NACE MR-01-75 standards. Rosemount and the Rosemount logotype are registered trademarks of Rosemount Inc. Coplanar is a trademark of Rosemount Inc. Monel is a registered trademark of International Nickel Co. Hastelloy is a registered trademark of Haynes International Teflon is a registered trademark of E.I. du Pont de Nemours & Co. Emerson Process Management Rosemount Inc. Emerson Process Management 8200 Market Boulevard GmbH & Co. Chanhassen, MN 55317 USA Argelsrieder Feld 3 T (U.S.) 1 800 999 9307 82234 Wessling T (International) (952) 906 8888 Germany F (952) 949 7001 T 49 (8153) 9390 F 49 (8153) 939172 www.rosemount.com ¢00813-0100-4733w¤ © 2004 Rosemount Inc. All rights reserved. Emerson Process Management Asia Pacific Private Limited 1 Pandan Crescent Singapore 128461 T (65) 6777 8211 F (65) 6777 0947 AP.RMT-Specialist@emersonprocess.com