Study in Batch Distillation using a Bubble Plate Column Particular safety issues: (1) Read the procedure below and then come back and re-read this paragraph. (2) Make sure that the chiller (for the condenser) is activated well before at least 20 minutes) the heating element for the still is switched on. (3) At the end of the experiment, the chiller must be allowed to run after the heating element is deactivated – until the still cools down enough that vapor is not leaving the top of the column. (4) Follow the instructor or assistants instructions about the distillate you’ve collected. It should either be poured into the appropriate waste vessel or returned to the still. (5) The heating element, column, and still will become very hot over the course of the experiment. Be careful not to touch them. 1. Calibrate the refractometer so you can identify compositions of samples. This is done by making samples of known composition by mixing known volumes of ethanol and water. Measure refractive index of each sample. Use a small syringe to extract a few drops needed for the refractometer. Immediately close the sample container to prevent vapor loss. Enter these measured refractive indices into Excel. Since the volume composition is known, you can calculate the mole fraction of each sample from the known densities. Show how you did this calculation in your report. Graph refractive index against mole fraction of ethanol and fit the data so you can relate refractive index to composition. Compare your values against those from the literature (see for example the Handbook of Chemistry and Physics). Note any differences and discuss that in your report. How can you improve the accuracy of your data? 2. Add sufficient amount of a 30 % (by volume) of ethanol-water mixture to the reboiler pot to bring the content of the reboiler pot to the 1000 ml mark. If the mixture in the pot does not look clear, use a fresh batch of solution. Take about 1 ml of the mixture in the pot for analysis. Start the heater at about 80% power and coolant circulation. (Note the full power rating of the heating mantle.) 3. Run the column at total reflux. Let system come to steady state. This can be verified by taking temperature data at the top and bottom of the column at 2 min intervals (It is suggested that you note the temperature to two decimal places even though the accuracy of the temperature sensor is only .1 deg C). When the changes in temperature are less than 1%/ min then steady state may be assumed. Please note the temperature data in your lab notebook. Take a sample from of the pot (1ml or less) and a sample from the top of the column (Make sure to flush the exit port of any liquid left over from the past). Set aside in a closed container to prevent evaporation losses. Measure the temperature at top and bottom. Repeat this again after 10 minutes, just to make sure that the composition has not changed. Note the pressure as well. Chemical and Biomedical Engineering Department USF © page 1 of 2 4. Determine the boil up rate in the column by shutting off the reflux and letting condensate accumulate at the top. Note the volume accumulated at 1 min intervals. Stop after 15 minutes or after the accumulator is half full whichever comes first. 5. Slowly release all the liquid in the accumulator back into the column. If you do it too fast, you can damage the equipment. Let system equilibrate again. 6. Now set the reflux ratio at about 50% by opening both valves approximately the same. You should get a distillate rate that is approximately half of the boilup rate measured above. Start collecting the distillate in a graduated cylinder. Note the amount accumulated as a function of time by taking readings every 5 minutes. Also note the bottom and top temperatures every minute. Stop after you have observed a measurable change in the top temperature. You should have collected about 100-150 ml of distillate. At this point you may proceed to turn off the power and shut down the column. 7. Use the refractometer to identify the compositions of the samples taken in step 2 after they have cooled to room temperature (why?). The inner diameter of the column is 3.0 cm. Clean up Clean the refractometer and switch it off. Shut of the power to the mantle. After the still has cooled, shut off the cooling system. Clean all glass ware with tap water first. Rinse once with distilled water and keep in the dishwasher for drying. 1. g at total reflux? Why or why not? Explain. Chemical and Biomedical Engineering Department USF © page 2 of 2