→→ HiQ® Gas Chromatography. Electron Capture Detector. Gas chromatography Information about gas chromatography in general can be found in the application sheet “Gas Chromatography” (GC). Environmental measurements of deposits, water and air are becoming more and more important in all parts of the world. One group of environmental threats are polyhalogenated organic compounds such as polychlorinated biphenyls (PCB’s), pesticides and other halogenated organics. Even small concentrations are causing damages and it is of great importance to be able to analyze those compounds correctly at low concentrations (ppm and ppb). GC with Electron Capture Detector (ECD) is a very sensitive method and is well suited for analysis of such polyhalogenated organic compounds. Analyze with the GC-ECD In the electron capture detector a beta emitter such as radioactive tritium or 63Ni is used to ionize the carrier gas. Fast beta particles generated by the radioactive source collide with the molecules of the carrier or make-up gas. By impact ionization,, free slow-moving electrons are produced which generate a measurable and steady current. If the GC effl uent contains organic molecules with electronegative functional groups, such as halogens, phosphorous and nitro groups, electrons will be captured and the current will be reduced. In comparison to a signal without sample compounds, the reduction in electron fl ow is proportional to the quantity of electrophile sample components. Gases An important facet of GC-ECD is the carrier gas. The carrier gas transfers the sample from the injector, through the column and into the EC-detector. As for all GC techniques the carrier gas must be inert and may not be adsorbed by the column material. Because the ECD is sensitive to water, the carrier gas must be dry. Besides, the halocarbon content must be as small as possible, since these are the typical compounds to be analyzed with the ECD. Halocarbon-free helium or nitrogen are therefore recommended as carrier gases for GC-ECD. To generate free slow-moving electrons the ECD requires nitrogen or methane, where methane is used in a form of a methane/argon mixture. Both nitrogen and the methane mixture are used as detector gases as well as carrier gases. Like all chromatographic analytical processes, gas chromatography is a relative method, i.e. calibration with a standard mixture is required. In a fi rst analysis a certifi ed standard mixture is measured. By comparable measurements, the sample components can be identifi ed and their proportion, and thereby their concentration, determined. The selection of gases, fi ttings and pipes as well as the installation is crucial to preserve the sensitivity, detection limits and reproducibility of a detector. Particular attention should be given to the choices. Gas outlet Signal Cathode Anode 63 Ni plated surface Detector gas Gas inlet from column HiQ® product program The HiQ® product program offers a wide range of gas qualities and equipment that fulfi ll the demands concerning analytical techniques of AAS. Carrier and detector gases Carrier gas Specifications Detector gas 10% n.a. n.a. Calibration mixtures Recommended central gas supply Recommended cylinder regulators To obtain optimal analytical results, AGA recommends the following gas qualities for GC-ECD analysis: Halocarbon free helium 5.0 or product code 6008 Halocarbon free nitrogen 5.5 product code 6014 O2 N2 Halocarbons as SF6 CnHm H2O Product code Halocarbon free helium He 5.0 ≤ 2 ppm ≤ 5 ppm ≤ 1 ppb ≤ 1 ppm ≤ 3 ppm 6008 Halocarbon free nitrogen 5.5 product code 6014 or Halocarbon free methane instarg 5% Component concentration 5% CH4 Balance gas Argon Halocarbons as SF6 ≤ 1 ppb Product code 6172 Halocarbon free nitrogen N2 5.5 ≤ 5 ppm ≤ 5 ppm ≤ 5 ppm ≤ 5 ppm 6014 Halocarbon free methane instarg 10% 10% CH4 Argon ≤ 1 ppb 6173 For calibration mixtures please look into the HiQ® product catalog or ask your local sales representative. HiQ® REDLINE central gas supply systems for inert and non-reactive gases. Group green for single gas supply panels designed for pure gases and mixtures. Group blue for single stage supply panels with internal purging designed for high purity gases and mixtures including flammable gases. HiQ® REDLINE single stage regulator, C200/1 for carrier and auxiliary gases. For calibration gases HiQ® REDLINE two stage regulator, C200/2 provides a stable secondary outlet pressure. C200 regulators can be plain or equipped with a shut-off valve (type A) or a needle valve (type B). For GC-ECD we recommend a C200 regulator in brass with a shut-off valve. Outlet pressure HiQ® REDLINE bar psi Product code Single stage C200/1 A, brass 0.2-3 3-45 3100 Single stage C200/1 A, brass 0.5-6 8-85 5467 Two stage C200/2 A, brass 0.2-3 3-45 5482 6 6 6 5% 5% 5% 5% 5% 5% Please look into our HiQ® catalog ‘Biotech, Chemical, Petrochemical & Pharma-ceutical’, look into our web site, http://hiq.aga.com, or contact your local AGA sales representative. Blending tolerance Analysis uncertainty Shelf life Sweden, AGA Gas AB, Phone +46 (0)8 706 95 00, Fax +46 (0)8 628 23 15, www.aga.se Finland, Oy AGA Ab, Phone +358 (0) 10 2421, Fax +358 (0) 10 242 0311, www.aga.fi Norway, AGA AS, Phone +47 23 17 72 00, Fax +47 22 02 78 04, www.aga.no Denmark, AGA A/S, Phone +45 32 83 66 00, Fax +45 32 83 66 01, www.aga.dk Iceland, ISAGA ehf., Phone +354 577 3000, Fax +354 577 3001, www.aga.is Estonia, AS Eesti AGA, Phone +372 6504 500, Fax +372 6504 501, www.aga.ee Latvia, AGA SIA, Phone +371 7023900, Fax +371 7023901, www.aga.lv Lithuania, AGA UAB, Phone +370 5 2787788, Fax +370 5 2701191, www.aga.lt ® HiQ and HiQ REDLINE are registered trademark of the Linde Group. More information