The Business Case for Public Funding of Remote Monitoring of Cardiac Implantable Electronic Devices Medical Technology Association of Australia Limited Level 12, 54 Miller Street North Sydney NSW 2066 Australia P: (02) 9900 0650 E: reception@mtaa.org.au www.mtaa.org.au Copyright © 2011 Medical Technology Association of Australia Limited (MTAA) To the extent permitted by law, all rights are reserved and no part of this publication covered by copyright may be reproduced or copied in any form or by any means except with the written permission of MTAA Limited. Contents Summary of the business case ................................................................................ 1 1. Overview .......................................................................................................5 2. About the medical technology industry .......................................................... 2 3. Background ...................................................................................................3 4. Patient Population and Clinical Need ............................................................. 3 5. Remote Monitoring for CIEDs ........................................................................ 3 6. Clinical Benefits of Remote Patient Monitoring .............................................. 5 7. Current Reimbursement for Remote Monitoring of CIED Patients.................. 6 8. Telehealth on the Medicare Benefit Schedule ................................................ 6 9. Economic Impact of Remote Monitoring ........................................................ 7 10. Cost Effectiveness of Remote Monitoring of Patients with CIEDs .................. 8 11. Current Cost to Medicare for CIED In-Office Follow-Up Checks .................... 9 12. Proposed Models for Public Funding of Remote Monitoring of CIEDs ......... 10 13. Conclusion ................................................................................................... 15 Appendix A .............................................................................................................. 16 Appendix B .............................................................................................................. 23 Appendix C .............................................................................................................. 24 Summary of the business case Remote monitoring of cardiac implantable electronic devices (CIED) is a health equity issue. It is also an issue about the future sustainability of the healthcare system. Currently rural and remote patients do not have equal access to cardiac monitoring. Further, remote monitoring contributes to judicious use of clinician time and resource. Remote monitoring of CIED patients: • Maintains patient safety and improves patient care • Eliminates the need for unnecessary in-office follow up visits • Involves no setup costs for physicians • Increases productivity • Addresses the current and future skilled workforce shortage • Offers potential cost savings to health care budget • Contributes to a culture of sustainability in health care provision. Remote monitoring is already endorsed by the Cardiac Society of Australia/New Zealand (also by the US Heart Rhythm Society and European Society of Cardiology). In Australia the only tangible barrier to the technology is lack of reimbursement. Two large clinical trials (Trust & Comps) showed that 70% to 90% of follow up checks are not actionable and in hindsight could have been skipped (wasted healthcare infrastructure, wasted physician, nurse and patient time, lost productivity of patients and carers, emissions from healthcare related travel). Demand and costs for CIED follow up will increase exponentially in the future due to wide-ranging device efficacy and population aging. Clinicians will support a new patient care model involving remote monitoring as long as the reimbursement covers their costs for providing the service. The objectives of MTAA are to: 1. Improve patient care, quality of life and access to modern technology 2. Provide a fair and adequate reimbursement to clinicians for this service 3. Achieve a cost saving to the Commonwealth budget by introducing efficiencyenhancing technology. Three options to achieve these outcomes are provided: 1. MBS item numbers based on existing items with an included loading for remote monitoring 2. Capitated costing model 3. Improved capitated model (flat fee per quarter). Option one, whilst simple, may result in over-servicing of patients. Capitated models, on the other hand, ensure there is no over servicing as physicians are paid for a service regardless of the number of data transmissions or reviews. Option 2, the capitated costing model, is budget neutral proposing an annual fee based on current utilisation. The flat fee per quarter model, option 3, proposes significant long term cost savings to the Commonwealth budget. Medical Technology Association of Australia Business case for public funding of remote monitoring of CIEDs Page 1 1. Overview The Medical Technology Association of Australia (MTAA) presents this business case to government for consideration for funding to address a healthcare equity and healthcare sustainability issue. The Federal Government Telehealth Initiative: Connecting Health Services with the Future: Modernising Medicare by Providing Rebates for Online Consultations provides people in remote, regional and outer metropolitan areas with access to telehealth through online consultations. This initiative provides patients with easier access to specialists, without the time and expense involved in travelling to major cities. MTAA applauds this policy initiative but raises concerns regarding a patient group that has been excluded in the range of new telehealth Medicare rebates. The exclusion of remote monitoring of CIEDs from the telehealth initiative effectively limits the potential benefits that could be realised by all stakeholders – payers, patients, clinics and doctors. A policy response is appropriate as remote monitoring is not a different service from inoffice monitoring; rather it is a way to deliver the same service from a remote location. National and international experience demonstrates that this service can be delivered safely and effectively The diagnosis, surgical intervention and on-going in-clinic care of Australian patients with CIED is well-established and sufficiently reimbursed by Medicare. However, more efficient and cost-effective use of existing resources leading to improved health outcomes for Australian patients with CIEDs can be achieved. Remote monitoring can be established as a publicly funded service in Australia with positive Federal Budget impact. There are no barriers to adoption except Medicare funding. Additional savings to the healthcare system are possible, due to prevention of costly episodes of care such as strokes and hospitalisations due to heart failure. Cost savings (especially for rural patients) are likely due to reduced travel to in-office visits by patients and their carers (time off work). The proposed Medicare rebate models address budget restraints and offer time and financial savings to patients and clinics. 2. About the medical technology industry MTAA represents the manufacturers, exporters and suppliers of medical technology products in Australia. Medical technologies are products used in the diagnosis, prevention, treatment and management of disease and disability. Products range from consumable items such as wound care products and syringes, to high technology implantable devices such as cochlear implants, cardiac defibrillators and orthopaedic joints, diagnostic imaging equipment, and products which use biological materials. The medical technology industry had sales in Australia of more than $7.5 billion in 200910 and employs more than 17,500 people. It is strongly research-based, often working closely with healthcare professionals to design and develop products for improved patient benefit. MTAA represents companies supplying approximately 70% of all nonpharmaceutical medical products on the Australian market. Medical Technology Association of Australia Business case for public funding of remote monitoring of CIEDs Page 2 3. Background CIEDs have expanded in number and complexity since their introduction in 1958 and include cardiac pacemakers, implantable cardioverter-defibrillators, implantable cardiac resynchronisation devices and implantable loop recorders. These devices are widely used for the treatment of chronic cardiac conditions such as bradycardia, tachycardia, heart failure and for arrhythmia monitoring i. There are approximately 3.5 million people in Australia with long-term cardiovascular disease (CVD) ii. CVD costs more lives in Australia and has the greatest level of health expenditure of any group of diseases. Deaths due to cardiac arrhythmias account for over half of all deaths attributable to CVD in Australia each year iii. Treatment for CVD costs over $5.9 billion in 2004-05 with over half of this money spent on patients admitted to hospital. The cost burden of CVD in Australia is expected to rise due to increasing incidence and prevalence. Telemetric remote monitoring of implanted devices represents a simple evolution in the method of delivering a standard medical service within a mature therapeutic field. The expertise, technology and infrastructure are all well-established and do not require additional expenditure to implement this telehealth service. However, the lack of Medicare rebates for remote monitoring of patients with a CIED prevents the widespread adoption of this telehealth service and encourages a continued reliance on traditional, inefficient in-office monitoring 4. Patient population and clinical need Implantation of cardiac devices in Australia has grown from 11,649 in 2000-01 to 21,223 in 2007-08 – representing an average growth rate of 11.74% per year. Therapeutic procedures for CVD and cardiac arrhythmias include pacemaker insertion and cardiac defibrillator implant. Between 2007 and 2008 there were 17,010 pacemaker insertion procedures and 3,318 cardiac defibrillator implantation procedures performed among hospitalised patients iv. Growth in the use of implantable devices to treat cardiac conditions is expected to rise with an increase in prevalence associated with Australia’s ageing population. Remote monitoring of patients with CIEDs is particularly well-suited to Australia which has approximately one-third of the population v, but only 12% of practicing cardiologists vi living in rural and remote regions. Furthermore, due to a shortage of cardiac clinics and electrophysiologists in Australia the time between in-clinic visits has increased for both urban and rural patients, which can compromise patient health outcomes vii. This forces local general practitioners to carry the burden of cardiac management for these patients viii. 5. Remote monitoring for CIEDs CIEDs are therapeutic devices used to treat a range of cardiac conditions. Long term (lifelong) follow-up of patients with CIEDs is necessary to monitor and optimize device function and to identify clinical and/or device-related problems. Most new CIEDs have the potential to be remotely monitored. While they cannot be reprogrammed remotely, this feature is being tested in pilots. There are two general types of remote monitoring systems, those that transmit data automatically and those where the patient is actively involved in data transmission. Medical Technology Association of Australia Business case for public funding of remote monitoring of CIEDs Page 3 Remote monitoring is the exchange of medical data between a patient who is at home and a healthcare professional, such as a cardiologist, based in a medical setting. Data from CIEDs can be automatically transferred using wireless technology and the internet. All manufacturers of CIEDs can now be monitored remotely for clinical or device assessment. Figures 1 and 2 below illustrate the difference between in-clinic and remote telemetric monitoring (follow-up). Figure 1: Telemetric in-clinic follow-up Figure 2: An example of telemetric remote follow-up* ix *Adapted from Orlov MV et al. (2009) - example of Medtronic’s CareLink remote follow-up system The purpose of both the remote and in-clinic monitoring is to assess the device performance noting that remotely transmitted data is identical to that available to the specialist during in-clinic monitoring. Remote monitoring of CIEDs removes the requirement for additional facilities, resources or technology and simply uses the existing resources more efficiently. This model would therefore be ideally suited for inclusion in the telehealth initiative. Medical Technology Association of Australia Business case for public funding of remote monitoring of CIEDs Page 4 Routine follow-up visits have traditionally required the patient to physically attend a clinic to allow specially trained staff to interrogate the device via telemetry and verify clinical events. These consultations can be time-consuming and place a substantial burden on already stretched healthcare resources x. Approximately 90% of patients who attend a clinic for routine monitoring do not require changes to either their device or their medical treatment xi. One of the key benefits of remote monitoring is that the clinician can be provided with automatic alerts of events that occur between routine visits. Devices are programmed to relay data if a certain heart condition is detected by the device, if therapy is delivered by the device for clinical purposes or if a safety issue is detected due to malfunctioning of a component. For many Australians with CIEDs remote monitoring would provide more practical and affordable access to their specialist. Remote monitoring systems include the implanted device, a remote sensor device (usually located in the patient’s home), a data transmission system and a remote monitoring service centre. Additional features of these systems may include networked monitors for vital signs such as blood pressure and weight. Patients may be actively involved in symptom monitoring and the system may prompt them to answer questions regarding their health. A recent meta-analysis of >6,000 patients found that remote monitoring of both symptoms and measures of heart failure status such as weight and blood pressure resulted in improved survival ratesxii. The goals for remote CIED monitoring have been outlined by an expert consensus group involving cardiac electrophysiologists representing 11 professional associations listed in Appendix Bxiii: There are a number of remote monitoring systems available in Australia. Some devices with remote monitoring capabilities listed on the Australian Register of Therapeutic Goods (ARTG) are shown in Appendix C. 6. Clinical benefits of remote patient monitoring The clinical benefits of remote monitoring for patients with CIEDs include: • • • • • • • • • • • • • 1 Lower mortality rates xiv 1 Reduced number of hospitalisations xv Shorter duration of hospitalisations xvi Discovery of previously undetected clinical symptoms xvii Increased adherence to treatment program and follow-up xviii Improved continuity of follow-up xix Reduction in the number of inappropriate shocks xx Improved quality of life Avoidance of further cardiac events xxi Reduction in anxiety associated with possible failure of the device xxii Efficient, timely follow-up for patients who live in remote locations xxiii Early detection of clinical events such cardiac arrhythmias and/or deterioration of clinical status xxiv Early detection of device related issues such as battery depletion, out-of-range leads or inappropriate delivery of shocks xxv Proof pending from randomized controlled trials Medical Technology Association of Australia Business case for public funding of remote monitoring of CIEDs Page 5 Appendix A contains a detailed list of clinical findings from relevant cohort studies and randomised control trials (RCTs) evaluating remote monitoring of patients with CIEDs. 7. Current reimbursement for remote monitoring of CIED patients Heart failure is increasingly being managed through specialist clinics. The diagnosis, surgical intervention and on-going in-clinic care of Australian patients with CIEDs are well-established and sufficiently reimbursed. Surgeons are reimbursed through Medicare for the implantation of the cardiac device and the actual device is reimbursed for eligible patients with private health insurance through the Prostheses List (manufacturers do not receive extra reimbursement for additional device capabilities such as remote monitoring). Medicare funding is available for routine, in-clinic monitoring of patients with an CIED (inclinic monitoring is covered by Medicare Benefits Schedule (MBS) item numbers 11718, 11721, 11722 and 11727). Despite a wealth of evidence for the value of remote monitoring for patients with CIEDs, Medicare only reimburses clinicians for face-to-face assessments. There has been a Medical Services Advisory Committee (MSAC) application (Application 1111) for public funding of remote monitoring of CIEDs in Australia. The 2008 MSAC review found that while remote monitoring is safe, clinical effectiveness was not demonstrated and there was insufficient data for a formal economic assessment xxvi. For these reasons, MSAC did not recommend public funding for remote monitoring of patients with CIEDs at that time. However, the results of several landmark trials in the field of remote monitoring (ALTITUDE, COMPAS, CONNECT and TRUST) have been published since. The results of these trials (which include over 188,000 patients with CIEDs) significantly addresses the uncertainty presented in the review of Application 1111 and establishes the evidence for clinical effectiveness xxvii. Canada has undertaken a recent health technology assessment of remote monitoring of CIED. The Ontario Health Technology Advisory Committee (OHTAC) reviewed the effectiveness and safety of internet-based device-assisted remote monitoring systems for therapeutic CIEDSxxviii. The review identified six systematic reviews, seven RCTs and 16 cohort studies and concluded that there was significant evidence for a reduction in in-office clinical follow-ups in the first year after implantation and that the rate of detecting clinically significant events is higher in patients followed up remotely. The review concluded that remote monitoring is a safe, effective substitute for in-office followup care and that these systems should be increasingly used in areas where geographical distance prevents regular follow-up. 8. Telehealth on the Medicare Benefit Schedule Public funding for telehealth was introduced in July 2011, with $620 million being allocated for telehealth services over 4 years. Currently the definition of telehealth has been restricted to videoconference consultations and reimbursement is only made in cases where there is an audiovisual link between the patient and practitioner. This proposal seeks the extension of public funding to the remote monitoring of cardiac implantable electronic devices. There are now 11 specialist telehealth items, which allow a specialist to consult with a patient by videoconference. These telehealth items are billed with an existing specialist consultation item under the MBS, and increase the base schedule fee for the Medical Technology Association of Australia Business case for public funding of remote monitoring of CIEDs Page 6 consultation item by 50%. Twenty-three new patient-end attendance items have also been introduced to support telehealth consultations with remote specialists. The schedule fees for these new patient-end item numbers effectively represent a 35% loading on the standard consultation item fees. General practitioners, (and practice nurses and Aboriginal health workers on behalf of GPs), midwives, nurse practitioners, and other medical practitioners will be able to bill the relevant patient end attendance item in combination with both a telehealth service incentive and a bulk billing incentive. MBS item numbers for telehealth (video consultations) are currently associated with a $6,000 incentive when a health practitioner provides their first video consult. Reimbursement for remote monitoring of CIEDs should not be associated with such an incentive as there are no set up costs. It is assumed that clinicians who assess CIED data already have access to computer equipment that is appropriate for data review. While remote monitoring of CIEDs can be considered a telehealth item, MTAA does not propose to adapt those MBS codes for this purpose. Remote monitoring of CIEDs does not rely on a video link between physician and patient. Hence, there are no setup costs involved. In addition, the health care specialist does not consider patient symptoms, merely the downloaded data from the implanted cardiac device. Hence, the existing items for in-office follow up provide a better starting point. These items concern the same specialist group (cardiologists and electrophysiologists) and offer efficiency gains similar time commitments. 9. Economic impact of remote monitoring Eucomed 2 recently submitted CIEDs as one of its four proposals to the European Commission’s Active and Healthy Ageing Innovation Union Strategy because it believes this form of treatment has not only the potential to significantly improve chronic cardiac care but also because this treatment can potentially save billions of Euros xxix. In Australia, direct cost savings may be achieved as the result of: • • • • • • Reducing the number of clinical visits (including specialist visits) xxx Reducing the distance travelled by healthcare professionals Early detection of symptom exacerbations and early intervention xxxi Fewer and/or shorter hospital stays xxxii Delaying the move into residential care Decreased utilisation of healthcare resources xxxiii. Indirect cost savings may be achieved as the result of: • • • 2 Reduced travel for patients and carers xxxiv Increased patient and carer productivity (i.e. less disruption to work schedule) Reduced number of inappropriate ICD shocks resulting in less battery drain and thus longer service time of the device xxxv. This leads to a smaller number of replacement devices per patient including a smaller number of replacement surgeries. EUCOMED is the European Medtech industry body Medical Technology Association of Australia Business case for public funding of remote monitoring of CIEDs Page 7 10. Cost effectiveness of remote monitoring of patients with CIEDs The costs of remote monitoring include the cost of the patient monitor device and the cost of data monitoring by appropriate staff. There have been few economic studies of the cost-effectiveness of remote monitoring for cardiac devices. Results are dated and not generally applicable to Australia. A brief summary of the available economic evidence is presented in Table 1. Table 1: Summary of economic data on remote monitoring of CIEDs Author Study Findings Elsner et al. (2006) xxxvi Assessed the costs of remote monitoring for 115 patients with ICDs There were significant cost reductions associated with a decrease in office visits for the RM group. For 100 patients a total of 81 physician hours and €712.31 could have been saved p.a. using RM. Replacing office visits with RM reduced transportation costs by 63% and time taken to analyze results by 41% Fauchier et al. (2005) xxxvii Assessed potential cost savings for the longterm care of ICDs using remote monitoring from a database of 502 patients Compared the costs of conventional follow-up and RM. Costs included: number of visits, clinical fees, ECGs, ICD surveillance and transportation. Results showed that RM may obviate up to 2 visits per year. Cost savings began at approximately 33.5 months. The time to onset of cost saving was dependent on distance, RM reduces overall costs, particularly when the distance between home and medical facility is >100 km Klersy et al. (2009) xxxviii Meta-analysis of 21 RCTs comparing remote monitoring with usual care in 5,715 patients with heart failure RM was associated with a significant decrease in the number of hospitalisations. Direct cost savings associated with RM ranged from €300 to €1000 and when combined with a QALYs gain of 0.06, the authors concluded that RM ‘dominated’ existing standard care for heart failure patients Raatikainen et al. (2008) xxxix Assessed the cost effectiveness of remote monitoring in 41 patients with ICDs Replacing two office visits with RM was associated with a 41% saving (the total annual saving associated with RM was €524–749 per patient). The authors concluded that the use of RM for all new ICD patients within Western Europe may save €16–23 million annually ECGs = electrocardiogram, QALY = quality-adjusted life year Medical Technology Association of Australia Business case for public funding of remote monitoring of CIEDs Page 8 11. Current cost to Medicare for CIED in-office follow-up checks The Medicare Benefits Schedule lists a number of items for the in-office testing of CIEDs. However, as the in-office check is performed by a specialist, other items are usually co-claimed for a patient visit. When looking at the total cost of in-office CIED checks to the Commonwealth, those costs must be taken into consideration as well. The items and their individual benefits (100% benefit level) are shown in Table 2. Table 2: Current benefits for in-office assessment of cardiac devices (100% benefit shown, benefits are current as of November, 2011) 3 MBS Item Service Benefit 11718* Implanted pacemaker pacemaker testing involving ECG, measurement of rate, width and amplitude of stimulus, including reprogramming when required $34.10 11721* Implanted pacemaker testing of atrioventricular sequential, rate responsive, or antitachycardia pacemakers, including reprogramming when required $68.45 11722* Implanted ECG loop recording, for investigation of recurrent unexplained syncope, including re-programming of device, retrieval of stored data, analysis, interpretation and report $34.10 11727* Implanted defibrillator testing involving electrocardiography, assessment of pacing and sensing thresholds for pacing and defibrillation electrodes, download and interpretation of stored events and electrograms, including programming when required $93.00 116 Subsequent attendance for follow-up with a consultant $74.10 * Note, other MBS item numbers can be claimed in conjunction with this item In the 2010-11 financial year the benefit paid by Medicare purely for CIED follow up was $8,456,202. This amount does not include co-claimed items. The costs are split as shown in Table 3 4. Table 3: Spending on current model: Patients are scheduled for periodic in-office consultations 3 MBS Item Device Count of Services 2010/11 Medicare contribution to service 2010/11* 11718 Single-chamber Pacemaker 8,968 $254,309 11721 Dual-chamber Pacemaker 99,164 $5,630,470 11722 Implantable ECG Loop Recorder 2,408 $68,833 http://www.mbsonline.gov.au/, data accessed November 2011 Medicare Benefits Schedule. 4 Statistics from www.medicareaustralia.gov.au/statistics/mbs_item.shtml accessed online 7 November 2011. Medical Technology Association of Australia Business case for public funding of remote monitoring of CIEDs Page 9 11727 ICD 32,480 $2,502,590 * Note, that the amount for Medicare contribution is smaller than the product of Count of Services and Item Benefit (Table 2) due to less than 100% benefit being paid by Medicare on some occasions. On the basis of MBS data from January 2008 to December 2010 and assuming only 50% of patients appear on MBS statistics data, the projected number of patients for implanted pacemakers 5 and implantable cardioverter/defibrillators (ICDs) 6 in 2015 will be 112,000. The average number of in-office checks performed during the financial years 2008, 2009 and 2010 for pacemaker and ICD patients in Australia is shown in Table 3. Table 3: Average number of in-office checks per patient per year 7 Average number of pacemaker checks per patient Average number of ICD checks per patient 2.4 3.0 During an in-office follow up other items are usually co-claimed depending on the additional diagnostic and therapeutic interventions by the health practitioner during the check. For simplicity, only the most frequently claimed item “Specialist Consultation” (MBS item 116) is considered when calculating the current average cost per patient per year. MTAA acknowledges that this might underestimate the actual cost: Current Cost per ICD patient p.a. Item code 11727 (ICD check) Item code 116 (specialist consultation) Average total cost Average cost per ICD patient per annum for follow up checks: 3.0 checks p.a. * $167.10 $93.00 $74.10 $167.10 $501.30 (Equation 1) Similarly, the current cost per pacemaker patient is $342 per year. 12. Proposed models for public funding of remote monitoring of CIEDs The following models are potential ways Medicare could implement the reimbursement of remote monitoring services for CIEDs. However, there may be alternative solutions that address the priorities of Medicare while supporting patient access and fair clinician reimbursement. MBS items for reimbursement of remote monitoring could be arranged in the following ways: 5 Projected for 2015 based on MBS data from Jan 08 to Dec 10 (MBS item 11718 and 11721) and assuming 50% of patients on MBS only. 6 Projected for 2015 based on MBS data from Jan 08 to Dec 10 (MBS item 11727) and assuming only 50% of patients appear on MBS statistics (no public patients). 7 Source: Medicare Australia, 2011 Medical Technology Association of Australia Business case for public funding of remote monitoring of CIEDs Page 10 Example A: MBS item numbers based on existing items with an included loading for remote monitoring Example B: Capitated costing model Example C: Improved capitated model (flat fee per quarter). Example A: CIED routine follow up PLUS MBS item numbers could be based on existing item numbers for CIED routine assessment with an included loading for remote assessment (Table 4). Table 4: New MBS item numbers for remote monitoring based on existing item numbers for CIED in-office assessment with an included loading for remote assessment MBS item Service 11721 Remote remote monitoring for implanted pacemaker testing 11722 Remote remote monitoring for implanted ECG loop recorders 11727 Remote remote monitoring for implanted defibrillator testing 116 remote Subsequent attendance Fee $102.67 [derived fee: using item 11721 $68.45 (implanted pacemaker testing) including similar loading as for telehealth items currently on MBS (50%)] $51.15 [derived fee: using item 11722 $34.10 (implanted ECG loop recorder testing) including similar loading as for telehealth items currently on MBS (50%)] $139.50 [derived fee: using item 11727 $93 (implanted defibrillator testing) including similar loading as for telehealth items currently on MBS (50%)] $111.15 [derived fee: using item 116 $74.10 (subsequent attendance for follow-up with a consultant) including similar loading as for telehealth items currently on MBS (50%)] It is anticipated that in most cases MBS item numbers for telehealth patient-end services would not be used as the monitoring takes place asynchronously, i.e. data transmission and evaluation by the specialist do not take place at the same time. While this model appears intuitive and straight forward, it has a few shortcomings, the main one being that there is no cap on the number of services provided using those item numbers. Potentially, this could lead to over servicing of patients with CIEDs with resulting cost blow outs of the Commonwealth budget. Using this model it would be hard to delineate between conventional in-office follow up and providing the service remotely as in theory an in-office follow up on one day could be followed by another claim for a remote check the following day (which may be legitimate to check whether a device reprogramming had the intended effect, for instance on atrial fibrillation recurrence). The loading of 50% seems somewhat arbitrary as telehealth items may not be the best comparator due to the missing same-time-interaction between patient and health care provider. Medical Technology Association of Australia Business case for public funding of remote monitoring of CIEDs Page 11 Example B: Capitated Costing Model An MBS item could be developed using a capitated model whereby a clinician is paid once per year per patient to provide the service. This would ensure there was no over servicing, i.e. the physician would be paid once regardless of how many data transmissions and reviews occur. Calculations for an annual fee could be made based on the number of online consultations expected per year (and could also include in-office visits). Remuneration model and budget impact Based on established MBS in-clinic monitoring item codes 11721, 11727 and 11722, there are various remuneration models that could be considered by Medicare for remote monitoring of CIEDs. The proposed model is designed to ensure that the overall budget impact is cost neutral and predictable. Remuneration models successfully implemented in other jurisdictions include capitation. A capitation model involves paying the clinician a set annual fee for each patient enrolled in a remote monitoring program. The clinician is then responsible for the management of the patient regardless of the amount of care required or whether the consultation is in-office or a remote monitoring service. Not all patients are suitable for remote monitoring; some may wish to move between remote and in-clinic monitoring, while others may be suitable for remote monitoring at one period but not in another. Such changes may be due to variations in a patient’s clinical status, their physical location or because their personal preferences change. Doctors also must have flexibility in the method of monitoring their patients and be able to move between models. However, the proposed capitation model is a single fee to cover inoffice and remote monitoring for an individual patient and is not intended to be used in addition to the current in-office items. Studies have shown that remote monitoring can be used to replace 50–63% of in-clinic visits without adversely affecting patient outcomes such as CV hospitalisations, ER visits and mortality xl. As a result, remote monitoring could safely and effectively replace a portion of scheduled and unscheduled services – thereby decreasing the demand on hospital resources, physician time and associated transportation costs xli . Table 5: Capitation model (example): Doctors receive a set annual fee for each patient with a CIED enrolled in a remote monitoring program MBS Item Follow-up for Device 11721c remote Pacemakers 11722c remote 11727c remote MBS Fee** Estimated Estimated Medicare Item Count of contribution to Services*** service**** $164.30 41,318 $ 6,788,547 Implantable Loop Recorder $137 1,911 $ ICD $279 10,827 $ 3,020,733 261,807 ** Formula: MBS Item for in-office multiplied by the Average Annual per Patient Utilisation in Table 3 *** Formula: Count of Services (Table 6) divided by the Average Annual per Patient Utilisation in 3 **** Formula: MBS Item Fee multiplied by Count of Services Medical Technology Association of Australia Business case for public funding of remote monitoring of CIEDs Page 12 The capitation model as presented does not require additional funding. The proposed annual fee is based on current utilisation. The annual capitation fee could be higher than proposed to be consistent with the recent Medicare initiative providing higher rebates for video consultations than for in-office consultations and considerable incentives to encourage the uptake of the remote monitoring service. Example C: Improved Capitated Model (Flat Fee per Quarter) This model takes the previous Example B and improves on possibilities for tracking of the new items as well as unscheduled in-office checks that may occur for remotely monitored patients. In addition by calculating the actual cost to the provider of providing the remote monitoring service, it will be possible to realize a cost benefit to the Commonwealth budget. This becomes possible by passing on savings created by efficiency improvements to the payer (Medicare) while at the same time providing a fair reimbursement to the health care provider. Under this model, two items for remote monitoring could be added to MBS with a flat fee per quarter. It might be advantageous to add two further items for in-office checks prompted by remote monitoring at the same benefit levels as existing items (11721 and 11727) to track spending (Table 6). This model would allow retention of the existing items for pacemaker, loop recorder and ICD checks (11718, 11721, 11722 and 11727) and their respective benefits for legacy devices and late technology adopters. Table 6: Proposed New Item Codes Pacemakers and CRT-Ps ICDs and CRT-Ds Remote Monitoring (flat fee per quarter) “Item Code 1” “Item Code 3” In-office Testing (indicated by Remote Monitoring) “Item Code 2” “Item Code 4” The new Item Code 1 and Item Code 3 would include the remote monitoring service and one optional in-office follow up per year. The landmark clinical trials for remote monitoring TRUST (ICDs) and COMPAS (pacemakers) used a setup where the regular in-office follow up was replaced by remote monitoring plus one in-office check after 12 months (TRUST) and 18 months (COMPAS). However, with additional clinical data becoming available, it is anticipated that the time between two in-office checks will increase. Hence, MTAA proposes to make the in-office check in Item Codes 1 and 3 optional. The benefits for those two items could be claimed quarterly rather than annually in order to better align the time of service delivery and benefit payment. That way, the health care provider is not forced into a position of having to provide a service in advance and only get paid after one year. Vice versa, Medicare does not have to pay a benefit in advance but only after the service has been provided for the quarter. Naturally, the same advantages of capitation as discussed for Example B do apply. During any quarter for which Item Code 1 or 3 are claimed, items 11721 or 11727 for inoffice checks cannot be claimed. This allows a clear transition of patients from the old to the new system. Item Code 2 and Item Code 4 are similar to the existing items for in-office checks of pacemakers and ICDs respectively. However, they are applied for in-office checks that became necessary after reviewing remote monitoring data transmitted by the device. Medical Technology Association of Australia Business case for public funding of remote monitoring of CIEDs Page 13 Using separate items for those checks would allow tracking of spending and gathering of valuable information on the real-life incidence of those checks. Documentation of the necessity of those checks is facilitated by all remote monitoring systems on the market. Again, potential for over servicing can be further reduced by limiting the number of claims possible for those items in any given year. Remuneration model and budget impact Following are calculation examples for the case of ICD patients only. However, the calculations for other patients with CIEDs are available. The cost for providing remote monitoring service is mainly influenced by the time that the specialist has to spend to review the data for each of his or her patients. In order to estimate the time of the health care provider required per remotely monitored patient, MTAA anticipates the following based on clinical evidence: • There will be 4 messages per year per ICD patient xlii • Dealing with one message will require an effort on part of the clinician similar to reading and reporting on a 12-lead ECG (MBS#11702: $15.25) • Remote monitoring takes about 10 minutes per 100 patients per day of clinician time, i.e. 20 minutes per patient per year (200 working days) xliii o Comparable MBS item is a 20 minute consultation by a specialist (MBS#133: $129.65) Therefore, the cost of providing the remote monitoring service (including one optional inoffice check annually) would be $283.65 per annum. Hence, the proposed benefit for Item Code 3 is $70.90 (quarterly). Similarly with slightly differing assumptions the proposed benefit for pacemaker patients (Item Code 1) would be $57.15. Following from that, the annual cost per remotely monitored patient (as example for ICD patients) can be calculated as follows: Remote Monitoring Fee (Item Code 3) Consultation fee (MBS item 116) 0.78 unscheduled checks (according to the TRUST RCT) reimbursed at proposed Item Code 4 ($93.00) Total cost (per annum) $283.65 $74.10 8 $72.54 $430.29 In order to calculate the cost savings by employing the model proposed in Example C the annual cost per ICD patient can be compared as follows: Current Model as per (Equation 1): $ 501.30 Proposed Model (Example C): $ 430.29 This model leaves some latitude for additional consultations. Assuming 34,000 ICD patients in Australia in 2015 9 (and 100% uptake of remote monitoring based on the availability of public funding), the cost savings to the Commonwealth budget would be in the order of $2.4 million. A similar calculation can be 8 This is for the specialist consultation provided as part of the optional in-office follow up once annually as part of the remote monitoring reimbursement. 9 Projected for 2015 based on MBS data from Jan 08 to Dec 10 (MBS item 11727) and assuming only 50% of patients appear on MBS statistics (no public patients). Medical Technology Association of Australia Business case for public funding of remote monitoring of CIEDs Page 14 made for pacemaker patients. The annual cost savings to the Commonwealth budget would be an additional $350,000. Therefore, MTAA argues that remote monitoring for CIED patients can be introduced in Australia to improve patient care while at the same time having a strongly positive effect on the Commonwealth budget (with cost savings of nearly $3 million). This is due to the reduction of unnecessary and resource-intense inoffice checks and their replacement by remote monitoring. 13. Conclusion For many Australians, the opportunity to use remote monitoring would provide a more practical, reliable and affordable access to medical care, without the time and expense involved in travelling to major cities. Preliminary evidence from clinical trials suggests that remote monitoring could mediate the need for costly episodes of care such as strokes and hospitalisations due to heart failure. The efficiency gains associated with remote monitoring are obvious from the perspective of the service provider and payer, but patients and their carers also benefit. When compared to in-clinic monitoring, the reductions in time, costs and opportunity costs (such as time off work) for both the patient and informal caregiver are particularly relevant for those living in rural and remote areas. Finding the best way forward can be readily included in current initiatives. In terms of quality of life outcomes, remote follow-up has been shown to be safe, and is preferred by both patients and physicians to in-office visits. In addition, remote monitoring ties in neatly with the Australian Government’s current telehealth Initiative. Clearly there are a variety of courses of action that can be taken to provide this patient-physician preferred service in a cost neutral manner or even realizing inherent cost savings to the Commonwealth budget. Medical Technology Association of Australia Business case for public funding of remote monitoring of CIEDs Page 15 Appendix A Cohort studies evaluating remote monitoring of ICDs – Arrhythmia and heart failure monitoring (modified from xliv) Author Country N Device Follow-Up Findings Burgada et al. (2006) xlv Belgium 271 ICD 12 months 908 pairs of RM and standard care data follow-up were compared. Physicians indicated that based on observation of remote data 81% of standard visits could have been avoided. The false-negative detection rate was 14% Ellery et (2006) xlvi al. United Kingdom 123 CRT CRT-D 3 months Assessed the potential of RM to predict hospitalization in HF patients. There were 11 unplanned readmissions during follow-up. In 70% of these cases a retrospective data review detected an increase in mean heart rate preceding hospital admission. A decrease in cardiac resynchronization therapy was reported in 43% of patients who were re-admitted Hauck et (2009) xlvii al. Germany 69 ICD CRT 18±9 months A total of 206 event reports were transmitted: 94% were prompted by clinical episodes, 3% by ineffective ICD shocks, 2% by abnormal pacing impedance and 1% by battery depletion. There were 8 serious adverse events, 6 of which were discovered by RM. Without RM there would have been a delay in clinical follow-up of 1.9±0.5 months in the first year (quarterly follow-up) and 4.9±0.5 months in following years (bi-annual follow-up) Joseph et (2004) xlviii al. United States 124 ICD 6 months Assessed physician and patient acceptability, diagnostic value and safety of RM. Patients were highly satisfied with the ease of learning to transmit data, time savings and convenience of routine follow-up in >90% of cases Medical Technology Association of Australia Business case for public funding of remote monitoring of CIEDs Page 16 Lazarus et al. (2007) xlix France (multicenter registry, global database covering 23 countries) 11,624 PM ICD CRT 10.5 months Transmitted events included medical (e.g. arrhythmias), configuration (e.g. pacing functions) and system related events (e.g. battery status). 86% of events were medical, 1.3% were alerts of abnormal function indicating device inactivation and 4% were alerts indicating an ineffective shock had been delivered. The mean interval between last follow-up and occurrence of unconfirmed asymptomatic events notified by RM was 26 days (this represented detection of an event 64 or 154 days earlier in patients usually followed up at 3 and 6 month intervals respectively). The authors concluded that the lifesaving potential of RM is due to early detection of clinical events Marzegalli et al. (2008) l Italy 67 CRT-D 3 months Assessed ease of use and patient satisfaction. RM data were judged as complete and as providing the same level of care as an in-clinic visit. 97% of patients would recommend RM to others and 78% had a preference for RM Masella et al. (2008) li Italy 67 CRT-D 3 months Follow-up by RM was more efficient than a standard face-to-face visit in terms of time savings for both physicians (5 vs 15 mins) and patients (6.6 vs 116 mins). Of a total of 23 clinical events detected by RM, only 2 necessitated a clinical visit, decreasing the number of unnecessary hospital admissions Mullens et al. (2010) lii United States 194 ICD CRT-D 4 months Assessed the feasibility of a nurse run RM program. Data were analysed daily and the patient was called if parameters crossed a pre-defined threshold (this occurred in 18% of patients). In 38% of cases where data crossed the threshold patients reported non adherence to drug and/or dietary plans. The authors concluded that RM of trends was feasible as part of routine daily management for HF Nielson et al. (2008) liii Germany 260 ICD 10.5 months The majority of events were medically related. Over 60% of new events occurred within the first month after follow up. RM was associated with the early detection of medical and technical events Raatikainen al. (2008) liv Finland 41 ICD 9 months Assessed whether RM is a safe, cost-effective alternative to standard in-clinic follow-up visits. In 99% of cases physicians found that data were comparable to a traditional device interrogation. All unscheduled data transmissions were able to be addressed remotely. Over 90% of patients found the system easy to use. In comparison to in-clinic visits, RM required significantly less time from physicians (8±5 vs 26±17 mins) and patients (7±5 vs 182±148 mins). The substitution of two routine in-office visits by RM reduced the overall cost of routine ICD follow-up by 41% per patient et (multicenter registry) Medical Technology Association of Australia Business case for public funding of remote monitoring of CIEDs Page 17 Ricci et (2008) lv al. Italy 117 PM ICD CRT 227 days Assessed the impact of RM on treatment and health care utilization. The device was programmed to transmit data daily. There were 2,249 analyses undertaken: 92% had no events and no action. 53 in-hospital visits were needed for unscheduled follow-up due to alerts transmitted (91% of these were medical and 9% were device transmission interruptions). There was an overall decrease in inhospital visits in the RM group. There was a significant learning curve on the use of the system: mean web connection times were significantly shorter for the last vs the first 50 connections. The authors concluded that RM had a positive effect due to early detection of life-threatening cardiac events and device failures Ricci et (2009) lvi al. Italy 166 PM ICD CRT 488 days Assessed the impact of RM on detection and treatment of AF: 26% (n=43) of patients had alerts for AF, 22 of who had no history of AF prior to implant. The median time to the first intervention for AF was 50 days (148 days before the scheduled follow-up) Ricci et lvii (2010) al. Italy 119 PM ICD CRT 12 months Evaluated patients’ acceptance of RM. There was a high level of acceptance and satisfaction. At 1-year follow-up, 93% of patients in the trial wished to continue with RM Sacher et (2009) lviii al. France (multicenter registry) 70 ICD 33±17 months Assessed RM in patients with Brugada syndrome who are at risk of sudden cardiac death. At follow-up the number of cardiology consultations was significantly lower in the RM group. Inappropriate shocks occurred in 8.5% of the RM group vs 17% of the standard care group (ns). The authors concluded that RM decreased outpatient consultations and may help prevent inappropriate shocks Santini et (2009) lix al. Italy 67 CRT-D 3 months Assessed whether RM improves clinical management of tachy-arrhythmias and HF episodes. In 70% of cases a patient could be managed remotely, avoiding a hospital visit. In a number of cases RM led to either adjustment of drug therapy over the phone or patient reassurance and no action Saxon et (2010) lx al. United States 185,778 ICD CRT Up to years ALTITUDE Survival Study Medical Technology Association of Australia Business case for public funding of remote monitoring of CIEDs 5 Assessed long term survival data and shock therapy with standard and RM follow-up. For the 69,556 ICD and CRT-D patients receiving remote follow-up, 1- and 5-year survival rates were significantly higher in comparison to the 116,222 patients who received follow-up in device clinics only (50% reduction in mortality). 4% of CRT-D patients also transmitted weight and blood pressure data on average 2.5 times per week. These patients had the lowest mortality risk (vs other networked CRT-D patients). The one year incidence of shock was 14% (8% were appropriate and 6% were inappropriate) Page 18 Schoenfeld al. (2004) lxi et United States 59 ICD 1 week Assessed the impact of RM on patients and clinicians. Practitioners reported that in 97% of occasions the service enabled them to provide care comparable to an in-office visit. Nurses reported a higher degree of satisfaction than physicians. 98% of patients indicated that the RM system was very easy or somewhat easy to use. RM detected a number of previously unobserved clinically significant findings, e.g. unobserved atrial undersensing Theuns et al. (2009) lxii Netherlands 146 ICD CRT 22±16 months Data were transmitted daily or in the case of pre-specified events (e.g. out-ofrange leads, arrhythmia). A total of 57,148 remote transmissions were recorded, 64% were classified as clinical and 6% as system-related. Only 1.8% of events were data transmissions triggered by critical events. The authors concluded that there was minimal additional burden on clinical workload despite the large number of data transmissions Varma et (2005) lxiii United States (multicenter) 107 PM 3 months Assessed RM in AF patients. Review of data was retrospective: 29 patients experienced 645 AF days (10.5% of implants). Data from 20 patients found that RM detected new-onset silent AF in 3 patients which resulted in anticoagulation therapy being started (tailored treatments were provided to the other 17 patients) al. Randomised controlled trials evaluating remote monitoring of ICDs – Arrhythmia and heart failure episodes (modified from lxiv) Medical Technology Association of Australia Business case for public funding of remote monitoring of CIEDs Page 19 Author Country System N Follow-Up Findings Al-Khatib et al. (2010) lxv United States Medtronic Care Link® 151 6, 12 months RM was compared to standard quarterly in-clinic device interrogations. The rate of the composite endpoint (hospitalization, emergency room visits, unscheduled clinic visits) did not differ significantly between groups. There was a significantly higher detection rate of AF at follow-up in the RM group. There were no between-group differences in cost Bikou et (2010) lxvi United Kingdom St Jude House Call Plus® 20 1, 3, months To determine work impact and cost of RM. Total time for RM follow-up was significantly shorter in the RM group, the in-clinic waiting time (30±18 mins) was eliminated in the RM group, the time for actual device interrogation did not differ between groups al. 6 ICD/CRT-D Crossley et (2009) lxvii al. United States Medtronic Care Link® 897 3 years Assessed whether RM decreases the duration of post-operative hospitalization, while maintaining safety levels. The event rate detected for clinically actionable events (CAEs) was higher in the RM group: 66% of CAE events were detected remotely, whereas for controls only 2% were detected during a standard transtelephonic transmission. The median time to detect a CAE was significantly faster in the RM group (4.9 vs 6.3 months) United States Medtronic Care Link Network® 1,997 15 months To determine whether RM with wireless (automatic) physician notification reduces the time from a clinical event to a clinical decision. There were no between-group differences in mortality. The mean time from a clinical event to the clinical decision was significantly faster (by 17 days) in the RM group. More alerts were noted for out-of-range lead impedance in the RM group and the time to detect these critical events was significantly faster. Replacing 4 clinic follow-up visits with RM reduced total clinical visits by 38% (6.3 to 3.3 patient visits per year). There were no between-group differences in health care utilization visits. The overall mean length of hospitalization was significantly shorter for the RM group 115 12 months Compared RM to conventional in office 3 month follow-up in ICD patients. There were no significant between-group differences in hospitalisations or patient mortality. There was a 63% reduction in office visits in the RM group. Over 80% PREFER (Pacemaker Remote Followup Evaluation and Review) Crossley et (2011) lxviii al. CONNECT (Clinical Evaluation of Remote Notification to Reduce Time to Clinical decision) Elsner et al. (2006) lxix REFORM Study ICD/CRT Germany Biotronik Home Monitoring® Medical Technology Association of Australia Business case for public funding of remote monitoring of CIEDs Page 20 ICD/CRT Halimi et (2008) lxx al. of the home monitoring-induced events had high need evaluation France/ Belgium Biotronik Home Monitoring® 379 United States Biotronik Home Monitoring® 1,450 OEDIPE Assessed the ability of RM to detect clinically actionable information sooner than with current transtelephonic transmission and in office follow-up. There was a significant decrease in medical reaction time to an event in the RM group. The duration of hospitalization was significantly shorter in the RM group with the majority of patients leaving hospital the same or next day. QOL was similar between the two groups at 1 month follow up. Costs were lower after one month in the RM group (but not significantly so) (Optimized postoperative surveillance of permanent pacemakers by home monitoring) Varma et al. (2010) lxxi TRUST (Lumos-T Safety Reduced Routine Office Device Follow-Up) Trial 15 months ICD Medical Technology Association of Australia Business case for public funding of remote monitoring of CIEDs Assessed whether RM could safely reduce in-hospital device evaluations. There were no between-group differences for morbidity. Adherence to the protocol follow-up schedule was significantly higher in the RM group. The mean number of in-clinic and in-hospital visits was significantly lower in the RM group (scheduled in-office visits were reduced by 61%). The overall rate of adverse events did not differ between groups. The time from onset of first arrhythmia to physician evaluation was significantly faster in the RM group (1 day vs 35.5 days). The time to detect clinically asymptomatic arrhythmia was also significantly shorter in the RM group. In-office follow-up visits were reduced by 45% at 1 year. The authors concluded that 1-2 minute follow-ups by RM provide a viable alternative to outpatient or doctor visits which typically take 15 minutes twice a year Page 21 Current (unpublished trials) Author Country System N Follow Up Findings Kacet et al. (2011) lxxii Effectiveness and Cost of ICD Follow-up Schedule with Telecardiology (ECOST) 43 French centres Biotronik Home Monitoring® 433 27 months Patients were randomly assigned to either daily RM or quarterly in-clinic follow-up. Device- or patient-related event notifications by the RM system could trigger in-clinic follow-up. RM was as safe as in-office visits and there were no between-group differences in major adverse events. The RM group had a 52% reduction in the number of inappropriate shocks (5% versus 10% in the usual care group) and a 72% reduction in the risk of hospitalisations related to inappropriate shock. RM had positive effects on battery longevity. A secondary analysis of costs will be performed Mabo (2010) lxxiii Comparative Follow-up Schedule with Home Monitoring (COMPAS) 43 French centres 538 18 months Randomized, multicenter clinical study evaluating the long-term follow up, safety and efficacy of RM of pacemakers. RM was found to be comparable to standard follow-up in terms of safety. Adverse events (e.g. death, cardiovascular or pacemaker-related events) occurred significantly less frequently in the RM group. There was a significant decrease in the time to event management with the average time dropping from 145 days with conventional follow-up to 28 days with RM. RM was associated with a 36% decrease in the number of in-clinic visits Mabo et al. (2011) lxxiv Evaluation of Tele Follow-up (EVATEL) study 30 French centres 1,501 12 months Patients were randomised to receive either conventional in-office follow-up visits or remote transmission of data from the ICD to the implant center every three months. RM was shown to be a safe alternative to in-office follow-up. There were no between-group differences in major adverse events (death, hospitalization). Significantly fewer patients in the RM group experienced inappropriate or ineffective therapy from their device. There was a 27% reduction in the number of inappropriate shocks delivered by the ICDs in the RM group (5% vs. 8% in controls) Year ICD Biotronik Home Monitoring® ICD All brands of devices (4 types of ICDs) AF = atrial fibrillation, CRT = cardiac resynchronization therapy, CRT-D = CRT defibrillator, HF = heart failure, ICD = implantable cardioverter defibrillators, PM = pacemaker, QOL = quality of life, RM = remote monitoring. Medical Technology Association of Australia Business case for public funding of remote monitoring of CIEDs Page 22 Appendix B Appendix B: Goals of CIED remote monitoring systems Patient related Optimize quality of life Optimize system to meet clinical requirements Identify at-risk patients and initiate appropriate follow-up with safety corrective actions and alerts Triage other health problems and make appropriate referrals CIED related Document function Identify and correct abnormal function Safely maximise longevity of pulse generator Identify when battery is approaching end of life, identify leads at risk of failure Disease related Document nature/frequency of arrhythmias, correlate with symptoms, determine appropriateness of device response Document (where possible) other physiologic parameters as part of chronic disease monitoring Monitor response to therapy Communication Maintain a patient database Communicate clinical and device-related information to patients and healthcare providers Provide technical expertise and education Medical Technology Association of Australia Business case for public funding of remote monitoring of CIEDs Page 23 Appendix C ARTG listing numbers for devices with remote monitoring capabilities ARTG # Manufacturer Description 116038 Medtronic CareLink programmer, model 2090-Pacemaker programmer 122545 Biotronik Cardio Messenger - Telemetry transmitter, instrument data 123599 Boston Scientific 130439 Biotronik ICS 3000 (with implantation module) - Pacemaker programmer 140355 Biotronik CardioMessenger II – Telemetry transmitter, instrument data 142199 St Jude Medical 155884 Biotronik 163472 St Jude Medical 174544 St Jude Medical 176910 Boston Scientific 179170 Biotronik ZOOM LATITUDE Programming System - Rapid Programmer model 3831 – Active implantable device communicator CardioMessenger II-S - Telemetry transmitter, instrument data Merlin@home Transmitter Model EX1100 - Pacemaker system analyser, transtelephonic Merlin Pacing System Analyzer Model EX3100 - Pacemaker system analyser, noninvasive LATITUDE AUNZ Communicator - transtelephonic CardioMessenger-S - Telemetry transmitter, instrument data Medical Technology Association of Australia Business case for public funding of remote monitoring of CIEDs Page 24 References i Wilkoff, B.L., Auricchio, A., Brugada, J., Cowie, M., Ellenbogen, K.A., Gillis, A.M., Hayes, D.L., Howlett, J.G., Kautzner, J., Love, C.J., Morgan, J.M., Priori, S.G., Reynolds, D.W., Schoenfeld, M.H., Vardas, P.E., & Goldberg, D.M. 2008. HRS/EHRA expert consensus on the monitoring of Cardiovascular Implantable Electronic Devices (CIEDs): Description of techniques, indications, personnel, frequency and ethical considerations - Developed in partnership with the Heart Rhythm Society (HRS) and the European Heart Rhythm Association (EHRA); and in collaboration with the American College of Cardiology (ACC). Europace, 10, (6) 707-725 Clarke R.A., Driscoll A. 2009. Access and quality of heart failure management programs in Australia. Australian Critical Care, 22: 111-6 Nielsen, J.C., Kottkamp, H., Zabel, M., Aliot, E., Kreutzer, U., Bauer, A., Schuchert, A., Neuser, H., Schumacher, B., Schmidinger, H., Stix, G., Clementy, J., Danilovic, D., & Hindricks, G. 2008. Automatic home monitoring of implantable cardioverter defibrillators. Europace, 10, (6) 729-735 Varma, N. 2009. Therapeutic implications of automatic home monitoring of implantable cardiac devices. Current Treatment Options in Cardiovascular Medicine, 11, (5) 366-372 ii Australia Institute of Health and Welfare. (2011). Cardiovascular disease: Australian facts, 2011. Cardiovascular disease series. Cat. no. CVD 53. Canberra: AIHW. iii Subbiah, R.N., Campbell, T.J., & Vandenberg, J.I. (2004). Inherited cardiac arrhythmia syndromes: what have they taught us about arrhythmias and anti-arrhythmic therapy? Clinical and Experimental Pharmacology and Physiology, 31(12):906-12. iv Australia Institute of Health and Welfare. (2011). Cardiovascular disease: Australian facts, 2011. Cardiovascular disease series. Cat. no. CVD 53. Canberra: AIHW. v National Health and Hospitals Reform Commission: A Healthier Future For All Australians – final report June 2009. Canberra: Department of Health and Ageing, 2009. vi Australian Medical Workforce Advisory Committee. (1999). The Cardiology Workforce In Australia, AMWAC Report 1999.5, Sydney. vii Nielsen, J.C., Kottkamp, H., Zabel, M., Aliot, E., Kreutzer, U., Bauer, A., Schuchert, A., Neuser, H., Schumacher, B., Schmidinger, H., Stix, G., Clémenty, J., Danilovic, D., & Hindricks G. (2008). Automatic home monitoring of implantable cardioverter defibrillators. Europace, 10(6):729-35. Varma N. (2009). Therapeutic Implications of Automatic Home Monitoring of Implantable Cardiac Devices. Current Treatment Options in Cardiovascular Medicine, 11:336-72. viii Clarke R.A., Driscoll A. 2009. Access and quality of heart failure management programs in Australia. Australian Critical Care, 22: 111-6 ix Orlov M.V., Szombathy T., Chaudhry G.M., Haffeajee C.I., 2009. Remote Surveillance of Implantable Cardiac Devices. PACE, 32: 928-39 x Clarke R.A., Driscoll A. 2009. Access and quality of heart failure management programs in Australia. Australian Critical Care, 22: 111-6 Nielsen, J.C., Kottkamp, H., Zabel, M., Aliot, E., Kreutzer, U., Bauer, A., Schuchert, A., Neuser, H., Schumacher, B., Schmidinger, H., Stix, G., Clementy, J., Danilovic, D., & Hindricks, G. 2008. Automatic home monitoring of implantable cardioverter defibrillators. Europace, 10, (6) 729-735 Varma, N. 2009. Therapeutic implications of automatic home monitoring of implantable cardiac devices. Current Treatment Options in Cardiovascular Medicine, 11, (5) 366-372 Medical Technology Association of Australia Business case for public funding of remote monitoring of CIEDs Page 25 xi Heidbüchel, H., Lioen, P., Foulon, S., Huybrechts, W., Ector, J., Willems, R., & Ector, H. (2008). Potential role of remote monitoring for scheduled and unscheduled evaluations of patients with an implantable defibrillator Europace, 10:351-7. xii Klersy, C., De Silvestri, A., Gabutti, G., Regoli, F., & Auricchio, A. (2009). A meta-analysis of remote monitoring of heart failure patients. Journal of the American College of Cardiology, 54:1683-94. xiii Wilkoff, B.L., Auricchio , A., Brugada, J., Cowie, M., Ellenbogen, K.A., Gillis, A.M., Hayes, D.L., Howlett, J.G., Kautzner, J., Love, C.J., Morgan, J.M., Priori, S.G., Reynolds, D.W., Schoenfield, M.H., & Vardas, P.E. (2008). HRS/EHRA Expert consensus on the monitoring of Cardiovascular Implantable Electronic Devices (CIEDs): Description of techniques, indications, personnel, frequency and ethical considerations. Europace, 10(6):707-25. xiv Saxon, L.A., Hayes, D.L., Gilliam, R., Heidenreich, P.A., Day, J., Seth, M., Meyer, T.E., Jones, P.W. & Boehmer, J.P. (2010). Long-term outcome after ICD and CRT implantation and influence of remote device follow-ip. The ALTITUDE Survival Study. Circulation, 122:2359-67. Burri, H.,Heidbüchel, H., Jung, W., & Brugada, P. Remote monitoring: a cost or an investment? Europace (2011) 13(suppl 2): ii44-ii48 doi:10.1093/europace/eur082 xv Klersy, C., De Silvestri, A., Gabutti, G., Regoli, F., & Auricchio, A. (2009). A meta-analysis of remote monitoring of heart failure patients. Journal of the American College of Cardiology, 54:1683-94. Heidbüchel, H., Lioen, P., Foulon, S., Huybrechts, W., Ector, J., Willems, R., & Ector, H. (2008). Potential role of remote monitoring for scheduled and unscheduled evaluations of patients with an implantable defibrillator Europace, 10:351-7. National Health and Hospitals Reform Commission: A Healthier Future For All Australians – final report June 2009. Canberra: Department of Health and Ageing, 2009. xvi Crossley, G.H. Chen, J., Choucair, W., Cohen, T.J., Gohn, D.C., Johnson, W.B., Kennedy, E.E., Mongeon, L., R., Serwer, G.A., Qiao, H., & WiIlkoff, B.W., for the PREFER Study Investigators. (2009). Clinical benefits of remote versus transtelephonic monitoring of implanted pacemakers. Journal of the American College of Cardiology, 54(22):2012-9. Crossley, G.H., Boyle, A., Vitense, H., Chang, Y., Mead, R.H., and CONNECT investigators. (2011). The CONNECT (Clinical Evaluation of Remote Notification to Reduce Time to Clinical Decision) Trial: The value of wireless remote monitoring with automatic clinician alerts. Journal of the American College of Cardiology, 57:1181-89. xvii Schoenfeld, M.H., Compton, S.J., Mead, R.H., Weiss, D.N., Sherfesee, L., Englund, J., & Mongeon, L.R. (2004). Remote monitoring of implantable cardioverter defibrillators: a prospective analysis. PACE - Pacing and Clinical Electrophysiology, 27(6 (Pt 1):757-63. Varma, N., Stambler, B., & Chun, S. (2005). Detection of atrial fibrillation by implanted devices with wireless data transmission capability. Pacing and Clinical Electrophysiology, 28(Suppl. 1):S133-S6. xviii Varma, N., Epstein, A.E., Irimpen, A., Schweikert, R., & Love, C. for the TRUST Investigators. (2010). Efficacy and safety of automatic remote monitoring for implantable cardioverter-defibrillator for follow-up. The Lumos-T safely reduces routine office device follow-up (TRUST) trial. Circulation, 122:325-32. xix Varma, N., Epstein, A.E., Irimpen, A., Schweikert, R., & Love, C. for the TRUST Investigators. (2010). Efficacy and safety of automatic remote monitoring for implantable cardioverter-defibrillator for follow-up. The Lumos-T safely reduces routine office device follow-up (TRUST) trial. Circulation, 122:325-32. xx Kacet, S., & Daubert, J.C. (2011). ECOST: Safety of implantable cardioverter defibrillator follow-up using remote monitoring: a randomized controlled trial. Program and abstracts of the European Society of Cardiology Congress; August 27-31, 2011; Paris, France. Session #707003 - 707004. 2011. Medical Technology Association of Australia Business case for public funding of remote monitoring of CIEDs Page 26 Mabo, P., & Auricchio, A. (2011). PURE: EVATEL: Remote follow-up of patients implanted with an ICD: the prospective randomized EVATEL study. Program and abstracts of the European Society of Cardiology Congress; August 27-31, 2011; Paris, France. Session #707001 - 707002. xxi Mabo, P. (2010). Home monitoring for pacemaker follow-up: Results of randomized COMPAS trial. France. Session Pacing and ICDs. Cardiostim, 2010; Nice, France. xxii Al-Khatib, S.M., Piccini, J.P., Knight, D., Stewart, M., Clapp-Channing, N., & Sanders, G.D. (2009). Remote Monitoring of Implantable Cardioverter Defibrillators versus Quarterly Device Interrogations in Clinic: Results from a Randomized Pilot Clinical Trial. Journal of Cardiovascular Electrophysiology,1-6. Marzegalli, M., Landolina, M., Lunati, M., Perego, G.B., Pappone, A., Guenzati, G., Campana, C., Frigerio, M., Parati, G., Curnis, A., Colangelo, I., & Valsecchi, S. (2009). Design of the evolution of management strategies of heart failure patients with implantable defibrillators (EVOLVO) study to assess the ability of remote monitoring to treat and triage patients more effectively. Trials, 10(42). xxiii Fauchier, L., Sadoul, N., Kouakam, C., Briand, F., Chauvin, M., Babuty, D., & Clementy, J. (2005). Potential cost savings by telemedicine-assisted long-term care of implantable cardioverter defibrillator recipients. PACE - Pacing and Clinical Electrophysiology, 28(Suppl 1):S255-9. xxiv Ricci, R.P., Morichelli, L., & Santini, M. (2009). Remote control of implanted devices through Home Monitoring technology improves detection and clinical management of atrial fibrillation. Europace, 11(1):54-61. Crossley, G.H., Boyle, A., Vitense, H., Chang, Y., Mead, R.H., and CONNECT investigators. (2011). The CONNECT (Clinical Evaluation of Remote Notification to Reduce Time to Clinical Decision) Trial: The value of wireless remote monitoring with automatic clinician alerts. Journal of the American College of Cardiology, 57:1181-89. Halimi, F., Clementy, J., Attuel, P., Dessenne, X., & Amara, W. (2008). Optimized postoperative surveillance of permanent pacemakers by home monitoring: the OEDIPE trial. Europace, 10(12):1392-9. Lazarus, A. (2007). Remote, wireless, ambulatory monitoring of implantable pacemakers, cardioverter defibrillators and cardiac resynchronization therapy systems: analysis of a worldwide database. PACE - Pacing and Clinical Electrophysiology, 30 (Suppl 1):S2-S12. Ricci, R.P., Morichelli, L., & Santini, M. (2008). Home monitoring remote control of pacemaker and implantable cardioverter defibrillator patients in clinical practice: Impact on medical management and health-care resource utilization. Europace, 10(2):164-70. xxv Hauck, M., Bauer, A., Voss, F., Weretka, S., Katus, H.A., & Becker, R. (2009). "Home monitoring" for early detection of implantable cardioverter-defibrillator failure: a single-center prospective observational study. Clinical Research in Cardiology, 98(1):19-24. Sacher, F., Probst, V., Bessouet, M., Wright, M., Maluski, A., Abbey, S., Bordachar, P., Deplagne, A., Ploux, S., Lande, G., Jaïs, P., Hocini, M., Haïssaguerre, M., Le Marec, H., & Clémenty, J. (2009). Remote implantable cardioverter defibrillator monitoring in a Brugada syndrome population. Europace, 11(4):489-94. xxvi Remote monitoring systems for patients with implanted cardiac devices, June 2008, MSAC application 1111. xxvii Saxon, L.A., Hayes, D.L., Gilliam, R., Heidenreich, P.A., Day, J., Seth, M., Meyer, T.E., Jones, P.W. & Boehmer, J.P. (2010). Long-term outcome after ICD and CRT implantation and influence of remote device follow-ip. The ALTITUDE Survival Study. Circulation, 122:2359-67. Varma, N., Epstein, A.E., Irimpen, A., Schweikert, R., & Love, C. for the TRUST Investigators. (2010). Efficacy and safety of automatic remote monitoring for implantable cardioverter- Medical Technology Association of Australia Business case for public funding of remote monitoring of CIEDs Page 27 defibrillator for follow-up. The Lumos-T safely reduces routine office device follow-up (TRUST) trial. Circulation, 122:325-32. Mabo, P. (2010). Home monitoring for pacemaker follow-up: Results of randomized COMPAS trial. France. Session Pacing and ICDs. Cardiostim, 2010; Nice, France. Crossley, G.H., Boyle, A., Vitense, H., Chang, Y., Mead, R.H., and CONNECT investigators. (2011). The CONNECT (Clinical Evaluation of Remote Notification to Reduce Time to Clinical Decision) Trial: The value of wireless remote monitoring with automatic clinician alerts. Journal of the American College of Cardiology, 57:1181-89. xxviii Medical Advisory Secretariat. (2011, April). Internet-based device-assisted remote monitoring of cardiovascular implantable electronic devices: an evidence-based analysis (DRAFT). Ontario Health Technology Assessment Series Retrieved from: http://www.health.gov.on.ca/english/providers/program/ohtac/draft_comment.html xxix Ecomed. (2011, 17 Feburary ). Who wants to save €5.6 billion? Telemonitoring and eHealth: more than fancy gadgetry Retrieved from http://eucomed.blogactiv.eu/2011/02/17/whowants-to-save/ xxx Crossley, G.H., Boyle, A., Vitense, H., Chang, Y., Mead, R.H., and CONNECT investigators. (2011). The CONNECT (Clinical Evaluation of Remote Notification to Reduce Time to Clinical Decision) Trial: The value of wireless remote monitoring with automatic clinician alerts. Journal of the American College of Cardiology, 57:1181-89. Schoenfeld, M.H., Compton, S.J., Mead, R.H., Weiss, D.N., Sherfesee, L., Englund, J., & Mongeon, L.R. (2004). Remote monitoring of implantable cardioverter defibrillators: a prospective analysis. PACE - Pacing and Clinical Electrophysiology, 27(6 (Pt 1):757-63. Ricci, R.P., Morichelli, L., & Santini, M. (2008). Home monitoring remote control of pacemaker and implantable cardioverter defibrillator patients in clinical practice: Impact on medical management and health-care resource utilization. Europace, 10(2):164-70. Sacher, F., Probst, V., Bessouet, M., Wright, M., Maluski, A., Abbey, S., Bordachar, P., Deplagne, A., Ploux, S., Lande, G., Jaïs, P., Hocini, M., Haïssaguerre, M., Le Marec, H., & Clémenty, J. (2009). Remote implantable cardioverter defibrillator monitoring in a Brugada syndrome population. Europace, 11(4):489-94. xxxi Ricci, R.P., Morichelli, L., & Santini, M. (2009). Remote control of implanted devices through Home Monitoring technology improves detection and clinical management of atrial fibrillation. Europace, 11(1):54-61. Crossley, G.H., Boyle, A., Vitense, H., Chang, Y., Mead, R.H., and CONNECT investigators. (2011). The CONNECT (Clinical Evaluation of Remote Notification to Reduce Time to Clinical Decision) Trial: The value of wireless remote monitoring with automatic clinician alerts. Journal of the American College of Cardiology, 57:1181-89. Varma, N., Epstein, A.E., Irimpen, A., Schweikert, R., & Love, C. for the TRUST Investigators. (2010). Efficacy and safety of automatic remote monitoring for implantable cardioverterdefibrillator for follow-up. The Lumos-T safely reduces routine office device follow-up (TRUST) trial. Circulation, 122:325-32. Kacet, S., & Daubert, J.C. (2011). ECOST: Safety of implantable cardioverter defibrillator follow-up using remote monitoring: a randomized controlled trial. Program and abstracts of the European Society of Cardiology Congress; August 27-31, 2011; Paris, France. Session #707003 - 707004. 2011. Mabo, P., & Auricchio, A. (2011). PURE: EVATEL: Remote follow-up of patients implanted with an ICD: the prospective randomized EVATEL study. Program and abstracts of the European Society of Cardiology Congress; August 27-31, 2011; Paris, France. Session #707001 - 707002. Mabo, P. (2010). Home monitoring for pacemaker follow-up: Results of randomized COMPAS trial. France. Session Pacing and ICDs. Cardiostim, 2010; Nice, France. Medical Technology Association of Australia Business case for public funding of remote monitoring of CIEDs Page 28 Al-Khatib, S.M., Piccini, J.P., Knight, D., Stewart, M., Clapp-Channing, N., & Sanders, G.D. (2009). Remote Monitoring of Implantable Cardioverter Defibrillators versus Quarterly Device Interrogations in Clinic: Results from a Randomized Pilot Clinical Trial. Journal of Cardiovascular Electrophysiology,1-6. xxxii Crossley, G.H. Chen, J., Choucair, W., Cohen, T.J., Gohn, D.C., Johnson, W.B., Kennedy, E.E., Mongeon, L., R., Serwer, G.A., Qiao, H., & WiIlkoff, B.W., for the PREFER Study Investigators. (2009). Clinical benefits of remote versus transtelephonic monitoring of implanted pacemakers. Journal of the American College of Cardiology, 54(22):2012-9. Ricci, R.P., Morichelli, L., & Santini, M. (2008). Home monitoring remote control of pacemaker and implantable cardioverter defibrillator patients in clinical practice: Impact on medical management and health-care resource utilization. Europace, 10(2):164-70. Santini, M., Ricci, R.P., Lunati, M., Landolina, M., Perego, G.B., Marzegalli, M., Schirru, M., Belvito, C., Brambilla, R., Guenzati, G., Gilardi, S., & Valsecchi, S. (2009). Remote monitoring of patients with biventricular defibrillators through the CareLink system improves clinical management of arrhythmias and heart failure episodes. Journal of Interventional Cardiac Electrophysiology, 24(1):53-61. xxxiii Klersy, C., De Silvestri, A., Gabutti, G., Raisaro, A., Curti, M., Regoli, F., & Auricchio, A. (2011). Economic impact of remote patient monitoring: an integrated economic model derived from a meta-analysis of randomized controlled trials in heart failure. European Journal of Heart Failure, 13(4):450-9. xxxiv Varma N. (2009). Therapeutic Implications of Automatic Home Monitoring of Implantable Cardiac Devices. Current Treatment Options in Cardiovascular Medicine, 11:33672. Al-Khatib, S.M., Piccini, J.P., Knight, D., Stewart, M., Clapp-Channing, N., & Sanders, G.D. (2009). Remote Monitoring of Implantable Cardioverter Defibrillators versus Quarterly Device Interrogations in Clinic: Results from a Randomized Pilot Clinical Trial. Journal of Cardiovascular Electrophysiology,1-6. Raatikainen M.J.P. Uusimaa, P., Van Ginneken, M.M.E., Janssen, J.P.G., & Linnaluoto, M. (2008). Remote monitoring of implantable cardioverter defibrillator patients: A safe, timesaving, and cost-effective means for follow-up. Europace, 10(10):1145-51. xxxv Kacet, S., & Daubert, J.C. (2011). ECOST: Safety of implantable cardioverter defibrillator follow-up using remote monitoring: a randomized controlled trial. Program and abstracts of the European Society of Cardiology Congress; August 27-31, 2011; Paris, France. Session #707003 - 707004. 2011. xxxvi Elsner, C.H., Sommer, P., Piorkowski, C., Taborsky, M., Neuser, H., Bytesnik, J., Geller, J.C., Kottkamp, H., Wiesmeth, H., & Hindricks, G. (2006). A prospective multicenter comparison trial of home monitoring against regular follow-up in MADIT II patients: additional visits and cost impact. Computational Cardiology, 33:241-4. xxxvii Fauchier L SN, Kouakam C, Briand F, Chauvin M, Babuty D, Clementy J. Potential cost savings by telemedicine-assisted long-term care of implantable cardioverter defibrillator recipients. PACE - Pacing and Clinical Electrophysiology. 2005;28(Suppl 1):S255-9. xxxviii Klersy, C., De Silvestri, A., Gabutti, G., Raisaro, A., Curti, M., Regoli, F., & Auricchio, A. (2011). Economic impact of remote patient monitoring: an integrated economic model derived from a meta-analysis of randomized controlled trials in heart failure. European Journal of Heart Failure, 13(4):450-9. xxxix Raatikainen M.J.P. Uusimaa, P., Van Ginneken, M.M.E., Janssen, J.P.G., & Linnaluoto, M. (2008). Remote monitoring of implantable cardioverter defibrillator patients: A safe, timesaving, and cost-effective means for follow-up. Europace, 10(10):1145-51. xl Crossley G.H., Boyle A., Vitense H., Chang Y., Mead R.H., and CONNECT investigators. The CONNECT (Clinical Evaluation of Remote Notification to Reduce Time to Clinical Decision) Trial: The Value of Wireless Remote Monitoring With Automatic Clinician Alerts. J. Am. Coll. Cardiol. published online Jan 19, 2011; doi:10.1016/j.jacc.2010.12.012 Medical Technology Association of Australia Business case for public funding of remote monitoring of CIEDs Page 29 Raatikainen M.J.P., et al. 2008. Remote monitoring of implantable cardioverter defibrillator patients: A safe, time-saving, and cost-effective means for follow-up. Europace, 10(10):1145– 51 Al-Khatib S.M., Piccini J.P., Knight D., Stewart M., Clapp-Channing N., Sanders G.D., 2009 Remote Monitoring of Implantable Cardioverter Defibrillators versus Quarterly Device Interrogations in Clinic: Results from a Randomized Pilot Clinical Trial. J Cardiovasc Electrophysiol (1-6) Elsner C., Sommer P., Piorkowski C. et al. 2006. A prospective multicenter comparison trial of Home Monitoring against regular follow-up in MADIT II patients: additional visits and cost impact. Comput Cardiol, 33: 241-4 xli Orlov M.V., Szombathy T., Chaudhry G.M., Haffeajee C.I., 2009. Remote Surveillance of Implantable Cardiac Devices. PACE, 32: 928-39 xlii Lazarus A.: Remote, Wireless, Ambulatory Monitoring of Implantable Pacemakers, Cardioverter Defibrillators, and Cardiac Resynchronization Therapy Systems: Analysis of a Worldwide Database. PACE 2007; 30:S2-S12. xliii Elsner CH et al. A Prospective Mulitcenter Comparison Trial of Home Monitoring against Regular Follow-up in MADIT II Patients: Additional Visits and Cost Impact. Computers in Cardiology 2006; 33: 241-244. xliv Medical_Advisory_Secretariat. Internet-based device-assisted remote monitoring of cardiovascular implantable electronic devices: an evidence-based analysis (DRAFT). 2011. xlv Brugada P. (2006). What evidence do we have to replace in-hospital implantable cardioverter defibrillator follow-up? Clinical Research in Cardiology, 95(Suppl 3):III3-III9. xlvi Ellery, S., Pakrashi, T., Paul, V., & Sack, S. (2006). Predicting mortality and rehospitalization in heart failure patients with home monitoring--the Home CARE pilot study. Clinical Research in Cardiology, 95(Suppl 3):III29-III35. xlvii Hauck, M., Bauer, A., Voss, F., Weretka, S., Katus, H.A., & Becker, R. (2009). "Home monitoring" for early detection of implantable cardioverter-defibrillator failure: a single-center prospective observational study. Clinical Research in Cardiology, 98(1):19-24. xlviii Joseph, G.K., Wilkoff, B.L., Dresing, T., Burkhardt, J., & Khaykin, Y. (2004). Remote interrogation and monitoring of implantable cardioverter defibrillators. Journal of Interventional Cardiac Electrophysiology, 11(2):161-6. xlix Lazarus, A. (2007). Remote, wireless, ambulatory monitoring of implantable pacemakers, cardioverter defibrillators and cardiac resynchronization therapy systems: analysis of a worldwide database. PACE - Pacing and Clinical Electrophysiology, 30 (Suppl 1):S2-S12. l Marzegalli, M., Lunati, M., Landolina, M., Perego, G.B., Ricci. R.P., Guenzati, G., Schirru, M., Belvito, C., Brambilla, R., Masella, C., Di Stasi, F., Valsecchi, S., & Santini, M.(2008). Remote monitoring of CRT-ICD: the multicentre Italian CareLink evaluation - ease of use, acceptance, and organizational implications. PACE - Pacing and Clinical Electrophysiology, 31(10):1259-64. li Masella, C., Zanaboni, P., Di, S.F., Gilardi, S., Ponzi, P., & Valsecchi, S. (2008). Assessment of a remote monitoring system for implantable cardioverter defibrillators. Journal of Telemedicine & Telecare, 14(6):290-4. lii Mullens, W., Oliveira, L.P.J., Verga, T., Wilkoff, B.L., & Wilson Tang, W.H. (2010). Insights from internet-based remote intrathoracic impedance monitoring as part of a heart failure disease management program. Congestive Heart Failure, 16(4):159-63. liii Nielsen, J.C., Kottkamp, H., Zabel, M., Aliot, E., Kreutzer, U., Bauer, A., Schuchert, A., Neuser, H., Schumacher, B., Schmidinger, H., Stix, G., Clémenty, J., Danilovic, D., & Hindricks G. (2008). Automatic home monitoring of implantable cardioverter defibrillators. Europace, 10(6):729-35. Medical Technology Association of Australia Business case for public funding of remote monitoring of CIEDs Page 30 liv Raatikainen M.J.P. Uusimaa, P., Van Ginneken, M.M.E., Janssen, J.P.G., & Linnaluoto, M. (2008). Remote monitoring of implantable cardioverter defibrillator patients: A safe, timesaving, and cost-effective means for follow-up. Europace, 10(10):1145-51. lv Ricci, R.P., Morichelli, L., & Santini, M. (2008). Home monitoring remote control of pacemaker and implantable cardioverter defibrillator patients in clinical practice: Impact on medical management and health-care resource utilization. Europace, 10(2):164-70. lvi Ricci, R.P., Morichelli, L., & Santini, M. (2009). Remote control of implanted devices through Home Monitoring technology improves detection and clinical management of atrial fibrillation. Europace, 11(1):54-61. lvii Ricci RP, Morichelli, L., Quarta, L., Sassi, A., Porfili, A., Laudadio, M.T., Gorgaro, A., Santini, M. Long-term patient acceptance of and satisfaction with implanted device remote monitoring. Europace. 2010;12(5):674-9. lviii Sacher, F., Probst, V., Bessouet, M., Wright, M., Maluski, A., Abbey, S., Bordachar, P., Deplagne, A., Ploux, S., Lande, G., Jaïs, P., Hocini, M., Haïssaguerre, M., Le Marec, H., & Clémenty, J. (2009). Remote implantable cardioverter defibrillator monitoring in a Brugada syndrome population. Europace, 11(4):489-94. lix Santini, M., Ricci, R.P., Lunati, M., Landolina, M., Perego, G.B., Marzegalli, M., Schirru, M., Belvito, C., Brambilla, R., Guenzati, G., Gilardi, S., & Valsecchi, S. (2009). Remote monitoring of patients with biventricular defibrillators through the CareLink system improves clinical management of arrhythmias and heart failure episodes. Journal of Interventional Cardiac Electrophysiology, 24(1):53-61. lx Saxon, L.A., Hayes, D.L., Gilliam, R., Heidenreich, P.A., Day, J., Seth, M., Meyer, T.E., Jones, P.W. & Boehmer, J.P. (2010). Long-term outcome after ICD and CRT implantation and influence of remote device follow-ip. The ALTITUDE Survival Study. Circulation, 122:2359-67. lxi Schoenfeld, M.H., Compton, S.J., Mead, R.H., Weiss, D.N., Sherfesee, L., Englund, J., & Mongeon, L.R. (2004). Remote monitoring of implantable cardioverter defibrillators: a prospective analysis. PACE - Pacing and Clinical Electrophysiology, 27(6 (Pt 1):757-63. lxii Theuns, D.A.M.J., Rivero-Ayerza, M., Knops, P., Res, J.C.J., & Jordaens, L. (2009). Analysis of 57,148 transmissions by remote monitoring of implantable cardioverter defibrillators. PACE - Pacing and Clinical Electrophysiology, 32(1). lxiii Varma, N., Stambler, B., & Chun, S. (2005). Detection of atrial fibrillation by implanted devices with wireless data transmission capability. Pacing and Clinical Electrophysiology, 28(Suppl. 1):S133-S6. lxiv Medical_Advisory_Secretariat. Internet-based device-assisted remote monitoring of cardiovascular implantable electronic devices: an evidence-based analysis (DRAFT). 2011. lxv Al-Khatib, S.M. Piccini, J.P., Knight D., Stewart M., Clapp-Channing N., & Sanders G.D. (2010). Remote monitoring of implantable cardioverter defibrillators versus quarterly device interrogations in clinic: results from a randomized pilot clinical trial. Journal of Cardiovascular Electrophysiology, 21(5):545-50. lxvi Bikou, O., Licka, M., Kathoefer, S., Katus, H.A., & Bauer, A. (2010). Cost savings and safety of ICD remote control by telephone: a prospective observational study. Journal of Telemedicine & Telecare, 16(7):403-8. lxvii Crossley, G.H. Chen, J., Choucair, W., Cohen, T.J., Gohn, D.C., Johnson, W.B., Kennedy, E.E., Mongeon, L., R., Serwer, G.A., Qiao, H., & WiIlkoff, B.W., for the PREFER Study Investigators. (2009). Clinical benefits of remote versus transtelephonic monitoring of implanted pacemakers. Journal of the American College of Cardiology, 54(22):2012-9. lxviii Crossley, G.H., Boyle, A., Vitense, H., Chang, Y., Mead, R.H., and CONNECT investigators. (2011). The CONNECT (Clinical Evaluation of Remote Notification to Reduce Time to Clinical Decision) Trial: The value of wireless remote monitoring with automatic clinician alerts. Journal of the American College of Cardiology, 57:1181-89. Medical Technology Association of Australia Business case for public funding of remote monitoring of CIEDs Page 31 lxix Elsner, C.H., Sommer, P., Piorkowski, C., Taborsky, M., Neuser, H., Bytesnik, J., Geller, J.C., Kottkamp, H., Wiesmeth, H., & Hindricks, G. (2006). A prospective multicenter comparison trial of home monitoring against regular follow-up in MADIT II patients: additional visits and cost impact. Computational Cardiology, 33:241-4. lxx Halimi, F., Clementy, J., Attuel, P., Dessenne, X., & Amara, W. (2008). Optimized postoperative surveillance of permanent pacemakers by home monitoring: the OEDIPE trial. Europace, 10(12):1392-9. lxxi Varma, N., Epstein, A.E., Irimpen, A., Schweikert, R., & Love, C. for the TRUST Investigators. (2010). Efficacy and safety of automatic remote monitoring for implantable cardioverter-defibrillator for follow-up. The Lumos-T safely reduces routine office device follow-up (TRUST) trial. Circulation, 122:325-32. lxxii Kacet, S., & Daubert, J.C. (2011). ECOST: Safety of implantable cardioverter defibrillator follow-up using remote monitoring: a randomized controlled trial. Program and abstracts of the European Society of Cardiology Congress; August 27-31, 2011; Paris, France. Session #707003 - 707004. 2011. lxxiii Mabo, P. (2010). Home monitoring for pacemaker follow-up: Results of randomized COMPAS trial. France. Session Pacing and ICDs. Cardiostim, 2010; Nice, France. lxxiv Mabo, P., & Auricchio, A. (2011). PURE: EVATEL: Remote follow-up of patients implanted with an ICD: the prospective randomized EVATEL study. Program and abstracts of the European Society of Cardiology Congress; August 27-31, 2011; Paris, France. Session #707001 - 707002. Medical Technology Association of Australia Business case for public funding of remote monitoring of CIEDs Page 32