DØ-note 4308 March 3, 2003 Characteristics of the Outer Layer Silicon Sensors for the Run IIb Silicon Detector M. Demarteau1, R. Demina2, S. Korjenevski2, F. Lehner3, R. Lipton1, H.S. Mao1, R. McCarthy4, R. Smith1 1) Fermilab, Batavia, USA 2) Kansas State University, Manhattan, USA 3) University of Zurich, Switzerland 4) State University of New York at Stony Brook, USA Abstract The characteristics of the first received prototype Hamamatsu silicon sensors for the outer layers for the Run IIb DØ silicon detector are described. The results of the electrical and mechanical characterizations indicate that the overall sensor quality is excellent. The sensors behave in all aspects very satisfactory and our specifications are well matched. 1 1. Introduction Extended running of the Tevatron at a center of mass energy approaching 2 TeV provides an enormous physics potential for the collider detectors. Our current theoretical prejudice indicates that the ability to identify heavy quarks in the final state is absolutely necessary to resolve these new physics channels. A silicon strip detector is currently the tool most suited to identify heavy flavors in proton-antiproton collisions. It is anticipated that the current silicon detector of the DØ experiment will start performing marginally at an integrated luminosity of 4 fb-1 due to radiation damage and low signal to noise ratio. It is for these reasons that the collaboration has designed a new silicon tracker [1]. The silicon tracking system will be a six layer device, divided into two radial groups. The inner two layers, covering a radius between ~18 mm and ~35 mm, will have axial readout only. These layers have a significantly reduced radius with respect to the current tracker. The outer group is comprised of layers 2 through 5. The basic building block of the outer layers is a stave. A stave is a two layer structure of silicon sensors. Silicon sensors are mounted on each side of a rohacell stave core, which has embedded cooling channels. One side of the stave will have axial readout, the other side stereo readout. The stereo angle is obtained by rotating the sensor. Each side of the stave is in turn populated with two readout modules. Closest to the interaction point is a 10-10 module, with two independent ~10 cm long readout sections. Next to it is mounted a 20-20 module, with two independent ~20 cm long readout sections. Each side of the stave thus has six silicon sensors for a total of twelve sensors per stave. Our design calls for 168 staves, for a total of 2016 outer layer sensors. This note described the results of the tests mainly done at Fermilab and KSU on a set of prototype outer layer sensors. 2. Sensor Specifications All silicon detectors are p+n type single sided sensors, with AC coupling and biased through poly-silicon resistors and need to withstand a dose of 2 x 1013 1 MeV equivalent neutrons per cm2. Only one vendor is being considered for the outer layer sensors, Hamamatsu Photonics. The silicon sensors will have a single-guard ring with peripheral n-well as designed by Hamamatsu in order to improve the high voltage stability after irradiation. The detailed specifications for the sensors can be found in [2]. Table 1 summarizes some of the main sensor characteristics. A detailed mechanical drawing of the sensors, drawing number 3823.210-ME-399565, on which the location of strips, fiducial marks, bonding and testing pads as well as the strip numbering definition and other features are indicated, is available and was provided to Hamamatsu (see Fig. 1). The sensor also has a 24-field scratch pad for unique identification. 2 Specifications: Wafer thickness Depletion voltage Leakage current Junction breakdown Implant width Al width Al strip resistivity Coupling capacitance Coupling capacitor breakdown Interstrip capacitance Polysilicon bias resistor Not working strips Active Length (mm) Active Width (mm) Cut Length (mm) Cut Width (mm) Strip Pitch (m) Readout Pitch (m) # of Readout strips Not working strips Layer 2-5 32020m, wafer warp less than m V<300V <100nA/cm2 at RT and FDV+10%V, total current < 16A at 350V >350V 8m 2-3 m overhanging metal < 20 /cm >12pF/cm >100V <1.2pF/cm 0.8 0.3 M <1% 98.33 38.34 100.00 40.34 30 60 639 <1% Table 1: Specifications for outer layer sensors 3. Prototype Sensors and Test Structures In July 2002 a set of 100 prototype sensors were ordered from Hamamatsu Photonics, which were shipped to Fermilab on November 29, 2002. The specifications of the sensors are detailed in reference 1. In Table 1 a summary of the specifications for the outer layer sensors is given. The sensors are produced on 6” wafers. Figure 2 shows a layout of the wafer. The sensors received have two different lot numbers, indicating that they were produced in two different batches. If the serial numbers on the sensors are an indication of the processing yield at the company, the yield is about 70%. In addition to the sensors, test structures are an integral part of the wafer. There are four separate test structures on a wafer of which we receive one of the two adjacent to the long side of the sensor. They are indicated by the serial numbers of the two sensors on the wafer. The test structure contains two ‘baby-sensors’. One is identical to the full-size sensor and is different only in the fact that it has eight readout strips. The other baby-sensor has no poly-silicon resistors. Adjacent to the baby-sensors are four implants with aluminization with separate contacts for each. As per our specifications, there are also silicon diodes on the test 3 Figure 1: Detailed layout of outer layer sensor 4 Figure 2: Layout of 6" Hamamatsu wafer structure, with and without guard ring structure. One area on the test structure contains a field MOS structure for flatband voltage measurements, and monitors for the implant resistance and coupling capacitance. The remaining features on the test structure are arrays of mainly poly-silicon resistors. The characterization of the sensors is divided into three distinct parts. In the next section the measurements on the test structures will be described. The section following it is the main section, describing the measurements performed on the sensors themselves. All measurements follow the procedures as outlined in reference [3]. The note concludes with a summary of the mechanical properties of the sensors. Both the test structures and sensors were also irradiated and characterized after irradiation. Those results are described in an accompanying note [4]. 4. Electrical Characterization of the Test Structures Two test structures, 35/36 and 65/66, were measured at Fermilab and test structures 67/68, 69/70 and 73/74 were measured at KSU. Four parameters are determined from measurements on the test structures: i. Coupling Capacitance 5 ii. Implant resistance iii. Aluminum strip resistance iv. Poly-silicon resistance v. Breakdown voltage of the coupling capacitor 4.1. Coupling Capacitance and Coupling Capacitor Breakdown On the test structure there is a series of four strips consisting of the p+ implant only, the coupling capacitor and the aluminization. Pads are connected to the implant (DC-pad) and the aluminum strip (AC-pad) at each end of the strip (see Figure 3). To facilitate the measurement, the corresponding pads of two adjacent strips are wirebonded at the far end of the strip. Using this configuration, the coupling capacitance of the strip, the implant resistance and the aluminum strip resistance were measured. Figure 3: Implant and Coupling Capacitor only structure on the Test Structure The capacitance of the coupling capacitor for two sets of two strips was measured to be 93 pF at a frequency of 1 kHz. This gives a value of 9.5 pF/cm. This is still large enough compared to the interstrip capacitance (see section 5.6), so that only a small capacitive crosstalk is expected. Several strips were selected on the baby sensor on the various test structures and a voltage applied across the coupling capacitor by placing a probe on the DC-pad and the AC-pad. Breakdown of the capacitor is defined as the voltage when the current reaches 100nA. Figure 4 shows the dependence of the current versus voltage for a few channels 6 on test structures 35/36 and 65/66. Test structure 65/66 breaks down at a slightly lower voltage than test structure 35/36, but both are well above our specified breakdown value of 100V. Ccc Breakdown Voltage (Test Structures) Current (nA) 4500 35/36, Ch. 8 4000 35/36, Ch. 5 3500 35/36, Ch. 3 35/36, Ch. 2 3000 65/66, Ch. 1 2500 65/66, Ch. 2 2000 65/66, Ch. 3 65/66, Ch. 6 1500 65/66, Ch. 7 1000 500 0 0 50 100 150 200 250 300 Voltage (V) Figure 4: IV-Curve for measurement of Ccc breakdown 4.2. Implant Resistance The implant resistance was measured by applying a voltage differential between two neighboring DC pads and mapping the current versus voltage. Sets of two strips were measured on test structure 35/36 and give an implant resistance of 104 k/cm. Figure 5 shows the measurement for the second set of strips. The implant resistance is not part of our specifications, but the measured value agrees with our expectations of a typical p+ doping. 7 Implant Resistance Measurement Voltage (V) 1.5 1 0.5 0 -0.5 -0.3 -0.1 0.1 0.3 0.5 -0.5 -1 -1.5 I ( A) Figure 5: Implant resistance measurement 4.3. Aluminum Strip Resistance The aluminum strip resistance was measured by applying a voltage differential between two neighboring AC pads and mapping the current versus voltage. Sets of two strips were measured on test structure 35/36 and give a resistance for the aluminization of 16 /cm, well within our specification which requires a resistivity of less than 20 /cm. Figure 6 shows the measurement for the set consisting of strips 6 and 8. The measurement on all other strips show identical results. The measured aluminum resistance corresponds to an aluminum thickness of about 1.5m. The introduced ENC noise of a total series resistance of around 160 for a strip length of 10cm in front of the preamplifier amounts to a noise contribution less than 200e. It can therefore be neglected. If the strip length is 20cm, as is the case for the 20-20 modules, the ENC noise is less than 500e. 8 Aluminum Strip Resistance Voltage (V) 1.5 Test Structure 35/36, Ch. 6-8 1 0.5 0 -0.004 -0.002 0 0.002 0.004 -0.5 -1 -1.5 I (A) Figure 6: Measurement of aluminum strip resistivity 4.4. Polysilicon Resistance 4.4.1. Strip Poly-silicon Resistance The resistance of the poly-silicon resistor was measured on the baby sensor. The baby sensor has exactly the same layout as the full strip detector. Also on the baby sensor there are intermediate and readout strips. Simply because of space constraints, the poly-silicon resistors cannot all reside on one end. The resistors alternate between strips. Consequently, the poly-silicon resistors for the intermediate strips are all at one end, and the resistors for the readout strip are all at the other end. The poly-silicon resistor value has been measured on the intermediate strips, by placing the probe on the DC-pad and the bias line and applying a negative voltage at the DC pad. Figure 7 shows the I-V curve as measured for the first channel on test structure 35/36. It yields a poly-silicon resistor value of 0.8 M. This measurement has been repeated for more channels and test structures and they all yield the same result within an error of 0.1 M. Performing the same measurement on the adjacent strip, the readout strip, still provides valuable information. Placing the probe on the DC pad measures the implant and polysilicon resistance in series. The average value of the resistance was then determined to be 2.4Mand hence rather consistent with our previous findings of polysilicon and implant resistors in series. 9 Rpoly Measurement on Test Structure 0 -3.0 -2.5 -2.0 -1.5 -1.0 -0.5 0.0 Voltage (v) -0.5 -1 -1.5 -2 -2.5 Current ( A) Figure 7: Measurement of the poly-silicon resistance for the first intermediate strip on test structure 35/36 4.4.2. Monitor Polysilicon Resistors The test structure also contains various arrays of poly-silicon resistors for monitoring purposes. These arrays are labeled ‘PSxx’, with xx a numeric identifier. The resistors labeled ‘ps20’ look identical to the strip poly-silicon resistors and were measured. Figure 8 shows one of the measurements. The measured resistance is 0.88 M. Resistance Measurement 'PS20' 0.6 Voltage (V) 0.4 0.2 0 -0.7 -0.5 -0.3 -0.1 0.1 0.3 0.5 0.7 -0.2 -0.4 -0.6 Current ( A) Figure 8: Measurement of resistance of monitoring resistor 'PS20' on test structure 10 5. Electrical Characterization of the Sensors Hamamatsu delivered 100 prototype sensors to Fermilab early in December 2002. The sensors were probed both at Fermilab and KSU. The following sets of measurements are performed on the sensors: i. I-V curve, to measure total detector leakage current as function of bias voltage up to 700V. ii. Long-term stability of detector total leakage current iii. C-V curve, to measure the total detector capacitance as function of bias voltage. The depletion voltage is extracted from this measurement iv. AC-scan, to measure the coupling capacitance and the current of individual strips. Sensor defects, most commonly shorted strips and pinholes, are identified with this measurement. v. DC-scan, to measure the leakage current of individual strips. Sensor defects, most commonly leaky strips, are identified with this measurement. vi. Interstrip isolation by measuring the interstrip resistance and measurement of resistance of bias resistor vii. Interstrip capacitance and total load capacitance Not all sensors undergo the full series of tests. Our quality assurance program calls for every sensor to undergo an IV-scan and a CV-scan. Due to scheduling pressures, at the time of writing this note, not all sensors have been submitted to an IV- and CV test yet. The emphasis was placed on mapping out a few sensors in detail. Table 3 shows a summary of which sensors were measured to date (see also section 5.7). Some of the sensors tested were irradiated and the results are described in reference [4]. Some other sensors were used to build detector modules. The results of the module readout noise tests are described in an accompanying note as well [5]. 5.1. Characterization at the Manufacturer The contract with the vendor calls for an electrical characterization of the sensors at the company. Each sensor should undergo: o I-V curve to 500 V at a temperature of 25 3C and a relative humidity of <50% o Optical inspection for defects, opens, shorts and defects, and verification of mask alignment to better than 2.5 m o Depletion voltage as determined from the C-V method o AC capacitance value measurement and pinhole determination 11 In addition, the manufacturer is asked to verify the poly-silicon resistor value, implant resistivity, coupling capacitor value and its breakdown as well as the aluminium resistance on the test structures. The vendor provides these results as average value per delivered batch (the 100 sensors came in two batches: SWA61737 and SWA61738). Our measured results on the teststructures as they are described in section 4 of this document, are in good agreement with Hamamatsu’s findings on the aforementioned quantities. Figure 9 shows, for completeness, the IV-curves for all 100 sensors as measured by Hamamatsu. There is one point that is off-scale. Sensor 103 has a leakage current of 731nA at 500V bias voltage. Although not visible in this graph, there is a slight increase in the leakage current for a fraction of sensors around 200 Volts. We will come back to this later. Total Leakage Current, Hamamatsu 500 I (nA) 400 300 200 100 0 0 100 200 300 400 500 Bias Voltage (V) Figure 9: Total detector leakage current as function of bias voltage as measured by Hamamatsu for all hundred prototype sensors All sensors have a depletion voltage between 110-130V, with a granularity of 10V. HPK measures the C-V dependence in steps of 10 Volts and defines the full depletion voltage as the lowest voltage where the increase in 1/C2 is found to be less than 2% [6]. This method tends to overestimate the depletion voltage. Table 2 lists all the sensor defects as noted by Hamamatsu. Only three sensors out of 100 (!) are flagged as having a defect. Sensor 51 and 65 have one and two occurrences of shorted aluminum strips, respectively. Sensor 108 has two shorted coupling capacitors. All three sensors were visually inspected and the shorts confirmed. 12 Lot No. Serial No. SWA61737 51 SWA61737 65 SWA61738 108 Ch. No. Type AC-AL short 20-21 AC-AL short 52-53 AC-AL short 39-40 Coupling short 94 Coupling short 95 Table 2: The defects on the three sensors as noted by Hamamatsu 5.2. IV-Scan All sensors that were tested have excellent detector leakage current. No sensor exceeded the specification of a total detector current in excess of 16A at 350V. In retrospect, it seems that this criterion was too loose and should be significantly tightened for the production order. Figure 10 shows the difference in total detector leakage current as measured at Hamamatsu and Fermilab for a subset of sensors. The difference is nearly always less than 50 nA, and independent of bias voltage. The offset in current can easily be attributed to the different environmental conditions for the measurements at the two different locations. The slight increase in current at 200V and 400V for some sensors is only present in the Hamamatsu data. Its origin is not clear. The measurements at KSU are done at a lower temperature, typically 1oC lower. HPK - FNAL Measurements 100 19 20 21 47 48 51 88 131 75 D Idet (nA) 50 25 0 -25 -50 -75 -100 0V 100V 200V 300V 400V 500V Vbias (V) Figure 10: Difference in leakage current for a subset of sensors as measured at Hamamatsu and Fermilab 13 The breakdown voltage is well above the specified 350V for all sensors. Figure 11 shows the detector currents up to a bias voltage of 700V. Most detectors show a gradual increase incurrent in the vicinity of 600V. Leakage Current 30000 19 20 25000 21 I (nA) 20000 47 48 15000 51 88 10000 131 5000 0 0 100 200 300 400 500 600 700 800 Vbias (V) Figure 11: Detector currents for a subset of sensors 5.3. Long Term Stability A long term test facility is available at Fermilab to monitor sensor behavior under bias for extended periods. The facility consists of a light and gas tight box which can accommodate up to six sensors in aluminum holders and a PC/Labview based readout system shown in figure 12. Data for sensor currents, box temperature and humidity are recorded at set intervals, typically every five minutes, during the burn-in period. Figure 13 shows the results of a burn-in of 5 layer 2 prototype sensors over a period spanning approximately six days. Detectors were biased to 150 V. After the first ten points the sensors are in dry air with a dew point below –40C. Room temperatures varied by 3.5C during the test and the currents are corrected to 20C. Some residual time dependence is visible. These variations are most prominent during periods when the temperatures change most rapidly. This residual variation appears to be due to a time lag between the air temperature recorded by the PC and the actual sensor temperature. However, no long-term drifts of the leakage currents have been observed and the sensors turned out to exhibit a rather stable bias behavior. 14 Temp, humidity HP 3456 DVM GPIB Keithley 706 switch box Keithley 485 pa meter Keithley 2410 ps Test Box with detector diodes PC Labview Figure 12: Setup for long term sensor current stability Sensor Burn-In 2/23/2003 corrected to 20 deg C 9.00E-08 8.00E-08 7.00E-08 Current 6.00E-08 5.00E-08 L2-47 4.00E-08 L2-48 3.00E-08 lL2-51 2.00E-08 L2-20 L2-21 1.00E-08 0.00E+00 0.E+00 5.E+04 1.E+05 2.E+05 2.E+05 3.E+05 3.E+05 4.E+05 4.E+05 5.E+05 5.E+05 Elapsed Tim e (sec) Figure 13: Six day burn-in results of five outer layer Hamamatsu sensors. 15 5.3. CV-Scan Depletion Voltage HPK (V) The CV-scan is used to determine the depletion voltage. The depletion voltage at our testing centers is extracted as the intercept of two straight lines in a 1/C2 - Vbias plot. Both Fermilab and KSU measure depletion voltage, which are consistently lower than Hamamatsu, which is attributed to Hamamatsu’s different method. Figure 14a shows a typical CV-curve for sensor 56. The depletion voltages as measured at the test centers and by Hamamatsu are summarized and compared in figure 14b. 700 Sensor 56 1/C2 (1/pF)2 600 500 400 Vdep = 96V 300 200 100 140 130 120 110 100 90 80 80 90 100 110 0 0 50 100 150 200 250 300 350 400 450 Depletion Voltage CV Scan (V) Vbias (V) Figure 14: a) CV-curve for sensor 56, and determination of depletion voltage; b) comparison of depletion voltage measured with the CV method and Hamamatsu measurements; the line indicates one-to-one correspondence. 5.4. AC-Scan The integrity of each strip is verified with the AC-scan. The measurement is performed by probing the AC-pad and the bias ring. A backside voltage of 20V is applied. The scan consists of two steps. First, a voltage of 80V is applied to the AC pad and the current drawn by the strip is measured. In a second step the voltage is lowered to 0V and the capacitance is measured at a low frequency either at 1kHz (KSU) or 10kHz (FNAL). The capacitance represents the coupling capacitance of the silicon oxide layer. Figure 15 shows a typical AC scan taken at 10kHz. The coupling capacitors are lower due to the frequency response of a low-pass filter consisting of coupling capacitor and implant resistor. At 1kHz the measured capacitances approach their zero-frequency limit and the so obtained values are then consistent with the teststructure measurements in section 4. Figure 16 shows the spread of the measured coupling capacitors values for 10 sensors. These data have been obtained at 1kHz and the mean value of the coupling capacitance is in agreement with our specifications. The spread (RMS-value) over the strips and sensors is less than 5%. The AC defects observed on three sensors by Hamamatsu could all be reproduced. 16 50 5.0 45 4.5 40 4.0 35 3.5 30 3.0 25 2.5 20 2.0 15 1.5 10 1.0 5 0.5 0 0.0 0 200 400 I (nA) C (pF) AC-Scan Sensor 21 (10 kHz) 600 Channel # Figure 15: AC-scan for sensor 21; the upper data set corresponds to the capacitance measurement. Figure 16: AC-scan for 10 sensors. The scans have been performed with the LCR meter at 1kHz. 17 5.5. DC-Scan A full DC-scan was also performed on most tested sensors, with the backside at full bias. No anomalies were observed. Sometimes a measurement gave a very low current reading, which normally could be attributed to a poor contact between the probe and the DC-pad. A repeat of the measurement usually confirmed that it was just a poor contact. Figure 17 gives an example of a DC-scan. DC-Scan, Sensor 28 ILeak (pA) 200 180 160 140 120 100 80 60 40 20 0 0 100 200 300 400 500 600 Channel # Figure 17: DC-scan results on sensor 28. 5.6. Polysilicon bias and interstrip resistors The DC-scans performed at KSU were also used to determine the polysilicon resistor values on all readout strips by applying a small voltage of 1V across the DC-pad and the bias rail while the backside was kept at a bias potential. Such a scan is sensitive to strip regions on the sensor with low interstrip resistances or to bias resistor inhomogeneities. Figure 18 shows the result of one scan performed on sensor 63. Since the polysilicon resistors are at the opposite end of the DC-pads, an effective series resistance of polysilicon and implant resistance is probed. The obtained values of 2.2-2.5 MOhm are in good agreement to the teststructure observations of section 4. The obtained values of 2.22.4 MOhm are in good agreement to the teststructure observations of section 4. Moreover, from the measurement of the strip resistances in Figure 18 we can conclude that the interstrip resistance is high enough (in the order of O(Gohm)) so that the strips are well isolated. 18 Figure 18: Measurement of the polysilicon resistors for sensor 63 5.7. Interstrip Capacitance The interstrip capacitance was measured extensively on sensors 47 and 88. The procedure as outlined in the Quality Assurance document [2] was followed. The setup with the equivalent circuit diagram is shown in figure 18. Two measurements are made. The first measurement involves a three probe setup. Probe 1 is placed on the AC pad of strip N; Probe 2 and 3 are placed on the AC pads of the adjacent strips, strip N-1 and N+1, respectively. The probes on strip N-1 and N+1 are tied together at the input of the LCR meter. The capacitance measured this way is, to first order, the total load capacitance the readout chip sees and is denoted by CL or C1. In a second measurement one of the probe tips adjacent to the center probe tip, is lifted. The capacitance measured in the second measurement is denoted C2. From the equivalent circuit diagram one finds that CL = C1 = 2 Ci + C b C2 = Ci + Cb . Here, Cb refers to the capacitance of the strip to the backplane and Ci is the interstrip capacitance. CL is to first order the total load capacitance on the readout chip. The second measurement can of course be performed for both neighboring strips and the results should be identical. In the results presented here, unless mentioned explicitly, the capacitance C2 is the average capacitance measured for the pair (N, N+1) and the pair (N, N-1): C2 = ( C2 (N, N-1) + C2 (N, N+1) ) / 2 . A few words about the limitations of the current setup are appropriate. The blocking capacitors used in the setup were 0.1 F. The results obtained for frequencies below 1 kHz were therefore not very stable. If measurements below 1 kHz are presented it is 19 implicit that they are accompanied with rather large error bars. The capacitance of the probes and leads was measured to be on average about 1 pF. In case ‘corrected’ results are shown, the measured capacitance of the leads, dependent on frequency of course, was subtracted from the measurement. Note that in the extraction of Ci no correction is needed. probe 2 on AC pad (strip N+1) probe 1 Ext. LCRprobe 3 on adapter meter AC pad (strip N-1) Cintrstr Cs Cs Cs Test chuck Figure 19: Setup for interstrip capacitance measurements Figure 20 shows the measurement of C1 as function of frequency for the triplet of strips (109, 110, 111) on sensor 88. At a frequency of 1 MHz, the frequency of relevance for operation with the SVX4 readout chip, the total capacitance is about 1 pF/cm. 20 Total Load Capacitance, Sensor #88 12.00 Strips 109-110-111 CL (pF) 10.00 8.00 6.00 4.00 2.00 0.00 1.0E+02 1.0E+03 1.0E+04 1.0E+05 1.0E+06 Frequency (Hz) Figure 20: Frequency dependence of C1 (see text for details) In figure 21 the frequency dependence of the capacitance on the backside bias voltage is shown. An initial backside voltage of 5 Volts was applied; the voltage was then increased to 10 Volts and then incremented in steps of 10 to 100 Volts. Figure 10 was taken at a backside bias voltage of 110 Volts, which corresponds to full depletion voltage plus 10%. Total Load Capacitance, Sensor #88 14.00 5V 10 V 12.00 20 V 30 V C L (pF) 40 V 10.00 50 V 60 V 8.00 70 V 80 V 90 V 6.00 100 V 4.00 2.00 0.00 1.0E+02 1.0E+03 1.0E+04 1.0E+05 1.0E+06 Frequency (Hz) 21 Figure 21: Frequency dependence of C1 for various backside bias voltage settings Total Load Capacitance, Sensor #88 12.00 CL (pF) 10.00 8.00 6.00 4.00 1 kHz 2.00 1 MHz 0.00 0 20 40 60 80 100 120 Backside Bias (V) Figure 22: Dependence of C1 on backside bias voltage for 1 kHz and 1 Mhz Figure 22 shows the dependence on the backside bias voltage for two frequencies. The dependence on the backside bias voltage decreases for increasing frequency. Moreover, there is a change in slope of the dependence on backside voltage at a frequency of about 5 kHz. The measurement of C2 for adjacent strips are in very good agreement, justifying the averaging of the two sets of strips to determine Ci and Cb. Figure 23 shows the relations of C2 corresponding to figures 21 and 22. The dependence of the capacitance C2 on the frequency and backside voltage is less than for the total capacitance. C2, Sensor #88 C2 (pF) 10 8 5V 10 V 20 V 30 V 40 V 50 V 60 V 70 V 80 V 90 V C2, Sensor #88 8 7 C2 (pF) 12 100 V 6 4 6 5 4 3 2 1 kHz 2 1 0 1.0E+02 1 MHz 0 1.0E+03 1.0E+04 1.0E+05 1.0E+06 Frequency (Hz) 0 20 40 60 80 100 120 Backside Bias (V) Figure 23: Dependence of C2 on frequency and backside voltage The measurements were repeated for three sets of triples, (109, 110, 111), (509, 510, 511) and (519, 520, 521), on sensor 88 and three sets of triplets on sensor 47. The results are all very consistent. Using the two equations given at the beginning of this section, the interstrip capacitance and backplane capacitance are extracted. Figure 24 shows the 22 results for the three sets of strips measured on sensor 88. The points labeled ‘background’ in the lower graph of figure 24 indicates the capacitance of the leads in the setup and ought to be subtracted from the plotted values to obtain the intrinsic backside capacitance. Interstrip Capacitance, Sensor 88 3.5 3.0 C i (pF) 2.5 2.0 1.5 1.0 Strips 109-110-111 Strips 509-510-511 Strips 519-520-521 0.5 0.0 1.0E+02 1.0E+03 1.0E+04 1.0E+05 1.0E+06 Frequency (Hz) Backside Capacitance, sensor 88 6.00 5.00 C b (pF) 4.00 3.00 2.00 1.00 0.00 1.0E+02 Strips 109-110-111 Strips 509-510-511 Strips 519-520-521 Background 1.0E+03 1.0E+04 1.0E+05 1.0E+06 Frequency (Hz) Figure 24: Interstrip and Backside capacitance as function of frequency 23 5.7. Summary Table 3 gives a summary of the measurements. If a dash is shown in a column, it indicates that the measurement was not performed for that particular sensor. The last column indicates if that sensor has been used to build a module. Strip numbers refer to strips that have been flagged as faulty by our testing procedures. The column labeled HPK strip# indicates the strip number(s) identified by Hamamatsu as having a defect. The type of defect was listed in Table 2. Two channels 290 and 299 in the AC-scan column are labeled with an asterisk. This means that a potential defect on these strips has to be confirmed, since the measured capacitance value is not conclusive and points to a bad contact between probe tip and pad. Sensor number Tested I-V scan I C-V Vdepl DCscan AC-scan (nA) @ 350V 3 5 6 8 10 19 20 21 28 29 33 47 48 51 52 54 56 57 58 59 62 63 64 65 88 131 KSU KSU KSU KSU KSU FNAL FNAL FNAL FNAL FNAL FNAL FNAL FNAL FNAL KSU KSU KSU KSU KSU KSU KSU KSU KSU KSU FNAL FNAL 31 34 34 38 88 174 132 128 ---------134 117 136 102 71 78 43 35 38 56 56 78 116 204 120 105 106 106 105 106 100 100 87 98 100 87 87 95 95 101 104 96 96 95 85 96 100 100 98 98 90 0 0 0 3 --0 0 0 0 308 0 328,567 0 0 0 0 0 0 0 0 0 0 0 0 --0 0 1 290* 1 --335 0 0 214 285,286 0 0 0 20,21,52,53 0 0 299* 0 0 0 40,512 0 0 0 --0 Comment HPK strip# Module 20-20 stereo 20-20 axial 20-20 axial 20-20 axial 20,21,52,53 irradiated irradiated irradiated irradiated 39,40 20-20 axial 10-10 stereo 10-10 stereo 10-10 axial 10-10 axial 20-20 stereo 20-20 stereo 20-20 stereo Table 3: Summary of measurements. 24 Note also in table 3, that the KSU I-V measurements have been performed at a lower temperature, so that in general the recorded leakage currents are somewhat smaller than at FNAL. All the four defects on sensor 51 which are claimed by HPK were found as well. However, we did not detect two shorted strips flagged by HPK on sensor 65 for yet unknown reasons. In addition we found out of 24 AC tested sensors nine additional bad strips (the two channels with an asterisk are not included here) on seven sensors during our AC-scans. These strips have not been discovered by HPK. 25 6. Mechanical Characterization of the Sensors The layout of the sensor as specified in drawing number 3823.210-ME-399565, provides a scratch pad field containing 6x4=24 pads. The vendor was asked to provide in three sets of 4 scratch pads a unique serial number coding for the sensors, with the rightmost pad representing the LSB. The remaining pads would be used for QC pass/fail marks. We expected the vendor to provide a binary serial number in the scratch pads. Figure 25 shows the rightmost pads for sensor 37. Hamamatsu has opted to use a binary decimal representation for the serial numbering, rather than a pure binary representation as we expected. For the production order we expect the serial numbering to be a binary representation and an additional footnote has been added to the drawing to that effect. Figure 25: Hamamatsu decimal binary serial numbering for sensor 37. A set of 10 sensors (serial numbers 66, 67, 69, 70, 71, 72, 73, 75, 76 and 77) was measured on an Optical Gage Products (OGP) coordinate measuring machine to verify the mechanical dimensions of the sensors. A reference system was established using the same convention as in the drawing with a corner fiducial defining the origin. The x-axis runs along the long side of the detector, parallel to the strip orientation; the y-axis is along the short axis, perpendicular to the strip orientation. The flatness of the sensors was measured by defining a grid of 11x11 points in x and y and measuring the z position of the top surface of the sensor. The OGP has an intrinsic z-resolution of a few m. Figure 26 shows the measurements for sensor 75. The sensor is warped along both the short and long axis. The difference between the minimum and maximum z-position on the sensor is then determined (see figure 27). The average of the ‘highest’ and ‘lowest’ point on a sensor is 67m, with a few sensors exceeding 70m. Our specifications call for a sensor warp of less than 50m, agreed to by Hamamatsu only on a best effort basis. Although the desired sensor warp is not obtained, the sensors are acceptable. 26 3D Graph L2-75 ta Z Da 0.08 0.06 0.04 0.02 0.00 50 40 30 20 10 0 0 20 120 100 80 4060 ata D X -10 Y Data Figure 26: Flatness measurements for outer layer sensor 75. Units along all axes are mm. Figure 27: Maximum variance in z for a set of ten Hamamatsu outer layer sensors. 27 The flatness data is analyzed further to see if there is a difference in sensor warp along either the x-, and y-axis. The sensor is divided into eleven slices along the x (y-) axis and the z-coordinate is plotted as function of the y- (x-) position along the sensor. The upper graphs in figure 28 show the measurements for sensor 73. The set of data points for each slice is fitted to a parabola, which fits the data points very well, and the coefficient of the quadratic term extracted. The lower two graphs in figure 28 show the values of the quadratic terms as determined by the fit for the ten sensors measured. The average values for the coefficient for the quadratic terms are -0.22 10-4 (mm-1) for the curvature along the x-axis, and -0.28 10-4 (mm-1) for the y-axis. The sensors have a slightly stronger warp along the short axis, perpendicular to the strips, than along the long axis. Figure 28: The upper two plots show the sensor z-position as function of x and y for eleven slices in y and x, respectively. The lower two graphs show the coefficient of the quadratic term, indicative of the sensor warp along the long and short axis. 28 In addition to the flatness measurements, a set of ten measurements each were taken along the two short edges of the sensor and a set of twenty measurements were taken for each of the long edges of the sensor. A straight line was then fitted to the data points. With this data four characteristics were verified: the sensor cut width and cut length, the accuracy of the cut edges, and the parallelism of the corresponding cut edges. In the determination of the cut width and length, the center point of the fitted line was used. The maximum difference between two measurements along each of the four sides for the ten sensors measured is shown in figure 29. It is obvious that the cut edges are extremely accurate. The absolute x or y positions of all edge measurements for the four edges for all ten sensors measured, are plotted in figure 30. There are in total 200 measurements for each of the two long edges, and 100 measurements for each of the short edges. Combining the measurements in x and y, gives an average cut width of the sensors of 40.348mm and an average cut length of 100.010mm, to be compared to the nominal values of 40.34mm and 100.00mm, respectively. The angle between the lines fitted to the cut edges averages 90.00 ± 0.002 degrees. All in all, the sensors are superior in all mechanical aspects, except for the sensor warp. X2m ax-X2m in (m m ) Y1m ax-Y1m in (m m ) 0.0035 0.0025 0.003 0.0025 0.002 0.002 0.0015 0.0015 0.001 0.001 0.0005 0.0005 0 0 1 2 3 4 5 6 7 8 9 10 1 2 3 4 5 6 7 8 9 10 8 9 10 X4m ax-X4m in (m m ) Y3m ax-Y3m in (m m ) 0.0025 0.0035 0.003 0.002 0.0025 0.002 0.0015 0.0015 0.001 0.001 0.0005 0.0005 0 0 1 2 3 4 5 6 7 8 9 10 1 2 3 4 5 6 7 Figure 29: Maximum difference between two measurements (in mm) along one of the four sensor cut edges. Each graph corresponds to a different cut edge. The horizontal axis denotes a sensor identifier. Y3 and X4 refer to the lines going through the origin of the coordinate system. 29 Figure 30: Absolute coordinates of x or y positions along the cut edges for ten sensors combined. Each graph corresponds to a different cut edge. 30 References [1] [2] [3] [4] [5] [6] DØ Run IIb Silicon Detector Technical Design Report, Chapter 3: Silicon Sensors Silicon Sensor Specifications for Layers 2-5, Version 3.0, April ‘02 Silicon Sensor quality assurance for the D0 RunIIb silicon detector: procedures and equipment, A. Bean et al., to be submitted as DØ note Radiation Characteristics of Hamamatsu Layer 2 Sensors for the DØ Run IIb Upgrade, T. Bolton et al., to be submitted as DØ note. Readout of Run IIb Outer Layer Silicon Sensors for the Run IIb Silicon Detector, A. Nomerotski and E. van Toerne K. Hara, Tsukuba University, private communication All documentations are available on: http://www.physik.unizh.ch/~lehnerf/dzero/run2b/prr/prr_l2.html 31