UV-VISIBLE SPECTROPHOTOMETER PNT17-PNG29 Data aprovació: Versió: Equipment identification: UV-Visible Spectrophotometer Brand: Hewlett Packard Model: 8453/G1103A Serial number: DE81801618 URV number: 002961 Provided by: Hewlett-Packard Española, S.A. Bought in: 1998 Interface with: PC, thermostatted water bath Location: University Rovira i Virgili, Chemical Engineering Department. Laboratory 222 Technical Specifications: Dimensions Weight Line Voltage Line Frequency Power consumption Ambient operating temperature Ambient non-operating temperature Humidity Operating altitude Non-operating altitude Safety standards: IEC, CSA, UL 34.4 cm wide 56.0 cm deep 18.5 cm high 16.5 kg 90-264 V AC, wide ranging capability 47-63 Hz, 220 VA, maximum 0-55ºC -40-70ºC <95%, at 25-40ºC, non-condensing Up to 2000 m Up to 4600 m for storing the instrument Installation Category II, Pollution Degree 2 Available techniques: Transmittance, Absorbance (Fixed wavelengths, spectrum/peaks, ration/equation, quantification) Kinetic measurements Thermal Denaturation Verification and Diagnostics Description of a UV-Visible Spectrophotometer: From: Beck, Shane. Across the Spectrum: Instrumentation for UV/Vis Spectrophotometry. The Scientist February 2, 1998. http://www.the-scientist.com/yr1998/feb/profile2_980202.html The measuring process of a UV-Visible spectrophotometer begins with a light source that generates light at a specific wavelength or wavelengths. Commonly, UV/Vis spectrophotometers utilize two light sources: a deuterium arc lamp for consistent intensity in the UV range (190 to 380 nm) and a tungsten-halogen lamp for consistent intensity in the visible PNG29: Manteniment i ús d’equipament científic PNT17: UV-Vis Spectrophotometer Ref PNT17-PNG29 Versió 1.0 1/9 UV-VISIBLE SPECTROPHOTOMETER PNT17-PNG29 Data aprovació: Versió: spectrum (380 to about 800 nm). Some spectrophotometers use xenon flash lamps, which offer decent intensity over the UV and visible regions. The source light is then directed to a dispersion device that causes different wavelengths of light to be dispersed at different angles. Two common dispersion devices used in UV/Vis spectrophotometers are prisms and holographic gratings. The angle of dispersion with a prism, however, can be nonlinear and sensitive to changes in temperature. Holographic gratings are glass blanks with narrow ruled grooves. The grating itself is usually coated with aluminum to create a reflecting source. Holographic gratings eliminate nonlinear dispersion and are not temperature sensitive. They do require filters, though, since light is reflected in different orders with overlapping wavelengths. Once the light has been passed through the dispersion device and the sample of interest it reaches a detector. Detectors in UV/Vis spectrophotometers come in a variety of shapes and sizes. Photomultiplier tubes are common; they provide good sensitivity throughout the UV/Visible spectral range and are highly sensitive at low light levels. Photodiodes have seen increasing use as detectors in spectrophotometers, bringing to the table a wider dynamic range. A photodiode is generally made up of a semiconductor and a capacitor to charge the semiconductor. As light hits the semiconductor, electrons flow through it, thereby lowering the charge on the capacitor. The intensity of light of the sample is proportional to the amount of charge needed to recharge the capacitor at predetermined intervals. As opposed to having single photodiodes, some spectrophotometers are composed of a photodiode array. Here, several photodiode detectors are arranged on a silicon crystal. The advantage of an array is the ability to do side-by-side readings, thus increasing speed. The entrance slit, dispersion device, and exit slit are referred to as the monochromator. Light passing through the monochromator exits as a band. The width of this band of light at half the maximum intensity is the spectral bandwidth. Bandwidth comes in to play with regard to accuracy, since the accuracy of any absorbance measurement is dependent on the ratio of the spectral bandwidth to the natural bandwidth of the substance being measured. The natural bandwidth is the width of the absorption band of the sample at half the absorption maximum. As a rule, a ratio between spectral bandwidth and natural bandwidth of 0.1 or less will generate absorbance measurements 99.5 percent accurate or better. Above this, accuracy deteriorates. In a conventional spectrophotometer, polychromatic light from the monochromator is transmitted through the sample, and the sample absorbance is determined by comparing the intensity of the light hitting the detector with just a sample blank with the intensity of light hitting the detector with the sample in place. With a diode array spectrophotometer, polychromatic light passes through the sample and is focused on the entrance slit of the polychromator (the entrance slit PNG29: Manteniment i ús d’equipament científic PNT17: UV-Vis Spectrophotometer Ref PNT17-PNG29 Versió 1.0 2/9 UV-VISIBLE SPECTROPHOTOMETER PNT17-PNG29 Data aprovació: Versió: 3/9 and dispersion device). The light is then dispersed onto the diode array with each diode measuring a portion of the spectrum. Single-beam spectrophotometers (which include conventional and diode array) are highly sensitive devices. Spectra can be acquired quickly in diode array spectrophotometers due to the single-beam design. In conventional spectrophotometers with a single beam, the blank and the sample are not measured simultaneously, and the interval between measurements-especially if measuring a broad spectrum-can subject results to lamp drift. Dual-beam spectrophotometers utilize a "chopper" that alternates the light path between the reference optical path and sample optical path to the detector at a speed that minimizes medium- or long-term effects of lamp drift. A variant of this is the split-beam design, which uses a beam splitter rather than a chopper. This allows light to be sent to the reference cell and sample cell simultaneously. Description of Our UV-Vis Spectrophotometer: The HP 8453 that we have is a diode array spectrophotometer with a graphical user interface and built-in GLP capabilities. This instrument allows better than 2 nm resolution and less than 0.03 percent stray light. It has various capabilities for standard measurement of absorbance and transmítanse as well as various modes of measurement including the kinetic mode, which allows us to measure the change in absorbance or transmítanse over time for specified wavelengths. The sample can be thermostatted. Diode array spectrophotometers are capable of acquiring complete UV/Visible absorbance spectra in as little as 100 msec. The key is that the grating of these instruments is fixed, and rather than moving the grating to acquire spectra, hundreds of detectors are placed at the exit of the monochromator. The detectors are all integrated on a single silicon chip called a photodiode array. The diodes act as capacitors that discharge in proportion to the incident light flux. The capacitance of each diode is converted to a binary word that is input to a computer. The HP 8452A simplifies the operation of the spectrophotometer even further by using a deuterium discharge lamp for the full UV and visible range, rather than a deuterium lamp for the UV and a tungsten incandescent lamp for the visible, as in done in most other instruments. The system is a single beam instrument, which means that you first run a scan on a cuvette containing just the solvent to determine the intensity of the lamp at each wavelength, Io(). Then you put in your sample in the same cuvette and scan the spectrum again. The absorbance is then calculated from the ratio of the two spectra: A() = log ( Io() / I()) The plot of A() versus is the spectrum of the solution. Operation and Measurement A. Start-up 1. Turn on the monitor, computer, and the spectrophotometer. There should not be a cuvette in the holder when you turn on the spectrophotometer since the instrument conducts its internal calibration at this time and calibration with a cuvette is erroneous. 2. Allow 15-20 minutes for the instrument to warm up before beginning measurement. 3. Clicking on the start menu, choose Programs, HP UV-Vis ChemStation, Instrument Online. The Windows based program will load. The system will then start the spectrum acquisition program and display the Main window. PNG29: Manteniment i ús d’equipament científic PNT17: UV-Vis Spectrophotometer Ref PNT17-PNG29 Versió 1.0 UV-VISIBLE SPECTROPHOTOMETER PNT17-PNG29 Data aprovació: Versió: B. Measuring Absorbance or Transmittance Spectra 1. In “Mode” on the top right of screen, pull down menu and choose “Standard”. 2. on left side of screen, choose “Task” from pull down menu (fixed wavelengths, spectrum/peaks, ratio/equation, quantification). 3. Click the “Setup” button. Specify the wavelength you wish to measure and other parameters such as data presentation and type (transmittance, absorbance). 4. Lift the cell holder lever to open the holder. Place a cuvette with your sample´s solvent in the sample compartment. Press the cell holder lever down to secure the cuvette. NOTE on CUVETTES: When using the spectrophotometer, you must first decide which type of cuvette is most suitable to your needs. If you are interested in wavelengths <300 nm, you must use a quartz crystal cuvette, since the polystyrene and plastic varieties have high noise in this range. At wavelengths >800 nm, plastic is more appropriate. There are also volume considerations. If you have very little sample, use the small quartz cuvette. When doing enzymatic assays, use the plastic 1 mL cuvettes so as not to waste reagent with the 3 mL cuvettes. 5. Click on the "Blank" menu button on the bottom left of the screen. The screen will change to the “Sample Spectra” window and the background scan will be displayed. With a polystyrene cuvette, there should be noise below 300 nm and a flat line above 300nm. If you see excessive noise, repeat the blank. If this does not solve the problem, try restarting the program and/or computer. 6. Gently clean with Kimwipe the sides of the cuvette that the light will pass through, and place the cuvette containing your sample, in the cell holder. 7. Click on the "Sample" menu button below the blank button on the bottom left of the screen. The spectrum will be displayed in a few seconds. The absorbance or transmittance at each wavelength will be displayed in the window “Sample table” below the “Sample Spectra” window. You can see a table of the absorbance or transmittance and standard deviation at each wavelength by double clicking on the spectrum itself. 8. You can measure all your samples and the spectra will overlay in the graphic window. If you wish to see only the current spectrum, click on the button “Last Spectrum” underneath the spectrum window. To return to overlaying spectra, click on “Overlay Spectra”, in the same location. 9. To save the spectrum, highlight the line containing the absorbance or transmittance information of the spectrum in the sample window, and go to File menu, “Save Samples As”. Choose file destination and an 8 character name for the file. Type in the file name you wish, however this file name should follow DOS naming conventions: 8 characters or less, no spaces, no punctuation symbols (no - / \ * . & _ , etc.), and start with a letter. 10. To print, highlight the spectrum window, and go to the File menu, then “Print”. C. To conduct an Enzymatic Assay using Kinetic Mode A spectrophotometric enzymatic assay is conducted by using a colorimetric cosubstrate of the enzyme. This means that the cosubstrate changes color or color intensity signal once acted upon by the enzyme, therefore changing the absorbance spectrum of the sample. To illustrate this point, let us take as an example the enzyme Horseradish Peroxidase (HRP). HRPactive + H2O2 → HRPinactive HRPinactive + ABTSred → HRPactive + ABTSox GREEN!!! HRP is a redox enzyme. As it catalyzes the reduction of hydrogen peroxide, it is oxidized. In order to further catalyze the reduction of hydrodrogen peroxide, it must be regenerated. The cosubstrate does this by reducing it back to its active form. ABTS is a good spectrophotometric enzymatic activity assay substrate because it turns color when it is oxidized by the inactive HRP. We can measure the activity of an HRP solution by determining how quickly it converts PNG29: Manteniment i ús d’equipament científic PNT17: UV-Vis Spectrophotometer Ref PNT17-PNG29 Versió 1.0 4/9 UV-VISIBLE SPECTROPHOTOMETER PNT17-PNG29 Data aprovació: Versió: reduced ABTS to oxidized ABTS, determined by measuring its change in absorbance over time. However, we must be careful not to relate the measured activity from the enzymatic activity assay with ABTS to the absolute activity of the enzyme, since the measured activity is also a measure of the kinetics between the enzyme and ABTS in the limiting reaction step. These assays are most useful for relative measurements of activity when using the same cosubstrate across all assays. There are several HRP cosubstrates we can use for enzymatic assays. There are also many enzymes whose activity can be measured with a spectrophotometrically active cosubstrate. Some examples include glucose oxidase and alkaline phosphatase, other commonly used enzymes in our lab. One good source of enzymatic assay protocols is the Sigma-Aldrich website. 1. To conduct an enzymatic assay, change the mode of the instrument to “Kinetic” in pull down menu window on the top right of the screen. 2. Blank the spectrophotometer as described above. 3. Measure a blank tailored to your assay in kinetic mode (usually contains all components of assay except the enzyme). 4. Measure your samples following the protocol. The data of interest is the absorbance change over the measured time. You can use this value to calculated the units of enzyme per mL, per mg, etc. Training Exercise Conduct an HRP assay with a range of concentrations of peroxide to find the saturated substrate condition. Calibration The HP 8453 conducts a wavelength calibration during the measurement and data processing of each sample. It does this by using the two emission lines from the deuterium lamp as calibration parameters. Performance Verification For the complete version, please refer to the manual “Operational Qualification/Performance Verification for HP 8453 UV-Visible Spectroscopy Systems”, pages 18-67. There are 7 tests that can be conducted on the spectrophotometer to verify its performance within the error laid out by published specifications. These tests are the photometric accuracy test, the wavelength accuracy test, the stray light test, the resolution test, the baseline flatness test, the photometric noise test, and the photometric stability test. The first four tests are done through exterior standards, and the last three though the software. The complete testing procedure can be completed within three hours. The procedure can be conducted by a qualified professional or by a worker with sufficient expertise. For details, please refer to the manual cited above. It is recommended that the OQ/PV procedure be conducted once a year, immediately after conducting user maintenance. It is important that the instrument be allowed 2 hours to warm up previous to the procedure. If the OQ/PV tests do not yield the published specifications, HP should be contacted immediately. PNG29: Manteniment i ús d’equipament científic PNT17: UV-Vis Spectrophotometer Ref PNT17-PNG29 Versió 1.0 5/9 UV-VISIBLE SPECTROPHOTOMETER PNT17-PNG29 Data aprovació: Versió: Maintenance and servicing: For the complete version, please refer to the HP manual “Reference Manual”, pages 94-134. This section of the manual describes maintenance procedures such as cleaning the instrument, exchanging the deuterium and tungsten lamps, and cleaning lenses. ALWAYS DISCONNECT THE INSTRUMENT FROM LINE POWER BEFORE MAINTENANCE. It is recommended especially that the spectrograph lens and the sources lens be cleaned once a year. The procedure can be found on pages 101-104 of this manual. The spectrophotometer case and sample compartment should be kept clean. Cleaning should be done with a soft cloth slightly dampened with water or a solution of water and a mild detergent. Do not let liquid drip into the instrument. Exchange the deuterium or tungsten lamp when the intensity test falls below the specified level or when one of the lamps no longer ignites. PNG29: Manteniment i ús d’equipament científic PNT17: UV-Vis Spectrophotometer Ref PNT17-PNG29 Versió 1.0 6/9 UV-VISIBLE SPECTROPHOTOMETER PNT17-PNG29 Data aprovació: Versió: Elaborat Revisat Aprovat Kristin Horton Data: 04/05/02 Comissió de Qualitat Data: Ioanis Katakis Director del Centre Data: 7/9 DOCUMENTACIÓ DE REFERÈNCIA Tipus Document Referència Títol Manual HP 8453 UV-Visible Spectrophotometer Reference Handbook Manual Understanding Your Advanced Software Manual Understanding your Biochemical Analysis Software Manual Installing Your UV-Visible Spectroscopy System Operational Qualification/Performance Verification for HP 8453 UVVisible Spectroscopy Systems Understanding Your UV-Visible Spectroscopy System Manual Manual QS QS01-PNT17 USE TRACKING QS QS02-PNT17 CALIBRATION TRACKING HISTÒRIC DE REVISIONS DE L’EQUIP DATA Punt revisat Descripció revisió PNG29: Manteniment i ús d’equipament científic PNT17: UV-Vis Spectrophotometer Ref PNT17-PNG29 Versió 1.0 GARANTIA UV-VISIBLE SPECTROPHOTOMETER PNT17-PNG29 Data aprovació: Versió: USE TRACKING (QS01-PNT17) DATE NAME PNG29: Manteniment i ús d’equipament científic PNT17: UV-Vis Spectrophotometer Ref PNT17-PNG29 Versió 1.0 TIME USED COMMENTS 8/9 UV-VISIBLE SPECTROPHOTOMETER PNT17-PNG29 Data aprovació: Versió: CALIBRATION TRACKING (QS02-PNT17) DATE NAME PNG29: Manteniment i ús d’equipament científic PNT17: UV-Vis Spectrophotometer Ref PNT17-PNG29 Versió 1.0 TEST OK COMMENTS 9/9