AirPurge™ System Quick Operating Instructions READ THE INSTRUCTIONS BELOW BEFORE USE This document is provided for reference purposes only. It is not a substitute for the Operator’s Manual. Caution: All users must be trained AirPurge System = AirPurge Device (“Device”) and Disposable Cartridge (“Cartridge”) May be used in the Operating Room or post anesthesia care area, with blood, blood components and intravenous infusions only. Do not use with arterial lines. May be used with pressure cuffs and/or fluid warmers only. Do not use with any other device. Inject critical drugs and contrast media downstream from the Device. Follow the setup below. Do not assemble Diposable Cartridge with I.V. lines prior to insertion into the Device. AirPurge System Setup (see Page 3 for Front Panel detail): Step 1: Mount the Device on an I.V. pole in a vertical position. Plug into an Green inlet cap cap Micro Bubble Chamber AC outlet (for non-battery operation). Fill level lines Red Retaining clip Blue patient side cap One Way Check valve Step 2: (Figure 1) Remove single use Cartridge from sterile pouch. Collection Bag connector must be tight; Blue and Green caps mounted in place. Step 3: (Figure 1) Remove red retaining clip; and discard. Open the Device door. With FRONT of the Cartridge facing you, place Cartridge on pegs in the device (Figure 2). Close; latch the door. Caution: Only use I.V. lines with Luer Lock connectors. When connecting the I.V. lines to the top and bottom of the Cartridge, do not over tighten Luer connections. Tighten connections only until slight resistance is felt. Figure 1 Collection Step 4: Remove Green inlet cap on top of the Cartridge and connect the bag primary I.V. to the connector of the Cartridge marked IN. Remove Blue Figure 2 outlet cap on the bottom of the Cartridge and connect the extension set to the connector at the bottom of the Cartridge marked OUT. Ensure there are no kinks in any tubing. button. Five (5) lights will illuminate. The “ON” button and 3 Battery lights will remain lit. The “PURGING” light only stays lit while purging. If any other Indicator lights stay on or there is an audible alarm, refer to page 3. Step 5: To turn on, push the Step 6: Start the fluid flow of the primary I.V line. The Cartridge will fill with I.V. fluid; any air in the line will be detected and diverted into the Collection bag. When free of air, flow will then be diverted back to the extension line. Warning – The Device cannot purge air from lines in the connected I.V. Extension Set between the AirPurge™ System and the patient. The I.V. set must be purged before the I.V. is flowing to the patient. Failure to follow the above warning could result in the introduction of air to the patient. Introduction of air to the patient could result in death or serious injury. Page 1 of 3 PN 1167 L rev F AirPurge™ System Quick Operating Instructions Once air free I.V. fluid clears the extension line, connect to the patient. Use with a fluid warmer: If the AirPurge System is to be used with a fluid warmer, please refer to the fluid warmer’s Instructions For Use. The Device should be connected distal to the fluid warmer with standard extension I.V. sets. Monitoring the System during a procedure: Caution: The AirPurge System is not a replacement for human vigilance. I.V. lines should always be monitored for signs of entrapped air. • Monitor the I.V. setup and the Device during use. • Monitor the Collection bag. If the Collection bag is full, replace the entire I.V. circuit (including Cartridge). Warning: If the Device is in a continuous “PURGING” mode and all I.V. lines and the Micro Bubble Chamber are filled with infusate, discontinue use of the AirPurge System – see directions below. Warning: When “TURN OFF – OPEN DOOR” or “TEMP” light/alarm is on discontinue use of the Device – see below. Do not attempt to troubleshoot. Return to the manufacturer. To discontinue use of the Device: Caution: Do not disconnect I.V. lines; fluid will leak from the Cartridge connection (Figure 3). This may require treatment as a Biohazard. The Cartridge should remain as part of the I.V. system Push the Open the door on the Device, and partially release the cartridge. The Device is now out of the I.V. circuit. Infusate will continue to the flow to the patient, but without air monitoring. Air must be monitored visually. When discarding, follow your institution’s procedures for disposing of hazardous biological material. (”OFF”) button. Tubing Holder Figure 3 Page 2 of 3 PN 1167 L rev F AirPurge™ System Quick Operating Instructions A AirPurge Device: Front Panel Detail (Figure 4) A: Battery Charge Indicator: Illuminated Lights, Green, Yellow, Orange = The AC adapter on the back of the Air-Purge has been plugged into an AC outlet. Green Light illuminated = Battery fully charged. Yellow Light illuminated = Battery will need charging. Orange Light blinking = Battery needs immediate charging. Plug into AC outlet. The battery alarm will continue until the battery is depleted. When the battery can no longer sustain system functionality, the green “ON” button indicator light will no longer be illuminated and the device will not remove (purge) air bubbles. B B: Indicator Lights/Alarms: TURN OFF – OPEN DOOR DOOR PURGING VERT TEMP Figure 4 C Internal diagnostics detect a system fault; alarm will also sound: DISCONTINUE USE. Door not securely closed or the Cartridge not seated correctly; alarm will also sound. Close and re-latch door. Illuminates during normal operation of the Device when actively purging air from the I.V. line (the patient is not being infused). The Device will alarm after Purging for more than one (1) minute. The Device is not vertical; alarm will also sound. Remount vertically Internal preset temperature limit is exceeded; alarm will also sound. DISCONTINUE USE. . C: Controls To turn the Device on, press this symbol To turn the Device off, press this symbol To manually purge the line, press and hold the “purge” symbol. Release symbol to stop manual purge. PN 1167 L rev F Anesthesia Safety Products 300 TradeCenter, Suite 5400 Woburn, MA 01801 Toll Free (866) 933-6550 Tel: (781) 933-6500 Fax: (781) 933-6515 www.ASPAMERICA.com Page 3 of 3