Annex A (normative) Optical Characterization A.1 General The manufacturer shall characterize the optical design of the EDF IOL and a monofocal IOL of the same distance power, against which the optical performance of the EDF IOL is to be compared, as described below. This characterization includes measurements of the optical performance of the EDF IOL that will assist in predicting performance of an individual with the EDF IOL implanted in their eye on visual acuity testing as specified in 10.2. In particular, this characterization will assess the optical resolution of the EDF IOL and the comparison monofocal IOL when used to view distant objects. It will assess the optical resolution of the EDF IOL and the comparison monofocal IOL when viewing objects at increasing near distances. A.2 Optical characterization A.2.1 Theoretical evaluation The manufacturer shall perform a theoretical evaluation (e.g., ray trace evaluation) of the optical design demonstrating the additional depth of focus of the EDF IOL compared to a monofocal IOL in an eye model with a physiologically appropriate cornea and white light at 3, and 4.5 mm apertures. An eye model cornea considered to be appropriate is described in Annex C. The theoretical evaluation shall include calculation of the modulation transfer function of the eye under entering vergence conditions, the conditions of lens tilt and de-centration specified in A.2.2.1. Using the calculated modulation transfer function data, the expected performance of the implanted lens during visual acuity testing shall be assessed with the method of A.7. A.2.2 Optical testing The following testing shall be performed on the EDF IOL and on the comparison monofocal IOL: A.2.2.1 Modulation transfer function (MTF) testing For this MTF testing, use 10 representative samples each of low, medium and high power for throughfrequency-response MTF testing, and 10 representative samples of medium power for through-focusresponse MTF testing. Modulation transfer function testing shall be performed with the lens placed in a model eye, as specified in Annex C or equivalent. The modulation transfer function shall be measured using the method specified in Annex C or equivalent. The modulation transfer function shall be measured at a fixed plane within the model eye, this position determined and set as specified in A.3. The modulation transfer function shall be measured for each of the following vergences of light entering the model eye from the target. - Range 0 D to -2.5 D - Increment or vergence step 0.25 D The subtended angle of the target features shall be the same for each step within the vergence range. Annex C gives an optical method to fulfill these vergence and target requirements. The modulation transfer function shall be measured over the entering vergence dioptric range specified above with a model eye having a pupil aperture of 3 mm and again with a model eye having a pupil aperture of 4.5 mm. The modulation transfer function shall be measured for the case of entering target vergence of 0 D with the lens centered in 4 mm diameter pupil aperture, de-centered 1 mm and centered but titled 5 degrees with respect to the pupil plane. The modulation transfer function measurements shall be done using white light as specified in A.4. A.2.2.2 Simulated depth of focus performance The simulated depth of focus performance of the EDF IOLs and the control monofocal IOL shall be determined in the following way. 1) With the IOL placed in the model test eye specified in Annex C or equivalent and with the vergence of the light entering the model eye from the test target equal to 0D adjust the image plane for best focus as specified in A.3. The test target shall be a 0.2 logMAR optotype. 2) Change the vergence of light entering the test eye in - 0.25 D steps while keeping the subtended size of test target constant until the image of the test target formed by the model can no longer be resolved. 3) The total negative vergence change to the end point of 2) is the simulated depth of focus comparison value that is used to compare the depth of focus value of the EDF IOL to the control monofocal IOL. Annex C gives an optical method that will fulfill the vergence and target requirements of A.2.2.2 A.2.2.3 Spurious imaging testing The lens shall be tested while in the model eye for the creation of spurious images, such as halos, various types of glare or secondary images, by examining the image of a ‘point source’ of light for concentrations of light other than the expected primary image. The lens shall be centered in the pupil aperture of the model eye, whose diameter is 4.5 mm, during testing. The vergence of light entering the eye from the point source of light shall be 0 D. Polychromatic, i.e. white light shall be used for the test. The image capture system shall be able to examine the image formed by the lens within the model eye over a visual angle of 5 degrees. The image capture system can be moved laterally, if needed, to cover this range. The image capture system shall be able to record image intensities over a range of so that spurious images that can be seen by a human observer at night but not under daylight conditions can be identified. Can be done, for instance, by using a neutral density filter to reduce the light from the target in conjunction with increasing the imaging camera gain. Spurious images observed to form shall be recorded by noting their relative visual angular position with respect to the primary image and their intensity relative to that of the primary image. The point source target shall subtend an angle of no more than 0.35 arc minutes. A.3 Determination of distance image plane within the model eye The distance image plane is defined as the plane within the model eye where the best image of the target forms when the vergence of light from the target, as it enters the model eye is 0 D. However, in the case of EDF-IOLs, where a very good image of a 0 D target forms over an axial range within the model eye, a method is needed to fix a single axial position for the image capture plane. Note that a microscope system for image transfer will have a very short depth of focus so object plane of the camera system is well defined. The distance image plane will therefore be experimentally found by first insuring that the target vergence is 0 D as it enters the model eye. While using a point target and while observing the image formed by the camera system, the distance between the camera system and the model eye is varied while noting the peak intensity of the captured image. The model eye is moved until the intensity is observed to take a maximum value. Then the camera system is moved away from the model eye until the peak intensity falls to 80% of the maximum peak intensity previous found. This procedure is equivalent to the procedure used in subjective eye refraction of increasing plus spherical power until blur is positively detected. This is done to insure that there is not an axial position giving good image resolution beyond the selected distance axial position. Finally the model eye is moved toward camera system until the peak intensity returns to 95 % of the maximum peak intensity found. A.4 White light For the purposes of this Annex white light is defined a polychromatic radiation covering at least the range of visible light, i.e. 380 nm to 700 nm, that has a spectral signature similar that used for illumination of visual acuity targets. An acceptable source of such radiation is a thermal source such as a quartz halogen lamp with a color temperature of at least 2800 K. A.5 Image resolution analysis using modulation transfer function measurement data The expected visual acuity for a given testing condition is found by using the measured modulation transfer function in the following way. Each modulation transfer function value whose spatial frequency is between 10 and 50 cycles per millimeter is first divided by its spatial frequency to form an array of weighted modulation transfer function values. The sum of the values in this array is found and multiplied by the spatial frequency increment squared of the measured data. This value is the weighted modulation transfer volume and it is well correlated to performance during visual acuity testing when the test IOL is implanted in a human eye and testing is done with an acuity target a distance such that the vergence of light entering the eye is the same as was the case for the model eye when the modulation transfer function was measured. The expected visual acuity value, given in logMAR units, is given by Eq. A.2. (to be added)