MRI Image Quality Assessment

advertisement

MRI Image Quality Assessment

David Collins

CR-UK Cancer Imaging Centre, The Institute of Cancer Research

Making the discoveries that defeat cancer in partnership with

Overview

•   Current Practice

•   Quality Assurance and Acceptance testing

•   Emerging applications

•   Quantitative DCE-MRI, DWI

•   Whole Body

Acceptance Testing Current Practice

We measure 1, 2 :

Resonance Frequency

Signal to Noise

Spatial Linearity

Spatial Resolution

Slice Thickness

Slice Position/Separation

Image Uniformity

Artifacts; Phase and other ghosts

Many others

Dynamic and Quantitative Measurements, are the above sufficient?

1. Price et al Med Phys 17(2) Mar/Apr/ 1990 2. Och Med Phys 19(1) Jan/Feb 1992

Acceptance Testing Current Practice

We rarely measure as specified 1, 2 :

•   Eddy Currents (sequence programming or integrator circuit

required)

•   Slice Profiles (1D FT method)

•   Nutation Angle calibration (vendor restrictions on user)

•   SNR (Modern scanners have many Rx channels 72-128)

•   B0 (Phase images not produced), spectroscopy not always an

option, vendors may provide field maps.

•   B1 uniformity

1. Price et al Med Phys 17(2) Mar/Apr/ 1990 2. Och Med Phys 19(1) Jan/Feb 1992

Acceptance Testing Current Practice

We rarely measure as specified 1, 2 :

•   Eddy Currents (sequence programming or integrator circuit

required)

•   Slice Profiles (1D FT method)

•   Nutation Angle calibration (vendor restrictions on user)

•   SNR (Modern scanners have many Rx channels 72-128)

•   B0 (Phase images not produced), spectroscopy not always an

option, vendors do provide field maps.

•   B1 uniformity

We rely on vendor installation tests and procedures, it is now vitally important to engage a site physicist in this process

1. Price et al Med Phys 17(2) Mar/Apr/ 1990 2. Och Med Phys 19(1) Jan/Feb 1992

Acceptance Testing Current Practice

•   How long should we spend performing acceptance testing?

•   At the Royal Marsden/ICR we allocate one week.

•   At least two senior clinical scientists with trainees.

•   We use a range of vendor, Eurospin test objects and home built test objects.

•   Some sites (DGH) perform zero testing.

Examples of gradient linearity correction

•   Previously gradient linearity required extensive evaluation and correction if MR images were used for R/T planning 1

•   Vendors now provide gradient linearity correction as either a default or selection option

1. Doran et al. Medical Physics 2005

Coronal

Orientation

Geometrical Distortions in Spin Echo

Sequences

NSA=1, Slice thickness=5mm NSA=2, Slice thickness=5mm

Sequence

5mm, 1NSA x y

5mm, 2NSA x y

3mm, 2NSA

3mm, 2NSA, without Dist Cor

3mm, 2NSA, with

Body Coil x y x x y y

NSA=2, Slice thickness=3mm NSA=2, Slice thickness=3mm,

Without Distortion Correction

% deviation

0.42

0.42

0.83

0.42

1.25

0.42

0.83

0

0.83

0

NSA=2, Slice thickness=3mm

With Body Coil

Custom-built Quality Assurance equipment 1 .

3T Achieva, without and with

3D distortion correction 2 .

Test object dimensions:

440mm LR, 400mm HF, 250mm AP

1.Tanner et al, Phys Med Biol 2000 Aug;45(8):2117-32 2.Doran et al. Medical Physics 2005

Gradient Linearity QA Test Object

Rapid 3D DCE-MRI protocol coronal acquisition

No Gradient Correction Vendor Gradient Correction

Dimensions 28 x 28cm Virtualscopics test object

Gradient Linearity QA Test Object

Rapid 3D DCE-MRI protocol coronal acquisition

No Gradient Correction Vendor Gradient

Correction

Dimensions 28 x 28cm Virtualscopics test object

SNR

•   Well established tests for conventional Spin Echo.

•   In practice single spin echo rarely used except acceptance testing

•   Parallel imaging is now ubiquitous in clinical protocols.

•   How do we assess SNR is a protocol specific way?

•   Is it a practical measure with multiple channels and flexible coils?

Evaluation of SNR with G-noise

Sum of Squares Adaptive Combine

Mean 826 Mean 894

2D EPI, parallel imaging factor 2, different signal combination options

All other measurement parameters fixed.

Evaluation of G-noise

Sum of Squares Adaptive Combine

Mean 22.278 Mean 6.187 snr = 36.25 snr =144.49

Note the spatial dependence of the noise distribution

Images acquired off resonance, alternatively transmitter at 0v.

Changing Role of MRI

•   Moving from purely morphological to include functional quantitative imaging

•   Moving from regional to whole body or large field of view imaging

32 Receiver channels

72 CP receiver elements in this configuration.

Functional Measures

Functional MRI (fMRI)

•   1-10+ minutes

Dynamic susceptibility contrast (DSC-MRI)

•   1-2 minutes

Dynamic Contrast Enhanced (DCE-MRI)

•   4-8 minutes

Diffusion Weighted Imaging (DWI)

•   1-30 minutes

Functional Measures

Functional MRI (fMRI)

•   1-10+ minutes

Dynamic susceptibility contrast (DSC-MRI)

•   1-2 minutes

Dynamic Contrast Enhanced (DCE-MRI)

•   4-8 minutes

Diffusion Weighted Imaging (DWI)

•   1-30 minutes

3 of the above typically use Echo Planar Imaging readout

Requirements for Functional

Measures

•   Temporal stability

•   Rapid data acquisition, high parallel imaging factors

•   Eddy Current compensation

•   Uniformity of both B0 and B1 +

DCE-MRI Protocol Schematic

MR images using different

T1 contrast

Dynamic Series

Contrast agent injection

Measurement of T

1

pre-injection Measurement of T

1

post-injection

Schematic diagram of the data acquisition

Quality Control for DCE-MRI

Acquisitions

Dynamic contrast enhanced imaging requires an accurate and precise measurement of T1 changes to model pharmacokinetic behaviour.

T1 measurements can be affected by many factors including

•   Slice profile

•   Scanner stability (RF and gradients)

•   B1 variations

•   K-space sampling

Evaluate how these factors can effect T1 measurements with a series of phantom experiments at 1.5T and 3.0T.

Phantoms

What should be assessed

•  

•  

•  

•  

•  

•  

∆ Signal(t)

∆ T1(t)

∆ [Gd](t) in true “dynamic” phantom that emulates tissue enhancement and is well characterized? … extremely difficult!

Range of T1 / T2 values of well characterized materials in “static” phantom. … feasible!

Spatial uniformity of transmit (B1) field

System SNR under test and clinical DCE scan conditions

CNR under DCE scan conditions

Repeatability and reproducibility

T1 Measurement Accuracy

Phased array coil

Sequence parameters of T1 measurement methods

•   Inversion recovery turbo-FLASH method

•   TR/TE/ θ /NSA = 10 000 ms/1.14 ms/2°/1

•   39 inversion times varying from 112 – 9000 ms

•   Variable flip angle method

•   3D GRE sequence

•   TR/TE/NSA = 4.36 ms/1.36 ms/3

•   3 varying θ s = 2°, 8° and 12°

•   Measurements were acquired with GRAPPA (iPAT = 2) and without GRAPPA

Siemens, Avanto (1.5T)

A quick, easy and accurate method of T1 measurement is the variable flip angle (VFA) method 1 where gradient echo MR images are acquired using different flip angles,

α i and T1 is calculated from the gradient of the linear plot of Si/sin( α i) vs. Si/tan( α i)

A range of [Gd] doped water phantoms with T1 values ranging from 55ms-2865ms were evaluated .

No major difference in T1 values obtained with and without partial parallel imaging

1. Fram et al Magn Reson Imaging. 1987;5(3):201-8.

Accuracy of Variable Flip Angle Method at 1.5T

T1 Map (200-1300)ms )

The accuracy and dynamic stability of clinical

DCE-MRI protocols are assessed with phantom measurements

Eurospin test object

•   Eurospin TO5 with 12 gels (T1’s ranging from 200 to 1300ms) were measured.

•   In each experiment the phantom temperature was equilibrated and recorded.

•   Dynamic stability was evaluated

GRE sequences frequently exhibited an initial (40 second to 2 minutes) signal drift in T1W acquisitions.

The signal drift in T1W images translates to 8-12% change in calculated T1 values

Stability of DCE-MRI measurements

Dynamic acquisitions may have a 2% T1 variation over the initial 10-20 seconds.

Calculated T1 values across a dynamic time series had different dynamic variability depending on the sampling scheme used.

Standard deviations ranged from 11.5ms

(elliptical encoding) to 4.7ms (elliptical encoding with phase and slice undersampling).

1.5T

No signal drift has been observed at 3T and dynamic variation is comparable with experiments at 1.5T.

The standard deviation of 3D GRE sequences was reduced from 12.4ms to

6.4ms with the use of phase stabilisation.

Eurospin test object is temperature sensitive ~3% change in T1 with 1 o C (20-23 o C)

3.0T

GRE Slice profiles

2° 5°

Slice profiles obtained from measurements at different nutation angles using a silicone oil test object.

12° 18°

Slice profile results can be used to provide correction terms within the post-processing software to correct these errors.

Essential to ensure that the slice positioning is centred on the region of interest.

Brookes JA et al; J Magn Reson Imaging. 1999 Feb;9(2):163-71.

24°

Sample T1 = 740ms

Inhomogeneous Transmit RF (B

1

) Field

RF transmit coils produce non-uniform field strengths

 FAs vary over the field of view

% dT

1 vs. T

1

40% error in α

30% error in α

20% error in α dT

1

≈ 2 × d α

10% error in α

T

1

(ms)

B

1

∞ θ  Correct δ T1 by performing in-vivo B

1

mapping

Sung et al Magn Reson Med. 2013 Oct;70(4):954-61

B1 Correction @3.0T

1

Typical value for breast at 3.0T

ROI

Fatty tissue ROI

T1 = 77ms

.

Original T1 estimate nutation angles 3 o and 16 o

1. Linderman et al JMRI (40); 171-80 2014 Data courtesy of Elizabeth O’Flynn

DCE-MRI QA Assessments

•   Noise Factor evaluated from full

DCE-MRI protocol

•   Useful for comparing protocols

•   Can be applied to pre-enhanced clinical data

•   SNR evaluated from full DCE-

MRI protocol

•   Useful for comparing protocols

•   Useful for routine QA

Kurland RJ MRM 2;136-158 1985 Imran J et al MRI 17;1347-1356 1999

What is diffusion MR?

Signal is exponentially dependent on the ADC and the sequence parameters (b-value):

b = 0 s mm -2 b = 50 s mm -2 b = 100 s mm -2 b = 250 s mm -2 b = 500 s mm -2 b = 750 s mm -2

ADC map

Diffusion Weighted Imaging Quality

Assurance

•   Use a range of phantoms:

•   ADC accuracy use Sucrose, PVP (temperature sensitive)

•   Ice water, temperature insensitive (long preparation >60mins)

•   Silicone oil, non diffusing material (artefact assessments)

•   Flood phantom ADC linearity

•   Fat/Water test object (evaluate fat suppression efficiency)

•   Others; alkanes for ADC, standard geometrical test objects

Diffusion Encoding pulse schemes

Unipolar Diffusion encoding scheme

Stejskal-Tanner

Scanning with shorter TE

Bipolar Diffusion encoding scheme

Reduction of eddy currents induced spatial distortions 1

1. Chan et al J Magn Reson. 2014 Jul;244:74-84

EPI N/2 Ghosting

Axial images of a PDMS phantom acquired using receive bandwidths (a) 1628

Hx/pixel, (b) 1776 Hx/pixel, (c) 1860 Hx/pixel, (d) 2170 Hx/pixel, (e) 2298 Hx/ pixel, (f) 2442 Hx/pixel, (g) 2790 Hx/pixel, (h) 3004 Hx/pixel b = 0 s mm -2 ).

Images have been windowed to enhance the visibility of the ghosts, at the same window width and level.

Target to reduce ghosting below 3%

EPI N/2 Ghosting

Axial images of a PDMS phantom acquired using receive bandwidths (a) 1628

Hx/pixel, (b) 1776 Hx/pixel, (c) 1860 Hx/pixel, (d) 2170 Hx/pixel, (e) 2298 Hx/ pixel, (f) 2442 Hx/pixel, (g) 2790 Hx/pixel, (h) 3004 Hx/pixel b = 0 s mm -2 ).

Images have been windowed to enhance the visibility of the ghosts, at the same window width and level.

What bandwidth should we use for N/2 ghost assessment ?

Optimization of protocol critical to success

Eddy currents: image distortion

b0 b0 – b1000

: b0

(background)

754 Hz/px 1450 Hz/px 2012 Hz/px 2416 Hz/px

Optimization of protocol critical to success

Eddy currents: image distortion

b0 b0 – b1000

: b0

(background)

754 Hz/px 1450 Hz/px 2012 Hz/px 2416 Hz/px

(a) monopolar

Axial images of a PDMS phantom

(a, b) b = 0 s mm -2 , (c, d) b = 1000 s mm -2 and (e, f) subtraction images. Images (a, c) acquired using a monopolar sequence and (b, d) using DSE

Subtraction images show the b = 0 s mm -2 image subtracted from the b = 1000 s mm -2 image.

(c)

(b) DSE

(d)

(e) (f)

8000

6000

4000

2000

0

5000

8000

6000

4000

2000

0

0

ï 5000

Quality Assurance ADC Accuracy

ADC water @0 o C = 1.1

× 10 − 3 mm 2 /s

Malyarenko D et al J Magn Reson Imaging. 2013 May;37(5):1238-46

Polyvinylpyrrolidone (PVP) phantom

The cylindrical ice –water phantom was used for the experiment, containing the vials with the different PVP concentrations.

#

ROI 1

ROI 2

ROI 3

ROI 4

ROI 5

ROI 6

ROI 7

PVP (% w/w)

0

2.5

5

10

15

20

25

45 min before the scanning the phantom was refilled with ice-water, in order to obtain a temperature of 0 0 C.

Dr Marianthi-Vasiliki Papoutsaki

ADC estimates from PVP phantom 1

PVP

(% w/w)

0 - ROI1

2.5 – ROI2

5 - ROI3

10 – ROI4

15 – ROI5

20 – ROI6

25 – ROI7

ADC median

(*10 -3 mm 2 /s)

1.12

1.05

0.99

0.85

0.71

0.60

0.50 ADC map

Linear relationship between PVP

R 2 = 0.996

0

0 C

Boss et al ISMRM 2013 Dr. Marianthi-Vasiliki Papoutsaki

ADC

ADC Uniformity Assessment

ADC profile

Evaluation of ADC uniformity is essential for quantitative WBDWI

Malyarenko et al. Magn Reson Med. 2013 May 13. Courtesy Dr. Jessica Winfield

ADC profile

Gradient

~2%

ADC profile

Non -Uniform

ADC ADC

ADC uniformity is image acquisition, reconstruction and post-processing dependent

Courtesy Dr. Jessica Winfield

ADC profiles along x and z axis of bipolar and unipolar diffusion pulse xaxis

Median ADC value +/-

2.5%

Centred at isocentre

ADC map bipolar zaxis

Median ADC value +/-

2.5%

Centred at isocentre

53% deviation from the median value in bipolar

Dr Marianthi-Vasiliki Papoutsaki

Fat-Water Test-object

Inner cylinder: water/NaCl/CuSO

4

Annulus:

Corn oil coronal

400 mm x 400 mm FOV

Gradient echo localiser images. Test object placed at

45 degrees to z-axis to create large ellipse in axial slices.

185 mm

140 mm axial

400 mm x 400 mm FOV

Winfield et al Phys Med Biol. 2014 May 7;59(9):2235-48

Upfield and downfield fat signals

Test object:

No fat suppression, b=900

Volunteer:

No fat suppression, b=900

•   upfield fat (1.3 ppm) shifted in anterior direction

•   downfield fat (5.3 ppm) shifted in posterior direction

Courtesy Dr. Jessica Winfield

Unsuppressed upfield fat at edges of

FOV using SPAIR at 1.5 T

Test object:

SPAIR, b=900

Volunteer:

SPAIR, b=900

•   Unsuppressed upfield fat at edges of FOV in test object and volunteer

•   SPAIR leaves downfield fat signal unsuppressed

Courtesy Dr Jessica Winfield

Whole Body Dixon (Fat/Water)

•   Whole Body Dixon registered and fused with WBDWI a) WBDWI, b) Water, c) Fused a)+b), d) Dixon T1 map, e) Water ratio

Whole Body Dixon (Fat/Water)

•   Whole Body Dual Contrast Dixon registered and fused with WBDWI a) WBDWI, b) Water, c) Fused a)+b), d) Dixon T1 map, e) Water ratio

Three quantitative whole body metrics ADC, T1 and Fat/Water ratio

Blackledge et al ISMRM 2009

Semi-automatic segmentation

+ b = 0 s/mm 2 ADC map

Acquired data

Select computed b-value and initial disease threshold

Smoothing of regions using

Markov random field model

(MRF)

Visualization/quantification of disease

User modifiable regions of interest

(ROI)

Blackledge et al PLoS One. 2014 Apr 7;9(4):e91779.

ADC stats

(x10 -3 mm 2 /s)

Mean

Variance

Skewness

Kurtosis

Pretreatment

0.88

0.05

0.46

1.00

Posttreatment

1.15

0.15

-0.02

-1.15

Discussion

•   Routine Acceptance testing and QA informs only on the important basic functionality

•   Broader range of test objects required

•   How well do phantom measurements translate into meaningful measures (eg ADC repeatabilty, DCE-MRI noise factors)?

•   QA has to move from the generic to the assessment of specific clinical protocols

Acknowledgements

Dr James d’Arcy

Dr Jessica Winfield

Dr Mihaela Rata

Dr Marianthi-Vasiliki Papoutsaki

Dr Matthew Blackledge in partnership with

Download