Simultaneous Capacitance Voltage (CV) Measurement

advertisement
Simultaneous Capacitance Voltage (CV) Measurement
FACULTY OF ENGINEERING
LAB SHEET
ENT 3036 SEMICONDUCTOR
DEVICES
TRIMESTER 2 2010-2011
SD2 – C-V MEASUREMENT OF MOS CAPACITOR
*Note: On-the-spot evaluation may be carried out during or at the end of the experiment.
Students are advised to read through this lab sheet before doing experiment. Your
performance, teamwork effort, and learning attitude will count towards the marks.
Page 1 / 17
Simultaneous Capacitance Voltage (CV) Measurement
1. Introduction
1.1 What is Capacitance Voltage (CV) Measurement
 It is the measurement of device capacitance against a sweep voltage
 The capacitance of the device is measured at quasistatic (very low frequency of mHz) and high
frequencies (either 100kHz or 1MHz)
 Best measurement results are obtained from measuring both low and high frequencies
simultaneously
 Measured data is used as information of device characterization and process adopted
1.2 MIS Capacitor
 Base on Metal Insulator Semiconductor structure
 Most common is Metal (Polysilicon), Oxide-Semiconductor Capacitor (MOSCAP)
 Devices can be transistors (MOSFET), P-N junctions or Schottky diodes
1.3 Why MIS Capacitor?
 Simple structure, easy to fabricate
 Readily integrate into process
 Similar to MOSFET structure, just add Source and Drain
1.4 Objectives:
1. To investigate the effect of low and high frequencies on the operating modes of a MOS capacitor
2. To analyse the device parameters of a MOS capacitor from C-V measurement
2. CV Measurement System
2.1 Keithley 82-WIN CV System
The Keithley model 82-WIN Simultaneous CV System consists of the following components:
 Keithley model 590 / 100k /1M CV Analyzer
o Source a high frequency of 100 kHz or 1 MHz selectable signal to the Device Under Test (DUT)
and measures the Capacitance, termed as CH.
 Keithley model 595 Quasi-static CV Meter
o Keithley model 595 Quasi-static CV Meter
o Source a low frequency (almost DC Voltage) signal to the DUT and measures the Capacitance,
termed as CQ.
o Provides the step voltage (max. of 40 V range) for simultaneous CV measurement
o Measures the leakage current through the DUT, (Q/t).
o Provides triggering to Keithley model 590 / 100k / 1M CV Analyzer
 Keithley model 230-1 Programmable Voltage Source
o Provides a DC offset bias voltage of +/- 100 V
 Keithley model 5951 Remote Coupler
Page 2 / 17
Simultaneous Capacitance Voltage (CV) Measurement
o Consists of tuned and resonant circuits to separate the low and high frequencies for
simultaneous CV measurements. Ensures minimal interaction between instruments when
performing measurement
o Acts as the interface to connect between DUT and instruments for simultaneous CV
measurement
 Metrics Interactive Characterization Software (ICS)
o Software control of instrument interface, data acquisition and analysis and test setup
 System Controller
o Installed with General Purpose Interface Bus (GPIB) conforming to IEEE488.2 for
communication between Software and instruments
2.2 Hardware Connection Diagram
Figure 2.2 shows the cable connection between the instruments.
Note: All connections to the model 5951 Remote Input Coupler are to use Keithley model
4801 low noise co-axial cables. Failure to comply with this will result in unreliable data
measurement.
Figure 2.2
2.3 GPIB Address of Instruments
Page 3 / 17
Simultaneous Capacitance Voltage (CV) Measurement
The GPIB addresses of the instruments are to be set as follow:
 Keithley model 590: Address 15
 Keithley model 595: Address 28
 Keithley model 230: Address 13
Please ensure that the addresses of the hardware and software are the same settings.
3. Typical CV Measurement Characteristic
 Two Curves
o Quasistatic CV (low frequency)
o High Frequency CV (100kHz or 1MHz)
Figure 3.0 shows a typical simultaneous CV measurement for a p-type semiconductor of a MOS-CAP
structure
Figure 3.0: Typical Simultaneous CV (p-type)
3.1 Fundamentals of CV Measurement
A typical CV measurement will consists of three regions:
 Accumulation region
In the accumulation region, the majority carriers will accumulate near the semiconductor surface.
Capacitance = Oxide Capacitance, Cox

Figure 3.1.1: Accumulation (p-type)
Depletion region
Page 4 / 17
Simultaneous Capacitance Voltage (CV) Measurement
When the device starts to deplete, the majority carriers will be push away from the semiconductor
surface. Separation of High and Low Frequency begins to take place due to generation of minority
carriers and Interface Traps.
Capacitance = Oxide Capacitance, Cox and Depletion Layer Capacitance in series
Figure 3.1.2: Depletion (p-type).
 Inversion region
During Inversion, minority carriers will be generated and dominant near the semiconductor surface.
Semiconductor Surface Charge is inverted. Minority carriers do not respond to high frequency stimulus
and generation are relatively slow.
Capacitance = Oxide Capacitance and Maximum Depletion Layer Capacitance in Series.
(For High Frequency Only)
Low Frequency Capacitance = Oxide Capacitance
Figure 3.1.3: Inversion (p-type).
3.2 Characteristic of CV
Page 5 / 17
Simultaneous Capacitance Voltage (CV) Measurement
Figure 3.2.1 shows a typical CV characteristic of p-type MOS-CAP (MOS-C) while Figure 3.2.2 is that
of an n-type MOS-CAP.
Figure 3.2.1: p-type material
Figure 3.2.2: n-type material
 Oxide Capacitance, Thickness and Gate Area
The oxide capacitance, Cox is the high frequency capacitance with the device biased in strong
accumulation. Oxide Thickness, Tox is calculated from Cox and the gate area as follows:
Tox= Aox / Cox
EQ (3.2.1)
Where: Tox is the Oxide Thickness in nm
A is the Gate Area in cm2
ox is permittivity of oxide material in F/cm (3.4 E-13)
Cox is oxide capacitance in F
Note: ox and other constants are initialized for use with silicon substrate, silicon dioxide
insulator and aluminum gate material, but may be changed for other materials.

Threshold Voltage
Page 6 / 17
Simultaneous Capacitance Voltage (CV) Measurement
The Threshold Voltage, VTH is obtained when the surface potential, S is twice that of the bulk potential,
B (Please refer to Figure 3.2.1 and 3.2.2). This corresponds to onset of strong inversion of MOS-CAP.
For an enhanced mode MOSFET, VTH correspond to the point where the device starts to conduct.
The VTH can be calculated from:
VTH = [ ± (A / Cox){(4Sq |Nbulk| |B|)} +2|B|] + VFB
EQ (3.2.2)
Where: A is the Gate Area in cm2
Cox is the oxide capacitance in Farads, F
S is the permittivity of substrate material (F/cm) [material constant=1.04E-12]
q is the electron charge of 1.60219 x 10-19 coul
B is the Bulk potential in V
Nbulk is the bulk doping concentration in cm-3
 Flatband Voltage (VFB) and Capacitance (CFB)
The Keithley model 82-WIN uses the flatband capacitance, CFB method of finding flatband voltage, VFB.
The extrinsic debye length, is used to calculate the ideal value of flatband capacitance and can be
calculated as follow:
= (SkT / q2Nx)
EQ (3.2.3)
Where: kT is the thermal energy at room temperature (4.046E-21) in Joules, J
Q is the electron charge (1.60219E-19) in columbs, coul
Nx is N at 90% Wmax or NA or ND when input by user
N at 90% Wmax is chosen to represent the bulk doping
Once the value of CFB is known, the value of VFB is interpolated from the closest Vgs value. Based on
doping, the calculation of CFB uses N at 90% Wmax, or user-applied NA (bulk doping for p-type,
acceptors) or ND (bulk doping for n-type, donors) The Flatband capacitance is first calculated as follows:
CFB = (CoxSA / ) / (Cox + SA / )
EQ (3.2.4)
Where: A is the Gate Area in cm2
Cox is the oxide capacitance in Farads, F
S is the permittivity of substrate material (F/cm) [material constant=1.04E-12]
is the extrinsic debye length
At Flatband, surface potential is zero.
3.3 Doping Profile
 Depletion Depth vs Gate Voltage
The doping profile is plotted as Depletion Depth vs Gate Voltage (Vgs). For the Keithley model 82-WIN,
the system computes the depletion depth, W from the high frequency capacitance and oxide
Page 7 / 17
Simultaneous Capacitance Voltage (CV) Measurement
capacitance at each measured value of Vgs. In order to graph this function, each W element needs to
be computed as follows:
W = AS [(1 / CH) – (1 / Cox)]
Where: W is the depletion depth expressed in um
S is the permittivity of substrate material, 1.04E-12 F/cm
CH is the high frequency capacitance in F
Cox is the oxide capacitance in F
A is the gate area in cm2
EQ (3.2.5)
 Doping Concentration vs Depth
The doping profile of a device is derived from the CV curve based on the definition of the differential
capacitance (measured by Keithley models 590 CV Analyzer and 595 Quasistatic CV Meter) as the
differential change in depletion region charge produced by a differential change in gate voltage.
Standard Doping (N) vs Depletion Depth(W) does not compensate for onset of accumulation and is
accurate only when in depletion. In the Keithley model 82-WIN, in order to correct for errors caused by
interface traps, the error term [{1 – (CQ / Cox)} / {1 – (CH / Cox)}] is included in the calculations in the
extraction of the doping concentration and termed as Corrected Doping Concentration, NCORR.
The extraction of the concentration level will be extracted at 90% depletion depth.
3.4 Interface Trap Density
The Interface Trap Density, DIT is calculated from Interface Trap Capacitance, CIT as follows:
Thus from EQ (3.4.1)
CIT = [(1 / CQ) – (1 / Cox)]-1 – [(1 / CH) – (1 / Cox)]-1
EQ (3.4.1)
DIT = CIT / A
EQ (3.4.2)
Where: CIT is the Interface Trap Capacitance in F
DIT is the Interface Trap Density in cm-2 eV-1
CQ is the Quasistatic Capacitance in F
CH is the High frequency Capacitance in F
Cox is the Capacitance Oxide in F
A is the gate Area in cm2
4. CV Measurement Procedure
4.1 Equipment Requirement

Experiment Wafer and Wafer Map Sheet
Page 8 / 17
Simultaneous Capacitance Voltage (CV) Measurement

Keithley model 82-WIN Simultaneous CV Measurement System consisting of components as
listed in Section 2
 MicroManipulator model S6-EVG probe station (vacuum type) with enclosure / light source
consisting of:
o 2 x co-axial manipulators / probes
o 1 x 6” chuck

Glove / Twister for Wafer handling
4.2 Connection
Please check the system connection with the diagram as in Figure 2.2. Once the connection is
established, power on the CV measurement system by pushing the power ON / OFF switch of the
respective instruments / equipments:





Keithley model 590 CV Analyzer
Keithley model 595 Quasistatic CV Meter
Keithley model 230-1 Programmable Voltage Source
The light source for MicroManipulator model S6-EVG probe station
Vacuum pump
4.3 Setup Procedure
Turn the vacuum switch to “OFF” position to disable the vacuum. (Switch is positioned on the right of
the chassis probe enclosure) Place the assigned experiment wafer on top of the chuck and turn “ON”
the vacuum switch. Ensure that the wafer is positioned properly on the chuck. Check the wafer map
sheet and select a appropriate site for your CV measurement from the listed wafer map sheet. Position
the tip of the probe that is connected to the OUTPUT of Keithley model 5951 (adjust using the X, Y
and Z-axis fine adjust) on the DOT wafer that CV measurement is to be performed. Ensure that the tips
are just contacting the DOT wafer and not scratching it. Similarly, position the tip of the probe that is
connected to the INPUT of Keithley model 5951 (adjust using X, Y and Z-axis fine adjust) on the top
surface of chuck. Once the probes are properly positioned to the DOT wafer, probe up again for both
left and right probes. (That is, the tips of the probe should not make contact with the DOT wafer or
chuck) Close the probe station enclosure.
4.4 Starting Metrics ICS and Setup of Experiment
Double-Click on the Metrics ICS icon
Measurement.
on the Desktop to execute Metrics ICS for CV
4.4.1 Selection of Experiment File



From the File Menu Options, Select “File Open”.
From the Attribute #1, Select “Expt” from the drop-down list and click “OK” when done.
The Project file that was opened will consist of the Setup for CV measurement for this
experiment.
Page 9 / 17
Simultaneous Capacitance Voltage (CV) Measurement
Figure 4.4.1A: File Open Select
Figure 4.4.1B shows the Project File named “Expt” of the CV measurement experiment
4.4.2 Loading of Calibration File Constant

From the icon option, click “Setup Editor”

From Setup Editor Menu, click “Opts” icon
Page 10 / 17
Simultaneous Capacitance Voltage (CV) Measurement

The KI82 options Setup Menu will be prompted
Figure 4.4.2: Setup Editor Menu



Select “Load cal….”
Select the most appropriate CAL file to use for the experiment. Please note that only
calibration constant files having the extension .CAL can be selected.
Once the appropriate calibration constant file is loaded, press “OK”.
Note: Calibration constant file is required to compensate for the High Frequency offset of the
connecting cables during measurement so as to obtain a more reliable and accurate
capacitance reading. Please check with relevant personnel on the correct calibration file
constant to be loaded.
4.4.3 Setup of Test Conditions for CV Measurement

 From the Setup Editor (See Figure 4.4.2), click on the OUT instrument icon (That is connected
to the Gate,G Terminal), the Source Setup for KI82-WIN will appear as in Figure 4.4.3.
 Depending on the wafer provided for the experiment, select an appropriate site for the CV
measurement that is listed on the wafer map sheet (As done in procedure 4.3).
 From the wafer map sheet, check the Start and Stop Voltage of the wafer site selected and key
the respective values to the “Start V” and “Stop V” accordingly.
 Select 20mV as the “Step V”
 Select “Mode” to Single Stair
 Select 100kHz for High Frequency Stimulus
Page 11 / 17
Simultaneous Capacitance Voltage (CV) Measurement
Figure 4.4.3: KI82-WIN Setup






Select “Delay” to be 0.1
From the Time measurement bias, set “Bias Voltage” equal to the voltage at “Start V”
Select Range of measurement for 590 to 2nF and 595 to 20nF
Check the “Leakage Correction” box
Select Filter to “3rdg”
Click “OK” When completed and return to Editor Setup Menu
4.4.4 Defining Pre-Stress Time





From the Editor Setup Menu, click on Time icon
to activate the Time
Parameter Setup Dialogue.
Check on the “Sec” box to select units of time measurement to be in seconds
Select “LIN” for Type to indicate a linear time measurement is preferred
On the “Wait” box, enter 30 (indicates 30 seconds of pre-stress time for device under test
before measurement will take place).
Press “Done” on the Setup Editor when completed and return to Data Sheet View.
Note: In CV measurement, a pre-stress time is allowed to bias the device to ascertain that the
device is in a state of equilibrium. That is the carriers are in the appropriate interface when
measurement is begin performed.
4.5 Performing CV Measurement
Page 12 / 17
Simultaneous Capacitance Voltage (CV) Measurement
Once the appropriate and necessary setup had been completed from procedures 4.1 to 4.4, CV
measurement can be made.




To perform CV measurement, select Measure icon
to activate the Measure Dialogue.
For CV measurement, select from the drop down list, “BIAS DELAY”. This will setup the CV
measurement to stress the device for 30 seconds at a voltage defined as in the Time Bias
Voltage
Ensure that both probes are not in contact with the wafer or chuck and close the probe shield
enclosure
Click on the ZERO CANCEL of the Measure Dialogue and follow the instructions as prompted.
This procedure is to offset any drift exhibit by the measuring system, interconnecting cables
and probes.
Figure 4.5: Measure Dialogue




Once the ZERO CANCEL had been completed, adjust the probes to contact the wafer and the
chuck according to procedure 4.3
Once the probes are in contact with the wafer, close the probe shield enclosure
Press on the “SINGLE” button of the Measure Dialogue to begin the CV Measurement
Once the CV measurement is completed, the results will be reflected in the Data Sheet View
4.6 Graph Plotting





Upon completion of the CV measurement, plot the following graphs:
Simultaneous CV measurements, VGS versus SIM CV (X-axis=VGS, Y1-axis=SIM CV)
Doping Concentration versus Depth (X-axis=DEPTHM, Y1-axis (log scale)=NCORR)
Interface Trap Density versus Trap Energy (X-axis=EIT, Y1-axis (log scale)=DIT)
Surface Potential versus Gate Voltage (X-axis=VGS, Y1-axis=PSISPSIO)
The graphs can be plotted as follows:

Click on the Plot View icon
Page 13 / 17
Simultaneous Capacitance Voltage (CV) Measurement



From the Plot View Setup Menu, select the appropriate measurement vectors from the drop
down list of Data Group for the respective axis
Ensure that selection of “Scale Type” for linear or Log scale is correct for each graph
Click “Apply” then “Done” when completed to exit from Plot View Setup
Figure 4.6: Plot View Setup Menu
4.7 Calculations and Extraction of CV Parameters
4.7.1 Minimum Capacitance, CH (min)
Extract the minimum High Frequency Capacitance from the CV plot
CH (min) = ____________ pF
4.7.2 Capacitance Oxide, Cox
Extract the Oxide Capacitance from the CV plot
Cox = _________________ nF
(Hint: Cox = Maximum Capacitance at High Frequency)
4.7.3 Oxide Thickness, Tox
Calculate the oxide thickness (Tox) from the given equation EQ (3.2.1)
Tox = __________________ nm
(Note: Please check the Area, A from the Wafer Map Sheet)
4.7.4 Depletion Depth, W
Find the maximum depletion depth, W(max) from Doping Concentration versus Depth graph or Data
View Spreadsheet and calculate depth at 90% level
W(max) = _________________ nm
W(90% ) = __________________ nm
Page 14 / 17
Simultaneous Capacitance Voltage (CV) Measurement
4.7.5 Doping Concentration, N
Given N90%W = 1.31 x 1013 cm-3
4.7.6 Extrinsic Debye Length, 
Calculate the extrinsic debye length, using the equation as in EQ (3.2.3)

= _______________________ nm
4.7.7 FlatBand Capacitance, CFB
Calculate the FlatBand Capacitance, CFB using equation as in EQ (3.2.4)
CFB = _____________________ nF
4.7.8 FlatBand Voltage, VFB
Extract the Flatband Voltage, VFB at CFB from the CV graph or Data View Spreadsheet
VFB = _____________________ mV
4.7.9 Bulk Potential, B
Calculate the Bulk Potential, B given the following:

B = (kT / q) [ln (N90%W / i)]
Where: kT is the thermal energy at room temperature (4.046E-21) in Joules, J
q is the electron charge (1.60219E-19) in columbs, coul
N90%W is the calculated doping concentration at 90% depletion depth
i is the intrinsic carrier concentration (1.45E10) in cm-3

B = ______________________ mV
4.7.10 Threshold Voltage VTH
Calculate Threshold Voltage, VTH using the formula in EQ (3.2.2)
VTH = _____________________ V
Note: Nbulk to take at N90%W
5. Questions
5.1
What is the three basic schemes of a MOS-C operation? Draw diagram (CV
Characteristics) and use only 1 sentence to describe each schemes of a p-typed MOS-
Page 15 / 17
Simultaneous Capacitance Voltage (CV) Measurement
C operation, draw the CV diagram for n-type MOS-C operations also. Indicate the
flat band and strong inversion positions. (18 points)
Derive an expression for the potential distribution (W) in an ideal MOS-C in the
depletion condition in terms of the surface potential, E s and the depletion width, Wmax
at the surface taking the zero potential in silicon bulk. The silicon is p-doped type and
x=0 at the oxide-silicon interface. (14 points)
Note: please approximate the space charge and field gradient are constant and negative in the
silicon for (0 ≤ W ≤ Wmax).
5.2
5.3
A MOS-C is maintained at T=300K, oxide thickness WOX = 90nm, and the silicon
doping is NA=1.5x1015/cm3. Compute (assuming surface potential equals to bulk
potential, SF): (8 points)
a)
Fin kT/q units and in volts
b)
depletion width, Wmax
c)
Surface electric field (oxide-semiconductor) Es
Marking Scheme
Lab
(10%)
Assessment Components
Hands-On & Efforts (2%)
On the Spot Evaluation
(2%)
Lab Report
(6%)
Details
The hands-on capability of the students and their efforts during the
lab sessions will be assessed.
The students will be evaluated on the spot based on the lab
experiments and the observations on the capacitor characteristics.
Each student will have to submit his/her lab final report within 7
days of performing the lab experiment. The report should cover the
Page 16 / 17
Simultaneous Capacitance Voltage (CV) Measurement
followings:
1. Introduction, which includes background information on
metal oxide semiconductor capacitor.
2. Experimental section, which includes the general summary
of the lab experiment work.
3. Results and Discussions, which include the measured
results, analysis, and evaluations, with neat graphs/images
of the results and recorded data.
4. Conclusion, which includes a conclusion on the
experimental.
5. List of References, which includes all the technical
references cited throughout the entire lab report.
The report must have references taken from online scientific
journals (e.g. www.sciencedirect.com,
http://ieeexplore.ieee.org/xpl/periodicals.jsp,
http://www.aip.org/pubs/) and/or conference proceedings (e.g.
http://ieeexplore.ieee.org/xpl/conferences.jsp).
Format of references: The references to scientific journals and text
books should follow following standard format:
Examples:
[1] William K, Bunte E, Stiebig H, Knipp D, Influence of low
temperature thermal annealing on the performance of
microcrystalline silicon thin-film transistors, Journal of
Applied Physics, 2007, 101, p. 074503.
[2] Hodges DA, Jackson HG, Analysis and design of digital
integrated circuits, New York, McGraw-Hill Book Company,
1983, p. 76.
Reports must be typed and single-spaced, and adopt a 12-point
Times New Roman font for normal texts in the report.
Any student found plagiarizing their reports will have the
assessment marks for this component (6%) forfeited.
The lab report has to be submitted to the Electronics lab staff.
Please make sure you sign the student list for your submission. No
plagiarism is allowed. Though the electrical characteristics of the
measured capacitor from the same group can be similar, the report
write-up cannot be duplicated for group members. The individual
report has to be submitted within 7 days from the date of your lab
session. Late submission is strictly not allowed.
Page 17 / 17
Download