Word - IECEE

advertisement
COMMITTEE OF TESTING LABORATORIES
IEC SYSTEM FOR CONFIRMITY TESTING AND
CERTIFICATION OF ELECTRICAL EQUIPMENT
IEC60825-1:1993+A1+A2
IEC60825-1:2007
TESTING AND MEASURING EQUIPMENT/ALLOWED BORROWING/RENTING
S=May be subconctracted (incl. borrowed or rented)
R=Required
Clause
Measurement/testing
9.2 and 9.3
9.2 and 9.3
9.2 and 9.3
9.2 and 9.3
9.2 and 9.3
9.2 and 9.3
9.2 and 9.3
Radiant power
Radiant Energy
Duration of the pulse
Relative spectral measurement
9.2 and 9.3
Testing / measuring equipment / material needed
Wavelength: 400 nm -1800 nm
Wavelength: 200 nm – 400 nm or > 1800 nm
Radiant Power Range: 1 nW to 500 mW
Radiant Power Range > 500 mW
See Note 2.
Photo detectors with rise/fall time adequate for the measured pulse
durations. Oscilloscope with speed adequate to match the detector.
wavelength measuring device (e.g. spectrometer, monochromator, spectrum
analyzer, etc.) as may be needed for the accurate operation of the
power/energy meter
Wavelength : 400 nm - 1100 nm
9.2 and 9.3
9.2 and 9.3
9.2 and 9.3
Data input
Signal generator suitable for the components evaluated
9.2 and 9.3
Angle of acceptance set up
Optics
9.2 and 9.3
Apertures
9.2 and 9.3
Angle of acceptance setup; radiant
power/energy/exposure/irradiance measurement
Positioning, alignment equipment
9.2 and 9.3
Precision measurement
9.2 and 9.3
Angle of subtense measurement
September 2008
Borrowing/
Renting
See Note 1
R
S
R
S
S
S
See Note 3
R
S
Wavelength: 200 nm - 400 nm or > 1100 nm
ver. 2.0
Optical positioning/mounting equipment (e. g. optical table, breadboard or
rail, optical mounts, holders, translation stages, etc.)
Caliper, meter stick and other suitable equipment as needed for distance and
aperture measurements. Caliper accuracy 0.1 mm
Lenses, apertures and an array sensor (e.g. CCD) as specified in 9.3.3 (a)
and fig. 5 of IEC 60825-1 Ed. 2.0
1 /2
IEC60825-1:1993+A1+A2 and IEC60825-1:2007
S
See Note 4
S
See Note 5
S
See Note 6
R
R
Optional.
If not
COMMITTEE OF TESTING LABORATORIES
IEC SYSTEM FOR CONFIRMITY TESTING AND
CERTIFICATION OF ELECTRICAL EQUIPMENT
9.2 and 9.3
Other equipment
9.2 and 9.3
Personal Protective Equipment
Other suitable measurement system such as knife edge type detector with
analysis software or beam profiler.
- Black/white CCD camera with monitor
- IR/UV viewers
Protective equipment as may be required by national/local regulations
available,
C6=1 must be
assumed
S
1) Required equipment represents the basic minimum necessary to perform measurement, which may be sufficient for many applications.
2) In some cases it is possible to derive the pulse energy from power measurements. Where appropriate, manufacturer’s specifications (e.g.
modulation frequency and duty cycle) may be used to calculate energy from power as an alternative to measurement. Participating laboratories shall
assure that such conversion is possible and accurate. The decision shall be based on the compatibility of the power detector specifications and the
measured laser characteristics such as peak power, pulse duration, repetition frequency, etc.
Participating laboratories shall assure that any borrowed energy meters are adequate for measurement of the laser beam in all aspects, including but
not limited to: wavelength, energy, pulse duration, repetition frequency, pulse damage threshold, etc.
3) Spectral range and resolution should be adequate so that the tested laser wavelength could be measured or verified with an accuracy of 1 nm.
4) Signal generators are only necessary when the modulated output is more hazardous than the emission at the same average power. Participating
laboratories shall assure that measurements done using modulated output represent the worst case condition
5) When lenses are used, optical aberrations shall be considered for suitability. When optics are used for angle of acceptance, the properties of the
optics shall be known and deemed suitable.
6) The use of additional apertures is only necessary when the required measurement geometry cannot be assured by the appropriate arrangement of
a circular sensor of a known diameter at an adjusted measurement distance.
7) Where hazardous levels of laser radiation are accessible during testing, laboratories must comply with national and other locally applicable safety
regulations.
Note: The presence of equipment alone does not indicate a satisfactory situation. Assessors must evaluate the equipment design, calibration,
uncertainty and documentation to ensure compliance with the directions of the standard. The requirements of ISO Guide 25 regarding validation are
applicable, as the tests of this standard are not standardised tests.
September 2008
ver. 2.0
2 /2
IEC60825-1:1993+A1+A2 and IEC60825-1:2007
See Note 7
Download