COMMITTEE OF TESTING LABORATORIES IEC SYSTEM FOR CONFIRMITY TESTING AND CERTIFICATION OF ELECTRICAL EQUIPMENT IEC60825-1:1993+A1+A2 IEC60825-1:2007 TESTING AND MEASURING EQUIPMENT/ALLOWED BORROWING/RENTING S=May be subconctracted (incl. borrowed or rented) R=Required Clause Measurement/testing 9.2 and 9.3 9.2 and 9.3 9.2 and 9.3 9.2 and 9.3 9.2 and 9.3 9.2 and 9.3 9.2 and 9.3 Radiant power Radiant Energy Duration of the pulse Relative spectral measurement 9.2 and 9.3 Testing / measuring equipment / material needed Wavelength: 400 nm -1800 nm Wavelength: 200 nm – 400 nm or > 1800 nm Radiant Power Range: 1 nW to 500 mW Radiant Power Range > 500 mW See Note 2. Photo detectors with rise/fall time adequate for the measured pulse durations. Oscilloscope with speed adequate to match the detector. wavelength measuring device (e.g. spectrometer, monochromator, spectrum analyzer, etc.) as may be needed for the accurate operation of the power/energy meter Wavelength : 400 nm - 1100 nm 9.2 and 9.3 9.2 and 9.3 9.2 and 9.3 Data input Signal generator suitable for the components evaluated 9.2 and 9.3 Angle of acceptance set up Optics 9.2 and 9.3 Apertures 9.2 and 9.3 Angle of acceptance setup; radiant power/energy/exposure/irradiance measurement Positioning, alignment equipment 9.2 and 9.3 Precision measurement 9.2 and 9.3 Angle of subtense measurement September 2008 Borrowing/ Renting See Note 1 R S R S S S See Note 3 R S Wavelength: 200 nm - 400 nm or > 1100 nm ver. 2.0 Optical positioning/mounting equipment (e. g. optical table, breadboard or rail, optical mounts, holders, translation stages, etc.) Caliper, meter stick and other suitable equipment as needed for distance and aperture measurements. Caliper accuracy 0.1 mm Lenses, apertures and an array sensor (e.g. CCD) as specified in 9.3.3 (a) and fig. 5 of IEC 60825-1 Ed. 2.0 1 /2 IEC60825-1:1993+A1+A2 and IEC60825-1:2007 S See Note 4 S See Note 5 S See Note 6 R R Optional. If not COMMITTEE OF TESTING LABORATORIES IEC SYSTEM FOR CONFIRMITY TESTING AND CERTIFICATION OF ELECTRICAL EQUIPMENT 9.2 and 9.3 Other equipment 9.2 and 9.3 Personal Protective Equipment Other suitable measurement system such as knife edge type detector with analysis software or beam profiler. - Black/white CCD camera with monitor - IR/UV viewers Protective equipment as may be required by national/local regulations available, C6=1 must be assumed S 1) Required equipment represents the basic minimum necessary to perform measurement, which may be sufficient for many applications. 2) In some cases it is possible to derive the pulse energy from power measurements. Where appropriate, manufacturer’s specifications (e.g. modulation frequency and duty cycle) may be used to calculate energy from power as an alternative to measurement. Participating laboratories shall assure that such conversion is possible and accurate. The decision shall be based on the compatibility of the power detector specifications and the measured laser characteristics such as peak power, pulse duration, repetition frequency, etc. Participating laboratories shall assure that any borrowed energy meters are adequate for measurement of the laser beam in all aspects, including but not limited to: wavelength, energy, pulse duration, repetition frequency, pulse damage threshold, etc. 3) Spectral range and resolution should be adequate so that the tested laser wavelength could be measured or verified with an accuracy of 1 nm. 4) Signal generators are only necessary when the modulated output is more hazardous than the emission at the same average power. Participating laboratories shall assure that measurements done using modulated output represent the worst case condition 5) When lenses are used, optical aberrations shall be considered for suitability. When optics are used for angle of acceptance, the properties of the optics shall be known and deemed suitable. 6) The use of additional apertures is only necessary when the required measurement geometry cannot be assured by the appropriate arrangement of a circular sensor of a known diameter at an adjusted measurement distance. 7) Where hazardous levels of laser radiation are accessible during testing, laboratories must comply with national and other locally applicable safety regulations. Note: The presence of equipment alone does not indicate a satisfactory situation. Assessors must evaluate the equipment design, calibration, uncertainty and documentation to ensure compliance with the directions of the standard. The requirements of ISO Guide 25 regarding validation are applicable, as the tests of this standard are not standardised tests. September 2008 ver. 2.0 2 /2 IEC60825-1:1993+A1+A2 and IEC60825-1:2007 See Note 7