Quantitative HBeAg assay

advertisement
Supporting Method
Quantitative HBeAg assay
The development of the quantitative HBeAg assay was based on the method of
Perrillo et al (14). Serum HBeAg was measured by EIA (ARCHITECT platform; Abbott
Laboratories). This commercial HBeAg kit is not marketed as a quantitative assay but
produces a signal which is linear within a restricted range. The HBeAg reference
preparation was obtained from the Paul Ehrlich (PE) Institute (Paul-Ehrlich Institute,
Langen, Germany). The preparation has a defined HBeAg activity of 100 PE IU/mL. An
in-house working HBeAg standard was then prepared from a pool of high-titre HBeAg
positive specimens received for routine diagnostic testing, and calibrated against the PE
reference preparation. The standard was subsequently determined to have an HBeAg titre
of 2900 PE IU/mL, and was then used to generate a calibration curve within the linear
range of the assay. Dilutions of the in-house working standard (diluted twofold, from 1 in
2 to 1 in 4,096, or 1450 to 0.7 PE IU/mL) were prepared using the ARCHITECT HBsAg
Manual Diluent (Abbott Diagnostics, Sligo, Ireland), which contains recalcified human
plasma that is non-reactive for HBsAg, HIV-1 RNA or HIV-1 Ag , anti-HIV-1/HIV-2,
anti-HCV, and anti-HBs. The standards were aliquoted and stored at -20C.
The performance characteristics of the assay for use as a quantitative tool were
evaluated. The in-house HBeAg working standard was used to generate a calibration
curve within the linear range of the assay. The linear range was approximately 0.5 – 100
PE IU/mL. For determination of the inter-assay variability, 8 dilutions of the standard (1
in 2 – 1 in 4096) were tested in four separate runs. No significant deviation from linearity
was present for any of the following dilutions: 1 in 16, 1 in 32, 1 in 64, 1 in 128, 1 in 256,
1 in 512, 1 in 1024 and 1 in 2048 (data not shown). Intra-assay variation was tested using
duplicate samples of the reference preparation at 1 in 64 and 1 in 2048 dilutions, in four
separate assay runs. The percentage coefficient of variation ranged from 0.68 – 7.6%,
with the higher coefficient of variation at lower HBeAg levels. The intra-assay variation
was less than 5% (range 0.54 – 2.7%) for all samples (results not shown).
Each day that clinical samples were tested, aliquots of the diluted reference
sample (1 in 16 – 1 in 2048) were used to generate a standard curve. Linear regression
was used to convert the assay result into PE IU/mL for each sample. Where samples fell
outside the linear range of the assay (approximately 0.5 – 90 PE IU/mL, but defined by
the standard curve for each assay run), serial dilution (1 in 10, 1 in 100, 1 in 1000, ± 1 in
10,000) was performed using the ARCHITECT HBsAg Manual Diluent. Non-reactive
human serum was used as the negative control.
Download