Advia - Red Blood Cell Laboratory

advertisement
Hematology Analyzer
Complete Blood Count (CBC)
The Advia 120 Hematology Analyzer yields data related to the cell counts, cell
size, cell type, and cell contents. A complete blood count (CBC/Diff/Retic)
produces data in three main categories: CBC includes the hemoglobin, red blood
cell, and platelet methods, the Diff includes the white blood cell differential, and
the Retic includes the reticulocyte analysis.
Figure 1. The Bayer/Siemens Advia 120 instrument.
The hemoglobin reagent lyses the red blood cells to release hemoglobin. The
heme iron in the hemoglobin is oxidized from the ferrous to the ferric state, and
then it is combined with cyanide to form the reaction product. The RBC/platelet
reagent causes isovolumetric sphering and fixation of the RBC and platelets.
Laser light detects the cells: the low angle scatter is used to measure volume and
the high angle scatter is used to detect the hemoglobin concentration.
Parameters measured or calculated: Total Hemoglobin, Mean Corpuscular
Hemoglobin, Mean Corpuscular Hemoglobin Concentration.RBC Count, Mean
RBC Volume, Hematocrit, Platelet Count, Mean Platelet Volume.
Figure 2 shows the RBC volume and hemoglobin concentration histograms as
they relate to the dot plot of hemoglobin concentration versus volume. The 9
quadrants marked in the dot plot correspond to the macrocytic, microcytic,
normocytic, hypochromic, hyperchromic, and normochromic cells. The values of
the quadrant markers are shown in blue.
Figure 3 shows the PLT scatter cytogram. This plot is the graphical
representation of two light-scatter measurements; the low-angle and high-angle
light scatter signals for each cell are transformed into volume and refractive index
values.
White Blood Cell (WBC) Differential (Diff)
The perox reagents lyse the RBC, fix the WBC, and stain the granules within the
WBC to complete the differentiation. The cells absorb light proportional to the
amount of peroxidase stain present, and scatter light proportional to their size.
The baso reagent lyses all cells except basophils. Cluster analysis of the
cytograms identifies each population based on its position, area, and density,
and then the number of cells/nuclei in each population is processed. The lines
that separate the different cell populations are calculated by the software on a
sample-by-sample basis. Parameters (# and % for each): total WBC, Neutrophils,
Lymphocyte, Monocyte, Eosinophil, Basophil, Mononuclear Cells, and
Polymorphnuclear Cells.
Figure 4. The perox cytogram shows the amount of peroxidase stain represented
on the x-axis and the light scatter represented on the y-axis.
Figure 5. The baso cytogram shows the high-angle light scatter (nuclear
configuration) on the x-axis, and the low-angle light scatter (cell size) on the yaxis.
Reticulocyte Analysis (Retic)
The Retic reagent contains a surfactant that isovolumetrically spheres the red
cells, and a dye that stains the RNA contained in the reticulocytes. The events
are plotted with maturation (RNA staining) along the x-axis and cell size (light
scatter) along the y-axis. Gates are set within the software to differentiate mature
RBC from the reticulocytes. Parameters: Retic Count (# and %), Retic Volume,
Retic Hemoglobin Content and Concentration.
Download