Agar Slant Innoculation

advertisement
Procedure: Basic Lab Skills
Micropipettes
The unit of measurement is the microliter (µl).
The capacity of each micropipette is indicated by
a number usually on the top, expressed as a
min/max or just a max.
Note:
Each micropipette is engineered to measure
within a particular range of volumes. The range
is usually given on the top of the pipet. For
example 2 – 20 designates a functional range of 2
– 20 µl. Sometimes only the maximum range is
given, i.e. P200 indicates that maximum range is
200 µl. In this case, the minimum range can often
be inferred from the accompanying pipets in the
set. If a P20 is being used, one would choose to
NOTES:
use the P20 instead of the P200 to measure
accurately below 20 µl. One must be familiar
with the pipets used in the particular lab to
determine the appropriate pipet to use.
CAUTION: DO NOT SET THE
MICROPIPETTES FOR VOLUMES OUTSIDE
THEIR RANGES. THIS WILL CAUSE
MISCALIBRATION AND INACCURATE
MEASUREMENTS.
Setting the micropipette:
Again, the technique for setting the micropipette
varies with the manufacturer. Often there is a dial
or in the housing of the micropipette and a widow
with three digits. Turning the dial rotates the
digits through cycles of ten to turn the next unit,
much like an odometer. The digits represent
different units by orders of ten depending on the
range of the micropipette. For example, on a P20
the top digit represents the 10’s, middle digit 1’s
and bottom digit 0.1’s. On the P200, the digit
values are 100’s, 10’s, and 1’s respectively.
Copyright © 2007 MassBioEd
1
Operation:
1. Always use a clean, disposable tip.
2. Depress the plunger to the first position.
3. Insert the tip in the liquid.
4. Slowly release the plunger.
5. Withdraw the pipette from the liquid.
6. Insert the pipette tip in the new tube.
7. Slowly depress the plunger to the second
position.
8. With the plunger depressed, remove the
pipette tip from the liquid.
9. Depress the eject button to remove the
disposable tip when finished.
Activity: Use the “Micropipette Challenge” to
practice your pipette skills.
Electronic Balances
Operation:
1. Check that the balance is plugged in and
turned on.
2. Position the balance away from drafts or
vibrations.
3. Press the tare (zero) button. The display
should read 0.00.
4. Place the object on the balance.
5. Read the mass in grams on the display.
NOTES:
If a container such as a weigh boat or beaker is used
to contain a substance during measurement, be sure to
tare the empty container before adding the substance.
Activity: Measure 2 g of NaCl in a test tube. Save the
NaCl for the next activity.
Serological Pipettes
The unit of measurement for serological pipets is
milliliters. Most serological pipettes require a pump.
Operation:
1. Insert the top of the pipette firmly into the
base of the pump.
2. Check the pump plunger is in the fully
depressed position.
3. Insert the tip of the pipet into a liquid.
Copyright © 2007 MassBioEd
NOTES:
2
4. Turn the wheel of the pump clockwise to draw
the liquid up.
5. Dispense the liquid by turning the wheel
counterclockwise.
Activity: Add 10mL water to the container
containing 2 g of NaCl.
Vortex
The vortex is used to mix solutions.
Operation:
1. Set the toggle switch to touch.
2. Adjust the speed control to 5.
3. Hold the upper third part of the test tube
firmly between your index finger and thumb.
4. Place the bottom of the test tube on the rubber
platform and push the test tube down firmly.
5. Hold for 5 – 10 seconds.
NOTES:
Activity: Vortex the salt solution.
Calculate the molarity of this solution?
(Molecular weight of NaCl= 58.5 grams/mole)
Molarity= Moles of solute / Liters of solution
Glassware washing
1.
2.
3.
4.
Pour the salt solution into a sink.
Scrub the test tube with a test tube brush.
Rinse the test tube 10 times with tap water.
Rinse the test tube 3 times with distilled
water.
NOTES:
Centrifuge
The centrifuge is used to spin samples at high speeds
resulting in a separation of components by mass.
Activity: Obtain a microcentrifuge tube of E.coli.
Centrifuge the E.coli solution to create a bacterial
pellet.
Copyright © 2007 MassBioEd
3
1. Check that the samples to be centrifuged are
of equal size, mass, and volume.
2. Close the cover of the tubes.
3. Place the tubes in the rotor receptacles in a
balanced configuration. Maintain an equal
space between samples. Hint: Align the tube
hinges facing out. The pellet should form on
the same side as the hinges.
4. Close the cover of the centrifuge.
5. Adjust the timer to the desired time, usually in
minutes.
Note: Setting the timer to pulse will allow you to
manually control the time of centrifugation. The
rotor will only spin while the start button is
depressed.
6. A red light will display when the rotor is
turning.
Note: If the centrifuge vibrates excessively, it is
not balanced. Turn of the centrifuge immediately
by turning the timer to 0 in timer mode or
releasing the start button in pulse mode.
7. Wait for the red light to turn off, indicating
that the rotor is stopped.
8. Open the cover and remove the samples.
Indicate the proper
balancing of a centrifuge
when you have 3
microcentrifuge tubes
Label the supernatant and
pellet in the diagram.
Sterile Technique
We avoid using flames with students in the lab.
Therefore our sterile technique uses disposable sterile
tips and transfer loops. A general rule of thumb for
sterile technique is to breach sterile boundaries as
briefly as possible.
Resuspension:
1. Obtain a container labeled “Bacterial Waste.”
2. Hold the tube containing the E.coli pellet at a
45 degree angle. (A 45 degree angle
minimizes exposure to particulates in the air.)
3. Open the tube and pour the supernatant into
the container labeled “Bacterial Waste.”
4. Close the tube.
Copyright © 2007 MassBioEd
4
Resuspension Con’t.:
5. Use a sterile pipette tip to add 200 µl of sterile
water to the E.coli pellet. Remember to hold
the tubes at 45 degrees.
6. Resuspend the E.coli pellet by gently drawing
the pellet in and out of the pipette tip until the
solution is cloudy with no visible E.coli
chunks.
7. Cap the tube and dispose of the pipette tip in
the bacterial waste container.
NOTES:
Pouring Agar:
1. Lift the lid of a Petri dish just enough to admit
the neck of the bottle. Avoid touching the
bottle to any surface.
2. Pour the enough agar to cover the bottom of
each dish.
3. Cover the dish and let it sit until the agar is
hardened. (about 30 – 45 min.)
Note: Cover should not be off
Inoculating liquid cultures
1. Obtain an agar dish with viable bacteria.
2. Obtain 1 ml of sterile Luria Broth (with
ampicillin for this exercise).
3. Obtain a packaged sterile transfer loop.
4. Open the end of the package that does not
contain the loop.
5. Remove the transfer loop, being careful to
touch only the handle.
Copyright © 2007 MassBioEd
5
6. Open the cover of the agar dish with bacteria
just enough to allow a sterile loop to touch the
surface of the agar.
7. Gently drag the transfer loop across the
surface of the agar, collecting bacterial
colonies on the loop. You may see clump of
bacteria on the end of the loop. However, it is
not necessary to see the bacteria.
8. Remove the loop and insert it into the 1 ml of
Luria Broth. Remember to hold the tube at an
angle.
9. Spin the loop in the broth to release bacteria
into the solution.
10. Remove the transfer loop, close the tube, and
dispose of the loop in the bacterial waste.
Sterile Technique Activity
Check your sterile technique by transferring 100 µl of
sterile water to a sterile agar dish.
Procedure:
1. Obtain a microcentrifuge tube of sterile water.
2. Obtain a sterile Petri dish of LB agar.
3. Label the bottom of the dish “sterile water”
along with your initials and today’s date.
4. Use a micropipette with a sterile tip to transfer
100 µl of sterile water to the agar.
5. Spread the water with a sterile transfer loop
by gently moving the sterile loop across the
agar.
Agar Slant Innoculation
a. Unwrap a sterile applicator swab.
b. Use it to collect some E. coli bacteria from the E.
coli agar plate.
c. Inoculate agar slants by gently swiping the swab on
the agar. Recap the agar slants and incubate at room
temperature for 4 days, or in a 37ºC incubator for 2
days, until bacterial colonies are visible. Once
colonies are present, the slants can be stored in a 4oC
refrigerator.
Copyright © 2007 MassBioEd
6
The Micropipette Challenge
Using 0.5-10µl and 10-100µl Micropipettes
Laboratory science often involves working with very small volumes of liquid; frequently
millionths of liters are used. One millionth of a liter is equal to one microliter,
abbreviated l.
You need a very accurate and precise instrument to measure such small volumes. The
instrument most often used to measure microliters is called a micropipette. You will be
using two types of micropipettes: a 0.5-10µl and a 10-100µl micropipette.. You can
distinguish them by the color of the cap on each pipette. A 0.5-10µl has a grey cap and a
10-100µl pipette has a yellow cap.
This challenge will help you become accustomed to using micropipettes. You have two
tubes of colored water: green and yellow. Practice using the micropipettes by adding the
amounts listed on the back of this paper to an empty microcentrifuge tube. Record which
micropipette you used and the numerical setting as it appears in the window of pipette;
darken the appropriate line to represent the decimal point. If the amount was greater than
the micropipette can hold and you needed to use it multiple times, record the last setting
as well as how many times you filled the micropipette to capacity.
Example:
Add the following
amounts to an
empty tube.
Use this color
water.
Record which
micropipette you
used.
235 l
Green
Yellow cap
Copyright © 2007 MassBioEd
Record the setting.
Darken the line
representing the
decimal point.
0
3
5
0
7
Amount to add in
empty tube
Color of Water
220 l
Green
7 l
Green
74.4 l
Yellow
.0183 ml
Green
.583 ml
Yellow
197 l
Yellow
.0047 ml
Green
48.6 l
Yellow
347 l
Green
Copyright © 2007 MassBioEd
Record
micropipette used
Record the setting
by darkening the
line representing
the decimal point
8
How many l should you have in test tube when you are finished?
How many ml should you have in the test tube when you are finished?
Can you think of a way to use the 10-100µl micropipette to check that the amount of
liquid in your test tube is what it should be? Did you pipet the correct amount of liquid?
If not, what happened?
Copyright © 2007 MassBioEd
9
Micropipette Challenge - Answer Key and Notes
Each group of students needs a 0.5-10µl and 10-100µl micropipe as well as the correct
tips for each micropipette. If p1000's are not available, you can modify the activity by
having students skip the volumes above 200 l, which are in the 1st, 5th, and 9th rows.
Answer
Correct pipetting will result in a final volume of 1500 l or 1.5 ml. Students can check
their pipetting accuracy in two ways:
1. Most microcentrifuge tubes have marks at 0.5 ml, 1ml, and 1.5 ml. Their liquid
should be at the level of the 1.5 ml mark, and the tube should be almost full.
2. Students can measure the liquid in their tube with a 1ml pipette. Withdraw 1 ml of
liquid from the tube. Expel it into a liquid waste container. Then withdraw the rest
of the liquid in the tube into the pipet. It should read 0.5ml.
The Micropipette Challenge
Using the p1000, p200 and p20 Series
Laboratory science often involves working with very small volumes of liquid; frequently
millionths of liters are used. One millionth of a liter is equal to one microliter,
abbreviated l.
You need a very accurate and precise instrument to measure such small volumes. The
instrument most often used to measure microliters is called a micropipette. You will be
using three types of micropipettes: a p1000, a p200, and a p20. You can distinguish them
by the label on each pipette. A p1000 is used to measure volumes from 201l to 1000l
(equal to 1ml), a p200 is used to measure volumes from 21l to 200l, and a p20 is used
to measure volumes from 2l to 20l.
This challenge will help you become accustomed to using micropipettes. You have two
tubes of colored water: green and yellow. Practice using the micropipettes by adding the
amounts listed on the back of this paper to an empty microcentrifuge tube. Record which
micropipette you used and the numerical setting as it appears in the window of pipette;
darken the appropriate line to represent the decimal point.
Example:
Add the following
amounts to an
empty tube.
Use this color
water.
Record which
micropipette you
used.
35 l
Green
p200
Copyright © 2007 MassBioEd
Record the setting.
Darken the line
representing the
decimal point.
0
3
5
0
10
Copyright © 2007 MassBioEd
11
Amount to add in
empty tube
Color of Water
220 l
Green
7 l
Green
74.4 l
Yellow
.0183 ml
Green
.583 ml
Yellow
197 l
Yellow
.0047 ml
Green
48.6 l
Yellow
347 l
Green
Record
micropipette used
Record the setting
by darkening the
line representing
the decimal point
How many l should you have in test tube when you are finished?
Copyright © 2007 MassBioEd
12
How many ml should you have in the test tube when you are finished?
Can you think of a way to use the p1000 to check that the amount of liquid in your test
tube is what it should be? Did you pipet the correct amount of liquid? If not, what
happened?
Copyright © 2007 MassBioEd
13
Micropipette Challenge - Answer Key and Notes
Each group of students needs a p20, a p200, a p1000, and the correct tips for each
micropipette. If p1000's are not available, you can modify the activity by having students
skip the volumes above 200 l, which are in the 1st, 5th, and 9th rows.
Answer - Using a p20, a p200, and a p1000
Correct pipetting will result in a final volume of 1500 l or 1.5 ml. Students can check
their pipetting accuracy in two ways:
3. Most microcentrifuge tubes have marks at 0.5 ml, 1ml, and 1.5 ml. Their liquid
should be at the level of the 1.5 ml mark, and the tube should be almost full.
4. Students can measure the liquid in their tube with the p1000. Set the p1000 to1000l
and withdraw 1000 l from the tube. Expel it into a liquid waste container. Now set
the p1000 to 500 l and withdraw 500 l from the tube. If they have air bubbles or
empty space in the tip, they did not pipet the full 1.5 ml. If they have extra liquid in
the tube after removing 1000 l and 500 l, they pipetted more than 1.5 ml.
Answer - Using a p20 and a p200
Correct pipetting of the six volumes less than 200 l will result in a final volume of 350
l. Students can check their pipetting accuracy in two ways:
1. Most microcentrifuge tubes have marks at 0.5 ml, 1ml, and 1.5 ml. Their liquid
should be a bit below the level of the 0.5ml mark.
2. Students can check their final volume using the p200. Set the p200 to 200 l and
withdraw 200 l from the tube. Expel it into a liquid waste container. Now set the
p200 to 150 l and withdraw 150 l from the tube. If they have air bubbles or empty
space in the tip, they did not pipet the full 350 l. If they have extra liquid in the tube
after removing 200 l and 150 l, they pipetted more than 350 l.
Copyright © 2007 MassBioEd
14
Download