USER’S MANUAL Important user information All users must read this entire manual to fully understand the safe use of EMMA Emergency Capnometer. Declaration of conformity 0413 Complies with 93/42/EEC Medical Device Directive. MEDICAL EQUIPMENT WITH RESPECT TO ELECTRICAL SHOCK, FIRE AND MECHANICAL HAZARDS ONLY IN ACCORDANCE WITH UL 60601-1 3SJV The obligations for PHASEIN under this guarantee shall be limited to repair, or at PHASEIN's option, replacement of necessary parts or assemblies and shall not include costs of shipping. Claims for damages during shipment must be filed promptly with the transportation company. All correspondence concerning the products must specify both the name of the product and its serial number as written on the label on the product. Use of the equipment for other than its intended use, or if it has been repaired by anyone except PHASEIN or a PHASEIN authorized service center, or altered or modified or used without following the instructions in the user manual, will void this warranty. Trademarks PHASEIN AB owns the following licensed trademarks: PHASEIN IRMA and PHASEIN XTP Windows and EMMA. Safety notices This user manual contains Warning notices and Caution notices. These notices shall be followed. WARNING! Warnings indicate a potential harmful condition that can possibly lead to injury or death. CAUTION! Cautions indicate conditions which may lead to the damage or malfunction of the device Liability PHASEIN AB shall in no event be liable for any direct, indirect, special or consequential damages including without limitation damages for loss of business profits, loss of income, business interruption, loss of business information, loss of use or other related exposures, however caused, arising from the faulty or incorrect use of the product. Statement PHASEIN AB guarantees that the product delivered has been thoroughly tested to ensure that it meets its published specifications Warranty PHASEIN AB warrants the products manufactured or distributed by them to be free from faulty materials and workmanship for a period of 12 months from date of original shipment to first end user except for disposable products or products which have a stated guarantee longer or shorter than 12 months. PHASEIN will perform warranty service at its factory. Patents PHASEIN AB holds the following patents regarding products described in this manual: SE519766; SE519779; SE523461; SE524086. Other patents pending. Copyright This document contains proprietary information that is protected by copyright. All rights are reserved. No part of this document may be photocopied, reproduced or translated to another language without prior written consent of PHASEIN AB. All Rights Reserved. © 2006 PHASEIN AB Contact information For help with this device, please contact PHASEIN AB, Svärdvägen 15, S-18233 Danderyd, Sweden Telephone: +46 8 544 98 150 Fax. +46 8 544 98 169 www.phasein.se, e-mail: emmasupport@phasein.se The information in this document is subject to change without notice. Article no: 0000-3804 Edition: 04 Released: February 2007 Revision history Edition Date Description 04 February 2007 Update for UL/CSA 03 January 2007 Corrected dimension and WEE symbol. 02 November 2006 Added explanation of symbols, catalog numbers corrected 01 October 2006 Updated Intended Use 00 September 2006 First Edition 2 (33) Contents 1 INTENDED USE .................................................................................................................................. 4 2 SAFETY INFORMATION .................................................................................................................... 5 2.1 2.2 2.3 3 W ARNINGS ..................................................................................................................................... 5 CAUTIONS ...................................................................................................................................... 6 SYMBOL DESCRIPTION .................................................................................................................... 7 DEVICE DESCRIPTION ...................................................................................................................... 8 3.1 EMMA EMERGENCY CAPNOMETER OVERVIEW ................................................................................. 8 3.2 PRINCIPLE OF OPERATION ............................................................................................................... 9 3.2.1 EMMA Airway Adapter ........................................................................................................ 10 4 PREPARATION FOR USE................................................................................................................ 11 4.1 4.2 4.3 4.4 5 SETTING UP .................................................................................................................................. 11 STARTING UP ................................................................................................................................ 12 SWITCHING OFF ............................................................................................................................ 12 CONNECTING THE EMMA EMERGENCY CAPNOMETER TO A BREATHING CIRCUIT .............................. 13 USER INTERFACE ........................................................................................................................... 14 5.1 OVERVIEW .................................................................................................................................... 14 5.2 CONTROLS ................................................................................................................................... 15 5.3 MONITORING ................................................................................................................................ 15 5.3.1 RESPIRATORY RATE Display ........................................................................................... 15 5.3.2 ETCO2 Display.................................................................................................................... 15 5.3.3 CO2 BAR GRAPH ............................................................................................................... 15 5.4 INDICATORS AND ALARMS.............................................................................................................. 16 5.4.1 Default Limits....................................................................................................................... 16 5.4.2 Battery Status Indicator ....................................................................................................... 16 5.4.3 Alarm Status Indicator (EMMA Monitor only) ...................................................................... 17 5.4.4 Alarm Silence (EMMA Monitor only) ................................................................................... 17 5.4.5 Display Indications .............................................................................................................. 18 5.4.6 Adjusting the ETCO2 Alarm Limits (EMMA Monitor only)................................................... 19 5.4.7 Test of Indicators and Alarms ............................................................................................. 20 6 EMMA EMERGENCY CAPNOMETER AND ACCESSORIES......................................................... 21 7 MAINTENANCE AND SERVICE ...................................................................................................... 22 7.1 7.2 7.3 7.4 7.5 7.6 7.7 8 BATTERY REPLACEMENT ............................................................................................................... 22 CLEANING .................................................................................................................................... 22 EMMA AIRWAY ADAPTER ............................................................................................................. 22 ZERO CALIBRATION....................................................................................................................... 22 RECYCLING .................................................................................................................................. 23 GAS SPAN CHECK ........................................................................................................................ 23 SERVICE AND PRODUCT RETURN REQUIREMENTS .......................................................................... 23 TECHNICAL SPECIFICATIONS ....................................................................................................... 24 8.1 8.2 8.3 8.4 GENERAL SPECIFICATIONS ............................................................................................................ 24 ELECTROMAGNETIC COMPATIBILITY (EMC) .................................................................................... 26 COMPLIANCE ................................................................................................................................ 30 CLASSIFICATIONS ......................................................................................................................... 30 3 (33) 1 Intended use The EMMA Emergency Capnometer Monitor measures, displays and monitors carbon dioxide concentration and respiratory rate during anesthesia, recovery and respiratory care. It may be used in the operating suite, intensive care unit, patient room, clinic, emergency medicine and emergency transport settings for adult and pediatric patients. The EMMA Emergency Capnometer Analyzer measures and displays carbon dioxide concentration and respiratory rate during anesthesia, recovery and respiratory care. It may be used in the operating suite, intensive care unit, patient room, clinic, emergency medicine and emergency transport settings for adult and pediatric patients. 4 (33) 2 Safety Information 2.1 Warnings Adhere to the following warnings for safe operation of EMMA Emergency Capnometer. WARNING! Throughout this Users Manual: EMMA Emergency Capnometer refers to both the EMMA Monitor and EMMA Analyzer models EMMA Monitor refers only to the EMMA Emergency Capnometer Monitor model EMMA Analyzer refers only to the EMMA Emergency Capnometer Analyzer model WARNING! EMMA Emergency Capnometer should only be used for the purpose and in the manner described in this manual. WARNING! For clinical applications EMMA Emergency Capnometer is intended for use only by authorized and trained health care professionals. WARNING! EMMA Emergency Capnometer must not be used with flammable anesthetic agents. WARNING! Use only PHASEIN manufactured EMMA Airway Adapters. WARNING! EMMA Airway Adapters shall not be reused. Used airway adapters shall be disposed of in accordance with local regulations for medical waste. WARNING! The user must be aware of that the EMMA Airway Adapter will add 6 ml dead space to the breathing circuit. WARNING! Measurements can be affected by mobile phones and RF communications equipment. It should be assured that the EMMA Emergency Capnometer is used in the electromagnetic environment specified. WARNING! EMMA Emergency Capnometer is intended only as an adjunct in patient assessment. It must be used in conjunction with the assessment of clinical signs and symptoms. WARNING! Remove batteries if EMMA Emergency Capnometer is not likely to be used for a period of time greater than 90 days. WARNING! If EMMA Emergency Capnometer is used with a respirator or with harmful gases such as N2O, always perform a pre-use tightness check of the patient circuit. 5 (33) WARNING! To prevent “rain-out” and moisture from draining into the EMMA Airway Adapter, place the adapter in the vertical position as shown in the picture below. WARNING! Do not use EMMA Emergency Capnometer with nebulized medications as this may affect the light transmission of the EMMA Airway Adapter windows. WARNING! Check all alarm settings before use. WARNING! For clinical applications do not silence the audible alarm. WARNING! Audible alarm volume of any monitor may not be heard in some loud environments, such as when sirens are in use and the care provider is more distant from the alarm source. Alarm volume should be tested with the extremes of your noise environment to confirm ability or limitations to hear an alarm in all circumstances of the environment. 2.2 Cautions CAUTION! If the EMMA Emergency Capnometer is used in a manner other than that for which it was intended, unpredictable behavior could result. CAUTION! The EMMA Airway Adapters are non-sterile devices. Do not autoclave the devices as this will damage them. CAUTION! Never sterilize or immerse EMMA Emergency Capnometer in liquid. CAUTION! Do not operate EMMA Emergency Capnometer at ambient temperatures less than 23°F (-5°C) or greater than 122 °F (50 °C). CAUTION! Federal law restricts this device to sale by or on the order of a physician. CAUTION! The use of EMMA Emergency Capnometer is restricted to one patient at a time. 6 (33) 2.3 Symbol Description Consult User’s Manual for detailed information. Patient Isolation – Identifies patient isolation type as BF applied part. Symbol for Waste Electrical and Electronic Equipment (WEEE). IPX1 According to the degree of protection against harmful ingress of water IPX1 (drip proof equipment). Catalog number. 7 (33) 3 Device Description 3.1 EMMA Emergency Capnometer overview The EMMA Emergency Capnometer is a quantitative carbon dioxide mainstream monitor comprised of a sensor body that snaps in place on top of a disposable EMMA Airway Adapter. The EMMA Emergency Capnometer is available in two different models: EMMA Analyzer with measurement and display functionality and EMMA Monitor with measurement, display and alarm functionality. Feature Battery Status Indicator (Green) No Adapter, Check Adapter, No Breath Detected EMMA Analyzer EMMA Monitor (Indicator) (Alarm) Alarms for High and Low ETCO2 with adjustable alarm limits Alarm sounder Alarm Status (Yellow) Alarm Silence button (EMMA Monitor only) Alarm Status (EMMA Monitor only) Battery Cover Carry strap Alarm Silence Indicator (EMMA Monitor only) Battery Cover release button Bar Graph Battery Status Indicator Power On button Respiratory Rate Display Figure 1. EMMA Monitor ETCO2 Display EMMA Sensor Body EMMA Airway Adapter Note: A figure of the EMMA Analyzer is shown in Section 5.1, Figure 9a. 8 (33) 3.2 Principle of operation The measurement of CO2 in the breathing gas mixture is based on the fact that different gas components absorb infrared light at specific wavelengths. A beam of invisible infrared light is directed through the respiratory gas flow in the EMMA Airway Adapter. As the beam passes through the airway adapter, some of the light is absorbed by the gas mixture. The amount of absorbed light is measured by a miniaturized two channel spectrometer positioned to receive the infrared light beam. The spectrometer incorporates a filter wheel fitted with two different optical "color" filters. The wavelength ranges of these filters are chosen such that one filters out colors where carbon dioxide has very strong absorption and the other filters out colors where carbon dioxide has no absorption. The spectrometer also incorporates an infrared detector that converts the light beam to an electrical signal. The electrical signal is converted to a digital value that is fed to a microprocessor. The ratio of the light measured through the two filters is then used by the microprocessor to calculate the carbon dioxide concentration in the breathing gas mixture. Spectrometer with optical filter wheel and infrared detector Infrared light source Respiratory gas Infrared light beam Figure 2. Principle of operation 9 (33) 3.2.1 EMMA Airway Adapter EMMA Emergency Capnometer snaps in place on top of the EMMA Airway Adapter. The airway adapter may, for example, be inserted between the endotracheal tube and the resuscitation bag or between the resuscitation bag and the patient mask. Respiratory gas measurements are, as described in the previous section, obtained by continuously measuring the infrared light absorption through the airway adapter. The EMMA Airway Adapter is therefore fitted with optical XTP™ windows that are transparent to light in the wavelength ranges of interest. XTP™ window Figure 3. EMMA Airway Adapter with XTP™ windows 10 (33) 4 Preparation for Use 4.1 Setting up Unpack and inspect the EMMA Emergency Capnometer for external damage. 1. Press the battery cover release button into the EMMA Emergency Capnometer body until the battery cover pops off. Figure 4. Releasing the battery cover 2. Open the battery compartment and insert two (2) AAA batteries. Make sure the batteries are fitted according to the indicated polarity. After battery installation, snap battery cover back into place. Figure 5. Inserting batteries 11 (33) 4.2 Starting up Attach the EMMA Airway Adapter to the EMMA Emergency Capnometer. The method to attach the EMMA Airway Adapter to the EMMA Emergency Capnometer Sensor Body is intuitive upon visual inspection. The EMMA Airway Adapter and EMMA Emergency Capnometer Sensor Body are designed so that they may only be connected in the proper orientation. Physical barriers prevent attachment in an improper orientation. The EMMA Airway Adapter and EMMA Emergency Capnometer Sensor Body will remain attached only if placed in the proper orientation and snapped into place, using light fingerpressure force. If the EMMA Airway Adapter and EMMA Emergency Capnometer Sensor Body are not properly fitted together they will not remain attached. It is also possible to first position the Emma Airway Adapter in the breathing circuit and then fit the EMMA Emergency Capnometer to the Adapter. Check that it is properly fitted and press the power on button. A start-up sequence will be initiated to check the proper functioning of all display segments. When the EMMA Emergency Capnometer is ready the endtidal CO 2 (ETCO2) display indicates 0, the Respiratory Display indicates “- -“, the first segment of the bar graph and the green battery status indicator are lit. If the ETCO2 display shows a non-zero value, ensure that it has not been an accumulation of CO2 between the EMMA Emergency Capnometer and the EMMA Airway Adapter by removing and reattaching the EMMA Airway Adapter. If the ETCO2 still displays a non-zero value after this procedure, perform a Zero Calibration as described in chapter 7.4 prior to using the EMMA Emergency Capnometer with a patient. Figure 6. Starting up the EMMA Emergency Capnometer 4.3 Switching off EMMA Emergency Capnometer switches off automatically according to the table below. Condition The EMMA Airway Adapter is removed. A No Breath Detected condition is present. EMMA Analyzer automatically switches off after: EMMA Monitor automatically switches off after: 15 sec. 15 sec. 2 min 2 min after Alarm Silence is pressed 12 (33) 4.4 Connecting the EMMA Emergency Capnometer to a breathing circuit The EMMA Emergency Capnometer can be connected to a patient circuit in different ways. The following pictures illustrate the most common methods of connection. Figure 7. EMMA Emergency Capnometer connected between an endotracheal tube and a resuscitation bag Figure 8. EMMA Emergency Capnometer connected to a mask 13 (33) 5 User Interface 5.1 Overview Alarm Silence Button (EMMA Monitor only) Upper (▲) Limit Adjustment for Hi/Lo ETCO2 (EMMA Monitor only) Alarm Indicator (EMMA Monitor only) Alarm Silence Indicator (EMMA Monitor only) ETCO2 Display ETCO2 Display CO2 Bar Graph CO2 Bar Graph Respiratory Rate Display Respiratory Rate Display Battery Status Indicator Battery Status Indicator Power On Button Power On Button Lower (▼) Limit Adjustment for Hi/Lo ETCO2 (EMMA Monitor only) Figure 9a. EMMA Analyzer Figure 9b. EMMA Monitor 14 (33) 5.2 Controls EMMA Emergency Capnometer has one Power On push button. On the EMMA Monitor, the Power On push button may also be used for adjusting the lower alarm limit for ETCO2 and the Alarm Silence Button may be used for adjusting the upper alarm limit for ETCO2. EMMA Analyzer EMMA Monitor Control Description Power On Turn on the device. Alarm Silence Silence active alarms for 2 minutes. Upper ETCO2 Limit Adjustment Allow the user to change the upper high and low ETCO2 limit. Lower ETCO2 Limit Adjustment Allow the user to change the lower high and low ETCO2 limit. 5.3 Monitoring The EMMA Emergency Capnometer has three displays: the ETCO2 display, the Respiratory Rate display and the CO2 Bar Graph. 5.3.1 RESPIRATORY RATE Display The RESPIRATORY RATE Display displays the Respiratory Rate (RR) in the unit breaths per minute (0 - 199 bpm) using red numbers. 5.3.2 ETCO2 Display The EMMA Emergency Capnometer is available in two versions displaying ETCO2 either in mmHg or kPa, as indicated on the device labeling. The ETCO2 Display will show the ETCO2 values in kPa (0.0 9.9 kPa) or mmHg (0 - 99 mmHg) displayed as red numbers. 5.3.3 CO2 BAR GRAPH The 14-segment red BAR GRAPH shows the current CO2 value. The table below illustrates the relation between the bar graph and the concentration of CO2 in kPa or mmHg. The bar graph provides qualitative information and is intended to be used as an adjunct to the two quantitative numeric displays. The bar graph is non-linear, using smaller units of measure at the bottom and larger units of measure at the top. The bar graph is intended to provide graphical visual communication in support of the two primary numeric displays. Segment 13 Segment 0 Segment kPa mmHg 13 12 11 10 9 8 7 6 5 4 3 2 1 0 ≥ 9.0 ≥ 8.0 ≥ 7.0 ≥ 6.0 ≥ 5.0 ≥ 4.0 ≥ 3.0 ≥ 2.0 ≥ 1.5 ≥ 1.0 ≥ 0.75 ≥ 0.5 ≥ 0.25 ≥ 0.0 ≥ 90 ≥ 80 ≥ 70 ≥ 60 ≥ 50 ≥ 40 ≥ 30 ≥ 20 ≥ 10 ≥8 ≥6 ≥4 ≥2 ≥0 Figure 10. EMMA Emergency Capnometer Monitor CO2 Bar Graph 15 (33) 5.4 Indicators and Alarms The EMMA Analyzer is equipped with an indicator system according to the table below. The EMMA Monitor is equipped with an Alarm Status Indicator and an audible alarm that may be silenced for a period of 2 minutes. The EMMA Monitor Status Indicator and audible alarm operate according to the table below. Condition Priority EMMA Analyzer EMMA Monitor Low Battery Advisory Indicator Indicator Check Adapter Advisory Indicator Alarm No Adapter Advisory Indicator Alarm No Breath Caution Indicator Alarm ETCO2 Low Caution Alarm ETCO2 High Caution Alarm 5.4.1 Default Limits The default factory settings for Respiratory Rate and ETCO2 (EMMA Monitor only) are as follows: Respiratory Rate (RR) ETCO2 (EMMA Monitor) Low High 3 bpm - OFF 50 mmHg (7.0 kPa) ETCO2 values are displayed after one breath and updated every breath. RR is displayed after two breaths and updated every breath. 5.4.2 Battery Status Indicator The Battery Status Indicator is normally lit with a steady green light. Approximately 30 minutes before depletion, the Battery Status Indicator starts blinking. Battery OK (> 30 min) Battery LOW (< 30 min) (Blinking) 16 (33) 5.4.3 Alarm Status Indicator (EMMA Monitor only) The Alarm Status Indicator is normally not lit. When an alarm is present it is lit with a steady yellow light. When the alarms No Breath or ETCO2 High/Low has been present for 40 seconds the Alarm Status Indicator will start to blink. Alarm Priority Alarm Status Indicator Audible Alarm Advisory Steady Yellow 1 tone beep and then silent for 24 seconds Caution Blinking Yellow 3 tone beep and then silent for 24 seconds 5.4.4 Alarm Silence (EMMA Monitor only) If an alarm is active, pressing the Alarm Silence button will silence the alarm for a period of 2 minutes. The Alarm Silence status is indicated by the upper segment of the bar graph being lit. If a No Breath alarm is generated, the alarm can be acknowledged and turned off by pressing the alarm silence button. The No Breath alarm should only be turned off when a patient is not connected to the EMMA Emergency Capnometer. 17 (33) 5.4.5 Display Indications ETCO2 Display -(Blinking) Blinking Numbers RR Display -(Blinking) Blinking Numbers Bar Graph Bar graph shows a center-out motion Meanings Suggested action(s) Check Adapter Check that the EMMA Airway Adapter is not occluded. Check that the EMMA Airway Adapter is correctly attached. No Adapter High / Low ETCO2 (EMMA Monitor only) Check patient CO2 out of range Check patient Meanings Suggested action(s) No Breath Check patient Check Adapter Check that the EMMA Airway Adapter is not occluded. No Adapter Check that the EMMA Airway Adapter is correctly attached. RR out of range Check patient Meanings Suggested action(s) Check Adapter Check that the EMMA Airway Adapter is not occluded. Change EMMA Airway Adapter No Adapter Connect Airway Adapter to EMMA Emergency Capnometer 18 (33) 5.4.6 Adjusting the ETCO2 Alarm Limits (EMMA Monitor only) 5.4.6.1 Adjusting the High ETCO2 Alarm Limit Press and hold the Alarm Silence button until the RR display shows “Hi” and the ETCO2 display shows the current high ETCO2 alarm limit. Release the button. To adjust the alarm limit: press the Alarm Silence button (▲) to increase, or the Power On button (▼) decrease the value. It is possible to switch off the high ETCO2 alarm by adjusting the limit above 99 mmHg (9.9 kPa). EMMA Monitor will indicate this setting by showing "- -" on the ETCO2 display during the adjustment routine. If no button has been activated for a short period of time, EMMA Monitor will automatically resume normal operation. Figure 11. Adjusting the High and Low ETCO2 alarm limits 5.4.6.2 Adjusting the Low ETCO2 Alarm Limit Press and hold the Power On button until the RR display shows “Lo” and the ETCO2 display shows the current low ETCO2 alarm limit. Release the button. To adjust the alarm limit: press the Alarm Silence button (▲) to increase, or the Power On button (▼) to decrease the value. It is possible to switch off the low ETCO2 alarm by adjusting the limit down to 0. EMMA Monitor will indicate this setting by showing "- " on the ETCO2 display during the adjustment routine. If no button has been activated for a short period of time, EMMA Monitor will automatically resume normal operation. 5.4.6.3 Alarm limit adjustment ranges The adjustment ranges for the ETCO2 alarm limits are as follows: ETCO2 displayed in mmHg ETCO2 displayed in kPa Low High OFF; 1 – 89 mmHg 11 – 99 mmHg; OFF OFF; 1 – 8.9 kPa 1.1 – 9.9 kPa; OFF If the high ETCO2 limit is decreased close to the low ETCO2 limit, the low limit will be automatically adjusted in order to maintain a minimum difference of 10 mmHg (1.0 kPa) between the high and low alarm limit. Similarly, if the low ETCO2 limit is increased close to the high ETCO2 limit, the high limit will be automatically adjusted to maintain a minimum difference of 10 mmHg (1.0 kPa) between the high and low alarm limit. 19 (33) 5.4.7 Test of Indicators and Alarms To test the visible status indicator and the audible alarm (EMMA Monitor only) perform the following steps. 1. Start-up the EMMA Emergency Capnometer as described in chapter 4.2. 2. Remove the Airway Adapter This will start the Indicator on EMMA Analyzer and the Alarm on EMMA Monitor. When the test is completed, attach the Airway Adapter to the Sensor Body as described in chapter 4.2. The EMMA Emergency Capnometer is ready for use. 20 (33) 6 EMMA Emergency Capnometer and Accessories Below is a list of device models, versions and approved accessories. For an up-to date list of accessories visit www.phasein.se EMMA Emergency Capnometer and Accessories Catalog number EMMA Analyzer (kPa) 601100 EMMA Analyzer (mmHg) 601102 EMMA Monitor (kPa) 605100 EMMA Monitor (mmHg) 605102 EMMA Airway Adapter, box of 25 100620 EMMA Airway Adapter, 6 boxes of 25 100621 EMMA Airway Adapter Infant, box of 25 100660 EMMA Airway Adapter Infant, 6 boxes of 25 100661 EMMA Pouch, Box of 10 100680 EMMA Lanyard, Bag of 10 100684 21 (33) 7 Maintenance and Service 7.1 Battery Replacement The EMMA Emergency Capnometer is equipped with a green Battery Status indicator. The Battery Status indicator starts blinking when the remaining lifetime of the batteries is approximately 30 min. To replace the batteries: Open the battery compartment by pressing the release button. Pull gently out the depleted batteries. Insert two new AAA type batteries into the battery compartment. NOTE: Make sure that the batteries are fitted according to the polarity marking! When the batteries are properly in place, gently snap the battery cover back into place. 7.2 Cleaning Remove the batteries before cleaning. EMMA Emergency Capnometer can be cleaned using a cloth moistened with isopropyl alcohol. CAUTION! DO NOT immerse EMMA Emergency Capnometer in any liquid. E 7.3 EMMA Airway Adapter The EMMA Airway Adapters are intended for single patient use. They are disposable and shall not be re-used. Treat used EMMA Airway Adapters in accordance with hospital protocol for disposable items. 7.4 Zero Calibration To zero calibrate EMMA Emergency Capnometer proceed as follows: Make sure that the EMMA Airway Adapter is properly attached. Start EMMA Emergency Capnometer by pressing the ON button. EMMA Analyzer: Press and hold down the ON button until the ETCO2 display shows “10” and the RR display shows “C0”. Keep the button depressed while the ETCO2 Display starts "counting down" i.e. displaying "9" - "8" - "7" etc and until "0" is displayed. EMMA Monitor: Press and hold down simultaneously the ON and Alarm Silence button until the ETCO2 display shows “10” and the RR displays “C0”. Keep both buttons depressed while the ETCO2 Display starts "counting down" i.e. displaying "9" - "8" - "7" etc and until "0" is displayed. Once “0” is displayed on the ETCO2 display, zero calibration of the EMMA Emergency Capnometer is completed. Gas readings should be verified with a reference instrument at regular intervals. A zero reference calibration of the IR measurement should be performed whenever an offset in gas readings is discovered. 22 (33) 7.5 Recycling The EMMA Emergency Capnometer and the EMMA Airway Adapter shall be disposed in accordance to local regulations for recyclable products. Dispose the old batteries according to local regulations. 7.6 Gas Span Check Gas readings should be verified at regular intervals with a reference instrument. 7.7 Service and Product Return Requirements Please contact PHASEIN about how to return a product for service using the contact details under Important User Information on the front inside cover of this manual. Products under warranty will be exchanged free of charge except for the cost of inbound freight. 23 (33) 8 Technical Specifications 8.1 General Specifications General EMMA Emergency Capnometer Description Compact, battery powered, quantitative capnometer for mainstream CO2 monitoring of adults and pediatrics. Measurements Non-dispersive IR absorption Models EMMA Analyzer (non-alarm) EMMA Monitor (full-alarm) Versions CO2 displayed in kPa or mmHg Warm up In operation and full accuracy within 5 sec. Calibration No routine calibration required Certifications CE marked per 93/42/EEC, FDA 510(k) and UL/CSA 60601-1 Dimensions 52 x 39 x 39 mm (2.1 x 1.5 x 1.5 inches) Weight Approximately 60 g (2.1 oz) with batteries Shock absorption Withstands repeated 1 m drops Environment Operating conditions Temperature: -5 - +50°C (23 to 122°F) Humidity: 10 - 95% RH (non-condensing) Atmospheric pressure: 70 - 120 kPa (1) Storage conditions Temperature: -30 - +70°C (-22 to 158°F) Humidity: 5 - 100% RH (condensing) Atmospheric pressure: 50 - 120 kPa Displays ETCO2 LED Numeric Display Respiratory rate LED Numeric Display CO2 Bar Graph 14 segment LED bar graph Controls Power ON key Alarm Silence 2 min. alarm silence key (EMMA Monitor only) ETCO2 Up/Down key for setting alarm limits (EMMA Monitor only) CO2 Range 0 - 99 mmHg (0 - 9.9 kPa) (2) Accuracy 0-40 mmHg ± 2 mmHg, 41-99 mmHg 6% of reading, during standard conditions.(3) Rise time ≤ 60 ms Total system response time < 0.5 sec. 24 (33) General EMMA Emergency Capnometer Respiratory Rate (RR) Respiratory rate 3 - 150 breaths/min Accuracy ± 1 bpm Breath detect Adaptive threshold, minimum 1 kPa CO2 change EMMA Airway Adapters Adult 6 ml dead space Indicators and Alarms EMMA Analyzer Indicators for: No Adapter, Check Adapter, No Breath Detected, Low Battery EMMA Monitor Alarms for: No Adapter, Check Adapter, No Breath Detected, Low Battery, Low ETCO2, High ETCO2 Sound Intensity Level ≥ 80 dBA (EMMA Monitor only) Power requirements Batteries Two (2) AAA Cell Alkaline Batteries (2 x 1.5VDC) (IEC Type LR03) Battery life time 8 hours of normal use Power status LED indicator Notes: (1) EMMA Emergency Capnometer displays CO2 in partial pressure units (kPa or mmHg) and thus does not require barometric pressure compensation. (2) Gas reading showing actual partial pressure at current humidity level. (3) To include quantitative effect on gas reading from variations in environment conditions and presence of N2O, anesthetic agents, Ethanol, Isopropyl alcohol, He, Acetone and Methane, then the CO2 accuracy range should be increased to ± 4 mmHg or 10% of reading whichever is the greater. 25 (33) 8.2 Electromagnetic Compatibility (EMC) Guidance and PHASEIN’s declaration – electromagnetic emissions – for EMMA Emergency Capnometer The EMMA Emergency Capnometer is intended for use in the electromagnetic environment specified below. The customer or the user of the EMMA Emergency Capnometer should assure that it is used in such an environment. Emissions test RF emissions CISPR 11 RF emissions CISPR 11 Compliance Group 1 Class B Electromagnetic environment – guidance The EMMA Emergency Capnometer uses RF energy only for its internal function. Therefore, its RF emissions are very low and are not likely to cause any interference in nearby electronic equipment. The EMMA Emergency Capnometer is suitable for use in all establishments including domestic establishments and those directly connected to the public low-voltage power supply network that supplies buildings used for domestic purposes. Harmonic emissions IEC 61000-3-2 Not applicable Not applicable Voltage fluctuations/ flicker emissions IEC 61000-3-3 Not applicable Not applicable 26 (33) Guidance and PHASEIN’s declaration – electromagnetic immunity – for EMMA Emergency Capnometer The EMMA Emergency Capnometer is intended for use in the electromagnetic environment specified below. The customer or the user of the EMMA Emergency Capnometer should assure that it is used in such an environment. Immunity test IEC 60601 test level Compliance level Electromagnetic environment – guidance Electrostatic discharge (ESD) IEC 61000-4-2 6kV contact 8 kV air 6 kV contact 8 kV air Floors should be wood, concrete or ceramic tile. If floors are covered with synthetic material, the relative humidity should be at least 30 %. Electrical fast transient/burst IEC 61000-4-4 2 kV for power supply lines 1 kV for input/output lines Not applicable Not applicable Surge IEC 61000-4-5 1 kV differential mode Not applicable Not applicable Voltage dips, short interruptions and voltage variations on power supply input lines IEC 61000-4-11 Not applicable Not applicable Not applicable Power frequency (50/60 Hz) magnetic field IEC 61000-4-8 3 A/m 30 A/m Power frequency magnetic fields should be at levels characteristic of a typical location in a typical commercial or hospital environment. Note: UT is the a.c. mains voltage prior to application of the test level. 27 (33) Guidance and PHASEIN’s declaration – electromagnetic immunity – for EMMA Emergency Capnometer The EMMA Emergency Capnometer is intended for use in the electromagnetic environment specified below. The customer or the user of the EMMA Emergency Capnometer should assure that it is used in such an environment. Immunity test IEC 60601 test level Compliance level Electromagnetic environment – guidance Portable and mobile RF communications equipment should be used no closer to any part of the EMMA Emergency Capnometer than the recommended separation distance calculated from the equation applicable to the frequency of the transmitter. Recommended separation distance Conducted RF IEC 61000-4-6 3V 150 kHz to 80 MHz 3V d = 1.17P d = 0.18P 80 MHz to 800 MHz Radiated RF IEC 61000-4-3 3 V/m 80 MHz to 2.5 GHz 20 V/m d = 0.35P 800 MHz to 2.5 GHz where P is the maximum output power rating of the transmitter in watts (W) according to the transmitter manufacturer and d is the recommended separation distance in meters (m). Field strengths from fixed RF transmitters, as determined by an electromagnetic site survey, a should be less than the compliance level in each frequency range.b Interference may occur in the vicinity of equipment marked with the following symbol: Note 1: At 80 MHz and 800 MHz, the higher frequency range applies. Note 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from structures, objects and people. a Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land mobile radios, amateur radio, AM and FM radio broadcast and TV broadcast cannot be predicted theoretically with accuracy. To assess the electromagnetic environment due to fixed RF transmitters, an electromagnetic site survey should be considered. If the measured field strength in the location in which the EMMA Emergency Capnometer is used exceeds the applicable RF compliance level above, the EMMA Emergency Capnometer should be observed to verify normal operation. If abnormal performance is observed, additional measures may be necessary, such as reorienting or relocating the EMMA Emergency Capnometer. b Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 20 V/m. 28 (33) Recommended separation distances between portable and mobile RF communications equipment and the EMMA Emergency Capnometer The EMMA Emergency Capnometer is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled. The customer or the user of the EMMA Emergency Capnometer can help prevent electromagnetic interference by maintaining a minimum distance between portable and mobile RF communications equipment (transmitters) and the EMMA Emergency Capnometer as recommended below, according to the maximum output power of the communications equipment. Rated maximum output power of transmitter W Separation distance according to frequency of transmitter m 150 kHz to 80 MHz d = 1.17P 80 MHz to 800 MHz d = 0.18P 800 MHz to 2.5 GHz d = 0.35P 0.01 0.12 0.018 0.035 0.1 0.37 0.057 0.11 1 1.2 0.18 0.35 10 3.7 0.57 1.1 100 12 1.8 3.5 For transmitters rated at a maximum output power not listed above, the recommended separation distance d in meters (m) can be estimated using the equation applicable to the frequency of the transmitter, where P is the maximum output power rating of the transmitter in watts (W) according to the transmitter manufacturer. Note 1: At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies. Note 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from structures, objects and people. Warning: Measurements can be affected by mobile phones and RF communications equipment. It should be assured that the EMMA Emergency Capnometer is used in the electromagnetic environment specified. 29 (33) 8.3 Compliance MDD 93/42/EEC ISO 21647:2004 EN/IEC 60601-1:1990, Amendment1 (1991), Amendment2 (1995) EN/IEC 60601-1-2:2001 ISO 5356-1 EN/IEC 60601-1-8:2003 prEN 1789:2004 8.4 Classifications According to the type of protection against electric shock INTERNALLY POWERED EQUIPMENT (Battery power) According to the degree of protection against electric shock TYPE BF APPLIED PART According to the degree of protection against harmful ingress of water IPX1 (drip proof EQUIPMENT) According to the mode of operation CONTINUOUS OPERATION According to the degree of safety of application in the presence of a FLAMMABLE ANESTHETIC MIXTURE WITH AIR, OR WITH OXYGEN OR NITROUS OXIDE EQUIPMENT not suitable for use in the presence of FLAMMABLE ANESTHETIC MIXTURE WITH AIR; OR WITH OXYGEN OR NITROUS OXIDE According to the susceptibility to vibrations MEDICAL DEVICE FOR USE IN ROAD AMBULANCES (ASD 5 - 20 Hz,: 00.5 g2/Hz, ASD 20 - 200 Hz: 0.05 g2/Hz -3 dB/octave) According to sterility No part of the EMMA Emergency Capnometer is sterile. 30 (33) Notes: 31 (33) Notes: 32 (33) Alarm Indicator (EMMA Monitor only) PHASEIN AB Svärdvägen 15 182 33 Danderyd Sweden www.phasein.se All Rights Reserved. © 2006 PHASEIN AB The information in this document is subject to change without notice. 33 (33)