Shawn Forrest Analysis of Optical and Electrochemical Options for a Microfabricated Potassium Sensor Introduction This paper summarizes a review of the various available options for fabricating a potassium-selective sensor for use in measuring the extracellular potassium concentrations around a single cell. The aim is to integrate this sensor into the nanophysiometer, a device being developed to measure the coupling of metabolic and electric activity in a single cardiac myocyte in a restricted extracellular space. A wide variety of mechanisms are used for ion sensing with varying degrees of applicability to microfabricated sensors. Several other relevant sensing parameters must also be considered for appropriateness of the mechanisms to the proposed application. Problem Statement The integration of the potassium sensor into the nanophysiometer introduces several requirements of varying importance. The most important requirements seem to be: 1) Small size – the sensor will be interfacing with the extracellular fluid in a chamber that has dimensions of 25 x 100 x 30 μm3 and therefore a volume of 75000 μm3 or 75 pL. The sensor must fit around this volume and leave room for other sensors for oxygen, pH, and voltage-sensing electrodes. This will likely constrain the surface area of the sensor to about 25 μm x 25 μm at the largest. 2) Good Selectivity – the aqueous solution surrounding the cell will be made up of Tyrode’s solution. Therefore the sensor will have to be reasonably selective for potassium over the other ions present. 3) Minimal sample perturbation – because the measurements are being taken outside a live cell and the subsequent effects of the measured concentrations of potassium on the cell are being studied, it is important that the sensor affect the potassium concentration very minimally. This is an especially stringent requirement due to the extremely small sample volume involved. 1 4) Fast temporal response – this would be desirable for resolving intra-beat fluctuations. 5) Sensitivity – this is less of an issue because the extracellular concentration of potassium in-vivo is generally in the 10 mM range. Most ion sensors are sensitive down to the μM range. 6) Biocompatibility – due to the simplicity of the Tyrode’s solution that will be used as the aqueous solution, biocompatibility issues that are often confronted in the use of ISEs should not be an issue. General Options The two main categories of options for ion sensing are optical and electrochemical sensors. Both categories of sensors have been well developed over the past few decades and have several specific mechanisms of transducing analyte concentrations for electronic recording and analysis. I will discuss the following options. 1) Optical Sensing a) Optical Fibers i) Optodes (1) Near field evanescent wave sensors (a) Covalently linked dyes (b) Dye-doped polymerization at tip ii) Bulk optode spectroscopic measurements b) Fluorescent Dyes i) Free dyes in solution ii) Encapsulated dyes (PEBBLEs) 2) Electrochemical Sensing a) Potentiometric i) Neutral-Carrier-based ISEs (1) Liquid junction (2) Solid state 2 (a) Self-assembling-monolayer To begin with, I list some of the pertinent details to optical measurements: Advantages and disadvantages of Fiber-based Optical sensors: (adapted from [11]) 1) They are not subject to electrical interferences. 2) No reference electrode is needed, although a reference source is useful. 3) An immobilized reagent does not have to be in contact with optical fibers. 4) Fiber-optic sensors can be miniaturized. 5) They are highly stable with respect to calibration, especially if ratiometric intensity measurements are used. 6) They have potential for higher information content than electrical transducers. Drawbacks: 1) They will only work if appropriate reagents can be developed. 2) They are subject to ambient light interference. This can be overcome by using a dark environment or modulated radiation. 3) They have a limited dynamic range compared to electrochemical sensors. 4) Immobilized chemistries are subjected to problems with inadequate path lengths, path length instability due to matrix swelling, reagent photolability, and reagent leaching. 5) A need still exists for more selective indicators and more reproducible immobilization procedures to enable high sensitivity and long-term stability. 6) Limited stability of the biological component immobilized onto a transducer surface. Near Field Optical Sensors The original application of optical sensors to extremely small-scale (submicron) ion sensing was reported by Kopelman and coworkers primarily for intracellular recording of pH.[17] The details of probe construction were published separately.[15] This method uses the covalent immobilization of reagents by photopolymerization on the end of a fiber. The fiber is drawn to a very small tip diameter (as small as 100 nm 3 diameter). Because this method uses near field optics, the diffraction limit does not apply to limit the degree of miniaturization. In this case miniaturization is limited by the size of the molecular probe. Because the diameter of the fiber tip is smaller than the wavelength of excitation light used, photons cannot pass into the tip. Yet the propagation continues as evanescent waves or excitons traveling to the end of the tip. Only fluorescent species very close to the end of the tip are excited, which avoids interference signals by other fluorescent molecules in the surrounding solution. Excitation is by argon laser and a measurement is by a fluorescent microscope and a photomultiplier tube. Two methods exist for the fabrication of the fibers. The first is to pull the heated fiber in a manner similar to micropipet drawing technique. This was the original method used. The second involves chemical etching of the glass fibers. This covalent immobilization of the fluorescent sensor to the fiber surface has great advantages for the time constant of the device. Because the molecular probe that does the actual Figure 1 (a and b) Examples of fiber tips fabricated. (c) Illustration of laserinduced polymerization at the fiber tip. sensing is essentially in solution, there is no significant mass transport event involved in binding of the analyte. The response time using this mechanism can reach as short as 500 msec as cited in [17]. In theory this mechanism could be used for potassium ion determinations, but I searched the literature in vain for an actual report of such a sensor. Such sensors have only been used for pH measurements because the reporter involved has variable fluorescence based on protonation. The author originally stated that “there is little difficulty in expanding these techniques into the fabrication of other sensors.” [15] Upon further searching I found that the reason for the lack of development of this technique for more generalized analyte sensing is that it requires derivatizing the fluorescent dye into a polymerizable version. [16] In the case of the pH sensitive dye fluoresceinamine, its derivative N-acryloylfluoresceinamine has an essential double bond that can afford polymerization. However, in many cases chemically sensitive dyes 4 cannot be derivatized in the same way and therefore cannot be immobilized to the fiber tip. One alternative to this covalent attachment of the dye is to enmesh the dye in a solgel matrix at the tip. This requires that the dye be reasonably hydrophilic and still requires that the dye be sufficiently selective. PBFI, the standard potassium probe from Molecular Probes, is only 1.5 times more selective for potassium than for sodium. This is much less selective than such ionophores for other ions, particularly H+ and Ca2+. Recently, a newly synthesized squarine dye was used in an optical potassium sensor. [3] This dye was found to have a similar selectivity coefficient of 1.4 and binding was only reported in the concentration range, which is much too low for the expected extracellular Na+ levels. These facts make such covalently immobilized optical sensors unfeasible for extracellular K+ measurements. Bulk Optodes As far as I have been able to find, the optical sensing of potassium and sodium have only been accomplished by coupling of the reporter to an ion exchange event with potassium according to the reaction pL HC R K C KLp R H where L is the potassium selective ionophore, C is the chromoionophore selective for hydrogen ions, R is the lipophilic salt incorporated into the membrane to allow electroneutrality, and p is a stoichiometric constant. Sensors using this technique are termed bulk optodes. It is based on the same ion binding mechanism as traditional ISEs using a polymer matrix based ionophore to provide good selectivity, but an ion exchange mechanism is used where protons are sensed by a chromoionophoric pH indicator. 5 Figure 2 Mechanism of K+-selective optode sensing. An ion-exchange event activates a chromoionophore to indicate the concentration of K+. M is the target ion, L is the ionophore, C is the chromoionophore, and R is the negative-site provider. Picture taken from [7]. The method was originally demonstrated by Seitz [20] and Wolfbeis [12], and the quantitation of the measurements is well-described by Simon [13]. The ion activities ratio in the aqueous sample is found based on the response function a eff aK H 1 eff K e Ro 1 eff Co p Lo p Ro 1 eff Co where Lo, Co and Ro are the analytical concentrations of ionophore, chromoionophore and lipophilic anion, respectively and Ke is the equilibrium exchange coefficient. The variable alpha is the normalized absorbance, which is the ratio of unprotonated form of the chromoionophore to the total chromoionophore concentration. Several potassium sensors based on this mechanism have been reported. The first optical K+ sensor was introduced by Wolfbeis [19] and used valinomycin as the ionophore and rhodamine B octadecyl ester as the reporter. Toth et. al. [18] later incorporated a plasticized polyvinyl chloride (PVC) membrane and the ionophore BME 44, an azacrown. Improvements have been made to the stability of these sensors by the use of ratiometric measurements, which compare the signals at two fluorescent wavelengths. [9] A Nd:YAG laser can be used for fluorescence measurements using Chromoionophore VI as the measuring dye and sometimes a reference dye such as Nile red is used for ratiometric measurements. The response time for these sensors is generally on the order of 30 sec or more. Two disadvantages of this coupling to a pH indicator are that the sensing becomes pH sensitive and the ion exchange process distributes protons into the sample volume, 6 possibly influencing the pH of the sample significantly. This will be a disadvantage for use in the nanophysiometer, where significant pH fluctuations are expected and influencing the extracellular pH by the sensor will be undesirable. A slight difficulty with the use of the fiber-optic would be the delivery of the fiber-optic tip to the chamber. Although the tip of the fiber can have a diameter on the order of 100 nm, the volume of the fiber increases as the cube of the distance from the tip and most fiber optic diameters are on the order of at least 100 μm. Integration of the fiber into the chamber would probably involve insertion of the fiber into a pre-fabricated channel. This would be a very exacting process but I know similar fiber optics have been inserted into channels of 40 μm. Dyes and PEBBLEs Another option that should at least be mentioned is the use of free dyes. However, as explained earlier, the dyes available for K+ are probably not sufficiently selective for extracellular measurements we are performing. PEBBLEs (Probes Encapsulated by Biologically Localized Embedding) are based on the same ion-exchange phenomenon as bulk optodes, but in this case they are beads in solution and their fluorescence is excited by diffuse light of the particular wavelength. The fact that these sensing methods are not localized is likely to interfere with the other sensors around the chamber. Delivery of the dyes or beads would be very difficult to quantify and keep uniform extracellularly. Traditional Electrochemical Ion Sensors Electrochemical ion sensors for potassium are almost exclusively in the form of potentiometric liquid-phase electrodes. The configuration of the measurement made by such an electrode is shown in Fig.3. Essentially the potential across the ion-selective membrane is measured, which is proportional to the target ion concentration in the sample solution due to the selective transport of that ion across the membrane by an ionophore. The membrane consists of a plasticized polymer matrix with an ionophore, and a lipophilic salt negative site provider. No ion-exchange process is necessary when a neutral ionophore carrier (like valinomycin) is used. So the sample is generally not 7 affected in the manner it is by bulk optodes, although there is often some slight leaching of the membrane components over long periods of time. Figure 3 Example of a liquid phase ion-selective electrode measurement. Picture taken from [10]. In turning to the more traditional electrochemical sensors, the main challenge becomes obtaining the extremely small size necessary to fit inside the cell chamber. The geometry involved requires a planar electrode configuration, which has been a research focus in recent years. [14] Several examples of miniature ion sensors have been published. Cosofret’s sensor [1] was Kapton-based and for heart measurements (diam=1mm), Dumschat’s [2] sensor was photolithographically patternable acrylate (linewidth = 300μm), Uhlig’s sensor used an underside cavity (surface area = 0.03 mm2, response time = 10-15 sec), Zielinska’s sensor Figure 4 Schematic of microfabricated potassium ISE by Knoll et. al. Illustration taken from [8]. has an area of 2.25 mm2, and Knoll’s sensor [8] had a side length of 70 μm. The latter sensor was the smallest and the only one close to our target size range. It is of the cavity variety and is shown in Fig. 4. That of Uhlig was similar and produced acceptably short response times. The ion selective membrane is placed directly over the Ag thin film wire and interfaces with the sample solution above through the small hole. 8 One important challenge that must be faced for miniaturization of any ISE is that of the required reference electrode. According to Suzuki, this is generally considered one of the most difficult issues in this field. [14] There are three proposed approaches for this. The first is a simple Ag/AgCl electrode in direct contact with the sample solution. The second is a solid-state electrode without a liquid junction. The third is a miniaturized version of a conventional liquid junction electrode. These options increase in stability with complexity, but it is not clear to me yet what degree of stability will be necessary for this application. Results vary widely between publications. I will need further study on this issue, but one consequence is fairly certain. The space required for the K+ sensor will be double that required for the K+ ISE because a reference electrode is necessary. Solid-State Electrochemical Sensors The size range makes the use of solid-state, or coated wire, ISEs desirable. These electrodes do not require an electrolyte filling solution. While such electrodes have been plagued in the past by extreme potential drift problems, more recent examples show very promising improvements. While it was originally thought that these potential drifts were due to an ill-defined redox couple, it was found in [6] that they are more likely due to water interactions between the metal and membrane. Improvements in the stability of solid-state electrodes has been accomplished by the use of a redox-active selfassembling-monolayer (SAM) on the metal surface. This layer successfully excludes water infiltration and produces much more stable potentials. Self-Assembling Monolayer ISEs One approach to creating a solid-state ISE is to use a self-assembling monolayer (SAM) of a redox-active compound. Successful use of this technique in producing a K+selective ISE has been reported.[4,5] There is some synthetic chemistry involved in obtaining the chemical species that was used, but the online supplementary material referred to in [4] concerning chemical synthesis was not available so I don’t know how involved this synthesis was. The brief description in the text indicated that it would not be prohibitively complex. 9 The SAM has a strong affinity for the gold electrode and is hydrophobic, allowing it to exclude water from the metal interface. It has also been shown to be redox-active, which allows the potential change of the inner membrane surface to be communicated to the gold electrode. The membrane composition of these ISEs is essentially the same as previously used, so the sensing parameters will be similar to other commonly used ISEs. The polymer matrix, however, is polyurethane rather than PVC because of the relatively poor adhesion characteristics of PVC. This will present a trade-off because PVC has better structural strength under high plasticizer concentration, which is generally necessary for producing a liquid phase-like environment for the ionophore. Other macroscale ISEs have been produced with this polymer with good results though so the polyurethane membrane should not be a concern. Other techniques for interfacial design are the use of silane coupling, sol-gel matrix, cellulose acetate, Nafion, poly(L-lysine), electropolymerized film, and plasmapolymerized films. I have not been able to fully research all of these methods to determine their usefulness for nanofabrication of an electrode. Proposed Electrode Fabrication A reasonable proposal for an electrochemical sensor would be to use a method similar to that of Uhlig and Knoll mentioned above by creating a silicon cavity of smaller size to place beneath the position of the myocyte. This would provide structural stability for the ion-selective membrane. If the electrodes were made of gold rather than silver, a SAM could be used to stabilize the potential drift in the ISE. The reference electrode can be fabricated in the same manner, but with a membrane without ionophore, the silver wire chloridized electrochemically, and a polyHEMA solution added around the Ag/AgCl wire to stabilize the reference electrode in the face of fluctuating Cl- ion concentrations. I’m not completely certain that the chloridization and polyHEMA addition steps will be feasible at such a small scale, but they could likely be accomplished using microcontact printing. 10 Conclusions Near-field optical sensing of potassium is not likely to be selective enough for our application because significant Na+ and Ca2+ concentrations are to be expected in the extracellular media. Bulk optode sensing may interfere with ion measurements due to the H+ ion exchange mechanism that must be used. This will likely cause some pHinfluenced shift in the potassium measurements and affect the local H+ somewhat. However, there is a degree of flexibility in the size of the optode matrix tip. Smaller tip sizes will have less affect on the solution and produce smaller signal, so it may be possible to strike a balance. It is difficult to determine what the characteristics of the SAM-based electrodes will be because there are no current examples of electrodes in this size range, but recent experiments are encouraging on their resulting stability are encouraging. The major difficulty with the electrochemical method is the need for a stable reference electrode, which has been a major obstacle for some time. I have proposed a possible solution. The sensing parameters such as selectivity and time response have less bearing in this case than the factors stated above. Both types of sensors provide similar responses. The optodes have a somewhat longer response time – 30sec vs. 15 min at the fastest, but most of the electrochemical measurements are more on the order of 1-2 min. At this point my conclusion is that the optode method would be quicker to develop because it has been used in this size range before and the method can be used essentially unchanged. But the dependence of the optode response on the pH of the environment is a fundamental drawback if significant pH shifts are expected in the extracellular space. I assume that this will be the case with the cell in such a small enclosed volume. The fluorescent lifetimes of often under 1 hr due to photobleaching may be quite inconvenient as well. Because of these considerations that the optode method is fundamentally flawed for the application at hand because we cannot use the a pH buffered solution (as is usually done), it seems that the best pursuit will be the pursuit of an electrode-based solution. The electrode method considered above is my current best guess as to a useful solution. There are several steps in this process that I am uncertain on though. 11 References 1. V. V. Cosofret, M. Erdosy, T. A. Johnson, R. P. Buck, R. B. Ash, and M. R. Neuman, "Microfabricated Sensor Arrays Sensitive to Ph and K+ for Ionic Distribution Measurements in the Beating Heart," Analytical Chemistry 67, 16471653 (1995). 2. C. Dumschat, R. Fromer, H. Rautschek, H. Muller, and H. J. Timpe, "Photolithographically Patternable Nitrate-Sensitive Acrylate-Based Membrane," Analytica Chimica Acta 243, 179-182 (1991). 3. K. Ertekin, B. Yenigul, and E. U. Akkaya, "Potassium sensing by using a newly synthesized squaraine dye in sol-gel matrix," Journal of Fluorescence 12, 263-268 (2002). 4. M. Fibbioli, K. Bandyopadhyay, S. G. Liu, L. Echegoyen, O. Enger, F. Diederich, P. Buhlmann, and E. Pretsch, "Redox-active self-assembled monolayers as novel solid contacts for ion-selective electrodes," Chemical Communications 339-340 (2000). 5. M. Fibbioli, K. Bandyopadhyay, S. G. Liu, L. Echegoyen, O. Enger, F. Diederich, D. Gingery, P. Buhlmann, H. Persson, U. W. Suter, and E. Pretsch, "Redox-active self-assembled monolayers for solid-contact polymeric membrane ion-selective electrodes," Chemistry of Materials 14, 1721-1729 (2002). 6. M. Fibbioli, W. E. Morf, M. Badertscher, N. F. de Rooij, and E. Pretsch, "Potential drifts of solid-contacted ion-selective electrodes due to zero-current ion fluxes through the sensor membrane," Electroanalysis 12, 1286-1292 (2000). 7. R. D. Johnson and L. G. Bachas, "Ionophore-based ion-selective potentiometric and optical sensors," Analytical and Bioanalytical Chemistry 376, 328-341 (2003). 8. M. Knoll, K. Cammann, C. Dumschat, J. Eshold, and C. Sundermeier, "Micromachined Ion-Selective Electrodes with Polymer Matrix Membranes," Sensors and Actuators B-Chemical 21, 71-76 (1994). 9. I. Koronczi, J. Reichert, G. Heinzmann, and H. J. Ache, "Development of a submicron optochemical potassium sensor with enhanced stability due to internal reference," Sensors and Actuators B-Chemical 51, 188-195 (1998). 10. E. Lindner and R. P. Buck, "Microfabricated potentiometric electrodes and their in vivo applications," Analytical Chemistry 72, 336A-345A (2000). 11. M. Mehrvar, C. Bis, J. M. Scharer, M. Moo-Young, and J. H. Luong, "Fiber-optic biosensors - Trends and advances," Analytical Sciences 16, 677-692 (2000). 12 12. B. P. H. Schaffar and O. S. Wolfbeis, "A Calcium-Selective Optrode Based on Fluorimetric Measurement of Membrane-Potential," Analytica Chimica Acta 217, 1-9 (1989). 13. K. Seiler and W. Simon, "Theoretical Aspects of Bulk Optode Membranes," Analytica Chimica Acta 266, 73-87 (1992). 14. H. Suzuki, "Advances in the microfabrication of electrochemical sensors and systems," Electroanalysis 12, 703-715 (2000). 15. W. H. Tan, Z. Y. Shi, and R. Kopelman, "Development of Submicron Chemical Fiber Optic Sensors," Analytical Chemistry 64, 2985-2990 (1992). 16. W. H. Tan, Z. Y. Shi, and R. Kopelman, "Miniaturized Fiberoptic Chemical Sensors with Fluorescent Dye-Doped Polymers," Sensors and Actuators BChemical 28, 157-163 (1995). 17. W. H. Tan, Z. Y. Shi, S. Smith, D. Birnbaum, and R. Kopelman, "Submicrometer Intracellular Chemical Optical Fiber Sensors," Science 258, 778-781 (1992). 18. K. Toth, G. Nagy, T. T. L. Bui, J. Jeney, and S. J. Choquette, "Planar waveguide ion-selective sensors," Analytica Chimica Acta 353, 1-10 (1997). 19. O. S. Wolfbeis and B. P. H. Schaffar, "Optical Sensors - An Ion-Selective Optrode for Potassium," Analytica Chimica Acta 198, 1-12 (1987). 20. Z. Zhujun, J. L. Mullin, and W. R. Seitz, "Optical Sensor for Sodium Based on IonPair Extraction and Fluorescence," Analytica Chimica Acta 184, 251-258 (1986). 13