X-ray photoelectron spectroscopy (XPS) report - AB-BDI-BL

advertisement
TS-MME-CCS
( Surface Analysis )
X-ray photoelectron spectroscopy (XPS) report
Client: Barbara Holzer, AB
Tel.:
Job No.: 2013-XPS-TS-MME-CCS-2007-04-24
Job opened by: M.Taborelli, D.Letant-Delrieux
Date: 2007-04-26
Number of samples: 4
Sample codification: 2013
2013-a-1: Stainless steel platelet contaminated at CERN and cleaned at IHEP (20/03/07)
2013-b-1: Stainless steel platelet contaminated at CERN and cleaned at IHEP (20/03/07)
2013-c-1: Aluminium platelet contaminated at CERN and cleaned at IHEP (20/03/07)
2013-d-1: Aluminium platelet contaminated at CERN and cleaned at IHEP (20/03/07)
Nature of samples:
Plates received wrapped in Aluminium foil (16/04/07)
Sample storage: air.
Aim of analysis: Verify the cleanness.
Results
Two samples were analyzed for each metal. The results obtained from the X-rays
photoelectron spectroscopy analysis are summarized in fig. 1. We recall that the
acceptance level of the amount of carbon contamination on stainless steel and aluminium
samples is fixed at 40 at. % and 29 at. %, respectively.
The stainless steel and aluminium samples are clean. Minor traces of P are detected on
the stainless steel surface.
For both type of metal a rather large signal of Si is measured; the latter is very likely
coming from silicates. Probably rinsing was insufficient to eliminate the silicates from
the cleaning detergent bath. The quality of rinsing is also responsible for the presence of
Ca on the surface.
Page 1 sur 2
XPS-report 2013-TS-MME-SC-2007-04-24
TS-MME-CCS
( Surface Analysis )
Conclusions:
The stainless steel and aluminium samples are sufficiently clean to be used in UHV.
70
C
O
Fe
Cr
Ni
Mo
Al
P
Si
Ca
N
surface concentration [at%]
60
50
40
30
20
10
0
2013-a-1-m
2013-b-1-m
2013-c-1-m
2013-d-1-m
Figure 1: So-called relative atomic surface concentrations calculated from the
photoelectron peak intensities in the XPS spectra -
Page 2 sur 2
XPS-report 2013-TS-MME-SC-2007-04-24
Download