V.4.4 Whole-Body MRI Technical Factors: Siemens 1.5 Tesla Sequence: Patient Position: Imaging Plane: Coil(s): Contrast: Anatomic coverage: TE (msec): TR (msec): TI (msec): Flip angle: Echo train length: Number of slices: Slice thickness (mm): Spacing/gap: (mm) Field of View (FOV) mm: Matrix (phase x frequency): Scan (Acquisition) Time: True Localizer FISP Turbo STIR OOPS T1 Supine, arms down at sides Coronal, sagittal or Coronal Coronal multiplane Body coil* Body coil* Body coil* None None None Whole body (cranial vertex to feet) 1.9-3.05 60-77 2.2-2.4 4-7 4200-6800 120-150 150 80 150-180 70-75 7-33 1 3-10 11-20 slices 1120 slices 5-10 4-6 4-6 2-5 1 1 500 200-500 200-500 128 x 256 128-140 x 256 150-180 x 256 6-20 sec. 2-3 minutes 15-25 sec. * Phased array coils may be used if their use does not lengthen the time of the examination. V.4.5 Whole-Body MRI Technical Factors: GE 1.5 Tesla Sequence: Patient Position: Imaging Plane: Coil(s): Contrast: Anatomic coverage: TE (msec): TR (msec): TI (msec): Bandwidth: Flip angle: Fat Saturation: Echo train length: Number of slices: Slice thickness (mm): Spacing/gap: (mm) Field of View (FOV) mm: Matrix: Scan (Acquisition) time: FIESTA FSE-XL IR OOPS T1 (SpGRE) Supine, arms down at sides Coronal Coronal Coronal Body coil* Body coil* Body coil* None None None Whole body (cranial vertex to feet) 1.5 27 2.3 4.5 Machine 120-250 determined (approx. 49006000) 120 32kHz 80 70 No no no 16 ~20 slices 11-20 slices 11-20 slices 6 4-6 4-6 1.0 1.0 1.0 500 200-480 200-480 128 x 256 20-40 sec. 128 x 256 2-3 min. 128-196 x 256 15-25 sec. Notes: *Phased array coils may be used if their use does not lengthen the time of the examination. V.4.6 Whole-Body MRI Technical Factors: Philips (Intera 1.5T @ R8.3 software) Sequence: Patient Position: Imaging Plane: Coil(s): Contrast: Anatomic coverage: TE (msec): TR (msec): TI (msec): Bandwidth: Flip angle: Fat Saturation: Echo train length: Number of slices: Slice thickness (mm): Spacing/gap: (mm) Field of View (FOV) mm: Matrix: Acquisitions (NSA) Scan (Acquisition) time: SCOUT IR TSE OOPS T1 (FFE) Supine, arms down at sides Coronal or Coronal Coronal multiplane Body coil* Body coil* Body coil* None None None Whole body (cranial vertex to feet) 28 2.3 Machine Machine determined determined (120(approx. 2000) 250) 150 32kHz 80 no no no 12 11-20 slices 11-20 slices 4-6 4-6 1.0 1.0 500 200-480 200-480 128 X 256 1 6-15 sec. 128 x 256 1 30 sec -1 minute 128-196 x 256 1 13-26 sec. Notes: SENSE can be applied on any of the above sequences to shorten the scan time *Phased array coils may be used if their use does not lengthen the time of the examination.