Force Comparison Protocol, 50 kN to 200 kN Force testing machines calibration, Compression mode SIM Cárdenas Moctezuma, Alejandro Torres Guzmán, Jorge C. Force and Pressure, CENAM June 2013 Content 1. 2. 3. 4. 5. 6. 7. 8. INTRODUCTIÓN OBJETIVE SCOPE REQUIREMENTS ORGANIZATION CRITERIA FOR EVALUATION OF RESULTS PRESENTATION OF RESULTS DATES OF TESTS Force Comparison Protocol SIM 3 3 3 3 3 6 6 6 June 2013 Sheet 2 of 6 1. INTRODUCTION To support the NMIs CMCs comparisons in force, at SIM area was proposed by the Centro Nacional de Metrología (CENAM). A comparison in calibration of force testing machines, in compression mode, within the interval of (50.0 to 200.0) kN was organized, where CENAM is the reference laboratory. 2. OBJECTIVE This comparison has the objective to determine the compatibility of results for the calibration of force testing machines. The reference values will be the results of the reference laboratory (CENAM). The values to be determined in this comparison are the average error and the uncertainty, for each point of force calibrated. 3. SCOPE The force testing machine will be calibrated in the interval from 50 kN to 200 kN in compression mode, using direct comparison method, according to the procedure described in this protocol. 4. REQUIREMENTS The laboratory must have technical competition to do calibrations of force testing machines, using a force transducer calibrated, to cover the points of force indicated in this protocol 5. 5.1 ORGANIZATION Contacts Logistics Coordination: CENAM Name Jorge C. Torres Guzmán e-mail jtorres@cenam.mx Reference Laboratory: CENAM, Force Laboratory. Name e-mail Alejandro Cárdenas Moctezuma acardena@cenam.mx 5.2 Phone 442-2110500, ext. 3 741 Fax. 442-2110568 Phone 442-2110500, ext. 3 749 Participants Laboratory (Country) Contact person 1 INDECOPI (Peru) Leonardo de la Cruz García 2 INM (Colombia) Aristides Dajer Espeleta adajer@inm.gob.co 3 Lanamme (Costa Rica) Humberto Tioli jose.tioli@ucr.ac.cr Force Comparison Protocol SIM e-mail ldelacruz@indecopi.gob.pe June 2013 Sheet 3 of 6 5.3 Description of equipment The arrangement of force testing machine, including force transducer, used as Transference Standard (TS) is showed in table 1. Table 1. Transfer Standard data (TS). Testing Machine Manufacturer Model Serial Number Satec Systems 60 HVL 1393 5.4 Manufacturer HBM Transducer Model Serial Number C3H3 G01639 Comparison procedure To realize the comparison, the TS will be calibrated by the reference laboratory and the participants. At the end of the measurements, each laboratory will deliver the readings and results. The reference laboratory will produce a report with the results obtained. The reference laboratory will calibrate the TS before and after the participation of the other laboratories. 5.4.1 Installation To install the participating laboratory calibration standard on the TS: Connect the participating laboratory standard and turn on the TS at least 1 hour before beginning the calibration. 5.4.2 Measurement Process Initiate the measurements applying three pre-loads, up to 200 kN. To perform the first round of measurements by increasing force, with the participant standard in initial position (0°). To do measurements in zero force and then the 5 target force points (50 kN, 75 kN, 100 kN, 150 kN, and 200 kN). The second round consists in rotating the participant standard 120° with respect to the initial position, to apply one preload and measure the zero force and then the 5 target force points (50 kN, 75 kN, 100 kN, 150 kN, and 200 kN). For the third round, rotate the participant standard 240° respect to the initial position, apply one preload and measure the zero force and then the 5 target force points. The next diagram describes the measurements process: Force Comparison Protocol SIM June 2013 Sheet 4 of 6 Pre-cargas Toma de lecturas 100 % 100 % Corrida 2 120 ° Corrida 1 0 ° 0% 0% 100 % Corrida 3 240 ° 0% 5.4.3 Calculations Each participant laboratory will carry out the necessary corrections to its measurements as it find suitable. 5.4.4 Uncertainty To estimate the combined uncertainty of the TS, the minimum components to consider are: The standard uncertainty due to the calibration system. The standard uncertainty due to reproducibility. The standard uncertainty due to resolution. The standard uncertainty due to zero error. The expanded uncertainty (U) of the measurements will be reported using a coverage factor of k = 2. The laboratory should report the uncertainty according to its CMCs declared (or the ones to be declared). Force Comparison Protocol SIM June 2013 Sheet 5 of 6 6. CRITERIA FOR EVALUATION OF RESULTS The criteria to evaluate the compatibility of results for the comparison are: Normalized Error (En), as shown in the next model. En xlab xref 2 2 U lab U ref Where: xlab is the error of the TS, determined by each participant laboratory. xref is the error of the TS, determined by the reference laboratory. Ulab is the expanded uncertainty, estimated by each participant laboratory. Uref is the expanded uncertainty, estimated by the reference laboratory. And according to the normalized error model: if |En| ≤ 1 the results are satisfactory, if |En| > 1 the results are not satisfactory. 7. PRESENTATION OF RESULTS The report of the results will be edited by CENAM and distributed among the participants firstly as report B and ones an agreement have been reached among all participants the final report (report A) will be edited. The final report will be sent to SIM and to BIPM for inclusion in their data bases. 8. TEST DATES The measurements are to be performed by each participating laboratory according to the scheduled presented in the following table. Table 2. Measurements dates. Laboratory (Country) Contact person Measurements date 1 CENAM (Mexico) Alejandro Cárdenas Moctezuma 2013-06-17 2 INDECOPI (Peru) Leonardo de la Cruz García 2013-06-18 3 INM (Colombia) Aristides Dajer Espeleta 2013-06-19 4 Lanamme (Costa Rica) Humberto Tioli 2013-06-20 5 CENAM (Mexico) Alejandro Cárdenas Moctezuma 2013-06-21 Force Comparison Protocol SIM June 2013 Sheet 6 of 6