Coating Overflow Removal Tool Team Vimage (Team #3) Mario H Gonzalez George Lao Vincent Hoang November 1, 2004 EXECUTIVE SUMMARY This purpose of this document is to describe why Guidant Corporation needs a Coating Overflow Removal Tool for Drug Eluting Stents (DES). There is brief description on what technologies led us to DES’s, which are considered by many physicians a revolutionary way to treat cardiovascular disease because they are coated by a drug that suppresses the immune system to avoid the bodies rejection of a foreign object.. Since DES’s are such a recent technology their manufacturing has not been mastered. The drug coating overflows onto unintended surfaces, namely the Inner Diameter (ID) resulting in an unsafe stent. Since unsafe stents cannot be sold the over flown stents are discarded and scrapped. This inefficient practice will be remedied by a tool that can remove any coating overflow from the ID. This tool’s scope will be designed to be implemented into the point were the stent has failed a quality inspection. From that point the tool will remove any coating overflow in order to pass a re-inspection for as many cycles until it passes quality inspection. There are several engineering requirements that have been set by Guidant, the most important being that outer diameter must not be contacted in any way to avoid any damage to the drug coating where it is intended to be. A brief description of Team Vimage’s visit to Guidant’s stent plant in Temecula is discussed at the end of the document. I Table of Contents EXECUTIVE SUMMARY ........................................................................................................ I INTRODUCTION ................................................................................................................... 1 PROBLEM STATEMENT ....................................................................................................... 2 PROBLEM SCOPE ................................................................................................................ 3 ENGINEERING REQUIREMENTS.......................................................................................... 4 PRELIMINARY BUDGET PLAN ............................................................................................ 5 PRELIMINARY PROJECT PLAN ........................................................................................... 6 TEAM VIMAGE’S VISIT TO GUIDANT ................................................................................. 9 i INTRODUCTION For people who suffer from heart disease a stent can save their lives. Stents were not the first treatment method for coronary disease. It was not until 1987 that the first coronary stent was reportedly used, but physicians had been trying to unblock arteries far before then. The technology that preceded stents is called angioplasty balloons. The procedure called, balloon dilation, consisted of a small balloon that would be fed through a main artery, such as the femoral artery, with a catheter to the problem area. Once the balloon is in place, it is inflated which compresses the plague against the wall of the artery. This led treatment directly led to stents because without a way to hold the artery open restenosis, the blockage an artery after corrective surgery, would occur. Stents were designed to act as a scaffold and hold the artery open after being expanded by a balloon. Stents have evolved from bare metal stents to today’s standard called Drug Eluting Stents (DES), which are coated with a drug. Our sponsor, Guidant, is planning to release a DES in the near future, but has not been unable to perfect the manufacturing of it. Currently, the drug coating process is flawed, which results in non-usable stents. At this point, the team has a clear understanding of the problem, which is described in detail in the following section. Since there are many ways to solve this problem the team has decided on a scope for the problem, which helps in narrowing design ideas by distinguishing tasks the design is to perform. The scope of the problem is closely related to the engineering requirements, which Guidant has set, because the scope must include any tolerances or constraints that are put in the engineering requirements. Once the problem is specifically stated the 1 report focuses on the scope and engineering requirements, which will then lead into budget planning followed by task planning. PROBLEM STATEMENT The problem occurs during the coating process. The solution can overflow due to gravity that pulls the drug layer downward and the surface tension of the drug cause it to creep around and into the Inner Diameter (ID). The amount coating that overflows depends on the machine accuracy and precision. The coating within the ID is not desired because it can affect the delivery process of the stent, in which it can damage the inflating balloon and causes asymmetric expansion of the stent. Also, the overflow can leave a scratch on the balloon and when the balloon is expanded to upwards of 200 psi that fine scratch can lead to a rupture in the balloon and put the patient’s life at risk. Due to these safety concerns, the stents that do have overflow on the ID are scrapped. The stents that the coating overflow removal tool will be designed for are made from 316L Stainless steel, with s-type round struts. The stents are laser cut from a block of stainless steel with the length ranging from 8mm to 28mm and a fixed diameter of 1.63 mm. The stent is polished until required smoothness is achieved. The drug solution, Everolimus, is then coated on the outer diameter of the stent. This drug is to suppress the body’s immune system from rejecting the stent. The complete stent is then inspected by a specialist to make sure the stent satisfies their quality control standards. Sometimes the stent cannot pass the inspection because the drug coating has overflow to an undesirable surface area, which is the inner diameter. 2 This overflow creates a problem that needs to be solved in order to maximize the yield produced. It is obvious that an overflow drug coating removal tool is needed to do this. It has to remove as much of the drug in the inner diameter with the least amount of damage to the stent. Our problem is to design and engineering such a device that will solve their problem while meeting their specifications. PROBLEM SCOPE Our goal is to design a machine that automatically cleans all the coating from the ID surface. We have decided that the tool will clean the entire ID rather than specific spots ID so that way the entire inner surface will have a uniform surface. It is also easier to clean the entire surface rather and pinpointing where overflow has occurred and then remove it from the exact site considering how small stents are. The scope of our project is going to start at the stage right after the stent has been coated has failed inspection and ends with another inspection. Our sponsor specified that we cannot alter the way they coat the stents in any way. We do have the option of changing the way the stents are held during the coating process to minimize the overflow. To avoid complication, our vision is that an operator has to feed the stent into the machine, and it will automatically remove all coating from the ID. After an initial “rough” cleaning there will a final cleaning to remove any lasting residues. Then everything after is out of the scope of our project. This scope is advantageous because it is recyclable. If the after the two step cleaning treatment there is still some overflow found in the re-inspection then the stent can be put into the coating overflow removal tool. 3 ENGINEERING REQUIREMENTS The requirement posted by our sponsor is very straight forward. 1) It cannot damage the stent during the coating removal process. Since the stent will expands 2 to 3 times its diameter after they are implanted, a little scratch can lead to major failure. 2) Our final design must be able to work with stents that are 8 mm to 28 mm long. 3) Our coating removal process cannot contact the outer surface of the stent. The coating they use are gummy and sticky even after they are dry. Any contact will embed impurity on the stent surface. For medical devices, this is definitely unacceptable. 4) Most importantly, our coating removal process cannot damage the coating located at the outer surface. It just destroys the purpose of using a drug eluting stent over the bare metal stent. 5) Each operation time must be less than 2 minutes. 6) The operation should be easily operated by 1 person. Labor intensive processes will dramatically increase the cost manufacturing of each stent. There are also some additional requests for our design: 1) If we are designing a machine that does the coating removal, it should have a footprint the size of an 8 ½” x 11” paper. Since there is a limited amount of space in their lab, the size of an additional machine is a very important factor. 2) If possible, remove all left over coating solution off the inner diameter. Since the DES is still under development, there are no absolute values of tolerance 4 we can or cannot have. During the course of our design process, we will have to report the estimated possible leftover to Guidant and let them decide the acceptability of our design. These two additional requests are not required to be satisfied by our final design, but our sponsor will be very happy if we can meet these two specifications. We hope to have as simple of a design as possible to minimize any necessary maintenance. Also, simple designs are more reliable than very complicated machines and since this machine should last indefinitely with proper maintenance we feel the simpler is better for its purpose. PRELIMINARY BUDGET PLAN With the standard budget of $400 dollars it seems unlikely to be able design and build a small and precise tool, but Guidant has agreed to provide, within reason, any tools or special equipment that may be necessary to work with stents to offset any spending. We have set a rough draft of our project budget by putting a price tag of about $1100 dollars on any donations Guidant will give. By taking into consideration of our limited time and resource, we will need $1500 dollar for our budget. The $1000 dollars is to be spending entirely on the manufacturing of the design. This includes buying pre-made parts such as simple machineries, specialize cutting tools or blades, heating or freezing equipment, and lastly any special material needed. Considering the fragility of the stent and it micro-size, or material need to be tough and yet small, so a one-third of the $1000 will be spend in material wise. The one third will be distributing to the machinery cost. Lastly, the rest will be use for building the supportive structure or outline of the design. The extra $500 dollar will be emergency money in case any unexpected problem arise 5 with our intial design and parameters need to be changed. The allocation is sufficient because we are aiming to make a simplest machine that will remove the coating only. Therefore, our budget plan is set within these constraints and yet will try to meet Guidant’s standard for quality. PRELIMINARY PROJECT PLAN Our project plan is divided into three phases; the first is project development, followed by project design and manufacture and lastly is project testing and analysis. In the project development phase, which covers the first quarter of school, the goal is to specify, justify and research the problem we are given by Guidant. Task fo the first quarter include background research, visiting Guidant, define our problem very specifically and create a list of potential conceptual design. By the end of this quarter we will have a clear project definition and requirement, and finalize a conceptual design. At the beginning of winter quarter, our design and manufacture phase will begin. This includes, research and development of similar conceptual design, a detail drawing of the design and production plan for manufacturing the design. The most important goal for this quarter is to be focus on the design layout, such detail drawing, and get them done by the end of the quarter. During spring break, we plan to work on some part of the manufacturing process of our product beside what is being made by the machine shop. We will assemble the parts during the beginning part of the spring quarter. We have two major goals for this quarter, one is to test and revised our product and the last is to analyze our project. The test and revised part will include sample testing and making 6 sure it functions properly and if not fixed it. In the analyzing part of the project we will look for specific requirements that were or was not met by or sponsor and find out why. The project plan is illustrate in the Gantt chart, which show the amount of time distribute for specify task. 7 8 10 04 3/ 1 / 11 04 2/ 1 / 12 04 2/ 1 / 05 1/ Complete 1 1/ 05 0/ Remaining 1 2/ 1 3/ 05 2/ Table 1.1 Gantt Chart 1 4/ 05 1/ 1 5/ 05 1/ Finalizing Product Analysis and Feedback Test and Adjust design Assemble of Parts Munafacturing Design Review Process Production Planning Detailed design analysis Finalize Conceptual Design Budget Planning Conceptual Design Narrow Design Possibilities Design Brain Storm Planing and Developement Problem Definition Report Background Research Contact Guidant TEAM VIMAGE’S VISIT TO GUIDANT Recently on a trip to Temecula, to visit Guidant’s stent plant our host Mark Abushabky gave us a brief tour of the facilities. This gave us the opportunity to see the specific stents that we will be working with and how they are processed. Through glass we saw the machines used to coat the stents and actually inspections taking place. This is when we found out that Evirolimus is an immunosuppressant that alters DNA, which is the main reason we will not be having any contact with the real coated stents. Contrary, we will be provided with stents that are “dummy” coated for testing purposes. An interesting procedure that Guidant insisted we do was to sign a confidentiality agreement in order to protect any confidential material. 9