MedPhys_Supplementary_Revised2

advertisement
Explanation of the nature of single-peak vs. double-peak signal in edgeillumination x-ray phase contrast imaging
Supplementary material to: Low-dose phase contrast mammography with
conventional x-ray sources, by A. Olivo, S. Gkoumas, M. Endrizzi, C. K. Hagen, M.
B. Szafraniec, P. C. Diemoz, P. R. T. Munro, K. Ignatyev, B. Johnson, J. A. Horrocks,
S. J. Vinnicombe, J. L. Jones, R. D. Speller
Edge-illumination (EI) X-ray phase contrast imaging (XPCi) was initially introduced
at synchrotrons as a scanning imaging method.1 It was later adapted to conventional
x-ray sources by means of x-ray masks replicating the EI principle for every single
row or column of a 2D area detector.2
In order to discuss image formation principles, it is simpler to refer to the original
“scanning” implementation of the method; the extension to static acquisition via
masks is then straightforward.
Figure S1. The Edge-Illumination x-ray phase contrast imaging method in its
“scanning” configuration (colour online). Two sample positions during the sample
scan are represented in panels (a) and (b).
1
The image formation mechanism is discussed in Figure S1. A row of pixels entering
the plane of the drawing (“linear detector”) is illuminated on its top edge by a beam
that is vertically thin, but as long as the linear detector itself in the direction entering
the plane of the drawing. For explanatory purposes, the beam has been represented
with two different colours to distinguish between the portion missing the pixel
(yellow) and the one hitting it (light blue), although in practice this is a single beam.
The differential phase contrast signal is created by refraction from the object or from
details inside the object. These can create positive and negative peaks; however, these
arise separately from the yellow and the light blue parts of the beam, respectively.
The creation of a positive peak is shown in figure 1S (a). When the detail touches the
upper part of the beam, x-rays that would normally miss the pixel are refracted inside
it, increasing the number of counts. This situation is encountered only for the
“yellow” part of the beam: since, in the absence of the sample, these photons miss the
detector, they cannot be used to decrease the number of counts i.e. contribute to the
creation of a negative signal.
Exactly the opposite happens for the “light blue” part of the beam. These photons
normally hit the detector and are thus counted: however, refraction from the object
can send them outside the pixel, as shown in figure 1S (b). When this happens, the
number of counts detected decreases, i.e. a negative peak is formed. Similarly to what
discussed for the yellow part of the beam, these photons can only create negative
peaks and cannot contribute to positive ones: in the absence of the sample, they are all
counted (compatibly with the efficiency of the used detector).
This means that, if the yellow part of the beam is eliminated and the detector is
illuminated only by the light blue beam, a scan of the sample will result in negative
2
peaks only; conversely, if the light blue part is eliminated, the scan will only yield
positive peaks. While the latter example has been reported previously,1 the situation
discussed in the paper falls in the former category i.e. elimination of the beam portion
falling outside the detector, leading to negative peaks only. In particular, figure 4 in
the main article shows a direct comparison between the “double peak” and the
“negative peak only” case.
Figure S2. The Edge-Illumination x-ray phase contrast imaging method in its static,
2D configuration (colour online).
Figure S2 shows the translation to the static, 2D implementation of the method
obtained through x-ray masks. The plurality of beamlets created by the pre-sample
mask effectively replaces the scanning, so that it is sufficient to place the sample
downstream of the pre-sample mask and image it in a single shot. The beamlets have
been colour-coded in yellow and light blue, with the two colours having exactly the
same meaning as for figure S1. Eliminating the yellow parts of the beam therefore
leads to images in which positive peaks disappear and only negative phase peaks are
observed. In the embodiment discussed in this article, this has been obtained simply
3
by reducing the size of the apertures in the pre-sample mask, and appropriately
positioning the two masks in such a way that the portion of each beamlet hitting the
solid septa on the detector mask is negligible. More quantitative details on the image
formation principles can be found in reference 3.
References
1
A. Olivo, F. Arfelli, G. Cantatore, R. Longo, R. H. Menk, S. Pani, M. Prest, P.
Poropat, L. Rigon, G. Tromba, E. Vallazza, and E. Castelli, “An innovative digital
imaging set-up allowing a low-dose approach to phase contrast applications in the
medical field,” Med. Phys. 28,1610-1619 (2001).
2
A. Olivo and R. Speller, “A coded-aperture approach allowing x-ray phase contrast
imaging with conventional sources,” Appl. Phys. Lett. 91, 074106 (2007).
3
A. Olivo and R. Speller, “Image formation principles in coded-aperture based x-ray
phase contrast imaging,” Phys. Med. Biol. 53, 6461-6474 (2008).
4
Download