NMR Safety and Etiquettes This document covers the minimal knowledge that must be shown before the university can responsibly allow access to the NMR spectrometer: · Personal Safety Near an NMR Spectrometer · Instrument Safekeeping · Locking and Shimming · The FACES on-line reservation system, and Etiquettes When the Schedule is Busy --- Personal Safety Near an NMR Spectrometer - No metal near the magnet. NMR Magnets are always live and always at field. They are never turned off. No ferromagnetic objects (i.e. all metal, until you know the exceptions) should be brought near them. The obvious things are tools, chairs, and gas cylinders. Equally important are small items that will fly down the bore: paper clips, jewelry, wallet chains. Barrettes, in particular, can be snatched from your hair if not secure. - No medical implants No persons with any medical implant should enter the NMR room without talking to the Director first. Some pacemakers (cardiac assist devices) have wire loops that become generators. Aneurism clips can be pulled through brain tissue. - Leave the door open This is a good habit around cryogens in general, but especially an NMR. If the magnet is disturbed and loses superconductivity, enough energy is released to boil off a swimming-pool volume of helium in a few seconds, displacing air. Such a "quench" happens rarely (about once per century per magnet) and usually in response to some abuse. But if you see the magnet “exploding slowly” or sending out a plume or acting like a steam whistle: follow your instinct to walk out the door. --- Spectrometer Safety for Beginners NMRs are expensive. They are delicate. The high cost of repairs (thousands of dollars) pales next to the cost of shutting down research and teaching for weeks or months. Fortunately there is no hazard to the NMR from the keyboard, and you are encouraged to explore NMR aggressively. Anyone in the spectrometer room, even a guest, must be able to answer the question “What is Rule 1”, or their access will be suspended. -- Rule 1: If anything might have fallen down the bore of the magnet, close the spectrometer and email Roger The primary threat to the instrument is if the probe gets broken, it takes two to six months to get it fixed and we don’t have an NMR in the interim, effectively shutting down all research programs doing organic synthesis. If your hand isn’t shaking when you open the door, you do not grasp the magnitude of this. The way this has happened is a sample has broken in or above the magnet: if there is a shard of glass in the probe, the next sample inserted will break it. The good news is the probe is trivial to clean if you close the spectrometer and call for help -- it takes me two minutes and incurs no expense to remove the probe and shake the glass out. However, severe damage will occur (the detector will be broken) if a sample is inserted while a shard of glass -- or any foreign object, large or small -- is down the bore. Samples are most often broken when being removed from the magnet: Climb the ladder. Lift the sample straight up. Most broken samples happen when being removed without climbing the ladder. Close the spectrometer if a sample breaks in or above the magnet Samples have also been broken when dropped into the magnet without the lift air. It can fall hard into the detector, the most fragile part of the spectrometer. (To avoid this, bounce the sample on the lift air before letting go of it.) It is also a good habit to turn of the spinner and wait a few seconds before ejecting a sample. You “close the spectrometer” by leaving a note on the keyboard “Broken Sample, Do Not Use”. Roger’s email address is posted on the magnet (r.kautz@neu.edu). -- Locking and Shimming Although it is possible to learn to acquire spectra by rote, you will be helpless and look foolish at your first job, where the software is different. If you start with just a little knowledge about what the spectrometer is doing, you’ll get a little smarter each time you use it. If you just get more used to not knowing, the opposite happens. Every sample perturbs the magnetic field differently, so the field has to be tuned up after each sample change. Z0 (“Z-naught”) is the average field; Z1 and Z2 are gradients to make the field at the ends of the sample the same as the field at the middle. -- Locking Lock is short for “Field Frequency Lock”. It continuously adjust Z0 for you during data acquisition, as trains go by or elevators go up and down. Without locking, the lines of your spectrum move during acquisition and would be a broad smear instead of a sharp line. Locking is done in two steps. The first step is to adjust Z0. Then you turn the lock on and adjust the other parameters. Z0 adjusts the static magnetic field. (Changing Z0 across its entire range would changes your sample’s frequency by about 10 ppm). You will adjust Z0 to tune your sample’s solvent line to the spectrometer frequency - which I will call a tuning fork inside the electronics cabinet. The correct Z0 will change significantly for different solvents; typical values are given on a chart by the spectrometer. The lock display shows the difference frequency between the solvent line and the tuning fork: Z0 too high or too low Z0 correct The more humps you see, the further out of tune your sample is. Adjust Z0 to get fewer humps … until there are no humps (a step function). After you turn the lock on, adjust lockpower to the value on the chart next to the spectrometer. Although you may increase the lock power to see the humps while adjusting Z0, if you leave the lock power too high it can suppress the solvent line instead of stimulate it. This won’t hurt the spectrometer or the sample, but the lock level will jump around and you won’t be able to shim. --- Shimming "Shimming" means making the magnetic field the same at the top, the middle, and the bottom of the sample. In contrast to Z0, which adjusted the magnetic field uniformly across the sample, Z1 and Z2 change the ends relative to the middle. Z1 is a linear correction, Z2 is quadratic (literally Z2), You adjust Z1 and Z2 in whatever direction increases the lock level. NMR Peak When Poorly Shimmed Well Shimmed The lock level is the height of the solvent line. Because the area of the solvent line is constant, if you make the line narrower it gets taller. The highest lock level you can obtain may be quite different for different samples -- you might max out at 60% for one sample and 90% for another. (And it of course also depends on the lock gain.) The true test of your shims is whether the peaks in your spectrum are sharp. Critical samples (multi-day acquisitions) get a final polish shimming on the spectrum, instead of the lock. Beginners should not mess with Z3 - Z5. In fact, because the spectrometer will have whatever shims the last user set, it’s a very good idea to load a good standard shimset (type bestshim) before you lock and shim your sample. - Common Problems If the lock level is jumping around, it is usually because the lock power is too high. If you optimize Z1 then optimize Z2, you may still have a lousy shim. Switch between Z1 and Z2 several times as you are shimming. Or walk Z2 through several steps, optimizing Z1 at each step, and keep the combination that gives the highest lock level. -- Other Important Concerns Learning NMR requires significant time at the console. There are a few things which you should get in the habit of doing right away, while your spectrum is acquiring, to increase the time available to explore, and to avoid taking incorrect or invalid spectra. --- Annotate spectrum As soon as you type ga to acquire a spectrum, type gettext and annotate it. Most chemists print out NMR spectra and paste them in their notebook (or e-notebook). The spectrum must say: what the sample is, the date, your name, and tell where the data is stored. (If not, your notebook is incomplete and you shouldn't get paid.) ELK-21-35 10 mg in CDCl3 (+TMS) (second TLC spot; not HPLC-purified) filename (optional notes about your sample) Ezra Kravitz, 13 October 2012 NMR400:hanson Name, Date Spectrometer: login If the work you are doing isn’t worth annotating, it’s not worth doing. I would suggest that if you haven’t annotated your spectrum and someone else needs the spectrometer, you should forfeit your time. - Set Reference Setting the zero ppm mark to TMS is the only kosher way to set the ppm scale. Calibrating the ppm scale to solvent peaks is acceptably close for casual work, if samples are a dilute solution. But WILL BE IN ERROR for mixed solvents or concentrated (> 10 mM) solutions. - While the spectrum is acquiring you can also Set integral resets; Choose line broadening (noise suppression)