User Notes for the Milton Multiband Radiometer Introduction The Milton Multiband Radiometer (MMR) was first developed in 1978 as a low-cost field portable radiometers sensing in the four Landsat MSS bands. An early version of the instrument is described by Milton (1980), but the present instrument differs from that early design in several ways, the most important of which are: Silicon photodiode detectors for all four bands, giving improved linearity and better signal-to-noise ratio; Digital display giving better precision and less chance of transcription errors; Internal data buffers, allowing near simultaneous measurements of target and reference panel with a single sensor head (single beam mode); The ability to use two sensor heads (dual beam mode) to make truly simultaneous measurements of the target and the reference panel. There have been many other improvements to the basic design, but the overriding aim has been to produce an instrument which is simple to use, reliable and adaptable to the varied needs of teachers and researchers in remote sensing. Inevitably, some compromises have had to be made along the way, some of which may affect how you use the instrument, so please read carefully through these notes before using the radiometer. Description of the Milton Multiband Radiometer The MMR comprises a meter unit, one or two sensor heads, a connecting cable and, optionally, a data logger. It is normally left to users to provide a suitable calibrated reference panel and a means of supporting the instrument over the surfaces of interest. The MMR can be operated hand-held, but it is better to ue it from a mast or a tripod and a fixing is provided on the sensor head for this purpose. Power for the MMR comes from a set of AA size batteries fitted within the meter unit. Rechargeable NiMH batteries are recommended, although alkaline cells may be substituted in an emergency (see important safety note below1). Many MMR supplied before 2001 were fitted with rechargeable nickel-cadmium cells but these are no longer recommended due to their well-known ‘memory’ effects. The MMR circuit draws very little current, so a fresh set of batteries will typically power the unit for over 48 hours continuously. 1 If non-rechargeable (i.e. alkaline) cells are fitted, a prominent label must be stuck on the outside of the instrument, warning users not to attempt to recharge them. 1 The MMR in use The MMR was designed to measure the spectral reflectance of spatially extensive targets. The standard instrument is not calibrated in radiance units because it is intended to be used with a calibrated reference panel to produce measurements of reflectance factor. Care must be exercised when using the MMR to measure the reflectance of small or spatially variable targets because the simple optics of the instrument mean that the field-of-view is not sharply defined, nor is the point spread function (psf) calibrated. With these limitations in mind, the MMR is well-suited to making rapid measurements of the spectral reflectance of many different terrestrial surfaces. The MMR may be operated in either single beam or dual beam mode. The principle is the same in each mode: the reflectance factor is defined by the amount of energy reflected from the target in a particular waveband and in a particular direction, as a proportion of that reflected from a calibrated reference panel. The panel must be calibrated to determine its reflectance in each of the wavebands sensed by the MMR, and it may also be desirable to calibrate the angular spectral response of the panel, depending upon the aims of the experiment and the conditions of measurement. This aspect is considered further below. Dual beam mode is more precise than single beam mode as it avoids errors introduced by temporal variations in the atmosphere. However, it involves the additional expense of a second sensor head, and the two heads must be accurately inter-calibrated (see below). The MMR is unique in that it allows measurements to be made either in dual beam mode or in single beam mode with minimal time delay between target and reference measurements. The normal delay inherent in the single beam method is reduced to the minimum by using internal data buffers to store the data values from the the target whilst those from the reference panel are acquired. Operation in single beam mode 1. Connect the sensor head to the socket marked ‘S’ (see Figure 1). 2. Switch on the radiometer and select channel 1 on the rotary switch. This switch controls which of the eight data buffers is displayed on the liquid crystal display (LCD). 3. Set the toggle switch to ‘low’. This switch controls the sensitivity of the MMR. The ‘low’ position is for low light levels, the ‘high’ position should be used if the meter goes over-range on the ‘low’ scale. 4. Block off all light from entering the sensor unit by pressing it against an optically dense object. It is not sufficient to press it into the palm of your hand as some near infra-red energy could be transmitted through your hand. Press it into your body or a thick piece of black cloth. 5. With the sensor head still blocked from receiving any light, press button ‘S’ for 3 seconds and then release. Then press button ‘R’ for a similar length of time and release. 6. Turn the rotary switch to each of the eight data buffers in turn and make a note of the reading shown on the LCD. These values are the ‘dark level offsets’ for each 2 band and they are a small error signal that must be removed from the data before calculating the reflectance factor. 7. Position the sensor head vertically over the centre of the reference panel, checking that it is not shading the panel and that the field-of-view is contained within the panel. 8. Press the button marked ‘R’ and hold it down until the reading on the display stabilises. If the illumination conditions are variable, the reading may never become completely stable, in which case a judgement should be made as to when to capture the data. When button ‘R’ is released, the data from all four spectral bands are transferred to the first four data buffers. 9. Position the sensor head vertically over the target surface, checking that it is not shaded by the operator or any of the equipment. 10. Press the button marked ‘S’ and, when the display stabilises, release the button. When button’S’ is released, the data from all four spectral bands are transferred to data buffers 5 to 8. 11. When the data from both reference panel and target have been sent to the data buffers they can be displayed in turn using the rotary switch. Take care not to press either of the buttons while reading out the eight data values. The data can be transcribed manually or can be entered into a suitable data logger. 12. It is normally advisable to make more than one measurement of the reflectance factor of a target, but it is not necessary to measure the dark level offsets every time. These only change significantly if the temperature of the instrument alters, so repeat the procedure from step (7). Operation in dual-beam mode 1. First, intercalibrate the two sensor heads as described below. 2. Connect the sensor head to be used with the reference panel to the socket marked ‘R’, and that to be used over the target surface to the socket marked ‘S’. 3. Switch on the radiometer and select channel 1 on the rotary switch. This switch controls which of the eight data buffers is displayed on the liquid crystal display (LCD). 4. Set the toggle switch to ‘low’. This switch controls the sensitivity of the MMR. The ‘low’ position is for low light levels, the ‘high’ position should be used if the meter goes over-range on the ‘low’ scale. 5. Take the reference panel sensor head and block off all light from entering it by pressing it against an optically dense object. It is not sufficient to press it into the palm of your hand as some near infra-red energy could be transmitted through your hand. Press it into your body or a thick piece of black cloth. With the sensor head still blocked from receiving any light, press button ‘S’ for 3 seconds and then release. 6. Turn the rotary switch to each of the first four data buffers in turn and make a note of the reading shown on the LCD. These values are the ‘dark level offsets’ for each band and they are a small error signal that must be removed from the data before calculating the reflectance factor. 7. Repeat step (5) for the target surface sensor head. 8. Turn the rotary switch to each of data buffers 5 to 8 in turn and make a note of the reading shown on the LCD. 3 9. Position the sensor heads over the reference panel and the target. Take the same precautions against shading and scattering from areas outside the field-of-view as in the single beam mode described above. 10. Press one of the buttons (‘S’ or ‘R’, it does not matter which) when you are ready to take a reading. Hold the button down until the reading on the LCD stabilises. If the illumination conditions are variable, the reading may never become completely stable, in which case a judgement should be made as to when to capture the data. When the button is released, the data from all four spectral bands from the reference panel and all four bands from the target are transferred to the data buffers. 11. When the data have been sent to the data buffers each band can be displayed in turn using the rotary switch. Take care not to press either of the buttons while reading out the eight data values. The data can be transcribed manually or can be entered into a suitable data logger. 13. It is normally advisable to make more than one measurement of the reflectance factor of a target, but it is not necessary to measure the dark level offsets every time. These only change significantly if the temperature of the instrument alters, so repeat the procedure from step (9). If the sensor heads are plugged into the sockets suggested above then the data presented to the LCD in both modes of operation are as follows: Rotary switch position 1 2 3 4 5 6 7 8 Data buffer number Data stored 1 2 3 4 5 6 7 8 Band 1 from target Band 2 from target Band 3 from target Band 4 from target Band 1 from reference panel Band 2 from reference panel Band 3 from reference panel Band 4 from reference panel Calculating the reflectance factor Single beam mode Reflectance factor = S VS k R VR Where: S = R = VS = digital number from surface in band digital number from the reference panel in band dark level offset from buffers 1-4 (surface, bands 1-4). 4 (1) VR = k = dark level offset from buffers 5-8 (reference panel, bands 1-4). absolute reflectance of the reference panel in band Dual beam mode Reflectance factor = S VS k P R VR (2) Where: S R VS VR k P = = = = = = digital number from surface in band digital number from the reference panel in band dark level offset from buffers 1-4 (surface, bands 1-4). dark level offset from buffers 5-8 (reference panel, bands 1-4). absolute reflectance of the reference panel in band panel calibration factor in band (see below). Figures 2 and 3 show worked examples of each of the modes of operation. Intercalibration of two sensor heads Before two sensor heads can be used together in dual beam mode it is necessary to intercalibrate them. First, you should check that the two heads have the same filters and detectors fitted2, and compare their spectral response curves if these are available. Assuming that the two sensor heads have well-matched spectral responses, all that remains is to relate their respective gains. The two sensor heads respond in a linear manner to changes in illumination so a linear regression equation is sufficient to achieve the intercalibration. The calibration equation can be most easily determined from a set of data collected on a day with intermittent bright sunshine. Clear blue sky interspersed with rapidly moving fair weather cumulus clouds is ideal. The two sensor heads should be mounted together pointing at a uniform, highly reflective diffuse target (a Spectralon reference panel is ideal). At least 50 measurements are then taken from the paired sensor heads, covering as wide a range of illumination conditions as possible. Don’t forget to measure the dark level offset. These data can then be plotted on a graph and P determined from the slope of the regression equation fitted to the data in each band. Figure 4 shows some typical results from this technique. Correlation values (r) between sensor heads of 0.99 or better are to be expected. Battery charging If rechargeable batteries are fitted, the need to recharge them is shown by a message appearing on the LCD. The mains adapter should then be connected to the socket on the right-hand side of the meter unit marked ‘AUX’ and the unit left on charge for up 2 This is usually shown on a label fixed to the sensor head. 5 to 14 hours. Please note that it is necessary to have the sensor head(s) plugged in while the unit is charging. Routine maintenance Checking the gelatin filters The MMR has two sets of filters: glass filters fitted internally and gelatin filters taped to the outside of the field stop. In time, the gelatin filters can become degraded, or they may be damaged, in which case it is possible to replace them. Hot temperatures can cause the gelatin filters to become brittle and crack, and hot, humid conditions can cause them to become cloudy. The first problem is usually evident by looking down the field stop and inspecting the surface of the filter. The second problem is only evident by removing the filter. If in doubt, it is best to fit new filters. The gelatin filters are available from the Kodak Wratten range (see Table) and are available through Kodak dealers worldwide. Band number 1 2 3 4 Wratten 45 Wratten 99 Wratten 29 Wratten 87 Kodak part number 75mm x 75mm 149-5761 149-6306 149-5621 149-6256 Colour blue green red near infra-red Kodak gelatin filters are made by dissolving organic dye in liquid gelatin and coating the proper amount of the solution onto prepared glass. After the coating is dry, the gelatin film is removed from the substrate and coated with a very thin lacquer film for protection. They should be handled with care and only held by the edges. They may be cut by placing the filter in a sandwich of thin card. Spare filters should be stored flat, interleaved with clean paper, in a cool, dark, dry place. To replace the gelatin filters, first remove the field stop by undoing the four small nuts. Note the orientation of the field stop relative to the sensor unit before you remove it. The four gelatin filters will be revealed held between the field stop and the front of the sensor unit, held by a piece of light trap material (black card or similar). Take care that the new filters are placed over the correct detectors and that the light trap is re-fitted and seated correctly, and then replace the field stop. Checking the connecting lead The multicore lead connecting the sensor unit to the meter unit can be damaged by repeated flexing or by accidents. It is recommended that users carry a spare lead during critical data collection. Take care not to damage the connection pins or sockets when inserting and removing the lead. If a faulty lead is suspected, check to see if any of the pins or sockets are bent or have become pushed back into the plastic housing. A pin that has become unseated and pushed back can often be cured by opening up the meter unit and applying gentle pressure from the inside. It should ‘click’ back into place. 6 Other adjustments Changing the sensor head gain The toggle switch on the front of the instrument marked ‘LOW’ and ‘HIGH’ switches the display between two sensitivity scales. It does not alter the basic gain of the MMR. Keep the MMR on the LOW setting, except if the display reads over-range (‘1’ in left-hand digit, others blank). As supplied, the MMR is suitable for field measurements over most surfaces with a grey reference panel, but there are occasions when greater sensitivity (higher gain) is required, such as if measurements are routinely made of dark soil surfaces under cloudy skies. It is possible to vary the gain of each band independently over a wide range to optimise the instrument for a range of applications. To do this you should open the sensor unit and identify the various switches and variable resistors shown in Figure ??. The following Table shows the relative gains possible by altering the position of each of the four switches and rotating the variable resistors. Variable resistor fully clockwise Variable resistor fully anti-clockwise Switch set at ON Gain = 1 (default) Gain = x 2.5 Switch set at OFF Gain = x2 Gain = x5 Answers to Frequently Asked Questions To be completed. See: http://www.soton.ac.uk/~ejm and: http://www.soton.ac.uk/~epfs References cited Milton, E.J., 1980. A portable multiband radiometer for ground data collection in remote sensing. International Journal of Remote Sensing, 1, 153-165. Nicodemus, F.F., Richmond, J.C., Hsia, J.J., Ginsberg, I.W. and Limperis, T.L., 1977. Geometrical considerations and nomenclature for reflectance, 1977, National Bureau of Standards Monograph 160, U.S. Govt. Printing Office, Washington D.C. 20402. Glossary anisotropy index: A number that represents the degree to which the 7 bidirectional reflectance distribution function (BRDF): bidirectional reflectance factor (BRF): goniometer: goniospectrometer: hot-spot: photometer: radiometer: solar principal plane: spectral library: Spectralon: spectrometer: directional reflectance of a surface departs from that expected from a perfectly diffuse, or Lambertian surface. Generally speaking, a Lambertian surface has an anisotropy index of 1.0. A mathematical function that describes the reflectance of a surface in terms of the angles that incident and reflected rays make with the normal to the surface and with a reference direction in the horizontal plane. The function is also wavelengthdependent. It has units of inverse steradians (sr-1). The ratio of the radiant flux reflected by a sample surface to that which would be reflected into the same reflected-beam geometry by an ideal, 100% reflective, perfectly diffuse (Lambertian) standard surface, irradiated in exactly the same way as the sample (Nicodemus et al., 1977, US National Bureau of Standards Monograph 160). A device to enable a sample to be positioned precisely in relation to the geometry of illumination and viewing. The combination of a goniometer and a spectrometer. The peak of reflectance observed in the solar principal plane at which the source of illumination (generally the Sun) is directly aligned with the sensor, and therefore the point at which the sensor views its own shadow and sees no shadows from objects on the surface. An radiometer with a spectral response deigned to match that of the human eye-brain system. Photometers are calibrated in photometric units (SI unit = ???). An instrument to measure the amount of radiant flux contained within a finite solid angle. Radiometers are calibrated in radiometric units (SI unit = mWsr-1nm-1) The plane joining the source of illumination, the sensor and the target. A collection of calibrated reflectance spectra, together with their associated documentation and metadata. Spectra should be traceable to a recognised national standard to be of value. A synthetic white plastic material made from sintered poly?? that is well-suited to use as a field reflectance standard. It is highly reflective throughout the optical region, chemically inert and resists surface contamination. Spectralon is a commercial product of Labsphere Inc. An instrument to measure the energy reflected from a surface into a solid angle in many different wavelengths, such that, for practical purposes, the measurement can be considered to be a continuous 8 spectrum. Many spectrometers (but not all) are calibrated in radiometric or photometric units. Further reading Clark, R.N., 1999, Spectroscopy of rocks and minerals, and principles of spectroscopy. Chapter 1 in Remote Sensing for the Earth Sciences: Manual of Remote Sensing, 3rd ed., Vol. 3, Rencz, A.N. (ed.), John Wiley & Sons Inc., New York, 3-58. [Excellent in-depth review of the principles of reflectance spectroscopy with many example spectra from rocks and minerals]. Deering, D.W., 1989, Field measurements of bidirectional reflectance. In: Theory and Applications of Optical Remote Sensing, Asrar, G., ed. John Wiley & Sons Inc., New York, 14-65. [Comprehensive, well-illustrated review of the subject from the perspective of a NASA scientist]. Duggin, M.J. and Philipson, W.R., 1982. Field measurement of reflectance: some major considerations. Applied Optics, 21, 2833-2840. [Describes some intercalibration issues involved in using the dual-beam method and proposes a methodology to achieve accurate BRF measurements using two spectroradiometers]. Milton, E.J., 1987. Principles of field spectroscopy. International Journal of Remote Sensing, 8, 1807-1827. [Review paper containing much practical advice and a list of instruments widely used in the 1980s]. Milton, E.J., Rollin, E.M. and Emery, D.R., 1995. Advances in field spectroscopy. In: Danson, F.M. and Plummer, S.E., ed. Advances in Environmental Remote Sensing, John Wiley & Sons Ltd, Chichester, 9-32. [Reviews the growth of field spectroscopy in the UK, especially thorough the activities of the UK Natural Environment Research Council Equipment Pool for Field Spectroscopy]. Robinson, B.F. and Biehl, L.L., 1979. Calibration procedures for measurement of reflectance factor in remote sensing field research. Society of Photo-Optical Instrumentation Engineers, 196-Measurement of Optical Radiations, 16-26. [Good introduction to the issues involved in making accurate field measurements]. Salisbury, J.W. 1998. Spectral measurements field guide. Schaepman, M.E., 1998. Calibration of a field spectroradiometer, Remote Sensing Series, Volume 31, Remote Sensing Laboratories, Department of Geography, University of Zurich, 146p. [Thorough field and laboratory evaluation of the GER3700 field spectrometer] Silva, L.F., 1978. Radiation and instrumentation in remote sensing. In: Swain, P.H. and Davis, S.M., ed. Remote Sensing: The Quantitative Approach, McGraw-Hill, New York, 21-135. [Important chapter in a highly influential book. Thorough, nonmathematical discussion of the physical principles of field spectroscopy] Steiner, D. and Guterman, T., 1966. Russian data on spectral reflectance of vegetation, soil and rock types, Final Technical Report on US Army Contract DA91-591-EUC-3863/OI-652-0106, Department of Geography, University of Zurich, 232p. [Comprehensive survey of Russian literature on field spectroscopy, including the extensive field data sets collected by Krinov and co-workers in the 1940s and 1950s]. 9